(Circulation. 2004;110:1814-1818.)
© 2004 American Heart Association, Inc.
Imaging |
From the Noninvasive Cardiac Imaging Laboratory, University of Chicago, Chicago, Ill (V.M.-A., L.S., L.W., P.M., E.G.C., R.K., R.M.L.); Dipartimento di Bioingegneria, Politecnico di Milano, Milan, Italy (E.G.C.); and Philips Medical Systems, Andover, Mass (I.S.S.).
Correspondence to Victor Mor-Avi, PhD, University of Chicago MC5084, 5841 S. Maryland Ave, Chicago Illinois 60637. E-mail vmoravi{at}medicine.bsd.uchicago.edu
Received February 3, 2004; de novo received March 17, 2004; revision received May 24, 2004; accepted May 26, 2004.
| Abstract |
|---|
|
|
|---|
Methods and Results Echocardiographic imaging was performed (Philips 7500) in 21 patients referred for CMR imaging (1.5 T, GE). Apical 2- and 4-chamber views and RT3D data sets were acquired and analyzed by 2 independent observers. The RT3D data sets were used to select nonforeshortened apical 2- and 4-chamber views (3DQ-QLAB, Philips). In both 2D and RT3D images, LV long axis was measured; endocardial and epicardial boundaries were traced, and mass was calculated by use of the biplane method of disks. CMR LV mass values were obtained through standard techniques (MASS Analysis, GE). The RT3D data resulted in significantly larger LV long-axis dimensions and measurements of LV mass that correlated with CMR better (r=0.90) than 2D (r=0.79). The 2D technique underestimated LV mass (bias, 39%), whereas RT3D measurements showed only minimal bias (3%). The 95% limits of agreement were significantly wider for 2D (52%) than RT3D (28%). Additionally, the RT3D technique reduced the interobserver variability (37% to 7%) and intraobserver variability (19% to 8%).
Conclusions RT3D imaging provides the basis for accurate and reliable measurement of LV mass.
Key Words: imaging echocardiography hypertrophy, left ventricular
| Introduction |
|---|
|
|
|---|
Accordingly, 3D imaging appears to be a potential alternative for accurate evaluation of LV mass, which can be measured directly from the 3D data sets without geometrical assumptions and modeling.3,4 Previous studies have demonstrated the feasibility of this approach from 3D data reconstructed from consecutive multiplane acquisitions.57 The recent development of real-time 3D (RT3D) imaging technology is especially appealing because it can potentially allow nearly online quantification of LV volumes and mass without the need for tedious reconstruction.
In the present study, we hypothesized that RT3D data sets could be used to provide fast and reliable estimates of LV mass that would be more accurate than those measured with conventional 2D techniques. We assumed that from the RT3D data sets it is possible to select anatomically correct apical long-axis views and thus avoid foreshortening. To test this hypothesis, we compared LV mass values obtained in a group of patients using the biplane method of disks from 2D apical 2- and 4-chamber views and those obtained with the biplane method from anatomically correct apical 2- and 4-chamber views selected offline from RT3D data sets against the current standard for LV mass values obtained from cardiac MR (CMR) images.
| Methods |
|---|
|
|
|---|
CMR Assessment of LV Mass
CMR images were obtained with a 1.5-T scanner (General Electric) with a phased-array torso coil. ECG-gated localizing spin-echo sequences were used to identify the LV long axis. Steady-state free-precession dynamic gradient-echo cine loops were obtained during 12-second breathholds. Data were reconstructed to give 20 frames per cardiac cycle. In all patients, 6 to 10 short-axis cine loops were obtained from the AV ring to the apex (9-mm slice thickness, no gaps).
CMR cine loops were analyzed offline with commercial software (MASS Analysis, General Electric). In every short-axis slice, endocardial and epicardial contours were manually traced at end diastole, while including the papillary muscles in the LV cavity. All tracings were performed by an investigator experienced in the interpretation of CMR images who had no knowledge of the echocardiographic measurements. For each patient, the traced contours were used to calculate LV mass, which served as the reference for comparisons against 2D and RT3D echocardiographic data.
2D Echocardiographic Assessment of LV Mass
The 2D harmonic imaging was performed by use of a commercial ultrasound system (SONOS 7500, Philips) with an S3 transducer from the apical window with the patient in the left lateral decubitus position. Five consecutive heartbeats were acquired at held respiration from each 4- and 2-chamber views while care was taken to avoid foreshortening. Images were stored digitally and analyzed offline (EnConcert, Philips). For both apical views, end-diastolic frames were selected as those captured at the peak of the R wave. In each view, endocardial and epicardial contours, including the papillary muscles in the LV cavity, were traced manually (Figure 1, top). The points of insertion of the mitral leaflets into the annulus were connected by a straight line, and the LV long-axis dimension was measured as the distance between the center of this line and the most distal point at the apical endocardium. The traced contours were used to calculate endocardial and epicardial LV volumes from Simpsons formula.8,9 The difference between the epicardial and endocardial volumes was computed for each view and multiplied by the specific mass of myocardial tissue (1.05 g/mL) to represent a biplane estimate of LV mass. The accuracy of these estimates was determined by comparing them with the CMR values.
|
RT3D Echocardiographic Assessment of LV Mass
Harmonic RT3D imaging was performed in the same setting with the fully sampled matrix-array transducer (X4, 2 to 4 MHz) that uses
3000 elements to obtain a pyramidal volume data set from a single window. Gain and compression controls, as well as the time gain compensation settings, were optimized for quality of 3D images. Care was taken to include the entire LV cavity within the pyramidal scan volume. RT3D data sets were then acquired with the wide-angled acquisition (93°x80°) mode in which 4 wedge-shaped subvolumes (93°x20° each) were obtained from 4 different cardiac cycles during held respiration. Acquisition was triggered to the R wave of every other cardiac cycle to allow time for storage of each subvolume, resulting in a total acquisition time of 8 heartbeats.
The RT3D data sets were analyzed with commercial software (3DQ-QLab, Philips). The pyramidal volume data (Figure 2, bottom right) were displayed in 3 different cross sections that could be modified interactively by use of the following color-coding convention. Apical 4-chamber view displayed in a green box (Figure 2, top left) could be controlled by manually shifting or rotating a green line shown in the apical 2-chamber view (Figure 2, top right) and the short-axis view (Figure 2, bottom left). Apical 2-chamber view displayed in a red box (figure 2, top right) could be controlled by shifting or rotating a red line shown in the apical 4-chamber and the short-axis views. The short-axis view displayed in a blue box (Figure 2, bottom left) could be controlled by shifting or rotating a blue line in both apical views.
|
Using this convention, we selected the anatomically correct 2- and 4-chamber views with the largest long-axis dimensions as follows. First, the orientation of the long axis of the 2-chamber view (Figure 2, top right) was determined by positioning the red line along the LV long axis in the 4-chamber view (Figure 2, top left). Then, the orientation of the long axis of the 4-chamber view (Figure 2, top left) was determined by positioning the green line along the LV long axis in the 2-chamber view (Figure 2, top right). Then, the midpapillary short-axis plane (Figure 2, bottom left) was defined by positioning the blue lines in both apical views (Figure 2 top) at the midpapillary level perpendicular to the long axis of the ventricle determined in the previous 2 steps. The final selection of the apical views was achieved by changing the position of the green and red lines in the short-axis view (Figure 2, bottom left). The green line was forced to pass near the center of the LV cavity and the most distant point on the right ventricular endocardium. Similarly, the red line was forced to pass near the center of the LV cavity perpendicular to the green line. We made these last 2 fine-tuning steps while carefully ensuring that the resulting apical views have the largest long-axis dimension of the ventricle; thus, they represent the anatomically correct apical views.
In these 2 planes, endocardial and epicardial contours were traced manually at end diastole with the papillary muscles included in the LV cavity (Figure 1, bottom). The LV long-axis dimension was measured the same way as in 2D. The traced contours were then used to calculate LV volumes by use of the biplane Simpson formula with 20 slices8,9 incorporated into the analysis software. The accuracy of these RT3D-derived LV mass values was determined by comparing them with the CMR values.
Interobserver and Intraobserver Variability
To determine the interobserver variability in the 2D and RT3D evaluations of LV mass, all measurements were repeated by a second observer blinded to the values obtained by the first observer. Interobserver variability was calculated for each patient as the absolute difference between the 2 observers in percent of their mean. To assess intraobserver variability, all measurements were repeated 1 month later by an observer blinded to the results of the previous measurements. Intraobserver variability was calculated as the difference between the 2 measurements in percent of their mean.
Statistical Analysis
All values were expressed as mean±SD. Agreement between each technique, 2D and RT3D, and CMR was evaluated by use of a paired t test and linear regression analysis with Pearsons correlation coefficient. In addition, Bland-Altman analysis was used to determine the bias and 95% limits of agreement between the echocardiographic measurements and their CMR counterparts. These analyses were performed for each observer individually and for the 2 observers measurements combined. The differences between LV long-axis dimension in the 2D and RT3D techniques were tested with a paired t test. Interobserver and intraobserver variability values were averaged for all patients and tested by use of a paired t test for significance of differences between techniques. Values of P<0.05 were considered significant.
| Results |
|---|
|
|
|---|
The 2D technique yielded an LV mass of 103±39 g (range, 38 to 165 g) by observer 1 and 72±22 g (range, 30 to 107 g) by observer 2, which were significantly lower than the CMR values (P<0.05). The combined readings of the 2 observers resulted in a correlation of r=0.79 with CMR values, a regression slope of 0.54, and an intercept of 19 g (Figure 3, left). Bland-Altman analysis (Figure 4, left) confirmed the underestimation by the 2D technique by demonstrating a bias of 39 g (39% of the mean) with 95% limits of agreement at ±58 g (±52% of the mean). The LV long-axis dimension was 9.2±0.7 cm in the 4-chamber view and 8.9±0.8 cm in the 2-chamber view.
|
|
Identification of nonforeshortened apical views from the RT3D data sets was achieved in all 19 patients, in most cases within 20 seconds. The RT3D-based technique yielded an LV mass of 120±35 g (range, 57 to 183 g) by observer 1 and 123±39 g (range, 57 to 213 g) by observer 2, which were similar to the CMR values (P=NS). The combined readings of the 2 observers resulted in a correlation with CMR values of r=0.90, a regression line slope of 0.86, and an intercept of 14 g (Figure 3, right). Bland-Altman analysis (Figure 4, right) showed no significant underestimation by the RT3D-based technique as reflected by only a minimal bias of 4 g (3% of the mean) with 95% limits of agreement at ±34 g (±28% of the mean). The LV long-axis dimension was 9.4±0.7 cm in the 4-chamber view and 9.4±0.8 cm in the 2-chamber view. These values were significantly larger than in the 2D measurements (Figure 5).
|
The interobserver variability was 37±19% of the measured LV mass values, ranging between 5% and 66% between patients, for the 2D technique and 7±7%, ranging between 0% and 23%, for the RT3D-based technique (P<0.05 between techniques). The intraobserver variability was 19±11% of the measured LV masses, ranging between 2% and 37% between patients, for the 2D technique and 8±5%, ranging between 1% and 22%, for the RT3D-based technique (P<0.05 between techniques).
| Discussion |
|---|
|
|
|---|
Despite these obvious advantages, LV mass measurements from 3D data have not become standard in clinical laboratories because of cumbersome and time-consuming data acquisition and analysis; thus, 3D assessment of LV mass remained limited mainly to the research arena. The recent development of the matrix-array transducer offers an opportunity for real-time imaging and online rendering of the heart without reconstruction, thus bringing the RT3D imaging into the mainstream.
Our goal was to explore the potential of this new technology for quantitative evaluation of LV mass. We hypothesized that data sets acquired with the matrix-array transducer could be cross sectioned offline to allow selection of anatomically correct apical views. Specifically, we assumed that tracing endocardial and epicardial boundaries on anatomically correct, nonforeshortened apical views would result in improved accuracy in the quantification of LV mass.2 We tested this hypothesis by acquiring and analyzing RT3D data in a group of consecutive patients referred for CMR imaging. This study design automatically provided us with validation of our measurements using an accepted standard reference method for LV mass.19 The interactive technique we used to identify anatomically correct, nonforeshortened apical views was based on that routinely applied to scout CMR images to identify anatomically correct short-axis views for imaging LV function.20
We found that RT3D data sets obtained with commercially available equipment provided images with sufficient detail to allow easy offline selection of nonforeshortened apical views, with the exception of 2 patients (10%) who were excluded because their hearts were too big to fit into the scan volume. This may indicate a limitation of this technique in patients with dilated cardiomyopathies.
The accuracy of the 3D technique could likely be improved further by increasing the number of planes used to trace the endocardial and epicardial boundaries, although adding planes beyond 3 was shown to have little if any added benefit in normally shaped ventricles.3 Another related limitation of this study is that no patients with nonconcentric hypertrophic cardiomyopathy were included. It is likely that in such patients, the RT3D measurements would have yielded lower levels of agreement with the reference method, because these measurements were performed in 2 orthogonal planes only. A more comprehensive approach such as detection of the endocardial and epicardial surfaces would be beneficial in these patients.
Nevertheless, to allow fair intertechnique comparisons, we used the same biplane method of disks to calculate LV mass from endocardial and epicardial boundaries traced on the 2D images and those selected from the RT3D data sets. Although this strategy might have compromised to some extent the accuracy of the 3D measurements compared with potentially more accurate albeit more complex and time-consuming multiplane analysis, it resulted in a quick, simple, and intuitive technique.
Despite these limitations, our RT3D-derived estimates of LV mass resulted in higher levels of agreement with the CMR reference values than the conventional 2D measurements obtained in the same subjects. Our measurements of the LV long-axis dimension showed that offline cross sectioning of the RT3D data sets resulted in significantly less foreshortened and thus more anatomically correct apical views than the conventional 2D measurements, thus explaining the improved accuracy of the former technique. In addition, our results indicate that the 3D technique provides more reproducible measurements than the conventional 2D technique. An important point is that the 3D measurements varied less between observers, reflecting reduced operator dependency. The facts that the interobserver variability of the RT3D technique is virtually equal to the intraobserver variability and that both are <10% indicate reproducibility that is acceptable for clinical use. Also of note is the fact that although CMR imaging is commonly used as a standard reference technique for LV mass measurements, its intermeasurement and intrameasurement reproducibility, albeit very high, is <100%. In this regard, the RT3D echocardiographic quantification of LV mass, even with variability <10%, remains inferior to CMR, with variability values <1% reported by multiple investigators.2124 This difference can be explained by the relatively limited ability of ultrasound imaging compared with MR to visualize the endocardial and, even more so, the epicardial boundaries.
| Conclusions |
|---|
|
|
|---|
| References |
|---|
|
|
|---|
2. Devereux RB, Pini R, Aurigemma GP et al. Measurement of left ventricular mass: methodology and expertise. J Hypertens. 1997; 15: 801809.[CrossRef][Medline] [Order article via Infotrieve]
3. Teupe C, Takeuchi M, Yao J, et al. Determination of left ventricular mass by three-dimensional echocardiography: in vitro validation of a novel quantification method using multiple equi-angular rotational planes for rapid measurements. Int J Cardiovasc Imaging. 2002; 18: 161167.[CrossRef][Medline] [Order article via Infotrieve]
4. Myerson SG, Montgomery HE, World MJ, et al. Left ventricular mass: reliability of M-mode and 2-dimensional echocardiographic formulas. Hypertension. 2002; 40: 673678.
5. Gopal AS, Keller AM, Shen Z et al. Three-dimensional echocardiography: in vitro and in vivo validation of left ventricular mass and comparison with conventional echocardiographic methods. J Am Coll Cardiol. 1994; 24: 504513.[Abstract]
6. Kuhl HP, Franke A, Frielingsdorf J, et al. Determination of left ventricular mass and circumferential wall thickness by three-dimensional reconstruction: in vitro validation of a new method that uses a multiplane transesophageal transducer. J Am Soc Echocardiogr. 1997; 10: 107119.[CrossRef][Medline] [Order article via Infotrieve]
7. Sapin PM, Gopal AS, Clarke GB, et al. Three-dimensional echocardiography compared to two-dimensional echocardiography for measurement of left ventricular mass anatomic validation in an open chest canine model. Am J Hypertens. 1996; 9: 467474.[CrossRef][Medline] [Order article via Infotrieve]
8. Helak JW, Reichek N. Quantitation of human left ventricular mass and volume by two-dimensional echocardiography: in vitro anatomic validation. Circulation. 1981; 63: 13981407.
9. Wyatt HL, Heng MK, Meerbaum S, et al. Cross-sectional echocardiography, I: analysis of mathematic models for quantifying mass of the left ventricle in dogs. Circulation. 1979; 60: 11041113.
10. Missouris CG, Forbat SM, Singer DR, et al. Echocardiography overestimates left ventricular mass: a comparative study with magnetic resonance imaging in patients with hypertension. J Hypertens. 1996; 14: 10051010.[Medline] [Order article via Infotrieve]
11. Kuhl HP, Bucker A, Franke A, et al. Transesophageal 3-dimensional echocardiography: in vivo determination of left ventricular mass in comparison with magnetic resonance imaging. J Am Soc Echocardiogr. 2000; 13: 205215.[Medline] [Order article via Infotrieve]
12. Kuhl HP, Hanrath P, Franke A. M-mode echocardiography overestimates left ventricular mass in patients with normal left ventricular shape: a comparative study using three-dimensional echocardiography. Eur J Echocardiogr. 2003; 4: 312319.[CrossRef][Medline] [Order article via Infotrieve]
13. Chuang ML, Beaudin RA, Riley MF, et al. Three-dimensional echocardiographic measurement of left ventricular mass: comparison with magnetic resonance imaging and two-dimensional echocardiographic determinations in man. Int J Card Imaging. 2000; 16: 347357.[CrossRef][Medline] [Order article via Infotrieve]
14. Kasprzak JD, Vletter WB, van Meegen JR, et al. Improved quantification of myocardial mass by three-dimensional echocardiography using a deposit contrast agent. Ultrasound Med Biol. 1998; 24: 647653.[CrossRef][Medline] [Order article via Infotrieve]
15. Gopal AS, Schnellbaecher MJ, Shen Z, et al. Freehand three-dimensional echocardiography for measurement of left ventricular mass: in vivo anatomic validation using explanted human hearts. J Am Coll Cardiol. 1997; 30: 802810.[Abstract]
16. Leotta DF, Munt B, Bolson EL, et al. Quantitative three-dimensional echocardiography by rapid imaging from multiple transthoracic windows: in vitro validation and initial in vivo studies. J Am Soc Echocardiogr. 1997; 10: 830839.[CrossRef][Medline] [Order article via Infotrieve]
17. Akinboboye O, Germano G, Idris O, et al. Left ventricular mass measured by myocardial perfusion gated SPECT: relation to three-dimensional echocardiography. Clin Nucl Med. 2003; 28: 392397.[CrossRef][Medline] [Order article via Infotrieve]
18. Takuma S, Cabreriza SE, Sciacca R, et al. Determination of left ventricular mass by real-time three-dimensional echocardiography: in vitro validation. Echocardiography. 2000; 17: 665674.[CrossRef][Medline] [Order article via Infotrieve]
19. Soler R, Rodriguez E, Marini M. Left ventricular mass in hypertrophic cardiomyopathy: assessment by three-dimensional and geometric MR methods. J Comput Assist Tomogr. 1999; 23: 577582.[CrossRef][Medline] [Order article via Infotrieve]
20. Lee VS, Resnick D, Bundy JM, et al. Cardiac function: MR evaluation in one breath hold with real-time true fast imaging with steady-state precession. Radiology. 2002; 222: 835842.
21. Nagel E, Schneider U, Schalla S, et al. Magnetic resonance real-time imaging for the evaluation of left ventricular function. J Cardiovasc Magn Reson. 2000; 2: 714.[Medline] [Order article via Infotrieve]
22. Myerson SG, Bellenger NG, Pennell DJ. Assessment of left ventricular mass by cardiovascular magnetic resonance. Hypertension. 2002; 39: 750755.
23. Moon JC, Lorenz CH, Francis JM, et al. Breath-hold FLASH and FISP cardiovascular MR imaging: left ventricular volume differences and reproducibility. Radiology. 2002; 223: 789797.
24. Vignaux O, Duboc D, Coste J, et al. Reproducibility of left ventricular mass measurement using a half-Fourier black-blood single-shot fast spin-echo sequence within a single breath hold: comparison with a conventional multiple breath-hold segmented gradient echo technique in patients. J Magn Reson Imaging. 2002; 15: 654660.[CrossRef][Medline] [Order article via Infotrieve]
This article has been cited by other articles:
![]() |
N. Tzemos, E. Lyseggen, C. Silversides, M. Jamorski, J. H. Tong, P. Harvey, J. Floras, and S. Siu Endothelial Function, Carotid-Femoral Stiffness, and Plasma Matrix Metalloproteinase-2 in Men With Bicuspid Aortic Valve and Dilated Aorta J. Am. Coll. Cardiol., February 16, 2010; 55(7): 660 - 668. [Abstract] [Full Text] [PDF] |
||||
![]() |
K.Y. E. Leung and J. G. Bosch Automated border detection in three-dimensional echocardiography: principles and promises Eur J Echocardiogr, February 6, 2010; (2010) jeq005v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
A Bhan, S Kapetanakis, and M J Monaghan Three-dimensional echocardiography Heart, January 15, 2010; 96(2): 153 - 163. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Muraru, L. P. Badano, G. Piccoli, P. Gianfagna, L. Del Mestre, D. Ermacora, and A. Proclemer Validation of a novel automated border-detection algorithm for rapid and accurate quantitation of left ventricular volumes based on three-dimensional echocardiography Eur J Echocardiogr, January 6, 2010; (2010) jep217v2. [Abstract] [Full Text] [PDF] |
||||
![]() |
C.-M. Yu, J. Y.-S. Chan, Q. Zhang, G. W.K. Yip, Y.-Y. Lam, A. Chan, D. Burkhoff, P.-W. Lee, and J. W.-H. Fung Impact of Cardiac Contractility Modulation on Left Ventricular Global and Regional Function and Remodeling J. Am. Coll. Cardiol. Img., December 1, 2009; 2(12): 1341 - 1349. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Fang, J. Y.-S. Chan, G. W.-K. Yip, J.-M. Xie, Q. Zhang, J. W.-H. Fung, Y.-Y. Lam, and C.-M. Yu Prevalence and determinants of left ventricular systolic dyssynchrony in patients with normal ejection fraction received right ventricular apical pacing: a real-time three-dimensional echocardiographic study Eur J Echocardiogr, November 20, 2009; (2009) jep171v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. A. Marsan, L. F Tops, P. Nihoyannopoulos, E. R Holman, and J. J Bax Real-time three dimensional echocardiography: current and future clinical applications Heart, November 15, 2009; 95(22): 1881 - 1890. [Full Text] [PDF] |
||||
![]() |
Y. Seo, T. Ishizu, Y. Enomoto, H. Sugimori, M. Yamamoto, T. Machino, R. Kawamura, and K. Aonuma Validation of 3-Dimensional Speckle Tracking Imaging to Quantify Regional Myocardial Deformation Circ Cardiovasc Imaging, November 1, 2009; 2(6): 451 - 459. [Abstract] [Full Text] [PDF] |
||||
![]() |
H.-J. Nesser, V. Mor-Avi, W. Gorissen, L. Weinert, R. Steringer-Mascherbauer, J. Niel, L. Sugeng, and R. M. Lang Quantification of left ventricular volumes using three-dimensional echocardiographic speckle tracking: comparison with MRI Eur. Heart J., July 1, 2009; 30(13): 1565 - 1573. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Mor-Avi, L. Sugeng, and R. M. Lang Real-Time 3-Dimensional Echocardiography: An Integral Component of the Routine Echocardiographic Examination in Adult Patients? Circulation, January 20, 2009; 119(2): 314 - 329. [Full Text] [PDF] |
||||
![]() |
J Pemberton, M Jerosch-Herold, X Li, L Hui, M Silberbach, W Woodward, K Thiele, A Kenny, and D J Sahn Accuracy of real-time, three-dimensional Doppler echocardiography for stroke volume estimation compared with phase-encoded MRI: an in vivo study Heart, September 1, 2008; 94(9): 1212 - 1213. [Full Text] [PDF] |
||||
![]() |
A-C Pouleur, J-B le Polain de Waroux, A Pasquet, B L Gerber, O Gerard, P Allain, and J-L J Vanoverschelde Assessment of left ventricular mass and volumes by three-dimensional echocardiography in patients with or without wall motion abnormalities: comparison against cine magnetic resonance imaging Heart, August 1, 2008; 94(8): 1050 - 1057. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Mor-Avi, C. Jenkins, H. P. Kuhl, H.-J. Nesser, T. Marwick, A. Franke, C. Ebner, B. H. Freed, R. Steringer-Mascherbauer, H. Pollard, et al. Real-time 3-dimensional echocardiographic quantification of left ventricular volumes multicenter study for validation with magnetic resonance imaging and investigation of sources of error. J. Am. Coll. Cardiol. Img., July 1, 2008; 1(4): 413 - 423. [Abstract] [Full Text] [PDF] |
||||
![]() |
S.-C. Yap, R.-J. M. van Geuns, A. Nemes, F. J. Meijboom, J. S. McGhie, M. L. Geleijnse, M. L. Simoons, and J. W. Roos-Hesselink Rapid and accurate measurement of LV mass by biplane real-time 3D echocardiography in patients with concentric LV hypertrophy: comparison to CMR Eur J Echocardiogr, March 1, 2008; 9(2): 255 - 260. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Agabiti-Rosei, M. L. Muiesan, and M. Salvetti Review: New approaches to the assessment of left ventricular hypertrophy Therapeutic Advances in Cardiovascular Disease, December 1, 2007; 1(2): 119 - 128. [Abstract] [PDF] |
||||
![]() |
J. N. Kirkpatrick, M. A. Vannan, J. Narula, and R. M. Lang Echocardiography in Heart Failure: Applications, Utility, and New Horizons J. Am. Coll. Cardiol., July 31, 2007; 50(5): 381 - 396. [Abstract] [Full Text] [PDF] |
||||
![]() |
A H Gershlick, M de Belder, J Chambers, D Hackett, R Keal, A Kelion, S Neubauer, D J Pennell, M Rothman, M Signy, et al. Role of non-invasive imaging in the management of coronary artery disease: an assessment of likely change over the next 10 years. A report from the British Cardiovascular Society Working Group Heart, April 1, 2007; 93(4): 423 - 431. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. D. Thomas and Z. B. Popovic Assessment of Left Ventricular Function by Cardiac Ultrasound J. Am. Coll. Cardiol., November 21, 2006; 48(10): 2012 - 2025. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. M. Lang, V. Mor-Avi, L. Sugeng, P. S. Nieman, and D. J. Sahn Three-Dimensional Echocardiography: The Benefits of the Additional Dimension J. Am. Coll. Cardiol., November 21, 2006; 48(10): 2053 - 2069. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. C. Houck, J. E. Cooke, and E. A. Gill Live 3D Echocardiography: A Replacement for Traditional 2D Echocardiography? Am. J. Roentgenol., October 1, 2006; 187(4): 1092 - 1106. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. L. Chuang and W. J. Manning Pitfalls in comparison of left ventricular mass measurements by echocardiography and cardiovascular magnetic resonance imaging Nephrol. Dial. Transplant., September 1, 2006; 21(9): 2668 - 2668. [Full Text] [PDF] |
||||
![]() |
L. Sugeng, V. Mor-Avi, L. Weinert, J. Niel, C. Ebner, R. Steringer-Mascherbauer, F. Schmidt, C. Galuschky, G. Schummers, R. M. Lang, et al. Quantitative Assessment of Left Ventricular Size and Function: Side-by-Side Comparison of Real-Time Three-Dimensional Echocardiography and Computed Tomography With Magnetic Resonance Reference Circulation, August 15, 2006; 114(7): 654 - 661. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. G. Caiani, L. Sugeng, L. Weinert, A. Capderou, R. M. Lang, and P. Vaida Objective evaluation of changes in left ventricular and atrial volumes during parabolic flight using real-time three-dimensional echocardiography J Appl Physiol, August 1, 2006; 101(2): 460 - 468. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Fukuda, G. Saracino, Y. Matsumura, M. Daimon, H. Tran, N. L. Greenberg, T. Hozumi, J. Yoshikawa, J. D. Thomas, and T. Shiota Three-Dimensional Geometry of the Tricuspid Annulus in Healthy Subjects and in Patients With Functional Tricuspid Regurgitation: A Real-Time, 3-Dimensional Echocardiographic Study Circulation, July 4, 2006; 114(1_suppl): I-492 - I-498. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. M. Lang, M. Bierig, R. B. Devereux, F. A. Flachskampf, E. Foster, P. A. Pellikka, M. H. Picard, M. J. Roman, J. Seward, J. Shanewise, et al. Recommendations for chamber quantification Eur J Echocardiogr, March 1, 2006; 7(2): 79 - 108. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. D. Jacobs, I. S. Salgo, S. Goonewardena, L. Weinert, P. Coon, D. Bardo, O. Gerard, P. Allain, J. L. Zamorano, L. P. de Isla, et al. Rapid online quantification of left ventricular volume from real-time three-dimensional echocardiographic data Eur. Heart J., February 2, 2006; 27(4): 460 - 468. [Abstract] [Full Text] [PDF] |
||||
![]() |
E G Caiani, C Corsi, L Sugeng, P MacEneaney, L Weinert, V Mor-Avi, and R M Lang Improved quantification of left ventricular mass based on endocardial and epicardial surface detection with real time three dimensional echocardiography Heart, February 1, 2006; 92(2): 213 - 219. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. M. Tsutsui, E. E. Martinez, C. E. Rochitte, J. F. Ramires, and W. Mathias Jr. Noninvasive evaluation of left circumflex coronary aneurysm by real-time three-dimensional echocardiography Eur J Echocardiogr, January 1, 2006; 7(1): 75 - 78. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J Monaghan Role of real time 3D echocardiography in evaluating the left ventricle Heart, January 1, 2006; 92(1): 131 - 136. [Full Text] [PDF] |
||||
![]() |
J. Chambers, K. Fox, and A. Fraser Recent advances in non-invasive cardiology: Article does not mention echocardiography BMJ, March 26, 2005; 330(7493): 731 - 731. [Full Text] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2004 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |