(Circulation. 2006;113:1305-1310.)
© 2006 American Heart Association, Inc.
Imaging |
From Klinik für Herz- und Kreislauferkrankungen (J.H., M.H., A.K., A.S.) and Institut für Radiologie und Nuklearmedizin (T.M., E.H., S.M.), Deutsches Herzzentrum München, Klinik an der TU München, Munich, Germany; and Institut für Strahlenschutz (M.Z.), GSF-Forschungszentrum für Umwelt und Gesundheit, Neuherberg, Germany.
Correspondence to Dr Jörg Hausleiter, Deutsches Herzzentrum, Lazarettstrasse 36, 80636 München, Germany. E-mail hausleiter{at}dhm.mhn.de
Received August 22, 2005; de novo received November 18, 2005; revision received January 5, 2006; accepted January 6, 2006.
| Abstract |
|---|
|
|
|---|
Methods and Results Radiation dose was estimated for 1035 patients undergoing coronary CTA. Scanning algorithms with and without an ECG-dependent dose modulation and with a reduced tube voltage were investigated on dose estimates and image quality. In the entire patient cohort, radiation dose estimates were 6.4±1.9 and 11.0±4.1 mSv for 16- and 64-slice CTA, respectively (P<0.01). The reduction in radiation dose estimates ranged between 37% and 40% and between 53% and 64% with the use of ECG-dependent dose modulation and with the combined use of the dose modulation and a reduced tube voltage, respectively. The reduction in dose estimates was not associated with a reduction in diagnostic image quality as assessed by the signal-to-noise ratio and by the frequency of coronary segments with diagnostic image quality.
Conclusions The increase in spatial and temporal resolution with 64-slice CTA is associated with an increased radiation dose for coronary CTA. Dose-saving algorithms are very effective in reducing radiation exposure and should be used whenever possible.
Key Words: imaging tomography angiography
| Introduction |
|---|
|
|
|---|
Clinical Perspective p 1310
Many of the technical factors that have enhanced image quality, particularly spatial and temporal resolution, also affect the radiation dose received from a CT examination of the heart.4 The rapid evolution of cardiac CT and its increasing availability, combined with the significance of the CT dose levels, have created a compelling need to assess the impact of different scanning protocols on CT dose. Previous studies have mainly focused on the theoretical and physical aspects of CT dose. Therefore, the aim of the present retrospective study was (1) to estimate the effective dose from cardiac CT angiography in daily practice, (2) to compare the dose estimates from 16- and 64-slice cardiac CT angiography, and (3) to evaluate the impact of 2 different strategies on dose savings and image quality with both scanner types.
| Methods |
|---|
|
|
|---|
Multislice Spiral Computed Tomography
Contrast-enhanced CT angiography was performed with 16- and 64-slice CT scanner systems (Somatom Sensation Cardiac 16 or 64; both Siemens Medical Solutions; Forchheim; Germany). With 16-slice CT, a collimation of 0.75 mm was used with a gantry rotation time of 420 ms. Additional acquisition parameters were adjusted to individually optimize CT angiographies: pitch range 0.16 to 0.21 and range of tube current 500 to 700 mA. With 64-slice CT, 64 overlapping 0.6-mm slices were acquired per rotation with the use of a focal spot that periodically moved in the longitudinal direction (z-flying focal spot; 0.6-mm collimation and 330-ms gantry rotation time, range of pitch 0.18 to 0.2, and range of tube current 633 to 945 mA). The resulting volume CT dose index (CTDIvol) ranged between 17.6 and 54.6 Gy and between 18.8 and 66.4 Gy for 16- and 64-slice CT, respectively. Contrast dye (350 mg of iodine per milliliter) was injected intravenously at flow rates between 4.0 and 5.0 mL/s. Transaxial images were reconstructed with kernel B30f with a slice width of 1.0 and 0.6 mm for 16- and 64-slice CT, respectively.
Strategies for Reduction of Radiation Dose
An algorithm also referred to as dose pulsing, which modulates the tube current according to the ECG during the spiral scan, was used whenever possible. Cardiac motion is least during diastole and greatest during systole; thus, the image data are most likely sharpest during the diastolic phase. Accordingly, this algorithm reduces the tube current by 80% during systole, in which image reconstruction is not likely to be of interest. Reasons for not using this algorithm were arrhythmias recorded during the preparation time before CT angiography or the need for optimal image quality during the entire cardiac cycle in selected patients.
Another approach for reducing radiation dose is reduction of the tube voltage, because the radiation dose varies with the square of the tube voltage. A reduced voltage may also influence image quality, especially when iodinated contrast media are used. Owing to the high relative atomic number of iodine, the attenuation of iodinated contrast media increases with reduced x-ray energy,5 which results in a higher contrast between contrast mediafilled vascular structures and the perivascular tissue. However, the reduction of radiation dose also correlates with increased image noise and potentially with decreased image quality. Thus, use of the scanning protocol with reduced tube voltage and a compensation for the reduced voltage by increasing the tube current were at the discretion of the coronary CT angiographer.
Estimation of CT Dose
The effective dose of CT angiography was estimated by a method proposed by the European Working Group for Guidelines on Quality Criteria in CT.6 This method has been shown to be reasonably robust and consistent for estimating the effective dose. The effective dose is derived from the product of the dose-length product and a conversion coefficient for the chest as the investigated anatomic region. This conversion coefficient (k=0.017 mSv · mGy1 · cm1) is averaged between male and female models. The parameters for dose-length product and CTDIvol, which averages radiation dose in the x-, y- and z-directions, were obtained from the protocol that summarized the relevant individual radiation exposure parameters for each step of the coronary CT investigation.
Assessment of Image Quality
To obtain objective indices of image quality, we determined image noise and contrast- and signal-to-noise ratios for different scanning protocols. Image noise was derived from the standard deviation of the density values (in Hounsfield units) within a large region of interest in the left ventricle. The contrast-to-noise ratio was defined as the difference between the mean density of the contrast-filled left ventricular chamber and the mean density of the left ventricular wall, which was divided by image noise. The signal-to-noise ratios were assessed in the proximal segments of the left and right coronary artery. This parameter, which may be more relevant in the detection of lumen narrowings and of atherosclerotic plaque burden, was calculated from the mean density values of the contrasted coronary lumen divided by the standard deviation of these density values. With 16-slice CT, image-quality parameters were obtained from 1.0-mm slices, whereas the slice thickness was 0.6 mm for 64-slice CT angiographic images.
Two reviewers independently evaluated the coronary CT angiographies by assessment of axial slices and 3 thin-slab maximum intensity projections. Orientated along the heart axis, the thin-slab (5-mm thickness, 1-mm increment) maximum intensity projections were reconstructed perpendicular to each other. The coronary artery tree was segmented according to modified American Heart Association classification,7 and the diagnostic image quality was investigated for each segment with a vessel diameter of at least 2.0 mm by caliper measurements. The grading criteria for image quality were as follows: Good diagnostic quality was defined for coronary segments without or with moderate calcifications, without motion artifacts, and with good vessel opacification. Limited image quality, but still diagnostic for the evaluation of lumen stenosis, was defined in the presence of moderate motion artifacts or a reduced vessel opacification. Coronary segments with extensive calcifications, stent placement, extensive motion artifacts, or poor opacification were considered nondiagnostic.
The scan protocol with the use of 120-kilovolt peak (kVp) tube voltage and without application of the ECG-dependent dose modulation was defined as the standard scan protocol for both 16- and 64-slice CT scanners. All image-quality parameters obtained with the algorithms for dose savings were compared with the parameters of the standard scan protocol. Because of the differences in image acquisition between 16- and 64-slice CT systems, no comparisons between the 2 systems were performed.
Statistical Analysis
Results are expressed as counts (or proportions in percentages) or as mean±SD. For discrete variables, comparisons of the 2 groups were performed with the
2 test or Fishers exact test, if appropriate. Continuous variables were analyzed with an unpaired, 2-tailed t test. Statistical significance was accepted for 2-sided probability values <0.025 (the usual
-level of 0.05 corrected for 2 planned comparisons between the standard CT scanning protocol and the 2 investigated protocols for effective dose reduction according to the Bonferroni method).
The authors had full access to the data and take responsibility for its integrity. All authors have read and agree to the manuscript as written.
| Results |
|---|
|
|
|---|
|
Algorithms for Dose Savings
To study the impact of different scan protocols on radiation dose and image quality, a subgroup of patients were randomly selected from the patient cohort. The examinations in these patients represented typical CT angiographic investigations for the evaluation of coronary artery disease, eg, a scan length of &125 mm and a heart rate of &60 bpm. The scan protocols were as follows: 120-kV scans with and without ECG-dependent dose modulation and 100-kV scans with dose modulation, each with 16- and 64-slice CT systems, respectively (Table 2). For 2 scan protocols, only 30 patients were included because of the limited availability of patients studied with the respective protocol. The selected patient population is characterized as being predominantly male patients (68%) with a mean body mass index of 26.6±3.7 kg/m2, without differences between the groups. With 16-slice CT, the use of ECG-dependent dose modulation resulted in a significant reduction of the effective dose estimate from 10.6±1.2 to 6.4±0.9 mSv (P<0.001). Similarly, the dose estimate was reduced from 14.8±1.8 to 9.4±1.0 mSv with the same algorithm in 64-slice CT (P<0.001). Therefore, the efficacy of the ECG-dependent dose modulation in dose reduction was 40% and 37% with 16-and 64-slice CT systems, respectively. The use of the scanning protocol with a 100-kV tube voltage in addition to ECG-dependent dose modulation resulted in a further reduction of the dose estimates. An effective dose of 5.0±0.3 and of 5.4±1.1 mSv was estimated for the 100-kV protocol with 16- and 64-slice CT, respectively. When we compared these estimates with the 120-kV protocol without ECG-dependent dose modulation, the overall relative dose reductions ranged between 53% and 64% (P<0.001 for 16- and 64-slice CT; Figure, panel A).
|
|
Image Quality
Image quality was assessed quantitatively by image noise, contrast-to-noise ratio, and signal-to-noise ratio. Although image noise did not differ between CT angiographies performed with and without ECG-dependent dose modulation, the use of the 100-kV protocols resulted in significantly increased image noise with 16- and 64-slice CT (Figure, panel C). Signal-to-noise and contrast-to-noise ratios did not differ between scan protocols within each CT system; however, a trend to a better contrast-to-noise ratio was observed with ECG-dependent dose modulation and with the 100-kV protocols with 16- and 64-slice CT. Table 2 summarizes the patient characteristics, relevant MSCT results, and results for image quality and for the estimate of the effective radiation dose in more detail.
To further assess the impact of different scan protocols on image and diagnostic quality, a qualitative assessment of coronary segments was performed. A total of 3492 coronary segments were evaluated for size, good or limited diagnostic quality, or nondiagnostic quality. Of these, 694 (19.9%) coronary segments presented with a small caliper and were not evaluated for diagnostic image quality. For the remaining segments, no deterioration of diagnostic image quality was observed with utilization of ECG-dependent dose modulation and of the 100-kV protocols with 16- and 64-slice CT (Figure, panel D; Table 3). The predominant reasons for nondiagnostic image quality were extensive motion artifacts or extensive coronary calcifications.
|
| Discussion |
|---|
|
|
|---|
The present study demonstrates that the effective dose for cardiac MSCT scanning in daily practice is estimated to be 6.4±1.9 and 11.0±4.1 mSv with 16- and 64-slice CT, respectively. The increase in dose estimates with the 64-slice technology is explained by the higher spatial and temporal resolution achieved with current CT scanner technology. The high spatial resolution, which is a prerequisite for accurate assessment of coronary artery narrowings and atherosclerotic plaque burden, increased from 16- to 64-slice CT owing to a decrease in collimation and the respective slice width (collimation 0.75 and 0.6 mm with 16- and 64-slice CT, respectively). To maintain a comparable image quality with the reduced slice width, the number of photons received by the detector array is usually increased by an escalation in the tube current. Such an escalation in tube current also results in increases in radiation exposure and in effective dose estimates. Furthermore, an increase in temporal resolution is achieved with newer 64-slice CT scanners by an increase in gantry rotation speed. To avoid a decrease in image quality, the rate of photons received by the detector array must also be increased, and therefore, the tube current is further increased, which results in a higher effective dose estimate.13 When one compares 16- and 64-slice CT dose estimates, another factor needs to be taken into account. With the increased number of simultaneously acquired slices, the scanning time of 64-slice CT is reduced, which allows for an increase in the scan length acquired with a single short breath hold. This development is also reflected in the increased number of patients who were scanned with longer scan ranges for the evaluation of bypass graft patency in the present study with 64-slice CT.
In the second part of this study, we investigated the impact of 2 different algorithms for reduction of the estimated radiation dose with both scanner types. The first algorithm used allows for ECG-dependent modulation of the tube current. Image reconstruction of cardiac MSCT examinations is usually performed in ventricular mid diastole, because cardiac motion and thus blurring of cardiac structures due to motion artifacts is least during that period of the cardiac cycle. Projection data that do not correspond to the desired time window of mid diastole are usually ignored by this postprocessing technique, called "retrospective gating." Therefore, the algorithm of ECG-dependent dose modulation allows for a high tube current with optimal image quality during mid diastole, while reducing the tube current by 80% in the remainder of the cardiac cycle. Because the onset of tube current reduction is based on the average of the 3 prior RR intervals, the algorithm may inappropriately reduce the current in the event of arrhythmias such as premature extra beats. Therefore, this algorithm may not be useful in patients with arrhythmias or in patients in whom the best image quality is needed throughout the entire cardiac cycle. However, such patients constitute a minority of patients studied currently by MSCT. In fact, the present study demonstrates that this algorithm had been used in 82.2% of patients investigated in our cardioradiologic MSCT laboratory. Although we cannot rule out that unexpectedly occurring premature extra beats deteriorated diagnostic image quality when the ECG-dependent dose algorithm was used, the likelihood of such arrhythmias occurring is very low, and the investigated parameters for image quality did not differ between patients in whom this algorithm was used or not used (Table 2). When we compared patients with typical coronary MSCT investigations, the use of this algorithm resulted in a significant reduction of the radiation dose estimate of &40% and 37% with 16- and 64-slice CT, respectively. Therefore, the algorithm for ECG-dependent dose modulation represents a very effective tool for limiting radiation dose in the vast majority of patients undergoing cardiac MSCT studies.
Recently, it has been established that lower tube voltages may allow for a reduction in effective dose in pediatric CT investigations without affecting diagnostic image quality.14 The decrease in radiation dose is achieved because the radiation dose varies with the square of the kilovoltage. Furthermore, the decreased tube voltage leads to increased opacification of vascular structures during contrast-enhanced CT angiography owing to an increase in the photoelectric effect and a decrease in Compton scattering.15 Therefore, we investigated in a further step the impact of a lower tube voltage on the effective dose estimates and image quality during cardiac 16- and 64-slice CT scanning. Because of a later availability of this scan protocol on the 64-slice CT system, the 100-kV tube voltage scan protocol was used less frequently in the present patient cohort. Compared with cardiac CT studies performed with 120 kV and without ECG-dependent dose modulation, the use of the reduced voltage of 100 kV together with ECG-dependent dose modulation resulted in the present study in an effective dose reduction of 53% and 64% with 16- and 64-slice CT, respectively. The small difference in relative dose reduction between 16- and 64-slice CT is mainly based on small variations in other scan parameters and in studied patient groups, including differences in pitch, scan length, and heart rate. As a consequence of the above-mentioned physical aspects, the signal improved with the lower tube voltage, but the image noise also increased significantly. With the increase of both parameters, the ratios for signal-to-noise and contrast-to-noise did not differ between scan protocols within each CT system. For noncardiac CT studies with kilovoltage reduction, an increase of the tube current by 50% has been proposed to maintain image quality and to reduce the dose estimates at the same time16; however, a further increase in tube current is limited with the available standard protocols for cardiac CT scanning on the studied CT scanners. Therefore, a tradeoff between dose saving and increased image noise has to be made with current cardiac CT protocols.
The best balance between optimal dose savings and maintained diagnostic, rather than esthetic, image quality for coronary CT angiographic investigations is difficult to assess; the present study compared the frequency of coronary segments with good and limited diagnostic image quality for the detection of stenosis between different scan protocols. In this qualitative analysis of image quality, no deterioration of diagnostic image quality was detected for scan protocols with the ECG-dependent dose modulation and with the 100-kV protocols with 16- and 64-slice CT. However, the value of this analysis is limited by a potential selection bias of used scan protocols, eg, the use of a 120-kV scan protocol without ECG modulation for patients with arrhythmias, who are likely to present with more nondiagnostic coronary segments, even when no dose-saving algorithms were applied. Finally, the impact of dose-saving algorithms on the detection of calcified and noncalcified plaques is unknown. Therefore, further studies are needed to investigate the balance between dose savings and maintained diagnostic image quality for coronary CT angiographic investigations.
Study Limitations
This study was not designed to compare diagnostic image quality and the assessibility of coronary segments between the 2 investigated CT systems because of the inherent differences in image acquisition. However, in the present study, a higher percentage of coronary segments were assessable by 64-slice CT owing to its higher spatial and temporal resolution. This finding may imply that the improvement in resolution is more important for the assessibility and discernment of coronary segments and stenoses than the observed differences in quantitative image-quality parameters. Therefore, further studies are needed to investigate the impact of temporal and spatial resolution, as well as of image-quality parameters, on the sensitivity and specificity of coronary stenosis detection by CT angiography to justify the increase in radiation dose when 64-slice CT is used.
In summary, it is important to be aware of the magnitude of the radiation dose received from coronary CT scanning and to realize that a careful selection of CT scanning protocols is needed to keep the radiation exposure "as low as reasonably achievable (ALARA)". Furthermore, cardiac CT angiographies should be performed with new scanning techniques that allow for a reduction in effective dose estimates, such as ECG-dependent dose modulation and reduced tube voltage, whenever the use of these techniques is possible without a deterioration of diagnostic image quality. Finally, future studies are needed to identify patients who truly benefit from coronary CT angiography to justify the potential hazards of radiation exposure.
| Acknowledgments |
|---|
Disclosures
None.
| References |
|---|
|
|
|---|
2. Achenbach S, Moselewski F, Ropers D, Ferencik M, Hoffmann U, MacNeill B, Pohle K, Baum U, Anders K, Jang IK, Daniel WG, Brady TJ. Detection of calcified and noncalcified coronary atherosclerotic plaque by contrast-enhanced, submillimeter multidetector spiral computed tomography: a segment-based comparison with intravascular ultrasound. Circulation. 2004; 109: 1417.
3. Leber AW, Knez A, von Ziegler F, Becker A, Nikolaou K, Paul S, Wintersperger B, Reiser M, Becker CR, Steinbeck G, Boekstegers P. Quantification of obstructive and nonobstructive coronary lesions by 64-slice computed tomography: a comparative study with quantitative coronary angiography and intravascular ultrasound. J Am Coll Cardiol. 2005; 46: 147154.
4. Morin RL, Gerber TC, McCollough CH. Radiation dose in computed tomography of the heart. Circulation. 2003; 107: 917922.
5. Huda W, Scalzetti EM, Levin G. Technique factors and image quality as functions of patient weight at abdominal CT. Radiology. 2000; 217: 430435.
6. Menzel HG, Schibilla H, Teunen D, eds. European guidelines on quality criteria for computed tomography. Luxembourg: European Commission; 2000. Publication No. EUR 16262 EN.
7. Austen WG, Edwards JE, Frye RL, Gensini GG, Gott VL, Griffith LS, McGoon DC, Murphy ML, Roe BB. A reporting system on patients evaluated for coronary artery disease: report of the Ad Hoc Committee for Grading of Coronary Artery Disease, Council on Cardiovascular Surgery, American Heart Association. Circulation. 1975; 51: 540.[Medline] [Order article via Infotrieve]
8. Mollet NR, Cademartiri F, Krestin GP, McFadden EP, Arampatzis CA, Serruys PW, de Feyter PJ. Improved diagnostic accuracy with 16-row multi-slice computed tomography coronary angiography. J Am Coll Cardiol. 2005; 45: 128132.
9. Lacomis JM, Wigginton W, Fuhrman C, Schwartzman D, Armfield DR, Pealer KM. Multi-detector row CT of the left atrium and pulmonary veins before radio-frequency catheter ablation for atrial fibrillation. Radiographics. 2003; 23: S35S48.
10. Schwartzman D, Lacomis J, Wigginton WG. Characterization of left atrium and distal pulmonary vein morphology using multidimensional computed tomography. J Am Coll Cardiol. 2003; 41: 13491357.
11. Ropers D, Ulzheimer S, Wenkel E, Baum U, Giesler T, Derlien H, Moshage W, Bautz WA, Daniel WG, Kalender WA, Achenbach S. Investigation of aortocoronary artery bypass grafts by multislice spiral computed tomography with electrocardiographic-gated image reconstruction. Am J Cardiol. 2001; 88: 792795.[CrossRef][Medline] [Order article via Infotrieve]
12. Nieman K, Cademartiri F, Raaijmakers R, Pattynama P, de Feyter P. Noninvasive angiographic evaluation of coronary stents with multi-slice spiral computed tomography. Herz. 2003; 28: 136142.[CrossRef][Medline] [Order article via Infotrieve]
13. McCollough CH, Zink FE. Performance evaluation of a multi-slice CT system. Med Phys. 1999; 26: 22232230.[CrossRef][Medline] [Order article via Infotrieve]
14. Suess C, Chen X. Dose optimization in pediatric CT: current technology and future innovations. Pediatr Radiol. 2002; 32: 729734.[CrossRef][Medline] [Order article via Infotrieve]
15. Ertl-Wagner BB, Hoffmann RT, Bruning R, Herrmann K, Snyder B, Blume JD, Reiser MF. Multi-detector row CT angiography of the brain at various kilovoltage settings. Radiology. 2004; 231: 528535.
16. Sigal-Cinqualbre AB, Hennequin R, Abada HT, Chen X, Paul JF. Low-kilovoltage multi-detector row chest CT in adults: feasibility and effect on image quality and iodine dose. Radiology. 2004; 231: 169174.
![]() |
J. M. van Werkhoven, J. D. Schuijf, O. Gaemperli, J. W. Jukema, L. J. Kroft, E. Boersma, A. Pazhenkottil, I. Valenta, G. Pundziute, A. de Roos, et al. Incremental prognostic value of multi-slice computed tomography coronary angiography over coronary artery calcium scoring in patients with suspected coronary artery disease Eur. Heart J., November 1, 2009; 30(21): 2622 - 2629. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Baumuller, S. Leschka, L. Desbiolles, P. Stolzmann, H. Scheffel, B. Seifert, B. Marincek, and H. Alkadhi Dual-Source versus 64-Section CT Coronary Angiography at Lower Heart Rates: Comparison of Accuracy and Radiation Dose Radiology, October 1, 2009; 253(1): 56 - 64. [Abstract] [Full Text] [PDF] |
||||
![]() |
J M van Werkhoven, O Gaemperli, J D Schuijf, J W Jukema, L J Kroft, S Leschka, H Alkadhi, I Valenta, G Pundziute, A de Roos, et al. Multislice computed tomography coronary angiography for risk stratification in patients with an intermediate pretest likelihood Heart, October 1, 2009; 95(19): 1607 - 1611. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. A. Rodriguez-Granillo, M. A. Rosales, S. Baum, P. Rennes, C. Rodriguez-Pagani, V. Curotto, C. Fernandez-Pereira, C. Llaurado, G. Risau, E. Degrossi, et al. Early Assessment of Myocardial Viability by the Use of Delayed Enhancement Computed Tomography After Primary Percutaneous Coronary Intervention J. Am. Coll. Cardiol. Img., September 1, 2009; 2(9): 1072 - 1081. [Abstract] [Full Text] [PDF] |
||||
![]() |
J Rixe, G Conradi, A Rolf, A Schmermund, A Magedanz, D Erkapic, A Deetjen, C W Hamm, and T Dill Radiation dose exposure of computed tomography coronary angiography: comparison of dual-source, 16-slice and 64-slice CT Heart, August 15, 2009; 95(16): 1337 - 1342. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Bischoff, F. Hein, T. Meyer, M. Hadamitzky, S. Martinoff, A. Schomig, and J. Hausleiter Impact of a Reduced Tube Voltage on CT Angiography and Radiation Dose: Results of the PROTECTION I Study J. Am. Coll. Cardiol. Img., August 1, 2009; 2(8): 940 - 946. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Fujioka, J. Horiguchi, M. Kiguchi, H. Yamamoto, T. Kitagawa, and K. Ito Survey of Aorta and Coronary Arteries With Prospective ECG-Triggered 100-kV 64-MDCT Angiography Am. J. Roentgenol., July 1, 2009; 193(1): 227 - 233. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Bettencourt, J. Rocha, M. Carvalho, D. Leite, A. M. Toschke, B. Melica, L. Santos, A. Rodrigues, M. Goncalves, P. Braga, et al. Multislice Computed Tomography in the Exclusion of Coronary Artery Disease in Patients With Presurgical Valve Disease Circ Cardiovasc Imaging, July 1, 2009; 2(4): 306 - 313. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. C. Weustink, N. R. Mollet, L. A. Neefjes, M. van Straten, E. Neoh, S. Kyrzopoulos, B. W. Meijboom, C. van Mieghem, F. Cademartiri, P. J. de Feyter, et al. Preserved Diagnostic Performance of Dual-Source CT Coronary Angiography with Reduced Radiation Exposure and Cancer Risk Radiology, July 1, 2009; 252(1): 53 - 60. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. L. Raff, K. M. Chinnaiyan, D. A. Share, T. Y. Goraya, E. A. Kazerooni, M. Moscucci, R. E. Gentry, A. Abidov, and for the Advanced Cardiovascular Imaging Consortium Radiation Dose From Cardiac Computed Tomography Before and After Implementation of Radiation Dose-Reduction Techniques JAMA, June 10, 2009; 301(22): 2340 - 2348. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. M. Takakuwa, E. J. Halpern, E. L. Gingold, D. C. Levin, and F. S. Shofer Radiation Dose in a "Triple Rule-Out" Coronary CT Angiography Protocol of Emergency Department Patients Using 64-MDCT: The Impact of ECG-Based Tube Current Modulation on Age, Sex, and Body Mass Index Am. J. Roentgenol., April 1, 2009; 192(4): 866 - 872. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Pflederer, L. Rudofsky, D. Ropers, S. Bachmann, M. Marwan, W. G. Daniel, and S. Achenbach Image Quality in a Low Radiation Exposure Protocol for Retrospectively ECG-Gated Coronary CT Angiography Am. J. Roentgenol., April 1, 2009; 192(4): 1045 - 1050. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Husmann, B. A. Herzog, O. Gaemperli, F. Tatsugami, N. Burkhard, I. Valenta, P. Veit-Haibach, C. A. Wyss, U. Landmesser, and P. A. Kaufmann Diagnostic accuracy of computed tomography coronary angiography and evaluation of stress-only single-photon emission computed tomography/computed tomography hybrid imaging: comparison of prospective electrocardiogram-triggering vs. retrospective gating Eur. Heart J., March 1, 2009; 30(5): 600 - 607. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Tatsugami, L. Husmann, B. A. Herzog, N. Burkhard, I. Valenta, O. Gaemperli, and P. A. Kaufmann Evaluation of a Body Mass Index-Adapted Protocol for Low-Dose 64-MDCT Coronary Angiography with Prospective ECG Triggering Am. J. Roentgenol., March 1, 2009; 192(3): 635 - 638. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. R. Mayo and J. A. Leipsic Radiation Dose in Cardiac CT Am. J. Roentgenol., March 1, 2009; 192(3): 646 - 653. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. M. van Werkhoven, J. D. Schuijf, O. Gaemperli, J. W. Jukema, E. Boersma, W. Wijns, P. Stolzmann, H. Alkadhi, I. Valenta, M. P.M. Stokkel, et al. Prognostic value of multislice computed tomography and gated single-photon emission computed tomography in patients with suspected coronary artery disease. J. Am. Coll. Cardiol., February 17, 2009; 53(7): 623 - 632. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Hausleiter, T. Meyer, F. Hermann, M. Hadamitzky, M. Krebs, T. C. Gerber, C. McCollough, S. Martinoff, A. Kastrati, A. Schomig, et al. Estimated Radiation Dose Associated With Cardiac CT Angiography JAMA, February 4, 2009; 301(5): 500 - 507. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. J. Einstein Radiation Protection of Patients Undergoing Cardiac Computed Tomographic Angiography JAMA, February 4, 2009; 301(5): 545 - 547. [Full Text] [PDF] |
||||
![]() |
G. M. Feuchtner, P. Stolzmann, W. Dichtl, T. Schertler, J. Bonatti, H. Scheffel, S. Mueller, A. Plass, L. Mueller, T. Bartel, et al. Multislice computed tomography in infective endocarditis comparison with transesophageal echocardiography and intraoperative findings. J. Am. Coll. Cardiol., February 3, 2009; 53(5): 436 - 444. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. A. Herzog, L. Husmann, N. Burkhard, O. Gaemperli, I. Valenta, F. Tatsugami, C. A. Wyss, U. Landmesser, and P. A. Kaufmann Accuracy of low-dose computed tomography coronary angiography using prospective electrocardiogram-triggering: first clinical experience Eur. Heart J., December 2, 2008; 29(24): 3037 - 3042. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Schmermund, B. Nowak, and T. Voigtlander Non-invasive CT coronary angiography: can high diagnostic image quality be achieved with less radiation exposure? Eur. Heart J., December 2, 2008; 29(24): 2955 - 2956. [Full Text] [PDF] |
||||
![]() |
A. J Einstein Radiation risk from coronary artery disease imaging: how do different diagnostic tests compare? Heart, December 1, 2008; 94(12): 1519 - 1521. [Full Text] [PDF] |
||||
![]() |
G Mowatt, J A Cook, G S Hillis, S Walker, C Fraser, X Jia, and N Waugh 64-Slice computed tomography angiography in the diagnosis and assessment of coronary artery disease: systematic review and meta-analysis Heart, November 1, 2008; 94(11): 1386 - 1393. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Stolzmann, S. Leschka, H. Scheffel, T. Krauss, L. Desbiolles, A. Plass, M. Genoni, T. G. Flohr, S. Wildermuth, B. Marincek, et al. Dual-Source CT in Step-and-Shoot Mode: Noninvasive Coronary Angiography with Low Radiation Dose Radiology, October 1, 2008; 249(1): 71 - 80. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. C. Weustink, N. R. Mollet, F. Pugliese, W. B. Meijboom, K. Nieman, M. H. Heijenbrok-Kal, T. G. Flohr, L. A. E. Neefjes, F. Cademartiri, P. J. de Feyter, et al. Optimal Electrocardiographic Pulsing Windows and Heart Rate: Effect on Image Quality and Radiation Exposure at Dual-Source Coronary CT Angiography Radiology, September 1, 2008; 248(3): 792 - 798. [Abstract] [Full Text] [PDF] |
||||
![]() |
H Scheffel, H Alkadhi, S Leschka, A Plass, L Desbiolles, I Guber, T Krauss, J Gruenenfelder, M Genoni, T F Luescher, et al. Low-dose CT coronary angiography in the step-and-shoot mode: diagnostic performance Heart, September 1, 2008; 94(9): 1132 - 1137. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. F Tops and M. J Schalij Multislice CT: is it essential before atrial fibrillation ablation? Heart, August 1, 2008; 94(8): 973 - 975. [Full Text] [PDF] |
||||
![]() |
W. P. Shuman, K. R. Branch, J. M. May, L. M. Mitsumori, D. W. Lockhart, T. J. Dubinsky, B. H. Warren, and J. H. Caldwell Prospective versus Retrospective ECG Gating for 64-Detector CT of the Coronary Arteries: Comparison of Image Quality and Patient Radiation Dose Radiology, August 1, 2008; 248(2): 431 - 437. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Hirai, J. Horiguchi, C. Fujioka, M. Kiguchi, H. Yamamoto, N. Matsuura, T. Kitagawa, H. Teragawa, N. Kohno, and K. Ito Prospective versus Retrospective ECG-gated 64-Detector Coronary CT Angiography: Assessment of Image Quality, Stenosis, and Radiation Dose Radiology, August 1, 2008; 248(2): 424 - 430. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. A. Bluemke, S. Achenbach, M. Budoff, T. C. Gerber, B. Gersh, L. D. Hillis, W. G. Hundley, W. J. Manning, B. F. Printz, M. Stuber, et al. Noninvasive Coronary Artery Imaging: Magnetic Resonance Angiography and Multidetector Computed Tomography Angiography: A Scientific Statement From the American Heart Association Committee on Cardiovascular Imaging and Intervention of the Council on Cardiovascular Radiology and Intervention, and the Councils on Clinical Cardiology and Cardiovascular Disease in the Young Circulation, July 29, 2008; 118(5): 586 - 606. [Full Text] [PDF] |
||||
![]() |
G. M. Feuchtner, W. Dichtl, S. Muller, D. Jodocy, T. Schachner, A. Klauser, and J. O. Bonatti 64-MDCT for Diagnosis of Aortic Regurgitation in Patients Referred to CT Coronary Angiography Am. J. Roentgenol., July 1, 2008; 191(1): W1 - W7. [Abstract] [Full Text] [PDF] |
||||
![]() |
W T Roberts, J J Bax, and L C Davies Cardiac CT and CT coronary angiography: technology and application Heart, June 1, 2008; 94(6): 781 - 792. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. M. Boxt CT angiography in children: it is accurate, but is it safe? J. Am. Coll. Cardiol. Img., May 1, 2008; 1(3): 340 - 342. [Full Text] [PDF] |
||||
![]() |
K. L. Gould, T. Pan, C. Loghin, N. P. Johnson, and S. Sdringola Reducing Radiation Dose in Rest-Stress Cardiac PET/CT by Single Poststress Cine CT for Attenuation Correction: Quantitative Validation J. Nucl. Med., May 1, 2008; 49(5): 738 - 745. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. M. Kramer All High-Risk Patients Should Not Be Screened With Computed Tomographic Angiography Circulation, March 11, 2008; 117(10): 1333 - 1339. [Full Text] [PDF] |
||||
![]() |
S. Achenbach, U. Ropers, A. Kuettner, K. Anders, T. Pflederer, S. Komatsu, W. Bautz, W. G. Daniel, and D. Ropers Randomized comparison of 64-slice single- and dual-source computed tomography coronary angiography for the detection of coronary artery disease. J. Am. Coll. Cardiol. Img., March 1, 2008; 1(2): 177 - 186. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. P. Earls, E. L. Berman, B. A. Urban, C. A. Curry, J. L. Lane, R. S. Jennings, C. C. McCulloch, J. Hsieh, and J. H. Londt Prospectively Gated Transverse Coronary CT Angiography versus Retrospectively Gated Helical Technique: Improved Image Quality and Reduced Radiation Dose Radiology, March 1, 2008; 246(3): 742 - 753. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Leschka, P. Koepfli, L. Husmann, A. Plass, R. Vachenauer, O. Gaemperli, T. Schepis, M. Genoni, B. Marincek, F. R. Eberli, et al. Myocardial Bridging: Depiction Rate and Morphology at CT Coronary Angiography--Comparison with Conventional Coronary Angiography Radiology, March 1, 2008; 246(3): 754 - 762. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Schroeder, S. Achenbach, F. Bengel, C. Burgstahler, F. Cademartiri, P. de Feyter, R. George, P. Kaufmann, A. F. Kopp, J. Knuuti, et al. Cardiac computed tomography: indications, applications, limitations, and training requirements: Report of a Writing Group deployed by the Working Group Nuclear Cardiology and Cardiac CT of the European Society of Cardiology and the European Council of Nuclear Cardiology Eur. Heart J., February 2, 2008; 29(4): 531 - 556. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Horiguchi, M. Kiguchi, C. Fujioka, Y. Shen, R. Arie, K. Sunasaka, and K. Ito Radiation Dose, Image Quality, Stenosis Measurement, and CT Densitometry Using ECG-Triggered Coronary 64-MDCT Angiography: A Phantom Study Am. J. Roentgenol., February 1, 2008; 190(2): 315 - 320. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Pugliese, N. R. Mollet, M. G. M. Hunink, F. Cademartiri, K. Nieman, R. T. van Domburg, W. B. Meijboom, C. Van Mieghem, A. C. Weustink, M. L. Dijkshoorn, et al. Diagnostic Performance of Coronary CT Angiography by Using Different Generations of Multisection Scanners: Single-Center Experience Radiology, February 1, 2008; 246(2): 384 - 393. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Husmann, I. Valenta, O. Gaemperli, O. Adda, V. Treyer, C. A. Wyss, P. Veit-Haibach, F. Tatsugami, G. K. von Schulthess, and P. A. Kaufmann Feasibility of low-dose coronary CT angiography: first experience with prospective ECG-gating Eur. Heart J., January 2, 2008; 29(2): 191 - 197. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Schoenhagen Back to the future: coronary CT angiography using prospective ECG triggering Eur. Heart J., January 2, 2008; 29(2): 153 - 154. [Full Text] [PDF] |
||||
![]() |
J. Hausleiter, T. Meyer, M. Hadamitzky, M. Zankl, P. Gerein, K. Dorrler, A. Kastrati, S. Martinoff, and A. Schomig Non-invasive coronary computed tomographic angiography for patients with suspected coronary artery disease: the Coronary Angiography by Computed Tomography with the Use of a Submillimeter resolution (CACTUS) trial Eur. Heart J., December 2, 2007; 28(24): 3034 - 3041. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Abdulla, S. Z. Abildstrom, O. Gotzsche, E. Christensen, L. Kober, and C. Torp-Pedersen 64-multislice detector computed tomography coronary angiography as potential alternative to conventional coronary angiography: a systematic review and meta-analysis Eur. Heart J., December 2, 2007; 28(24): 3042 - 3050. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. M. Hurwitz, R. E. Reiman, T. T. Yoshizumi, P. C. Goodman, G. Toncheva, G. Nguyen, and C. Lowry Radiation Dose from Contemporary Cardiothoracic Multidetector CT Protocols with an Anthropomorphic Female Phantom: Implications for Cancer Induction Radiology, December 1, 2007; 245(3): 742 - 750. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Hamon, R. Morello, J. W. Riddell, and M. Hamon Coronary Arteries: Diagnostic Performance of 16- versus 64-Section Spiral CT Compared with Invasive Coronary Angiography Meta-Analysis Radiology, December 1, 2007; 245(3): 720 - 731. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. B Meijboom, N. R Mollet, C. A Van Mieghem, A. C Weustink, F. Pugliese, N. van Pelt, F. Cademartiri, E. Vourvouri, P. de Jaegere, G. P Krestin, et al. 64-Slice CT coronary angiography in patients with non-ST elevation acute coronary syndrome Heart, November 1, 2007; 93(11): 1386 - 1392. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. B. Meijboom, C. A.G. van Mieghem, N. R. Mollet, F. Pugliese, A. C. Weustink, N. van Pelt, F. Cademartiri, K. Nieman, E. Boersma, P. de Jaegere, et al. 64-Slice Computed Tomography Coronary Angiography in Patients With High, Intermediate, or Low Pretest Probability of Significant Coronary Artery Disease J. Am. Coll. Cardiol., October 9, 2007; 50(15): 1469 - 1475. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. J. Einstein, K. W. Moser, R. C. Thompson, M. D. Cerqueira, and M. J. Henzlova Radiation Dose to Patients From Cardiac Diagnostic Imaging Circulation, September 11, 2007; 116(11): 1290 - 1305. [Full Text] [PDF] |
||||
![]() |
G. M. Feuchtner, T. Schachner, J. Bonatti, G. J. Friedrich, P. Soegner, A. Klauser, and D. z. Nedden Diagnostic Performance of 64-Slice Computed Tomography in Evaluation of Coronary Artery Bypass Grafts Am. J. Roentgenol., September 1, 2007; 189(3): 574 - 580. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Malagutti, K. Nieman, W. B. Meijboom, C. A.G. van Mieghem, F. Pugliese, F. Cademartiri, N. R. Mollet, E. Boersma, P. P. de Jaegere, and P. J. de Feyter Use of 64-slice CT in symptomatic patients after coronary bypass surgery: evaluation of grafts and coronary arteries Eur. Heart J., August 1, 2007; 28(15): 1879 - 1885. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. J. Einstein, M. J. Henzlova, and S. Rajagopalan Estimating Risk of Cancer Associated With Radiation Exposure From 64-Slice Computed Tomography Coronary Angiography JAMA, July 18, 2007; 298(3): 317 - 323. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Ector, O. Dragusin, B. Adriaenssens, W. Huybrechts, R. Willems, H. Ector, and H. Heidbuchel Obesity Is a Major Determinant of Radiation Dose in Patients Undergoing Pulmonary Vein Isolation for Atrial Fibrillation J. Am. Coll. Cardiol., July 17, 2007; 50(3): 234 - 242. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. M. Feuchtner, S. Muller, J. Bonatti, T. Schachner, C. Velik-Salchner, O. Pachinger, and W. Dichtl Sixty-Four Slice CT Evaluation of Aortic Stenosis Using Planimetry of the Aortic Valve Area Am. J. Roentgenol., July 1, 2007; 189(1): 197 - 203. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Cademartiri, J. D. Schuijf, F. Pugliese, N. R. Mollet, J. W. Jukema, E. Maffei, L. J. Kroft, A. Palumbo, D. Ardissino, P. W. Serruys, et al. Usefulness of 64-Slice Multislice Computed Tomography Coronary Angiography to Assess In-Stent Restenosis J. Am. Coll. Cardiol., June 5, 2007; 49(22): 2204 - 2210. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Kumar, K. G. Bis, A. Shetty, A. Vyas, A. Anderson, M. Balasubramaniam, W. O'Neill, and W. Stein Aortic Root Catheter-Directed Coronary CT Angiography in Swine: Coronary Enhancement with Minimum Volume of Iodinated Contrast Material Am. J. Roentgenol., May 1, 2007; 188(5): W415 - W422. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Leschka, E. Oechslin, L. Husmann, L. Desbiolles, B. Marincek, M. Genoni, R. Pretre, R. Jenni, S. Wildermuth, and H. Alkadhi Pre- and Postoperative Evaluation of Congenital Heart Disease in Children and Adults with 64-Section CT RadioGraphics, May 1, 2007; 27(3): 829 - 846. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Ropers, F.-K. Pohle, A. Kuettner, T. Pflederer, K. Anders, W. G. Daniel, W. Bautz, U. Baum, and S. Achenbach Diagnostic Accuracy of Noninvasive Coronary Angiography in Patients After Bypass Surgery Using 64-Slice Spiral Computed Tomography With 330-ms Gantry Rotation Circulation, November 28, 2006; 114(22): 2334 - 2341. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Ghostine, C. Caussin, B. Daoud, M. Habis, E. Perrier, D. Pesenti-Rossi, A. Sigal-Cinqualbre, C.-Y. Angel, B. Lancelin, A. Capderou, et al. Non-Invasive Detection of Coronary Artery Disease in Patients With Left Bundle Branch Block Using 64-Slice Computed Tomography J. Am. Coll. Cardiol., November 21, 2006; 48(10): 1929 - 1934. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Achenbach Computed Tomography Coronary Angiography J. Am. Coll. Cardiol., November 21, 2006; 48(10): 1919 - 1928. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Vogel-Claussen, H. Pannu, P. J. Spevak, E. K. Fishman, and D. A. Bluemke Cardiac Valve Assessment with MR Imaging and 64-Section Multi-Detector Row CT RadioGraphics, November 1, 2006; 26(6): 1769 - 1784. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. B. Meijboom, N. R. Mollet, C. A.G. Van Mieghem, J. Kluin, A. C. Weustink, F. Pugliese, E. Vourvouri, F. Cademartiri, A. J.J.C. Bogers, G. P. Krestin, et al. Pre-Operative Computed Tomography Coronary Angiography to Detect Significant Coronary Artery Disease in Patients Referred for Cardiac Valve Surgery J. Am. Coll. Cardiol., October 17, 2006; 48(8): 1658 - 1665. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. Garcia, J. Lessick, M. H. K. Hoffmann, and for the CATSCAN Study Investigators Accuracy of 16-row multidetector computed tomography for the assessment of coronary artery stenosis. JAMA, July 26, 2006; 296(4): 403 - 411. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2006 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |