| Introduction |
|---|
|
|
|---|
Recently, we reported the death of a 21-year-old patient who received a Prizm 2 DR model 1861 ICD pulse generator (Guidant, Inc) in 2001 to prevent sudden cardiac death resulting from hypertrophic cardiomyopathy.3 In March 2005, this young man experienced a witnessed arrest and could not be resuscitated. His ICD was returned to Guidant, which found that the device had failed during the delivery of a shock. The cause of failure was massive electronic damage caused by electrical overstress that occurred when a short circuit developed between a high-voltage wire and a tube used to test the housing during manufacturing (see the Figure). At the time of our patients death, Guidant had knowledge of 25 similar Prizm 2 DR model 1861 failures in patients, 3 of whom had required rescue defibrillation. Indeed, Guidant had first observed this mode of failure 3 years earlier, in February 2002, when 2 returned Prizm 2 DR pulse generators exhibited the same short circuiting that caused our patients device to fail. Guidant was sufficiently concerned about these failures that manufacturing changes were made in April and November of 2002, which allegedly prevented short circuiting. Nevertheless, Guidant chose not to inform patients or physicians about these failures or the manufacturing changes designed to prevent them. Moreover, Guidant continued to sell pulse generators that were built before the 2002 manufacturing changes. Unknowingly, therefore, we and other physicians implanted Prizm 2 DR ICDs in 2002 and 2003 that Guidant knew were prone to sudden unexpected failure.4
|
In May 2005, after the death of our patient, we recommended during meetings with Guidant officials that the company promptly communicate the detailed nature of this flaw to physicians and patients. That communication, it was noted, should also emphasize that no test or monitoring technique could predict if or when a suspect Prizm 2 DR device may fail. This recommendation was based on the view that physicians and patients should have all critical information so that they can decide whether prophylactic ICD replacement was prudent. Guidants responsibility, in our opinion, was to disclose these vital data completely and expeditiously.
However, Guidant believed that such a communication was inadvisable and unnecessary. The company maintained that because the observed failure rate was very low, physicians could unnecessarily expose patients to the risks of device replacement surgery, including infection. Guidants statistical argument ignored the basic tenet that patients have a fundamental right to be fully informed when they are exposed to the risk of death no matter how low that risk may be perceived. Furthermore, by withholding vital information, Guidant had in effect assumed the primary role of managing high-risk patients, a responsibility that belongs to physicians. The prognosis of our young, otherwise healthy patient for a long, productive life was favorable if sudden death could have been prevented.5 Certainly, this was the rationale for implanting his Prizm 2 DR in 2001. If we had known that the Prizm 2 DR was prone to sudden failure as a result of short circuiting or another mechanism, his device would have been replaced promptly.
Because Guidant declined to inform patients and their physicians, we regarded it as our moral and ethical obligation to disclose the problem to the medical community and the public.3,6 On May 23, 2005, the day before the New York Times reported these events, Guidant sent a letter to physicians describing the Prizm 2 DR flaw and suggested the failures were random events; Guidant recommended that physicians continue normal monitoring.7
On June 17, 2005, after alerting the Food and Drug Administration (FDA), Guidant recalled 26 000 Prizm 2 DR devices manufactured before April 2002.4 Simultaneously, Guidant issued 3 additional recalls affecting >50 000 of its cardiac resynchronization ICDs.810 One of these recalls was prompted by 15 Contak Renewal and Contak Renewal 2 ICD failures8 that were also caused by short circuiting that Guidant had known about for at least a year. Similar to its management of the Prizm 2 DR situation, Guidant had implemented manufacturing changes in August 2004 and did not disclose the problem until a patient died when a Contak Renewal ICD short-circuited during shock delivery on May 30, 2005, while the FDA was investigating the Prizm 2 DR.
The FDA classified Guidants Prizm 2 DR and Contak Renewal and Contak Renewal 2 notifications as class I recalls, which indicate that a reasonable probability exists that the use of these devices will cause serious adverse health consequences or death. In contrast, Medtronics voluntary February 2005 advisory with regard to a battery short in its Marquis models11 was an FDA class II recall, which denotes that the probability of serious adverse health consequences such as death are remote. Collectively, these 4 Guidant recalls, which affected >70 000 implanted devices worldwide, were among the largest such industry regulatory actions in the past 25 years.
One can only speculate as to what Guidant or the FDA may have done or the events that may have transpired if the Prizm 2 DR problem had not been disclosed publicly.12 Nevertheless, there are important lessons to be learned from the failure and recall of these ICDs.
| Lesson 1: The FDAs Postmarket ICD Device Surveillance System Is Broken |
|---|
|
|
|---|
Guidant reported 2 adverse events to the FDA in 2004 that described the same Prizm 2 DR defect that caused our patients device to fail and further indicated that the company had made manufacturing changes to prevent it. The first report14 was received by the FDA >1 year before to the death of our patient. The second report15 was received by the FDA in August 2004 after a patient received an ineffective shock; the second report15 was explicit: "Analysis confirmed an electrical short between two components, specifically the feedthrough wire and backfill tube. It was concluded that the shorted condition involving the header df-1 feed-thru wires resulted in the clinical observations. Although the occurrence of this failure has been very low, Guidant implemented manufacturing enhancements in April and November 2002 to correct this issue."
The FDA, therefore, was in possession of important information about the safety of the Prizm 2 DR many months before the death of our patient and, to the best of our knowledge, took no action. The explanation for the FDAs inaction is unknown, but it may be that the agency was not prepared for the upsurge in ICD technology and the extraordinary growth in the number of ICD implantations that has occurred during the past 5 years.
Fixing the postmarket surveillance system must be a high priority for the FDA and Congress. The public should have full access to all of the FDAs medical device safety and efficacy data. This should include manufacturers annual reports that may contain failure data and manufacturing changes, which the FDA is presently withholding from public scrutiny presumably because they may reveal a companys trade secrets. For government to keep such vital product safety information from patients or physicians for any reason should be unlawful.
A physician or patient should be able to learn quickly from the FDA if a particular ICD or other critical medical device has exhibited quality issues that may affect the performance of a product. Moreover, the FDAs postmarket surveillance system should alert physicians and user facilities when manufacturers report the type of life-threatening failure modes exemplified by the Prizm 2 DR and the Contak Renewal and Contak Renewal 2 ICDs. Currently, the publics only source of postmarket product performance data are the FDAs Manufacturers and User Facility Device Experience (MAUDE) database. However, MAUDE is not designed to be a routine surveillance tool for physicians or a useful source of information for patients. Moreover, the MAUDE database may not be a reliable source of information; eg, it contained only 10 of the 28 Prizm 2 DR failures at the time these devices were recalled.
| Lesson 2: Physicians Do Not Have the Data Necessary to Assess Device Problems and to Make Rational Clinical Judgments |
|---|
|
|
|---|
Without precise failure rate data, physicians and patients cannot make prudent management decisions. In the aftermath of the Prizm 2 DR and Contak Renewal recalls, physicians and patients have had to choose between prophylactic replacement or continued follow-up. Because accurate failure rate data are unavailable for these devices, management decisions are being made according to the perceived rather than the actual risk of catastrophic ICD failure. Caught in this conundrum, and wishing to avoid surgical complications, eg, infection, and to "do no harm," physicians may be hesitant to replace these devices. For ICD patients at high risk for sudden cardiac arrest, however, the low likelihood of a treatable infection is acceptable when the alternative is sudden death should the device malfunction.
Because precise failure rate data are needed, a reformed postmarket surveillance system must include government-mandated prospective follow-up studies of market-released devices. These studies should be sufficiently powered to detect low-frequency device failures and to provide accurate estimates of ICD longevity.
| Lesson 3: Quality Standards for ICDs and Guidelines for Managing Device Recalls Are Needed |
|---|
|
|
|---|
Additionally, no universally accepted definitions exist for such critical device events as "random component failure." Guidant has stated that short circuiting in the Prizm 2 DR was due to a rare random component failure and implied that such failures occurred despite industrys best efforts to mitigate them.7 To suggest that the death of our patient or the death of the Contak Renewal patient was not due to a specific, avoidable failure mode is misleading and incorrect.
Despite the volume of recalls and advisories and the number of patients affected by them, the appropriate clinical strategies for managing ICD recalls and advisories are uncertain. A recent survey suggested that experienced physicians differ significantly in their approach to ICD recalls.16 The factor that most strongly influenced a physicians decision to replace a suspect device was the manufacturers estimated risk of sudden device failure. In our judgment, the patients underlying heart disease and prognosis should be the deciding factors favoring prophylactic device replacement. For example, the ICD may be truly life-saving for patients with genetic heart diseases because, for many of them, the only risk of cardiovascular death is ventricular fibrillation.
Given the large number of patients who have or will receive ICDs and the inevitability of future device problems, the Heart Rhythm Society, whose declared mission is to improve the care of patients by promoting optimal healthcare policies and standards, should take this unique opportunity to collaborate with other professional societies to establish realistic quality standards for ICDs and practical guidelines for managing device recalls.
| Conclusions |
|---|
|
|
|---|
Although ICDs are highly effective and generally dependable, the recent Prizm 2 DR model 1861 experience and the recalls of 2005 demand that ICD quality and reliability improve. The Heart Rhythm Society should lead the development of quality standards for ICDs and guidelines for managing device recalls and safety alerts. Congress and the FDA must develop and apply an effective postmarket surveillance system that improves the safety of medical devices for all patients. Finally, a major goal of these reforms is to reassure patients that ICD therapy is reliable and effectively regulated.
| Acknowledgments |
|---|
Dr Maron is a grantee from Medtronic, Inc.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
R. Passman and A. Kadish Sudden Death Prevention With Implantable Devices Circulation, July 31, 2007; 116(5): 561 - 571. [Full Text] [PDF] |
||||
![]() |
B. J. Maron, P. Spirito, W.-K. Shen, T. S. Haas, F. Formisano, M. S. Link, A. E. Epstein, A. K. Almquist, J. P. Daubert, T. Lawrenz, et al. Implantable Cardioverter-Defibrillators and Prevention of Sudden Cardiac Death in Hypertrophic Cardiomyopathy JAMA, July 25, 2007; 298(4): 405 - 412. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. J. Maron Surgical Myectomy Remains the Primary Treatment Option for Severely Symptomatic Patients With Obstructive Hypertrophic Cardiomyopathy Circulation, July 10, 2007; 116(2): 196 - 206. [Full Text] [PDF] |
||||
![]() |
M. Shuchman Debating the Risks of Drug-Eluting Stents N. Engl. J. Med., January 25, 2007; 356(4): 325 - 328. [Full Text] [PDF] |
||||
![]() |
B. J. Maron, I. Olivotto, and M. S. Maron The dilemma of left ventricular outflow tract obstruction and sudden death in hypertrophic cardiomyopathy: do patients with gradients really deserve prophylactic defibrillators? Eur. Heart J., August 2, 2006; 27(16): 1895 - 1897. [Full Text] [PDF] |
||||
![]() |
M. M. Scheinman and E. Keung The Year in Clinical Electrophysiology J. Am. Coll. Cardiol., March 21, 2006; 47(6): 1207 - 1213. [Full Text] [PDF] |
||||
![]() |
Z. Goldberger and R. Lampert Implantable Cardioverter-Defibrillators: Expanding Indications and Technologies JAMA, February 15, 2006; 295(7): 809 - 818. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Auricchio, M. Gropp, S. Ludgate, P. Vardas, J. Brugada, S. G. Priori, and Writing Committee for the European Heart Rhythm As European Heart Rhythm Association Guidance Document on cardiac rhythm management product performance. Europace, January 1, 2006; 8(5): 313 - 322. [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2005 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |