(Circulation. 2005;111:3435-3442.)
© 2005 American Heart Association, Inc.
Interventional Cardiology |
From the Brigham and Womens Hospital and Harvard Medical School, Boston, Mass.
Correspondence to Laura Mauri, MD, MSc, BC 3-012K, Brigham and Womens Hospital, 75 Francis St, Boston, MA 02116. E-mail lmauri1{at}partners.org
Received November 11, 2004; revision received March 3, 2005; accepted March 9, 2005.
| Abstract |
|---|
|
|
|---|
Methods and Results We examined the relationship between incremental changes in observed late loss in lumen diameter and the probability of restenosis using reported late loss from 22 published trials of various types of stents (bare-metal, drug-eluting, and small-vessel stents). Next, the power of late loss differences was compared with that of corresponding binary restenosis rates. The relationship between mean late loss and its SD was linear and did not vary with stent type (drug-eluting or bare-metal) or vessel diameter. At all levels of late loss examined (0 to 1 mm), incremental changes were associated with increasing restenosis risk (with an increasing magnitude of effect at higher levels of late loss). The power to detect a treatment effect was greater for late loss than for binary angiographic restenosis (
32% relative increase in power,
24% absolute increase for late loss between 0.2 and 0.6 mm).
Conclusions Late loss is monotonically related to restenosis risk in published stent trials. It is a generalizable and powerful angiographic end point in early or small trials of new drug-eluting stents.
Key Words: angioplasty coronary disease restenosis stents trials
| Introduction |
|---|
|
|
|---|
6 to 9 months after stenting with the use of quantitative angiographic and intravascular ultrasound metrics such as percent diameter stenosis or percent volume obstruction.14 The assessment of coronary restenosis as a failure event utilizes dichotomous end points, such as the occurrence of a 50% diameter stenosis or clinically driven repeated target lesion intervention at follow-up, that signify a critical amount of renarrowing associated with resting flow or flow reserve obstruction.
Successful drug-eluting stents have reduced angiographic restenosis rates by
75% compared with bare-metal stents, resulting in binary angiographic and clinical restenosis rates of only 5% to 10%.5,6 Because low binary event rates result in decreased statistical power in randomized trials, the number of patients needed to make comparisons between new and proven drug-eluting stents has increased substantially. Continuous end points such as follow-up minimum lumen diameter (MLD), percent diameter stenosis, and late loss have been used to test new stent technologies, particularly in early-phase, smallsample size trials, because of their inherent greater statistical power.79
Whereas relative measures of restenosis, such as percent diameter stenosis, depend on vessel diameter for their calculation and interpretation, absolute measures of restenosis magnitude can be more easily compared across trials of varying vessel size. Late loss, defined as the difference between immediate postprocedure MLD and MLD 6 to 9 months after percutaneous coronary intervention,10 is an angiographic measure of the absolute amount of renarrowing. Specifically, it measures the change in MLD of the treated coronary segment due to vascular contraction and neointimal hyperplasia.11 A general model adjusted for acute gain was developed in 1993 that directly related late loss to the incidence of binary restenosis.1,12 For coronary stents, which by design resist vascular contraction and generally achieve a uniform lumen diameter within the stented segment, late loss is an angiographic surrogate for neointimal hyperplasia, the target of drug-eluting stents.11
Within the sirolimus-eluting stent trials, mean late loss is correlated with the probability of angiographic and clinical restenosis.13 We sought to determine (1) whether this correlation between mean late loss and restenosis rates was generalizable to other stent studies and (2) whether the use of late loss as an end point in modest-sized clinical trials would be more powerful than the traditional binary restenosis end points. Specifically, we hypothesized that late loss in drug-eluting stent trials is a monotonic measure of restenosis risk, and we developed a model of late loss for evaluating contemporary drug-eluting stents.
| Methods |
|---|
|
|
|---|
Definitions
Late loss was defined as the difference between the MLD immediately after the procedure and the MLD at 6- to 8-month follow-up. Percent diameter stenosis was defined as [1(MLD/reference vessel diameter)]x100. Binary angiographic restenosis was defined as a >50% diameter stenosis at follow-up. To evaluate the relationship of mean late loss to both binary angiographic restenosis and its standard deviation (SD), the in-stent rather than in-segment (including the 5-mm margins proximal and distal to the stent) late loss was used because in-segment measurement was not available in most studies (Tables 1 and 2
).
|
|
Statistical Analysis
Baseline clinical and angiographic variables were recorded as mean and SD for continuous variables and percentage for binary variables.
Evaluation of Monotonicity of Relationship Between Mean and SD of Late Loss
We performed regression analysis on the SD of in-stent late loss for each trial with mean late loss as the predictor variable. We tested linear, nonlinear (using polynomial terms), and interaction effects (for drug-eluting versus bare-metal and for small-vessel versus nonsmall-vessel stenting). A 2-sided P value
0.05 was considered significant.
Power Transformation for Late Loss Distribution
In drug-eluting stent trials, distributions with low late loss (in which the mean late loss value is a fraction of the SD) require transformation to account for right-skewed data, whereas higher late loss distributions (as in bare-metal stents with mean late loss values
1 mm and relatively smaller SDs) do not require transformation for normality.13 We estimated the power transformation necessary to overcome the magnitude of rightward skew by interpolating the optimum power between those of the drug-eluting and control arms of the SIRIUS trial, anchored at mean late losses of 0.17 and 1.0 mm, respectively.13
Predication of Binary Restenosis Rate From Mean In-Stent Late Loss
The power transformation model was used to predict the binary angiographic restenosis rate from late loss. Essentially, the transformed mean and SD were used to estimate the probability of exceeding the transformed late loss threshold for binary restenosis by integrating the area under the cumulative distribution function. The predicted probability of restenosis, assuming a given mean MLD and reference diameter, was calculated for varying mean late loss from 0 to 1.0 mm, in 0.1-mm increments.
Power Comparison (Late Loss Versus Binary Angiographic Restenosis)
The powers of the end points late loss and binary angiographic restenosis were compared with the example of a hypothetical trial designed to detect a 35% treatment effect between 2 stents using 200 subjects per arm. In the range of late loss values from 0 to 1.0 mm, power was estimated for a 35% treatment effect on the end point of late loss and compared with the power of the corresponding predicted restenosis rates and treatment effect from the model.
| Results |
|---|
|
|
|---|
Late Loss Is Positively Correlated With Restenosis Rate
A generalizable correlation across bare-metal stent and drug-eluting stent trials would be required for late loss to serve as a valid surrogate for binary restenosis. Two analyses were performed to establish the nature of this relationship: (1) an analysis of the relationship between observed mean late loss from each trial and the corresponding observed restenosis rates and (2) an analysis of the relationship between the mean late loss and the corresponding SDs.
Late Loss Is Related Monotonically to Binary Restenosis
The relationship between mean in-stent late loss and restenosis rate for all study samples was found to be curvilinear and monotonic (Figure 1), that is, higher mean late losses were associated with higher reported rates of binary angiographic restenosis. Interaction terms for bare-metal versus drug-eluting and small-vessel versus nonsmall-vessel stenting were nonsignificant, with no apparent heterogeneity of the relationship across these categories.
|
SD of Late Loss Is Related Monotonically to Its Mean
If the relationship of the mean and SD of late loss does not follow a monotonic relationship, identification of a population with a higher mean but lower SD than an existing reference could occur, creating the theoretical possibility that a higher mean would be associated with a lower restenosis rate. Such a paradoxical situation would make interpretation of mean late loss values complex. In contrast, we found that the relationship between the late loss mean and SD was monotonic across the 22 trials with reported in-stent late loss (Figure 2), suggesting that conditions for such a paradoxical relationship did not exist. Linear regression revealed that the relationship between mean and SD was best described with the following function: SD of late loss=0.33+0.31xmean late loss in millimeters.
|
Comparative Efficiency of Late Loss and Binary Angiographic Restenosis
Power calculations for the restenosis end points late loss and binary angiographic restenosis were compared for a hypothetical stent study of modest sample size (200 subjects per arm). As generally expected with the comparison of continuous and binary variables measuring similar processes, late loss was consistently found to be a more efficient end point than binary angiographic restenosis across mean late loss, ranging from 0 to 1.0 mm (Figure 3). The largest difference in power between the 2 end points occurred at mean late loss levels of 0.1 to 0.7 mm.
|
Relationship of Late Loss to Binary Restenosis Rate
Using the relationship between the mean in-stent late loss and its SD defined above, one can predict the SD for a given hypothetical mean. We previously described a statistical method to overcome the greater degree of right skew in the late loss distribution at lower means by using a power transformation varying with the mean late loss.13 Combination of these 2 methods yields a model to predict binary angiographic restenosis rates from late loss for a study population of any given reference vessel diameter and residual stenosis (Table 3).37 This model describes a monotonic and curvilinear relationship between mean late loss and restenosis rate (Figure 4). The expected incremental restenosis rate between 2 mean late losses of 0.2 and 0.4 mm is 3.1%, whereas the incremental restenosis rate between 2 mean late losses of 0.4 and 0.6 mm is 6.4% in a reference population with mean reference vessel diameter of 2.79 mm and residual stenosis of 6% (Table 3).
|
|
| Discussion |
|---|
|
|
|---|
We sought to evaluate the generalizability and power of the angiographic end point late loss as a summary statistic for study samples. We demonstrated that across the published stent trials, a strong positive association exists between the mean in-stent late loss estimates and binary angiographic restenosis rates. Specifically, over the range of trials observing late loss from 0 to 1.3 mm, higher sample mean late losses were associated with higher sample restenosis rates. Consequently, the conditions for a paradoxical result in which an increasing increment of late loss might correspond to a decreasing increment of binary restenosis risk did not exist. Further statistical support for this positive monotonic (nondecreasing) correlation was the association of late loss sample mean and SD. The theoretical possibility that an increasing increment of mean late loss could correspond to a decreasing increment of SD, and thus paradoxically lower restenosis risk, was also not demonstrated in the empirical observations of trials reporting late loss to date.
The utility of late loss as a measure of restenosis in clinical trials is supported by the following observations. First, if a uniform stent diameter at deployment is assumed, late loss is a consistent measure of the magnitude of the renarrowing process. Although it is estimated from 2 angiograms, usually separated over a 6- to 9-month period, the estimates have shown a consistency of correlation between its value and restenosis rate (Figure 1). Because the renarrowing process within a stent is due solely to the hyperplastic component, with no contribution from the vascular contraction component seen in nonstent procedures, late loss is an intuitive measure of neointimal hyperplasia, the pathological target of drug-eluting stents.11
Second, late loss is a powerful measure of binary restenosis proclivity. The predictable monotonic relationship of late loss to restenosis allows estimation of restenosis probability in a given study population with known reference vessel diameter and poststent MLD. Observed results (Figure 1) and our model (Figure 4) both show substantial increased risk of binary angiographic restenosis when observed mean late loss increases over the range of 0.1- to 0.7-mm mean in-stent late loss. This generalized model of late loss is not a regression of late loss on binary restenosis but rather an empirical model relating mean values of late loss to their predictable distributions. Thus, the curvilinear predictive model, based on expected late loss probability densities, is similar in shape to the observed data.
Additionally, when late loss is expected to range from 0.1 to 0.7 mm, as in trials designed to compare new drug-eluting stents, late loss has greater power than binary angiographic restenosis to discriminate restenosis tendency for moderate-sized (
200 subjects) trials (Figure 3) and even more so for smaller sample sizes. Although larger pivotal trials (of
1500 subjects) would be required to power dichotomous end points such as target vessel failure, the confidence bounds of the late loss estimate are narrower than those of the binary angiographic restenosis end points at any sample size.
If there is substantial tapering of a vessel over the length of the analysis segment, then the true magnitude of neointimal hyperplasia may be underestimated by late loss. This is a particular problem for the "in-segment" measurement of late loss in which subtraction of the MLD obtained from the 5-mm "step-up" and "step-down" shoulders may systematically underestimate the amount of neointimal hyperplasia compared with the in-stent measurement. If there were significant poststent variability of MLD along the length of the stent, then the same limitation would be true for the in-stent measurement. In the present stenting trials with an average residual stenosis of 5%, however, the residual diameter is relatively uniform throughout the stent (ie, with little tapering of the stent itself); therefore, most variability in the in-stent late loss measurement is created by variations in the MLD at late follow-up, not after the procedure.
Late Loss and Clinical Trials of Drug-Eluting Stents
New drug-eluting stents are best evaluated by execution of 1 or 2 large "pivotal" randomized trials in which comparison is made to an accepted standard. Pivotal trials seek to estimate efficacy (freedom from restenosis) and detect important rare serious adverse events, such as coronary thrombosis or aneurysm formation. End points that combine safety and efficacy measures, such as target vessel failure or major adverse cardiac events (both of which include death, myocardial infarction, and clinically driven repeated coronary revascularization), have been deemed to be the most inclusive and remain important for pivotal stent trials. Large sample sizes (generally >1000 to 2000 subjects) are required for contrasting target vessel failure rates between the competing randomized arms, expected to range from 5% to 15%. In smaller pilot, nonpivotal, or follow-up trials, these end points lose power. The strong positive correlation and generalized monotonic relationship between late loss and restenosis, combined with its high statistical power, make late loss an attractive end point for nonpivotal drug-eluting stent trials.
We have estimated that even small differences in mean in-stent late loss can translate to important differences in binary restenosis (Table 3, Figure 4). Small differences in clinical restenosis are important for several reasons. First, cost-effectiveness analysis suggests that small differences in restenosis rates can translate to meaningful differences in cost.32 Second, the small differences in restenosis risk seen in pivotal study populations are likely amplified in practice, where the magnitude of clinical benefit is expected to increase with the risk profile of the target patient population.34,35 Regardless of the actual (and currently unknown) threshold of late loss difference that is clinically relevant, increasing values of late loss are associated with increasing risk of binary restenosis. These findings support the notion that late loss performance can reliably predict the restenosis propensity for new drug-eluting stents.
Limitations
We chose binary angiographic restenosis as the outcome variable used to validate late loss. Our choice of angiographic binary restenosis was based on the wide availability of this standard variable in the literature and the lack of a consistently reported clinical restenosis end point. Although the 50% diameter stenosis cutoff definition of binary angiographic restenosis may not define all patients with physiological coronary flow obstruction, it is a robust measure of clinical restenosis, with high correlation between binary angiographic restenosis and target lesion or target vessel revascularization.36 Finally, the end point percent diameter stenosis is an excellent measure of the final follow-up result in a given patient. However, percent diameter stenosis has a wide range (eg, 15% to 35% for percent diameter stenosis versus 0 to 1.0 for late loss), and its calculation and interpretation are dependent on reference vessel size, making its use in nonrandomized comparisons more problematic than that of late loss.
| Acknowledgments |
|---|
| References |
|---|
|
|
|---|
2. Nobuyoshi M, Kimura T, Nosaka H, Mioka S, Ueno K, Yokoi H, Hamasaki N, Horiuchi H, Ohishi H. Restenosis after successful percutaneous transluminal coronary angioplasty: serial angiographic follow-up of 229 patients. J Am Coll Cardiol. 1988; 12: 616623.[Abstract]
3. Serruys PW, Luijten HE, Beatt KJ, Geuskens R, de Feyter PJ, van den Brand M, Reiber JH, ten Katen HJ, van Es GA, Hugenholtz PG. Incidence of restenosis after successful coronary angioplasty: a time-related phenomenon: a quantitative angiographic study in 342 consecutive patients at 1, 2, 3, and 4 months. Circulation. 1988; 77: 361371.
4. Sonoda S, Morino Y, Ako J, Terashima M, Hassan AH, Bonneau HN, Leon MB, Moses JW, Yock PG, Honda Y, Kuntz RE, Fitzgerald PJ. Impact of final stent dimensions on long-term results following sirolimus-eluting stent implantation: serial intravascular ultrasound analysis from the SIRIUS trial. J Am Coll Cardiol. 2004; 43: 19591963.
5. Moses JW, Leon MB, Popma JJ, Fitzgerald PJ, Holmes DR, OShaughnessy C, Caputo RP, Kereiakes DJ, Williams DO, Teirstein PS, Jaeger JL, Kuntz RE. Sirolimus-eluting stents versus standard stents in patients with stenosis in a native coronary artery. N Engl J Med. 2003; 349: 13151323.
6. Stone GW, Ellis SG, Cox DA, Hermiller J, OShaughnessy C, Mann JT, Turco M, Caputo R, Bergin P, Greenberg J, Popma JJ, Russel ME. A polymer-based, paclitaxel-eluting stent in patients with coronary artery disease. N Engl J Med. 2004; 350: 221230.
7. Sousa JE, Costa MA, Abizaid A, Abizaid AS, Feres F, Pinto IM, Seixas AC, Staico R, Mattos LA, Sousa AG, Falotico R, Jaeger J, Popma JJ, Serruys PW. Lack of neointimal proliferation after implantation of sirolimus-coated stents in human coronary arteries: a quantitative coronary angiography and three-dimensional intravascular ultrasound study. Circulation. 2001; 103: 192195.
8. Grube E, Silber S, Hauptmann KE, Mueller R, Buellesfeld L, Gerckens U, Russell ME. TAXUS I: six- and twelve-month results from a randomized, double-blind trial on a slow-release paclitaxel-eluting stent for de novo coronary lesions. Circulation. 2003; 107: 3842.
9. Meredith IT. ENDEAVOR I: Multicenter evaluation of ABT-578 elution from a phosphorylcholine-coated stent. Presented at: EuroPCR; May 2528, 2004; Paris, France.
10. Kuntz RE, Safian RD, Levine MJ, Reis GJ, Diver DJ, Baim DS. Novel approach to the analysis of restenosis after the use of three new coronary devices. J Am Coll Cardiol. 1992; 19: 14931499.[Abstract]
11. Post MJ, Borst C, Kuntz RE. The relative importance of arterial remodeling compared with intimal hyperplasia in lumen renarrowing after balloon angioplasty: a study in the normal rabbit and the hypercholesterolemic Yucatan micropig. Circulation. 1994; 89: 28162821.
12. Kuntz RE, Safian RD, Carrozza JP, Fishman RF, Mansour M, Baim DS. The importance of acute luminal diameter in determining restenosis after coronary atherectomy or stenting. Circulation. 1992; 86: 18271835.
13. Mauri L, Orav EJ, OMalley AJ, Moses JW, Leon MB, Holmes DR Jr, Teirstein PS, Schofer J, Breithardt G, Cutlip DE, Kereiakes DJ, Shi C, Firth BG, Donohoe DJ, Kuntz RE. Relationship of late loss in lumen diameter to coronary restenosis in sirolimus-eluting stents. Circulation. 2005; 111: 321327.
14. Baim DS, Cutlip DE, Midei M, Linnemeier TJ, Schreiber T, Cox D, Kereiakes D, Popma JJ, Robertson L, Prince R, Lansky AJ, Ho KK, Kuntz RE. Final results of a randomized trial comparing the MULTI-LINK stent with the Palmaz-Schatz stent for narrowings in native coronary arteries. Am J Cardiol. 2001; 87: 157162.[CrossRef][Medline] [Order article via Infotrieve]
15. Baim DS, Cutlip DE, OShaughnessy CD, Hermiller JB, Kereiakes DJ, Giambartolomei A, Katz S, Lansky AJ, Fitzpatrick M, Popma JJ, Ho KK, Leon MB, Kuntz RE. Final results of a randomized trial comparing the NIR stent to the Palmaz-Schatz stent for narrowings in native coronary arteries. Am J Cardiol. 2001; 87: 152156.[CrossRef][Medline] [Order article via Infotrieve]
16. Serruys PW, de Jaegere P, Kiemeneij F, Macaya C, Rutsch W, Heyndrickx G, Emanuelsson H, Marco J, Legrand V, Materne P, Berlardi J, Sigwart U, Colombo A, Goy J, van den Heuvel P, Delcan J, Morel MA, for the BENESTENT Study Group. A comparison of balloon-expandable-stent implantation with balloon angioplasty in patients with coronary artery disease. N Engl J Med. 1994; 331: 489495.
17. Serruys PW, van Hout B, Bonnier H, Legrand V, Garcia E, Macaya C, Sousa E, van der Giessen W, Colombo A, Seabra-Gomes R, Kiemeneij F, Ruygrok P, Ormiston J, Emanuelsson H, Fajadet J, Haude M, Klugmann S, Morel MA. Randomised comparison of implantation of heparin-coated stents with balloon angioplasty in selected patients with coronary artery disease (BENESTENT II). Lancet. 1998; 352: 673681.[CrossRef][Medline] [Order article via Infotrieve]
18. Lansky AJ, Roubin GS, OShaughnessy CD, Moore PB, Dean LS, Raizner AE, Safian RD, Zidar JP, Kerr JL, Popma JJ, Mehran R, Kuntz RE, Leon MB. Randomized comparison of GR-II stent and Palmaz-Schatz stent for elective treatment of coronary stenoses. Circulation. 2000; 102: 13641368.
19. Schofer J, Schluter M, Gershlick AH, Wijns W, Garcia E, Schampaert E, Breithardt G. Sirolimus-eluting stents for treatment of patients with long atherosclerotic lesions in small coronary arteries: double-blind, randomised controlled trial (E-SIRIUS). Lancet. 2003; 362: 10931099.[CrossRef][Medline] [Order article via Infotrieve]
20. Colombo A, Drzewiecki J, Banning A, Grube E, Hauptmann K, Silber S, Dudek D, Fort S, Schiele F, Zmudka K, Guagliumi G, Russell ME. Randomized study to assess the effectiveness of slow- and moderate-release polymer-based paclitaxel-eluting stents for coronary artery lesions. Circulation. 2003; 108: 788794.
21. Kastrati A, Schomig A, Dirschinger J, Mehilli J, Dotzer F, von Welser N, Neumann FJ, for the ISAR-SMART Study Investigators (Intracoronary Stenting or Angioplasty for Restenosis Reduction in Small Arteries). A randomized trial comparing stenting with balloon angioplasty in small vessels in patients with symptomatic coronary artery disease. Circulation. 2000; 102: 25932598.
22. Park SW, Lee CW, Hong MK, Kim JJ, Cho GY, Nah DY, Park SJ. Randomized comparison of coronary stenting with optimal balloon angioplasty for treatment of lesions in small coronary arteries. Eur Heart J. 2000; 21: 17851789.
23. Koning R, Eltchaninoff H, Commeau P, Khalife K, Gilard M, Lipiecki J, Coste P, Bedossa M, Lefevre T, Brunel P, Morice MC, Maillard L, Guyon P, Puel J, Cribier A. Stent placement compared with balloon angioplasty for small coronary arteries: in-hospital and 6-month clinical and angiographic results. Circulation. 2001; 104: 16041608.
24. Doucet S, Schalij MJ, Vrolix MC, Hilton D, Chenu P, de Bruyne B, Udayachalerm W, Seth A, Bilodeau L, Reiber JH, Harel F, Lesperance J. Stent placement to prevent restenosis after angioplasty in small coronary arteries. Circulation. 2001; 104: 20292033.
25. Moer R, Myreng Y, Molstad P, Albertsson P, Gunnes P, Lindvall B, Wiseth R, Ytre-Arne K, Kjekshus J, Golf S. Stenting in Small Coronary Arteries (SISCA) trial: a randomized comparison between balloon angioplasty and the heparin-coated beStent. J Am Coll Cardiol. 2001; 38: 15981603.
26. Morice MC, Serruys PW, Sousa JE, Fajadet J, Ban Hayashi E, Perin M, Colombo A, Schuler G, Barragan P, Guagliumi G, Molnar F, Falotico R. A randomized comparison of a sirolimus-eluting stent with a standard stent for coronary revascularization. N Engl J Med. 2002; 346: 17731780.
27. Fischman DL, Leon MB, Baim DS, Schatz RA, Savage MP, Penn I, Detre K, Veltri L, Ricci D, Nobuyoshi M, Cleman M, Heuser R, Almond D, Teirstein PS, Fish RD, Colombo A, Brinker J, Moses J, Shaknovich A, Hirshfeld J, Bailey S, Ellis S, Rake R, Goldberg S, for the Stent Restenosis Study Investigators. A randomized comparison of coronary-stent placement and balloon angioplasty in the treatment of coronary artery disease. N Engl J Med. 1994; 331: 496501.
28. Kereiakes DJ, Cox DA, Hermiller JB, Midei MG, Bachinsky WB, Nukta ED, Leon MB, Fink S, Marin L, Lansky AJ. Usefulness of a cobalt chromium coronary stent alloy. Am J Cardiol. 2003; 92: 463466.[CrossRef][Medline] [Order article via Infotrieve]
29. Kastrati A, Mehilli J, Dirschinger J, Dotzer F, Schuhlen H, Neumann FJ, Fleckenstein M, Pfafferott C, Seyfarth M, Schomig A. Intracoronary Stenting and Angiographic Results: Strut Thickness Effect on Restenosis Outcome (ISAR-STEREO) trial. Circulation. 2001; 103: 28162821.
30. Grube E, Sonoda S, Ikeno F, Honda Y, Kar S, Chan C, Gerckens U, Lansky AJ, Fitzgerald PJ. Six- and twelve-month results from first human experience using everolimus-eluting stents with bioabsorbable polymer. Circulation. 2004; 109: 21682171.
31. Gershlick A, De Scheerder I, Chevalier B, Stephens-Lloyd A, Camenzind E, Vrints C, Reifart N, Missault L, Goy JJ, Brinker JA, Raizner AE, Urban P, Heldman AW. Inhibition of restenosis with a paclitaxel-eluting, polymer-free coronary stent: the European evaLUation of pacliTaxel Eluting Stent (ELUTES) trial. Circulation. 2004; 109: 48793.
32. Schampaert E, Cohen EA, Schluter M, Reeves F, Traboulsi M, Title LM, Kuntz RE, Popma JJ. The Canadian study of the sirolimus-eluting stent in the treatment of patients with long de novo lesions in small native coronary arteries (C-SIRIUS). J Am Coll Cardiol. 2004; 43: 11101115.
33. Lansky AJ, Costa RA, Mintz GS, Tsuchiya Y, Midei M, Cox DA, OShaughnessy C, Applegate RA, Cannon LA, Mooney M, Farah A, Tannenbaum MA, Yakubov S, Kereiakes DJ, Wong SC, Kaplan B, Cristea E, Stone GW, Leon MB, Knopf WD, ONeill WW. Nonpolymer-based paclitaxel-coated coronary stents for the treatment of patients with de novo coronary lesions: angiographic follow-up of the DELIVER clinical trial. Circulation. 2004; 109: 19481954.
34. Hong MK, Park SJ. Long-term benefits of drug-eluting stents for challenging patients. Presented at: Transcatheter Cardiovascular Therapeutics; October 2630, 2004; Washington, DC.
35. Kastrati A, Mehilli J, von Beckerath N, Dibra A, Hausleiter J, Pache J, Schuhlen H, Schmitt C, Dirschinger J, Schomig A. Sirolimus-eluting stent or paclitaxel-eluting stent vs balloon angioplasty for prevention of recurrences in patients with coronary in-stent restenosis: a randomized controlled trial. JAMA. 2005; 293: 165171.
36. Cutlip DE, Chauhan MS, Baim DS, Ho KK, Popma JJ, Carrozza JP, Cohen DJ, Kuntz RE. Clinical restenosis after coronary stenting: perspectives from multicenter clinical trials. J Am Coll Cardiol. 2002; 40: 20822089.
37. Lachin JM. Biostatistical methods: the assessment of relative risks. In: Wiley Series in Probability and Statistics. New York, NY: John Wiley & Sons Inc; 2000: 455456.
This article has been cited by other articles:
![]() |
R A Byrne, S Eberle, A Kastrati, A Dibra, G Ndrepepa, R Iijima, J Mehilli, and A Schomig Distribution of angiographic measures of restenosis after drug-eluting stent implantation Heart, October 1, 2009; 95(19): 1572 - 1578. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Latib, J. Cosgrave, and A. Colombo Bimodal distribution of angiographic measures of restenosis: what does it mean? Heart, October 1, 2009; 95(19): 1556 - 1558. [Full Text] [PDF] |
||||
![]() |
J. J. Wykrzykowska, P. W. Serruys, Y. Onuma, T. de Vries, G.-A. van Es, P. Buszman, A. Linke, T. Ischinger, V. Klauss, R. Corti, et al. Impact of Vessel Size on Angiographic and Clinical Outcomes of Revascularization With Biolimus-Eluting Stent With Biodegradable Polymer and Sirolimus-Eluting Stent With Durable Polymer: The LEADERS Trial Substudy J. Am. Coll. Cardiol. Intv., September 1, 2009; 2(9): 861 - 870. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Verheye, P. Agostoni, K. D. Dawkins, J. Dens, W. Rutsch, D. Carrie, J. Schofer, C. Lotan, C. L. Dubois, S. A. Cohen, et al. The GENESIS (Randomized, Multicenter Study of the Pimecrolimus-Eluting and Pimecrolimus/Paclitaxel-Eluting Coronary Stent System in Patients with De Novo Lesions of the Native Coronary Arteries) Trial J. Am. Coll. Cardiol. Intv., March 1, 2009; 2(3): 205 - 214. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. R. Costa Jr, A. Abizaid, R. Costa, F. Feres, L. F. Tanajura, A. Abizaid, L. A. Mattos, R. Staico, D. Siqueira, A. G.M.R. Sousa, et al. Preliminary Results of the Hydroxyapatite Nonpolymer-Based Sirolimus-Eluting Stent for the Treatment of Single De Novo Coronary Lesions: A First-in-Human Analysis of a Third-Generation Drug-Eluting Stent System J. Am. Coll. Cardiol. Intv., October 1, 2008; 1(5): 545 - 551. [Abstract] [Full Text] [PDF] |
||||
![]() |
S.-W. Lee, S.-W. Park, Y.-H. Kim, S.-C. Yun, D.-W. Park, C. W. Lee, M.-K. Hong, H.-S. Kim, J.-K. Ko, J.-H. Park, et al. Drug-Eluting Stenting Followed by Cilostazol Treatment Reduces Late Restenosis in Patients With Diabetes Mellitus: The DECLARE-DIABETES Trial (A Randomized Comparison of Triple Antiplatelet Therapy With Dual Antiplatelet Therapy After Drug-Eluting Stent Implantation in Diabetic Patients) J. Am. Coll. Cardiol., March 25, 2008; 51(12): 1181 - 1187. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. L. van der Hoeven, S.-S. Liem, J. W. Jukema, N. Suraphakdee, H. Putter, J. Dijkstra, D. E. Atsma, M. Bootsma, K. Zeppenfeld, P. V. Oemrawsingh, et al. Sirolimus-eluting stents versus bare-metal stents in patients with ST-segment elevation myocardial infarction: 9-month angiographic and intravascular ultrasound results and 12-month clinical outcome results from the MISSION! Intervention Study. J. Am. Coll. Cardiol., February 12, 2008; 51(6): 618 - 626. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. J. Pocock, A. J. Lansky, R. Mehran, J. J. Popma, M. P. Fahy, Y. Na, G. Dangas, J. W. Moses, T. Pucelikova, D. E. Kandzari, et al. Angiographic surrogate end points in drug-eluting stent trials: a systematic evaluation based on individual patient data from 11 randomized, controlled trials. J. Am. Coll. Cardiol., January 1, 2008; 51(1): 23 - 32. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. J. Kereiakes The Emperor's New Clothes: Another Cypher Versus Taxus Post-Hoc Meta-Analysis J. Am. Coll. Cardiol., October 2, 2007; 50(14): 1381 - 1385. [Full Text] [PDF] |
||||
![]() |
M. Togni, S. Eber, J. Widmer, M. Billinger, P. Wenaweser, S. Cook, R. Vogel, C. Seiler, F. R. Eberli, W. Maier, et al. Impact of Vessel Size on Outcome After Implantation of Sirolimus-Eluting and Paclitaxel-Eluting Stents: A Subgroup Analysis of the SIRTAX Trial J. Am. Coll. Cardiol., September 18, 2007; 50(12): 1123 - 1131. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Moreno, C. Fernandez, A. Sanchez-Recalde, G. Galeote, L. Calvo, F. Alfonso, R. Hernandez, R. Sanchez-Aquino, D. J. Angiolillo, S. Villarreal, et al. Clinical impact of in-stent late loss after drug-eluting coronary stent implantation Eur. Heart J., July 1, 2007; 28(13): 1583 - 1591. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Camenzind, P. G. Steg, and W. Wijns A Cause for Concern Circulation, March 20, 2007; 115(11): 1440 - 1455. [Full Text] [PDF] |
||||
![]() |
E. Camenzind Treatment of In-Stent Restenosis -- Back to the Future? N. Engl. J. Med., November 16, 2006; 355(20): 2149 - 2151. [Full Text] [PDF] |
||||
![]() |
Y.-H. Kim, S.-W. Park, S.-W. Lee, D.-W. Park, S.-C. Yun, C. W. Lee, M.-K. Hong, H.-S. Kim, J.-K. Ko, J.-H. Park, et al. Sirolimus-Eluting Stent Versus Paclitaxel-Eluting Stent for Patients With Long Coronary Artery Disease Circulation, November 14, 2006; 114(20): 2148 - 2153. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. H. Strauss All Drug-Eluting Stents Are Equal, But Some Drug-Eluting Stents Are More Equal Than Others J. Am. Coll. Cardiol., October 3, 2006; 48(7): 1310 - 1311. [Full Text] [PDF] |
||||
![]() |
S. Elezi, A. Dibra, J. Mehilli, J. Pache, R. Wessely, A. Schomig, and A. Kastrati Vessel Size and Outcome After Coronary Drug-Eluting Stent Placement: Results From a Large Cohort of Patients Treated With Sirolimus- or Paclitaxel-Eluting Stents J. Am. Coll. Cardiol., October 3, 2006; 48(7): 1304 - 1309. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Saia, G. Piovaccari, A. Manari, A. Santarelli, A. Benassi, E. Aurier, P. Sangiorgio, F. Tarantino, G. Geraci, G. Vecchi, et al. Clinical Outcomes for Sirolimus-Eluting Stents and Polymer-Coated Paclitaxel-Eluting Stents in Daily Practice: Results From a Large Multicenter Registry J. Am. Coll. Cardiol., October 3, 2006; 48(7): 1312 - 1318. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Fajadet, W. Wijns, G.-J. Laarman, K.-H. Kuck, J. Ormiston, T. Munzel, J. J. Popma, P. J. Fitzgerald, R. Bonan, R. E. Kuntz, et al. Randomized, Double-Blind, Multicenter Study of the Endeavor Zotarolimus-Eluting Phosphorylcholine-Encapsulated Stent for Treatment of Native Coronary Artery Lesions: Clinical and Angiographic Results of the ENDEAVOR II Trial Circulation, August 22, 2006; 114(8): 798 - 806. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Tung, S. Kaul, G. A. Diamond, and P. K. Shah Narrative review: drug-eluting stents for the management of restenosis: a critical appraisal of the evidence. Ann Intern Med, June 20, 2006; 144(12): 913 - 919. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Kastrati, A. Dibra, J. Mehilli, S. Mayer, S. Pinieck, J. Pache, J. Dirschinger, and A. Schomig Predictive Factors of Restenosis After Coronary Implantation of Sirolimus- or Paclitaxel-Eluting Stents Circulation, May 16, 2006; 113(19): 2293 - 2300. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. R. Dixon, C. L. Grines, and W. W. O'Neill The Year in Interventional Cardiology J. Am. Coll. Cardiol., April 18, 2006; 47(8): 1689 - 1706. [Full Text] [PDF] |
||||
![]() |
R. Seabra-Gomes Percutaneous coronary interventions with drug eluting stents for diabetic patients. Heart, March 1, 2006; 92(3): 410 - 419. [Full Text] [PDF] |
||||
![]() |
J. Mehilli, A. Dibra, A. Kastrati, J. Pache, J. Dirschinger, A. Schomig, and for the Intracoronary Drug-Eluting Stenting to Abr Randomized trial of paclitaxel- and sirolimus-eluting stents in small coronary vessels Eur. Heart J., February 1, 2006; 27(3): 260 - 266. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. J. Kereiakes, R. E. Kuntz, L. Mauri, and M. W. Krucoff Reply J. Am. Coll. Cardiol., January 17, 2006; 47(2): 471 - 471. [Full Text] [PDF] |
||||
![]() |
P. Agostoni, L. Mauri, E. J. Orav, and R. E. Kuntz Letter Regarding Article by Mauri et al, "Late Loss in Lumen Diameter and Binary Restenosis for Drug-Eluting Stent Comparison" * Response Circulation, December 13, 2005; 112(24): e358 - e358. [Full Text] [PDF] |
||||
![]() |
L. Mauri, E. J. Orav, S. C. Candia, D. E. Cutlip, and R. E. Kuntz Robustness of Late Lumen Loss in Discriminating Drug-Eluting Stents Across Variable Observational and Randomized Trials Circulation, November 1, 2005; 112(18): 2833 - 2839. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2005 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |