(Circulation. 2004;110:2498-2505.)
© 2004 American Heart Association, Inc.
Special Report |
From the Minneapolis Heart Institute and Minnesota Cardiovascular Research Institute, Minneapolis, Minn (R.S.S.); Harvard-Massachusetts Institute of Technology Division of Health Sciences and Technology, Cambridge (E.R.R.); Borgess Medical Center, Kalamazoo, Mich (A.C.); Brigham and Womens Hospital, Boston, Mass (E.R.R., C.R.); University of British Columbia, Vancouver, Canada (L.M.); American Cardiovascular Research Institute, Atlanta, Ga (N.A.C., K.A.R.); Washington Hospital Center, Washington, DC (R.W.); Columbia University, New York, NY (J.W.); University of Pennsylvania Health System, Philadelphia (R.L.W.); Armed Forces Institute of Pathology, Bethesda, Md, and US Food and Drug Administration, Rockville, Md (J.L.G., O.D.H., B.D.Z., R.V.).
Correspondence to Robert S. Schwartz, MD, Minnesota Cardiovascular Research Institute, Minneapolis Heart Institute Foundation, 920 E 28th St, Minneapolis, MN 55407. E-mail rss{at}rsschwartz.com
Key Words: stents restenosis drugs
| Introduction |
|---|
|
|
|---|
This document presents an integrated recommendation set for evaluating drug-eluting stents in peripheral vessels with preclinical models. The recommendations encompass study design, experimental performance, and histopathologic evaluations and emphasize the need to evaluate safety and efficacy at multiple points in time. The present document is a consensus of clinical, academic, and commercial groupsall experts in the evaluation of preclinical investigational or interventional devices. Because preclinical peripheral studies are less well understood than are coronary models, the recommendations do not prescribe a single method for device evaluation; they instead provide broad suggestions for evaluation. Peripheral vascular knowledge will increase as correlations of preclinical with clinical data become available. Future versions of this document will necessarily reflect such enhanced knowledge.
Drug-eluting stents are implantable devices that present or release single or multiple bioactive agents into blood vessels after implantation. A formal definition for the drug-eluting stent is as follows: A stent combined with therapeutic agent(s) physically affixed, often by embedding the agent in a carrier coating such as a polymer matrix and/or by directly applying the agent to the stent surface or within the stent. Following implantation, the therapeutic agent(s) is/are released into the blood stream as well as locally into tissue adjacent to the stent.
Drug formulations, matrices, and release are as described in the 2002 coronary artery guidelines.1
One major purpose of drug-eluting stents is to limit neointimal hyperplasia. Before clinicians consider undertaking preclinical studies, they should have strong reasons for considering stents and agents for an application such as this. Preclinical studies should follow cell culture work or work that suggests efficacy in other applications (eg, malignancies). In vitro testing should be a fundamental requisite for characterizing the elution kinetics of the agent from the stent and devices.
Peripheral stent drug dosing and kinetics should be studied as they are with coronary stents. The proposed clinical doses and stent kinetics should be justified by in vivo preclinical data. Multiple doses should be examined in peripheral stents to determine a safe range and, if possible, to determine where toxicity occurs. Ideally, a toxic dose should be 3 to 10 times higher than the anticipated clinical dose giving a comparable safety margin.
Systemic and "downstream" drug effects and toxicity that occur with higher dosages in peripheral stents should be carefully documented. Intravenous injection studies should be performed to estimate systemic toxicity. Long-term stented animal studies should be performed with monitoring plasma drug levels. Assessment should be per the previous guideline document, and at a minimum should include nephrotoxicity, hepatotoxicity, neurotoxicity, and possible immune compromise. Perfusion magnetic resonance imaging (MRI) and positron emission tomography (PET) also may be useful in evaluating the central nervous system for carotid stenting, in addition to Doppler ultrasound of extracranial vessels.
The characterization of in vitro and in vivo drug release should follow that described for drug-eluting stents for coronary application (Table 1). In vivo drug release can be characterized according to the guidelines for coronary stents. As with these guidelines,
6 stents per time point should be used. Blood levels should be used to estimate a "systemic pK" for the drug.
|
Additional measurements for estimating in vivo pharmacokinetics can be obtained. These include adding time points to better characterize drug release into arterial tissue, better defining a reasonable safety margin dose, and estimating drug levels in arterial and other tissues proximal and distal to the stent. These measurements also are summarized in the coronary artery guidelines. As with those guidelines, in the case of bioerodable coatings, information about disappearance should be presented.
Peripheral stents appear to have an increased risk for fracture in proportion to increase in their diameter. Fracture appears to occur in both drug-eluting and bare metal stents.2 A polymer-based drug-eluting stent with fracture might theoretically enhance local fluid, corrosion, and possibly stent fracture. A systematic investigation of stent breakage should be included in research done on peripheral drug-eluting stents.
Stent drug-release characterization both in vitro and in vivo should be studied before undertaking human clinical testing and should be understood before performing animal testing. The reproducibility of results should be established to ensure the interpretability of subsequent clinical and animal evaluations. Drug elution and its characterization are inexact and technically difficult, but careful attempts with documentation should be made regardless. The important drug-eluting stent applications are for in-stent restenosis, and experience suggests that testing should determine the healing response for single and overlapped stents.
| Animal Models |
|---|
|
|
|---|
The ideal animal model for the evaluation of peripheral drug-eluting stents is uncertain and, at present, less well understood than it is for coronary models. Drug deposition and pharmacological responses vary with arterial site and lesion morphology. Devices used for normal peripheral vessels in animals may be used as models for the design of devices for human peripheral clinical application. Histopathology suggests that the peripheral arteries of domestic crossbred swine and the iliac arteries of rabbits may be best suited because their size and access are similar to those of human vessels. Both exhibit substantial elastin and show comparability of the intima, media, and adventitia. The injury response with neointimal formation of peripheral porcine or rabbit vessels is not presently well understood.
Device selection and testing should take place in vessels with comparable dimensions, and stents should be appropriately sized to the artery in preclinical studies. Too much mechanical stent injury confounds safety and efficacy results. To obtain a comparable model in larger-vessel studies, it may be necessary to ligate vascular branches. Safety and efficacy should be examined in a comparative study with several time points. All tissue surrounding the stent, including muscle, nerves, and vessels, should be examined for toxic effects, especially if the drug used is known to be toxic to certain types of tissues and cells.
As with the coronary arteries, the most important safety concern in peripheral drug-eluting stents is acute and subacute stent thrombosis. All animals that experience adverse consequences, limb ischemia, death, or other untoward clinical events should be examined and their stent status carefully documented. Because standard stent practice in patients entails oral aspirin plus clopidogrel or ticlopidine, these agents should be administered throughout the preclinical study.
The underlying tissues (eg, muscular or elastic arteries, veins) in superficial femoral artery and vascular access stents may respond differently to drug-eluting stents. One limit in model planning is that peripheral porcine arteries appear not to develop neointima as vigorously as do coronary arteries (Schwartz, Edelman, Carter, and Chronos, unpublished observations, 19902000). Accordingly, when possible, stent testing in the vascular bed or in the same type of artery or vein where stents will be used clinically is preferable. Similarly, the use of maximum-length stents at clinical doses/unit area, total doses, and release rates intended for clinical use is indicated. Drug toxicity, thrombus, and end-organ infarct are of particular concern when stenting renal and carotid arteries.
Porcine Peripheral Artery Model
The preferred porcine model is the normocholesterolemic domestic crossbred, or mini-swine artery. Stents may be placed in the carotid, renal, or iliofemoral systems. Stents should be appropriately sized for the artery (stent:artery ratio of between 1.0 and 1.1) and implanted into arteries with no previous injury.
Rabbit Iliac Artery and Other Models
The Expert Consensus Group considers the rabbit iliac artery to be less acceptable as a model for peripheral studies because the vessels are smaller and the reaction to injury is less well defined. The canine model has limited inflammatory responses and may be less sensitive as a safety model. Other species that may provide longer and larger vessels in the neck and legs include the goat and sheep. Calf arteries grow quickly and larger vessel diameter changes occur over short time periods (A.P. Moody, unpublished data, 1991), which may confound researchers conducting studies of
6 months duration. Unfortunately, experience with and knowledge of stenting in these alternative species are limited or unknown at this time.
Drug-Eluting and Control Stents
Only 1 stent should be implanted per artery except when questions of stent overlap or multiple-stent dosing are considered. Stents may be placed in multiple different arteries in the same animal, and an adequately controlled study should include bare, carrier-only, and carrier-plus drug-eluting stents. The stents should be appropriately sized (stent:artery ratio of
1:1). When possible, stents should be tested in the arterial bed corresponding to that in which the stents will be clinically deployed. This consideration includes the need in preclinical studies to implant the stents in arteries of similar type as much as possible, whether it is muscular, elastic, or conduit.
Preclinical trials must be adequately controlled. Polymeric materials for drug elution frequently affect the arterial repair process, generally in a toxic manner. When a polymer or carrier of any sort is present, it is imperative to include as controls stents that are coated with polymer alone, especially because polymer coatings that do not contain drug may react differently from coatings that are devoid of drug after complete release. This may reflect different surface characteristics (eg, porosity or texture), especially when matrix-type devices are used. In addition, changes in the drug-coating procedure should merit additional studies of the polymer alone because the coating procedure may substantially change the biological activity of the device. When drug is bound directly to a stent, the stent without drug can be a satisfactory control, although any changes to the surface of the metal should be applied to the control devices. When drugs are attached directly to metal stents without a polymer, the stent metal is activated by creating a metal oxide layer. In this case, the control devices for testing should be subject to the identical steps used in the manufacture of drug-eluting stents before attachment of the active drug. This should include identical cleaning, packaging, and sterilization procedures.
| Testing Boundaries |
|---|
|
|
|---|
Overlapping Stents
Stent overlap is sometimes needed during clinical implantation. Because overlapping drug-eluting stents may cause an additive or combined effect from the 2 stents, intentional overlap should be evaluated in preclinical studies. The overlap distance should be roughly one third the length of a stent, or
4 mm, and the number of overlapping stent-implant pairs evaluated should be no less than 5. Histopathologic processing should be done with sections taken from the reference segments, the single (nonoverlapping) stented region, and the overlapping stented region.
Sampling Time Points and Size
Stent efficacy or relative vascular patency at the stent site should be evaluated. Neointima should not be obstructive. Thrombus must be absent and data should be obtained early (3 or 7 days) to determine subacute thrombosis risk and early histopathology. Long-term neointimal thickening and other arterial effects should be monitored at 28 days and at least 1 late time point. The late monitoring point (3 or 6 months) depends on when healing and drug release both are complete for the model chosen (for rabbits, see references 3 through 5; data for other species are pending). Note that the US Food and Drug Administration typically recommends monitoring for
6 months. Three-month follow-up data are in general acceptable for initiating clinical feasibility trials if no adverse findings are noted at the time. Six-month data should be available at the time of Investigational Device Exemption submission for a pivotal study. These later time points are especially important given the impact of peri-stent late remodeling as an additional cause of the peri-stent effects that would affect the clinical outcome.
Long-term time points for animal studies (
6 months) are important, but results may be less rigorously interpreted until global understanding is obtained about the relationship between animals and humans at longer implantation times. Such studies on coronary and peripheral vessels by several investigative groups are pending or under way. The choice of a long-term end point becomes more complicated when dealing with drug-eluting stents. Whereas injury after balloon angioplasty or stent implantation may peak and then resolve over weeks to months, the presence of a drug might change that dynamic.
Arterial reaction to the device and general animal health should be reviewed for a practical multiple of the drug residence time within the primary target tissue. Elution observation thus corresponds to 4 elution half-lives in the tissue, after the drug is no longer measurable from the release devices. Because the tissue will be nearly devoid of drug within 4 half-lives, waiting for this period of time is thought to be appropriate. Another important time point is drug metabolic half-life, a parameter that is less easily measured.
The number of stents for study should be determined by a power calculation for the predetermined expected difference in key parameters. Sample size power calculations are not yet well defined; therefore, estimation is appropriate. Typically, 7 to 10 drug-eluting stents per time point are satisfactory in most peripheral models, but this amount depends on the variability seen in the data; however, this number should be sufficient to draw conclusions. One method for analyzing coronary artery results uses a comparison of regression models. This model should be applicable to peripheral vessel studies.6
| Implantation Procedure |
|---|
|
|
|---|
| Stent-Related Antiplatelet Medication |
|---|
|
|
|---|
| Autopsy/Necropsy Evaluation |
|---|
|
|
|---|
Necropsy should be performed by qualified personnel to determine the cause of death for all animals dying after entry into the study, regardless of whether or not the study animal completed the allotted survival time. An expert opinion should be rendered as to the cause of death when it does not result from euthanasia. The performance of all stents in such early or unexpected deaths should be determined, recorded, and reported. The target organ or tissues of the vessel bed downstream of the stent implants should be examined for any evidence of infarct (and distribution), as well as for any fibrosis, especially in the perivascular region.
Histopathology should be performed on all implanted stent-artery segments, including segments in those animals dying any time after the implantation procedure has begun. All implanted stents should be sectioned, regardless of how long they were implanted. The thoracic cavity (pigs) or abdominal cavity/retroperitoneum (rabbits) should be examined for effusion, inflammation, infection, perforation, or other systemic problems.
| Tissue Processing and Fixation |
|---|
|
|
|---|
100 mm Hg. Immersion fixation should be performed at a volume that is sufficiently large to allow complete and rapid fixative percolation through the tissue and without alteration of the tissue shape. After removal, the target organs, tissues, or limbs containing the stent should be sectioned transaxially (short axis sections) at minimum intervals of 1 cm. These sections should be examined grossly for target organ or muscle infarction. All such infarctions should be included in the final report. | Histopathologic Stains, Histopathology, and Histomorphometry |
|---|
|
|
|---|
| Clinical and Blood Parameter Evaluations |
|---|
|
|
|---|
Blood parameters should be tested for findings that are suggestive of allergy or hepatic or renal dysfunction. These measures should be taken at baseline and at euthanasia and include a complete blood count with differential blood count, liver enzymes (alanine aminotransferase, aspartate aminotransferase), and creatinine. Specific chemical tests may need to be performed or monitoring laboratory parameters assayed if idiosyncratic effects on tissues and cells are suspected or if clinical symptoms develop.
| Arteriography and Intravascular Ultrasound |
|---|
|
|
|---|
| Stent Evaluation |
|---|
|
|
|---|
Semiquantitative Histopathology
Inflammation and Injury
Inflammation by histopathologic evaluation might include an injury score (value 0 to 3) at each stent strut site, an inflammation description (absent, or cell types and location), and an inflammation score (value 0 to 3) for the overall vessel as well as the adventitia media, neointima, and at stent strut sites (see, eg, reference 10). When possible, cell density in specific tissue compartments should be recorded as number of cells per area.
Angiogenesis and Other Histopathology
Angiogenesis can be assessed (scored 0 to 3) and reported for the adventitia, media, and neointima.11 Other histopathologic features should be sought in the adventitia, media, and neointima and assigned a value of 0 through 3 or be described quantitatively as the number and size of vessels per unit area. These observations include fibrin or fibrinoid deposits, hemorrhage, and necrosis.
Observational Histopathologic Data
Endothelialization, Reendothelialization and Vessel Healing
A stented vessel should be considered healed when it shows endothelialization or a healthy-appearing layer of near-complete periluminal cells. Endothelialization should be recorded as absent, partial, or complete in all sections. A semiquantitative analysis can be performed and presented as the percentage of circumference covered by endothelium. The time to reendothelialization should be estimated. Scanning electron microscopy from
3 stents is recommended to assess endothelial recovery. The Expert Consensus Group suggests that a completely healed vessel exhibits reendothelialization and no evidence of fibrin, fibrinoid deposits, excessive inflammation, or hemorrhage.
Stent Strut Position, Apposition/Malapposition, and Adjacent Tissue
Other observations should include the extent to which stent struts are apposed to the vessel wall (percentage of wires in contact) and covered by tissue or endothelium (as a percentage). A subjective description also should be created for adjacent tissue, including medial necrosis or thinning, loss of cellularity, and hyalinization. Stents should be observed carefully to determine whether strut malapposition exists.
Organ Evaluation and Histopathology
Whether renal, cerebral, visceral, or muscle, tissue histology directly beneath, distal to, and supplied by the stented artery should be observed and recorded as normal or abnormal and same or different from control stents. Histopathology is a standard. Because of its sampling nature, however, it may not be as sensitive as is necessary to detect stent-induced organ damage, particularly in the extremities and in solid organs. Vasoconstriction or embolic phenomena induced by the stent may be a concern with certain drug-eluting technologies, and it is therefore recommended that an evaluation method capable of 3D analysis be used in a fraction of devices. This method might take the form of computed tomography or MRI scanning of the intact organ before histopathology, and if abnormalities are found, the histopathology may be guided by such imaging methods. Serial complete gross sectioning at 5-mm increments with a gross dissection blade also can be used to locate organ pathology. Specific additional but optional end-organ evaluations follow.
Extremities
The postmortem evaluation of limb muscles may be performed with radioactive microspheres. Perfusion assessment would include measured radioactive counts of homogenized tissue or MRI methods.12,13 In vivo muscle studies could be performed with standard radioisotope-based perfusion measures. Histopathology of distal muscle should be evaluated for abnormal findings of fiber size variation, nuclear abnormalities, fiber angulation, pyknosis of nuclei, and fiber atrophy.
Kidneys
Radionuclide renal studies in vivo may be performed with labeled microspheres or radioisotopes such as 99mTc dimercaptosuccinic acid, used for comparative imaging of stented versus nonstented kidneys to assess parenchymal abnormalities. Renal histopathology should be examined for evidence of glomerulonephritis, sclerosis, hemorrhage, fibrosis, and inflammation. Attention should be paid to glomerular hyper- and hypocellularity and mesangial appearance, as well as to the search for interstitial fibrosis.
Brain
The determination of embolic effects on the brain also may be performed via perfusion scanning. PET or MRI, if available, can be used to measure brain perfusion in vivo, and radionuclide imaging also can be used with single photon emission computed tomography.14,15 Histopathology of the brain should be examined for plaques, evidence of amyloid, neurofibrillary tangles and neuropil threads, and angiopathy including hemorrhage.
Quantitative Histomorphometry
Morphometry of histopathologic sections is essential for stent evaluation. Measurement systems should be calibrated and documented against a traceable standard before each measurement session. Measurements of all sections should include medial area, area within the internal elastic lamina, area within the external elastic lamina, lumen area, and stent area (area within the stent itself).
Neointimal measurement tests efficacy and should include thickness at each stent strut site and in between the struts, the mean of these values, and total neointimal area. The average neointimal thickness (average for all strut sites) should be calculated for each section. If the stent struts separate from the area within the internal elastic lamina, then each site should be measured and reported for distance of the strut separation.
Calculations Derived from Quantitative Histomorphometry
The following calculations from histopathologic information should be derived for each stent.
| Statistical Analysis of Safety and Efficacy |
|---|
|
|
|---|
|
|
| Report Summary and Conclusions |
|---|
|
|
|---|
Study conclusions based on a critical analysis of blinded data from an adequate number of animals studied are crucial for understanding device safety and efficacy. Conclusions should simply restate the data, but they should be an ordered interpretive list of stent safety, toxicity (including proposed toxicity margins), and efficacy. The appropriate use of representative graphics (charts, tables) should be included to support and simplify the conclusions.
Conclusions should be summarized and then synthesized. Conclusions should be drawn indicating whether the devices performed better than control (bare stent) and polymer or carrier-only devices (ie, relatively suppressed neointima formation and no excess neointima by histopathology). If such a statement is not possible because of mixed results, then the efficacy statement should be explicitly enumerated separately for all of the above parameters. Ideally, the drug-eluting device should perform better than either controls or the polymer or carrier-only devices. At a minimum, the performance of the drug-eluting devices should be no worse than the controls and polymer or carrier-only devices. If no efficacy difference is found in the preclinical morphometric studies between drug eluting versus bare and polymer-coated controls, then proceeding to human clinical trials may be considered from a risk:benefit perspective. This decision should be based on whether the device appears to be safe in preclinical studies after evaluating drug biological effects such as delayed healing. Finally, the report should reference whether the stented vessels healed completely, partially, or not at all.
| Consensus Opinion: Satisfactory Findings and Outcomes |
|---|
|
|
|---|
Stent Thrombosis
Experience with stented porcine coronary arteries suggests that sudden death may be more common in drug-eluting devices, principally because of platelet-rich coronary stent thrombosis.16,17 This problem is uncommon in peripheral stents, and thus local or embolic thrombotic events are not likely to be seen in peripheral stenting studies. Nevertheless, stent thrombosis should be considered in all cases of poor outcome. The early mortality rate for pigs should be
15%. Therefore, unexpected death occurring later than 24 hours after implantation should be vigorously investigated for cause. Overall mortality including early and late deaths should be <25%, which is a high number. A study with such high mortality suggests that a problem exists with either the devices or the implantation methods and techniques. Any percentage of deaths higher than this number should be a warning of a substantial problem somewhere in the study.
Inflammation, Fibrin Deposits, and Fibrinoid Formation
Polymer coatings by their nature may induce fibrin deposits, inflammatory responses, and fibrinoid formation.18 Such responses may be acceptable if the reaction is minimal or mild, and the responses do not accelerate, extend, or cause substantial vascular injury or stenosis. It is crucial, however, that investigators demonstrate that later time points have inflammatory reactions that meet the same safety criteria as do early time points.
Neointima and Arterial Injury
Stent neointima should be thin, with adequate lumen at all time points. Deviations from these occurrences should be quantified and an assessment made by the pathologist who will determine whether such injury resulted from the drug- or polymer-associated inflammation or the mechanical injury of the stent itself. If conclusions of the study are made without including such severely injured sections, then the number and locations of such excluded sections should be stated.
| Overall Conclusions |
|---|
|
|
|---|
| Acknowledgments |
|---|
Dr Schwartz receives research funding from Guidant, Inc, and Boston Scientific. Dr Edelman is a speaker and does occasional research for Biocompatibles, Guidant, Medtronic, and Orbus. Dr Carter receives research support from Cordis Inc, Boston Scientific, Medtronic, Guidant, and Orantel. Dr Rogers is a speaker for Cordis, BSC, Guidant, Medtronic, Cor Pharmaceuticals, and Millennium. Dr Robinson receives research support from Guidant, Medtronic AVE, Boston Scientific, and Abbott Laboratories. Dr Waksman is a consultant to Guidant and Novoste and receives royalties from his work for Emory University/Novoste. Dr Machan receives research support from Angiotech Pharmaceuticals, Cook Inc, and Boston Scientific and is a consultant to Angiotech. Dr Wilensky holds equity interest in Johnson & Johnson.
| Footnotes |
|---|
| References |
|---|
|
|
|---|
2. Duda SH, Pusich B, Richter G, et al. Sirolimus-eluting stents for the treatment of obstructive superficial femoral artery disease: six-month results. Circulation. 2002; 106: 15051509.
6. Huckle WR, Drag MD, Acker WR, et al. Effects of subtype-selective and balanced angiotensin II receptor antagonists in a porcine coronary artery model of vascular restenosis. Circulation. 1996; 93: 10091019.
7. Virmani R, Guagliumi G, Farb A, et al. Localized hypersensitivity and late coronary thrombosis secondary to a sirolimus-eluting stent: should we be cautious? Circulation. 2004; 109: 701705.
8. Farb A, Burke AP, Kolodgie FD, et al. Pathological mechanisms of fatal late coronary stent thrombosis in humans. Circulation. 2003; 108: 17011706.
9. Virmani R, Kolodgie FD, Farb A, et al. Drug eluting stents: are human and animal studies comparable? Heart. 2003; 89: 133138.
11. Pelisek J, Fuchs A, Engelmann MG, et al. Vascular endothelial growth factor response in porcine coronary and peripheral arteries using nonsurgical occlusion model, local delivery, and liposome-mediated gene transfer. Endothelium. 2003; 10: 247255.[Medline] [Order article via Infotrieve]
12. Meller J, Sahlmann CO, Becker W. Nuclear medicine studies in the dialysis patient. Semin Dial. 2002; 15: 269276.[CrossRef][Medline] [Order article via Infotrieve]
13. Bennett HF, Li D. MR imaging of renal function. Magn Reson Imaging Clin N Am. 1997; 5: 107126.[Medline] [Order article via Infotrieve]
14. Hyun Y, Lee JS, Rha JH, et al. Different uptake of 99mTc-ECD and 99mTc-HMPAO in the same brains: analysis by statistical parametric mapping. Eur J Nucl Med. 2001; 28: 191197.[CrossRef][Medline] [Order article via Infotrieve]
15. Ding HJ, Huang YF, Tzeng CC, et al. Synthesis of Tc-D,D-HMPAO and Tc-L,L-HMPAO and their comparison of chemical and biological properties. Bioorg Med Chem Lett. 1999; 9: 31993202.[CrossRef][Medline] [Order article via Infotrieve]
16. Ahn YK, Jeong MH, Kim JW, et al. Preventive effects of the heparin-coated stent on restenosis in the porcine model. Cathet Cardiovasc Interv. 1999; 48: 324330.[CrossRef][Medline] [Order article via Infotrieve]
17. Jeong MH, Owen WG, Staab ME, et al. Porcine model of stent thrombosis: platelets are the primary component of acute stent closure. Cathet Cardiovasc Diagn. 1996; 38: 3843.[CrossRef][Medline] [Order article via Infotrieve]
18. van der Giessen W, Lincoff A, Schwartz, R, et al. Marked inflammatory sequelae to implantation of biodegradable and nonbiodegradable polymers in porcine coronary arteries. Circulation. 1996; 94: 16901697.
19. Janicki C, Hwang CW, Edelman ER. Dose model for stent-based delivery of a radioactive compound for the treatment of restenosis in coronary arteries. Med Phys. 2003; 30: 26222628.[CrossRef][Medline] [Order article via Infotrieve]
20. Welt FG, Tso C, Edelman ER, et al. Leukocyte recruitment and expression of chemokines following different forms of vascular injury. Vasc Med. 2003; 8: 17.
21. Hwang CW, Wu D, Edelman ER. Impact of transport and drug properties on the local pharmacology of drug-eluting stents. Int J Cardiovasc Intervent. 2003; 5: 712.[CrossRef][Medline] [Order article via Infotrieve]
22. Hwang CW, Edelman ER. Arterial ultrastructure influences transport of locally delivered drugs. Circ Res. 2002; 90: 826832.
23. Lowe HC, Schwartz, RS, Mac Neill BD, et al. The porcine coronary model of in-stent restenosis: current status in the era of drug-eluting stents. Catheter Cardiovasc Interv. 2003; 60: 515523.[CrossRef][Medline] [Order article via Infotrieve]
24. Rogers JH, Chang D, Lasala JM. Percutaneous repair of coronary artery bypass graft-related pseudoaneurysms using covered JOSTENTs. J Invasive Cardiol. 2003; 15: 533535.[Medline] [Order article via Infotrieve]
25. Lepor NE, Rogers CD. Coronary artery disease treatment in the catheterization laboratory: the promise of an evolving new pharmaco-mechanical approach. Introduction. Rev Cardiovasc Med. 2002; 3: S1S3.[CrossRef][Medline] [Order article via Infotrieve]
26. Machan L, Burt HM, Hunter WL. Local delivery of chemotherapy: a supplement to existing cancer treatments. A case for surgical pastes and coated stents. Adv Drug Deliv Rev. 1997; 26: 199207.[CrossRef][Medline] [Order article via Infotrieve]
27. Guagliumi G, Virmani R, Musumeci G, et al. Drug-eluting versus bare metal coronary stents: long-term human pathology. Findings from different coronary arteries in the same patient. Ital Heart J. 2003; 4: 713720.[Medline] [Order article via Infotrieve]
28. Cheneau, E, John MC, Fournadjiev J, et al. Time course of stent endothelialization after intravascular radiation therapy in rabbit iliac arteries. Circulation. 2003; 107: 21532158.
29. Guagliumi G, Farb A, Musumeci G, et al. Images in cardiovascular medicine. Sirolimus-eluting stent implanted in human coronary artery for 16 months: pathological findings. Circulation. 2003; 107: 13401341.
30. Finn AV, Gold HK, Tang A, et al. A novel rat model of carotid artery stenting for the understanding of restenosis in metabolic diseases. J Vasc Res. 2002; 39: 414425.[CrossRef][Medline] [Order article via Infotrieve]
31. Hiatt BL, Carter AJ, Yeung AC. The drug-eluting stent: is it the Holy Grail? Rev Cardiovasc Med. 2001; 2: 190196.[Medline] [Order article via Infotrieve]
32. Carter AJ. Drug-eluting stents for the prevention of restenosis: standing the test of time. Catheter Cardiovasc Interv. 2002; 57: 6971.[CrossRef][Medline] [Order article via Infotrieve]
33. Serruys PW, Regar E, Carter AJ. Rapamycin eluting stent: the onset of a new era in interventional cardiology. Heart. 2002; 87: 305307.
34. Suzuki T, Kopia G, Hayashi S, et al. Stent-based delivery of sirolimus reduces neointimal formation in a porcine coronary model. Circulation. 2001; 104: 11881193.
35. Farb A, Heller PF, Shroff S, et al. Pathological analysis of local delivery of paclitaxel via a polymer-coated stent. Circulation. 2001; 104: 473479.
36. Virmani R, Farb A, Carter AJ, et al. Pathology of radiation-induced coronary artery disease in human and pig. Cardiovasc Radiat Med. 1999; 1: 98101.[CrossRef][Medline] [Order article via Infotrieve]
37. Klugherz BD, Llanos G, Lieuallen W, et al. Twenty-eight-day efficacy and phamacokinetics of the sirolimus-eluting stent. Coron Artery Dis. 2002; 13: 183188.[CrossRef][Medline] [Order article via Infotrieve]
38. Klugherz BD, Song C, DeFelice S, et al. Gene delivery to pig coronary arteries from stents carrying antibody-tethered adenovirus. Hum Gene Ther. 2002; 13: 443454.[CrossRef][Medline] [Order article via Infotrieve]
39. Nasser TK, Mohler ER III, Wilensky RL, et al. Peripheral vascular complications following coronary interventional procedures. Clin Cardiol. 1995; 18: 609614.[Medline] [Order article via Infotrieve]
This article has been cited by other articles:
![]() |
Y. Suzuki, A. C. Yeung, and F. Ikeno The Pre-Clinical Animal Model in the Translational Research of Interventional Cardiology J. Am. Coll. Cardiol. Intv., May 1, 2009; 2(5): 373 - 383. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. M M Pires, D. Eefting, M. R de Vries, P. H A Quax, and J W. Jukema Sirolimus and paclitaxel provoke different vascular pathological responses after local delivery in a murine model for restenosis on underlying atherosclerotic arteries Heart, August 1, 2007; 93(8): 922 - 927. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Camenzind, P. G. Steg, and W. Wijns A Cause for Concern Circulation, March 20, 2007; 115(11): 1440 - 1455. [Full Text] [PDF] |
||||
![]() |
E. Camenzind Treatment of In-Stent Restenosis -- Back to the Future? N. Engl. J. Med., November 16, 2006; 355(20): 2149 - 2151. [Full Text] [PDF] |
||||
![]() |
U. Speck, B. Scheller, C. Abramjuk, C. Breitwieser, J. Dobberstein, M. Boehm, and B. Hamm Neointima Inhibition: Comparison of Effectiveness of Non-Stent-based Local Drug Delivery and a Drug-eluting Stent in Porcine Coronary Arteries. Radiology, August 1, 2006; 240(2): 411 - 418. [Abstract] [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|