(Circulation. 2004;110:e572-e576.)
© 2004 American Heart Association, Inc.
AHA Conference Proceedings |
Abstract
Inflammation plays a key role in atherosclerosis. A number of different biomarkers of inflammation are measurable in blood. These include cytokines, chemokines, soluble adhesion molecules, and acute-phase reactants. The first 3 groups of molecules are not routinely measured in clinical laboratories. The acute-phase reactants include C-reactive protein (CRP). This analyte is stable and is readily measured by several commercial high-sensitivity (hs) assays, and numerous clinical studies have demonstrated its usefulness as an atherosclerotic risk marker. For these reasons, CRP is currently the inflammatory marker of choice. Comparison of commercial hsCRP assays has demonstrated that many are capable of imprecision of <7% at a CRP concentration of <1 mg/L. Method comparability studies have shown good agreement among some commercial hsCRP assays in terms of quartile assignments, whereas additional standardization efforts are necessary for others. The first goal of standardization is to develop a secondary reference material with a CRP of
4 mg/L. This material can be used to assist diagnostic vendors in calibrator value assignment and assay harmonization. Another project is to define clearly what performance characteristics are necessary for hsCRP assays. Preliminary estimates based on the biological variability of CRP indicate that the total allowable analytical error is between 15% and 32%. The former requirement can be met with accuracy and imprecision of <5% and the latter requirement with accuracy and imprecision of <11%. If 2 hsCRP results are averaged, then the imprecision requirements are relaxed slightly. Clinical validation of these performance requirements is also required.
Key Words: AHA Scientific Statements atherosclerosis inflammation risk factors
Inflammation plays a key role in the pathophysiology of atherosclerotic disease. A number of inflammatory markers that are measurable in blood have been investigated for their ability to predict the risk of future atherosclerotic events. These markers can be divided into 3 groups of molecules: cytokines and chemokines, soluble adhesion molecules, and acute-phase reactants. A number of preanalytical and analytical issues including specimen type and stability, assay imprecision, commercial availability, and standardization are reviewed here for selected members of each of these 3 groups of markers.
Cytokines and Chemokines
A large number of cytokines and chemokines have been described. Among those that have been investigated for a relationship to atherosclerosis are interleukin-1ß (IL-1ß), IL-6, IL-8, IL-10, tumor necrosis factor-
(TNF-
), and monocyte chemoattractant protein (MCP-1). All of these molecules are relatively unstable in serum. Serum or plasma must be rapidly separated from the cellular components of blood and assayed rapidly or frozen to prevent analyte degradation.1 Typically, these assays are performed in research laboratories via ELISA techniques. For some cytokines, more automated methods are becoming available for research. An automated microplate system for IL-6 has been described.2 Chemiluminescent assays for IL-6, IL-8, and TNF-
are available on an automated immunoassay analyzer for research use only in the United States.3 Multiplex assays for several cytokines also have been described.4 This methodology shows great promise, but at present it is still a research tool rather than a routine part of laboratory medicine practice. Most of the cytokine assays that have been used show imprecision, as assessed by a coefficient of variation between 2% and 15%, according to the concentration of analyte being measured.
Two general limitations of many cytokine assays are the lack of a sufficiently low limit of quantification for use in apparently healthy subjects and problems with assay standardization. Many of the commercial cytokine assays are designed to quantify the relatively high concentrations of cytokines encountered in acute inflammatory processes. For cytokines to be useful for atherosclerotic risk prediction, the assay must be able to accurately and precisely quantify cytokine levels within the reference interval (ie, the concentrations encountered in an apparently healthy population) so that subjects can be divided into tertiles, quartiles, or quintiles. Standardization of cytokine and chemokine assays is problematic. Although international standards exist for several cytokines, including IL-1ß, IL-6, IL-8, and TNF-
, standardization of cytokine immunoassays is a difficult process.58 One reason for this may be that the nature of the different monoclonal antibody pairs used for each assay does not permit comparable recognition in individual patient samples.7
Soluble Adhesion Molecules
A number of soluble adhesion molecules have been examined as possible markers for atherosclerotic risk. E-selectin, P-selectin, intracellular adhesion molecule-1 (sICAM-1), and vascular adhesion molecule-1 (sVCAM-1) are some of the better-studied molecules. Little information is available about preanalytical conditions affecting their recovery. Typically, like most cytokines, they are unstable unless frozen. ELISA methods generally are used to measure these analytes in research laboratories. The availability of commercial kits is limited. These analytes are poorly standardized. Overall, these limitations preclude their consideration for routine use in the clinical laboratory.
Acute-Phase Reactants
Acute-phase reactants are a group of proteins that are synthesized by the liver in response to cytokines, primarily IL-6. Fibrinogen, serum amyloid A (SAA), and C-reactive protein (CRP) are 3 acute-phase reactants that have been best studied as inflammatory markers of atherosclerosis.
Fibrinogen is a clotting factor produced by the liver that has been widely studied as an inflammatory marker of atherosclerotic events. It usually is measured in sodium citrate plasma, although EDTA plasma can be used for some assays. Fibrinogen is relatively unstable, and plasma should be separated from cells and frozen as soon as possible if not analyzed immediately.9 Two general classes of methods are available: immunoassays and functional assays. Many automated commercial methods are available, and most hospital laboratories have the capability to measure fibrinogen. The imprecision of most assays is in the range of 4% to 8%. Despite that several studies have indicated that fibrinogen is a good inflammatory marker for atherosclerotic risk and the assay is widely available, difficulties with assay standardization makes this analyte unattractive for routine clinical use as a marker of atherosclerotic risk.10 An international plasma standard (World Health Organization [WHO] 2nd International Standard [IS] Fibrinogen, Plasma 98/612) is available for fibrinogen.11 Because several different methodologies are used, however, standardization is problematic. The fibrinogen concentration can be estimated from the clotting curve of the prothrombin time on automated photo-optical coagulometers. These results can be significantly higher than values derived from Clauss measurements, especially in patients receiving oral anticoagulants, in some patients with documented hypodysfibrinogenemia, and in patients with high fibrinogen concentrations.12,13 The Clauss method can underestimate the fibrinogen concentrations in the presence of hirudin peptide 54-65 and fibrinogen fragment E.13 Assignment of fibrinogen values to commercially available plasma quality control materials also can be problematic, and significant variability from instrument to instrument exists.14 The use of a single type of functional assay, or perhaps better yet the use of immunoassays directed toward a specific epitope, may facilitate standardization of fibrinogen assays and improve the feasibility of use of this marker.
SAA protein is another acute-phase reactant that has been examined as a risk factor for atherosclerotic events. SAA in serum samples is stable for 8 days if refrigerated.15 A single commercial nephelometric assay is available in the United States for research use only. The imprecision of this assay is <9%.16 An international standard for SAA (WHO 1st IS 92/680) has been established.17 Multiple phenotypes of SAA are known to exist that have the potential to complicate future immunoassay standardization efforts for this analyte.
High-Sensitivity CRP
High-sensitivity measurement of CRP (hsCRP) has received a great deal of attention recently for use as an atherosclerotic risk marker. CRP is a 120-kDa pentamer that can be measured by a variety of methods including radioimmunoassay, immunonephelometry, immunoturbidimetry, immunoluminometry, and ELISA. High sensitivity is loosely defined as having sufficient precision at low concentrations of CRP to allow stratification of the results from apparently healthy subjects into tertiles, quartiles, or quintiles. Practically speaking, this translates into a lower limit of quantification of
0.3 mg/L and assay imprecision of <10% at a CRP concentration <0.5 to 1.0 mg/L, although exact performance requirements have not been fully defined.
The preanalytical and analytical sources of variations in hsCRP measurements have been reviewed.18 CRP has several advantages over other markers discussed to this point. First, the analyte is quite stable. Whole-blood samples are stable for 3 days at ambient temperature. Serum, plasma, and whole blood are stable when refrigerated for at least 7 days and for extended periods of time when frozen.19 Second, many automated hsCRP methods are commercially available (>20 in the United States). This plethora of methods greatly facilitates the widespread use of this marker in clinical laboratories. Third, an international secondary serum reference material (CRM 470, developed by the International Federation for Clinical Chemistry) is available for standardization of kit calibrators. The CRP concentration in CRM 470 is 39.2 mg/L.
Several studies have examined assay imprecision and method comparability of several commercial hsCRP assays.1923 The imprecision of hsCRP methods at low CRP concentrations varies widely (Table), although a number of methods are capable of imprecision of <7% at CRP concentrations as low as 0.5 mg/L, a value that demarcates the 25th to 35th percentile in healthy populations. The standardization of the assays based on assignment to the appropriate quartile on the basis of a nephelometric comparison method is good for many methods, whereas other methods require additional standardization efforts (Figure). One of the challenges to standardization with CRM 470 is that the CRP concentration is 10 times higher than is appropriate for standardization of hsCRP. Development of a secondary standard reference material would benefit continued standardization efforts. The Centers for Disease Control and Prevention have initiated a project to standardize hsCRP assays. The first phase of this project is to identify a secondary reference material with characteristics that are the same as fresh patient samples. Results from this first phase indicate that it is feasible to use diluted CRM 470 as a reference material for hsCRP assays.24 The second phase of the standardization project will involve working with diagnostic vendors to ensure accurate calibrator value assignment and to normalize assay results. When this project is completed, it should minimize assay inaccuracy and permit the use of hsCRP results generated by different assays in different laboratories to be used interchangeably with a common set of clinical decision thresholds. The College of American Pathologists introduced proficiency-testing material for hsCRP assays in 2002. This proficiency-testing program should aid in both the assessment of assay standardization and the performance of hsCRP methods in individual laboratories.
|
|
As mentioned above, the performance requirements for hsCRP assays are not clearly defined. The total analytical error for a given analyte typically can be up to 50% of the intraindividual biological variation.25 The total analytical error is equal to systematic error plus the random error.26 The systematic error is equivalent to the assay accuracy, and the random error equals 1.96 times the assay imprecision.27 Published intraindividual biological variability for CRP is 30% to 63%.2830 On the basis of these 3 studies, the total analytical error is between 15% and 32%. Additional studies on intraindividual variation are necessary to further refine the estimate of true CRP biological variability. If the accuracy of hsCRP assays is ±10%, a goal that is achievable with a standardization program, then the allowable imprecision is 2.6% to 11% for a single hsCRP measurement. If multiple hsCRP measurements are made, then the allowable imprecision is greater. If 2 measurements are averaged, then the allowable imprecision of the assay becomes 3.7% to 15.6%. The overall imprecision of many current commercial hsCRP assays is <7%. This degree of imprecision may be adequate clinically, but this estimate remains to be refined and formally validated.
Conclusions
A number of atherosclerotic risk markers have been measured in research laboratories. On careful review of the status of these assays in terms of clinical outcome studies, analyte stability under ambient or refrigerated conditions, availability of multiple automated methods that are cleared by the US Food and Drug Administration for clinical use, adequate precision and accuracy to be clinically useful, and standardization, it was determined that hsCRP is the best available marker. Further research is required to better define the performance characteristics necessary for assays bearing the designation hsCRP. These characteristics include developing guidelines for total analytical error from a careful review of the intraindividual biological variability of the analyte under conditions that will be encountered in clinical practice, defining allowable random and systematic error limits based on this information, validating these guidelines in the clinical setting, and completing the standardization efforts that are under way at the Centers for Disease Control and Prevention.
Footnotes
This paper represents a summary of a scientific conference sponsored by the American Heart Association. The opinions expressed in this paper are those of the authors and do not necessarily represent those of the editor or the American Heart Association. The publication of these proceedings was approved by the American Heart Association Science Advisory and Coordinating Committee on August 4, 2004. All writing group members were required to complete and submit shortly before the workshop a Faculty Disclosure Questionnaire. These disclosures are available as an online appendix at http://www.circulationaha.org (Circulation. 2004;110:e577e578).
A single reprint is available by calling 800-242-8721 (US only) or by writing the American Heart Association, Public Information, 7272 Greenville Ave, Dallas, TX 75231-4596. Ask for reprint No. 71-0306. To purchase additional reprints: up to 999 copies, call 800-611-6083 (US only) or fax 413-665-2671; 1000 or more copies, call 410-528-4121, fax 410-528-4264, or e-mail kgray@lww.com. To make photocopies for personal or educational use, call the Copyright Clearance Center, 978-750-8400.
The workshop overview; the reports of the Laboratory Science, the Clinical Practice, and the Population Science Discussion Groups; and 2 other background papers are available online at http://www.circulationaha.org (Circulation. 2004;110:e543e544; e545e549; e550e553; e554e559; e560e567; and e568e571).
References
1. Flower L, Ahuja RH, Humphries SE, Mohamed-Ali V. Effects of sample handling on the stability of interleukin 6, tumour necrosis factor-alpha and leptin. Cytokine. 2000; 12: 17121716.[CrossRef][Medline] [Order article via Infotrieve]
2. Fraunberger P, Pfeiffer M, Cremer P, Holler E, Nagel D, Dehart I, Thein M, Walli AK, Seidel D. Validation of an automated enzyme immunoassay for interleukin-6 for routine clinical use. Clin Chem Lab Med. 1998; 36: 797801.[CrossRef][Medline] [Order article via Infotrieve]
3. Berthier F, Lambert C, Genin C, Bienvenu J. Evaluation of an automated immunoassay method for cytokine measurement using the Immulite Immunoassay system. Clin Chem Lab Med. 1999; 37: 593599.[CrossRef][Medline] [Order article via Infotrieve]
4. Prabhakar U, Eirikis E, Davis HM. Simultaneous quantification of proinflammatory cytokines in human plasma using the LabMAP assay. J Immunol Methods. 2002; 260: 207218.[CrossRef][Medline] [Order article via Infotrieve]
5. Bienvenu J, Coulon L, Doche C, Gutowski MC, Grau GE. Analytical performances of commercial ELISA-kits for IL-2, IL-6 and TNF-alpha. A WHO study. Eur Cytokine Netw. 1993; 4: 447451.[Medline] [Order article via Infotrieve]
6. Gaines Das RE, Poole S. The international standard for interleukin-6. Evaluation in an international collaborative study. J Immunol Methods. 1993; 160: 147153.[CrossRef][Medline] [Order article via Infotrieve]
7. Ledur A, Fitting C, David B, Hamberger C, Cavaillon JM. Variable estimates of cytokine levels produced by commercial ELISA kits: results using international cytokine standards. J Immunol Methods. 1995; 186: 171179.[CrossRef][Medline] [Order article via Infotrieve]
8. Mire-Sluis AR, Gaines Das R, Thorpe R. Implications for the assay and biological activity of interleukin-8: results of a WHO international collaborative study. J Immunol Methods. 1997; 200: 116.[CrossRef][Medline] [Order article via Infotrieve]
9. Jacobs DS, Demott WR, Grady HJ, Horvat RT, Huestis DW, Kasten BI Jr, eds. Laboratory Test Handbook. 4th ed. Hudson, Ohio: Lexi-Comp; 1996: 246247.
10. Heinrich J, Assmann G. Fibrinogen and cardiovascular risk. J Cardiovasc Risk. 1995; 2: 197205.[CrossRef][Medline] [Order article via Infotrieve]
11. Whitton CM, Sands D, Hubbard AR, Gaffney PJ. A collaborative study to establish the 2nd International Standard for Fibrinogen, Plasma. Thromb Haemost. 2000; 84: 258262.[Medline] [Order article via Infotrieve]
12. Chitolie A, Mackie IJ, Grant D, Hamilton JL, Machin SM. Inaccuracy of the derived fibrinogen measurement. Blood Coagul Fibrinolysis. 1994; 5: 955957.[Medline] [Order article via Infotrieve]
13. De Cristofaro R, Landolfi R. Measurement of plasma fibrinogen concentration by the prothrombin-time-derived method: applicability and limitations. Blood Coagul Fibrinolysis. 1998; 9: 251259.[Medline] [Order article via Infotrieve]
14. Hirai N, Tatsumi N, Hino M, Yamane T, Ohta K. Problems with assays based on fibrin clot formation principle: analysis of multilaboratory data. Osaka City Med J. 1998; 44: 5571.[Medline] [Order article via Infotrieve]
15. N Latex SAA [package insert]. Newark, Del: Dade Behring; 2000.
16. Ledue TB, Weiner DL, Sipe JD, Poulin SE, Collins MF, Rifai N. Analytical evaluation of particle-enhanced immunonephelometric assays for C-reactive protein, serum amyloid A and mannose-binding protein in human serum. Ann Clin Biochem. 1998; 35: 745753.
17. Poole S, Walker D, Gaines Das RE, Gallimore JR, Pepys MB. The first international standard for serum amyloid A protein (SAA): evaluation in an international collaborative study. J Immunol Methods. 1998; 214: 110.[CrossRef][Medline] [Order article via Infotrieve]
18. Ledue TB, Rifai N. Preanalytic and analytic sources of variations in C-reactive protein measurement: implications for cardiovascular disease risk assessment. Clin Chem. 2003; 49: 12581271.
19. Roberts WL, Schwarz EL, Ayanian S, Rifai N. Performance characteristics of a point of care C-reactive protein assay. Clin Chim Acta. 2001; 314: 255259.[CrossRef][Medline] [Order article via Infotrieve]
20. Roberts WL, Sedrick R, Moulton L, Spencer A, Rifai N. Evaluation of four automated high-sensitivity C-reactive protein methods: implications for clinical and epidemiological applications. Clin Chem. 2000; 46: 461468.
21. Roberts WL, Moulton L, Law TC, Farrow G, Cooper-Anderson M, Savory J, Rifai N. Evaluation of nine automated high-sensitivity C-reactive protein methods: implications for clinical and epidemiological applications. Part 2. Clin Chem. 2001; 47: 418425.
22. Hamwi A, Vukovich T, Wagner O, Rumpold H, Spies R, Stich M, Langecker C. Evaluation of turbidimetric high-sensitivity C-reactive protein assays for cardiovascular risk estimation. Clin Chem. 2001; 47: 20442046.
23. Rothkrantz-Kos S, Schmitz MP, Bekers O, Menheere PP, van Dieijen-Visser MP. High-sensitivity C-reactive protein methods examined. Clin Chem. 2002; 48: 359362.
24. Kimberly MM, Vesper HW, Caudill SP, Cooper GR, Rifai N, Dati F, Myers GL. Standardization of immunoassays for measurement of high-sensitivity C-reactive protein: phase I: evaluation of secondary reference materials. Clin Chem. 2003; 49: 611616.
25. Koch DD, Peters T. Selection and evaluation of methods. In: Burtis CA, Ashwood ER, eds. Tietz Textbook of Clinical Chemistry. 3rd ed. Philadelphia, Pa: W.B. Saunders; 1999.
26. Ehrmeyer SS, Laessig RH. The relationship of intralaboratory bias and imprecision on laboratories ability to meet medical usefulness limits. Am J Clin Pathol. 1988; 89: 1418.[Medline] [Order article via Infotrieve]
27. Carey RN, Garber CC. Evaluations of methods. In: Kaplan LA, Pesce AJ, eds. Clinical Chemistry, Theory, Analysis and Correlation. St. Louis, Mo: C.V. Mosby; 1989.
28. Clark GH, Fraser CG. Biological variation of acute phase proteins. Ann Clin Biochem. 1993; 30: 373376.
29. Macy EM, Hayes TE, Tracy RP. Variability in the measurement of C-reactive protein in healthy subjects: implications for reference intervals and epidemiological applications. Clin Chem. 1997; 43: 5258.
30. Kluft C, de Maat MP. Determination of the habitual low blood level of C-reactive protein in individuals. Ital Heart J. 2001; 2: 172180.[Medline] [Order article via Infotrieve]
This article has been cited by other articles:
![]() |
D Gimeno, J E Ferrie, M Elovainio, L Pulkki-Raback, L Keltikangas-Jarvinen, C Eklund, M Hurme, T Lehtimaki, J Marniemi, J S A Viikari, et al. When do social inequalities in C-reactive protein start? A life course perspective from conception to adulthood in the Cardiovascular Risk in Young Finns Study Int. J. Epidemiol., April 1, 2008; 37(2): 290 - 298. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Wei, A. T. Whaley-Connell, K. Chen, J. Habibi, G. M.-E. Uptergrove, S. E. Clark, C. S. Stump, C. M. Ferrario, and J. R. Sowers NADPH Oxidase Contributes to Vascular Inflammation, Insulin Resistance, and Remodeling in the Transgenic (mRen2) Rat Hypertension, August 1, 2007; 50(2): 384 - 391. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Jeevanantham, N. Singh, K. Izuora, J. P. D'souza, and D. H. Hsi Correlation of High Sensitivity C-Reactive Protein and Calcific Aortic Valve Disease Mayo Clin. Proc., February 1, 2007; 82(2): 171 - 174. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Juonala, J. S.A. Viikari, T. Ronnemaa, L. Taittonen, J. Marniemi, and O. T. Raitakari Childhood C-Reactive Protein in Predicting CRP and Carotid Intima-Media Thickness in Adulthood: The Cardiovascular Risk in Young Finns Study Arterioscler Thromb Vasc Biol, August 1, 2006; 26(8): 1883 - 1888. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Mora, N. Rifai, J. E. Buring, and P. M Ridker Additive Value of Immunoassay-Measured Fibrinogen and High-Sensitivity C-Reactive Protein Levels for Predicting Incident Cardiovascular Events Circulation, August 1, 2006; 114(5): 381 - 387. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. S. Vasan Biomarkers of Cardiovascular Disease: Molecular Basis and Practical Considerations Circulation, May 16, 2006; 113(19): 2335 - 2362. [Full Text] [PDF] |
||||
![]() |
T. A. Pearson, G. A. Mensah, Y. Hong, and S. C. Smith Jr CDC/AHA Workshop on Markers of Inflammation and Cardiovascular Disease: Application to Clinical and Public Health Practice: Overview Circulation, December 21, 2004; 110(25): e543 - e544. [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2004 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |