(Circulation. 2006;113:2054-2061.)
© 2006 American Heart Association, Inc.
Coronary Heart Disease |
From the Division of Cardiovascular Medicine (M.K.C.N., A.C.Y., W.F.F.), Stanford University Medical Center, Stanford, Calif, and the Department of Cardiology (M.K.C.N.), Royal Prince Alfred Hospital, Sydney, Australia.
Correspondence to William F. Fearon, Stanford University Medical Center, H3554, 300 Pasteur Dr, Stanford, CA 94305. E-mail wfearon{at}stanford.edu
Received November 23, 2005; revision received February 2, 2006; accepted March 2, 2006.
| Abstract |
|---|
|
|
|---|
Methods and Results Using a pressure-temperature sensor-tipped coronary wire, thermodilution-derived CFR and IMR were measured, along with fractional flow reserve (FFR), in 15 coronary arteries (15 patients) under the following hemodynamic conditions: (1) twice at baseline; (2) during right ventricular pacing at 110 bpm; (3) during intravenous infusion of nitroprusside; and (4) during intravenous dobutamine infusion. Mean CFR did not change during baseline measurements or during nitroprusside infusion but decreased during pacing (from 3.1±1.1 at baseline to 2.3±1.2 during pacing, P<0.05) and during dobutamine infusion (from 3.0±1.0 to 1.7±0.6 with dobutamine, P<0.0001). By comparison, mean values for IMR and FFR remained similar throughout all hemodynamic conditions. The mean coefficient of variation between 2 baseline measurements was significantly lower for IMR (6.9±6.5%) and FFR (1.6±1.6%) than for CFR (18.6±9.6%; P<0.01). Mean correlation between baseline measurements and each hemodynamic intervention was superior for IMR (r=0.90±0.05) and FFR (r=0.86±0.12) compared with CFR (r=0.70±0.05; P<0.05).
Conclusions Compared with CFR, IMR provides a more reproducible assessment of the microcirculation, which is independent of hemodynamic perturbations. Simultaneous measurement of FFR and IMR may provide a comprehensive and specific assessment of coronary physiology at both epicardial and microvascular levels, respectively.
Key Words: microcirculation pressure coronary disease
| Introduction |
|---|
|
|
|---|
Clinical Perspective p 2061
Guidewire-based measurement of coronary flow reserve (CFR), either by Doppler flow or thermodilution techniques, has become an increasingly important invasive method for assessing the physiological significance of coronary disease.7,8 However, use of CFR to interrogate the microcirculation independently is limited because CFR interrogates the flow status of both the epicardial artery and the microcirculation but does not allow discrimination between these 2 components.7 Furthermore, CFR is limited by its dependence on heart rate and blood pressure, thereby calling into question its reproducibility.9
With recent technological advances, it is now possible to measure pressure and to estimate coronary artery flow simultaneously with a single pressure-temperature sensor-tipped coronary wire.10,11 By the thermodilution technique, the mean transit time (Tmn) of room-temperature saline injected down a coronary artery can be determined and has been shown to correlate inversely with absolute flow.10 From this technique, a thermodilution-based CFR can be derived that has been shown to correlate well with Doppler velocity wire-derived CFR and with absolute flow as measured by a flow probe but that has the same conceptual disadvantages as Doppler-derived CFR.8,11 Using this thermodilution method, we reently proposed and validated a novel index of microcirculatory resistance (IMR) for assessing the status of the microcirculation independent of the epicardial artery.12 In an animal model, IMR, defined as the distal coronary pressure divided by the inverse of the hyperemic mean transit time, correlated well with an accepted experimental method for measuring microvascular resistance.12
Unlike CFR, IMR is derived at peak hyperemia, thereby eliminating the variability of resting vascular tone and hemodynamics. We therefore hypothesized that IMR would not only be a more specific but also a more reproducible measure of coronary microcirculatory status that may be less subject to hemodynamic variation. This is especially relevant in the catheterization laboratory, particularly during interventional procedures, during which changes in heart rate, blood pressure, and cardiac contractility are likely to occur. The goal of the present study was to evaluate the feasibility, reproducibility, and hemodynamic dependence of IMR compared with CFR in humans.
| Methods |
|---|
|
|
|---|
50% and fractional flow reserve [FFR] >0.75) were included in the study. Patients were excluded if they had significant renal insufficiency (serum creatinine >1.5 mg/dL), recent acute myocardial infarction (within 1 week), or congestive heart failure. The study protocol was approved by Stanford Universitys Administrative Panel on Human Subjects. Every patient provided informed written consent. All patients were brought to the cardiac catheterization laboratory in a fasting state without discontinuation of their cardiac medications. After conventional diagnostic coronary angiography was performed, 3000 to 5000 U of intravenous heparin was administered, and a 6F coronary guiding catheter was used to engage the coronary artery of interest. Intracoronary nitroglycerin (200 µg) was given. A 0.014-in coronary pressure wire (Radi Medical Systems, Wilmington, Mass) was calibrated, equalized to the guiding catheter pressure with the sensor positioned at the ostium of the coronary artery, and then advanced to the distal coronary artery (at least two thirds of the way down the vessel). CFR, IMR and FFR were then measured under a variety of hemodynamic conditions as described below.
Coronary Physiology Measurements
CFR, IMR, and FFR were measured by methods described previously.12,14 Briefly, with commercially available software (Radi Medical Systems) the shaft of the pressure wire can act as a proximal thermistor by detecting changes in temperature-dependent electrical resistance. The sensor near the tip of the wire simultaneously measures pressure and temperature and can thereby act as a distal thermistor. The transit time of room-temperature saline injected down a coronary artery can be determined with a thermodilution technique.10,11 Three injections of 3 mL of room-temperature saline were made down the coronary artery, and the resting mean transit time (Tmn) was measured. Intravenous infusion of adenosine (140 µg · kg1 · min1) was then administered to induce steady state maximal hyperemia, and 3 more injections of 3 mL of room-temperature saline were made, and the hyperemic Tmn was measured. Simultaneous measurements of mean aortic pressure (Pa, by guiding catheter) and mean distal coronary pressure (Pd, by pressure wire) were also made in the resting and maximal hyperemic states. CFR was calculated as resting Tmn divided by hyperemic Tmn. IMR was calculated as the distal coronary pressure at maximal hyperemia divided by the inverse of the hyperemic Tmn. FFR was calculated by the ratio of Pd/Pa at maximal hyperemia.
To investigate the effects of hemodynamic changes on coronary physiology measurements, CFR, IMR, and FFR were measured under 4 different hemodynamic conditions: (1) Baseline: To study the intrinsic variability of the coronary physiology indices, measurements were made twice under baseline conditions, once before any hemodynamic intervention compared with a second measurement after all hemodynamic interventions described below had been completed. (2) Tachycardia: The measurements were repeated during right ventricular pacing at 110 bpm (with a 5F bipolar pacemaker lead introduced via the right femoral vein). (3) Hypotension: The measurements were repeated during intravenous infusion of nitroprusside (0.5 to 2 µg · kg1 · min1) titrated to achieve a reduction of systolic blood pressure of
20 mm Hg. (4) Increased cardiac contractility: Measurements were repeated during a 5-minute intravenous dobutamine infusion starting at 10 µg · kg1 · min1, which then was increased to 20 µg · kg1 · min1 provided there was no significant increase in heart rate at the lower dose (Figure 1). After each intervention, heart rate, mean aortic pressure, mean coronary transit time, and distal coronary pressure were allowed to return to their baseline values before the next intervention. In 4 patients, the resting blood pressure was too low to allow nitroprusside infusion. In 4 patients, a repeat baseline measurement after all interventions was not performed: 2 (who also underwent percutaneous intervention during the same procedure) because of excessive procedure time and 2 because of a combination of dyspnea arising from intravenous adenosine infusion and prolonged procedure time. Three patients were excluded from the dobutamine arm because of concurrent ß-blocker usage. Hence, the baseline variability protocol was completed in 11 patients; the effect of tachycardia was studied in 15 patients; the effect of nitroprusside was studied in 11 patients; and the effect of dobutamine was studied in 12 patients.
|
Statistical Analysis
Continuous values are presented as mean±SD. Mean values were compared with Student paired and unpaired t tests. Differences in study variables between baseline and different hemodynamic conditions were compared by 2-way ANOVA for repeated measures with assessment of intergroup differences by Bonferroni multiple comparisons test. Linear regression analysis was used to compare the relationship between IMR, CFR, and FFR values under different hemodynamic conditions. Probability values of <0.05 were considered statistically significant.
The authors had full access to the data and take full responsibility for its integrity. All authors have read and agree to the manuscript as written.
| Results |
|---|
|
|
|---|
|
Reproducibility of Thermodilution Measurements
IMR and thermodilution-derived CFR could be easily measured in all cases. The coefficient of variation of 3 consecutive thermodilution-derived coronary transit time measurements at baseline was 7.2±3.0% at rest and 4.4±3.0% at maximal hyperemia (P=0.02). The distance of the pressure-temperature sensor of the coronary wire from the guide catheter tip has been noted as a potential source of error for obtaining coronary thermodilution curves, with short distances being associated with time intervals that are too short to allow for adequate curve fitting.11 In the present study, the wire sensor was located in the distal third of the coronary artery in all cases; the mean transducer distance from the guide tip was 9.6±2.0 cm, and there was no significant correlation between transducer distance and mean resting or hyperemic coronary transit times (P>0.1).
Baseline Hemodynamics and Intrinsic Variability of CFR and IMR
To evaluate the intrinsic variability of CFR and IMR under baseline conditions, we compared 2 baseline measurements, 1 taken before hemodynamic interventions (baseline 1 [B1]) and a second after a series of hemodynamic interventions (baseline 2 [B2]) that included right ventricular pacing at 110 bpm, nitroprusside infusion, and dobutamine infusion. The hemodynamic variables and calculated indices for both baseline conditions are shown in Table 2. In the resting state, heart rate, mean coronary transit times, and systemic and distal coronary pressures were similar between the 2 baseline measurements. During B1, intravenous adenosine infusion increased heart rate (from 69±10 to 76±13 bpm) and reduced both mean systemic pressure (mean reduction 19±13 mm Hg) and mean distal coronary pressure (mean reduction 22±12 mm Hg; P<0.05 for hyperemia versus rest). Similar responses to adenosine infusion were observed during the second baseline measurements (Table 2). Mean values for CFR (2.8±0.8 and 2.7±1.1 for B1 and B2, respectively) and IMR (22.6±6.0 and 21.9±7.2 for B1 and B2, respectively) were similar between the 2 baseline measurements. IMR values between the first and second baseline measurements were very highly correlated (r=0.96, y=4.0+1.1x, P<0.0001). The correlation between first and second baseline CFR values was significant but weaker (r=0.77, y=0.20+1.1x, P=0.0055). The coefficient of variation between first and second baseline measurements was significantly better for IMR (6.9±6.5%) than for CFR (18.6±9.6%; P<0.01; Figure 2).
|
|
Hemodynamic Dependence of CFR and IMR
To evaluate the hemodynamic dependence of CFR and IMR, we also measured these indices during a variety of hemodynamic interventions including right ventricular pacing at 110 bpm, nitroprusside infusion, and dobutamine infusion. A representative series of measurements for 1 patient is shown in Figure 1.
To study the effects of heart rate changes on CFR and IMR, we measured these indices before and during right ventricular pacing at 110 bpm. The hemodynamic variables and calculated indices for baseline and pacing conditions are shown in Table 3. Right ventricular pacing increased the mean heart rate by 39±10 bpm at rest and by 32±14 bpm at peak hyperemia (P<0.0001 versus baseline). Resting systemic and coronary pressures were similar between baseline and pacing conditions. Pacing was associated with a reduction in resting Tmn from 0.97±0.33 to 0.83±0.30 seconds, but this was not statistically significant. At peak hyperemia, mean values for Tmn and distal coronary pressure for baseline and pacing conditions were similar (Table 3). CFR decreased significantly from 3.1±1.1 at baseline to 2.3±1.2 during pacing (P<0.05). In contrast, IMR during pacing (22.9±6.9 U) was similar to that at baseline (21.8±6.5 U; P=NS; Figure 3). IMR values during pacing were highly correlated to those at baseline (r=0.91, y=2.0+0.96x, P<0.0001). In comparison, CFR values during pacing had a weaker correlation with baseline CFR values (r=0.65, y=0.21+0.72x, P=0.009).
|
|
To evaluate the effects of blood pressure alterations on CFR and IMR, we measured these indices at baseline and during nitroprusside infusion. The hemodynamic changes induced by nitroprusside infusion (mean rate of infusion 0.6±0.2 µg · kg1 · min1) are shown in Table 4. At rest, nitroprusside infusion reduced both mean systemic pressure and distal coronary pressure by an average of 19±9 mm Hg without significantly increasing heart rate (P<0.001 versus baseline; Table 4). During nitroprusside infusion, maximal hyperemia induced by intravenous adenosine infusion produced further incremental reductions in mean systemic pressure (by 8±9 mm Hg) and distal coronary pressure (by 13±7 mm Hg; P<0.01 for hyperemia versus rest). Reduction in blood pressure due to nitroprusside infusion was associated with a longer hyperemic Tmn than at baseline, but this was not statistically significant (0.33±0.08 for baseline versus 0.40±0.13 for nitroprusside, P=NS). Compared with baseline, nitroprusside infusion had no significant effect on either CFR or IMR (P=NS). IMR values during nitroprusside infusion correlated well with those at baseline (r=0.87, y=2.5+1.1x, P=0.0006). CFR values during nitroprusside infusion also correlated with those at baseline, but not as strongly (r=0.71, y=0.26+0.96x, P=0.01).
|
To evaluate the effects of changes in cardiac contractility on CFR and IMR, we measured these indices at baseline and during dobutamine infusion. The hemodynamic changes produced by dobutamine infusion are shown in Table 5. Dobutamine infusion (mean rate of infusion 15.3±5.2 µg · kg1 · min1) increased heart rate at rest (from 71±11 bpm at baseline to 85±17 bpm with dobutamine) and during peak hyperemia (from 79±14 bpm at baseline to 97±16 bpm with dobutamine; P<0.01 versus baseline). Resting and hyperemic mean systemic and distal coronary pressures were unaffected by dobutamine infusion (Table 5). At rest, dobutamine infusion reduced Tmn compared with baseline (0.99±0.36 seconds at baseline versus 0.55±0.17 seconds during dobutamine infusion, P<0.001 versus baseline) but had no effect on Tmn during maximal hyperemia induced by adenosine infusion (0.34±0.11 at baseline versus 0.35±0.13 with dobutamine, P=NS). As a consequence of increased resting coronary flow induced by dobutamine, CFR values measured during dobutamine infusion were significantly lower than those at baseline (3.0±1.0 at baseline versus 1.7±0.6 with dobutamine, P<0.0001). In contrast, IMR values were unchanged by dobutamine infusion (22.2±6.0 at baseline versus 23.6±8.2 with dobutamine, P=NS). IMR values during dobutamine infusion correlated well with those at baseline (r=0.84, y=1.7+1.2x, P<0.0001). In comparison, CFR values during dobutamine infusion exhibited a significant but less robust correlation with those at baseline (r=0.67, y=0.40+0.47x, P=0.007). The mean correlation between baseline measurements and each hemodynamic intervention (including repeat baseline measurements) was superior for IMR compared with CFR (r=0.90±0.05 for IMR versus r=0.70±0.05 for CFR, P<0.05; Figure 3).
|
Hemodynamic Dependence of FFR
FFR was also simultaneously measured under all hemodynamic conditions described above. Mean FFR values were very similar under all hemodynamic conditions (Tables 2 through 5![]()
![]()
), a finding consistent with previous literature.9 The coefficient of variation between the first and repeat baseline measurements was 1.6±1.6%. The mean correlation between baseline measurements and each hemodynamic intervention (including repeat baseline measurements) was 0.86±0.12, which was similar to the mean correlation for IMR and superior to that for CFR (P<0.05; Figure 3).
| Discussion |
|---|
|
|
|---|
The salient findings of the present study are as follows: (1) IMR demonstrates less intrinsic variability and better reproducibility at baseline than CFR, and (2) whereas CFR is very sensitive to hemodynamic changes, IMR is largely independent of variations in hemodynamic state. Furthermore, FFR (when simultaneously measured with IMR and CFR) is highly reproducible and also largely independent of hemodynamic state. These findings suggest that IMR could be reliably applied in the catheterization laboratory for interrogation of microcirculatory resistance. Furthermore, simultaneous measurement of FFR and IMR with a single pressure-temperature sensor-tipped coronary wire may provide a simple means for comprehensive and specific assessment of coronary physiology at both epicardial and microvascular levels, respectively.
In a recent study using a porcine animal model, we found that IMR distinguished between normal and abnormal microcirculatory function and correlated well with true microcirculatory resistance as measured by an external flow probe and pressure wire.12 Furthermore, IMR, in its simplest form, was not significantly affected by the presence of a moderate to severe epicardial stenosis and is therefore a specific measure of the state of the microcirculation, unlike CFR. In more severe stenoses, in which collateral flow may be contributing to myocardial perfusion, a more complex form of IMR, which incorporates the coronary wedge pressure, is necessary to accurately determine microvascular resistance.13,15
Because IMR is derived at peak hyperemia, we postulated that it would be independent of resting vascular tone and hemodynamics. The present study demonstrates that IMR is easily measured in humans with a commercially available pressure-temperature sensor-tipped coronary wire and that its values are highly reproducible. Furthermore, variations in hemodynamic status, including changes in heart rate, blood pressure, and contractility, do not significantly affect IMR measurements. During all hemodynamic interventions in the present study (rapid right ventricular pacing, nitroprusside infusion, and dobutamine infusion), IMR exhibited greater hemodynamic stability than CFR.
Use of CFR to evaluate the microcirculation is limited by the fact that CFR interrogates the entire coronary system, including the epicardial artery and the microcirculation.7 For this reason, a patient with epicardial disease but with normal microcirculatory function can have an abnormal CFR, which potentially limits the applicability of CFR when assessing microvascular disease. Furthermore, because CFR represents a ratio between peak hyperemic and resting coronary flow, factors that affect resting hemodynamics, such as heart rate and contractility, may affect the reproducibility of CFR.9 Previous studies have shown that Doppler flow velocityderived CFR is significantly reduced by tachycardia9,16,17 and by increased contractility9 but is not significantly affected by changes in blood pressure due to compensatory changes in coronary blood flow.9,16,17 Consistent with these previous studies, the present study documents the hemodynamic dependence of thermodilution-derived CFR. In the present study, hypotension induced by nitroprusside infusion had no effect on thermodilution-derived CFR. In contrast, right ventricular pacinginduced tachycardia and dobutamine infusion were both associated with significant reductions in thermodilution-derived CFR values, largely due to an increase in resting coronary blood flow (and hence a reduction in resting Tmn). Hence, like Doppler flow velocityderived CFR, interpretation of serial measurements of thermodilution-derived CFR are limited by a high degree of hemodynamic variability, largely as a result of changes in basal coronary blood flow.
FFR is measured during maximal hyperemia and therefore is not prone to variability due to fluctuations in baseline hemodynamic state. In the present study, we show that when peak hyperemia is induced by intravenous adenosine, the intrinsic variability of FFR is very low (<2%) and may be superior to that for FFR measured with intracoronary adenosine,9 a route of administration that may not induce as strong a hyperemic response and that is associated with a very short half-life and hence is potentially subject to greater variability. FFR measurements, like IMR, demonstrated no significant variation during any of the hemodynamic conditions in the present study.
Study Limitations
IMR, FFR, and CFR are limited by their reliance on the achievement of maximal hyperemia. Failure to achieve peak hyperemia, by not achieving maximal reduction in microvascular resistance, may result in overestimation of IMR. Conversely, CFR will be underestimated in the absence of maximal hyperemia. In the case of FFR, if maximal hyperemia does not occur, the pressure gradient across a stenosis will be underestimated and the FFR overestimated. For these reasons, intravenous adenosine, considered the reference standard for induction of peak hyperemia, was used for the present study. Use of less effective hyperemic agents/protocols may affect the reproducibility of these indices.
In the present study, dobutamine infusion was associated with an increase in heart rate, which in itself, can produce increased contractility.18 However, although the influences of contractility and heart rate were not divorced from each other, the heart rate increase observed with dobutamine infusion was much less than that observed with right ventricular pacing, which suggests that an increase in contractility had indeed been tested.
The effect of the severity of epicardial stenosis on measurement of microvascular resistance is controversial. Some have suggested that the minimum achievable microvascular resistance increases with the increasing severity of an epicardial artery stenosis.19,20 In contrast, we and others have reported that microvascular resistance is not affected by increasing epicardial artery stenosis if collateral flow is taken into account.13,15 In the present study, all coronary physiological measurements were made in arteries that were either normal or had only minor angiographic stenoses. In cases with severe epicardial stenosis, the simplified measurement of IMR, as used in the present study, may overestimate resistance because it does not account for collateral flow, and a more complex measurement of IMR that incorporates the coronary wedge pressure is necessary.13
Microvascular resistance has also been determined invasively by measuring distal pressure and estimating flow with a Doppler wire.21 Because of the additional complexity of using a Doppler wire, we did not test the reproducibility or hemodynamic dependence of this technique.
Lastly, the distance of the pressure wire down a vessel will impact the measured hyperemic transit time and the IMR. The variability of IMR depending on the distance of the wire down the vessel was not tested in the present study, although in this study, there was no correlation between transducer distance and transit times.
Conclusions
Despite the importance of the status of the microcirculation in determining clinical outcomes in a wide variety of cardiovascular conditions, a simple and reproducible method for invasively assessing the state of the coronary microcirculation has been lacking. Measurement of CFR for assessing the coronary microcirculation has been limited by its lack of specificity and its high degree of hemodynamic variability. IMR is a new index for specific and quantitative assessment of coronary microcirculatory resistance that can be measured easily in the cardiac catheterization laboratory. IMR, by virtue of being more reproducible and less hemodynamically dependent than CFR, appears to be superior to CFR for assessing the coronary microcirculation. Finally, simultaneous measurement of FFR and IMR, by specifically quantifying the status of the epicardial artery and microcirculation, respectively, may provide a simple, comprehensive means of evaluating the physiological state of a coronary artery in the catheterization laboratory.
| Acknowledgments |
|---|
Disclosures
None.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
N. Beohar, A. K. Erdogan, D. C. Lee, H. N. Sabbah, M. J. Kern, J. Teerlink, R. O. Bonow, and M. Gheorghiade Acute heart failure syndromes and coronary perfusion. J. Am. Coll. Cardiol., July 1, 2008; 52(1): 13 - 16. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Cuisset, M. Hamilos, N. Melikian, E. Wyffels, J. Sarma, G. Sarno, E. Barbato, J. Bartunek, W. Wijns, and B. De Bruyne Direct stenting for stable angina pectoris is associated with reduced periprocedural microcirculatory injury compared with stenting after pre-dilation. J. Am. Coll. Cardiol., March 18, 2008; 51(11): 1060 - 1065. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. F. Fearon, M. Shah, M. Ng, T. Brinton, A. Wilson, J. A. Tremmel, I. Schnittger, D. P. Lee, R. H. Vagelos, P. J. Fitzgerald, et al. Predictive value of the index of microcirculatory resistance in patients with ST-segment elevation myocardial infarction. J. Am. Coll. Cardiol., February 5, 2008; 51(5): 560 - 565. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Sezer, H. Oflaz, T. Goren, I. Okcular, B. Umman, Y. Nisanci, A. K. Bilge, Y. Sanli, M. Meric, and S. Umman Intracoronary Streptokinase after Primary Percutaneous Coronary Intervention N. Engl. J. Med., May 3, 2007; 356(18): 1823 - 1834. [Abstract] [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2006 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |