Donate Help Contact The AHA Sign In Home
American Heart Association
Circulation
Search: search_blue_button Advanced Search
Circulation. 2004;110:2609-2617
Published online before print October 18, 2004, doi: 10.1161/01.CIR.0000146818.60588.40
This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
110/17/2609    most recent
01.CIR.0000146818.60588.40v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow Request Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Weinberg, C. E.
Right arrow Articles by Shandas, R.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Weinberg, C. E.
Right arrow Articles by Shandas, R.
Related Collections
Right arrow Pulmonary circulation and disease
Right arrow Echocardiography
Right arrow Pediatric and congenital heart disease, including cardiovascular surgery

(Circulation. 2004;110:2609-2617.)
© 2004 American Heart Association, Inc.


Congenital Heart Disease

Extraction of Pulmonary Vascular Compliance, Pulmonary Vascular Resistance, and Right Ventricular Work From Single-Pressure and Doppler Flow Measurements in Children With Pulmonary Hypertension: a New Method for Evaluating Reactivity

In Vitro and Clinical Studies

Craig E. Weinberg, PhD; Jean R. Hertzberg, PhD; D. Dunbar Ivy, MD; K. Scott Kirby, RDCS; K. Chen Chan, MD; Lilliam Valdes-Cruz, MD; Robin Shandas, PhD

From the Department of Mechanical Engineering (C.E.W., J.R.H., R.S.), University of Colorado, Boulder, and the Children’s Hospital/University of Colorado Health Sciences Center (D.D.I., K.S.K., K.C.C., L.V.-C., R.S.), Denver, Colo.

Correspondence to Robin Shandas, PhD, Cardiovascular Flow Research Laboratory, The Children’s Hospital/UC Health Sciences Center, 1056 E 19th Ave, B-100, Denver, CO 80218. E-mail shandas.robin{at}tchden.org

Received May 22, 2004; revision received August 2, 2004; accepted August 18, 2004.


*    Abstract
up arrowTop
*Abstract
down arrowIntroduction
down arrowMethods
down arrowResults
down arrowDiscussion
down arrowReferences
 
Background— Current evaluation of pulmonary hypertension (PH) in children involves measurement of pulmonary vascular resistance (PVR); however, PVR neglects important pulsatile components. Pulmonary artery (PA) input impedance and ventricular power (VP) include mean and pulsatile effects and have shown promise as alternative measures of vascular function. Here we report the utility of pulsed-wave (PW) Doppler-measured instantaneous flow and pressure measurements for estimation of input impedance and VP and use this method to develop a novel parameter: reactivity in compliance.

Methods and Results— An in vitro model of the general pulmonary vasculature was used to obtain impedance and VP, measured by PW Doppler and a reference flow meter. The method was then tested in a preliminary clinical study in subjects with normal PA hemodynamics (n=4) and patients with PH undergoing reactivity evaluation (8 patients; 23 data points). In vitro results showed good agreement between the impedance spectra computed from both flow-measurement methods. Excellent correlation was seen in vitro between actual resistance and the zero-frequency (Zo) impedance value (r2=0.984). Excellent agreement was also found between Zo and PVR in the clinical measurements (y=1.075x+0.73; r=0.993). Furthermore, total VP and VP/cardiac output increased significantly with hypertension (128.73 to 365.91 mW and 2.42 to 6.69 mW · mL–1 · s–1, respectively). The first-harmonic value of impedance (Z1) was used as a measure of compliance reactivity; older patients exhibited markedly less compliance reactivity than did younger patients.

Conclusions— Input impedance and VP calculated from Doppler measurements and a single-catheter pressure measurement provide comprehensive characterization of PH and reactivity.


Key Words: hypertension, pulmonary • pulmonary heart disease • pediatrics • echocardiography • hemodynamics


*    Introduction
up arrowTop
up arrowAbstract
*Introduction
down arrowMethods
down arrowResults
down arrowDiscussion
down arrowReferences
 
The management of children with primary and secondary pulmonary hypertension (PH) continues to be a challenge for clinical cardiologists. Several recent advances such as the use of pharmaceutical agents to decrease pulmonary artery (PA) pressures and pulmonary vascular resistance (PVR) have facilitated the treatment of such patients.1–5 Although assessment of the hemodynamic status of the pulmonary vasculature is a major component in treatment evaluation, the question of how best to do this is still unanswered.6,7 Current methods evaluate hypertension and vascular reactivity through measurement of PVR, which is the ratio of mean pressure drop across the pulmonary vasculature to mean pulmonary flow and which is based on the assumption of steady hemodynamics.8 This technique can be problematic, because right ventricular afterload (RVA) and consequently, PA hemodynamics can change owing to acute and chronic variations in proximal artery compliance and pressure and flow pulse reflections, neither of which is considered in current measurements.9–11

The acquisition and processing of instantaneous pressure and flow measurements allow a more comprehensive characterization of PH and reactivity. Two parameters in particular, input impedance and PV power, can be obtained from instantaneous flow and pressure measurements. The input impedance spectrum, defined as the ratio of harmonic pressure to flow, characterizes RVA comprehensively, reflecting proximal arterial compliance, distal VR, and wave-transmission properties.9,10,12–14 Changes in these factors affect the impedance spectrum by altering the frequency response in both magnitude and phase.9,13 RV power (the time rate of RV work), consisting of both a mean component and an oscillatory (pulsatile) component, quantifies the ventricular power needed to move blood through the pulmonary vasculature and may be a useful measure of RV pump capacity.9,12,14,15 These parameters, by themselves or in combination, should provide a more comprehensive means of evaluating PH and assessing vascular reactivity than is currently available through PVR measurements alone. However, although instantaneous pressure is currently measured in the catheterization laboratory during evaluation, instantaneous flow is not measured; both are needed to compute the aforementioned factors. Prior studies measuring impedance and RV power used invasive cuff-type flow probes, which are not an option for routine clinical studies. The goals of this investigation were (1) to determine whether the addition of a simple-to-use Doppler method for measuring instantaneous flow would provide a clinically viable means of evaluating PA input impedance and RV power and (2) to develop and test a new parameter, reactivity in compliance, obtained from the impedance data, to assess vascular compliance changes with clinical challenge.


*    Methods
up arrowTop
up arrowAbstract
up arrowIntroduction
*Methods
down arrowResults
down arrowDiscussion
down arrowReferences
 
In Vitro Study
The experimental system consisted of multiple components working synchronously to reproduce hypertensive PA hemodynamics (see Figure 1). A model RV driven by a computer-controlled pump (VSI Superpump, Vivitro Systems Inc) delivered pulsatile flow at 90 beats/min and variable stroke volumes (16 to 26 mL) to the mock pulmonary test section. Distal flow resistance was adjusted with a hydraulic resistance section downstream of the compliant mock artery.16 To study different compliance values, 3 mock arteries of varying thickness (0.20, 0.27, and 0.51 mm) and compliance (38.31%, 22.58%, and 10.81% diametric change/100 mm Hg) and constant internal diameter (1.9 cm) were constructed from a silicone elastomer, as described previously.16 Local compliance was quantified by dynamic compliance, Cdyn=(DSDD)/[DD(PSPD)]x104, where D=diameter, P=pressure, and subscripts S and D indicate systolic and diastolic conditions, respectively.17 A 33% (vol/vol) glycerin-water mixture with a viscosity of 3.5 to 4.0 cP was used to simulate blood.18 Cornstarch (1% by weight) was added to the solution to enhance ultrasound backscatter.



View larger version (21K):
[in this window]
[in a new window]
 
Figure 1. Schematic of in vitro experimental apparatus constructed to simulate pediatric PA system. Abbreviations are as defined in text.

Pressure, velocity, and flow were measured simultaneously at the same location in the proximal mock artery. Time calibration was performed for all acquisition systems to correct any time lags introduced by each measurement technique. The pressure measurements were made with a solid-state transducer (Millar Inc). Instantaneous flow was monitored by an ultrasonic transit-time flowmeter (Transsonics). Instantaneous velocity was acquired though PW Doppler ultrasound measurements with a commercially available ultrasound scanner (Vivid 5, GE Medical Systems Inc) and postprocessing software (EchoDisp, GE Medical Systems Inc). Actual VR for the in vitro model was computed from pressure-drop and flow measurements. All pressure and flow measurements were digitized with a data acquisition system (LabView, National Instruments) sampled at 500 Hz. A total of 22 experimental conditions were produced with the 3 tube models and varying mean pressures.

To compute the instantaneous flow rate from the PW Doppler measurements, the measured velocity must be multiplied by a scaling factor that corresponds to local instantaneous cross-sectional area and the coefficient of velocity. A correction factor for area, Acorr, was computed by recognizing that the mean flow during 1 cardiac cycle must be equal between the 2 measuring devices; thus, Acorr=Qm/Vpw, where Qm is the time-averaged flow from the perivascular flow meter and Vpw is the time-averaged PW Doppler velocity. Doppler flow [Q(t)PW] was then calculated as Down


{12MM1}

Clinical Study
Patient Selection
After institutional review board approval and informed consent and assent (where appropriate) had been obtained, patients underwent cardiac catheterization for routine evaluation and treatment at The Children’s Hospital, Denver, Colo. A limited number of patients were considered; these were divided into 2 groups. Group 1 studies were conducted after closure of a small patent ductus arteriosis. These patients had normal mean PA pressures and were considered the control group (n=4). Group 2 consisted of patients with PH who were undergoing tests for PV reactivity with the use of oxygen and/or nitric oxide (8 patients; n=23 conditions). Table 1 provides hemodynamic data for each subject. For the reactivity studies, the initial data point was considered the baseline value, and each subsequent challenge was denoted as challenge 1, challenge 2, etc.


View this table:
[in this window]
[in a new window]
 
TABLE 1. Patient Hemodynamic Data, Including Clinical Challenge (Name and Type)

Clinical Data Acquisition
PW Doppler velocity measurements were taken within the middle section of the main PA with a commercial ultrasound scanner (Vivid 5) from a parasternal short-axis window. Two-dimensional echo and color Doppler flow imaging modalities were used to align the Doppler sample volume parallel to the main flow direction. For group 2 patients, data were obtained before and during nitric oxide and/or oxygen treatment. To determine the physiologic value of Acorr, cardiac output was used instead of Qm in the aforementioned equation. Cardiac output was measured by Fick’s method with measured oxygen consumption in cases where intracardiac shunts were in place and by thermodilution otherwise. Main PA pressure was measured with standard fluid-filled catheters (Transpac IV, Abbott Critical Care Systems). To facilitate temporal alignment of velocity and pressure signals after acquisition, pressure traces were digitized directly into the auxiliary input of the ultrasound scanner simultaneously with the PW Doppler information. Tracings were obtained for 16 seconds and transferred to an optical disc for offline analysis. All hemodynamic data were recorded before and during treatment for calculation of PVR.

Data Analysis
The recorded pressure and flow signals were separated into n individual cardiac cycles for analysis (n>35 for the in vitro flow meter system; n>20 for the in vitro PW Doppler system; and n>20 for clinical data). After filtering (low pass, with a 30-Hz cutoff) and temporal alignment of the signals, a fast Fourier transform was applied to the pressure and flow data, and the moduli and phase along with means and standard deviations of the pressure and flow spectra were calculated for each frequency bin.

Calculation of Input Impedance Spectrum
The input impedance modulus Z and phase {phi} are defined by Down


{12MM2}

Down


{12MM3}

where {omega} is the frequency, |P| is the modulus of the pressure spectrum, |Q| is the modulus of the flow spectrum, {theta} is the pressure phase, and {varphi} is the flow phase.

Four parameters were extracted from the impedance data: distal VR, characteristic resistance, an index of compliance, and hydraulic reflection coefficient. Distal VR was obtained from the modulus data at the zero harmonic (Z0). The characteristic resistance (Zc) was estimated by averaging the impedance modulus from its first minimum up to the eighth harmonic.19,20 The index of compliance was obtained from the modulus of the impedance data at the first harmonic. The reflection coefficient {Gamma} was calculated by21 Down


{12MM4}

Calculation of Ventricular Power
Ventricular power (the rate of ventricular work) comprises potential energy (pressure energy) and kinetic energy. Kinetic energy has been shown to account for a small fraction of the total ventricular power and is neglected here.9,15 The mean power (mean energy per unit time) was calculated as WM = P0·Q0, where P0 and Q0 are the zero-frequency components of the pressure and flow moduli, respectively. The oscillatory power (pulsatile energy per unit time) was calculated as Down


{12MM5}

where l represents the harmonic of the fundamental frequency.15 Finally, total RV power was calculated as WT = WM + WO. To quantify the total power expended per unit of forward flow, total power was divided by cardiac output (WT/CO)15

Statistics
With the exception of the input impedance modulus and phase (see next section), all data are presented as mean±SD. The in vitro instantaneous flow data computed from PW Doppler measurements were compared with the results from the flow meter with Student’s t test. The same t test was used to identify differences between patients with normal and high mean pulmonary pressures. The level of significance for all statistical tests was set at 5%.

Experimental Uncertainty
For impedance modulus and phase calculations, it was necessary to determine the complete experimental uncertainty to ensure that subsequent statistical comparisons were robust. Uncertainty (u) in the mean modulus due to variance in the data at a specific frequency is defined as Down


{12MM6}

where {varsigma} is the standard deviation of the data, n is the number of data points, and tn,95 is the t value calculated from Student’s t distribution with {nu} degrees of freedom and 95% confidence. The number of degrees of freedom was calculated from the Welch-Satterthwaite formula.22 The uncertainty in the impedance phase was calculated in a similar manner as described earlier.


*    Results
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
*Results
down arrowDiscussion
down arrowReferences
 
In Vitro Study
Mean pressures within the in vitro models ranged from 26.64 to 43.99 mm Hg for tube 1, 30.67 to 53.45 mm Hg for tube 2, and 29.49 to 59.56 mm Hg for tube 3. An average pulse pressure of 16.81±3.55 mm Hg was observed for all models throughout all conditions. The mean flow rate was kept constant at 1.88±0.23 L/min. Actual VR, measured as the mean pressure drop divided by mean flow within the in vitro models, ranged from 11.15 to 37.06 mm Hg · L–1 · min–1.

Figure 2 shows input impedance spectra (modulus, A; phase, B) obtained from the flow probe and PW Doppler data for tube 2 at a mean pressure of 41.98 mm Hg and a resistance of 22.35 mm Hg · L–1 · min–1. Good agreement was found in the impedance magnitudes calculated by flow probe and PW Doppler data for the Z0 component and the minimum impedance value for all tubes and conditions. However, PW Doppler overestimated the impedance modulus at higher harmonics. Similar results were seen for the other tube models. Excellent correlation was seen between Z0 and the actual VR calculated from the drop in pressure across the system (see Figure 3A).



View larger version (16K):
[in this window]
[in a new window]
 
Figure 2. Comparison of experimental input impedance spectrum modulus (A) (Z; mm Hg · L–1 · min–1) and phase (B) ({phi}, radians) generated by perivascular flowmeter (•) and spectrum calculated by PW Doppler ultrasound ({blacksquare}). Uncertainties calculated for both techniques are plotted as error bars. Abbreviations are as defined in text.



View larger version (14K):
[in this window]
[in a new window]
 
Figure 3. A, Correlation between actual VR and Z0 of input impedance (mm Hg · L–1 · min–1) from in vitro study. B and C, Impedance spectral response to changes in compliance (B, modulus; C, phase) is presented over first 8 harmonics. Uncertainties are presented as error bars along with means. •Indicates tube 1; {blacksquare}, tube 3. Abbreviations are as defined in text.

To investigate the impedance spectral response to changes in compliance independent of all other parameters, the most- and least-compliant tubes were exposed to equal hemodynamics and resistance. Figure 3B and 3C shows the impedance modulus and phase for the most-compliant and least-compliant in vitro tube models. The mean values and their uncertainties are presented. It can be seen that as compliance decreased, the impedance values at both the first and second harmonics increased. This indicates that either of these values may be useful for assessing compliance in the clinical situation.

No differences in mean power between the 2 flow measurement methods were seen (mean difference of 0.02±0.44%; P=NS). However, PW Doppler underestimated the oscillatory component by 25.10±19.14% (P≤0.05). No statistical difference was seen between total RV power calculated from the 2 flow measurement methods (mean difference of –0.26±0.48%; P=NS). Measurement of total RV power expelled for fluid motion agreed between the 2 methods (mean difference of –0.33±0.28%; P=NS).

Clinical Data
Patients 1 through 4 had normal pulmonary pressures (group 1), whereas patients 5 through 12 had PH (see Table 1). The mean cardiac index was 5.76±0.94 L · min–1 · m–2 for group 1 patients and 4.62±2.49 L · min–1 · m–2 for group 2 (P=NS). Figure 4A shows the impedance modulus, and Figure 4B shows the phase for group 1 subjects. Input impedance (magnitude and phase) for the PH patients are presented in Figure 5. For brevity, only 4 patients are presented here, but these values were representative of the population. Z0 increased with hypertension, as would be expected (see Figure 5A).



View larger version (16K):
[in this window]
[in a new window]
 
Figure 4. Pulmonary input impedance modulus (A) (Z; mm Hg · L–1 · min–1) and phase (B) ({phi}, radians) for group 1 patients are presented as function of harmonic. Means are shown with error bars illustrating uncertainty in means. •indicates patient 1; {blacksquare}, patient 2; {blacklozenge}, patient 3; and {blacktriangleup}, patient 4. Abbreviations are as defined in text.



View larger version (18K):
[in this window]
[in a new window]
 
Figure 5. Input impedance for limited number of group 2 patients is shown. A, Impedance modulus (Z; mm Hg · L–1 · min–1) and B, phase ({phi}, radians) as function of harmonics of heart rate. Means are shown, with error bars as uncertainty in means. For patient 5, 35% O2 (•); patient 8, room air ({blacksquare}); patient 9, 100% O2+40 ppm NO ({blacklozenge}); and patient 10, room air ({blacktriangleup}). Abbreviations are as defined in text.

Significant differences were seen in Z0 between normal and hypertensive patients (P≤0.05). As with the in vitro data, excellent agreement was found between Zo and hemodynamic measurement of PVR (y=1.07x+0.73, r=0.993; Figure 6). On the basis of the in vitro observations of changes in impedance values at the first and second harmonics with compliance changes, we used the percent change in impedance value at the first harmonic (Z1) with challenge as a measure of reactivity in compliance for the clinical studies. These data, along with reactivity in PVR, are presented in Figure 7, sorted by patient number, challenge, and mean PA pressure for each condition.



View larger version (21K):
[in this window]
[in a new window]
 
Figure 6. Zo vs hemodynamically measured values of PVR for all subjects. Excellent agreement was seen, indicating value of Zo in following PVR clinically. Abbreviations are as defined in text.



View larger version (31K):
[in this window]
[in a new window]
 
Figure 7. Reactivity in PVR (A) and reactivity in compliance (B, measured here as change in Z1) for PH patients subjected to various clinical challenges (see Table 1), plotted with mean PA pressure at each condition. In general, reactivity in PVR (defined as >–20% change) was correlated to reactivity in compliance (defined here as >–10% change) except for patients 8 and 9. Abbreviations are as defined in text.

All challenges except 1 produced negative changes (ie, decrease in Z1), indicating some increase in compliance with challenge. How much the compliance increased varied from patient to patient and from challenge to challenge. A reactivity threshold of –10% change in Z1 change could be used to gauge the change in compliance with reactivity. When reactivity in PVR was examined along with reactivity in compliance, as defined earlier, the data showed that most patients who were reactive in PVR were also reactive in compliance (Figure 7), with 2 exceptions: patients 8 and 9, who are discussed later. These data provide initial evidence for using input impedance to gauge reactivity in both resistance (based on changes in Zo) and compliance (based on Z1).

No differences in Zc were seen (P=NS) between group 1 and 2 data. However, the PA reflection coefficient increased significantly from 0.45±0.11 to 0.75±0.15 with PH (P≤0.05). Significant increases were seen in both mean power (group 1, 120±42 mW; group 2, 346±391 mW) and oscillatory power (group 1, 8.45±3.27 mW; group 2, 19.4±19.0 mW) with hypertension (P≤0.05 for both). Furthermore, the average total power increased from 128.73±44.84 to 365.91±409.38 mW (P≤0.05) between the 2 groups. Table 2 summarizes these data. Maximum interobserver and intraobserver variabilities in impedance were 3.2% and 3.9%, respectively.


View this table:
[in this window]
[in a new window]
 
TABLE 2. Impedance (Z0, Zc) and Power Values for the Clinical Data


*    Discussion
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
up arrowResults
*Discussion
down arrowReferences
 
This project addressed several difficulties related to the evaluation of PH in pediatric subjects, including patients undergoing reactivity testing and evaluation of new treatments. These patients are already in the catheterization laboratory during such evaluation; the intent here was to determine whether the addition of an easy-to-use Doppler measurement would provide important additional data on PV status. The input impedance (and parameters derived from impedance data) and RV power were chosen as key markers. Previous investigators, using more invasive cuff-type flow probes, have identified these as useful parameters of vascular function.23–28

Utility of PW Doppler for Impedance and Power Measurements
Although the utility of PW Doppler for cardiac output and pulmonary-to-systemic flow-ratio measurements has been examined by investigators in our laboratory and others,29–31 evaluation of Doppler as a tool for impedance and power calculations, especially in the pediatric pulmonary vasculature, has not been performed. The first concern in using Doppler methods to obtain instantaneous flow involves conversion of velocity to flow through a correction factor for local area. In the clinical studies, a mean value of 1.70±1.28 cm2 was found for Acorr. The assumption of constant area will affect the pulsatile content of the calculated flow and therefore, the final response of the impedance spectra. For most accurate calculations, the instantaneous area of the PA must be measured over time.

The second measurement problem involves the inherent higher-frequency noise found in the velocity data, which results in disagreement in high-frequency impedance spectra calculated from Doppler versus flow meter data. These differences affect parameters containing significant oscillatory components, such as oscillatory power, characteristic resistance, and reflection coefficient. No differences between Doppler-derived and flow meter-obtained data were found in mean and total ventricular power, input impedance at lower harmonics, and mean VR. Thus, it is in the evaluation of these latter parameters for which PW Doppler should be useful in the clinical setting.

Evaluation of Impedance and Power in PH
No problems were encountered in the use of PW Doppler for impedance and power measurements in the preliminary clinical studies. Although lack of prior impedance data after closure of patent ductus arteriosus prevents direct comparison of our data with those in the literature, the input impedance spectra in our clinical studies display general frequency-response characteristics and patient-to-patient variance representative of previously published in vivo data for children who were evaluated with perivascular cuff-type flow probes.23–25

Clinically, no differences were seen in characteristic resistance between patients with or without PH. In previously published studies with cuff-type flow meters by Lucas et al24 and Wilcox and Lucas, 25 the characteristic resistance for children with normal mean pulmonary pressures was reported as 1.46±0.75 and 1.43±0.88 L · min–1 · m–2, respectively. Our values were higher (normal patients at Zc=2.27±0.96 L · min–1 · m–2), presumably because of the Doppler-based overestimation that was also seen in vitro. The reflection coefficient was seen to increase significantly (from 0.43 to 0.76) with PH, although no prior data are available in the literature to verify the hypertensive value. The normal value (0.43) is consistent with that reported in the literature.14,21

Our clinical measurements of mean RV power were consistent with those in the study by Lucas et al,23 who used the more invasive flow meters. Normalization of total power by cardiac output may provide a better means of comparing values among various patients. Lucas et al23 obtained an average of 2.77 mW · mL–1 · s–1 in normalized power for patients with normal mean pulmonary pressures. This value is similar to that obtained in the group 1 subjects (=2.60 mW · mL–1 · s–1). Hypertensive patients displayed a significant increase in the ratio (6.96 mW · mL–1 · s–1). It is thus clear that significant changes in RV workload occur with the onset of PH in children; this parameter will also need to be examined in a larger clinical study.

The ability to obtain the type of data presented here from a single pressure measurement (ie, without wedge pressure measurements) and Doppler allows evaluation of not only distal vascular bed reactivity but also proximal vascular reactivity through the examination of how compliance varies with mean PA pressure. With PH, compliance of the proximal PAs decreases because of 2 mechanisms: an acute decrease as a result of increased PA pressures (strain stiffening) and a chronic decrease because of structural changes (fibroblast proliferation, increase in medial thickness, overexpression of structural proteins, etc) in the arterial wall as the PA adapts to the chronically higher pressures. Compliance can be increased to normal values in strain-stiffened arteries simply by lowering mean PA pressure; however, chronically adapted arteries will require long-term treatment to reverse the structural remodeling. Measuring PVR alone will not differentiate these scenarios; however, input impedance should allow evaluation of proximal PA reactivity (through compliance) as well as distal bed reactivity (through Z0). If compliance changes with vasodilator therapy, this would indicate that the proximal PAs have not undergone structural remodeling. Conversely, if compliance does not change with mean PA pressure, this would indicate that structural alterations have taken place. Here we propose use of the first-harmonic value of impedance (Z1) as an index of pulmonary vascular compliance. Based on the in vitro data, it is clear that this value (as well as the second-harmonic value) varies with compliance.

When applied to the reactivity studies, we found that Z1 decreased for all but 1 challenge, indicating increasing compliance with vasodilator challenge. The degree to which compliance increased varied depending on the challenge; based on these data, a nominal value of –10% can be proposed as a figure of merit. Patients in whom Z1 decreased more than –10% with challenge can then be classified as reactive in compliance. It is interesting to note that the oldest patients (No. 7, 14 years; No. 8, 6 years; and No. 12, 12 years) exhibited the smallest changes in Z1 with challenge. This fits with our hypothesis that the pulmonary vasculature of these patients exposed to PH over a longer time period would have undergone chronic structural adaptation. When compared with reactivity in PVR, we found a general correlation between a patient’s being reactive in compliance and reactive in PVR. Two exceptions, patients 8 and 9, were found. Patient 8 was reactive in compliance by our classification (although the actual percent change was only –20%) but would not be classified as reactive in PVR (percent change in PVR=–15%). For patient 9, true baseline (ie, room air or reduced oxygen) values could not be obtained and thus, the baseline value reported here is that with a nitric oxide plus oxygen challenge, making it difficult to gauge true reactivity.

We conclude that the use of a single pressure measurement combined with Doppler measurement of flow in the main PA provides a promising and relatively easy-to-implement method to quantify both PV compliance and PVR for reactivity studies. Such additional information may provide a means to further categorize patients for subsequent treatment, including surgical correction of congenital heart defects, as recently described in the INOP study.32 Studies to further evaluate these parameters in a larger clinical population and in animal models of PH are under way.


*    Acknowledgments
 
This project was supported in part by grants from the National Institutes of Health (HL 67393 and HL 72738) and the General Clinical Research Centers Program, National Center for Research Resources, National Institutes of Health (MO1 RR00069).


*    References
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
up arrowResults
up arrowDiscussion
*References
 

  1. Chen EP, Bittner HB, Tull F, Craig D, Davis RD, Van Trigt P. Nitric oxide improves pulmonary vascular impedance, transpulmonary efficiency, and left ventricular filling in chronic pulmonary hypertension. Thorac Cardiovasc Surg J. 1997; 113: 849–857.
  2. Ivy DD. Diagnosis and management of severe pediatric pulmonary hypertension. Cardiol Rev. 2001; 9: 227–237.[CrossRef][Medline] [Order article via Infotrieve]
  3. Barst RJ, Ivy DD, Dingemanse J, Widlitz A, Schmitt K, Doran A, Binaman D, Nguyen N, Gaitonde M, van Giersberg PL. Pharmacokinetics, safety, and efficacy of Bosentan in pediatric patients with pulmonary arterial hypertension. Clin Pharmacol Ther. 2003; 73: 372–382.[CrossRef][Medline] [Order article via Infotrieve]
  4. Ivy DD, Parker D, Doran A, Parker D, Kinsella JP, Abman SH. Acute hemodynamic effects and home therapy using a novel pulsed nasal nitric oxide delivery system in children and young adults with pulmonary hypertension. Am J Cardiol. 2003; 92: 886–890.[CrossRef][Medline] [Order article via Infotrieve]
  5. Turanlahti MI, Laitinen PO, Sarna SJ, Pesonen E. Nitric oxide, oxygen, and prostacyclin in children with pulmonary hypertension. Heart. 1998; 79: 169–174.[Abstract/Free Full Text]
  6. Shandas R, Weinberg C, Ivy D, Nicol E, DeGroff CG, Hertzberg J, Valdes-Cruz L. Development of a noninvasive ultrasound color M-mode means of estimating pulmonary vascular resistance in pediatric pulmonary hypertension: mathematical analysis, in vitro validation, and preliminary clinical studies. Circulation. 2001; 104: 908–913.[Abstract/Free Full Text]
  7. Abbas AE, Fortuin FD, Schiller NB, Appleton CP, Moreno CA, Lester SJ. A simple method for noninvasive estimation of pulmonary vascular resistance. J Am Coll Cardiol. 2003; 41: 1021–1027.[Abstract/Free Full Text]
  8. Hoffman JIE. Diagnosis and treatment of pulmonary vascular disease. Birth Defects: Orig Art Ser. 1972; 8: 9–18.
  9. Milnor WR, Conti R, Lewis KB, O’Rourke MF. Pulmonary arterial pulse wave velocity and impedance in man. Circ Res. 1969; 25: 637–649.[Abstract/Free Full Text]
  10. Grant BJB, Lieber BB. Clinical significance of pulmonary arterial input impedance. Eur Respir J. 1996; 9: 2196–2199.[CrossRef][Medline] [Order article via Infotrieve]
  11. Harned HS Jr. Pediatric Pulmonary Heart Disease. Boston, Mass: Little, Brown & Co; 1990: 55–77.
  12. Nichols WW, O’Rourke MF. McDonald’s Blood Flow in Arteries: Theoretical, Experimental, and Clinical Principles. 4th ed. New York, NY: Arnold; 1998.
  13. Taylor MG. An approach to an analysis of the arterial pulse wave, I: oscillations in an attenuating line. Phys Med Biol. 1957; 1: 258–269.[CrossRef][Medline] [Order article via Infotrieve]
  14. Bergel DH, Milnor WR. Pulmonary vascular impedance in the dog. Circ Res. 1965; 16: 401–415.[Abstract/Free Full Text]
  15. Milnor WR, Bergel DH, Bargainer JD. Hydraulic power associated with pulmonary blood flow and its relation to heart rate. Circ Res. 1966; 19: 467–480.[Abstract/Free Full Text]
  16. Weinberg CE, Hertzberg JR, Shandas R. Utility of intravascular ultrasound to measure local compliance of the pediatric pulmonary artery: in vitro studies. J Am Soc Echocardiogr. 2002; 15: 1507–1514.[CrossRef][Medline] [Order article via Infotrieve]
  17. Association for the Advancement of Medical Instrumentation (AAMI). Cardiovascular Implants—Vascular Prosthesis. standard VP20, section 8.10, June 1994.
  18. Milnor WR. Hemodynamics. Baltimore, Md: Williams & Wilkins; 1982.
  19. Taylor MG. The input impedance of an assembly of randomly branching elastic tubes. Biophys J. 1966; 6: 29–51.[Abstract/Free Full Text]
  20. O’Rourke MF. Vascular impedance in studies of arterial and cardiac function. Physiol Rev. 1982; 62: 570–623.[Free Full Text]
  21. Westerhof N, Sipkema P, van Den Bos GC, Elzinga G. Forward and backward waves in the arterial system. Cardiovasc Res. 1972; 6: 648–656.[Medline] [Order article via Infotrieve]
  22. Figliola RS, Beasley DE. Theory and Design for Mechanical Measurements. New York, NY: John Wiley & Sons, Inc; 1991: 105–177.
  23. Lucas CL, Wilcox BR, Coulter NA. Pulmonary vascular response to atrial septal defect closure in children. J Surg Res. 1975; 18: 571–586.[CrossRef][Medline] [Order article via Infotrieve]
  24. Lucas CL, Radke N, Wilcox BR, Henry GW, Keagy BA. Maturation of pulmonary input impedance spectrum in infants and children with ventricular septal defect. Am J Cardiol. 1986; 57: 821–827.[CrossRef][Medline] [Order article via Infotrieve]
  25. Wilcox BR, Lucas CL. Pulmonary input impedance in children with left-right shunt. J Surg Res. 1980; 29: 40–49.[CrossRef][Medline] [Order article via Infotrieve]
  26. Kussmaul WG, Mieland JM, Laskey WK. Pressure-flow relations in the pulmonary artery during myocardial ischaemia: implications for right ventricular function in coronary disease. Cardiovasc Res. 1988; 22: 627–638.[Medline] [Order article via Infotrieve]
  27. Kussmaul WG, Altschuler JA, Herrmann HC, Laskey WK. Effects of pacing tachycardia and balloon valvuloplasty on pulmonary artery impedance and hydraulic power in mitral stenosis. Circulation. 1992; 86: 1770–1779.[Abstract/Free Full Text]
  28. Sharp MK, Pantalos GM, Minich L, Tani LY, McGough EC, Hawkins JA. Aortic input impedance in infants and children. J Appl Physiol. 2000; 88: 2227–2239.[Abstract/Free Full Text]
  29. Valdes-Cruz LM, Horowitz S, Mesel E, Sahn DJ, Fisher DC, Larson D, Goldberg SJ, Allen HD. A pulsed Doppler echocardiographic method for calculation of pulmonary and system flow: accuracy in a canine model with ventricular septal defect. Circulation. 1983; 68: 597–602.[Free Full Text]
  30. Sahn DJ, Valdes-Cruz LM. Ultrasound Doppler methods for calculating cardiac volume flows, cardiac output, and cardiac shunts, Cardiovasc Clin. 1986; 17: 19–31.[Medline] [Order article via Infotrieve]
  31. Siegel J, DePetro M, Fontanilla L, Sierzenski PR, McGraw P, O’Connor RE. Doppler ultrasound as a non-invasive means to determine stroke volume and cardiac output. Acad Emerg Med. 2003; 10: 427. Abstract.[CrossRef]
  32. Balzer DT, Kort HW, Day RW, Corneli HM, Kovalchin JP, Cannon BC, Kaine SF, Ivy DD, Webber SA, Rothman A, Ross RD, Aggarwal S, Takahashi M, Waldman JD. Inhaled nitric oxide as a preoperative test (INOP test I): the INOP Test Study Group. Circulation. 2002; 106 (suppl I): I-76–I-81.[Medline] [Order article via Infotrieve]



This article has been cited by other articles:


Home page
Eur Heart JHome page
J.-W. Lankhaar, N. Westerhof, T. J.C. Faes, C. Tji-Joong Gan, K. M. Marques, A. Boonstra, F. G. van den Berg, P. E. Postmus, and A. Vonk-Noordegraaf
Pulmonary vascular resistance and compliance stay inversely related during treatment of pulmonary hypertension
Eur. Heart J., July 1, 2008; 29(13): 1688 - 1695.
[Abstract] [Full Text] [PDF]


Home page
Asian Cardiovasc. Thorac. Ann.Home page
M. Atiq, H. Tasneem, and K. Aziz
Estimation of Pulmonary Vascular Resistance with Doppler Diastolic Gradients
Asian Cardiovasc Thorac Ann, June 1, 2008; 16(3): 221 - 225.
[Abstract] [Full Text] [PDF]


Home page
Am. J. Physiol. Heart Circ. Physiol.Home page
V. Muthurangu, D. Atkinson, M. Sermesant, M. E. Miquel, S. Hegde, R. Johnson, R. Andriantsimiavona, A. M. Taylor, E. Baker, R. Tulloh, et al.
Measurement of total pulmonary arterial compliance using invasive pressure monitoring and MR flow quantification during MR-guided cardiac catheterization
Am J Physiol Heart Circ Physiol, September 1, 2005; 289(3): H1301 - H1306.
[Abstract] [Full Text] [PDF]


Home page
HypertensionHome page
L. Mazzaro, S. J. Almasi, R. Shandas, D. R. Seals, and P. E. Gates
Aortic Input Impedance Increases With Age in Healthy Men and Women
Hypertension, June 1, 2005; 45(6): 1101 - 1106.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
110/17/2609    most recent
01.CIR.0000146818.60588.40v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow Request Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Weinberg, C. E.
Right arrow Articles by Shandas, R.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Weinberg, C. E.
Right arrow Articles by Shandas, R.
Related Collections
Right arrow Pulmonary circulation and disease
Right arrow Echocardiography
Right arrow Pediatric and congenital heart disease, including cardiovascular surgery