From the Cardiovascular Branch (A.N.R., P.V.K., C.O., R.D., R.J.A., V.J.W., W.H.S., O.K., A.J.D., V.K.R., R.J.L.) and the Laboratory of Cardiac Energetics (M.A.G., S.S., E.R.M.), Division of Intramural Research, National Heart, Lung, and Blood Institute, National Institutes of Health, Bethesda, Md; Cardiovascular Section, Department of Medicine, University of Wisconsin, Madison (A.N.R.); and Department of Radiology, The Johns Hopkins University, Baltimore, Md (P.V.K., M.X., E.A.).
Correspondence to Robert J. Lederman, MD, Cardiovascular Branch, Division of Intramural Research, National Heart, Lung, and Blood Institute, National Institutes of Health, Building 10, Room 2c713, MSC 1538, Bethesda, MD 20892-1538. E-mail lederman{at}nih.gov
Received September 29, 2005; revision received December 13, 2005; accepted December 19, 2005.
| Abstract |
|---|
|
|
|---|
Methods and Results Carotid artery CTO was created by balloon injury in 19 lipid-overfed swine. After 6 to 8 weeks, 2 underwent direct necropsy analysis for histology, 3 underwent primary x-rayguided CTO recanalization attempts, and the remaining 14 underwent rtMRI-guided recanalization attempts in a 1.5-T interventional MRI system. Real-time MRI intervention used custom CTO catheters and guidewires that incorporated MRI receiver antennae to enhance device visibility. The mean length of the occluded segments was 13.3±1.6 cm. The rtMRI-guided CTO recanalization was successful in 11 of 14 swine and in only 1 of 3 swine with the use of x-ray alone. After unsuccessful rtMRI (n=3), x-rayguided attempts were also unsuccessful.
Conclusions Recanalization of long CTO is entirely feasible with the use of rtMRI guidance. Low-profile clinical-grade devices will be required to translate this experience to humans.
Key Words: angioplasty catheterization magnetic resonance imaging occlusion peripheral vascular disease
| Introduction |
|---|
|
|
|---|
Editorial p 1053
Clinical Perspective p 1107
X-ray fluoroscopic angiography with radiocontrast identifies the occluded inflow artery and possibly collateral-dependent outflow beyond the occlusion but cannot discriminate CTO arterial wall and lumen. In these cases, the entire occluded arterial segment remains invisible to the operator. Virtually "blind" device manipulation risks procedural failure and arterial perforation. Furthermore, ionizing radiation and nephrotoxic contrast exposure may become excessive during lengthy CTO recanalization procedures.
MRI can image the entire peripheral artery occlusion, including arterial wall, lumen content, and adjacent structures, without ionizing radiation or iodinated radiocontrast. Advances in rapid imaging, combined with development of catheter devices visible under MRI, have made real-time MRI (rtMRI)guided therapeutic interventions feasible.319 We hypothesize that rtMRI can guide catheter navigation within an occluded artery and that combined CTO recanalization and percutaneous transluminal angioplasty can be conducted entirely under rtMRI guidance with the use of custom MRI-visible devices in a swine model.
| Methods |
|---|
|
|
|---|
Histology was performed on 2 uninstrumented CTO arteries 8 weeks after balloon injury. Formalin-fixed, paraffin-embedded, transverse sections (4 µm) were taken at 5-mm intervals along the artery length and stained with hematoxylin and eosin and Massons trichrome.
Interventional MRI Suite
Procedures were performed in a previously described combined x-ray/MRI interventional suite (Axiom Artis/1.5-T Sonata, Siemens).20 In-room hemodynamics, scan control, and rendered images were displayed with the use of shielded LCD projectors. Optical microphones (Phone-Or) and filtered headsets (Magnacoustics) permitted staff communication.
Digital subtraction x-ray and CEMRA of the arch and great vessels were performed immediately before and after intervention. Typical parameters (repetition time [TR] [ms]/echo time [TE] [ms]/flip angle [FA] [degrees]/bandwidth [Hz per pixel]/voxel size [VOX] [mm]) were as follows: TR/TE/FA/bandwidth/VOX=3.7/1.4/25/375/0.3x0.5x2, slab 300x244x128 mm. Real-time steady state free precession (rtSSFP) and 2D T1-weighted (T1w) spin echo axial images representing key anatomic planes were stored for uninterrupted retrieval during the intervention. Typical parameters were as follows: rtSSFP: TR/TE/FA/bandwidth/VOX=3.5/1.7/45/975/1.0x1.4x4; T1w: TR/TE/FA/bandwidth/VOX=800/12/180/130/0.5x0.5x3, 2 averages.
During rapid device manipulation in the aortic arch, rtSSFP frame rate was accelerated 3:1 with the use of echo sharing, wherein image data are interleaved over adjacent frames. Saturation prepulses suppressed background tissue to better visualize inflation of balloons filled with gadopentetate dimeglumine (0.8% Gd-DTPA, Magnevist, Berlex) and selective intra-arterial angiography (2.4% Gd-DTPA). Uninterrupted real-time subtraction angiography was achieved by displaying the difference between maximum intensity projections in time before and after the injection. Voxel size was reduced to increase spatial resolution, and temporal image filtering (averaging) was applied to improve signal-to-noise ratio (SNR) during slow device advancement within the CTO. Multiple oblique slices could be acquired simultaneously, repositioned interactively, and individually disabled and enabled as desired. Custom scan features allowed adjustment of pulse sequence, reconstruction, and display parameters without scan interruption. Finally, signals from separate device antennae could be independently displayed along their entire length, color-highlighted, and combined in real time with anatomic images from surface coils.
Interventional Devices
Active devices were prepared and tested in our laboratory. The intent was to create devices that were conspicuous along their entire length, including the tip.
Custom MRI Active CTO Support Catheter
Distal 1-cm microcoils were mounted on 6F x100-cm angled, tapered, custom tungsten-braided catheters (Minnesota MedTec). Coil signals are transmitted along the catheter shaft under insulating polyimide. Tuning, matching, and decoupling circuitry were attached at the proximal hub to prevent heating during MRI. The device was connected to a separate MR scanner receiver channel so that device-related images could be displayed in color.
Custom MRI Active CTO Guidewire
Consecutive gold-silver-goldplated nitinol core wires were inserted within nitinol hypotubes, separated by polymer insulation. Closely wound MP35N (cobalt-chromium alloy) microcoils were attached distally to add wire flexibility, and 2.5-cm distal microcoils enhanced MRI receiver sensitivity at the tip. The final receiver-coil guidewire was 0.032 inchx155 cm, straight, and stiff. The proximal end of the guidewire had detachable connectors, permitting catheter exchange (Figure 1).
|
Custom Active Flexible "Workhorse" Guidewire
Flexible-tip nitinol guidewires (Nitrex, ev3, 0.032 inchx155 cm) were altered to serve as active, "loopless" design21 receiver coils that were visible along their entire length. These were used as "workhorse" wires to safely deliver the CTO catheter to the origin of the occluded carotid artery.
Delivery Sheath
10F delivery sheaths (Fast Cath, St Jude; Brite Tip, Cordis) were trimmed to 75 cm. Stainless steel MRI markers (1 mmx0.014 inch) were bonded to the sheath tip, creating passive markers (discrete signal void) during MRI.
Device Testing
MRI heating tests (Appendix) were performed on CTO support catheters and guidewires with the use of previously described methods.22,23
rtMRI-Guided CTO Recanalization
The primary study end point was successful traversal of occluded arteries from their patent origins to their patent outflow regions. The secondary study end point was restoration of antegrade flow by adjunctive balloon angioplasty.
Animals were pretreated with aspirin 325 mg and clopidogrel 300 mg. Intravenous heparin 100 U/kg bolus was given initially and supplemented with 50 U/kg every 2 hours. After sheath insertion, the active CTO support catheter was navigated to the brachiocephalic artery over the active floppy guidewire and positioned proximal to the occluded LCA. The floppy wire was exchanged for the active CTO wire. The CTO guidewire and catheter were steered intraluminally through the CTO with the use of multislice rtMRI. Two parallel slices orthogonal to the CTO artery centerline were used to ensure an intraluminal device course. One slice tracked the wire tip, and the second was advanced ahead of the wire tip, providing a target for the operator. Once the CTO was traversed, a 260-cm polytetrafluoroethylene guidewire (Glidewire, Terumo) was used to exchange the CTO guide catheter for a 100-mm-long peripheral angioplasty balloon sized 1:1 to the artery wall. After several balloon inflations with 5 mmol/L Gd-DTPA, selective angiography was performed with the use of 5 mL of 10 mmol/L Gd-DTPA. Intra-arterial nitroglycerin 200 to 400 µg was administered as needed to treat spasm.
Comparative X-RayGuided Recanalization of CTO Model
X-rayguided recanalization was attempted in 3 swine by experienced interventionists to assess the suitability and procedural difficulty of the CTO model. Contemporary techniques and devices were used (Shinobi 0.014 inch, Cordis; Cross-It 0.010 to 0.014 inch, Guidant; Glidewire Straight Stiff and Super-Stiff 0.035 inch; 4F Glidecath, Terumo). Each attempt used high-performance x-ray imaging (Artis FC, Siemens) with 13-cm field of view, 15 frames per second, appropriate collimation, and at least 60 minutes of fluoroscopy. In 3 additional swine, after unsuccessful MRI-guided recanalization, x-rayguided recanalization also was attempted.
Statistics and Analysis
Continuous variables were reported as mean±SD and were compared with a 2-tailed Student t test (Excel 2003, Microsoft). Discrete variables (procedure success) were compared by Fisher exact test.24 The authors had full access to the data and take responsibility for its integrity. All authors have read and agree to the manuscript as written.
| Results |
|---|
|
|
|---|
CTO Model Geometry and Histopathology
Native LCA diameter by x-ray was 4.6±0.7 mm before balloon injury, measured 70 mm cranial to the carotid bifurcation. Immediate preintervention CTO diameter (T1w) at this level and CTO length (CEMRA) were 4.7±1.3 and 133±16 mm, respectively. By comparison, the contralateral carotid artery diameter (T1w) at this level, the CTO minimal luminal diameter (T1w), and CTO maximal external wall diameter (T1w) were 5.0±0.7, 3.1±1.3, and 6.7±1.2 mm, respectively. These measurements suggest zones of positive (outward) and negative (inward) remodeling in this swine CTO model and are consistent with histology. MRI identified the entire CTO artery wall endoluminal and external boundaries and occasional proximal dissection flaps resulting from initial balloon injury (Figure 2). Histology of the explanted, naive CTO arteries revealed dense luminal collagen deposition, disrupted intima, microchannels, and distal organized thrombus (Figure 3).
|
|
rtMRI-Guided Recanalization
The primary end point, rtMRI-guided CTO recanalization, was achieved in 11 of 14 animals. The shaft profile and tips of the active devices were conspicuous while navigating the aorta and brachiocephalic arteries (Figure 4). The occluded artery walls and active interventional devices were easily visualized under MRI (Figure 5). Multiplanar imaging facilitated continuous assessment of device-anatomic relationships, enabling intraluminal device traversal. Mean time to wire traversal was 55±22 minutes (spread, 25 to 99 minutes), with an observed learning curve between the first and last third of interventions performed (69±23 versus 34±7 minutes; P=0.048). After balloon angioplasty, real-time selective MRA demonstrated brisk antegrade flow in 6 of 11 of those successfully recanalized with a guidewire (Figure 6). Gross necropsy of all successfully recanalized animals revealed mild artery thickening with no extramural hematoma (Figure 7) and appeared similar to the uninstrumented CTO lesions.
|
|
|
|
In 3 of 14 animals, the active guidewire would not advance beyond the proximal one third of the occluded segment despite an apparently intraluminal trajectory. After careful MRI interrogation to rule out any adverse sequelae such as hematoma, x-rayguided attempts in all 3 were similarly unsuccessful.
In 2 of 11 successfully recanalized under rtMRI, antegrade blood flow was not achieved despite intra-arterial nitroglycerin and several prolonged balloon inflations because of refractory recoil, spasm, and unavailability of suitable stents. Intraluminal position was confirmed by selective real-time MR and x-ray angiography through the wire port of balloons positioned distal to the lesion.
Homemade active CTO support catheters would not advance over active CTO guidewires that successfully traversed the occlusions in 3 of 11, preventing balloon exchange and subsequent angioplasty. In 1 of these 3 (the very first experimental animal), overaggressive attempts at advancing the active catheter over the wire resulted in clear real-time visualization of extraluminal CTO wire exit followed by extravascular hematoma accumulation. This complication was immediately recognized in rtMRI, and the procedure was aborted.
Comparative X-RayGuided Recanalization
Primary x-rayguided CTO recanalization was successful in only 1 of 3 different animals attempted (requiring 45 minutes of fluoroscopy time). Crossover x-ray recanalization was also attempted in 3 of 14 animals after unsuccessful recanalization under rtMRI; none were successful. When these were combined, 1 of 6 x-rayguided procedures were successful.
Successful CTO recanalization was more likely (79% versus 17%; P=0.02) during rtMRI attempts (11 of 14) compared with all x-rayguided attempts (1 of 6). During extended primary x-ray recanalization attempts, wire exit caused contrast extravasation in one animal, and dissection was associated with lethal mediastinal hematoma in another.
rtMRI Enhancements
Subtraction real-time MRA improved visibility of intra-arterial Gd-DTPA injections compared with saturation preparation alone (Figure 6E and 6F). We increased spatial resolution by reducing the voxel size, but this degraded the SNR. We used temporal image filtering (averaging) to improve this reduced SNR during periods of slow device motion.
| Discussion |
|---|
|
|
|---|
This experience demonstrates successful wholly rtMRI-guided endovascular recanalization of long and challenging CTO with the use of custom devices in a suitable animal model. In this model, MRI was successful in 79% of attempts with the use of homemade devices and in 17% of all x-ray attempts with the use of higher-performance commercial devices.
The model CTO artery diameter and length resemble long (albeit rigid) human iliofemoral artery occlusions. These lesions would be considered TransAtlantic Inter-Society Consensus (TASC) type D, or best suited for surgical rather than endovascular recanalization.1 Unlike x-ray, MRI delineates the occluded artery walls, permitting truly image-guided wire navigation. Custom "active" MRI CTO devices prove conspicuous and distinct from adjacent tissues. Procedure duration declined throughout this experience because of a "learning" phenomenon.
Device conspicuity, simultaneous display of multiple real-time imaging slices, and uninterrupted interactive features contributed to procedural success. Real-time selective MRA techniques with small volumes of dilute Gd-DTPA confirmed artery patency. This work represents an advance because a broad assortment of imaging and device features are fully integrated to support complex and completely rtMRI-guided vascular interventions.
Long conductive devices may heat significantly during MRI radiofrequency excitation and pose a safety concern. The custom active devices used here are appropriately tuned, matched, and decoupled and do not heat during scanning in a suitable test phantom (Appendix).
Our CTO model differs from tortuous, lipid, and calcium-rich atherosclerotic human iliofemoral disease. Severe arterial recoil and spasm prevented antegrade flow (n=2) or guiding catheter traversal (n=3) despite successful wire recanalization. The lesions were sufficiently long and challenging for a feasibility experiment. Of note, x-rayguided recanalization attempts were less successful, even though they used higher-performance commercial clinical devices. Indeed, MRI guidance offers the potential to visualize the entire occluded segment and luminal boundaries, particularly in severely tortuous lesions often found in human peripheral CTO.
This experience adds to that of other published reports of rtMRI-guided procedures that traverse anatomic boundaries such as transjugular intrahepatic portosystemic shunt14,44 and atrial transseptal puncture.16,19 CTO recanalization is a more complex procedure because the vascular lumen is smaller, the diseased vessels tend to be tortuous, and the risk of lethal perforation is higher when catheters are manipulated within the high-pressure arterial system. These support the principle that MRI has potential utility to guide conventional and novel catheter-based interventions because of superior tissue visualization.
Because the SNR of proton MRI is very low, there is inadequate temporal and spatial resolution to image coronary arteries in real-time with the use of MRI. Barring unforeseen technical breakthroughs, it is unlikely that such recanalization procedures can be conducted on occluded coronary arteries.
Limitations
The comparison of x-ray and MRI-guided recanalization attempts should be interpreted with caution because of the imbalance in the number of animals in each group and because primary and crossover attempts were combined for analysis.
Repeated use of recycled devices, abrupt transition points, high friction, and poor mechanical responsiveness account for failure to recanalize and deliver catheter devices into the target arterial occlusion in a few animals. Our homemade devices had inferior capacity for pushing, torque, and tip-bending performance compared with clinical x-ray devices in bench testing (data not shown). Low-profile, tapered, trackable, steerable, and hydrophilic-coated active MRI devices can be manufactured with available technology but would have been prohibitively expensive to manufacture in the small quantities tested here. In addition, suitable stents would have treated flow-limiting artery recoil and spasm but were not used here. Metal alloys such as nitinol, platinum-iridium, or nickel-cobalt-chromium have less MRI susceptibility artifact than stainless steel. Numerous reports already have been published of experimental rtMRI-guided deployment of such devices.7,17,18,45 Nevertheless, our primary end point, rtMRI-guided CTO recanalization, was successfully accomplished with the use of our rudimentary device prototypes. Subsequent therapy to maintain antegrade flow and patency such as stenting could be conducted under x-ray, immediately after successful rtMRI CTO recanalization.
Conclusions
This work demonstrates the feasibility of rtMRI guidance to recanalize arterial CTO with the use of custom active MRI visible devices in a suitably challenging swine model. Future image-guided human peripheral artery CTO interventions may be possible with the use of this technology. However, clinical-grade catheter devices must be manufactured to translate these findings into patients.