(Circulation. 2006;114:1962-1967.)
© 2006 American Heart Association, Inc.
Interventional Cardiology |
From Royal Brompton Hospital (M.J.M.), London, United Kingdom; Royal Sussex County Hospital (D.H.-S.), Brighton, United Kingdom; University Hospital (J.V.D.G.), Birmingham, United Kingdom; Glenfield Hospital (C.D.), Leicester, United Kingdom; Western Infirmary (W.S.H.), Glasgow, United Kingdom; Cardiothoracic Centre (W.L.M.), Liverpool, United Kingdom; and Georg-August University (C.J.), Goettingen, Germany.
Correspondence to Dr Michael J. Mullen, Consultant Cardiologist, Royal Brompton Hospital, Sydney Street, London, SW3 6NP, England. E-mail m.mullen{at}rbht.nhs.uk
Received September 16, 2006; revision received September 30, 2006; accepted October 3, 2006.
| Abstract |
|---|
|
|
|---|
Methods and Results We conducted a prospective, open-label, multicenter clinical study in 58 patients aged 28 to 68 years who had a clinically significant ASD or PFO. Percutaneous shunt closure was undertaken with the BioSTAR septal repair implant. Successful device implantation was achieved in 57 (98%) of 58 patients. Closure at 30 days and 6 months, assessed by contrast transthoracic echocardiography, was 48 (92%) of 52 and 54 (96%) of 56, respectively. There was no evidence of a clinically significant response to the device. Transient atrial arrhythmia occurred in 5 patients after implantation. No major safety issues were observed.
Conclusions This study demonstrates the feasibility, safety, and effectiveness of BioSTAR for the closure of ASD and PFO in humans with a high rate of early and complete shunt closure. BioSTAR is a novel septal repair implant designed to provide biological closure of atrial-level defects using the patients natural healing response. Because 90% to 95% of the implant is absorbed and replaced with healthy native tissue, future access to the left atrium may be achieved.
Key Words: heart septal defects heart defects, congenital pediatrics stroke
| Introduction |
|---|
|
|
|---|
Clinical Perspective p 1967
Long-term follow-up of these permanent implants remains limited, and the potential for late unexpected complications remains long after the defect is effectively sealed and the device function has become redundant. Furthermore, it has recently been reported that a persistent low-grade inflammatory response to such devices persists for many years after implantation and might result in late and unexpected complications.18 In addition, a permanent synthetic device will obstruct access to the left atrium that might be necessary for the later treatment of acquired heart disease. The development of an absorbable device that utilizes endogenous healing to effect a rapid and complete sealing of the defect, before being absorbed and replaced by host tissue, is therefore both intuitively attractive and offers the potential for long-term benefit.
The BioSTAR septal repair implant (NMT Medical, Boston, Mass) is a novel, bioabsorbable device specifically designed for the closure of ASD and PFO. BioSTAR uses an acellular porcine intestinal collagen layer (ICL) matrix (Organogenesis, Canton, Mass), mounted on an MP35N STARFlex (NMT Medical) "double-umbrella" framework (Figure 1).3,1921 The device is coated with a heparin benzalkonium chloride complex (Celsus, Cincinnati, Ohio), which in animal experiments reduced protein and blood cell deposition and thrombus formation.22 BioSTAR has a self-centering mechanism that consists of nitinol microsprings connected between the left and right atrial umbrellas (Figure 1). The collagen matrix is rapidly incorporated into the atrial septum, which results in a low profile and early sealing of the defect.22,23 Gradual remodeling occurs over a period of
24 months, during which the collagen is absorbed and replaced by host tissue.22,23 The BioSTAR Evaluation STudy (BEST) was designed to assess for the first time in humans the feasibility, safety, and efficacy of the BioSTAR septal occluder in the closure of PFO and ASD.
|
| Methods |
|---|
|
|
|---|
Implantation Procedure
Procedures were performed with patients under general anesthesia with fluoroscopic and transesophageal echocardiography (TEE) imaging. Before the procedure, patients were preloaded with aspirin 300 mg and clopidogrel 300 mg. The morphology and excursion of the intra-atrial septum was initially assessed with TEE. An aneurysm was defined as total excursion of
15 mm measured by M-mode in the long axis. After cannulation of the right femoral vein, a soft-tipped guidewire was advanced through the defect and positioned within a left-sided pulmonary vein. All patients received intravenous heparin (100 IU/kg) during the procedure and intravenous antibiotic prophylaxis as per the centers standard operating procedures. The size and anatomy of the defect were determined by gentle inflation of a compliant sizing balloon (NMT Medical) across the defect until a waist was apparent. On the basis of this measurement, a BioSTAR implant was selected that was approximately 1.5 to 2 times the balloon-stretched diameter. Delivery and deployment of the device are similar to that previously described for the STARFlex device.3,1921 Devices available during the study were 23, 28, and 33 mm. After rehydration in heparinized saline, the BioSTAR implant was loaded into a proprietary delivery catheter and advanced to the left atrium via an 11F transeptal sheath. The distal umbrella was opened in the left atrium, and the device and sheath were retracted until the distal umbrella was opposed against the left atrial wall of the septum. The proximal umbrella was then deployed by further withdrawal of the sheath while keeping gentle tension on the device. Correct positioning of the device was confirmed by 2-dimensional and color Doppler TEE imaging in multiple planes, after which the delivery system was activated to release the implant.
Follow-Up
Before discharge, patients underwent transthoracic echocardiography (TTE) to confirm correct device position. All patients continued taking aspirin 75 mg once daily and clopidogrel 75 mg once daily, each for 90 days after the implant procedure. Contrast TTE was performed at baseline (before closure) and at 30 days and 6 months after closure. Residual shunts were assessed by contrast TTE, at rest and during Valsalva maneuver, according to a specified protocol. Shunt size was graded as negative (no bubbles), trivial (few [<10] scattered bubbles seen in the left heart), moderate (obvious shunts with >10 bubbles at any 1 time seen in the left heart), and large (complete opacification [>20 bubbles] of a section or all of the left heart). The TEE was also repeated at 30 days to assess the healing response and thrombus formation. Echocardiograms were recorded onto videotape and were subsequently reviewed by an independent core laboratory not involved with the closure procedures (Goettingen, Germany). At each visit, blood was taken for measurement of hematologic and biochemical parameters, C-reactive protein, and erythrocyte sedimentation rate.
A Data, Safety, and Adverse Events Monitoring Committee reviewed and evaluated all reported adverse events. This committee comprised an immunologist, 2 cardiologists, a medical ethicist, and a statistician who were all independent of the study and the sponsor.
The primary end point for the study was defect closure at 6 months, defined as procedural success with no shunt or trivial shunt on TTE. Patients who did not receive a BioSTAR implant or for whom follow-up was not available were excluded from the analysis of the primary end point. The number and percentage of subjects with closure at 6 months are reported. Secondary end points were device success, defined as successful delivery and deployment of the BioSTAR at the intended site and removal of the delivery system, and procedural success, defined as device success and completion of the procedure without the occurrence of major adverse events, including death, major bleeding, thromboembolism, loss of device placement or structural integrity, and the need for cardiac surgery. Secondary end-point analysis was performed on all patients for whom treatment was attempted. Safety analysis was performed on all patients treated in the study. Prespecified safety end points were death due to any cause, clinically relevant immune reaction to the device, clinically relevant serum C-reactive protein and erythrocyte sedimentation rate levels, clinically relevant device related thrombus, systemic thromboembolism, major bleeding, and need for cardiac surgery to explant the device.
The authors had full access to the data and take full responsibility for its integrity. All authors have read and agree to the manuscript as written.
| Results |
|---|
|
|
|---|
|
|
Fifty-nine BioSTAR implants were attempted in 58 patients (Table 2). All devices were successfully delivered and deployed (100%) with no procedural complications related to the device or delivery system. Overall device success was 97% (57/59), and procedural success was 98% (57/58). In 1 patient with a hemodynamically significant ASD, a 33-mm BioSTAR was initially deployed. On TEE interrogation and before detachment from the delivery system, a single arm of the distal umbrella appeared to have prolapsed into the right atrium. This device was successfully retrieved, and because no larger BioSTAR was available, the defect was closed with an alternative nonstudy device. In a second patient with a PFO and an unusually thick secundum septum, a 28-mm device was initially deployed. On TEE interrogation, this device also appeared to be inadequately positioned and was therefore retrieved and successfully replaced with a 33-mm device. The remaining 56 patients had successful deployment of a single BioSTAR device with low profile and excellent conformability to the anatomy noted on TEE (Figure 2).
|
Of the 57 patients who had closure with BioSTAR, 1 patient was lost to follow-up, and 56 patients completed the study. Closure of the defect was achieved in 48 (92%) of 52 patients at 30 days and 54 (96%) of 56 patients at 6 months. Six patients did not have contrast studies performed at 30 days because this was introduced as a modification to the original protocol. In 1 of the 2 patients with a significant residual shunt at 6 months, a continuous stream of contrast was noted in the left heart that was suggestive of an additional pulmonary shunt.
Sixty-one patients were in the safety population, including all patients in whom a BioSTAR implant was attempted and 3 in whom no PFO was crossed and no implant performed. There were no major adverse events at implant or during follow-up; no devices required explantation, and no major safety issues with BioSTAR were identified.
Five patients required treatment for transient atrial arrhythmia after implantation (1 cardioversion and 4 medical treatment). These patients tended to be older, with an average age of 50 years, and 3 had an atrial septal aneurysm, both of which are factors known to predispose to arrhythmia.10,24
The 30-day postimplant TEE was performed in 54 (96%) of 56 patients. In all patients, the device was noted to be well positioned, with a low profile on the atrial septum. In 1 patient, a mobile echogenic mass was noted on the right atrial side of the device. This was initially interpreted as part of the collagen matrix; as a precaution, however, the patient received anticoagulation therapy, and on repeat TEE, on completion of the study, the mass had fully resolved.
The results of analyses of markers are given in Table 3. There was no evidence of a systemic adverse response in any patient. One patient developed urticaria after device implant, with no increase in C-reactive protein or erythrocyte sedimentation rate levels. The symptoms resolved spontaneously.
|
| Discussion |
|---|
|
|
|---|
Atrial septal repair is the most common cardiac intervention for structural heart disease, with procedures predominantly being performed at a young age. Currently available devices are composed mainly of metallic wires with synthetic fabrics. After a variable period of time, these devices become encapsulated by fibrous tissue, which effectively seals the defect, thereafter rendering the implant redundant. In humans, histopathological assessment of explanted devices several years after implantation has shown persistence of a local inflammatory response characterized by lymphocytic tissue infiltration and multinucleated foreign-body giant cells,18 and the potential for late unexpected complications thus remains. The well-described occurrence of late erosion in some devices is of particular concern.1214 Furthermore, because of their permanent nature, these implants will obstruct transeptal access to the left atrium should it be required in the future for the treatment of acquired heart disease. Technologies now evolving for the treatment of left-sided heart disease include percutaneous heart valve repair and replacement, left atrial appendage closure, and arrhythmia intervention, with further interventions likely to emerge over the course of current patients lives.25 A septal repair implant that is absorbed and remodels into the patients own tissue is appealing because little or no foreign material remains, thereby minimizing the potential for late complications and preserving access to the left atrium.
A key advantage of the highly purified, acellular type I collagen matrix used in BioSTAR over synthetic scaffold materials is its unique ability to induce a host connective tissue response that results in site-specific tissue regeneration as opposed to scar tissue formation.2629 It is derived from porcine intestinal collagen and is only mildly immunoreactive because of its phylogenetically well-conserved primary sequence and helical structure.30 In addition, the proprietary cleaning mechanism applied to the ICL used in BioSTAR devices removes cells, cellular debris, and other noncollagenous components (DNA, RNA, lipids, and glycosaminoglycans) that may also cause an inflammatory response. The process is detergent- and enzyme-free, thus preserving the structural integrity, cell compatibility, and mechanical strength of the matrix and its ability to remodel.30,31 The material is also unique in that cross-linking with carbodiimide can specifically alter the mechanical and biological characteristics of the material. Increasing the degree of cross-linking increases biodegradation time by making the collagen less susceptible to enzymatic degradation, reduces the antigenicity, increases the tensile strength, and decreases the physical softness of the scaffold.31,32 Moreover, the ICL matrix can incorporate bioactive substances. The BioSTAR device makes use of this property in that it has a heparin-coated surface. Preclinical experiments have demonstrated more rapid and complete endothelialization of this matrix than with nonheparin-coated devices, with a low immune response during gradual absorption and replacement by host tissue.22,23
In the present study, we report the first clinical use of the BioSTAR device in patients with a clinically significant atrial-level shunt. Our results demonstrate a high rate of shunt closure at both 30 days and 6 months. Residual shunts were assessed by contrast TTE after a specified protocol of provocative maneuvers. The high early closure rate may be due in part to the impermeable nature of the collagen matrix. In addition, biological interaction between the ICL and host tissue, augmented by what we believe to be an intermolecular attraction, incorporates the device into the atrial septum at an early stage, which results in an effective and complete edge-to-edge seal. This process was evident from the exceptional low profile and conformability of the device noted on TEE immediately after implantation and at 30 days (Figure 2).
In the present study, no attempt was made to select patients with more favorable anatomy, and no patient with a PFO was excluded on the basis of defect size or presence of an aneurysm. Additionally, no modification of the septal anatomy was performed (ie, septal puncture or balloon pull-through).33,34 The high closure rate demonstrated by BioSTAR confirms its effectiveness in treating the majority of anatomic variants.
There were no major adverse events during the study, and no major safety issues related to BioSTAR were identified. The porcine ICL used in this implant is acellular and RNA-, DNA-, and pyrogen-free.2932 This and similar materials have been successfully used in tissue repair elsewhere in the body.26,35,36 We found no evidence of a systematic inflammatory response to the collagen. One patient developed urticaria after implant that resolved spontaneously. Inflammatory markers were not raised in this patient, and on specialist review and adjudication by the Data, Safety, and Adverse Events Monitoring Committee, this was considered most likely a medication reaction. No other clinically relevant responses were observed.
The Rapid Transport delivery system and the MP35N framework are identical to those used on STARFlex, which has a proven safety record.3,21 The framework holds the ICL in place with little risk of erosion, as is reported with woven-wireconfigured devices.1214 The BioSTAR implant is easily deployed and, if required, retrievable before release, as was demonstrated in 2 patients in the present study.
Atrial arrhythmia is a common complication after placement of all transeptal devices.811 Symptoms normally develop within the first 2 weeks after implantation and usually resolve within 1 to 2 months. In the present study, atrial arrhythmia that required treatment occurred in 5 patients. This rate is consistent with previous studies.8,10,11 On completion of the study, all symptoms had resolved in these patients, and only 1 was still taking antiarrhythmic medication. The patients who developed arrhythmia in the present study were on average 50 years old, and 3 had atrial septal aneurysms, which are more common in older patients and when an aneurysmal septum is present.10,24
Thromboembolic events are a recognized complication of all currently used devices.15 The BioSTAR implants used in the present study had a heparin coating. In preclinical experiments, this has been shown to significantly reduce protein and blood cell deposition and to enhance healing.22 Although 1 patient in the present study had an echogenic mass noted on the right atrial side of the device on TEE at 30 days, the precise nature of this mass was not clear. As a precaution, the patient was given anticoagulation therapy, and on completion of the study, the mass was noted to have resolved with no clinical consequences.
| Conclusions |
|---|
|
|
|---|
| Acknowledgments |
|---|
Sources of Funding
This trial was supported by a grant provided by NMT Medical (Boston, Mass).
Disclosures
Drs Mullen, Hildick-Smith, De Giovanni, Duke, and Jux have received payment for teaching at NMT Medical. Drs Mullen and Hildick-Smith have an ownership interest in NMT Medical. The other authors report no conflicts.
| References |
|---|
|
|
|---|
2. Perloff JK, Child JS, eds. Congenital Heart Disease in Adults. Philadelphia, Pa: WB Saunders; 1991.
3. Nugent AW, Britt A, Gauvreau K, Piercey GE, Lock JE, Jenkins KJ. Device closure rates of simple atrial septal defects optimized by the STARFlex device. J Am Coll Cardiol. 2006; 48: 538544.
4. Billinger K, Ostermayer S, Carminati M, DeGiovanni J, Ewert P, Hess J, Maymone-Martins F, Qureshi S, Salmon A, Schneider M, Wilson N, Sievert H. HELEX septal occluder for transcatheter closure of patent foramen ovale: multicentre experience. EuroIntervention. 2006; 1: 465471.
5. Buscheck F, Sievert H, Kleber F, Tiefenbacher C, Krumsdorf U, Windecker S, Uhlemann F, Wahr DW. Patent foramen ovale using the Premere device: the results of the CLOSEUP trial. J Interv Cardiol. 2006; 19: 328333.[CrossRef][Medline] [Order article via Infotrieve]
6. Purcell IF, Brecker SJ, Ward DE. Closure of defects of the atrial septum in adults using the Amplatzer device: 100 consecutive patients in a single center. Clin Cardiol. 2004; 27: 509513.[Medline] [Order article via Infotrieve]
7. Meier JM, Berger A, Delabays A, Girod G, Graf D, Lyon X, Roguelov C, Vogt P, Stauffer JC, Eeckhout E. Percutaneous closure of patent foramen ovale: head-to-head comparison of two different devices. EuroIntervention. 2005; 1: 2226.
8. Alaeddini J, Feghali G, Jenkins S, Ramee S, White C, Abi-Samra F. Frequency of atrial tachyarrhythmias following transcatheter closure of patent foramen ovale. J Invasive Cardiol. 2006; 18: 365368.[Medline] [Order article via Infotrieve]
9. Hill SL, Berul CI, Patel HT, Rhodes J, Supran SE, Cao QL, Hijazi ZM. Early ECG abnormalities associated with transcatheter closure of atrial septal defects using the Amplatzer septal occluder. J Interv Card Electrophysiol. 2000; 4: 469474.[CrossRef][Medline] [Order article via Infotrieve]
10. Kiblawi FM, Sommer RJ, Levchuck SG. Transcatheter closure of patent foramen ovale in older adults. Catheter Cardiovasc Interv. 2006; 68: 136142.[CrossRef][Medline] [Order article via Infotrieve]
11. Post MC, Van Deyk K, Budts W. Percutaneous closure of a patent foramen ovale: single-centre experience using different types of devices and mid-term outcome. Acta Cardiol. 2005; 60: 515519.[CrossRef][Medline] [Order article via Infotrieve]
12. Amin Z, Hijazi ZM, Bass JL, Cheatham JP, Hellenbrand WE, Kleinman CS. Erosion of Amplatzer septal occluder device after closure of secundum atrial septal defects: review of registry of complications and recommendations to minimize future risk. Catheter Cardiovasc Interv. 2004; 63: 496502.[CrossRef][Medline] [Order article via Infotrieve]
13. Divekar A, Gaamangwe T, Shaikh N, Raabe M, Ducas J. Cardiac perforation after device closure of atrial septal defects with the Amplatzer septal occluder. J Am Coll Cardiol. 2005; 45: 12131218.
14. Preventza O, Sampath-Kumar S, Wasnick J, Gold JP. Late cardiac perforation following transcatheter atrial septal defect closure. Ann Thorac Surg. 2004; 77: 14351437.
15. Sherman JM, Hagler DJ, Cetta F. Thrombosis after septal closure device placement: a review of the current literature. Catheter Cardiovasc Interv. 2004; 63: 486489.[CrossRef][Medline] [Order article via Infotrieve]
16. Lai DW, Saver JL, Araujo JA, Reidl M, Tobis J. Pericarditis associated with nickel hypersensitivity to the Amplatzer occluder device: a case report. Catheter Cardiovasc Interv. 2005; 66: 424426.[CrossRef][Medline] [Order article via Infotrieve]
17. Wertman B, Azarbal B, Riedl M, Tobis J. Adverse events associated with nickel allergy in patients undergoing percutaneous atrial septal defect or patent foramen ovale closure. J Am Coll Cardiol. 2006; 47: 12261227.
18. Sigler M, Jux C. Biocompatibility of septal defect closure devices. Heart. October 11, 2006. DOI: 10.1136/hrt.2006.098103. Available at: http://heart.bmjjournals.com. Accessed October 15, 2006.
19. Hausdorf G, Kaulitz R, Paul T, Carminati M, Lock J. Transcatheter closure of atrial septal defect with a new flexible, self-centering device (the STARFlex Occluder). Am J Cardiol. 1999; 84: 11131116.[CrossRef][Medline] [Order article via Infotrieve]
20. Hausdorf G. StarFlex ASD closure: deployment, techniques, equipment. J Interv Cardiol. 2001; 14: 6976.[Medline] [Order article via Infotrieve]
21. Carminati M, Chessa M, Butera G, Bini RM, Giusti S, Festa P, Spadoni I, Redaelli S, Hausdorf G. Transcatheter closure of atrial septal defects with the STARFlex device: early results and follow-up. J Interv Cardiol. 2001; 14: 319324.[Medline] [Order article via Infotrieve]
22. Jux C, Bertram H, Wohlsein P, Bruegmann M, Paul T. Interventional atrial septal defect closure using a totally bioresorbable occluder matrix: development and preclinical evaluation of the BioSTAR device. J Am Coll Cardiol. 2006; 48: 161169.
23. Jux C, Wohlsein P, Bruegmann M, Zutz M, Franzbach B, Bertram H. A new biological matrix for septal occlusion. J Interv Cardiol. 2003; 16: 149152.[Medline] [Order article via Infotrieve]
24. Berthet K, Lavergne T, Cohen A, Guize L, Bousser MG, Le Heuzey JY, Amarenco P. Significant association of atrial vulnerability with atrial septal abnormalities in young patients with ischemic stroke of unknown cause. Stroke. 2000; 31: 398403.
25. Solomon SB. The future of interventional cardiology lies in the left atrium. Int J Cardiovasc Intervent. 2004; 6: 101106.[CrossRef][Medline] [Order article via Infotrieve]
26. Badylak SF. Xenogeneic extracellular matrix as a scaffold for tissue reconstruction. Transpl Immunol. 2004; 12: 367377.[CrossRef][Medline] [Order article via Infotrieve]
27. Badylak SF, Tullius R, Kokini K, Shelbourne KD, Klootwyk T, Voytik SL, Kraine MR, Simmons C. The use of xenogeneic small intestinal submucosa as a biomaterial for Achilles tendon repair in a dog model. J Biomed Mater Res. 1995; 29: 977985.[CrossRef][Medline] [Order article via Infotrieve]
28. Rosen M, Roselli EE, Faber C, Ratliff NB, Ponsky JL, Smedira NG. Small intestinal submucosa intracardiac patch: an experimental study. Surg Innov. 2005; 12: 227231.
29. Huynh T, Abraham G, Murray J, Brockbank K, Hagen PO, Sullivan S. Remodeling of an acellular collagen graft into a physiologically responsive neovessel. Nat Biotechnol. 1999; 17: 10831086.[CrossRef][Medline] [Order article via Infotrieve]
30. Abraham GA, Murray J, Billiar K, Sullivan SJ. Evaluation of the porcine intestinal collagen layer as a biomaterial. J Biomed Mater Res. 2000; 51: 442452.[CrossRef][Medline] [Order article via Infotrieve]
31. Hardin-Young J, Carr R, Downing G, Condon K, Termin P. Modification of native collagen reduces antigenicity but preserves cell compatibility. Biotechnol Bioeng. 1996; 49: 675682.[CrossRef]
32. Billiar K, Murray J, Laude D, Abraham G, Bachrach N. Effects of carbodiimide crosslinking conditions on the physical properties of laminated intestinal submucosa. J Biomed Mater Res. 2001; 56: 101108.[CrossRef][Medline] [Order article via Infotrieve]
33. Ruiz CE, Alboliras ET, Pophal SG. The puncture technique: a new method for transcatheter closure of patent foramen ovale. Catheter Cardiovasc Intervent. 2001; 53: 369372.[CrossRef][Medline] [Order article via Infotrieve]
34. Chintala K, Turner D, Leaman S, Rodriquez-Cruz E, Wynne J, Greenbaum A, Forbes T. Use of balloon pull-through technique to assist in CardioSEAL device closure of patent foramen ovale. Catheter Cardiovasc Intervent. 2003; 60: 101106.[CrossRef][Medline] [Order article via Infotrieve]
35. Franklin ME Jr, Gonzalez JJ Jr, Glass JL. Use of porcine small intestinal submucosa as a prosthetic device for laparoscopic repair of hernias in contaminated fields: 2-year follow-up. Hernia. 2004; 8: 186189.[Medline] [Order article via Infotrieve]
36. Mostow EN, Haraway GD, Dalsing M, Hodde JP, King D. Effectiveness of an extracellular matrix graft (OASIS Wound Matrix) in the treatment of chronic leg ulcers: a randomized clinical trial. J Vasc Surg. 2005; 41: 837843.[CrossRef][Medline] [Order article via Infotrieve]
![]() |
A Wahl and B Meier Patent foramen ovale and ventricular septal defect closure Heart, January 1, 2009; 95(1): 70 - 82. [Abstract] [Full Text] [PDF] |
||||
![]() |
Z. M. Hijazi and S. M. Awad Pediatric Cardiac Interventions J. Am. Coll. Cardiol. Intv., December 1, 2008; 1(6): 603 - 611. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. P. Graham Jr The Year in Congenital Heart Disease J. Am. Coll. Cardiol., July 24, 2007; 50(4): 368 - 377. [Full Text] [PDF] |
||||
![]() |
S. R. Dixon, C. L. Grines, and W. W. O'Neill The Year in Interventional Cardiology J. Am. Coll. Cardiol., July 17, 2007; 50(3): 270 - 285. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2006 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |