(Circulation. 2004;110:2389-2394.)
© 2004 American Heart Association, Inc.
Heart Failure |
From the Department of Medicine (P.B.A.), Cardiovascular Disease and Physiology, University of Oklahoma Health Sciences Center, Oklahoma City, Okla; the Carlyle Fraser Heart Center/Division of Cardiology (A.L.S.), Emory University School of Medicine, Atlanta, Ga; Ohio State University Heart Center (W.T.A.), Columbus, Ohio; and Medtronic, Inc (K.J.K., R.W.S., A.S., M.M.R.), Minneapolis, Minn.
Correspondence to Philip B. Adamson, MD, University of Oklahoma Health Sciences Center, 920 SL Young Blvd, WP3120, Oklahoma City, OK 73104. E-mail Philip-adamson{at}ouhsc.edu
Received June 7, 2004; accepted June 30, 2004.
| Abstract |
|---|
|
|
|---|
Methods and Results A total of 397 patients with New York Heart Association class III or IV heart failure were studied. Of these, 370 patients had information from their implanted cardiac resynchronization device for mortality risk stratification, and 288 patients had information for measured parameters (ie, HRV, night heart rate, and patient activity) and clinical event analyses. Continuous HRV was measured as the standard deviation of 5-minute median atrial-atrial intervals (SDAAM) sensed by the device. SDAAM <50 ms when averaged over 4 weeks was associated with increased mortality risk (hazard ratio 3.20, P=0.02) and SDAAM were persistently lower over the entire follow-up period in patients who required hospitalization or died. SDAAM decreased a median of 16 days before hospitalization and returned to baseline after treatment. Automated detection of decreases in SDAAM was 70% sensitive in detecting cardiovascular hospitalization, with 2.4 false-positives per patient-year of follow-up.
Conclusions This study demonstrates that SDAAM continuously measured from an implanted cardiac resynchronization device is lower in patients at high mortality and hospitalization risk. SDAAM declines as patient status decompensates. Continuous long-term SDAAM may be a useful tool in the clinical management of patients with chronic heart failure.
Key Words: heart failure heart rate mortality trials nervous system, autonomic
| Introduction |
|---|
|
|
|---|
Heart rate variability (HRV) is a physiological marker of cardiac autonomic control that provides prognostic information about mortality risk.610 Recently, HRV was measured with an implanted device, demonstrating that long-term HRV measurement was possible and that cardiac resynchronization therapy resulted in a shift of autonomic control away from sympathetic dominance.4 However, the prognostic value of long-term HRV measurements derived from an implanted device remains unclear, especially when one considers the role of HRV in regard to patient management. The present study examined a long-term measure of HRV, the standard deviation of a 5-minute median atrial-atrial sensed interval (SDAAM), with correlation to mortality and hospitalization risk.
| Methods |
|---|
|
|
|---|
130 ms, left ventricular ejection fraction
35%, and left ventricular end-diastolic dimension
55 mm. Inclusion and exclusion criteria for the present study were identical to those of the Multicenter InSync Randomized Clinical Evaluation (MIRACLE) trial.11 Patients provided informed consent to be involved in this trial, and each sites Institutional Review Board reviewed and approved the study. Study enrollment began on November 30, 2000, and the final patient was enrolled June 4, 2002. There were 397 successful implantations of the InSync III system. Patients with save-to-disk data available for specified time periods were included in the study. Save-to-disk data were available if at least 4 daily values were available for applicable weeks.
HRV, Night Heart Rate, and Patient Activity Parameters
HRV was measured prospectively from the atrial sensed cycle length acquired continuously from the implanted cardiac resynchronization device. All patients in the present study received cardiac resynchronization therapy continuously throughout the follow-up period. Individuals responsible for heart rate parameter measurements were not aware of patients clinical status. In brief, atrial sensed cycle lengths were measured, and 5-minute A-A interval median values were computed. HRV was then defined as the SDAAM values for each entire 24-hour period. The daily SDAAM was stored and included as a component of a rolling trend analysis, which continuously accumulated data for 6 months. If the patient experienced atrial high-rate episodes (including atrial fibrillation) or had atrial pacing for more than 80% of the 24-hour period, SDAAM from that day was excluded from the analysis. The first 4 weeks after implantation were excluded to allow patients to return to normal lifestyle after the surgery. Data from the SDAAM analysis were uplinked by a programmer at a clinic visit.
The potential clinical value of the SDAAM was evaluated in 3 ways. First, the average SDAAM from weeks 5 to 8 after implantation was used to test the hypothesis that longer-term, device-based SDAAM values are associated with risk for mortality or hospitalizations. For mortality risk stratification, patients were stratified on the basis of their 4-week SDAAM values into 3 groups: (1) low SDAAM (<50 ms), (2) intermediate SDAAM (between 50 and 100 ms), and (3) high SDAAM (>100 ms). These values were chosen on the basis of previous studies6,7 with short-term HRV to determine whether long-term HRV performed in a similar fashion as short-term HRV. Deaths were recorded, and Kaplan-Meier survival curves were constructed.
Second, to determine whether SDAAM was different between groups at high and low risk for subsequent clinical decompensation, SDAAM was compared among patients who were classified into 3 groups on the basis of their 18-month outcomes: (1) no clinical decompensation, (2) minor exacerbations not requiring hospitalization, and (3) cardiovascular hospitalization or death. Events were recorded prospectively. An event was classified as a minor exacerbation if the patient experienced shortness of breath, dyspnea, peripheral edema, dehydration, inadequate cardiac output, or syncope and was treated with an adjustment of diuretic doses or other clinical measures commonly used to treat heart failure acute decompensation.
The third evaluation tested the hypothesis that continuously acquired SDAAM changes as heart failure patients decline clinically. To test this hypothesis, an algorithm was developed to detect decreases in the trend of SDAAM daily values over time. For this analysis, patients were classified into 2 groups: (1) hospitalized for cardiovascular reasons or (2) no clinical decompensation/not hospitalized in the 18-month follow-up period. Patients were excluded from the latter group if they had frequent changes in ß-blocker therapy during the 12 months after device implantation. Patients were included in the former group if they had at least 100 days after implantation before hospitalization and if they had save-to-disk data available for the 12 weeks before hospitalization.
For comparison purposes, the association between hospitalization events and changes in the night heart rate and patient activity trends was also determined. Night heart rate was defined as the average ventricular rate between midnight and 4 AM, excluding atrial high-rate episodes. The activity sensor in the implanted device accumulated the number of activity counts sensed in each minute. A minute was considered "active" if the counts exceeded a threshold that corresponded to a walk rate of
70 steps per minute.
SDAAM, Night Heart Rate, and Activity Change Detection Algorithm
To determine the clinical value of device-based SDAAM, night heart rate, and patient activity, a detection algorithm was developed with a randomly selected subset of the patients with and without hospitalizations in the 18-month follow-up period (development database) and then validated in the remaining patients (validation database). The algorithm was applied retrospectively to SDAAM, night heart rate, and activity data but was designed for daily prospective application as the data were acquired. The algorithm consisted of 3 processing steps that were computed daily, summarized in Figure 1. First, an adaptive reference value was updated daily to track recently measured values and approximate the patients expected value (Figure 1, dashed line, upper panel). Second, the accumulated difference between the reference value and the measured value was computed (Figure 1, bottom panel, dotted line). The accumulated difference decreased over time whenever the measured value was less than the reference value. Finally, a threshold for the accumulated difference was established (Figure 1, bottom panel, dashed line), and a change in the parameter was defined when the threshold was exceeded.
|
The detection algorithm was evaluated by calculating sensitivity (ie, algorithm detection in advance of hospitalization events) and false-positive rates (ie, algorithm detection in the absence of hospitalization events) in the validation database. Detection was considered a true positive when the accumulated difference exceeded the threshold value before hospitalization and remained in excess of the threshold for each day before hospitalization. The exemplary threshold used in this analysis for SDAAM was 200 ms days. This threshold was selected to optimize the predictive value of the algorithm determined as the best trade-off between sensitivity and false-alarm rate. If the accumulated difference did not exceed the threshold on the day of hospitalization it was considered a false-negative detection. False-positive detections were considered as any time the accumulated difference exceeded the threshold in a patient who did not have a hospitalization. If false-positive detections were continuous over days, the initial detection was considered the false-positive event. A new false-positive event was considered only when the accumulated difference returned to baseline and subsequently crossed the threshold again. The resulting number of false-positive detections was divided by the total duration of the data set and was expressed as the number of false-positives per patient-year of follow-up. Modified receiver operating characteristic (ROC) curves were constructed for SDAAM, night heart rate, and patient activity from the sensitivity and false-positive rate results.
Statistical Analysis
Differences in baseline demographics among the groups were analyzed with 1-way ANOVA for continuous variables and the Pearson
2 test for discrete variables. The Cox proportional hazards model was used to estimate the hazard ratio in the survival analysis. The associated probability value corresponds to the null hypothesis that the hazard was the same between groups (ie, hazard ratio=1). Differences in mean values of SDAAM, night heart rate, and patient activity between the groups were compared by ANOVA. Mean differences in SDAAM, night heart rate, and patient activity leading up to hospitalization were evaluated by paired t test. Significance was established at the P=0.05 level. The software used for these analyses was SAS (Statistical Analysis System) version 8.2.
| Results |
|---|
|
|
|---|
|
SDAAM and Mortality Risk
Of the 397 patients in the trial, 370 had SDAAM values from weeks 5 to 8 after implantation. The 27 patients not included had no save-to-disk information available. Thirty-six patients had SDAAM values <50 ms, 227 were in the intermediate group (SDAAM 50 to 100 ms), and 107 had SDAAM values >100 ms. There were 34 deaths among these patients in the first 12 months after implantation (9% absolute 1-year mortality). All-cause mortality was higher in the low-SDAAM group (SDAAM <50 ms), with a hazard ratio of 3.20 (P=0.02), as shown in Figure 2. The critical events committee adjudicated 25 of these 34 deaths as cardiovascular in origin. For cardiovascular death specifically, the hazard ratio was 4.43 (P=0.01).
|
Measured Parameters and Clinical Events
In the 288 patients with SDAAM data available, absolute SDAAM values were persistently lower in patients who were hospitalized or who died (74±22 ms) when averaged over weeks 5 to 52 after implantation compared with the other groups (88±25 ms, P=0.0007 for minor event; 90±22 ms, P<0.0001 for no event compared with hospitalization; no statistical difference between minor-event and no-event groups) as shown in Figure 3. SDAAM was not different between hospitalized patients and patients who died (75±25 versus 73±20 ms, P=0.6). Nighttime heart rate was higher in the patients who were hospitalized or died (73±11 bpm) than in those with minor exacerbations (69±8 bpm, P=0.05) and individuals without events (69±9 bpm, P=0.014). Nighttime heart rate was not different between patients who were hospitalized and patients who died (73±10 versus 72±12 bpm, P=0.5).
|
Thirty-four patients had cardiovascular hospitalizations at least 100 days after implantation and save-to-disk data for the 12 weeks preceding hospitalization. SDAAM declined before hospitalization from 76±27 ms at baseline (9 to 12 weeks before hospitalization) to 64±26 ms at the time of hospitalization (P<0.0001). The change in SDAAM before hospitalization was apparent up to 3 weeks before the event, as summarized in the aggregate trends for the 34 patients requiring hospitalization in Figure 4.
|
Of the 34 patients with hospitalization events, 14 were used to develop the detection algorithm, and 20 were used to validate algorithm performance. The ROC curve for the SDAAM detection algorithm is shown in Figure 5. At a threshold of 200 ms days, a sensitivity rate of 70% was associated with 2.4 false-positive events per patient-year of follow-up. True-positive detections occurred for 70% (7 of 10) of patients undergoing ß-blocker therapy and 70% (7 of 10) of patients not receiving ß-blocker therapy. The median time between the threshold crossing and hospital admission was 16 days. These results were obtained for a single detection threshold applied to all patients.
|
Night heart rate increased and patient activity declined as patients approached the time of hospitalization, but the changes were less sensitive than SDAAM (Figure 4). Specifically, night heart rate increased from 75±11 bpm at baseline to 78±11 bpm at the time of hospitalization (P=0.014). Patient activity declined from 188±109 to 164±118 min/d when baseline values were compared with values at the time of hospitalization (P=0.028). The ROC curves for night heart rate and patient activity demonstrated lower sensitivity than SDAAM over the entire range of false-positive rates (Figure 5).
| Discussion |
|---|
|
|
|---|
Autonomic Mechanisms of Heart Failure
The pathophysiology underlying heart failure is traditionally thought to be a condition of invariable sympathetic activation and parasympathetic withdrawal arising from decreased cardiac output.13,14 Circulating plasma norepinephrine values directly correlate to mortality risk and symptom severity,15 but recent basic science data and clinical observations suggest that the reality of autonomic derangements is likely much more complex and individual.16 The recent MOXonidine CONgestive heart failure (MOXCON) trial demonstrated harm in patients who had central inhibition of sympathetic outflow even though plasma norepinephrine levels declined dramatically.17 In dogs with ischemic left ventricular dysfunction, sympathetic activation only occurred in a group at high risk for sudden death despite similar degrees of left ventricular injury in the low-risk animals.18 Therefore, the autonomic response to heart failure represents a complex pathophysiology with mechanisms that have proved difficult to study over time, especially with regard to other hormonal influences on heart rate control. Longer-term HRV, as measured by SDAAM in the present study, may more accurately reflect very-low-frequency oscillations seen in power spectral analysis of heart rate.19 Very-low-frequency heart rate oscillations mostly arise from parasympathetic cardiac control but also reflect influences from the renin-angiotensin system.19 The SDAAM decline observed in the present study as patients became clinically decompensated likely reflects a combination of sympathetic activation, parasympathetic withdrawal, and renin-angiotensin activation. Which component predominated cannot be determined from this study.
Sympathetic activation was important, however, because patients with more severe symptoms, who were less likely to tolerate ß-blocker therapy, had low long-term SDAAM persistently over the 12-month follow-up period. Night heart rate also increased as patients neared decompensation, which supports the possibility of both sympathetic activation and parasympathetic withdrawal, in agreement with other studies.20 A decline in patient activity was also observed in the present trial as patient status deteriorated, which may have played a role in the changes observed in the SDAAM parameter.
SDAAM declined several days before hospitalization, which suggests that the neurohormonal control system of the heart detects a change in status and responds long before the patient develops symptoms that would traditionally bring them to the attention of the physician. If this type of insight into the hearts control system can be monitored continuously, especially with a reliable automatic detection system similar to the one used in the present study, patient monitoring and follow-up could be simplified by obtaining the information from the patients home through an Internet link. This type of information processing system is currently available.
Clinical Utility of Continuous SDAAM Measurement
Early warning of potential clinical decompensation can be obtained from continuous measurement of right ventricular pressures with a permanently implanted hemodynamic monitoring system.5 Implantable hemodynamic monitoring systems are still experimental and are in the midst of clinical validation, but the preliminary data suggest that pressure changes can be detected 4 to 5 days before hospitalization.5 Continuous SDAAM measurements may have similar utility and may serve as an accurate early warning system. The clinical response to a decline in SDAAM may simply be to increase the frequency of patient assessment to ensure appropriate volume management. If this is possible, then a reliable means to stratify patient risk may identify which individuals need frequent follow-up. This utility would be very helpful in heart failure treatment programs that serve large patient populations. Automatic surveillance of downloaded information from the patients home could flag changes that portend a poor immediate prognosis and notify the physician only when a problem arises. The use of continuous SDAAM would complement traditional use of physical examination, with the possibility of stratifying follow-up needs and clinical assessments when dealing with large numbers of heart failure patients.
Conclusions
This study demonstrated that patients with heart failure have stable autonomic activity until they start to decline clinically. Neurohormonal activation, as suggested by a decrease in SDAAM, occurs several days to weeks before patients decompensate enough to require hospitalization. Changes in SDAAM can be tracked with information derived from permanently implanted cardiac resynchronization devices and may provide another means to remotely acquire important physiological information about heart failure patients, which may be helpful in day-to-day management of the syndrome.
| References |
|---|
|
|
|---|
2. Fonarow G, Stevenson LW, Walden JA, et al. Impact of a comprehensive heart failure management program on hospital readmission and functional status of patients with advanced heart failure. J Am Coll Cardiol. 1997; 30: 725732.[Abstract]
3. Hanumanthu S, Butler J, Chomsky D, et al. Effect of a heart failure program on hospitalization frequency and exercise tolerance. Circulation. 1997; 96: 28422848.
4. Adamson PB, Kleckner K, VanHout WL, et al. Cardiac resynchronization therapy improves heart rate variability in patients with symptomatic heart failure. Circulation. 2003; 108: 266269.
5. Adamson PB, Magalski A, Braunschweig F, et al. Ongoing right ventricular hemodynamics heart failure: clinical value of measurements derived from an implantable monitoring system. J Am Coll Cardiol. 2003; 41: 563567.
6. Kleiger RE, Miller JP, Bigger JT Jr, et al. Decreased heart rate variability and its association with increased mortality after acute myocardial infarction. Am J Cardiol. 1987; 59: 256262.[CrossRef][Medline] [Order article via Infotrieve]
7. Task Force of the European Society of Cardiology and the North American Society of Pacing and Electrophysiology. Heart rate variability: standards of measurement, physiological interpretation and clinical use. Circulation. 1996; 93: 10431065.
8. La Rovere MT, Pinna GD, Maestri R, et al. Short-term heart rate variability strongly predicts sudden cardiac death in chronic heart failure patients. Circulation. 2003; 107: 565570.
9. La Rovere MT, Pinna GD, Hohnloser SH, et al. Baroreflex sensitivity and heart rate variability in the identification of patients at risk for life-threatening arrhythmias: implications for clinical trials. Circulation. 2001; 103: 20722077.
10. Bigger JT, Fleiss JL, Rolnitzky LM, et al. The ability of several short-term measures of RR variability to predict mortality after myocardial infarction. Circulation. 1993; 88: 927934.
11. Abraham WT, Fisher WG, Smith AL, et al, for the MIRACLE Study Group. Cardiac resynchronization in chronic heart failure. N Engl J Med. 2002; 346: 18451853.
12. Malik M, Camm AJ. Components of heart rate variability: what they really mean and what we really measure. Am J Cardiol. 1993; 72: 821822.[CrossRef][Medline] [Order article via Infotrieve]
13. Packer M. The neurohormonal hypothesis: a theory to explain the mechanism of disease progression in heart failure. J Am Coll Cardiol. 1992; 20: 248254.[Abstract]
14. Cohn JN. The management of chronic heart failure. N Engl J Med. 1996; 335: 490498.
15. Cohn JN, Levine TB, Olivari MT, et al. Plasma norepinephrine as a guide to prognosis in patients with chronic congestive heart failure. N Engl J Med. 1984; 311: 819823.[Abstract]
16. Cohn JN. Sympathetic nervous system in heart failure. Circulation. 2002; 106: 24172418.
17. Cohn JN, Pfeffer MA, Rouleau J, et al. Adverse mortality effect of central sympathetic inhibition with sustained-release moxonidine in patients with heart failure (MOXCON). Eur J Heart Fail. 2003; 5: 659667.[CrossRef][Medline] [Order article via Infotrieve]
18. Adamson PB, Vanoli E. Early autonomic and repolarization abnormalities contribute to lethal arrhythmias in chronic ischemic heart failure: characteristics of a novel heart failure model in dogs with postmyocardial infarction left ventricular dysfunction. J Am Coll Cardiol. 2001; 37: 17411748.
19. Taylor JA, Carr DL, Myers CW, et al. Mechanisms underlying very-low-frequency RR-interval oscillations in humans. Circulation. 1998; 98: 547555.
20. Moller M, Nielsen BL, Fabricius J. Paroxysmal ventricular tachycardia during repeated 24-hour ambulatory electrocardiographic monitoring of postmyocardial infarction patients. Br Heart J. 1980; 43: 447453.
This article has been cited by other articles:
![]() |
M. Landolina, M. Gasparini, M. Lunati, M. Santini, R. Rordorf, A. Vincenti, P. Diotallevi, A. S. Montenero, C. Bonanno, T. De Santo, et al. Heart rate variability monitored by the implanted device predicts response to CRT and long-term clinical outcome in patients with advanced heart failure Eur J Heart Fail, November 1, 2008; 10(11): 1073 - 1079. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Arya, M. Block, J. Kautzner, T. Lewalter, H. Mortel, S. Sack, B. Schumacher, P. Sogaard, M. Taborsky, D. Husser, et al. Influence of Home Monitoring on the clinical status of heart failure patients: Design and rationale of the IN-TIME study Eur J Heart Fail, November 1, 2008; 10(11): 1143 - 1148. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. R. Zile, T. D. Bennett, M. St. John Sutton, Y. K. Cho, P. B. Adamson, M. F. Aaron, J. M. Aranda Jr, W. T. Abraham, F. W. Smart, L. W. Stevenson, et al. Transition From Chronic Compensated to Acute Decompensated Heart Failure: Pathophysiological Insights Obtained From Continuous Monitoring of Intracardiac Pressures Circulation, September 30, 2008; 118(14): 1433 - 1441. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. J. Schwartz, G. M. De Ferrari, A. Sanzo, M. Landolina, R. Rordorf, C. Raineri, C. Campana, M. Revera, N. Ajmone-Marsan, L. Tavazzi, et al. Long term vagal stimulation in patients with advanced heart failure First experience in man Eur J Heart Fail, September 1, 2008; 10(9): 884 - 891. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Braunschweig, I. Ford, V. Conraads, M. R. Cowie, G. Jondeau, J. Kautzner, M. Lunati, R. M. Aguilera, C. M. Yu, M. Marijianowski, et al. Can monitoring of intrathoracic impedance reduce morbidity and mortality in patients with chronic heart failure? Rationale and design of the Diagnostic Outcome Trial in Heart Failure (DOT-HF) Eur J Heart Fail, September 1, 2008; 10(9): 907 - 916. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Fantoni, F. Regoli, A. Ghanem, S. Raffa, C. Klersy, A. Sorgente, F. Faletra, M. Baravelli, L. Inglese, J. A. Salerno-Uriarte, et al. Long-term outcome in diabetic heart failure patients treated with cardiac resynchronization therapy Eur J Heart Fail, March 1, 2008; 10(3): 298 - 307. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Ritzema, I. C. Melton, A. M. Richards, I. G. Crozier, C. Frampton, R. N. Doughty, J. Whiting, S. Kar, N. Eigler, H. Krum, et al. Direct Left Atrial Pressure Monitoring in Ambulatory Heart Failure Patients: Initial Experience With a New Permanent Implantable Device Circulation, December 18, 2007; 116(25): 2952 - 2959. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Stellbrink and H.-J. Trappe The follow-up of cardiac devices: what to expect for the future? Eur. Heart J. Suppl., December 1, 2007; 9(suppl_I): I113 - I115. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. E. Wolfel Can We Predict and Prevent the Onset of Acute Decompensated Heart Failure? Circulation, October 2, 2007; 116(14): 1526 - 1529. [Full Text] [PDF] |
||||
![]() |
Authors/Task Force Members, P. E. Vardas, A. Auricchio, J.-J. Blanc, J.-C. Daubert, H. Drexler, H. Ector, M. Gasparini, C. Linde, F. B. Morgado, et al. Guidelines for cardiac pacing and cardiac resynchronization therapy: The Task Force for Cardiac Pacing and Cardiac Resynchronization Therapy of the European Society of Cardiology. Developed in Collaboration with the European Heart Rhythm Association Europace, October 1, 2007; 9(10): 959 - 998. [Full Text] [PDF] |
||||
![]() |
Authors/Task Force Members, P. E. Vardas, A. Auricchio, J.-J. Blanc, J.-C. Daubert, H. Drexler, H. Ector, M. Gasparini, C. Linde, F. B. Morgado, et al. Guidelines for cardiac pacing and cardiac resynchronization therapy: The Task Force for Cardiac Pacing and Cardiac Resynchronization Therapy of the European Society of Cardiology. Developed in Collaboration with the European Heart Rhythm Association Eur. Heart J., September 2, 2007; 28(18): 2256 - 2295. [Full Text] [PDF] |
||||
![]() |
E. Page, S. Cazeau, P. Ritter, D. Galley, and C. Casset Physiological approach to monitor patients in congestive heart failure: application of a new implantable device-based system to monitor daily life activity and ventilation Europace, August 1, 2007; 9(8): 687 - 693. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. A. McAlister, J. Ezekowitz, N. Hooton, B. Vandermeer, C. Spooner, D. M. Dryden, R. L. Page, M. A. Hlatky, and B. H. Rowe Cardiac Resynchronization Therapy for Patients With Left Ventricular Systolic Dysfunction: A Systematic Review JAMA, June 13, 2007; 297(22): 2502 - 2514. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. R. Raj, D. E. Roach, M. L. Koshman, and R. S. Sheldon Standard Deviation of Sequential Five-Minute R-R Interval Means (SDANN) Is a Prognostic Marker, but Not Necessarily an Autonomic Marker J. Am. Coll. Cardiol., September 19, 2006; 48(6): 1285 - 1286. [Full Text] [PDF] |
||||
![]() |
J. M. Aranda Jr, G. W. Woo, R. S. Schofield, E. M. Handberg, J. A. Hill, A. B. Curtis, S. F. Sears, J. S. Goff, D. F. Pauly, and J. B. Conti Management of Heart Failure After Cardiac Resynchronization Therapy: Integrating Advanced Heart Failure Treatment With Optimal Device Function J. Am. Coll. Cardiol., December 20, 2005; 46(12): 2193 - 2198. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Fantoni, S. Raffa, F. Regoli, F. Giraldi, M. T. La Rovere, J. Prentice, F. Pastori, S. Fratini, J. A. Salerno-Uriarte, H. U. Klein, et al. Cardiac Resynchronization Therapy Improves Heart Rate Profile and Heart Rate Variability of Patients With Moderate to Severe Heart Failure J. Am. Coll. Cardiol., November 15, 2005; 46(10): 1875 - 1882. [Abstract] [Full Text] [PDF] |
||||
![]() |
C.-M. Yu, L. Wang, E. Chau, R. H.-W. Chan, S.-L. Kong, M.-O. Tang, J. Christensen, R. W. Stadler, and C.-P. Lau Intrathoracic Impedance Monitoring in Patients With Heart Failure: Correlation With Fluid Status and Feasibility of Early Warning Preceding Hospitalization Circulation, August 9, 2005; 112(6): 841 - 848. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. R. Raj, D. E. Roach, R. S. Sheldon, P. B. Adamson, A. L. Smith, W. T. Abraham, K. J. Kleckner, R. W. Stadler, and M. M. Rhodes Letter Regarding Article by Adamson et al, "Continuous Autonomic Assessment in Patients With Symptomatic Heart Failure: Prognostic Value of Heart Rate Variability Measured by an Implanted Cardiac Resynchronization Device" * Response Circulation, July 12, 2005; 112(2): e37 - e38. [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2004 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |