(Circulation. 2004;110:3003-3010.)
© 2004 American Heart Association, Inc.
Arrhythmia/Electrophysiology |
From the Division of Cardiology (L.L., C.V., D.B., V.J., Z.E., T.D., D.D., R.B., H.C.) and the Department of Radiology and Radiological Science (M.M., D.K.), The Johns Hopkins Hospital, Baltimore, Md, and the Department of Medicine-Cardiology, University of Bonn, Bonn, Germany (L.L., B.L.).
Correspondence to Mahadevappa Mahesh, MS, PhD, The Russell H. Morgan Department of Radiology and Radiological Science, Johns Hopkins University, 601 N Caroline St, Baltimore, MD 21287-0856. E-mail mmahesh{at}jhmi.edu
Received March 23, 2004; revision received June 16, 2004; accepted June 17, 2004.
| Abstract |
|---|
|
|
|---|
Methods and Results The study included 15 patients with AF and 5 patients each with atrial flutter and atrioventricular nodal reentrant tachycardia (AVNRT) who underwent fluoroscopically guided procedures on a biplane x-ray system operated at a low-frame pulsed fluoroscopy (7.5 frames per second). Radiation exposure was measured directly with 50 to 60 thermoluminescent dosimeters (TLDs). Peak skin doses (PSDs), effective radiation doses, and risk of fatal malignancies were all computed. Mean fluoroscopy durations for AF procedures were 67.8±21 minutes in the right anterior oblique (RAO) and 61.9±16.6 minutes in the left anterior oblique (LAO) projection, significantly different from that required for atrial flutter and AVNRT. The mean PSDs measured with the TLDs were 1.0±0.5 Gy in the RAO and 1.5±0.4 Gy in the LAO projection. The lifetime risk of excess fatal malignancies normalized to 60 minutes of fluoroscopy was 0.07% for women and 0.1% for men.
Conclusions The relatively small amounts of the patients radiation exposure in this study, despite the prolonged fluoroscopy durations, can be attributed to the use of very-low-frame pulsed fluoroscopy, the avoidance of magnification, and optimal adjustments of the fluoroscopy exposure rates. The resulting lifetime risk of fatal malignancy is within the range previously reported for standard supraventricular arrhythmias.
Key Words: catheter ablation radiography radiation
| Introduction |
|---|
|
|
|---|
| Methods |
|---|
|
|
|---|
All patients were asked to contact us if a skin rash developed. In addition, each patient was contacted by phone 2 weeks after ablation to inquire regarding evidence of radiation skin injury. Patients were also seen in the clinic 2 to 3 months after their ablation procedure.
Electrophysiology Testing and RF Ablation
Catheter ablation was performed by use of the double transseptal approach. Contrast venography of all PVs was performed.14 A 64-pole basket catheter was used to map the PVs. PV potentials were targeted with RF energy.15 RF was also applied to the cavotricuspid isthmus. The approach used for AF ablation did not differ regardless of whether a patient had paroxysmal or persistent AF. RF ablation of atrial flutter and AVNRT was performed in the standard manner.
Fluoroscopic System and Radiation Measurement
Ablations were performed on a biplane fluoroscopy system (BICOR HS, Siemens Medical Systems). The system was set to pulsed mode at 7.5 frames per second. The image intensifier diameter of 23 cm was used for all subjects. Collimation and magnification was not used during the procedure. The x-ray beam size on the back of the patients skin surface was
12 and 13 cm in diameter for the right anterior oblique (RAO) and left anterior oblique (LAO) projections. Tube potential ranged from 70 to 109 kV peak (kVp) and tube current of 1 to 10 mA with maximum output of 8.7 R/min as measured at 30 cm from the image intensifier. The half-value layer (HVL) was 4.1 and 4.8 mm of aluminum at 80 kVp, respectively, for anteroposterior (RAO) and lateral (LAO) planes. Physics evaluations were performed at the onset of the study and every 6 months thereafter to ensure proper operational conditions. The radiation dose distribution on subjects was measured by use of thermoluminescent dosimeter (TLD) sensors. The TLD sensors were single-element TLD-100 lithium fluoride (LiF) nominally 3x3x1 mm in size (Landauer, Inc), calibrated to diagnostic energies (70 to 140 kVp). Individually wrapped TLD sensors were placed in the pockets of a custom-made vest (Figure 1A) and positioned on the back of the patient. We refer to this as the "TLD vest." Depending on patient size, between 50 and 60 TLDs were distributed at regular intervals (3 to 5 cm) to cover the entire region (Figure 1B) facing the fluoroscopy source. In addition, 2 TLD sensors were taped to the center of the collimator on each x-ray tube, because the system lacked any type of online dose monitors. To verify that the radiation field was covered completely by TLDs during the ablation procedure, Ready-Pack (Kodak X-Omat) film was placed behind the vest. The top and bottom parts of the vest edges were marked with a pin on the film for later verification.
|
The patients demographic information is recorded in Table 1. The patients anteroposterior and lateral thickness in the chest region was measured with a caliper. During the ablation procedure, the source-toimage intensifier distance (SID) and source-to-skin distance (SSD) were recorded along with the tube angulations (Table 1). We adjusted the LAO and RAO angulations anatomically in accordance with the patients orientation of the interventricular septum. As a result, there is a slight variation in the RAO and LAO angulations between patients. The voltage (kVp), current (mA), and total fluoroscopy duration for each plane were recorded.
|
Estimation of Deterministic Risk From Radiation Exposure
The TLD sensors from subjects and collimator were sent out for analysis (Landauer, Inc). The risks for radiation-induced skin injuries were calculated by 3 methods. The "direct method" measured radiation exposure with the TLD sensors placed on the subjects by use of the TLD vest. The maximum value for each projection is tabulated as peak skin dose (PSD). The "extrapolation method" is the dose measured with the TLD sensors positioned on the collimator and extrapolated to the subjects skin surface on the basis of measured SSD. The "estimation method" was based on fluoroscopic times and technical parameters such as kVp, mA, SSD, SID, and others12 recorded for the procedure.
Estimation of Carcinogenic Risk From Radiation Exposure
Organ doses were computed for average men and women by the organ-dose methodology.16 For men, an HVL of 4.0 mm Al was assumed, corresponding to a 90-kVp beam with 4.0 mm Al total filtration. For women, an HVL of 3.5 mm Al was used, corresponding to 80 kVp with a total filtration of 3.9 mm Al. Three different sets of organ doses were calculated on the basis of the radiation estimates derived with the direct, extrapolation, and estimation approaches as described earlier. Once estimated organ doses had been computed for each subject, the effective doses were computed by use of International Commission on Radiological Protection 6017 tissue-weighing factors. The risk of a fatal malignancy in specific tissues was obtained from organ-dose data by use of risk estimates compiled in the Biological Effects of Ionizing Radiation Committee Report.18 The risk of fatal cancers was also estimated from the total effective dose using an average lifetime risk factor of 4% per Sv.19
Statistical Analysis
All variables are given as mean±SD. ANOVA was used to determine the significance level of differences in several parameters with regard to the ablation target. Bonferronis method was used for post hoc comparisons in ANOVA. A value of P<0.05 was considered statistically significant. All statistical analyses were performed using STATA software, version 8.0 (STATA Corp). Radiation doses are expressed in standardized international units.
| Results |
|---|
|
|
|---|
Fluoroscopy Time
Shown in Figure 2A and summarized in Table 2 are the fluoroscopy durations. The total fluoroscopy duration received by patients undergoing ablation of AF was at least 4-fold greater than during ablation of atrial flutter and at least 8-fold greater than during ablation of AVNRT.
|
|
Radiation Exposure
The PSDs as measured with the TLD vest are shown in Figure 2B. The mean PSD received by patients undergoing AF ablation (1.00±0.5 and 1.48±0.37 Gy in the RAO and LAO projections, respectively) was at least 3-fold greater than that received by patients undergoing ablation of atrial flutter and AVNRT. Despite the prolonged fluoroscopy times during AF ablation, only 1 exceeded the 2-Gy threshold dose required for the onset of radiation-induced skin injuries.20 Despite this, no patient in our study had any clinical evidence of radiation skin injury. The higher PSD in the LAO compared with the RAO projection is because of the orientation of the x-ray tube, resulting in a longer attenuation path for the x-ray beam in the LAO than in the RAO projection.
The skin dose distribution for a typical AF ablation procedure as measured with TLD sensors is shown in Figure 3A. The Ready-Pack direct exposed film (Figure 3B) verified the positioning of the TLD sensors within the x-ray field and demonstrates that the radiation entered the skin covering a small region on the back of the subject
16x10 cm in the RAO projection and 14x15 cm in the LAO projection. In Figure 3A, the fluoroscopic duration was 77.6 minutes in the RAO and 76.5 minutes in LAO, and the corresponding PSDs were 0.90 and 1.67 Gy in the RAO and LAO projections, respectively.
|
In addition to direct measurements of skin doses by use of the TLD vest, the skin doses were also estimated by the extrapolation and estimation methods described previously. The mean skin doses during AF, atrial flutter, and AVNRT ablation in both the RAO and LAO projections are summarized in the top part of Table 3. The absorbed skin doses as estimated by the extrapolation method overestimated by 72% and 66% for the RAO and LAO projections, respectively, compared with direct method. The absorbed skin doses as estimated by the estimation method overestimated by 57% and 108% for the RAO and LAO projections, respectively, compared with direct method. An excellent correlation between the 3 methods was observed in both the RAO and LAO projections (Figure 4).
|
|
Cancer Risk
The effective radiation doses calculated on the basis of the PSD measured with the TLD vest as well as the extrapolation and estimation methods is tabulated in the bottom part of Table 3.
Estimates of organ doses and the associated lifetime risk of excess fatal malignancies per 1 000 000 patients undergoing RF catheter ablation of AF is shown in Table 4. Data for risk estimates normalized to 60 minutes of fluoroscopy are based on the BEIR V report for leukemia, breast, lung, and stomach cancers using estimated doses to the organs.18 Radiation exposures received during catheter ablation may also result in other types of fatal cancers, such as bone or liver; however, the incidence would be much lower than for those included in Table 4.
|
| Discussion |
|---|
|
|
|---|
1 hour of fluoroscopy exposure was delivered to patients in both the RAO and LAO projections. A second finding is that skin radiation exposure peaks at small regions
3 to 5 cm in diameter in the LAO and RAO projections. Despite similar fluoroscopy durations in both projections, we found that radiation dose was greater in the LAO than in the RAO projection, primarily because of a longer attenuation path for the x-ray beam entering in the LAO projection. Third, despite very long fluoroscopy times of up to 99 minutes in one projection, the resulting PSD approached the threshold for acute skin injury (2 Gy) in only 1 of 15 patients. Fourth, although approaches used to estimate radiation exposure result in a 50% to 100% overestimation of true radiation exposure, a good correlation exists between the skin dose measured directly with TLD sensors on the skin and the extrapolated dose based on a limited number of TLDs attached to the collimator of the x-ray tube and the estimated dose based on technical factors. And finally, the estimated lifetime risk of a fatal malignancy after PV ablation using a modern low-frame pulsed fluoroscopy system is relatively low.
Fluoroscopy Exposure
The duration of fluoroscopy for patients undergoing catheter ablation of AF exceeded that of patients undergoing catheter ablation of atrial flutter or AVNRT by >4-fold. This finding most likely reflects 2 main factors. First, ablation of AF is far more complex compared with ablation of atrial flutter and AVNRT, requiring a double transseptal puncture and extensive mapping and ablation. Second, ablation of AF is a relatively new procedure, whereas ablation of atrial flutter and AVNRT has become extremely routine. The amount of fluoroscopy received by patients undergoing catheter ablation of AF in the present study (LAO, 67.8±21 minutes; RAO, 61.9±16.6 minutes) is greater than but in a similar range to that previously been reported (57±30 minutes,21 79±29 minutes22). This difference can probably be attributed to the experience with AF ablation at the center at the time these studies were performed, as well as the fact that our ablation approach was somewhat more extensive, requiring both isolation of all 4 PVs and creation of a cavotricuspid isthmus ablation lesion.
Radiation Exposure
Skin injuries are well-recognized complications for catheter ablation procedures.68 Acute radiation-induced skin injury is subdivided into first-, second-, and third-degree reactions.7 The threshold for transient erythema and epilation is 2 to 3 Gy.7 In our study, only 1 of 15 patients reached a PSD of 2 Gy, despite prolonged fluoroscopy durations. This finding is striking and speaks to the effectiveness of the systems that use several technologies, such as last image hold, pulsed fluoroscopy, and additional filters to reduce radiation exposure. This also explains why radiation skin damage is an extremely rare complication of ablation procedures. In this study, the mean PSD for catheter ablation of AF was 1.00±0.5 and 1.48±0.37 Gy in the RAO and LAO projections, respectively. To date, only 1 previous study by Macle et al21 has reported radiation exposure for patients undergoing ablation of AF. They reported a median radiation exposure nearly 1000-fold lower than reported in our study (0.0011 versus 1 Gy [RAO] and 1.48 Gy [LAO]). Although it is difficult to reconcile these differences, the most likely basis for this is the differing approaches to the measurement of radiation. In the present study, a large number of TLDs with a TLD vest was used to measure PSD, whereas Macle et al21 used a single electronic personal dosimeter positioned on the anterior chest at the level of the xiphoid process. It is therefore highly likely that the single electronic dosimeter was not at the focal point of the radiation beam and therefore greatly underestimated the true PSD. One of the major concerns for studies using TLDs is that the radiation field may not be fully covered by dosimeters, and therefore, a complete reading may not be provided. This limitation was overcome in this study by placing a radiographic film (Ready-Pack) behind the vest. The exposure on the film ensured that the TLDs were well within the radiation field.
Methods to Determine Radiation Exposure
A variety of methods for measurement or estimation of radiation exposure during catheter ablation procedures has been used. Calkins et al9 and Koovor et al11 used 6 and 41 TLDs, respectively, attached to the skin for direct recording of radiation exposure. A third study by Rosenthal et al12 used fluoroscopy times and technical factors of the x-ray equipment for estimation of radiation exposure. We refer to this method as the estimation method and used this calculation to compare with the direct and extrapolation methods. Yet another approach was used in studies by Lindsay et al10 and Perisinakis et al,13 in which fluoroscopy time was recorded in the individual projections during ablation, and in a separate step, an anthropomorphic phantom was loaded with dosimeters and radiated with the fluoroscopy parameters found in the human studies.
In our study, the estimation method overestimated the skin dose by 57% in the RAO and by 108% in the LAO projection compared with the direct method. However, the direct method is very time-consuming and costly because of the large numbers of TLDs needed to cover the complete skin area potentially irradiated during the ablation procedure. We therefore introduced the extrapolation method. This method uses 2 TLDs placed at the center of the x-ray tube collimator. We demonstrated that there is a significant correlation between direct PSD measurements as obtained by analysis of all 50 to 60 TLDs and the extrapolated radiation doses obtained from readings from the 2 TLDs on the collimator of each x-ray tube. In the absence of online dose monitors, this method, which can be used to estimate PSD exposure quickly and inexpensively, may be of clinical value, particularly in patients undergoing complex ablation procedures such as AF ablation, but also among those patients undergoing repeat procedures in a relatively short period of time.8
Radiation-Induced Cancer
Radiation-induced cancer is an important delayed risk associated with diagnostic and therapeutic procedures that require fluoroscopic imaging. This is of particular concern after procedures, such as ablation of AF, that require prolonged fluoroscopy guidance. The additional lifetime risk for a fatal malignancy associated with ablation of AF was 0.15% for female patients and 0.21% for male patients in our study. The respective risk normalized to 60 minutes of fluoroscopy was 0.07% for female and 0.1% for male patients. Organs contributing most to the total risk were lungs, stomach, and active bone marrow in all patients and, in addition, breast tissue in female patients. The risk for a fatal malignancy resulting from ablation of AF in this study is slightly greater, although within the range previously reported to result from ablation of standard types of supraventricular arrhythmias (0.03% to 0.26%).913 It is reassuring that despite markedly prolonged fluoroscopy durations, the advances in techniques to reduce radiation exposure have allowed the risk of a fatal malignancy resulting from catheter ablation to remain at this low level.
Study Limitations
There are several limitations in this study. First, catheter ablation of AF was performed without electroanatomic mapping systems. Ablation of AF guided by electroanatomic mapping systems has been reported to require significantly less fluoroscopy exposure. Second, as pointed out previously, this series of patients was among the first 100 patients to undergo catheter ablation of AF at our institution. As reported previously by Knight et al,22 reductions in procedure and fluoroscopy duration can be anticipated after the first 70 AF ablation procedures. Third, dose-area-product (DAP) meters were not available on the system. DAP meters are available on new interventional systems. These monitors record fluoroscopy times and provide an estimate of radiation exposure (dosearea product [cGy-cm2]), which is the total radiation dose multiplied by the radiation field area. Although this parameter correlates with total patient radiation exposure, it provides little information about the PSD, which is the cause of skin injuries. Modified versions of the DAP meters are now available that provide cumulative dose at the center of the beam, which correlates more closely with PSD. We believe that these monitors will be better suited to assess risk for skin injuries, because the user can keep track of the cumulative dose to the patients skin surface in any projection. In fact, the Food and Drug Administration is considering mandating such a real-time dose display on fluoroscopy units under the new regulations currently under review. In the absence of DAP, we placed 2 TLDs on each of the collimators and extrapolated to the skin surface, which correlated with the direct measurements. The correlation shown in our study between these measurements and PSD supports the potential clinical usefulness of on-line DAP monitors.
Clinical Implications
The results of this study demonstrate that catheter ablation of AF requires several orders of magnitude greater fluoroscopy duration and radiation exposure than conventional catheter ablation procedures. This finding, together with the fact that repeat ablation procedures are commonly required among AF patients, should compel electrophysiologists to make every attempt to minimize radiation exposure. Radiation exposure to a patient can be minimized by reducing fluoroscopy time, by using low-frame pulsed fluoroscopy, and also by reducing the rate of radiation exposure. Fluoroscopy time, as well as the need for biplane fluoroscopy, can be reduced significantly with the use of electroanatomic mapping systems.2,23 The increasing availability and familiarity of electrophysiologists with the systems are likely to have a major impact on fluoroscopy time. The relatively small amounts of a patients radiation exposure in this study, despite the extremely prolonged fluoroscopy durations, can be attributed in large measure to the use of extremely low (7.5 frames per second) fluoroscopy rates, the avoidance of magnification, and optimal adjustments of the fluoroscopy unit to minimize exposure rates. The results of this study also validate the effectiveness of methods that estimate or extrapolate patient radiation exposure. Although each of these methods results in a 50% to 100% overestimation of actual patient radiation exposure, their excellent correlation with actual patient radiation exposure speaks to their potential clinical role.
| Footnotes |
|---|
| References |
|---|
|
|
|---|
2. Pappone C, Rosanio S, Oreto G, Tocchi M, Gugliotta F, Vicedomini G, Salvati A, Dicandia C, Mazzone P, Santinelli V, Gulletta S, Chierchia S. Circumferential radiofrequency ablation of pulmonary vein ostia: a new anatomic approach for curing atrial fibrillation. Circulation. 2000; 102: 26192628.
3. Chen SA, Hsieh MH, Tai CT, Tsai CF, Prakash VS, Yu WC, Hsu TL, Ding YA, Chang MS. Initiation of atrial fibrillation by ectopic beats originating from the pulmonary veins: electrophysiological characteristics, pharmacological responses, and effects of radiofrequency ablation. Circulation. 1999; 100: 18791886.
4. Oral H, Knight BP, Tada H, Ozaydin M, Chugh A, Hassan S, Scharf C, Lai SW, Greenstein R, Pelosi F Jr, Strickberger SA, Morady F. Pulmonary vein isolation for paroxysmal and persistent atrial fibrillation. Circulation. 2002; 105: 10771081.
5. Vasamreddy C, Lickfett L, Jayam V, Nasir K, Bradley D, Eldadah Z, Dickfeld T, Berger RD, Calkins H. Predictors of recurrence following catheter ablation of atrial fibrillation using an irrigated-tip ablation catheter. J Cardiovasc Electrophysiol. 2004; 15: 692697.[CrossRef][Medline] [Order article via Infotrieve]
6. Rosenthal LS, Beck TJ, Williams J, Mahesh M, Herman MG, Dinerman JL, Calkins H, Lawrence JH. Acute radiation dermatitis following radiofrequency catheter ablation of atrioventricular nodal reentrant tachycardia. Pacing Clin Electrophysiol. 1997; 20: 18341839.[CrossRef][Medline] [Order article via Infotrieve]
7. Nahass GT. Fluoroscopy and the skin: implications for radiofrequency catheter ablation. Am J Cardiol. 1995; 76: 174176.[CrossRef][Medline] [Order article via Infotrieve]
8. Mahesh M. Fluoroscopy: patient radiation exposure issues. Radiographics. 2001; 21: 10331045.
9. Calkins H, Niklason L, Sousa J, el-Atassi R, Langberg J, Morady F. Radiation exposure during radiofrequency catheter ablation of accessory atrioventricular connections. Circulation. 1991; 84: 23762382.
10. Lindsay BD, Eichling JO, Ambos HD, Cain ME. Radiation exposure to patients and medical personnel during radiofrequency catheter ablation for supraventricular tachycardia. Am J Cardiol. 1992; 70: 218223.[CrossRef][Medline] [Order article via Infotrieve]
11. Kovoor P, Ricciardello M, Collins L, Uther JB, Ross DL. Risk to patients from radiation associated with radiofrequency ablation for supraventricular tachycardia. Circulation. 1998; 98: 15341540.
12. Rosenthal LS, Mahesh M, Beck TJ, Saul JP, Miller JM, Kay N, Klein LS, Huang S, Gillette P, Prystowsky E, Carlson M, Berger RD, Lawrence JH, Yong P, Calkins H. Predictors of fluoroscopy time and estimated radiation exposure during radiofrequency catheter ablation procedures. Am J Cardiol. 1998; 82: 451458.[CrossRef][Medline] [Order article via Infotrieve]
13. Perisinakis K, Damilakis J, Theocharopoulos N, Manios E, Vardas P, Gourtsoyiannis N. Accurate assessment of patient effective radiation dose and associated detriment risk from radiofrequency catheter ablation procedures. Circulation. 2001; 104: 5862.
14. Vasamreddy C, Jayam V, Lickfett L, Nasir K, Bradley D, Eldadah Z, Dickfeld T, Donahue K, Halperin H, Berger RD, Calkins H. Technique and results of pulmonary vein angiography in patients undergoing catheter ablation of atrial fibrillation. J Cardiovasc Electrophysiol. 2004; 15: 2126.[CrossRef][Medline] [Order article via Infotrieve]
15. Haissaguerre M, Shah DC, Jais P, Hocini M, Yamane T, Deisenhofer I, Chauvin M, Garrigue S, Clementy J. Electrophysiological breakthroughs from the left atrium to the pulmonary veins. Circulation. 2000; 102: 24632465.
16. Stern SH, Rosenstein M, Renaud L, Zankl M. Handbook of Selected Tissue Doses for Fluoroscopic and Cineangiographic Examination of the Coronary Arteries. Rockville, Md: US Department of Health and Human Services; 1995.
17. International Commission on Radiological Protection. 1990 Recommendations of the International Commission on Radiological Protection. ICRP Publication 60. New York, NY: Pergamon Press; 1991: 46.
18. Committee on the Biological Effects of Ionizing Radiation. Health Effects of Exposure to Low Levels of Ionizing Radiation. BEIR V Report. Washington, DC: National Academy Press; 1990: 175.
19. Hall EJ. Radiobiology for the Radiologist. Baltimore, Md: Lippincott Williams & Williams; 2000.
20. US Food and Drug Administration. Avoidance of Serious X-Ray Induced Skin Injuries to Patients During Fluoroscopically-Guided Procedures. Washington, DC: 1994.
21. Macle L, Weerasooriya R, Jais P, Scavee C, Raybaud F, Choi KJ, Hocini M, Clementy J, Haissaguerre M. Radiation exposure during radiofrequency catheter ablation for atrial fibrillation. Pacing Clin Electrophysiol. 2003; 26: 288291.[CrossRef][Medline] [Order article via Infotrieve]
22. Knight BP, Oral H, Chugh A, Scharf C, Lai SW, Pelosi F Jr, Strickberger SA, Morady F. Effects of operator experience on the outcome and duration of pulmonary vein isolation procedures for atrial fibrillation. Am J Cardiol. 2003; 91: 673677.[CrossRef][Medline] [Order article via Infotrieve]
23. de Groot N, Bootsma M, van der Velde ET, Schalij MJ. Three-dimensional catheter positioning during radiofrequency ablation in patients: first application of a real-time position management system. J Cardiovasc Electrophysiol. 2000; 11: 11831192.[CrossRef][Medline] [Order article via Infotrieve]
This article has been cited by other articles:
![]() |
E. M. Aliot, W. G. Stevenson, J. M. Almendral-Garrote, F. Bogun, C. H. Calkins, E. Delacretaz, P. D. Bella, G. Hindricks, P. Jais, M. E. Josephson, et al. EHRA/HRS Expert Consensus on Catheter Ablation of Ventricular Arrhythmias: Developed in a partnership with the European Heart Rhythm Association (EHRA), a Registered Branch of the European Society of Cardiology (ESC), and the Heart Rhythm Society (HRS); in collaboration with the American College of Cardiology (ACC) and the American Heart Association (AHA) Europace, June 1, 2009; 11(6): 771 - 817. [Full Text] [PDF] |
||||
![]() |
T. Moritake, Y. Matsumaru, T. Takigawa, K. Nishizawa, A. Matsumura, and K. Tsuboi Dose Measurement on Both Patients and Operators during Neurointerventional Procedures Using Photoluminescence Glass Dosimeters AJNR Am. J. Neuroradiol., November 1, 2008; 29(10): 1910 - 1917. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. A. Mettler Jr, W. Huda, T. T. Yoshizumi, and M. Mahesh Effective Doses in Radiology and Diagnostic Nuclear Medicine: A Catalog Radiology, July 1, 2008; 248(1): 254 - 263. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Ector, O. Dragusin, B. Adriaenssens, W. Huybrechts, R. Willems, H. Ector, and H. Heidbuchel Obesity Is a Major Determinant of Radiation Dose in Patients Undergoing Pulmonary Vein Isolation for Atrial Fibrillation J. Am. Coll. Cardiol., July 17, 2007; 50(3): 234 - 242. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Handke, G. Heinrichs, U. Moser, F. Hirt, F. Margadant, F. Gattiker, C. Bode, and A. Geibel Transesophageal Real-Time Three-Dimensional Echocardiography: Methods and Initial In Vitro and Human In Vivo Studies J. Am. Coll. Cardiol., November 21, 2006; 48(10): 2070 - 2076. [Abstract] [Full Text] [PDF] |
||||
![]() |
N Theocharopoulos, J Damilakis, K Perisinakis, E Manios, P Vardas, and N Gourtsoyiannis Occupational exposure in the electrophysiology laboratory: quantifying and minimizing radiation burden Br. J. Radiol., August 1, 2006; 79(944): 644 - 651. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. P. Efstathopoulos, D. G. Katritsis, S. Kottou, N. Kalivas, E. Tzanalaridou, E. Giazitzoglou, S. Korovesis, and K. Faulkner Patient and staff radiation dosimetry during cardiac electrophysiology studies and catheter ablation procedures: a comprehensive analysis Europace, June 1, 2006; 8(6): 443 - 448. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. M. Scheinman and E. Keung The year in clinical electrophysiology J. Am. Coll. Cardiol., March 1, 2005; 45(5): 790 - 795. [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2004 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |