| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
(Circulation. 2006;114:1321-1341.)
© 2006 American Heart Association, Inc.
AHA Scientific Statement |
| Abstract |
|---|
|
|
|---|
Key Words: AHA Scientific Statements angiography catheterization
| Introduction |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
The writing committee consists of internationally recognized experts in coronary physiology. Both academic and private practice sectors are represented. The material contained herein is intended not only for the coronary angiographer and the interventional cardiologist but also referring cardiologists. The reader should view the experts consensus opinions in this statement as the best attempt of the American Heart Association to guide clinical practice in those areas where evidence-based practice exists.
| 1. Fundamental Concepts and Techniques |
|---|
|
|
|---|
1.2. The Coronary PressureFlow Relationships
To appreciate the derivations of the commonly used physiology measurements, a brief review of coronary flow and pressure will be helpful. Myocardial ischemia results from an imbalance between myocardial oxygen supply and demand. Coronary blood flow provides the needed oxygen supply for any given myocardial oxygen demand and normally increases automatically from a resting level to a maximum level in response to increases in myocardial oxygen demand from exercise and neurohumoral or pharmacological hyperemic stimuli. This increase from baseline to maximal flow has been termed coronary flow reserve (CFR).3,4 Blood flow has 3 major resistance components: the epicardial vessel (R1), the small arteries and arterioles (R2), and the intramyocardial capillary system (R3). When coronary reserve is normal, these 3 resistances are assumed to be functioning normally. In patients without atherosclerosis, the large epicardial vessel resistance (R1) is trivial. Arteries with diameters >400 µm have only minimal resistance. Adjustment of coronary resistance occurs principally at the R2 resistance (vessels <400 µm in diameter) and is due to the integrated action of several control mechanisms.47 Across a normal epicardial artery, supplying normal myocardium, coronary blood flow can increase >3-fold in adults.3,8
Autoregulation automatically maintains the basal flow at a constant level in response to changing pressure and oxygen demand. Atherosclerotic narrowings produce epicardial vessel resistance and, after a critical reduction in vessel lumen area, can abolish not only coronary reserve but also autoregulation, thus reducing resting coronary blood flow. Moreover, several conditions, including left ventricular hypertrophy, myocardial ischemia, and diabetes, can affect the microcirculation (R3 resistance), blunting the maximal increase in coronary blood flow in the absence of epicardial vessel narrowing.
Coupled with reduced flow is the loss of pressure distal to a stenosis. The resistance to flow through a stenosis caused by viscous friction, flow separation, turbulence, and eddies at the site of the stenosis results in energy loss. Energy loss produces pressure loss distal to the stenosis and thus a pressure gradient across the narrowed segment (Figure 1). The pressure loss or gradient increases with increasing coronary flow along a quadratic pressure dropflow relationship of the specific coronary stenosis, as illustrated in a patient shown in Figure 2.
|
|
In vessels without a stenosis, the pressureflow curve of maximal vasodilation is linear in the physiological pressure range (Figure 3). However, when a stenosis is present, the maximal flow at any given arterial pressure is lower. In this setting, the coronary pressureflow line at maximum vasodilation is no longer straight but curvilinear because stenosis resistance is flow dependent (Figure 3, dotted line, maximally dilated with stenosis). The clinical significance of these observations is that 2 nearly identical angiographic stenoses (eg, 60% diameter narrowing) can have a dramatically different clinical impact for the patient. One stenosis would limit flow for increasing demand and would produce angina, whereas the other would be angiographically apparent but would remain clinically unimportant.
|
1.3. Coronary Flow and Flow Reserve
As noted above, as stenosis severity increases, maximal coronary flow becomes attenuated and CFR decreases. CFR is a combined measure of the capacity of the major resistance components (the epicardial coronary artery and supplied vascular bed) to achieve maximal blood flow in response to hyperemic stimulation (see Figure 3). A normal CFR implies that both the epicardial and minimally achievable microvascular bed resistances are low and normal. However, when abnormal, CFR does not indicate which component is affected, thus limiting the clinical applicability of this measurement. Although early studies in animals and patients suggested an absolute CFR of 3.5 to 5,8,9 in adult patients with chest pain syndromes and CAD risk factors undergoing cardiac catheterization with angiographically normal vessels, the normal CFR was 2.7±0.6,10 which suggests a degree of patient variability and microvascular disease. In patients with essential hypertension and normal coronary arteries or in patients with aortic stenosis and normal coronary arteries, CFR may be reduced, in part because of myocardial hypertrophy and an abnormal microvasculature.
CFR can be altered by changes in either basal or hyperemic flow, which are influenced by hemodynamics, loading conditions, and contractility. For example, tachycardia increases basal flow and decreases hyperemic flow, thus reducing CFR 10% for each 15-beat increase in heart rate.11,12
Because CFR is a summed response of both the epicardial and microvascular flow, clinicians are reluctant to use CFR as the sole indicator of lesion significance except when it is normal. To increase confidence in CFR as a measure of lesion severity, the determination of relative CFR (rCFR) has been proposed by Gould et al,3 who defined rCFR as the ratio of maximal flow in a coronary artery with stenosis (QS) to maximal flow in a normal coronary artery without a stenosis (QN). It was shown that rCFR was independent of the aortic pressure and ratepressure product and was well suited to assess the physiological significance of coronary stenoses when an adjacent nondiseased coronary artery was available. For invasive catheterization laboratory flow studies, rCFR was defined as the ratio of CFRtarget to CFRnormal reference vessel: equation
|
|
The normal range for rCFR is 0.8 to 1.0.13,14 Because of the variability of CFR and limitations in patients with multivessel CAD, rCFR is not commonly used. Likewise, rCFR relies on the assumption that the microvascular circulatory response is uniformly distributed among the myocardial beds; thus, rCFR is of no value in patients with myocardial infarction (MI) or left ventricular regional dysfunction or in patients in whom the microcirculatory responses may be heterogeneous (eg, those with myocardial fibrosis or asymmetric hypertrophy).15
In clinical terms, CFR is best used to assess the microcirculation in the absence of epicardial artery narrowings. CFR is not used to assess stenosis significance because of the influence of hemodynamics and the unknown impact of the microcirculation.
1.4. Coronary Pressure and Fractional Flow Reserve
Myocardial perfusion pressure, normally the diastolic coronary pressure, equals aortic pressure minus the left ventricular diastolic pressure or central venous pressure. Across normal coronary arteries, aortic pressure is transmitted completely, without appreciable pressure loss even to the most distal regions. As noted earlier, the distal coronary pressure in arteries with an atherosclerotic narrowing is decreased in relationship to the degree of stenosis resistance. Pijls et al16,17 related the distal coronary pressure to the ischemic potential of a stenosis by calculating a value called the fractional flow reserve (FFR). By taking the ratio of the coronary pressure measured distal to the stenosis to aortic pressure as the normal perfusion pressure (distal coronary pressure/aortic pressure) and obtaining these measurements when the microvascular resistance was minimal and assumed to be constant (that is, at maximal hyperemia), the percentage of normal coronary flow, or a fraction of normal flow (ie, FFR), can be calculated. The FFR purportedly measures the maximum achievable myocardial blood flow in the presence of a coronary artery stenosis as a percentage of the maximum blood flow in the hypothetical case of a completely normal artery.16,17 The FFR model assumes that under maximum arterial vasodilation, the resistance of the myocardium is minimal and constant across different myocardial vascular beds, and thus blood flow to the myocardium is proportional to the driving pressure (myocardial perfusion pressure).
FFR can be derived separately for the myocardium, for the epicardial coronary artery, and for the collateral supply. Calculations of myocardial, coronary, and collateral FFR from pressure measurements taken during maximal arterial vasodilation (ie, hyperemia) are as follows: equation
|
|
|
|
|
|
where Pa is mean aortic pressure; Pd, mean distal coronary pressure;
P, mean translesional pressure gradient; Pv, mean right atrial pressure; and Pw, mean coronary wedge pressure or distal coronary pressure during balloon inflation.18 Because FFRcor uses Pw, it can be calculated only during balloon coronary angioplasty. For daily clinical practice, FFR can be easily calculated by a simplified ratio of pressures and expressed as: equation
|
|
The FFR is simplified to Pd/Pa given the assumption that Pv is negligible relative to Pa. An example of FFR measurement is shown in Figure 4. An FFR value of 0.6 means that the maximum myocardial flow across the stenosis is only 60% of what it should be without the stenosis (Figure 4). An FFR of 0.9 after percutaneous coronary intervention (PCI) means that the maximum flow to the myocardium is 90% that of a completely normal vessel. The concept of FFR has been thoroughly examined in both experimental and clinical studies.19,20
|
Unlike most other physiological indices, FFR has a normal value of 1.0 for every patient and every coronary artery. Although findings in animal studies showed an effect on FFR from heart rate and arterial pressure, human studies did not show significant changes in FFR with changes in heart rate, blood pressure, or contractility.18 FFR has a high reproducibility and low intra-individual variability (Figure 5). Moreover, FFR, unlike CFR, is independent of gender and CAD risk factors such as hypertension and diabetes, and it varies less with common doses of adenosine than does CFR.21
|
A nonischemic threshold value has been prospectively confirmed19 and was compared with noninvasive stress testing.20,22 An FFR <0.75 is associated with inducible ischemia (specificity, 100%), whereas a value >0.80 indicates absence of inducible ischemia in the majority of patients (sensitivity, 90%).
In summary, for the assessment of an epicardial stenosis, the distal pressuretoaortic pressure ratio at maximal hyperemia, or FFR, is a measurement of lesion significance that, unlike CFR, has low variability and high reproducibility and is relatively unaffected by changes in hemodynamics.
1.5. Combined Pressure and Flow Measurements
Recent advancements in sensor guidewire technology allow a more complete interrogation of the coronary circulation with a single wire (by pressure and velocity or by pressure and thermodilution). With combined distal measurements, a clinician has all the relevant hemodynamic information to make an informed decision about the physiological condition of the entire coronary circulation. Currently, 2 methods are used for these calculations. The first method combines simultaneous measurements of the phasic pressure and Doppler velocity distal to a stenosis. These measurements permit the calculation of stenosis and microvascular resistances and have been applied clinically. By using distal pressure and velocity during hyperemic conditions, a hyperemic stenosis resistance (HSR) index can be calculated as HSR=
P/v, where
P represents the hyperemic pressure gradient (PaPd) and v represents the average peak velocity at hyperemia.23,24 This index purportedly provides a refined physiological measurement quantifying the impediment to maximal flow caused exclusively by the stenosis. Like FFR, HSR has a normal reference value (HSR=0) and is independent of basal hemodynamic conditions, with high reproducibility and low variability.23,24 Simultaneous measurement of distal pressure and flow velocity also allows the separate assessment of microvascular resistance25,26 and the construction of stenosis pressure gradientvelocity curves (Figure 2) that uniquely characterize the hemodynamics of any lesion and associated physiological responses.24,27 The pressure dropvelocity relations are also well suited to visualizing the effect of percutaneous intervention.24 An HSR value >0.8 mm Hg · cm1 · s is the threshold for prediction of reversible ischemia when compared with noninvasive stress testing.23 HSR had a higher diagnostic accuracy than either pressure (FFR) or flow velocity (CFR) alone, especially in cases where outcomes of FFR and CFR do not agree.28 HSR requires the use of both a distal pressure and flow velocity signal, which can be obtained with a dual-sensor (pressure and Doppler velocity) guidewire.24,26
The second method uses distal pressure and thermodilution flow, as assessed by the inverse of the arrival (transit) time of a room-temperature saline bolus to the distal coronary artery segment.2932 By measuring the mean transit time at rest and comparing it to the mean transit time at peak hyperemia, a thermodilution CFR can be calculated. Experimental animal and human studies have validated CFRthermo against CFRDoppler.3032 The ability to measure distal pressure and estimate flow by using the thermodilution technique with a single wire also allows independent assessment of the microvasculature by calculating the index of microcirculatory resistance.33 Although principally a research tool, the simultaneous measurement of FFR, CFR, and the index of microcirculatory resistance permits a unique characterization of the epicardial and microvascular resistances. The effect of changes in flow and pressure on the 3 parameters, FFR, CFR, and HSR, is illustrated in Figure 6. The significant features of available physiological measurements in the catheterization laboratory are listed in Table 1.
|
|
1.6. Techniques of Sensor-Wire Measurements
Relevant physiological data require accurate signal acquisition. Pressure and flow velocity signals are acquired with nearly identical methods. After diagnostic angiography or during angioplasty, a sensor guidewire is passed through an angioplasty Y-connector attached to a diagnostic or guiding catheter. Heparin (40 to 60 U/kg IV) and nitroglycerin (100 to 200 µg IC) (to minimize vasomotion and measurement variability) are given several minutes before the measurements.
1.6.1. Flow Velocity
The flow velocity of red blood cells moving past the ultrasound emitter/receiver on the end of a guidewire is determined from the frequency shift, defined as the difference between the transmitted and returning frequency: equation
|
|
where v represents velocity of blood flow; f0, transmitting (transducer) frequency; f1, returning frequency; c, constant for the speed of sound in blood; and
, angle of incidence. If needed, volumetric flow can be calculated as the product of vessel area (in square centimeters) and mean cross-sectional flow velocity: cm/sxcm2=cm3/s. Doppler flow velocities can be used to represent volumetric coronary flow when the vessel cross-sectional area remains constant over the measurement period. Compared with volumetric measurements, CFR by velocity alone may underestimate the volumetric CFR in vessels with intact endothelial function and flow-mediated vasodilation. For intermediately narrow lesions, sensor guidewires are considered to be nonobstructive to flow. The cross-sectional area of a 0.014-inch (0.356-mm)diameter sensor guidewire is 0.099 mm2, only 1.4% of the cross-sectional area of a 3-mmdiameter vessel and approximately 12% of the lumen area in a vessel with a diameter of 1 mm. Although it is true that a small guidewire does add cross-sectional area to a stenosis that is being interrogated, for intermediately severe lesions the addition of the small cross-sectional area is relatively negligible. However, if the cross-sectional area of the guidewire does reduce the true lumen area, this area is indeed already very small, and when FFR is abnormal the lesion is nearly always associated with abnormal stress imaging. For a scientist needing to know the precise effect of a stenosis on blood flow, a transstenotic measurement wire is not the most precise tool. However, in the clinical arena of the cardiac catheterization laboratory, the error contributed by the measurement wire in this system favors overestimating severity in the marginal cases over missing hemodynamically significant lesions.
To measure a flow velocity signal, the Doppler sensor at the tip of the wire is advanced
5 to 10 artery-diameter lengths (>2 cm) beyond the stenosis, acquiring measurements in an area of reestablished laminar flow. Resting flow velocity data are recorded. Maximal coronary hyperemia is then induced by intracoronary or intravenous pharmacological stimulus (eg, adenosine), with average peak flow velocity (APV) continuously recorded. CFR is computed as the ratio of maximal hyperemic to basal APV (Figure 7). Poor Doppler signal acquisition may occur in 10% to 15% of patients. To optimize the velocity signal, the operator must adjust the guidewire position (sample volume). Several different tip orientations interrogating the maximal velocity spectra may be tested. A full discussion of coronary Doppler flow measurement can be found elsewhere.22
|
1.6.2. Pressure Measurements
For coronary pressure measurements, 2 pressure-wire systems are available at this time. Both wires can be used as angioplasty guidewires with mechanical properties close to those of standard clinical guidewires. For both wires, a pressure sensor is located 3 cm from the tip at the junction of the radioopaque and radiolucent portion of the wire. The pressure sensor therefore can be moved across a coronary artery stenosis and back again (showing a pressure drop) without recrossing the stenosis with the wire tip. An interface is available to record and analyze the pressure signals or to transfer the data to the regular catheterization laboratory physiological monitoring system.
Before inserting the guidewire into the patient, the sensor-wire and guide catheter pressure signals are calibrated and zeroed. The patient is given heparin and nitroglycerin as noted earlier. The sensor-wire is then introduced and positioned at the tip of the guiding catheter where the guiding catheter and wire pressures are equalized. The wire is then advanced across the stenosis or to the most distal part of the coronary artery for serial lesions or diffuse disease measurement.
Next, as described for CFR, a pharmacological hyperemic stimulus is administered via an intracoronary route (through the guide catheter) or intravenously. The mean and phasic pressure signals are continuously recorded. At peak hyperemia, represented by the nadir or lowest distal pressure, the FFR is then calculated as the ratio between the mean distal coronary pressure (measured by the pressure wire) and mean aortic pressure (measured by the guiding catheter).
1.6.2.1. Pressure Pullback Technique for Serial or Diffuse Stenoses
To study the distribution of abnormalities along a diseased coronary artery, the pressure wire can be pulled back slowly during hyperemia, precisely indicating at which particular locations hemodynamically significant abnormalities are present (Figure 8). To obtain a pullback recording, the sensor is placed in the distal segment of the coronary artery, and sustained maximum hyperemia is induced either by adenosine (140 µg/kg per minute IV) or, in rare cases, by papaverine (10 to 15 mg intracoronary). The sensor is then pulled back slowly by hand under fluoroscopic guidance, with the pressure curves being recorded simultaneously. Pressure gradients and FFR at various segments over the length of the artery are calculated. Pressure loss due to diffuse atherosclerosis is gradual, whereas a focal stenosis can be differentiated by an abrupt increase in pressure proximal to the lesion. By moving the sensor back and forth, the exact location of a pressure drop representing a focal obstruction to flow can be determined.34
|
After pressures are measured, the interface coupler is disconnected and a balloon catheter, stent, or other interventional equipment can be advanced over the pressure guidewire. Six-French guiding catheters are commonly used, but diagnostic catheters as small as 4F have been be used as well, with appreciation for the limitations in the smaller-caliber catheters with regard to pressure damping.
At the end of the procedure, the wire is withdrawn, and the sensor is positioned again at the tip of guiding catheter to verify equal guiding catheter and guidewire pressures, thus ensuring that no pressure signal drift has occurred. A complete description of the application and pitfalls of coronary pressure measurements can be found elsewhere.34
1.6.3. Pharmacological Agents for Coronary Hyperemia: Adenosine and Papaverine
The achievement of maximal hyperemia is critical to the physiological measurements as required by the underlying models. The measurement of the stenosis resistance index appears to be somewhat less sensitive to this requirement.23 The pharmacological agents most often used are intracoronary or intravenous adenosine and intracoronary papaverine (Table 2).
|
Adenosine is a naturally occurring nucleoside, formed within the myocytes from dephosphorylation of adenosine 5'-triphosphate (ATP) or cyclic adenosine monophosphate or from S-adenosyl homocysteine. It is rapidly removed from the blood by both a high-affinity red blood cell uptake mechanism and a direct deamination, with a half-life in human blood of <20 seconds. Adenosine interacts with purinergic subclass A2 receptors to increase cytosolic cyclic adenosine monophosphate as the second messenger for vasorelaxation. Because of its unique position in ATP metabolism, adenosine is thought to be an important endogenous regulator of coronary blood flow during both stress and ischemia.
The vasodilatory effects of adenosine are primarily on the microcirculation, with little effect on the epicardial conduit arteries. Both intracoronary and intravenous adenosine infusions produce maximal hyperemia in humans. To ensure maximal hyperemia, serial incremental intracoronary adenosine doses should be given until a plateau of response is achieved.21,35,36 However, in a small percentage of cases, coronary hyperemia has been suspected to be suboptimal with intracoronary adenosine, which suggests that a repeated higher intracoronary adenosine dose may be helpful. Adenosine is generally safe in the recommended intracoronary or intravenous dosages.
During intravenous adenosine administration, some patients may experience chest discomfort or dyspnea-like complaints that often resemble angina pectoris because of stimulation of free nerve fibers in the heart by the adenosine (which is also the intrinsic transmitter of angina pectoris), but these symptoms are not reflective of true ischemia. Such complaints should not be a reason to stop the infusion and should not be confused with bronchospasm, which has been infrequently reported in some patients with severe obstructive pulmonary disease. Asthma is a contraindication for intravenous adenosine administration.
During an intravenous adenosine infusion, the patient should avoid Valsalva maneuvers because straining interrupts continuous drug infusion and may produce fluctuations in both coronary blood flow and distal pressure. During exaggerated respiratory activity, FFR should always be taken as the lowest value of Pd/Pa during the respiratory cycle. When possible, intravenous adenosine should be infused through a large, preferably central, vein.
ATP, a precursor of adenosine, has the same effect in humans and, although less expensive than adenosine,37,38 is not available in the United States.
Intracoronary papaverine increases coronary blood flow velocity 4 to 6 times over resting values in patients with normal coronary arteries. Papaverine (10 to 15 mg) produces a response equal to that of an infusion of adenosine (140 µg/kg per minute IV) or dipyridamole (0.56 to 0.84 mg/kg). The total duration of the hyperemic response to intracoronary papaverine is 60 to 90 seconds, or about 4 times longer than the response to intracoronary adenosine. Because papaverine can occasionally cause QT prolongation and ventricular tachycardia or fibrillation and has been reported to generate lactate and interact with ginko, adenosine is the preferred agent for induction of coronary hyperemia.
1.6.4. Safety of Intracoronary Sensor-Wire Measurements
Qian et al39 examined the safety of intracoronary Doppler wire measurements in 906 patients. Complications with the procedure were rare: transient bradycardia (1.7%), coronary spasm (2%), and ventricular fibrillation (0.2%). The clinical practice of using sensor-wire measurements with pharmacologically induced hyperemia has been applied in thousands of patients in the past decade and is generally considered safe.
1.6.5. Limitations of Physiological Measurements
Several potential pitfalls and confounding conditions can complicate or produce erroneous coronary physiological measurements (Table 3). The 3 most common major technical problems are guiding catheter obstruction to flow, poor zeroing/calibration, and signal drift (pressure) or signal loss (Doppler). Additionally, for both pressure and flow measurements, suboptimal guide catheter engagement may result in inadequate delivery of bolus adenosine, producing submaximal hyperemia and thus limiting the accuracy of the FFR and CFR.
|
Guiding catheter obstruction of the coronary ostium will limit hyperemic blood flow and is generally indicated by arterial pressure wave deformity (eg, damping). Although guiding catheter side holes permit better flow and pressure transmission into the artery, some degree of obstruction may remain. This issue is avoided by not seating the guiding catheter and by using intravenous adenosine rather than intracoronary bolus. A smaller caliber guide catheter may also obviate this problem. An artificial difference between aortic and distal coronary pressures may appear because of a damped guiding catheter pressure signal (often in association with small caliber catheters caused by contrast media in the catheter) and can be recognized by the shape of the pressure waveform. Flushing the guiding catheter with saline will restore a reliable aortic pressure.
Although uncommon, pressure signal drift may occur during a procedure. In such cases, true coronary wire pressure can be verified by pulling the sensor back to the tip of the guiding catheter. In addition, the preservation of the dicrotic notch of distal pressure is a marker of adequate distal pressure transmission and is always associated with FFR >0.80.40
For the coronary Doppler flow, technical artifacts of signal acquisition also must be recognized and then minimized to obtain reliable flow data.22
In summary, the technique for sensor-wire pressure and flow measurements is identical to angioplasty guidewire placement. The induction of maximal hyperemia with intracoronary or intravenous adenosine is important for obtaining accurate data. When used by trained operators, sensor-wire measurement is generally considered safe and valuable for the important clinical data obtained.
| 2. Clinical Applications of Coronary Physiology |
|---|
|
|
|---|
|
2.1. Ischemic Thresholds of Coronary Physiological Measurements
Testing modalities for myocardial ischemia have been compared with all of the physiological indices described above. Studies attempting to limit false-positive and false-negative results provide the best correlative data. Whether radionuclide perfusion abnormalities represent true ischemia is controversial, but this method is among the most commonly used for decision-making about lesion severity and indications for intervention and thus has served as a common standard against which the physiological measurements have been assessed.
An FFR <0.75 identified coronary stenoses in patients with inducible myocardial ischemia with high sensitivity (88%), specificity (100%), positive predictive value (100%), and overall accuracy (93%). A CFR of <2.0 corresponded to reversible myocardial perfusion imaging defects with high sensitivity (86% to 92%), specificity (89% to 100%), predictive accuracy (89% to 96%), and positive and negative predictive values (84% to 100% and 77% to 95%, respectively). For intermediate lesions, the sensitivity and specificity of HSR (>0.8 mm Hg · cm1 · s) were 79% and 90%, respectively, with a predictive accuracy of 87%.23
A summary of studies of the correlations between stress imaging and coronary physiological measurements is provided in Table 5.
|
2.2. Angiographic Subsets
2.2.1. The Intermediate Stenosis
The intermediate lesion, usually reported in the range of 40% to 70% narrowing, is the most frequently encountered stenosis in patients with CAD and has been associated with a large interobserver and intraobserver variability in the reported angiographic interpretation. When FFR is <0.75 or CFR is <2.0, the stenosis is considered to be hemodynamically significant and a PCI can be supported. If FFR is >0.80 or CFR >2.5, the clinical benefit of PCI can be questioned. The FFR range 0.75 to 0.80 can be considered a gray zone in which clinical judgment must complement quantitative assessments in forming the final treatment decision. Although most studies identify a distinct threshold, this value varies slightly depending on multiple study design and patient factors. For this reason, deferral of intervention can be confidently considered with FFR values >0.80. A number of studies4144 have shown that for a given intermediate coronary stenosis and FFR >0.75, the combined risk for death or acute MI is only 1% per year with medical treatment alone. In the DEFER (Deferral versus Performance of PTCA in patients without Documented Ischemia)41 study, the risk of an adverse event increased if a vessel with an intermediate lesion and an FFR >0.75 was treated percutaneously with a bare-metal stent (Figure 9). In a similar fashion, the ILIAS (Intermediate Lesion: Intracoronary flow ASsessment versus 99mTc-MIBI SPECT) study,45 using CFR and technetium 99mTc sestamibi single-photon emission computed tomography (MIBI SPECT), examined intermediate lesions in 191 patients with multivessel CAD. PCI was performed when both CFR and SPECT were indicative of myocardial ischemia. PCI was deferred when both CFR and SPECT were normal. When disparity existed between CFR and SPECT, PCI was deferred (deferred patients n=182). At the end of 1-year follow-up, 19 events had occurred. CFR was a more accurate predictor of events than was SPECT (relative risk: 3.9).2630 The risk for death and acute MI exceeded the risk of medical treatment compared with PCI for nonhemodynamically significant lesions.
|
At this time, implantation of stents in nonsignificant plaques, even drug-eluting stents, is not supported by evidence-based studies. For an asymptomatic patient with mild stenosis and an FFR >0.75 (or CFR >2.5), one would consider noncardiac causes of the chest pain.
Despite excellent safety, some patients with deferred procedures may still have recurrent angina requiring continued medical therapy. Nonetheless, when a lesion is physiologically normal, the clinical impact of deferring intervention is associated with a satisfactory clinical outcome.4144 Like other medical tests obtained at a single point in time, in-laboratory translesional physiology may not reflect the episodic, dynamic ischemia-producing conditions of daily life, particularly those related to vasomotor changes during exercise or emotional stress, conditions often highly responsive to medical therapy. Physiological thresholds, validated by stress imaging modalities and clinical outcomes, support decisions to defer intervention while continuing medical therapy for endothelial dysfunction, hypertension, hyperlipidemia, and episodic coronary vasoconstriction. Despite normal physiological measurements at one point in time, mild hemodynamically nonsignificant lesions may progress and become significant, leading to new cardiac events. Continued surveillance and medical therapy for atherosclerosis are recommended in all such patients.
Table 6 summarizes results of studies deferring PCI of intermediate lesions on the basis of physiological end points.
|
2.2.2. Multivessel Disease
With the increasing use of coronary stents in an ever more complex patient population, a frequent application of physiological assessment involves lesion selection in patients with multivessel disease. Accurate lesion selection is important because noninvasive studies have demonstrated that MIBI SPECT fails to correctly indicate all ischemic areas in 90% of patients.46 In 35% of such patients, no perfusion defect was present, possibly because of balanced ischemia. Often, one ischemic area was masked by another, more severely underperfused area. Furthermore, when several stenoses or diffuse disease is present within one coronary artery, an abnormal MIBI SPECT hypoperfusion image cannot discriminate among the different stenoses along the length of that vessel. For clinical practice, these observations highlight that regions that are not responsible for ischemia may contain significant-appearing narrowings, whereas other, more severe-appearing lesions may not be hemodynamically important. Coronary pressure measurements are particularly useful for localizing regions of suspected ischemia.
Several small nonrandomized studies have reported the use of coronary physiology in multivessel disease.45,47,48 Recently, in patients with multivessel disease referred for bypass surgery, patients who underwent selective PCI of hemodynamically significant stenoses had a prognosis similar to that of patients who had coronary artery bypass surgery of all angiographic diseased vessels.46
2.2.3. Left Main and Ostial Lesions
Angiography cannot reliably characterize many ostial lesions, an especially critical problem when the left main coronary artery (LMCA) is involved. Furthermore, in patients with multivessel disease, uncertainty about the contribution of the LMCA to the clinical syndrome may confuse the issue of whether to perform PCI or surgery.
FFR is used to assess LMCA narrowings, with specific technical considerations for guiding catheter seating and intravenous adenosine. Because of the potential for the guiding catheter to obstruct blood flow across an ostial narrowing, FFR measurements should be performed with the guiding catheter disengaged from the coronary ostium and with hyperemia induced by intravenous adenosine. Initially, the guiding catheter and wire pressures should be matched (equalized) before the guiding catheter is seated. Then, the guiding catheter is seated, and the pressure wire is advanced into the left anterior descending or left circumflex artery. The guiding catheter is then disengaged and the intravenous adenosine infusion initiated. After 1 to 2 minutes, FFR is calculated, and thereafter the wire can be pulled back slowly to identify the exact location of the pressure drop. In case of a distal left main narrowing, this procedure may be performed twice, once with the pressure wire in the left anterior descending artery and again with the wire in the circumflex artery. Figure 10 illustrates FFR in a patient with an ostial LMCA narrowing.
|
The use of FFR in an LMCA stenosis has been examined in 54 patients.49 In those 30 patients with an FFR <0.75, surgery was performed, and in those 24 patients with an FFR
0.75, medical therapy was chosen. After a follow-up of 3 years, no differences in event-free survival rate or functional class were seen between the groups. None of the patients in the medical group experienced MI or died.
2.2.4. Serial Stenosis and Diffuse Disease
When
2 discrete stenoses are present in the same vessel, the hyperemic flow and pressure gradient through the first one will be attenuated by the presence of the second one and vice versa. One stenosis will mask the true effect of its serial counterpart by limiting the maximum hyperemia that can be achieved. This fluid, dynamic interaction between 2 serial stenoses depends on the sequence, severity, and distance between the lesions, as well as on the flow rate. When the distance between 2 lesions is >6 times the vessel diameter, the stenoses generally behave independently, and the overall pressure gradient is the sum of the individual pressure losses at any given flow rate.
The calculation of the exact FFR of each lesion separately is possible but remains academic, as the coronary wedge pressure (during coronary balloon occlusion) is needed to perform these calculations.50,51 In clinical practice, the use of the pressure pullback recording is particularly well suited for identifying the several regions of a vessel with large pressure gradients that may benefit from treatment. The stenosis with the largest gradient can be treated first, and the FFR can be remeasured for the remaining stenoses to determine the need for further treatment.
In arteries with diffuse atherosclerotic disease, unlike in normal vessels, coronary pressure noticeably declines along the length of the artery, indicating an increase in resistance caused by diffuse diameter changes. De Bruyne et al52 demonstrated that in 37 strictly normal coronary arteries, FFR was 0.97±0.02, whereas in 106 nonstenotic arteries in patients with atherosclerosis elsewhere in the coronary circulation, FFR was 0.89±0.08, with frank ischemic values <0.75 in 8% of them (Figure 11).
|
2.2.5. Diffuse Disease and Long Lesions
Considerations similar to those for patients with several discrete stenoses in 1 coronary artery can be applied to patients with diffuse CAD or long lesions. The pressure pullback recording at maximal hyperemia will provide the information that is needed to decide if and where stent implantation may be useful. The location of a focal pressure drop superimposed on the diffuse disease can be identified as an appropriate location for treatment. In some cases, the gradual decline of pressure along the vessel occurs over a very long segment, such that interventional treatment is not possible. Medical treatment (or bypass surgery) can then be elected.
2.3. Prognostic Value of Physiological Parameters
2.3.1. Percutaneous Coronary Interventions
Because of the limitations of the angiogram for precisely identifying luminal abnormalities, the angiographic criteria before and after balloon angioplasty are poor predictors of immediate and long-term prognosis. However, when coupled with a satisfactory angiographic result, coronary physiological indices were predictive of the short-term and long-term clinical outcomes after balloon angioplasty.42,53,54,54a,54b For example, in patients with single-vessel disease, a CFR
2.5 in combination with diameter stenosis
35% after balloon angioplasty alone was associated with a favorable long-term outcome (major adverse cardiac event [MACE] rates <20%).53 When compared with coronary stent implantation, physiologically guided balloon angioplasty (with <35% diameter stenosis) had similar event-free survival rates at 1 year: 86.6% and 85.6% for primary stent implantation and optimal balloon angioplasty, respectively.54c As expected, coronary stent implantation after a suboptimal balloon angioplasty was associated with a significantly improved outcome, decreasing the MACE rate from 26.7% to 10.7% (odds ratio [OR]: 3.0, P=0.005).54d Interestingly, coronary stent implantation after an optimal balloon angioplasty also improved outcome, decreasing the MACE rate from 15.9% to 6.5% (OR: 2.7, P=0.066). Although use of physiological information for guidance of balloon angioplasty has been clinically valuable, this approach has yielded to near-universal stent implantation as the treatment of choice.54c,55
For the practice of stent implantation, physiological measurements, specifically FFR, do not address adequacy of implantation but do provide prognostic information about the patients long-term results. In a multicenter trial, Pijls et al56 examined 750 patients with poststenting FFR data and found that the FFR immediately after stent implantation was an independent variable related to all MACE (Figure 12). The lowest MACE rates occurred in patients with the highest FFR values. FFR normalized (>0.95) in 36% of patients, a finding associated with an event rate of 5%. For patients with FFRs between 0.90 and 0.95 (32% of patients), the event rate was 6%. In the 32% of patients with FFRs <0.90, event rates were 20%. In the 6% of patients with FFRs <0.80, the event rate was 30%. The lower the FFR, the more likely the patient would have late events (>6 months after the procedure), which suggests that both procedurally related problems (eg, dissections) and untreated diffuse disease were associated with worse long-term outcomes. The use of FFR after stent implantation, although not routine, can provide insight into the patients prognosis.
|
2.3.2. Acute Coronary Syndromes
2.3.2.1. Acute Myocardial Infarction
Because acute coronary syndromes involve dynamic changes in both the active lesion and the supplied myocardial bed, the clinical validity of applying physiological measurements in such a setting remains uncertain.
However, the use of FFR in patients with a previous MI has been studied. De Bruyne et al57 compared SPECT imaging and FFR in 57 patients >6 days after acute MI. FFR was measured before and after the PCI. The sensitivity and specificity of FFR >0.75 for detecting normal SPECT imaging were 82% and 87%, respectively (Figure 13), with a concordance between FFR and SPECT imaging of 85% (P<0.001). When only true positive and true negative SPECT imaging studies were used, the corresponding values increased to 87%, 100%, and 94%, respectively. It was also of interest that patients with higher FFRs had higher left ventricular ejection fractions despite similar stenosis diameters. FFR could distinguish patients with positive SPECT imaging
6 days after acute MI who could benefit from revascularization.
|