Implantable Cardioverter-Defibrillators and Biventricular Pacing Therapy
Alan Kadish, MD;
Mandeep Mehra, MD
From the Division of Cardiology and Bluhm Cardiovascular Institute, Feinberg School of Medicine, Chicago, Ill (A.K.), and the Division of Cardiology, University of Mayland, Baltimore (M.M.).
Correspondence to Alan Kadish, MD, 251 E Huron, Suite 8-536, Chicago, IL 60611. E-mail a-kadish{at}northwestern.edu
Key Words: death, sudden defibrillation electrophysiology tachyarrhythmias
 |
Introduction
|
|---|
Although the age-adjusted mortality from heart disease has declined
in the United States, cardiovascular disease remains the No.
1 cause of death.
1,2 Patients with cardiac disease generally
die of one of 2 causes: sudden, unexpected cardiac death or
progressive heart failure. Chronic heart failure (CHF) has become
an epidemic in the United States. Implantable cardioverter-defibrillator
(ICD) therapy has had a major impact on the treatment of heart
failure in the United States. The ICD has been shown to decrease
mortality relative to the best medical therapy in patients who
have survived an episode of sustained ventricular tachycardia
(VT) or ventricular fibrillation (VF).
3 Unfortunately, only
a small minority of patients who experience an out-of-hospital
cardiac arrest in the United States are successfully resuscitated.
4 Thus, ICD therapy has been applied for the primary prevention
of sudden death in patients at high risk of cardiac arrest.
Once thought to be a "bystander" curiosity in the progression
of heart failure, electrical-conduction disturbances are now
recognized to be important causes of left ventricular (LV) dysfunction.
Cardiac resynchronization therapy (CRT) can restore more-normal
electrical contraction and, when combined with defibrillation
(CRTD), can have a major impact on the mortality and morbidity
of heart failure. Although there remain many limitations and
challenges to the appropriate application of ICDs or CRTD, there
is no question that device therapy has had a major impact on
the management of patients with LV dysfunction.
 |
ICD Therapy
|
|---|
The ICD was developed in the 1970s as a device to detect and
automatically terminate ventricular tachyarrhythmias.
5,6 Early
ICDs were large devices that required a thoracotomy to implant.
Improvements in batteries, capacitors, and microprocessors
7 have resulted in the rapid evolution of ICD technology. Current-generation
ICDs are small enough to be easily implanted in the pectoral
region and can perform a variety of sophisticated functions,
including atrial
8 and ventricular defibrillation,
6 antitachycardia
pacing (ATP),
1 backup bradycardia pacing, electrogram storage,
and biventricular pacing.
9 ICDs can be implanted under local
anesthesia and conscious sedation in a procedure that often
lasts <1 hour.
The development of the ICD was a revolutionary advance in the treatment of heart disease. Its development represented something of a paradox. It was the work of a small group of individuals, but its evolution has represented an extraordinary collaboration between technology, industry, and medicine. The ICD was conceived and developed by Dr Michel Mirowski (Figure 1) and Dr Morton Mower working in a small laboratory at a hospital affiliated with a major medical center. Dr Mirowskis thoughts were widely criticized in the mainstream medical literature as being unrealistic. Dr Mirowskis vision and perseverance in the face of criticism were remarkable, and the ICD represented the culmination of the work of lone visionaries. In contrast, the development of the ICD has also been emblematic of the tremendous change that advanced technology has produced in medical care. The ICD incorporates technological advances, such as tiny microprocessors with hundreds of programmable features, batteries, and capacitors specifically designed for the ICD and novel miniaturization techniques. These advances would have been completely unimaginable 2 decades ago and exemplify the role that industry and technology play in modern medical care.
 |
Basic ICD Components and Function
|
|---|
The basic components of ICDs have not changed in the past 20
years.
7,9 ICDs contain batteries, capacitors, a microprocessor,
and a "header" that contains a site to attach leads. Although
ICDs have millions of programmable features, advanced microprocessor
technology has allowed these features to be seamlessly incorporated
without an increase in device size. ICD leads are constructed
in a fashion similar to pacemaker leads, except that a shocking
coil or coils are also included. In a single-chamber ICD, one
lead is placed in the right ventricle at a site that can pace
and also defibrillate the heart. The ICD generator and microprocessors
are encased in a titanium can that may be as small as 30 cm
3 and can weigh as little as 50 g.
The essential features of ICD function include detecting tachyarrhythmia, classification of a tachyarrhythmia as a treatable arrhythmia, delivering therapy, monitoring of heart rhythm after treatment, and storage of diagnostic results from the episode. Sensing of tachyarrhythmias occurs via a dedicated bipole of small electrodes near the catheter tip (dedicated sensing) or via a bipole from a tip electrode to a coil electrode that is also used to deliver defibrillation shocks (integrated sensing). After appropriate detection, ATP may be initiated for appropriate rhythm disturbances, or a defibrillation shock may be delivered between 2 coil electrodes or between a coil and the ICD generator box (active can). Arrhythmia detection can be based purely on heart rate, as in early-generation devices, or on more sophisticated algorithms. An example of appropriate arrhythmia detection is shown in Figure 2. After arrhythmia detection and fulfillment of appropriate diagnostic criteria (A), therapy is initiated. In B, ATP was used in an attempt to terminate an episode of VT. ATP was not successful. In C, electrical cardioversion occurred after the failure of ATP to terminate the arrhythmia.

View larger version (99K):
[in this window]
[in a new window]
|
Figure 2. Electrograms recorded during interrogation of ICD from patient who had sustained arrhythmia episode. Each of 3 panels is similarly organized. Top trace shows atrial electrogram. Second trace shows ventricular electrogram from rate-sensing lead. Third panel shows marker channel. In A, tachycardia is detected as indicated by markers labeled TS (tachycardia sense). VT is present at cycle length of 370 ms, which is in VT detection zone. Note that diagnosis is confirmed by presence of ventriculoatrial dissociation and sinus rhythm in atrium. In B, multiple trains are delivered but fail to terminate VT. ICD is programmed to deliver cardioversion shock after failed ATP events. C shows cardioversion shock, resulting in delayed VT and atrioventricular paced rhythm that follows. Abbreviations are as defined in text.
|
|
 |
ICD Features and Types
|
|---|
ICDs can be divided into 3 broad categories: single-chamber
ICDs, dual-chamber ICDs, and CRT ICDs. Single-chamber ICDs have
only a right ventricular lead, dual-chamber ICDs have a right
atrial and a right ventricular lead, and CRTDs also have an
LV lead that is placed via the coronary sinus (
Figure 3) or
epicardially. At present, no firm guidelines exist for implantation
of single-versus dual-chamber ICDs. Although several clinical
trials have compared single-chamber with dual-chamber pacemakers,
10 few trials have examined comparisons between single-chamber
and dual-chamber defibrillators. Although dual-chamber defibrillators
provide additional flexibility when added to old diagnostic
information,
11 clear clinical benefit has not been demonstrated
in prospective, randomized trials, except for patients who have
bradycardia pacing indications.
12 It has also become clear that
electrical disturbances may be a cause of CHF, rather than just
an effect of CHF. This results in progressive LV dysfunction,
not due to permanently necrotic myocardium but due rather to
electrical dyssynchrony.
13,14 CRT, by restoring mechanical synchrony,
can reverse signs and symptoms of heart failure and in some
patients improve ejection fraction.
15,16

View larger version (84K):
[in this window]
[in a new window]
|
Figure 3. Fluoroscopic images of CRT device insertion. Both images are right interior oblique images. In both panels, right ventricular ICD lead has already been placed in right ventricular apex. Figure 3A shows angiogram of coronary sinus obtained with balloon occlusion catheter. Anterolateral branches of coronary sinus are identified. Figure 3B shows LV lead placed within anterolateral coronary sinus branch. Abbreviations are as defined in text.
|
|
 |
Complications
|
|---|
A number of complications have been described with the use of
ICD therapy, and these need to be carefully considered in evaluating
the risk-benefit ratio of ICD implantation for individual patients.
The most common complication of ICD therapy in almost all series
is inappropriate ICD shocks that range from 21% to 25%.
17,18 The most common cause of these shocks is atrial fibrillation,
other supraventricular tachyarrhythmias, or sinus tachycardia.
Although atrial fibrillation is not a contraindication to ICD
implantation, a careful assessment of heart rate during fibrillation
should be performed to minimize the incidence of inappropriate
shocks. Inappropriate ICD shocks due to oversensing can also
occur but are less frequent in current ICDs. Although the effects
of inappropriate shocks can be psychologically devastating in
individual patients, in most cases, changes in ICD programming
or adjustment of medical therapy can eliminate or decrease the
incidence of inappropriate shocks. More serious complications
can occur during ICD implantation, such as pneumothorax, hemothorax,
infection, lead dislodgment, and cardiac tamponade.
1721 The overall incidence of significant complications, with the
exception of inappropriate ICD shocks, should be well under
5%. Unless patients are acutely ill, have severe CHF, or are
undergoing lead extraction, death from ICD implantation should
be uncommon.
 |
Cost-Effectiveness
|
|---|
Healthcare costs in the United States have risen at a rate faster
than inflation for the past 20 years. Although ICD therapy accounts
for only a very small percentage of current healthcare costs,
the cost of ICD implantation including the device, lead or leads,
and hospitalization for ICD implantation may be as high as $30
000 to $40 000. Thus, careful analysis of ICD cost-effectiveness
is appropriate. At present, there are no hard guidelines for
cost-effectiveness application of medical therapy. However,
a carefully thought-out, case-by-case approach should be utilized
when considering expanding ICD indications to patients in whom
the clinical benefit may be small.
2225 If ICD costs can
be decreased, it will allow the therapy to be applied to larger
numbers of patients without dramatic increases in healthcare
expenditures for ICDs.
 |
ICD Therapy for Secondary Prevention of Sudden Cardiac Death
|
|---|
Initial observational studies suggested that the ICD was effective
in preventing recurrent cardiac arrest in survivors of sudden
arrest.
26,27 However, no comparative trials validating the relative
effectiveness of ICD therapy compared with other forms of therapy
were performed until the late 1990s. Three randomized, controlled
trials compared the best antiarrhythmic therapy with ICD therapy
for the treatment of ventricular tachyarrhythmias in patients
resuscitated from a cardiac arrest.
28,29 Although the trials
differed slightly in inclusion criteria and study design, the
major results were similar. In the Antiarrhythmic versus Implantable
Defibrillator trial (AVID), 1016 patients with life-threatening
arrhythmias were randomized to receive an ICD or antiarrhythmic
therapy.
19 More than 75% of the patients who were randomized
to receive antiarrhythmic therapy were treated with amiodarone.
29 There was an

30% relative decrease in overall mortality in the
group who received the ICD compared with antiarrhythmic drugs.
Absolute mortality at 2 years was 25.3% in the control (drug)
group and 18.4% in the ICD group. The Canadian Implantable Defibrillator
trial (CIDS) randomized 328 patients to receive an ICD and 331
to receive amiodarone. There was a 20% relative risk reduction
and a 2% absolute annual mortality benefit in the group who
received an ICD. However, this difference did not reach significance.
In the Cardiac Arrest Study Hamburg (CASH), patients were randomized
to receive an ICD, amiodarone, or metoprolol.
30 Two hundred
twenty-eight patients were included in the study. There was
an

30% reduction in mortality in the group who received the
ICD. The absolute mortality benefit was difficult to calculate
because the primary end point of the study was sudden death
mortality rather than total mortality. There was no significant
difference between the effect of amiodarone and that of metoprolol.
Although the results of the study were of borderline significance,
the patient population was small. Taken together and evaluated
in a meta-analysis, the results suggest that ICDs are superior
to medical therapy for the treatment of patients who have experienced
a cardiac arrest.
3 Subgroup analysis has suggested that some
patients who have survived a cardiac arrest, such as those >65
years or with well-preserved ventricular function, may not have
as great a benefit from ICD therapy.
3136 However, trials
that are adequately powered to detect differences among such
subgroups have not been completed.
36 Although the data suggesting
lower mortality in patients who receive an ICD are clear, there
are many patients who experience a cardiac arrest in whom less-aggressive
medical therapy may be appropriate because of an overall poor
prognosis or comorbid conditions.
 |
ICD Therapy for Primary Prevention of Sudden Cardiac Death
|
|---|
Risk Stratification
A multifactorial approach to the prevention of sudden cardiac
death is clearly needed. Components of this approach include
modifying risk factors for coronary disease, identifying genetic
predispositions to sudden death, improving community-based cardiopulmonary
resuscitation training, and using technological advances, such
as the automatic external defibrillator
37 and the wearable defibrillator.
Despite substantial efforts, however, survival rates from cardiac
arrest in the United States remain low.
1 Although survival rates
>30% were briefly achieved in some local areas such as Seattle
and continue to be achieved in areas such as casinos and airports,
the overall survival rate from cardiac arrest in the United
States is probably <10%.
38 Thus, identifying patients at
high risk for cardiac arrest and treating them with pharmacological
or nonpharmacological therapy to prevent cardiac arrest is appropriate.
The most important risk factor for determining the risk of cardiac arrest in patients with acquired heart diseases is the extent of myocardial damage.39 The risk of cardiac arrest increases moderately with ejection fractions <40% and more significantly with ejection fractions <30%. Other factors that have been shown to be associated with the risk of sudden cardiac death include the extent of heart failure, QRS width, and neurohormone levels. Some studies have suggested that the inducibility of ventricular arrhythmias at electrophysiological testing, heart rate variability, signal-averaged ECGs, and T-wave alternans may provide independent prognostic information. However, for many of these risk-stratification tests, high positive and negative predictive values have not been consistently obtained. A multifactorial model40 may be scientifically most appropriate to determine the risk of cardiac arrest, but it is unclear whether such a model can be widely applied clinically. Thus, selection of patients for trials examining the efficacy of the ICD in the primary prevention of sudden death has typically been based on ejection fraction and, at most, 1 or 2 other risk-stratification tests.
Clinical Trials
Several recently completed, large-scale trials have examined the use of ICD therapy in patients with underlying structural heart disease at risk for sudden death. Six large trials have examined the use of the ICD in patients with coronary artery disease. The CABG-PATCH trial41 randomized 900 patients who were undergoing cardiac surgery for adjunctive epicardial ICD placement. The inclusion criteria for the CABG-PATCH trial were the presence of LV dysfunction and an abnormal signal-averaged ECG. There was no difference in survival between the patients who received the epicardial ICD and those who did not. The negative results of this study could potentially be attributed to deleterious effects of epicardial ICD implantation or to improvements in ventricular function occurring as a result of bypass surgery. In contrast, 4 other trials that used the ICD for the primary prevention of sudden death in patients with chronic coronary disease and LV dysfunction all showed benefits of the ICD. The MADIT-I study, published in 1996, randomized patients with an LV ejection fraction of
35%, coronary artery disease, nonsustained VT, and inducible VT to receive an ICD or standard therapy.42 Only 196 patients were enrolled. There were 15 deaths in the ICD arm and 39 in the conventional-therapy arm. The hazard ratio for overall mortality was 0.46 (P=0.009). Although the study was small and limited by an imbalance in ß-blocker use in the 2 groups, the dramatic clinical benefit of the ICD in this patient population led to approval of ICD use for this indication. The Multi-center, Unsustained Tachycardia Trial (MUSTT) had a complex study design.43 Two thousand two hundred two patients were screened and 704 were randomized to be treated with an electrophysiology-guided approach or a medical approach. No difference was seen between the 2 arms. However, a subpopulation who was randomized to the electrophysiologically guided approach received an ICD. The risk of cardiac arrest or death from arrhythmia among patients who received the ICD was significantly lower than in those without the ICD (hazard ratio, 0.24; P<0.001). Analyzing the effect of the ICD in a trial that was designed to compare an electrophysiologically to a nonelectrophysiologically guided approach is complex. However, the dramatic reduction in mortality seen and the consistency with the results of the MADIT-I trial were important observations that the ICD would benefit patients with coronary disease and LV dysfunction. The MADIT-II trial did not use electrophysiological testing or signal-averaged ECGs to stratify patients for enrollment.18 Patients were randomized if they had a prior myocardial infarction and an LV ejection fraction
30%. The crude mortality rate was 19.8% in the control arm and 14.2% in the ICD arm. The hazard ratio for reduction of all-cause mortality was 0.69 (P=0.016). These results confirmed that patients with coronary disease and LV dysfunction would benefit from the ICD for the primary prevention of sudden death. One limitation to these trials is that in some, medical therapy was not optimal by current standards. The most recent trial examining the effectiveness of the ICD in the primary prevention of sudden death in patients without coronary artery disease was the Sudden Cardiac Death in Heart Failure Trial (SCD-HeFT), which has been presented in abstract form.44 This trial randomized patients with all types of LV dysfunction to an ICD, placebo, or conventional therapy. The results in patients with nonischemic cardiomyopathy are discussed in the next section. Approximately 1300 patients with coronary disease were randomized. Amiodarone had no effect on mortality. The ICD reduced overall mortality by 23%. The ICD had a significant benefit in the overall patient population. In patients with coronary disease, the hazard ratio crossed 1, but an impressive trend to reduction in overall mortality was seen.
One study on ICD therapy soon after infarction failed to show a benefit. In the Defibrillator in Acute Myocardial infarction Trial (DINAMIT), 675 patients were randomized to receive an ICD or standard medical therapy.45 Patients were randomized 4 to 40 days after myocardial infarction and had LV dysfunction and abnormal heart rate variability. Survival was similar in the 2 groups. The reasons for the failure to demonstrate benefits of prophylactic ICD implantation are not yet fully apparent but suggest that ICD implantation soon after myocardial infarction does not yield dramatic clinical benefit. Taken together, the results of these trials suggest that the transvenous ICD substantially reduces mortality in patients with chronic coronary disease and severe LV dysfunction.
Although small trials failed to show ICD benefit in patients with nonischemic dilated cardiomyopathy,30,46 2 large, multicenter, randomized trials have been performed. The Defibrillators in Non Ischemic Cardiomyopathy Treatment Evaluation (DEFINITE) study randomized 458 subjects with nonischemic dilated cardiomyopathy and LV ejection fractions <36% to receive the best medical therapy or the best medical therapy plus an ICD.17 After 2 years, 14.1% of the patients in the standard-therapy group and 7.9% of the patients in the ICD group had died of all causes (6.2% absolute mortality reduction, P=0.08). Although statistical significance was not reached in the primary end point, the results of the study did suggest that the ICD might have utility in selected patients with nonischemic dilated cardiomyopathy. In the SCD-HeFT trial, >1100 patients with nonischemic cardiomyopathy were randomized into 3 arms.44 There was a 27% relative risk reduction in patients who received an ICD. The difference in patients with nonischemic cardiomyopathy did not quite reach significance, but a strong trend to mortality reduction was observed.
The results of these studies suggest that in judiciously selected patients with significant LV dysfunction that persists despite optimal medical therapy, ICD implantation may be appropriate. A group with moderately high risk for cardiac arrest in whom ICD implantation may be justified is probably defined by those patients with ejection fractions
35%. In patients with less severe ventricular function, additional assessments such as electrophysiological or noninvasive testing may be appropriate to further risk-stratify patients.
Primary Prevention of Sudden Death in Other Diseases
A number of diseases other than dilated cardiomyopathy have been associated with an increased incidence of sudden cardiac death. These include inherited diseases of ion channels, such as long-QT syndrome,47 Brugada syndrome,48 and catecholaminergic VT.49 In addition, other structural heart diseases, such as right ventricular dysplasia,50 hypertrophic cardiomyopathy,51 and certain types of congenital heart disease, may be associated with an increased risk of sudden death. The frequency of these conditions is not as high as that of ischemic or nonischemic dilated cardiomyopathy, making prospective, randomized, control trials difficult to perform. Nonrandomized, observational studies suggest that subgroups of high-risk patients may benefit from ICD therapy. One randomized trial was performed in Thailand on patients with an inherited risk of sudden cardiac death (DEBUT).52 In that trial, patients with an inherited syndrome of sudden death that probably represents the Brugada syndrome were randomized to receive ß-blockers or an ICD. Patients who received an ICD had better survival.
In the absence of large-scale trials for most of these conditions, risk-stratification algorithms based on retrospective studies have been created in an attempt to help select appropriate patients for ICD therapy. Although criteria suggesting an increased risk differ among different conditions, a family history of sudden death in almost all of the diseases mentioned earlier appears to be associated with an increased risk of sudden death. In addition, there are disease-specific markers, such as septal wall thickness in hypertrophic cardiomyopathy,51 the length of the QT interval in the inherited long-QT syndrome,47 and the presence of persistent rather than intermittent right precordial ST-segment elevation in Brugada syndrome,48 that appear to portend an increased risk of sudden death and may make aggressive use of the ICD for the primary prevention of sudden death appropriate.
 |
Cardiac Resynchronization Therapy
|
|---|
Physiologically, the presence of electrical-conduction defects
in heart failure is associated with a decrease in contractile
performance, development or prolongation of mitral regurgitation,
and wasted cardiac work as a result of development of mechanical
asynchrony.
53 Intraventricular asynchrony, characterized by
regional contractile phase delay in the anteroseptal and posterolateral
walls of the LV, and interventricular asynchrony, characterized
by a phase delay in contraction between the right and left ventricles,
are the typical manifested consequences of electrical remodeling
in the cardiac conduction system. These electrical alterations
translate into abnormal myocardial metabolism, redirection of
regional coronary perfusion, and cellular changes characterized
by abnormalities in cellular calcium handling and upregulation
of stress kinases.
54,55 Thus, restoring electrical synchrony
could potentially improve cardiac function by reversing these
cellular, physiological, and hemodynamic aberrations.
 |
Resynchronization Pacing in Heart Failure
|
|---|
The first clinical report of biventricular pacing is credited
to Cazeau et al
56 in 1994. That study was followed by a series
of uncontrolled studies demonstrating marked short-term improvements
in hemodynamics and symptoms with biventricular stimulation.
57,58
The European and Canadian InSync Study was designed to test the safety and efficacy of a multisite pacemaker (Medtronic InSync) in the treatment of refractory heart failure. The trial was conducted without a control group and thus, was observational. Significant improvements in exercise capacity, New York Heart Association (NYHA) functional class, and quality-of-life score were also noted at each of these time points.59 The MIRACLE (Multicenter InSync Randomized Clinical Evaluation) trial randomized 453 patients to activation or nonactivation of cardiac resynchronization in a double-blind study design.60 Inclusion criteria were NYHA class III/IV, QRS >130 ms, ejection fraction <35%, and LV end-diastolic dimension >55 mm (echocardiographic). Resynchronization improved symptoms, quality of life, and exercise capacity (6-minute walk test increased by 39 m and
O2 increased by 1 mL · min1 · kg1) while reducing LV dimension (LV end-diastolic dimension decreased by 5 mm) and improving ejection fraction. There were also improvements in the numbers of patients requiring hospitalization (8% versus 15%) or intravenous medications (7% versus 15%) for the treatment of heart failure (P<0.05). The biggest drawback of that study is that it was not subjected to an intention-to-treat analysis, because randomization assignment was made only after initial successful device implantation was secured.
 |
Long-Term Studies of Multisite Pacing (With ICD)
|
|---|
The CONTAK CD trial
61 enrolled 581 patients, who also met conventional
indications for ICD implantation, to active CRT or not. Initially,
patients were randomized (in a double-blind design) to either
biventricular or no pacing for 3 months and then crossed over
to the opposite assignment for the next 3 months (n=248), but
the scheme was later changed to a 6-month parallel control design
(n=333). The primary end point was a composite of mortality,
heart failure hospitalizations, and episodes of ventricular
arrhythmias. The primary end point favored CRT but was not statistically
significant. However, the 6-minute walk test, NYHA class, and
peak aerobic capacity were significantly improved with CRT,
particularly in patients with NYHA III/IV heart failure.
The MIRACLE ICD trial was designed in a similar manner to the original MIRACLE study, the only exception being that all patients were to have an indication for ICD.62 In addition to the primary heart failurerelated efficacy variables, this trial also assessed the efficacy of ICD function in the presence of CRT. At 6 months, significant improvements in NYHA class and peak aerobic capacity (but not in the 6-minute walk test) were noted. Furthermore, no significant differences were noted in overall heart failure status, LV size or function, hospitalizations, or survival. More important, this study demonstrated that arrhythmia recognition and ICD function were preserved, and no proarrhythmia was noted. The CRTD trials as a group showed functional improvement after biventricular pacing.
Morbidity and Mortality Studies of Multisite Pacing
COMPANION (comparison of medical therapy, pacing, and defibrillation in CHF) enrolled patients with CHF and NYHA class III or IV symptoms, despite maximized medical therapy.15 Inclusion criteria included a QRS duration >120 ms and a PR interval >150 ms. The trial had 3 treatment arms: One of 5 patients was to receive optimal pharmacological therapy, 2 of 5 were to receive optimal pharmacological therapy plus biventricular pacing, and the remaining 2 of 5 were to receive biventricular pacing plus backup ICD therapy. In contrast to all others, this study was sufficiently powered to evaluate a primary end point of combined all-cause mortality and hospitalization. Peak oxygen consumption was the other (functional) primary end point. Data were analyzed by an intention-to-treat statistical approach. Patients (n=1520) were randomized (93%), and 1080 patients were implanted with a CRT pacer or defibrillator (CRTD); 118 patients failed the initial implantation (88% implantation success for CRT and 92% for CRTD). LV lead dislodgment was seen in 2% and 2.5% in the CRT and CRTD groups, respectively. Compared with patients treated by medical therapy only, there was a statistically significant event rate reduction in the primary combined end point of total hospitalization and total mortality at 1 year in the CRT/CRTD group (odds ratio, 0.82; P=0.05 and 0.81, P=0.015, respectively), as well as in the combined end point of hospitalization for CHF and death (odds ratio, 0.64 and 0.60, respectively; P=0.05). Mortality at 1 year decreased by 24% (P=0.059, NS) in the CRT group and by 36% (P=0.003) in the CRTD group. The effects of CRT on hospitalization due to CHF appeared to be more pronounced in patients with left bundle-branch block (as opposed to intraventricular conduction defect or right bundle-branch block), patients with a longer QRS duration (>148 ms), and patients receiving ß-blockers.
Meta-Analyses of CRT
Two published meta-analyses have attempted to evaluate the collective effect of CRT on the end points of progressive heart failure mortality or total mortality. The first meta-analysis by Bradley and colleagues16 suggested that biventricular pacing decreased deaths due to progressive heart failure. The other analysis by McAlister and colleagues63 reviewed 9 trials (3216 patients) for efficacy and suggested that this technique improves LV ejection fraction and quality of life and reduces heart failure hospitalizations. Additionally, these authors suggested an improvement in survival, but this was largely accounted for by the inclusion of the ICD arm from the COMPANION trial. Thus, the meta-analyses suggest but do not definitively answer the question of mortality reduction by CRT when used in isolation from an ICD.
Unresolved Issues With CRT
The evidence supporting CRT in severe heart failure is compelling; however, device placement exerts a substantial placebo response, with evidence of improved functional capacity and quality-of-life parameters in those randomized to the control group and carries with it a significant nonresponder rate (no improvement in functional capacity and well-being). Furthermore, a marked heterogeneity of response (differences in magnitude of observed benefit between trials presumably enrolling similar heart failure populations) has also been noted. One of the most important advances in CRT must accrue from diminishing the implantation risks and redefining successful implantation to include correlation of the site and magnitude of mechanical asynchrony.64
An increasing evidence base points to the use of echocardiography (and, in particular, tissue Doppler imaging) to help identify potential responders to CRT based on assessment of mechanical asynchrony. At present, several echocardiographic methods65,66 to assess dyssynchrony have been proposed, varying from conventional to advanced approaches, primarily involving tissue Doppler imaging, strain, strain rate, and tissue tracking. It is currently unclear which of these parameters provides optimal information on asynchrony and which parameters may actually allow prospective identification of responders to CRT. It is uncertain whether optimization of atriventricular delay, interventricular resynchronization, or intraventricular resynchronization should be the principal goal of CRT.
The underlying etiology of heart failure may be important. Patients with ischemic cardiomyopathy frequently have large areas of scar tissue, and not infrequently, these are the areas of latest activity. It is unclear whether pacing of nonviable (scar) tissue results in clinical improvement.67
Another issue of importance relates to extrapolation of resynchronization therapy to mild (NYHA II) heart failure. Abraham and colleagues68 have reported on a subgroup of NYHA II patients in sinus rhythm enrolled at the Multicenter InSync ICD II trial. In that study, a significant improvement in cardiac structure and function was noted, but no improvement in exercise capacity or neurohormones was evident. As more information about mechanisms of benefit from resynchronization accumulate and patient selection evolves, further large-scale, adequately powered, clinical trials will be needed to assess effects on morbidity and mortality as well as functional outcomes in patients with less severe heart failure.
It is also unclear whether an ICD must accompany a cardiac resynchronization device (CRTD therapy) or whether it should be performed in isolation. Taking the recent SCD-HeFT results in context, one could argue that CRT therapy should ideally be accompanied by ICD placement in those with NYHA III symptoms. As in the SCD-HeFT investigation, results also suggest that ICDs might be beneficial in heart failure patients with advanced symptoms, irrespective of ischemic or nonischemic etiology.44 In the COMPANION trial, there was a trend for CRT to decrease mortality, but CRTD significantly decreased mortality. The point estimate for mortality reduction by the addition of the ICD was similar in class III and class IV heart failure. Thus, carefully selected patients with NYHA class IV heart failure may also derive benefit from concomitant ICD therapy.60
 |
Atrial Defibrillation
|
|---|
ICD therapy is effective for the treatment of atrial fibrillation.
An atrial-only defibrillator was developed and tested in early
clinical trials.
69 However, development of a stand-alone atrial
ICD has not progressed. Instead, atrial ATP and defibrillation
capabilities have been incorporated in current-generation dual-chamber
ICDs. Adequate atrial defibrillation thresholds have been obtained
in almost all patients undergoing implantation.
70,71 The main
limitation to the use of the ICD is the painful nature of ICD
shocks. Self-administered sedation and shock timing have been
used to minimize the morbidity of atrial defibrillation shocks.
In addition, atrial fibrillation may often begin as atrial tachycardia,
and thus, atrial ATP may be effective at terminating arrhythmias
soon after they begin and prevent the progression to atrial
fibrillation.
 |
The Future
|
|---|
A summary of current recommendations for ICD and CRT is shown
in the
Table. The ICD (including CRT) has already had a major
impact on the treatment of heart failure. Future-generation
ICDs will likely become cardiac disease management devices.
Sensors that can monitor the progression of heart failure and
even follow intracellular and extracellular ion concentrations
are under development. Devices that are currently in clinical
trials can monitor the extent of heart failure and potentially
be developed to provide feedback management of heart failure
by changing pacing rate, atrioventricular intervals, VV (intervals
in biventricular devices), or other parameters. Transtelephonic
or wireless communication from ICDs to nurses and physicians
caring for patients has become a reality. Other features such
as ischemia detection can potentially be incorporated in future
generations of ICDs. Each of these features should be carefully
evaluated for its clinical benefit and cost-effectiveness before
routine application to the management of all patients.
 |
References
|
|---|
- Zipes DP, Wellens HJ. Sudden cardiac death. Circulation. 1998; 98: 23342351.[Free Full Text]
- Myerburg RJ, Interian AJ, Mitrani RM, Kessler KM, Castellanos A. Frequency of sudden cardiac death and profiles of risk. Am J Cardiol. 1997; 80: 10F19F.[CrossRef][Medline]
[Order article via Infotrieve]
- Connolly SJ, Hallstrom AP, Cappato R, Schron EB, Kuck KH, Zipes DP, Greene HL, Boczor S, Domanski M, Follmann D, Gent M, Roberts RS. Meta-analysis of the implantable cardioverter defibrillator secondary prevention trials: AVID, CASH and CIDS studies; Antiarrhythmics vs Implantable Defibrillator study; Cardiac Arrest Study Hamburg; Canadian Implantable Defibrillator Study. Eur Heart J. 2000; 21: 20712078.[Abstract/Free Full Text]
- Huikuri HV, Castellanso A, Myerburg RJ. Sudden death due to cardiac arrhythmias. N Engl J Med. 2001; 345: 14731482.[Free Full Text]
- Mirowski M, Mower MM. Transvenous automatic defibrillator as an approach to prevention of sudden death from ventricular fibrillation. Heart Lung. 1973; 2: 867869.[Medline]
[Order article via Infotrieve]
- Mirowski M, Reid PR, Mower MM, Watkins L, Gott VL, Schauble JF, Langer A, Heilman MS, Kolenik SA, Fischell RE, Weisfeldt ML. Termination of malignant ventricular arrhythmias with an implanted automatic defibrillator in human beings. N Engl J Med. 1980; 303: 322324.[Medline]
[Order article via Infotrieve]
- Cannom DS, Prystowsky EN. The evolution of the implantable cardioverter defibrillator. Pacing Clin Electrophysiol. 2004; 27: 419431.[CrossRef][Medline]
[Order article via Infotrieve]
- Wellens HJ, Lau CP, Luderitz B, Akhtar M, Waldo AL, Camm AJ, Timmermans C, Tse HF, Jung W, Jordaens L, Ayers G. Atrioverter: an implantable device for the treatment of atrial fibrillation. Circulation. 1998; 98: 16511656.[Abstract/Free Full Text]
- DiMarco JP. Implantable cardioverter-defibrillators. N Engl J Med. 2003; 349: 18361847.[Free Full Text]
- Lamas G, Lee K, Sweeney M, Silverman R, Leon A, Yee R, Marinchak R, Flaker G, Schron E, Orav E, Hellkamp A, Greer SG, McAnulty J, Ellenbogen K, Ehlert F, Freedman R, Estes NI, Greenspon A, Goldman L. Ventricular pacing or dual-chamber pacing for sinus-node dysfunction. N Engl J Med. 2002; 346: 18541862.[Abstract/Free Full Text]
- Swerdlow CD, Chen PS, Kass RM, Allard JR, Peter CT. Discrimination of ventricular tachycardia from sinus tachycardia and atrial fibrillation in a tiered-therapy cardioverter-defibrillator. J Am Coll Cardiol. 1994; 23: 13421355.[Abstract]
- Rinaldi CA, Simon RD, Baszko A, Bostock J, Elliot D, Bucknall CA, Gill JS. A 17-year experience of inappropriate shock therapy in patients with implantable cardioverter-defibrillators: are we getting any better? Heart. 2004; 90: 330331.[Free Full Text]
- Leclercq C, Walker S, Linde C, Clementy J, Marshall A, Ritter P, Djiane P, Mabo P, Levy T, Gadler F, Bailleul C, Daubert J. Comparative effects of permanent biventricular and right-univentricular pacing in heart failure patients with chronic atrial fibrillation. Eur Heart J. 2002; 23: 17801787.[Abstract/Free Full Text]
- Ukkonen H, Beanlands RS, Burwash IG, de Kemp RA, Nahmias C, Fallen E, Hill MR, Tang AS. Effect of cardiac resynchronization on myocardial efficiency and regional oxidative metabolism. Circulation. 2003; 107: 2831.[Abstract/Free Full Text]
- Bristow MR, Saxon LA, Boehmer J, Krueger S, Kass DA, De Marco T, Carson P, Dicarlo L, DeMets D, White BG, DeVries DW, Feldman AM, for the Comparison of Medical Therapy, Pacing, and Defibrillation in Heart Failure (COMPANION) Investigators. Cardiac-resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure. N Engl J Med. 2004; 350: 21402150.[Abstract/Free Full Text]
- Bradley DJ, Bradley EA, Baughman KL, Berger R, Calkins H, Goodman SN, Kass DA, Powe NR. Cardiac resynchronization and death from progressive heart failure: a meta-analysis of randomized controlled trials. JAMA. 2003; 289: 730740.[Abstract/Free Full Text]
- Kadish A, Dyer A, Daubert JP, Quigg R, Estes M, Anderson KP, Calkins H, Hoch D, Goldberger J, Shalaby A, Sanders WE, Schaechter A, Levine JH. Prophylactic defibrillator implantation in patients with nonischemic dilated cardiomyopathy. N Engl J Med. 2004; 350: 21512158.[Abstract/Free Full Text]
- Moss AJ, Zareba W, Hall WJ, Klein H, Wilber DJ, Cannom DS, Daubert JP, Higgins SL, Brown MW, Andrews ML. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med. 2002; 346: 877883.[Abstract/Free Full Text]
- The Antiarrhythmics versus Implantable Defibrillators (AVID) Investigators. A comparison of antiarrhythmic-drug therapy with implantable defibrillators in patients resuscitated from near-fatal ventricular arrhythmias. N Engl J Med. 1997; 337: 15761583.[Abstract/Free Full Text]
- Gradaus R, Block M, Brachmann J, Breithardt G, Huber HG, Jung W, Kranig W, Mletzko RU, Schoels W, Seidl K, Senges J, Siebels J, Steinbeck G, Stellbrink C, Andresen D, German ER. Mortality, morbidity, and complications in 3344 patients with implantable cardioverter defibrillators: results fron the German ICD Registry EURID. Pacing Clin Electrophysiol. 2003; 26: 15111518.[CrossRef][Medline]
[Order article via Infotrieve]
- Kron J, Herre J, Renfroe EG, Rizo-Patron C, Raitt M, Halperin B, Gold M, Goldner B, Wathen M, Wilkoff B, Olarte A, Yao Q. Lead- and device-related complications in the antiarrhythmics versus implantable defibrillators trial. Am Heart J. 2001; 141: 9298.[CrossRef][Medline]
[Order article via Infotrieve]
- Tunis SR. Why Medicare has not established criteria for coverage decisions. N Engl J Med. 2004; 350: 21962198.[Free Full Text]
- Sheldon R, Connolly S, Krahn A, Roberts R, Gent M, Gardner M. Identification of patients most likely to benefit from implantable cardioverter-defibrillator therapy: the Canadian Implantable Defibrillator Study. Circulation. 2000; 101: 16601664.[Abstract/Free Full Text]
- Sanders GD, Hlatky MA, Every NR, McDonald KM, Heidenreich PA, Parsons LS, Owens DK. Potential cost-effectiveness of prophylactic use of the implantable cardioverter defibrillator or amiodarone after myocardial infarction. Ann Intern Med. 2001; 135: 870883.[Abstract/Free Full Text]
- Special report: cost-effectiveness of implantable cardioverter-defibrillators in a MADIT-II population. Technology Evaluation Center Assessment Program. 2004; 19: 12.
- Winkle RA, Mead RH, Ruder MA, Gaudiani VA, Smith NA, Buch WS, Schmidt P, Shipman T. Long-term outcome with the automatic implantable cardioverter-defibrillator. J Am Coll Cardiol. 1989; 13: 13531361.[Abstract]
- Mirowski M, Reid PR, Winkle RA, Mower MM, Watkins L Jr, Stinson EB, Griffith LS, Kallman CH, Weisfeldt ML. Mortality in patients with implanted automatic defibrillators. Ann Intern Med. 1983; 98: 585588.[Medline]
[Order article via Infotrieve]
- Lee DS, Green LD, Liu PP, Dorian P, Newman DM, Grant FC, Tu JV, Alter DA. Effectiveness of implantable defibrillators for preventing arrhythmic events and death: a meta-analysis. J Am Coll Cardiol. 2003; 41: 15731582.[Abstract/Free Full Text]
- Connolly S, Krahn A, Roberts R, Gent M, Gardner M, Connolly SJ. Canadian implantable defibrillator study (CIDS): a randomized trial of the implantable cardioverter defibrillator against amiodarone. Circulation. 2000; 101: 16601664.[Abstract/Free Full Text]
- Kuck KH, Cappato R, Siebels J, Ruppel R. Randomized comparison of antiarrhythmic drug therapy with implantable defibrillators in patients resuscitated from cardiac arrest: the Cardiac Arrest Study Hamburg (CASH). Circulation. 2000; 102: 748754.[Abstract/Free Full Text]
- Reynolds MR, Josephson ME. MADIT II (second Multicenter Automated Defibrillator Implantation Trial) debate: risk stratification, costs, and public policy. Circulation. 2003; 108: 17791783.[Free Full Text]
- Owens DK, Sanders GD, Heidenreich PA, McDonald KM, Hlatky MA. Effect of risk stratification on cost-effectiveness of the implantable cardioverter defibrillator. Am Heart J. 2002; 144: 440448.[CrossRef][Medline]
[Order article via Infotrieve]
- Owens DK, Sanders DG, Harris RA, McDonald KM, Heidenreich PA, Dembitzer A, Hlatky MA. Cost-effectiveness of implantable cardioverter defibrillators relative to amiodarone for prevention of sudden cardiac death. Ann Intern Med. 1997; 126: 112.[Abstract/Free Full Text]
- Oseroff O, Retyk E, Bochoeyer A. Subanalyses of secondary prevention implantable cardioverter-defibrillator trials: Antiarrhythmics Versus Implantable Defibrillators (AVID), Canadian Implantable Defibrillator Study (CIDS), and Cardiac Arrest Study Hamburg (CASH). Curr Opin Cardiol. 2004; 19: 2630.[CrossRef][Medline]
[Order article via Infotrieve]
- Hallstrom AP, McAnulty JH, Wilkoff BL, Follmann D, Raitt MH, Carlson MD, Gillis AM, Shih HT, Powell JL, Duff H, Halperin BD; Antiarrhythmics Versus Implantable Defibrillator (AVID) Trial Investigators. Patients at lower risk of arrhythmia recurrence: a subgroup in whom implantable defibrillators may not offer benefit. J Am Coll Cardiol. 2001; 37: 10931099.[Abstract/Free Full Text]
- Goldberger JJ, Parker MA, Kadish AH. Over-AVID subgroup analysis. J Am Coll Cardiol. 2001; 38: 15861587.[Free Full Text]
- Myerburg RJ, Velez M, Rosenberg DG, Fenster J, Castellanos A. Automatic external defibrillators for prevention of out-of-hospital sudden death: effectiveness of the automatic external defibrillator. J Cardiovasc Electrophysiol. 2003; 14: S108S116.[CrossRef][Medline]
[Order article via Infotrieve]
- Myerburg RJ, Fenster J, Velez M, Rosenberg D, Lai S, Kurlansky P, Newton S, Knox M, Castellanos A. Impact of community-wide police car deployment of automated external defibrillators on survival from out-of-hospital cardiac arrest. Circulation. 2002; 106: 10581064.[Abstract/Free Full Text]
- Kadish A, Schmaltz S, Calkins H, Morady F. Management of nonsustained ventricular tachycardia guided by electrophysiological testing. Pacing Clin Electrophysiol. 1993; 16: 10371050.[CrossRef][Medline]
[Order article via Infotrieve]