(Circulation. 2005;112:3009-3017.)
© 2005 American Heart Association, Inc.
Contemporary Reviews in Cardiovascular Medicine |
From the Cardiovascular Branch, Division of Intramural Research, National Heart, Lung, and Blood Institute, National Institutes of Health, Bethesda, Md.
Correspondence to Robert J. Lederman, MD, Cardiovascular Branch, Clinical Research Program, Division of Intramural Research, National Heart, Lung, and Blood Institute, National Institutes of Health, Building 10, Room 2c713, MSC 1538, Bethesda, MD 20892-1538, USA. E-mail Lederman{at}nih.gov
Key Words: imaging magnetic resonance imaging catheterization peripheral vascular disease
| Introduction |
|---|
|
|
|---|
Clinical investigational procedures have begun at several centers. The chief limitation to clinical translation, at present, appears to be the availability of clinical-grade catheter devices. This brief review will survey interventional cardiovascular imaging, treatment, and patient handling considerations; unique iCMR catheter design requirements; proof-of-concept animal and clinical experiments conducted to date; and novel applications we can expect in the near future.
| How MRI Creates Images for Intervention |
|---|
|
|
|---|
iCMR is now possible because technical advances permit rapid imaging on clinical scanners. These include highly uniform magnetic fields, rapidly changeable magnetic field gradients, multichannel receivers, and advanced computing systems.5,6 Newer systems have permitted clinical implementation of rapid and efficient pulse sequences such as steady-state free precession (SSFP; also known commercially as TrueFISP, FIESTA, or Balanced FFE) that provide excellent contrast between blood and myocardium. Imaging speed is limited by the physiological limits of stimulating patients peripheral nerves (from rapidly changing magnetic fields) and of heating (from prolonged RF excitation).7,8 These problems are more pronounced at higher field strengths. Faster pictures come at the expense of reduced spatial resolution. For comparison, x-rayguided interventions are currently performed using 512x512- or 1024x1024-pixel matrices; most rtMRI interventional procedures use 256x192 or smaller pixel matrices. However, the information content of these images is comparable to or higher than x-ray because of tissue and blood visualization (Figure 1).
|
Configuring a Suite for Clinical Interventional MRI
Investigational iCMR may require adjunctive or emergency bailout x-ray fluoroscopy (XRF); the two are combined into so-called XMR suites. For example, the NHLBI intervention suite9 consists of independently operating MRI and XRF systems (Sonata 1.5T and Axiom Artis FC, Siemens, Figure 2) separated by RF-shielded sliding doors. A motorized transport system (Miyabi, Siemens) permits rapid, smooth intermodality patient transport. In-lab image display can employ either shielded LCD screens or projectors. To attenuate acoustic noise and permit open-microphone communication among the operator, staff and patient, we use directional optical microphones (Phone-Or) attached to RF-filtered headsets (Magnacoustics) using two communications channels. Modifications to the way magnetic gradients are changed10 and newer, quieter MR systems may ameliorate the acoustic noise problem.
|
iCMR poses new challenges to patient handling, monitoring, and safety.7,8 We use custom multiple-lead RF-shielded ECG cables (In Vivo Research) because the rapidly switching magnetic gradients and RF pulses used to create images can heat standard cables, causing skin burns, or can induce signal artifacts resembling arrhythmias. Moreover, cardiac and aortic blood flow within the main magnetic field irrevocably distorts ECG ST segments, so at present ECG is nondiagnostic and is used only for identifying heart rate and periodicity. Other systems use vectorcardiographic signals to optimize cardiac rhythm detection to gate image acquisition and to monitor patients. Inside the interventional suite, all ferromagnetic equipment (ie, infusion pumps, ventilator) should be substituted with MRI-compatible equipment. A pathway must be available to evacuate the patient rapidly from the high magnetic field during emergencies. For example, discharging cardiac defibrillation paddles within the high magnetic field can easily knock down emergency caregivers, and the defibrillator or other resuscitation equipment moved in proximity to the MRI bore can become missiles.
Sedated or otherwise noncommunicative patients (ie, during general anesthesia) also need to be protected from heating during prolonged rapid MRI examinations, either by monitoring predicted specific absorption rate (SAR) during RF excitation or by actual temperature monitoring. Perhaps the most important safety consideration is careful training of all personnel who might enter the suite.11
All contemporary MRI scanner systems offer basic real-time mode, with online interactive slice prescription during scanning.12 Meaningful therapeutic procedures may require more advanced functionality, including real-time multislice acquisition and 3D display and color highlighting of catheter channels.13 These features are available or under development by most hardware vendors.
In addition to anatomic images, investigational systems can provide additional functional information to guide procedures, such as real-time ("Doppler-like") flow mapping.14,15
Intravascular Devices
Most contemporary XRF catheter devices employ ferrous braiding to improve catheter stiffness and torque responsiveness. Ferrous materials degrade MR images unacceptably. Substitution with nonferrous materials usually renders the devices poorly visible under MRI and often degrades mechanical performance. A number of approaches to this problem are depicted in Figure 3 and reviewed briefly below.
|
Passive Catheter Designs
"Passive" catheters, visualized based on intrinsic materials properties, are attractive in their simplicity and safety but overall are not sufficiently visible.
Polymer catheters16,17 or guidewires18 with embedded metals can be tracked on the basis of signal voids (dark spots) induced by magnetic susceptibility artifacts.
Real-world navigation of passive catheters is challenging. Volume-averaging refers to detail lost when image voxels are larger than small volumes of interest. The dark spots created by passive catheters can be obscured by volume-averaging and therefore require imaging slices too thin to work with. Moreover, catheter elements can be "lost" when they move out of the selected imaging slice, and the "bloomed" signal voids caused by embedded metals can obscure nearby tissues. Nevertheless, clinical catheterization has successfully been conducted by tracking signal voids, such as those induced by CO2-filled balloon catheters.19,20
Investigators have successfully navigated passive catheters in vivo that are brightened because they are filled with gadolinium contrast.21,22 A shortcoming of this approach is that the lumen is no longer available to deliver other devices or agents. There has been little in vivo success coating catheters with gadolinium-like agents.23
Active Catheter Designs
"Active" catheters incorporate MRI receiver coils for imaging, tracking, or both. In early designs, catheter coils were placed directly against tissue to produce high-resolution imaging or spectroscopy.2430
Some active designs induce electronically controlled magnetic field inhomogeneities that cause signal voids,3134 visible in part because they can be turned on and off. These suffer similar limitations to passive designs because they use signal voids to indicate their position.
Active catheters that contain MR receiver coils are more conspicuous as they move within the body. Typically these designs make the entire length of catheters visible3540 (Figure 3C). An added benefit of these designs is that catheters can be visualized even if they move outside the selected imaging slice41 (Figure 4). Some approaches use the catheter coil both for signal detection and for RF excitation.37
|
Alternatively, active catheters with focal coils have been used to track individual points on the catheter.29,4244 This requires minor modifications of the pulse sequences used to acquire the images (ie, a small number of nonselective excitations intermittently to locate the markers in space). Cross-hairs indicating coil position can be overlaid on top of road-map or image frames as they are updated (Figure 3E). Position or motion of such tracking coils can be used automatically to specify MRI parameters such as scan plane37,45 or field of view.46 Similarly, surgical (and soon catheter) devices are commercially available incorporating multiple orthogonal magnetometers that, used in conjunction with known gradient sequences, can be used to determine catheter position.47 Such active tracking coils and magnetometers are attractive in their simplicity but insufficient to visualize the entire length of catheters, a feature that may be important for navigation or to recognize, for example, catheter devices that kink or otherwise fail. One possible solution would be to embed multiple tracking coils along the length of the catheters.
By virtue of incorporating long conductive elements, active catheter designs risk heating during the RF excitation required for MRI.4850 Several investigators have addressed this problem by incorporating circuitry for detuning or decoupling the conductive wire during RF excitation,5154 thereby preventing heating. Some promising recent decoupling approaches incorporate transformers into the transmission lines55 or replace transmission lines with fiberoptic connections.33,34,56,57
One active device, an "open loop" 0.030'' x 100-cm guidewire coil,9,58,59 has US Food and Drug Administration approval to market for intravascular MR imaging.59a Although it could be used to deliver devices for therapeutic intravascular procedures, the receiver sensitivity of this design falls to zero at the tip, which limits its use in navigating diseased blood vessels. More complex designs address this shortcoming40 but are not yet commercially available.
Hybrid and Multispectral Catheter Designs
Hybrid designs have been demonstrated that incorporate tuned receiver coil elementsbut not connecting wiresinto catheters60,61 or even stents.62,63 These inductively coupled designs can provide much of the visibility of traditional active devices without the potential heating problem of conductive transmission lines (Figure 3F).
Multispectral MRI, visualizing spins from species other than water protons, is an exciting new concept being explored for passive catheter tracking by several groups. Kozerke et al64 have demonstrated MRI tracking of 19F-filled catheters in phantoms. Golmans team has demonstrated in vivo catheter tracking and angiography using hyperpolarized 13C-hydrocarbons.65,66 Concurrent or interleaved proton and nonproton MRI might combine the superior tracking ability of active devices with the safety of passive devices.
Applications
A compelling array of applications has been reported for interventional cardiovascular MRI in animal models.
Cardiac
Schalla et al have demonstrated transvenous and transarterial cardiac catheterization in a porcine model of atrial septal defect wholly using iCMR and tracking receiver microcoils to mark catheter tips.67 They navigated catheters among cardiac chambers to obtain invasive pressure and oximetric measurements. The strength of this demonstration is its excellent simulation of clinical procedures; the weakness is the inability to visualize the length of their catheters, an important safety limitation in clinical procedures. Volz et al68 have recently demonstrated local flow assessment using intravascular receiver coils, and Thompson et al69 assessed flow noninvasively using complex-difference phase contrast, both of which might prove useful adjuncts during cardiovascular catheterization.
Our group has conducted percutaneous transcatheter myocardial injection of gadolinium contrast,70 and in a later iteration, targeted delivery of iron-labeled mesenchymal stromal cells to myocardial infarct border targets.41 These used multiple active intravascular devices, visible along their entire length, combined with rapid sequential imaging of multiple slices. The slices were displayed in 3D and used colors to highlight catheter-related signal (Figure 1B and Figure 4). The strength of the demonstration is the simulation of clinical procedures, with excellent target and catheter visualization, and instantaneous rather than roadmap myocardial targeting, such as that provided by electromagnetic positioning (eg, Biosense NOGA). The weaknesses are that the utility of precise cell delivery remains unproven and that catheters are not available for clinical testing. Other groups also have used retrograde transaortic access to conduct image-guided myocardial injections.7174,74a
Kuehne et al75 demonstrated MRI-guided transcatheter aortic valve replacement in swine using passive nitinol devices. This application is attractive because of the critical importance of image-guided placement of the stent-valve in relation to the coronary arteries and aortic root. However, such susceptibility-based device imaging alone probably will not be satisfactory in this application.
Three groups7678 have deployed passive nitinol atrial septal defect (ASD) occluders in porcine models. Although these are good demonstrations, it remains unclear whether iCMR adds clinical value in positioning ASD occluders compared with XRF and concurrent intravascular or transesophageal ultrasound guidance.
Cardiac electrophysiologists ablate myocardium to alter local myocardial conduction and interrupt symptomatic atrial or ventricular arrhythmias. Currently these are guided functionally using electrograms and electromagnetic mapping, with adjunctive image guidance using XRF or ultrasound. Primarily image-guided myocardial ablation is an interesting alternative that currently is conducted under direct surgical guidance79 and that might be conducted under iCMR. Preliminary catheter tracking with acquisition of filtered local electrograms has been reported,80 as has MRI characterization of ablated myocardium.81 The Massachusetts General Hospital team has conducted real-time MRI positioning of an electrophysiological catheter overlaid onto high-resolution time-resolved (4-dimensional) images of infarcted pig myocardium82 (Figure 5).
|
Several groups54,83 have reported percutaneous coronary artery intervention in healthy animals, some showing spectacular images of intracoronary stent artifacts.83 These works, though impressive, may not readily be translated into clinic. Using contemporary proton MRI, it is not clear that iCMR-guided coronary intervention ever will be feasible. Traversing complex coronary lesions require delicate guidewire and device manipulation, and currently this is conducted under XRF guidance with device imaging at 200-µm spatial resolution in a 5121024 square pixel matrix at 15 to 60 frames/second. Comparable spatial or temporal resolution does not appear accessible with available MRI technology.
A promising recent development is the report of successful iCMR-guided atrial septal puncture by Arepally et al.84 Our labs also have successfully employed this procedure to enable balloon atrial septostomy85 (Figure 6). These iCMR demonstrations are noteworthy because they show cardiovascular procedures need no longer be confined to vascular chambers and lumens. In particular, this technology likely will soon enable extra-anatomic bypass procedures such as palliative left-right extracardiac shunt or aortofemoral bypass, procedures that at present require open surgical exposure. Even catheter-based valve repair procedures may prove feasible.
|
Extra-Cardiac
Several groups have conducted angioplasty8690 and stenting62,63,9195 (Figure 7A7C) using passive and active catheter techniques in animal models of arterial stenosis. Similarly, groups have positioned aortic aneurysm endografts,96,97 inferior vena cava filter98100 devices, and have embolized renal artery segments101,102 using iCMR. These are important proofs of concept toward clinical application, but such demonstrations in healthy animals or simple disease models are relatively limited: For the most part, they require simple axial displacement of devices compared with difficult device navigation and positioning in patients with complex atherosclerotic disease.
|
Perhaps more exciting are recent novel procedures that demonstrate the real strength of iCMR to provide surgical-grade exposure for minimally invasive procedures that are not adequately guided by XRF or that require open surgery. Our group103 has recanalized long segments of chronic total arterial occlusions in an animal model. Using custom catheter and guidewire coils, we were able to traverse long segments of occlusion while keeping within arterial adventitial borders, an important clinical challenge (Figure 7D7E). Arepally et al104 have made a transcatheter mesocaval shunt, or extrahepatic connection between portal and systemic venous circulations, first using a septostomy needle and then deploying a novel custom vascular connector (Figure 8). A team at Stanford integrated a flat-panel XRF system into a double-doughnut operative MRI system and conducted dual-modality transjugular intrahepatic portosystemic shunting (TIPS) in animals105 and in patients106 and showed a reduction in number of punctures required compared with historical controls.
|
Early Human Experiments
A few human iCMR procedures have been conducted. Razavi et al20 reported a landmark series of cardiac catheterizations in children using a combined XMR environment. The same group is conducting x-ray fused with MRI (XFM) procedures, in which prior MRI datasets are combined with real-time XRF to conduct therapeutic procedures (Figure 9). The Hopkins team59 and others9,59a have reported invasive receiver coil imaging of human peripheral artery atheromata. The Regensburg team has conducted high-quality selective intra-arterial MR angiography107 and has reported some preliminary revascularization procedures using passive devices in the iliac108 and femoral109 arteries. As described above, the Stanford team has conducted iCMR-assisted TIPS procedures in patients.106
|
| Future Prospects and Conclusion |
|---|
|
|
|---|
However, iCMR offers single-modality solutions that may prove superior. In particular, iCMR may enable a range of breakthrough applications, such as image-guided ablation to treat cardiac rhythm disorders, minimally invasive extra-anatomic bypass, and beating-heart valve repair. Not only can iCMR provide surgical-grade exposure for clinicians, it can visualize the immediate effects of treatment upon target tissue. Overall, iCMR guidance for transcatheter cardiovascular procedures has the potential to revolutionize the way interventionists operate, enabling minimally invasive procedures otherwise impossible with conventional imaging guidance.
| Acknowledgments |
|---|
| Footnotes |
|---|
The online-only Data Supplement, which contains a video file, can be found at http://circ.ahajournals.org/cgi/content/full/112/19/3009/DC1.
| References |
|---|
|
|
|---|
2. Bashore TM, Bates ER, Berger PB, Clark DA, Cusma JT, Dehmer GJ, Kern MJ, Laskey WK, OLaughlin MP, Oesterle S, Popma JJ, ORourke RA, Abrams J, Brodie BR, Douglas PS, Gregoratos G, Hlatky MA, Hochman JS, Kaul S, Tracy CM, Waters DD, Winters WL Jr. American College of Cardiology/Society for Cardiac Angiography and Interventions clinical expert consensus document on cardiac catheterization laboratory standards: a report of the American College of Cardiology Task Force on Clinical Expert Consensus Documents. J Am Coll Cardiol. 2001; 37: 21702214.
3. McCullough PA, Wolyn R, Rocher LL, Levin RN, ONeill WW. Acute renal failure after coronary intervention: incidence, risk factors, and relationship to mortality. Am J Med. 1997; 103: 368375.[CrossRef][Medline] [Order article via Infotrieve]
4. Goldstein JA, Balter S, Cowley M, Hodgson J, Klein LW. Occupational hazards of interventional cardiologists: prevalence of orthopedic health problems in contemporary practice. Catheter Cardiovasc Interv. 2004; 63: 407411.[CrossRef][Medline] [Order article via Infotrieve]
5. Duerk JL, Butts K, Hwang KP, Lewin JS. Pulse sequences for interventional magnetic resonance imaging. Top Magn Reson Imaging. 2000; 11: 147162.[CrossRef][Medline] [Order article via Infotrieve]
6. Duerk JL, Wong EY, Lewin JS. A brief review of hardware for catheter tracking in magnetic resonance imaging. Magma. 2002; 13: 199208.[CrossRef][Medline] [Order article via Infotrieve]
7. Shellock FG. Magnetic Resonance Procedures: Health Effects and Safety. Boca Raton: CRC Press; 2001.
8. Shellock FG, Crues JV. MR procedures: biologic effects, safety, and patient care. Radiology. 2004; 232: 635652.
9. Dick AJ, Raman VK, Raval AN, Guttman MA, Thompson RB, Ozturk C, Peters DC, Stine AM, Wright VJ, Schenke WH, Lederman RJ. Invasive human magnetic resonance imaging during angioplasty: feasibility in a combined XMR suite. Catheter Cardiovasc Interv. 2005; 64: 265274.[CrossRef][Medline] [Order article via Infotrieve]
10. Hennel F, Girard F, Loenneker T. "Silent" MRI with soft gradient pulses. Magn Reson Med. 1999; 42: 610.[CrossRef][Medline] [Order article via Infotrieve]
11. Kanal E, Borgstede JP, Barkovich AJ, Bell C, Bradley WG, Felmlee JP, Froelich JW, Kaminski EM, Keeler EK, Lester JW, Scoumis EA, Zaremba LA, Zinninger MD. American College of Radiology white paper on MR safety. AJR Am J Roentgenol. 2002; 178: 13351347.
12. Debbins JP, Riederer SJ, Rossman PJ, Grimm RC, Felmlee JP, Breen JF, Ehman RL. Cardiac magnetic resonance fluoroscopy. Magn Reson Med. 1996; 36: 588595.[Medline] [Order article via Infotrieve]
13. Guttman MA, Lederman RJ, Sorger JM, McVeigh ER. Real-time volume rendered MRI for interventional guidance. J Cardiovasc Magn Reson. 2002; 4: 431442.[CrossRef][Medline] [Order article via Infotrieve]
14. Riederer SJ, Wright RC, Ehman RL, Rossman PJ, Holsinger-Bampton AE, Hangiandreou NJ, Grimm RC. Real-time interactive color flow MR imaging. Radiology. 1991; 181: 3339.
15. Nayak KS, Pauly JM, Kerr AB, Hu BS, Nishimura DG. Real-time color flow MRI. Magn Reson Med. 2000; 43: 251258.[CrossRef][Medline] [Order article via Infotrieve]
16. Bakker CJ, Hoogeveen RM, Weber J, van Vaals JJ, Viergever MA, Mali WP. Visualization of dedicated catheters using fast scanning techniques with potential for MR-guided vascular interventions. Magn Reson Med. 1996; 36: 816820.[Medline] [Order article via Infotrieve]
17. Bakker CJ, Bos C, Weinmann HJ. Passive tracking of catheters and guidewires by contrast-enhanced MR fluoroscopy. Magn Reson Med. 2001; 45: 1723.[CrossRef][Medline] [Order article via Infotrieve]
18. Buecker A, Spuentrup E, Schmitz-Rode T, Kinzel S, Pfeffer J, Hohl C, van Vaals JJ, Gunther RW. Use of a nonmetallic guide wire for magnetic resonanceguided coronary artery catheterization. Invest Radiol. 2004; 39: 656660.[CrossRef][Medline] [Order article via Infotrieve]
19. Miquel ME, Hegde S, Muthurangu V, Corcoran BJ, Keevil SF, Hill DL, Razavi RS. Visualization and tracking of an inflatable balloon catheter using SSFP in a flow phantom and in the heart and great vessels of patients. Magn Reson Med. 2004; 51: 988995.[CrossRef][Medline] [Order article via Infotrieve]
20. Razavi R, Hill DL, Keevil SF, Miquel ME, Muthurangu V, Hegde S, Rhode K, Barnett M, van Vaals J, Hawkes DJ, Baker E. Cardiac catheterisation guided by MRI in children and adults with congenital heart disease. Lancet. 2003; 362: 18771882.[CrossRef][Medline] [Order article via Infotrieve]
21. Unal O, Korosec FR, Frayne R, Strother CM, Mistretta CA. A rapid 2D time-resolved variable-rate k-space sampling MR technique for passive catheter tracking during endovascular procedures. Magn Reson Med. 1998; 40: 356362.[Medline] [Order article via Infotrieve]
22. Omary RA, Unal O, Koscielski DS, Frayne R, Korosec FR, Mistretta CA, Strother CM, Grist TM. Real-time MR imagingguided passive catheter tracking with use of gadolinium-filled catheters. J Vasc Interv Radiol. 2000; 11: 10791085.[Medline] [Order article via Infotrieve]
23. Frayne R, Strother CM, Unal O, Wehelie A, Yu H. MR signalemitting coatings. United States Patent No. 6,361,759. United States; 2002.
24. Kantor HL, Briggs RW, Balaban RS. In vivo 31P nuclear magnetic resonance measurements in canine heart using a catheter-coil. Circ Res. 1984; 55: 261266.
25. Martin AJ, Plewes DB, Henkelman RM. MR imaging of blood vessels with an intravascular coil. J Magn Reson Imaging. 1992; 2: 421429.[Medline] [Order article via Infotrieve]
26. Hurst GC, Hua J, Duerk JL, Cohen AM. Intravascular (catheter) NMR receiver probe: preliminary design analysis and application to canine iliofemoral imaging. Magn Reson Med. 1992; 24: 343357.[Medline] [Order article via Infotrieve]
27. Kandarpa K, Jakab P, Patz S, Schoen FJ, Jolesz FA. Prototype miniature endoluminal MR imaging catheter. J Vasc Interv Radiol. 1993; 4: 419427.[Medline] [Order article via Infotrieve]
28. Atalar E, Bottomley PA, Ocali O, Correia LC, Kelemen MD, Lima JA, Zerhouni EA. High resolution intravascular MRI and MRS by using a catheter receiver coil. Magn Reson Med. 1996; 36: 596605.[Medline] [Order article via Infotrieve]
29. Hillenbrand CM, Elgort DR, Wong EY, Reykowski A, Wacker FK, Lewin JS, Duerk JL. Active device tracking and high-resolution intravascular MRI using a novel catheter-based, opposed-solenoid phased array coil. Magn Reson Med. 2004; 51: 668675.[CrossRef][Medline] [Order article via Infotrieve]
30. Worthley SG, Helft G, Fuster V, Fayad ZA, Shinnar M, Minkoff LA, Schechter C, Fallon JT, Badimon JJ. A novel nonobstructive intravascular MRI coil: in vivo imaging of experimental atherosclerosis. Arterioscler Thromb Vasc Biol. 2003; 23: 346350.
31. Glowinski A, Kursch J, Adam G, Bucker A, Noll TG, Gunther RW. Device visualization for interventional MRI using local magnetic fields: basic theory and its application to catheter visualization. IEEE Trans Med Imaging. 1998; 17: 786793.[CrossRef][Medline] [Order article via Infotrieve]
32. Adam G, Glowinski A, Neuerburg J, Bucker A, van Vaals JJ, Gunther RW. Visualization of MR-compatible catheters by electrically induced local field inhomogeneities: evaluation in vivo. J Magn Reson Imaging. 1998; 8: 209213.[Medline] [Order article via Infotrieve]
33. Konings MK, Bartels LW, van Swol CF, Bakker CJ. Development of an MR-safe tracking catheter with a laser-driven tip coil. J Magn Reson Imaging. 2001; 13: 131135.[CrossRef][Medline] [Order article via Infotrieve]
34. Eggers H, Weiss S, Boernert P, Boesiger P. Image-based tracking of optically detunable parallel resonant circuits. Magn Reson Med. 2003; 49: 11631174.[CrossRef][Medline] [Order article via Infotrieve]
35. Ladd ME, Erhart P, Debatin JF, Hofmann E, Boesiger P, von Schulthess GK, McKinnon GC. Guidewire antennas for MR fluoroscopy. Magn Reson Med. 1997; 37: 891897.[Medline] [Order article via Infotrieve]
36. Ocali O, Atalar E. Intravascular magnetic resonance imaging using a loopless catheter antenna. Magn Reson Med. 1997; 37: 112118.[Medline] [Order article via Infotrieve]
37. Atalar E, Kraitchman DL, Carkhuff B, Lesho J, Ocali O, Solaiyappan M, Guttman MA, Charles HK Jr. Catheter-tracking FOV MR fluoroscopy. Magn Reson Med. 1998; 40: 865872.[Medline] [Order article via Infotrieve]
38. Burl M, Coutts GA, Herlihy DJ, Hill-Cottingham R, Eastham JF, Hajnal JV, Young IR. Twisted-pair RF coil suitable for locating the track of a catheter. Magn Reson Med. 1999; 41: 636638.[CrossRef][Medline] [Order article via Infotrieve]
39. Rivas PA, Nayak KS, Scott GC, McConnell MV, Kerr AB, Nishimura DG, Pauly JM, Hu BS. In vivo real-time intravascular MRI. J Cardiovasc Magn Reson. 2002; 4: 223232.[CrossRef][Medline] [Order article via Infotrieve]
40. Susil RC, Yeung CJ, Atalar E. Intravascular extended sensitivity (IVES) MRI antennas. Magn Reson Med. 2003; 50: 383390.[CrossRef][Medline] [Order article via Infotrieve]
41. Dick AJ, Guttman MA, Raman VK, Peters DC, Pessanha BS, Hill JM, Smith S, Scott G, McVeigh ER, Lederman RJ. Magnetic resonance fluoroscopy allows targeted delivery of mesenchymal stem cells to infarct borders in swine. Circulation. 2003; 108: 28992904.
42. Dumoulin CL, Souza SP, Darrow RD. Real-time position monitoring of invasive devices using magnetic resonance. Magn Reson Med. 1993; 29: 411415.[Medline] [Order article via Infotrieve]
43. Leung DA, Debatin JF, Wildermuth S, McKinnon GC, Holtz D, Dumoulin CL, Darrow RD, Hofmann E, von Schulthess GK. Intravascular MR tracking catheter: preliminary experimental evaluation. AJR Am J Roentgenol. 1995; 164: 12651270.
44. Ladd ME, Zimmermann GG, McKinnon GC, von Schulthess GK, Dumoulin CL, Darrow RD, Hofmann E, Debatin JF. Visualization of vascular guidewires using MR tracking. J Magn Reson Imaging. 1998; 8: 251253.[Medline] [Order article via Infotrieve]
45. Zhang Q, Wendt M, Aschoff AJ, Zheng L, Lewin JS, Duerk JL. Active MR guidance of interventional devices with target-navigation. Magn Reson Med. 2000; 44: 5665.[CrossRef][Medline] [Order article via Infotrieve]
46. Wacker FK, Elgort D, Hillenbrand CM, Duerk JL, Lewin JS. The catheter-driven MRI scanner: a new approach to intravascular catheter tracking and imaging-parameter adjustment for interventional MRI. AJR Am J Roentgenol. 2004; 183: 391395.
47. Nevo E. Method and apparatus to estimate location and orientation of objects during magnetic resonance imaging. United States Patent No. 6,516,213. United States; 2003.
48. Konings MK, Bartels LW, Smits HF, Bakker CJ. Heating around intravascular guidewires by resonating RF waves. J Magn Reson Imaging. 2000; 12: 7985.[CrossRef][Medline] [Order article via Infotrieve]
49. Nitz WR, Oppelt A, Renz W, Manke C, Lenhart M, Link J. On the heating of linear conductive structures as guide wires and catheters in interventional MRI. J Magn Reson Imaging. 2001; 13: 105114.[CrossRef][Medline] [Order article via Infotrieve]
50. Liu CY, Farahani K, Lu DS, Duckwiler G, Oppelt A. Safety of MRI-guided endovascular guidewire applications. J Magn Reson Imaging. 2000; 12: 7578.[CrossRef][Medline] [Order article via Infotrieve]
51. Ladd ME, Quick HH. Reduction of resonant RF heating in intravascular catheters using coaxial chokes. Magn Reson Med. 2000; 43: 615619.[CrossRef][Medline] [Order article via Infotrieve]
52. Lee C, McNamara C, Viohl I. Connector and guidewire connectable thereto. United States Patent No. 6,714,809. United States: Surgi-Vision; 2004.
53. Yeung CJ, Susil RC, Atalar E. RF safety of wires in interventional MRI: using a safety index. Magn Reson Med. 2002; 47: 187193.[CrossRef][Medline] [Order article via Infotrieve]
54. Serfaty JM, Yang X, Foo TK, Kumar A, Derbyshire A, Atalar E. MRI-guided coronary catheterization and PTCA: a feasibility study on a dog model. Magn Reson Med. 2003; 49: 258263.[CrossRef][Medline] [Order article via Infotrieve]
55. Weiss S, Vernickel P, Schaeffter T, Schultz V, Gleich B. A safe transmission line for interventional devices. In: Jolesz F, Kahn T, Lewin J, eds. 5th International MRI Symposium. Boston, Mass; 2004.
56. Wong EY, Zhang Q, Duerk JL, Lewin JS, Wendt M. An optical system for wireless detuning of parallel resonant circuits. J Magn Reson Imaging. 2000; 12: 632638.[CrossRef][Medline] [Order article via Infotrieve]
57. Weiss S, Schaeffter T, Brinkert F, Kuhne T, Bucker A. [An approach for safe visualization and localization of catheter during MR-guided intravascular procedures]. Z Med Phys. 2003; 13: 172176.[Medline] [Order article via Infotrieve]
58. Hofmann LV, Liddell RP, Arepally A, Montague B, Yang X, Bluemke DA. In vivo intravascular MR imaging: transvenous technique for arterial wall imaging. J Vasc Interv Radiol. 2003; 14: 13171327.[Medline] [Order article via Infotrieve]
59. Hofmann LV, Liddell RP, Eng J, Wasserman BA, Arepally A, Lee DS, Bluemke DA. Human peripheral arteries: feasibility of transvenous intravascular MR imaging of the arterial wall. Radiology. 2005; 235: 617622.
59. Larose E, Yeghiazarians Y, Libby P, Yucel EK, Aikawa M, Kacher DF, Aikawa E, Kinlay S, Schoen FJ, Selwyn AP, Ganz P. Characterization of human atherosclerotic plaques by intravascular magnetic resonance imaging. Circulation. 2005; 112: 23242331.
60. Kuehne T, Fahrig R, Butts K. Pair of resonant fiducial markers for localization of endovascular catheters at all catheter orientations. J Magn Reson Imaging. 2003; 17: 620624.[CrossRef][Medline] [Order article via Infotrieve]
61. Quick HH, Zenge MO, Kuehl H, Kaiser G, Aker S, Massing S, Bosk S, Ladd ME. Interventional magnetic resonance angiography with no strings attached: wireless active catheter visualization. Magn Reson Med. 2005; 53: 446455.[CrossRef][Medline] [Order article via Infotrieve]
62. Quick HH, Kuehl H, Kaiser G, Bosk S, Debatin JF, Ladd ME. Inductively coupled stent antennas in MRI. Magn Reson Med. 2002; 48: 781790.[CrossRef][Medline] [Order article via Infotrieve]
63. Kivelitz D, Wagner S, Schnorr J, Wetzler R, Busch M, Melzer A, Taupitz M, Hamm B. A vascular stent as an active component for locally enhanced magnetic resonance imaging: initial in vivo imaging results after catheter-guided placement in rabbits. Invest Radiol. 2003; 38: 147152.[CrossRef][Medline] [Order article via Infotrieve]
64. Kozerke S, Hegde S, Schaeffter T, Lamerichs R, Razavi R, Hill DL. Catheter tracking and visualization using 19F nuclear magnetic resonance. Magn Reson Med. 2004; 52: 693697.[CrossRef][Medline] [Order article via Infotrieve]
65. Svensson J, Mansson S, Johansson E, Petersson JS, Olsson LE. Hyperpolarized 13C MR angiography using trueFISP. Magn Reson Med. 2003; 50: 256262.[CrossRef][Medline] [Order article via Infotrieve]
66. Mansson S, Johansson E, Magnusson P, Chai CM, Hansson G, Petersson JS, Stahlberg F, Golman K. 13C imaging: a new diagnostic platform. Eur Radiol. Epub ahead of print June 14, 2005. DOI: 10.1007/s00330-005-2806-x.
67. Schalla S, Saeed M, Higgins CB, Martin A, Weber O, Moore P. Magnetic resonanceguided cardiac catheterization in a swine model of atrial septal defect. Circulation. 2003; 108: 18651870.
68. Volz S, Zuehlsdorff S, Umathum R, Hallscheidt P, Fink C, Semmler W, Bock M. Semiquantitative fast flow velocity measurements using catheter coils with a limited sensitivity profile. Magn Reson Med. 2004; 52: 575581.[CrossRef][Medline] [Order article via Infotrieve]
69. Thompson RB, McVeigh ER. High temporal resolution phase contrast MRI with multiecho acquisitions. Magn Reson Med. 2002; 47: 499512.[CrossRef][Medline] [Order article via Infotrieve]
70. Lederman RJ, Guttman MA, Peters DC, Thompson RB, Sorger JM, Dick AJ, Raman VK, McVeigh ER. Catheter-based endomyocardial injection with real-time magnetic resonance imaging. Circulation. 2002; 105: 12821284.
71. Saeed M, Lee R, Martin A, Weber O, Krombach GA, Schalla S, Lee M, Saloner D, Higgins CB. Transendocardial delivery of extracellular myocardial markers by using combination x-ray/MR fluoroscopic guidance: feasibility study in dogs. Radiology. 2004; 231: 689696.
72. Karmarkar PV, Kraitchman DL, Izbudak I, Hofmann LV, Amado LC, Fritzges D, Young R, Pittenger M, Bulte JW, Atalar E. MR-trackable intramyocardial injection catheter. Magn Reson Med. 2004; 51: 11631172.[CrossRef][Medline] [Order article via Infotrieve]
73. Corti R, Badimon J, Mizsei G, Macaluso F, Lee M, Licato P, Viles-Gonzalez JF, Fuster V, Sherman W. Real time magnetic resonance guided endomyocardial local delivery. Heart. 2005; 91: 348353.
74. Krombach GA, Pfeffer JG, Kinzel S, Katoh M, Gunther RW, Buecker A. MR-guided percutaneous intramyocardial injection with an MR-compatible catheter: feasibility and changes in T1 values after injection of extracellular contrast medium in pigs. Radiology. 2005; 235: 487494.
74. Rickers C, Gallegos R, Seethamraju RT, Wang X, Swingen C, Jayaswal A, Rahrmann EP, Kastenberg ZJ, Clarkson CE, Bianco R, OBrian T, Verfaillie C, Bolman RM 3rd, Wilke N, Jerosch-Herold M. Applications of magnetic resonance imaging for cardiac stem cell therapy. J Interv Cardiol. 2004; 17: 3746.[CrossRef][Medline] [Order article via Infotrieve]
75. Kuehne T, Yilmaz S, Meinus C, Moore P, Saeed M, Weber O, Higgins CB, Blank T, Elsaesser E, Schnackenburg B. Magnetic resonance imaging-guided transcatheter implantation of a prosthetic valve in aortic valve position: feasibility study in swine. J Am Coll Cardiol. 2004; 44: 22472249.
76. Buecker A, Spuentrup E, Grabitz R, Freudenthal F, Muehler EG, Schaeffter T, van Vaals JJ, Gunther RW. Magnetic resonanceguided placement of atrial septal closure device in animal model of patent foramen ovale. Circulation. 2002; 106: 511515.
77. Rickers C, Jerosch-Herold M, Hu X, Murthy N, Wang X, Kong H, Seethamraju RT, Weil J, Wilke NM. Magnetic resonance imageguided transcatheter closure of atrial septal defects. Circulation. 2003; 107: 132138.
78. Schalla S, Saeed M, Higgins CB, Weber O, Martin A, Moore P. Balloon sizing and transcatheter closure of acute atrial septal defects guided by magnetic resonance fluoroscopy: assessment and validation in a large animal model. J Magn Reson Imaging. 2005; 21: 204211.[CrossRef][Medline] [Order article via Infotrieve]
79. Hazel SJ, Paterson HS, Edwards JR, Maddern GJ. Surgical treatment of atrial fibrillation via energy ablation. Circulation. 2005; 111: e103e106.
80. Susil RC, Yeung CJ, Halperin HR, Lardo AC, Atalar E. Multifunctional interventional devices for MRI: a combined electrophysiology/MRI catheter. Magn Reson Med. 2002; 47: 594600.[CrossRef][Medline] [Order article via Infotrieve]
81. Lardo AC, McVeigh ER, Jumrussirikul P, Berger RD, Calkins H, Lima J, Halperin HR. Visualization and temporal/spatial characterization of cardiac radiofrequency ablation lesions using magnetic resonance imaging. Circulation. 2000; 102: 698705.
82. Reddy V, Malchano Z, Dukkipati S, Holmvang G, Schmidt E, Dumoulin C, Mallozzi R, Ruskin J. Interventional MRI: electroanatomical mapping using real-time MR tracking of a deflectable catheter. Heart Rhythm. 2005; 2: S279S280.Abstract.
83. Spuentrup E, Ruebben A, Schaeffter T, Manning WJ, Gunther RW, Buecker A. Magnetic resonanceguided coronary artery stent placement in a swine model. Circulation. 2002; 105: 874879.
84. Arepally A, Karmarkar PV, Weiss C, Rodriguez ER, Lederman RJ, Atalar E. Magnetic resonance imageguided trans-septal puncture in a swine heart. J Magn Reson Imaging. 2005; 21: 463467.[CrossRef][Medline] [Order article via Infotrieve]
85. Raval AN, Karmarkar PV, Guttman MA, Ozturk C, DeSilva R, Aviles RJ, Wright VJ, Schenke WH, Atalar E, McVeigh ER, Lederman RJ. Real-time MRI-guided atrial septal puncture and balloon septostomy in swine. Catheter Cardiovasc Interv. In press.
86. Wildermuth S, Dumoulin CL, Pfammatter T, Maier SE, Hofmann E, Debatin JF. MR-guided percutaneous angioplasty: assessment of tracking safety, catheter handling and functionality. Cardiovasc Intervent Radiol. 1998; 21: 404410.[CrossRef][Medline] [Order article via Infotrieve]
87. Yang X, Bolster BD Jr, Kraitchman DL, Atalar E. Intravascular MR-monitored balloon angioplasty: an in vivo feasibility study. J Vasc Interv Radiol. 1998; 9: 953959.[Medline] [Order article via Infotrieve]
88. Buecker A, Adam GB, Neuerburg JM, Kinzel S, Glowinski A, Schaeffter T, Rasche V, van Vaals JJ, Guenther RW. Simultaneous real-time visualization of the catheter tip and vascular anatomy for MR-guided PTA of iliac arteries in an animal model. J Magn Reson Imaging. 2002; 16: 201208.[CrossRef][Medline] [Order article via Infotrieve]
89. Godart F, Beregi JP, Nicol L, Occelli B, Vincentelli A, Daanen V, Rey C, Rousseau J. MR-guided balloon angioplasty of stenosed aorta: in vivo evaluation using near-standard instruments and a passive tracking technique. J Magn Reson Imaging. 2000; 12: 639644.[CrossRef][Medline] [Order article via Infotrieve]
90. Omary RA, Frayne R, Unal O, Warner T, Korosec FR, Mistretta CA, Strother CM, Grist TM. MR-guided angioplasty of renal artery stenosis in a pig model: a feasibility study. J Vasc Interv Radiol. 2000; 11: 373381.[Medline] [Order article via Infotrieve]
91. Buecker A, Neuerburg JM, Adam GB, Glowinski A, Schaeffter T, Rasche V, van Vaals JJ, Molgaard-Nielsen A, Guenther RW. Real-time MR fluoroscopy for MR-guided iliac artery stent placement. J Magn Reson Imaging. 2000; 12: 616622.[CrossRef][Medline] [Order article via Infotrieve]
92. Dion YM, Ben El Kadi H, Boudoux C, Gourdon J, Chakfe N, Traore A, Moisan C. Endovascular procedures under near-real-time magnetic resonance imaging guidance: an experimental feasibility study. J Vasc Surg. 2000; 32: 10061014.[CrossRef][Medline] [Order article via Infotrieve]
93. Kuehne T, Saeed M, Higgins CB, Gleason K, Krombach GA, Weber OM, Martin AJ, Turner D, Teitel D, Moore P. Endovascular stents in pulmonary valve and artery in swine: feasibility study of MR imagingguided deployment and postinterventional assessment. Radiology. 2003; 226: 475481.
94. Feng L, Dumoulin CL, Dashnaw S, Darrow RD, Delapaz RL, Bishop PL, Pile-Spellman J. Feasibility of stent placement in carotid arteries with real-time MR imaging guidance in pigs. Radiology. 2005; 234: 558562.
95. Raval AN, Telep JD, Guttman MA, Ozturk C, Jones M, Thompson RB, Wright VJ, Schenke WH, DeSilva R, Aviles RJ, Raman VK, Slack MC, Lederman RJ. Real-time magnetic resonance imagingguided stenting of aortic coarctation with commercially available catheter devices in swine. Circulation. 2005; 112: 699706.
96. Mahnken AH, Chalabi K, Jalali F, Gunther RW, Buecker A. Magnetic resonanceguided placement of aortic stents grafts: feasibility with real-time magnetic resonance fluoroscopy. J Vasc Interv Radiol. 2004; 15: 189195.[Medline] [Order article via Infotrieve]
97. Raman VK, Karmarkar P, Guttman MA, Dick AJ, Peters DC, Ozturk C, Raval AN, Pessanha BS, Thompson RB, De Silva R, Aviles RJ, Atalar E, McVeigh ER, Lederman RJ. Fluoroscopic magnetic resonanceguided endovascular repair of experimental abdominal aortic aneurysm in swine. J Am Coll Cardiol. 2005; 45: 20692077.
98. Frahm C, Gehl HB, Lorch H, Zwaan M, Drobnitzky M, Laub GA, Weiss HD. MR-guided placement of a temporary vena cava filter: technique and feasibility. J Magn Reson Imaging. 1998; 8: 105109.[Medline] [Order article via Infotrieve]
99. Bartels LW, Bos C, van Der Weide R, Smits HF, Bakker CJ, Viergever MA. Placement of an inferior vena cava filter in a pig guided by high-resolution MR fluoroscopy at 1.5 T. J Magn Reson Imaging. 2000; 12: 599605.[CrossRef][Medline] [Order article via Infotrieve]
100. Bucker A, Neuerburg JM, Adam GB, Glowinski A, Schaeffter T, Rasche V, van Vaals JJ, Gunther RW. Real-time MR guidance for inferior vena cava filter placement in an animal model. J Vasc Interv Radiol. 2001; 12: 753756.[Medline] [Order article via Infotrieve]
101. Bucker A, Neuerburg JM, Adam G, Glowinski A, van Vaals JJ, Gunther RW. [MR-guided coil embolisation of renal arteries in an animal model]. Rofo. 2003; 175: 271274.[Medline] [Order article via Infotrieve]
102. Fink C, Bock M, Umathum R, Volz S, Zuehlsdorff S, Grobholz R, Kauczor HU, Hallscheidt P. Renal embolization: feasibility of magnetic resonance-guidance using active catheter tracking and intraarterial magnetic resonance angiography. Invest Radiol. 2004; 39: 111119.[CrossRef][Medline] [Order article via Infotrieve]
103. Raval AN, Karmarkar PV, Ozturk C, Guttman MA, Xu M, DeSilva R, Aviles RJ, Dick AJ, Raman VK, Atalar E, McVeigh ER, Lederman RJ. Chronic total peripheral artery occlusion recanalization using interactive real-time magnetic resonance imaging guidance is feasible in a swine model. J Cardiovasc Magn Reson. 2005. Abstract.
104. Arepally A, Kamarkar P, Weiss C, Atalar E. Percutaneous MR-guided transvascular access of the mesenteric venous systemstudy in a swine model. Radiology. In press.
105. Kee ST, Rhee JS, Butts K, Daniel B, Pauly J, Kerr A, OSullivan GJ, Sze DY, Razavi MK, Semba CP, Herfkens RJ, Dake MD. 1999 Gary J. Becker Young Investigator Award. MR-guided transjugular portosystemic shunt placement in a swine model. J Vasc Interv Radiol. 1999; 10: 529535.[Medline] [Order article via Infotrieve]
106. Kee ST, Ganguly A, Daniel BL, Wen Z, Butts K, Shimikawa A, Pelc NJ, Fahrig R, Dake MD. MR-guided transjugular intrahepatic portosystemic shunt creation with use of a hybrid radiography/MR system. J Vasc Interv Radiol. 2005; 16: 227234.[Medline] [Order article via Infotrieve]
107. Paetzel C, Zorger N, Seitz J, Volk M, Nitz WR, Herold T, Feuerbach S, Lenhart M. Intraarterial contrast materialenhanced magnetic resonance angiography of the aortoiliac system. J Vasc Interv Radiol. 2004; 15: 981984.[Medline] [Order article via Infotrieve]
108. Manke C, Nitz WR, Djavidani B, Strotzer M, Lenhart M, Volk M, Feuerbach S, Link J. MR imagingguided stent placement in iliac arterial stenoses: a feasibility study. Radiology. 2001; 219: 527534.
109. Paetzel C, Zorger N, Bachthaler M, Hamer OW, Stehr A, Feuerbach S, Lenhart M, Volk M, Herold T, Kasprzak P, Nitz WR. Magnetic resonanceguided percutaneous angioplasty of femoral and popliteal artery stenoses using real-time imaging and intra-arterial contrastenhanced magnetic resonance angiography. Invest Radiol. 2005; 40: 257262.[CrossRef][Medline] [Order article via Infotrieve]
This article has been cited by other articles:
![]() |
R. Benza, R. Biederman, S. Murali, and H. Gupta Role of Cardiac Magnetic Resonance Imaging in the Management of Patients With Pulmonary Arterial Hypertension J. Am. Coll. Cardiol., November 18, 2008; 52(21): 1683 - 1692. [Abstract] [Full Text] [PDF] |
||||
![]() |
R M Duerden, K S Pointon, and S Habib Review of clinical cardiac MRI Imaging, September 1, 2006; 18(3): 178 - 186. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. S. Baim and R. Y. Kwong Is Magnetic Resonance Image Guidance the Key to Opening Chronic Total Occlusions? Circulation, February 28, 2006; 113(8): 1053 - 1055. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2005 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |