(Circulation. 2004;110:826-834.)
© 2004 American Heart Association, Inc.
Original Articles |
From the Cardiac MR Research Group, Division of Imaging Sciences, Kings College (V.M., A.T., R.A., S.H., M.E.M., D.L.G.H., R.S.R.); the Department of Congenital Heart Disease, Guys Hospital (R.T., E.B., R.S.R.); and the Cardiothoracic Unit, Institute of Child Health and Great Ormond Street Hospital (A.T.), London, United Kingdom.
Correspondence to Dr R. Razavi, 5th Floor, Guys Hospital, London SE1 9RT, UK. E-mail reza.razavi{at}kcl.ac.uk
Received October 2, 2003; de novo received December 12, 2003; revision received March 30, 2004; accepted April 2, 2004.
| Abstract |
|---|
|
|
|---|
Methods and Results In 24 patients with either suspected pulmonary hypertension or congenital heart disease requiring preoperative assessment, PVR was calculated by the use of simultaneously acquired MR flow and invasive pressure measurements (condition 1). In 19 of the 24 patients, PVR was also calculated at 20 ppm nitric oxide +30% (condition 2) and at 20 ppm nitric oxide +100% oxygen (condition 3), with the use of the MR method. This method proved safe and feasible in all patients. In 15 of 19 patients, PVR calculated by Fick flow was compared with the MR method. At condition 1, Bland-Altman analysis revealed a bias of 2.3% (MR > Fick) and limits of agreement of 50.2% to 45.5%. At condition 2, there was poorer agreement (bias was 28%, and the limits of agreement were 151.3% to 95.2%). At condition 3, there was very poor agreement (bias was 54.2%, and the limits of agreement were 174.4% to 66.0%).
Conclusions We have demonstrated the feasibility of using simultaneous invasive pressure measurements and MR flow data to measure PVR in humans.
Key Words: vasculature magnetic resonance imaging catheterization
| Introduction |
|---|
|
|
|---|
Currently, catheter manometers are used to measure pressure, whereas the Fick principle is used to quantify flow. However, the Fick principle cannot be considered a "gold standard."46
The Fick principle requires measurement of hemoglobin, aortic and pulmonary artery oxygen saturations and partial pressures, and oxygen consumption. This dependency on multiple measurements leads to the accumulation of individual measurement errors and is a considerable source of inaccuracy.46 The accuracy and precision of the Fick principle is further reduced in patients with large intracardiac shunts and high pulmonary blood flow as the result of the reduced pulmonary arteriovenous oxygen content difference.46 Reliable PVR quantification is vital in this group of patients because they are likely to develop pulmonary vascular disease. The use of high-concentration oxygen as a pulmonary vasodilator further reduces the reliability of the Fick principle. This is a severe limitation because response to vasodilators is important in making treatment decisions about medical therapy.3
Other methods of invasive flow quantification (eg, indicator dilution and thermodilution techniques) also suffer from reduced reliability in the presence of cardiac shunts or valvular regurgitation.7
An additional problem with current techniques is the requirement for fluoroscopic guidance when placing and manipulating catheters. Exposure to medical radiation is increasingly being recognized as a significant health risk for both patients and healthcare professionals.8,9
There is a need for a method of flow quantification that allows accurate and reproducible measurement of PVR, with less exposure to x-ray radiation. Velocity-encoded phase-contrast magnetic resonance (MR) enables quantification of blood flow in major vessels. Cardiac output and the pulmonary-to-systemic flow ratio (Qp:Qs) measured with the use of this technique have been shown to be accurate.10,11 In addition, phase-contrast MR has been validated in numerous phantom experiments.10,12 However, it has not been previously possible to obtain invasive pressure and MR flow data simultaneously.
We have set up a program of MR-guided diagnostic cardiac catheterization in children and adults with suspected pulmonary hypertension. This technique allows simultaneous acquisition of invasive pressure and MR flow data with less exposure to x-ray radiation. The aim of the present study is to evaluate the practicality of this novel technique and compare it with the traditional method. We describe the first 24 patients to be studied in our facility.
| Methods |
|---|
|
|
|---|
|
Cardiac Catheterization
Subjects underwent cardiac catheterization in an MR interventional suite (1 · 5 T Intera I/T MRI scanner; maximum gradient performance, 30 mT/m amplitude; slew rate, 150 T/m per second; Philips Medical Systems) with x-ray backup (BV Pulsera cardiac x-ray unit, Philips Medical Systems) in the same room (XMR suite). General anesthesia was induced and maintained with propofol, and uncuffed endotracheal tubes were used in patients younger than 12 years of age. Catheter advancement was carried out either under MR guidance or by using a combination of x-ray and MR guidance, depending on ease of catheter manipulation. Angiographic catheters (4F to 6F) with carbon dioxidefilled balloons were used for all procedures, allowing visualization in both imaging modalities.
MR Catheter Visualization
All imaging was performed with the use of an adult or pediatric 2-element phase-array coil (Flex L, Flex M; Philips Medical Systems) for signal reception and the body coil for signal transmission. An interactive steady-state free precession sequence (echo time, 1.45 ms; repetition time, 2.9 ms; matrix, 128x128; field of view, 250 to 350 mm; 10 to 14 frames per second; flip angle, 45°; slice thickness, 6 to 8 mm; free breathing acquisition) was used to visualize the catheter balloon during manipulation within the heart and great vessels. The imaging plane could be manipulated in real time into any plane to follow the catheter.
PVR Data Collection
PVR data were collected in all 24 patients at baseline (30% oxygen) by means of the MR and Fick methods; 30% oxygen was used at baseline to compensate for the relative hypoxia associated with anesthesia.13 In 19 patients in whom assessment of response to vasodilators was clinically indicated according to our hospital policy, PVR was also calculated at 30% oxygen and 20 ppm of NO (condition 2) and at 100% oxygen and 20 ppm of NO (condition 3), by using both the MR and Fick methods.
Pressure data (main pulmonary artery and pulmonary wedge or left atrium) was collected in all 24 patients during image acquisition with the use of standard fluid-filled catheters coupled to a computerized data recording system (5S collect system, Datex Ohmeda).
MR Flow Quantification
Pulmonary artery flow data were acquired through the use of a flow-sensitive fast-field echo sequence (repetition time,
9 ms; echo time,
5 ms; matrix, 128 to 192x256, field of view, 250 to 350 mm; flip angle, 15; slice thickness, 5 to 7 mm; free breathing acquisition, number signal averages of 3; retrospective gating, 20 to 40 phases; in-plane and through-plane resolution optimized to patient size). A dedicated nonlinear phase correction filter, based on Chebichev polynomials (Philips Medical Systems), was used to correct for phase errors introduced by eddy currents and Maxwell terms. In addition, the image plane was centered in the bore of the magnet to further reduce the phase errors. Image planes were located at the midpoint of the pulmonary artery or at the midpoint of the superior vena cava in patients with bidirectional cavopulmonary shunts. In the patient with a Waterston shunt, the pulmonary flow was defined as the difference between the preshunt and postshunt flow in the ascending aorta.
Fick Flow Measurement
In the 19 patients who underwent Fick flow measurement in response to vasodilators, blood samples were obtained from the pulmonary artery and the aorta or left atrium pulmonary vein, depending on anatomy. The oxygen saturations and partial pressures of these samples were measured (A-vox Systems Inc). This allowed quantification of oxygen content incorporating oxygen bound to hemoglobin and dissolved plasma oxygen. In conditions 1 and 2, oxygen consumption was measured with the use of a paramagnetic metabolic monitor14 (Deltatrac, Datex Ohmeda); however, in condition 3, exposure to 100% oxygen made it impossible to measure oxygen consumption with the use of the metabolic monitor. Therefore, the oxygen consumption immediately before commencement of 100% oxygen was used. In 4 patients, technical difficulties with acquisition of oxygen consumption (ie, endotracheal leak) or oxygen content measurement invalidated calculation of pulmonary artery flow. These 4 patients were excluded from the comparison between Fick and MR quantification of PVR.
MR Validation In Vitro
A 25-mm-diameter distensible rubber tube embedded in methylcellulose gel was used to assess the accuracy and precession of phase-contrast MR. Pulsatile flow (1 to 8.7 L/min) and ECG gating were generated with the use of an animal bypass machine (Harvard Apparatus Inc). A solution of glycerol (40%) and water (60%) was used to mimic the MR (T1=850 ms, T2=170 ms) and viscous properties of blood. Accuracy was assessed by comparing a range of flow rates, quantified by means of the same MR sequence as used in the human study and with the use of the graduated cylinder and stopwatch method.10 For a given flow rate, the heart rate and stroke volume corresponded to expected values for age-appropriate patients with similar cardiac output. Precision was assessed by comparing repeated phase-contrast MR flow quantification at low, medium, and high flow rates. In addition, 10 phase-contrast MR measurements at high flow rate were made with the catheter in situ to assess the effect of the catheter on accuracy and precision.
Data Analysis
Averaged pulmonary blood stroke volume was calculated with phase-contrast MR data by means of a semiautomatic vessel edge detection algorithm (Flow, Medis), with operator correction. Pulmonary artery blood flow is the product of pulmonary blood stroke volume and heart rate. In all patients, each scan (n=62) was analyzed by 2 different researchers (blinded to each others results and to results derived from invasive oximetry) to test interoperator variability. The Fick equation was used to calculate pulmonary blood flow from invasive oximetry data. Flow was divided by surface area to produce the indexed flow. PVR was calculated by dividing transpulmonary artery pressure gradient by indexed pulmonary artery flow: PVR = (mean pulmonary artery pressure pulmonary wedge or left atrial pressure) ÷ (pulmonary blood flow/body surface area).
X-Ray Dose Comparison
X-ray dose was recorded in all 24 study subjects and compared with x-ray dose in age- and procedure-matched control subjects who underwent PVR studies in our traditional catheterization laboratory. Control patients underwent PVR studies either because of suspected pulmonary hypertension or as part of preoperative assessment.
Statistical Analysis
Data are expressed as means (±SD) unless otherwise stated. A 2-tailed t test was used to compare x-ray dose in the study and control population, the response to NO, and in vitro flow data. Correlation coefficients and Bland-Altman analysis were used to compare the different methods phantom flow quantification (phase-contrast MR versus stopwatch) and interoperator variability. Bias was the mean of the difference between the 2 methods, and agreement was the mean±2 SD.15 Precision was quantified by use of the 95% confidence intervals of the bias and the agreements.15 Correlation coefficients and Bland-Altman analysis were also used to compare the different methods of PVR quantification (Fick versus phase-contrast MR). To allow comparison of the differences between the 2 methods at each of the conditions, the bias, level of agreements, and precision were expressed as percentages. A probability value of <0.05 was taken as statistically significant. Statistical analysis was performed with the use of Matlab (Mathworks) (Figures 1, 2, 3, and 4![]()
![]()
).
|
|
|
|
| Results |
|---|
|
|
|---|
X-Ray Exposure
There was a significant difference between the mean x-ray dose received by the 24 patients in this study who underwent XMR-guided cardiac catheterization and control subjects, who underwent traditional cardiac catheterization (mean, 5.8±11.3 Gy cm2 versus mean, 38.2±39.2 Gy cm2, respectively, P<0.005).
PVR: Phase-Contrast MR
The mean baseline PVR for all 24 patients calculated with the use of phase-contrast MR flow data was 5.1±6.3 WU (Woods units)/m2. In the 19 patients in whom the response to vasodilators was assessed, the baseline PVR was 5.9±6.8 WU/m2. At condition 2 (20 ppm of NO), the mean PVR fell to 5.2±6.2 WU/m2, a fall from baseline of 12% (P=0.049), and at condition 3 (20 ppm NO +100% O2), the mean PVR fell to 5.0±6.0 WU/m2, a fall from baseline of 15% (P=0.044).
Phase-Contrast MR Validation
The correlation coefficient between flow quantified by use of phase-contrast MR and by use of a graduated cylinder and stopwatch method was 0.99 (P<0.001) (Figure 1). The bias was negligible (0.02 L/min; precision, 0.10 to 0.06 L/min); an upper level of agreement was 0.26 L/min (precision, 0.11 to 0.40 L/min) and a lower level of agreement was 0.30 L/min (precision, 0.16 to 0.44 L/min). Precision was assessed by measuring the mean and standard error (as a percentage of the mean) of 10 repeated phase-contrast MR measurement at 3 flow rates. At the low flow rate, the phase-contrast MR flow was 0.74 L/min ±0.7%; at the medium flow rate, the flow was 2.92 L/min ±0.4%; and at the high flow rate, the flow was 5.05 L/min ±1.2%. The mean flow rate with the catheter in situ was 5.19 L/min ±0.9%, which was not statistically different from the flow rate without the catheter (P=0.1). Interoperator variability was minimal, with a bias of 0.02 L/min (precision, 0.01 to 0.05 L/min), an upper level of agreement of 0.22 L/min (precision, 0.16 to 0.27 L/min), and a lower level of agreement of 0.18 L/min (precision, 0.13 to 0.24 L/min).
PVR: Phase-Contrast MR Versus Fick
In the 15 patients in whom comparison between phase-contrast MR and Fick was made (Table 2), the mean PVR at baseline was 4.7±4.6 WU/m2 with the use of Fick-derived flow data and 4.6±3.5 WU/m2 with the use of phase-contrast MR flow data. At condition 2, the mean PVR measured with the use of Fick was 3.3±3.2 WU/m2 and the PVR with the use of phase-contrast MR was 3.7±2.1 WU/m2. At condition 3, mean PVR fell to 2.4±2.3 WU/m2 with the use of Fick and 3.0±1.9 WU/m2 with the use of phase-contrast MR flow data. At baseline, the correlation coefficient between the 2 methods was 0.91 (P<0.05) (Figure 2). Bland-Altman analysis revealed a bias of 2.3% (MR PVR > Fick PVR), with a precision of 10.9% to 15.6%, an upper limit of agreement of 50.2% (precision 27.2% to 73.2%), and a lower limit of agreement of 45.5% (precision, 22.6% to 68.5%). At condition 2, the two methods correlated less well (r=0.78, P<0.05) (Figure 3). The bias was 28% (precision, 6.1% to 62.1%), the upper limit of agreement was 151.3% (precision, 92.2% to 210.5%), and the lower limit of agreement was 95.2% (precision, 36.2% to 154.4%). At condition 3, the two methods showed poor correlation (r=0.59, P=0.02) (Figure 4) and very poor agreement. The bias was 54.2% (MR PVR > Fick PVR), with a precision of 20.1% to 87.5%; the upper limit of agreement was 174.4% (precision, 116.7% to 232.0%) and the lower limit of agreement was 66.0% (precision, 8.4% to 123.6%).
|
| Discussion |
|---|
|
|
|---|
Calculation of PVR is essential in the treatment of patients with suspected pulmonary hypertension.1,2 Unfortunately, the Fick principle, which is used to quantify pulmonary artery blood flow, is an indirect and potentially flawed means of calculating pulmonary flow.4 This may lead to inaccuracies in PVR quantification, which could adversely affect patient treatment. The development of new methods of simultaneous flow and pressure quantification are therefore an important step.
Velocity-encoded phase-contrast MR is a validated technique that enables quantification of blood flow in major vessels. Our novel facility allows catheter placement under both x-ray and MR within the same interventional suite, allowing the safe combination of invasive pressure measurements and MR flow quantification. In all patients, we demonstrated the feasibility and safety of this new method. Furthermore, in 10 patients, catheter placement was accomplished wholly under MR guidance, thus eliminating x-ray exposure.
Validation of Phase-Contrast MR
Assessment of the accuracy and precision of a diagnostic method requires comparison with a gold standard. There is no in vivo gold standard for flow quantification.4 There are, however, a number of in vitro studies with nondistensible phantoms that have demonstrated the accuracy and precision of this technique.10,12 We have further demonstrated the in vitro accuracy and precision of phase-contrast MR in our facility by using a flow phantom that replicates in vivo conditions more closely than previous studies. In vivo internal validation of phase-contrast MR by comparison of aortic and pulmonary blood flow or comparison of phase-contrast flow through the pulmonary artery or aorta with the ventricular output measured from short-axis images has also been previously performed.16
Discrepancy Between Fick and Phase-Contrast MR
We have shown reasonable agreement between Fick and phase-contrastderived PVR at baseline. This corroborates a number of other studies that have found reasonable agreement between phase-contrast MR flow and invasive method of flow quantification (eg, Fick).10,11 However, we found that in the presence of NO, there was less agreement between the 2 methods, and with 100% oxygen and NO there was a large bias and worsening agreement. The Fick principle is known to be inaccurate and imprecise in the presence of high pulmonary blood flow and high concentrations of oxygen.46 Our phantom studies demonstrate that phase-contrast MR is accurate at high flow rates. In addition, the oxygen content of pulmonary arterial blood at 100% oxygen is similar to the oxygen content of aortic blood at baseline, where it has been shown that phase-contrast MR is accurate.10 We therefore believe that the worsening agreement between the 2 methods in response to pulmonary vasodilation is due to errors in the Fick method rather than phase-contrast MR and that the Fick method underestimates PVR in the presence of 100% oxygen and 20 ppm of NO. This has important implications for patient treatment because response to vasodilators is integral to the assessment of patients with pulmonary hypertension.
Reduced X-Ray Dose
A further benefit of our MR method is the reduction in exposure to x-ray radiation. This is particularly important in children, who are more susceptible to the carcinogenic effects of x-ray dose. The reduced x-ray dose also allows more frequent invasive assessment of PVR. This will enable better monitoring of the effectiveness of new drug therapies. It also has implications for staff in the catheterization laboratory who may receive a significant cumulative dose in their working lifetime.
Limitations
Phase-contrast MR flow is less accurate in patients with either arrhythmias during acquisition or turbulent blood flow; the presence of these is a general limitation of this technique. We did not repeat previous work16 internally validating phase-contrast MR in our patient population because we did not wish to prolong the procedure times. However, we did demonstrate accuracy and precision of this technique in vitro. We also did not repeat previous work4 looking at the repeatability of the Fick method to minimize blood sampling, particularly in our younger patients. Finally, MR-guided cardiac catheterization takes longer than traditional techniques. This is partly due to the extra time required for patient preparation and transfer between modalities. However, the procedure time has fallen, as we have gained experience, to
2 hours.
Conclusions
We have demonstrated the feasibility of performing PVR quantification with the use of a novel MR method. Our facility includes an MR scanner with x-ray backup. We believe that phase-contrast MR is more accurate than Fick in measuring pulmonary blood flow. Although we have gone some way in demonstrating this, definitive proof requires invasive in vivo validation in an animal model of pulmonary hypertension. Reduction or elimination of x-ray radiation, added anatomic and functional information available with MR, and the relative ease and accuracy of phase-contrast MR flow quantification may make this technique the method of choice for invasive measurement of PVR.
| Acknowledgments |
|---|
| References |
|---|
|
|
|---|
2. Chemla D, Castelain V, Herve P, et al. Haemodynamic evaluation of pulmonary hypertension. Eur Respir J. 2002; 20: 13141331.
3. Atz AM, Adatia I, Lock JE, et al. Combined effects of nitric oxide and oxygen during acute pulmonary vasodilator testing. J Am Coll Cardiol. 1999; 33: 813819.
4. Cigarroa RG, Lange RA, Hillis LD. Oximetric quantitation of intracardiac left-to-right shunting: limitations of the Qp/Qs ratio. Am J Cardiol. 1989; 64: 246247.[CrossRef][Medline] [Order article via Infotrieve]
5. Dhingra VK, Fenwick JC, Walley KR, et al. Lack of agreement between thermodilution and Fick cardiac output in critically ill patients. Chest. 2002; 122: 990997.
6. Hillis LD, Firth BG, Winniford MD. Analysis of factors affecting the variability of Fick versus indicator dilution measurements of cardiac output. Am J Cardiol. 1985; 56: 764768.[CrossRef][Medline] [Order article via Infotrieve]
7. Razavi RS, Baker A, Qureshi SA, et al. Hemodynamic response to continuous infusion of dobutamine in Alagilles syndrome. Transplantation. 2001; 72: 823828.[CrossRef][Medline] [Order article via Infotrieve]
8. Guidelines on patient dose to promote the optimisation of protection for diagnostic medical exposures. National Radiation Protection Board. 1999; 10: 1.
9. Recommendations of the International Commission on Radiological Protection ICRP 60. Ann ICRP. 1990; 21: 13.[Medline] [Order article via Infotrieve]
10. Beerbaum P, Korperich H, Barth P, et al. Noninvasive quantification of left-to-right shunt in pediatric patients: phase-contrast cine magnetic resonance imaging compared with invasive oximetry. Circulation. 2001; 103: 24762482.
11. Hundley WG, Li HF, Hillis LD, et al. Quantitation of cardiac output with velocity-encoded, phase-difference magnetic resonance imaging. Am J Cardiol. 1995; 75: 12501255.[CrossRef][Medline] [Order article via Infotrieve]
12. Robertson MB, Kohler U, Hoskins PR, et al. Quantitative analysis of PC MRI velocity maps: pulsatile flow in cylindrical vessels. Magn Reson Imaging. 2001; 19: 685695.[CrossRef][Medline] [Order article via Infotrieve]
13. Aitkenhead AR, Rowbotham DJ, Smith G. Textbook of Anesthesia. 4th ed. London: Churchill Livingstone; 2001.
14. Tissot S, Delafosse B, Bertrand O, et al. Clinical validation of the Deltatrac monitoring system in mechanically ventilated patients. Intensive Care Med. 1995; 21: 149153.[CrossRef][Medline] [Order article via Infotrieve]
15. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986; 1: 307310.[CrossRef][Medline] [Order article via Infotrieve]
16. Firmin DN, Nayler GL, Klipstein RH, et al. In vivo validation of MR velocity imaging. J Comput Assist Tomogr. 1987; 11: 751756.[Medline] [Order article via Infotrieve]
This article has been cited by other articles:
![]() |
A. Bell, P. Beerbaum, G. Greil, S. Hegde, A. M. Toschke, T. Schaeffter, and R. Razavi Noninvasive Assessment of Pulmonary Artery Flow and Resistance by Cardiac Magnetic Resonance in Congenital Heart Diseases With Unrestricted Left-to-Right Shunt J. Am. Coll. Cardiol. Img., November 1, 2009; 2(11): 1285 - 1291. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. E. Saikus and R. J. Lederman Interventional Cardiovascular Magnetic Resonance Imaging: A New Opportunity for Image-Guided Interventions J. Am. Coll. Cardiol. Img., November 1, 2009; 2(11): 1321 - 1331. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. M. Müller, P. H. Rehak, M. Puchinger, D. Wagner, W. Marte, and K.-H. Tscheliessnigg Measurement of cardiac output and pulmonary transit time for assessment of pulmonary vascular resistance in domestic piglets Exp Physiol, June 1, 2009; 94(6): 659 - 664. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Crean Cardiovascular MR and CT in congenital heart disease Heart, December 1, 2007; 93(12): 1637 - 1647. [Full Text] [PDF] |
||||
![]() |
A. Vonk-Noordegraaf, J.-W. Lankhaar, M. J.W. Gotte, J. T. Marcus, P. E. Postmus, and N. Westerhof Magnetic resonance and nuclear imaging of the right ventricle in pulmonary arterial hypertension Eur. Heart J. Suppl., December 1, 2007; 9(suppl_H): H29 - H34. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Muthurangu, A. M. Taylor, S. R. Hegde, R. Johnson, R. Tulloh, J. M. Simpson, S. Qureshi, E. Rosenthal, E. Baker, D. Anderson, et al. Cardiac Magnetic Resonance Imaging After Stage I Norwood Operation for Hypoplastic Left Heart Syndrome Circulation, November 22, 2005; 112(21): 3256 - 3263. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Alphonso, M. Baghai, P. Sundar, R. Tulloh, C. Austin, and D. Anderson Intermediate-term outcome following the fontan operation: a survival, functional and risk-factor analysis Eur. J. Cardiothorac. Surg., October 1, 2005; 28(4): 529 - 535. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Muthurangu, D. Atkinson, M. Sermesant, M. E. Miquel, S. Hegde, R. Johnson, R. Andriantsimiavona, A. M. Taylor, E. Baker, R. Tulloh, et al. Measurement of total pulmonary arterial compliance using invasive pressure monitoring and MR flow quantification during MR-guided cardiac catheterization Am J Physiol Heart Circ Physiol, September 1, 2005; 289(3): H1301 - H1306. [Abstract] [Full Text] [PDF] |
||||
![]() |
V Muthurangu and R S Razavi The value of magnetic resonance guided cardiac catheterisation Heart, August 1, 2005; 91(8): 995 - 996. [Full Text] [PDF] |
||||
![]() |
T Kuehne, S Yilmaz, I Schulze-Neick, E Wellnhofer, P Ewert, E Nagel, and P Lange Magnetic resonance imaging guided catheterisation for assessment of pulmonary vascular resistance: in vivo validation and clinical application in patients with pulmonary hypertension Heart, August 1, 2005; 91(8): 1064 - 1069. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|