Circulation. 2004;110:II-74-II-78
doi: 10.1161/01.CIR.0000138947.63799.89
This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Walther, T.
Right arrow Articles by Mohr, F. W.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Walther, T.
Right arrow Articles by Mohr, F. W.
Related Collections
Right arrow CV surgery: valvular disease

(Circulation. 2004;110:II-74 – II-78.)
© 2004 American Heart Association, Inc.


Surgery for Valvular Heart Disease

Prospectively Randomized Evaluation of Stented Xenograft Hemodynamic Function in the Aortic Position

Thomas Walther, MD PhD; Sven Lehmann, MD; Volkmar Falk, MD PhD; Sebastian Metz, MD; Nicolas Doll, MD; Ardawan Rastan, MD; Marika Viehweg, RN; Markus Richter, MD; Jan Gummert, MD PhD; Friedrich W. Mohr, MD PhD

From Universität Leipzig, Herzzentrum, Klinik für Herzchirurgie, Germany.

Correspondence to Dr. Thomas Walther, Universität Leipzig, Herzzentrum, Klinik für Herzchirurgie, Strümpellstr. 39, 04289 Leipzig, Germany. E-mail walt{at}medizin.uni-leipzig.de


*    Abstract
up arrowTop
*Abstract
down arrowIntroduction
down arrowMethods
down arrowResults
down arrowDiscussion
down arrowReferences
 
Background— Standard stented aortic xenograft valves have not yet been compared regarding their hemodynamic function using a stratified intraoperative randomization protocol.

Methods and Results— 100 patients were prospectively included after intraoperative metric sizing of the decalcified aortic annulus. They received Mosaic (M) or Perimount (P) aortic valve replacement. Patient age was 73±5 years, 51 were female, and New York Heart Association (NYHA) functional class was 2.8±0.5. The 21-mm annulus group consisted of 5 (M)/7 (P) patients, the 23-mm annulus group of 20 (M)/20 (P), the 25-mm annulus group of 18 (M)/19 (P), and the 27-mm annulus group of 4 (M)/7 (P) patients, respectively. Hemodynamic function was evaluated using transthoracic echocardiography before discharge and at follow-up (438±352 days). Surgery was uncomplicated in all patients. Labeled valve sizes were 0.93 (M) and 1.05 (P) mm smaller than the annulus diameters (p=NS). In-hospital mortality was 5%, all nonvalve-related. Transvalvular blood flow velocities and transvalvular pressure gradients were significantly lower in the 25 P versus the 25 M group at baseline and in the 23 P and 25 P groups at follow-up. There was a significant regression of left ventricular mass index in all patients at follow-up. However, left ventricular mass regression was more pronounced after P aortic valve replacement.

Conclusion— Labeled sizes of prosthetic heart valves implanted are smaller than the true aortic annulus. Both standard aortic xenografts compared in this prospectively randomized trial provide a sufficient hemodynamic and functional outcome.


Key Words: aortic valve replacement • stented bioprosthesis • aortic annulus diameter • hemodynamics


*    Introduction
up arrowTop
up arrowAbstract
*Introduction
down arrowMethods
down arrowResults
down arrowDiscussion
down arrowReferences
 
Aortic stenosis is the most common acquired heart valve lesion in the Western societies; it is usually caused by degenerative changes with complex calcification of the native leaflets and the aortic annulus. In symptomatic patients or in the presence of severe stenosis with significant additional left ventricular hypertrophy, aortic valve replacement (AVR) is indicated1.

During the past decades, AVR using mechanical valves or stented xenografts has become a routine procedure with low perioperative risk.2–10 Prosthetic heart valves have undergone continuous modifications to reach a good level of product quality and to provide an acceptable postoperative hemodynamic performance over the past years. However, an optimal device, possibly mimicking the native physiological valve, is neither available nor in view for AVR.

While facing an aging patient population requiring AVR, there is an increasing interest in xenografts; patients aged 65 years and older are considered suitable candidates.1,2,4,7,8 Stented valves usually provide a sufficient functional outcome, as proven by several clinical studies.4–6,10

During AVR after complete decalcification, sizing is performed to select the largest fitting and thus hemodynamically most suitable valve size. Differences in valve sizes and in sizer dimensions exist between the clinically available stented xenografts.11–16 These may lead to misunderstandings, suboptimal valve selection, misinterpretation of clinical results, and significant discrepancies between hemodynamic outcome data. No randomized comparison of different stented xenografts has been published to our knowledge thus far. Furthermore, intraoperative stratified randomization according to the true aortic annulus diameter has not been described in the literature to our knowledge. Aim of the present study was to independently evaluate 2 standard stented aortic xenografts regarding their hemodynamic outcome in relation to metric intraoperative sizing.


*    Methods
up arrowTop
up arrowAbstract
up arrowIntroduction
*Methods
down arrowResults
down arrowDiscussion
down arrowReferences
 
One hundred patients with relevant aortic valve disease received AVR between March 2000 and April 2003. Patients were randomized to receive 1 of 2 standard stented xenografts (Mosaic, Medtronic Inc, Minneapolis, Minn or Perimount, Edwards Lifesciences, Irvine, Calif), and both were approved for routine clinical implantation (Figure 1). This prospective clinical trial was approved by the local ethical committee; all patients gave informed consent after the study protocol had been outlined in detail. Patients with an indication for xenograft AVR (age older than 65 years or special request or contraindication for anticoagulation) were included.



View larger version (101K):
[in this window]
[in a new window]
 
Stented aortic xenografts implanted in this prospectively randomized trial. A, The MosaicTM porcine xenograft (Medtronic Inc). B, The Perimount pericardial xenograft (Edwards).

The 2 devices provide the following features: Mosaic (M) xenograft is a porcine valve mounted on a thin flexible stent. The stent had been originally designed for the Hancock II valve and implies a reduced profile height of 13.5 mm. The xenogenic tissue is processed under physiological pressure fixation, allowing the leaflets to float freely at zero pressure differential using diluted glutaraldehyde. After final rinsing, amino-oleic acid, a surfactant, is applied as additional anticalcification treatment. Hemodynamic data available from the company indicate an effective orifice area of 1.22 cm2 for the 21-mm, 1.38 cm2 for the 23-mm, 1.65 cm2 for the 25-mm, and 1.8 cm2 for the 27-mm prosthesis, respectively.17

The Perimount (P) xenograft is manufactured from bovine pericardium. A neutralogic stress-free fixation process in which the tissue is fixed in a bath of glutaraldehyde solution with no applied pressure is used. The stent profile height is 15 mm. The proprietary XenoLogiX treatment consisting of a cumulative, multistep process that begins after glutaraldehyde fixation and implies exposure to a combination of agents, including ethanol and Tween-80, is used as anticalcification treatment. Hemodynamic data available from the company indicate an effective orifice area of 1.3 cm2 for the 21-mm, 1.5 cm2 for the 23-mm, 1.8 cm2 for the 25-mm, and 1.8 cm2 for the 27-mm prosthesis, respectively.17 A significant reduction of tissue calcification has been shown on animal experiments for the amino-oleic acid and XenoLogiX treatments.18,19

Surgery was performed by experienced surgeons only via standard median sternotomy. Extracorporeal circulation was initiated by right atrial and ascending aorta cannulation. A left atrial vent was inserted via the right upper pulmonary vein. Antegrade cold crystalloid (Bretschneider HTK solution, Köhler Chemie) or warm blood cardioplegia according to surgeon personal preference and moderate hypothermia were applied. After transverse aortotomy, the diseased aortic valve was completely excised, including removal of calcification from the aortic annulus and eventually the anterior mitral leaflet. Then the stratified randomization process was applied: The aortic annulus diameter was carefully measured using metric sizers. Then the patient was assigned to the individual annulus diameter group according to the metric annulus diameter. The type of prosthesis to be implanted was then selected from a by-chance randomization list. Then the prosthesis related sizer was used for valve size selection. Valve implantation was performed using standard horizontal mattress sutures supported by Teflon felt. This resulted in a slightly supra annular position of the valve.

Hemodynamic evaluation was solely performed according to the intraoperatively measured annulus diameter groups. Follow-up was performed at the hospitals outpatient clinic. Mean follow-up is 438±352 days (range, 180 to 1133), 77 patients were seen in-hospital. Total follow-up consists of 124 patient-years. Transthoracic echocardiographic examinations were performed preoperatively, before discharge, and at follow-up. Multiplane transesophageal echocardiography was used intraoperatively or whenever additional information besides transthoracic echocardiographic examinations measurements was required. Cardiac morphology and function as well as valve hemodynamics were assessed using standard measurements. Echocardiograms were performed according to standard guidelines and were interpreted by a single physician. Maximum pressure gradients were calculated according to the complete Bernoulli equation and left ventricular mass index according to the Devereux formula.20,21 All measurements were performed according to the Penn convention.

Valve-related morbidity and mortality were evaluated according to standard guidelines.22 Absolute and relative frequencies were calculated. Results are given as mean±SD. After assessing for normal distribution, the Student t test for matched pairs was applied. P<0.05 was considered significant.


*    Results
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
*Results
down arrowDiscussion
down arrowReferences
 
From the 100 patients, 51 were female and mean age was 72.8±4.7 years. Predominant aortic valve disease was stenosis in 96 and severe incompetence in 4 patients, respectively. Preoperative New York Heart Association (NYHA) functional class was 2.5±0.6 and body surface area 1.8±0.2 m2. Preoperatively, 83 patients were in sinus rhythm, 15 were in atrial fibrillation, and 2 had permanent pacemaker. The additive Euroscore was 5±1.9, and there were no significant differences between the annulus size groups. Preoperative hemodynamic measurements at cardiac catheterization were as follows: left ventricular ejection fraction 61±14%; left ventricular end-diastolic pressure 23±13 mmHg, mixed venous oxygen saturation 66±7%, and cardiac index 2.2±0.5 l/min per m2.

Intraoperatively cross-clamp duration was 57±14 minutes, with no relevant differences between the prostheses. All patients had uneventful primary valve implantation. Additional procedures were as follows: myocardial revascularization in 18, replacement of the ascending aorta in 2, mitral valve repair in 1, and left atrial ablation therapy in 2 patients, respectively. Repeat exploration for bleeding had to be performed in 2 patients.

Five patients died in hospital because of sudden cardiac death on postoperative day (POD) 7, multiple organ failure on POD 11, sepsis on POD 15, stroke on POD 30, and pneumonia on POD 45. Patients had received 21-mm (3), 23-mm (1), and 25-mm (1) valves, respectively. Effective orifice area divided by body surface area was 0.74 (3), 0.78 (1), and 0.93 (1) cm2/m2 in these 5 patients, with no significant difference to the other 95 patients.

During follow-up, 3 patients died because of lung cancer (2) after 9 and 24 months, and because of heart failure after 30 months, respectively. Effective orifice area divided by body surface area was 0.68 (1), 0.96 (1), and 0.8 (1) cm2/m2 in these 3 patients.

Mean aortic annulus diameter was 23.9±1.7 mm and mean implanted aortic valve prosthesis size (as labeled) was 23.0±1.8 mm. A total of 12 patients had a 21-mm annulus (5 of those received M and 7 received P prosthesis); 40 patients had a 23-mm annulus (M=20 and P=20); 37 had a 25-mm annulus (M=18 and P=19), and 11 patients had a 27-mm annulus (M=4 and P=7), respectively.

Within those annulus size groups of 21, 23, 25, and 27 mm, patients received differently labeled valve sizes. Details are given in Table 1. However, there were no significant differences in labeled valve sizes in relation to annulus diameters between the 2 (M versus P) prostheses. Results on indicated effective valve orifice area divided by body surface area and on the patient annulus index are given in Table 2.


View this table:
[in this window]
[in a new window]
 
TABLE 1. Implanted Labeled Aortic Valve Xenograft Size in Relation to the True Annulus Diameter Measured From 21 to 27 mm


View this table:
[in this window]
[in a new window]
 
TABLE 2. Effective Valve Orifice Area (cm2) Divided by Body Surface Area (m2) and Patient Annulus Index (Annulus Diameter Divided by Body Surface Area) According to Different Annulus Size Groups Given for the 2 Stented Xenografts

Further evaluations are purely performed between the different annulus diameter groups. Because of the fact that a small number of patients had a 21-mm or a 27-mm annulus, evaluation will furthermore concentrate on the 23- and 25-mm annulus groups with 20 versus 20 and 18 versus 19 implants, respectively.

Circulatory function was within normal limits in all patients at postoperative baseline, as well as at follow-up echocardiographic examinations. Details on heart rate, left ventricular ejection fraction, and cardiac index are given in Table 3.


View this table:
[in this window]
[in a new window]
 
TABLE 3. Circulatory Function Given for the 23-mm and the 25-mm Annulus Groups for Mosaic (M) vs. Perimount (P) Aortic Valve Implants

Hemodynamic function at baseline as well as at follow-up echocardiographic measurements is given in Table 4. Maximum transvalvular blood flow velocity was within normal limits for stented xenografts in all patients. However, blood flow velocity was lower in all patients after P AVR. There was a significantly lower blood flow velocity in the 25 P group at baseline as well as in the 23 P group at follow-up measurements. At follow-up, differences in the 25-mm annulus groups were close to statistical significance (P=0.06). Consecutively, maximum transvalvular pressure gradients were lower after P AVR in those groups. Differences reached statistical significance in the 25-mm annulus group at baseline and in the 23- and 25-mm annulus groups at follow-up, respectively.


View this table:
[in this window]
[in a new window]
 
TABLE 4. Hemodynamic Function Given for the 23-mm and the 25-mm Annulus Groups for Mosaic (M) vs. Perimount (P) Aortic Valve Implants

Left ventricular internal diameters and left ventricular mass index are given in Table 5. There was no significant change in left ventricular diameters and a significant regression of left ventricular mass index in all groups between baseline and follow-up measurements. However, regression was more pronounced after P AVR.


View this table:
[in this window]
[in a new window]
 
TABLE 5. Left Ventricular End-Diastolic Internal Diameter (LVDd) and Left Ventricular Mass Index (LVMI) Given for the 23-mm and the 25-mm annulus Mosaic (M) vs Perimount (P) Groups at Baseline and at Follow-up Measurement


*    Discussion
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
up arrowResults
*Discussion
down arrowReferences
 
Aortic valve disease is the most common acquired heart valve disease in Western civilizations. AVR is the standard therapy leading to a persistent improvement in hemodynamic function as well as regression of left ventricular hypertrophy. The implantation of stented xenografts can be considered the standard approach from age 65 years onward.

Both stented xenografts compared in this study have been evaluated thoroughly and have a reliable short-term and long-term clinical function.17,23–27 Furthermore, good postoperative hemodynamic function has been shown. However, as yet, no prospectively randomized comparison of hemodynamic function has been published.

There are several important findings. This is the first prospective study comparing different standard stented xenografts using intraoperative metric sizing and a stratified randomization protocol. Intraoperative stratified randomization according to true aortic annulus diameter measured after complete decalcification is the only measure allowing for objective comparison between different aortic valve implants. Thus it should become a standard approach for future comparative clinical trials.

From the results it became obvious that there are discrepancies between the true metric aortic annulus diameter and the labeled diameter of the prosthetic valve finally implanted. Differences in metric versus product-related sizer dimensions have been described previously without any subsequent adjustments to clinical practice.11–16 Now, for the first time to our knowledge, a prospective comparison of true metric aortic annulus diameters to labeled aortic valve diameters becomes available. Not surprisingly there were no significant differences between the 2 xenograft valve types regarding labeled valve diameters implanted in the different aortic annulus diameter groups (Table 1). Neither the superstructure nor the size of the coronary sinuses dictated any downsizing of the prostheses in relation to the measured sizes of the prostheses and the intra-annular diameter.

Results in Table 2 indicate that patients with 21- and 23-mm annuli are at risk for moderate patient prosthesis mismatch. The finding of moderate patient prosthesis mismatch has been associated with slightly increased mortality of 2.1-fold in a larger patient population.17 The more frequent occurrence of moderate patient prosthesis mismatch in the smaller size annulus groups may be an explanation for the increased mortality observed throughout this study. However, only 1 patient died primarily because of a cardiac cause. A more frequent use of stentless aortic valves would probably be a useful strategy to avoid or at least decrease the incidence of moderate patient prosthesis mismatch.

Circulatory function was comparable between groups and within normal limits at postoperative baseline as well as at follow-up evaluations. Predischarge baseline echocardiographic examinations should be the common standard for the evaluation of valve implants in the short-term and the long-term.

Hemodynamic function was assessed by comparing the standard measure of maximum transvalvular blood flow velocity. This allowed for further calculation of transvalvular pressure gradients given in Table 4. The observed results indicate the hemodynamic advantage of the P aortic xenograft versus the M valve.

Postoperative left ventricular mass regression can be regarded as one of the most important parameters to define a normally functioning prosthetic aortic valve. As such, significant left ventricular mass regression was observed in all patients, as shown in Table 5. This indicates that all xenografts provided sufficient hemodynamic function for myocardial reverse remodeling and thus functional recovery in the long-term. However, differences were observed between the 2 prostheses, yielding a more pronounced left ventricular mass regression after P xenograft implantation.

Are there any explanations for the differences in hemodynamic function between the 2 xenografts examined? The most striking difference of the 2 xenografts, their construction from porcine valves in comparison to bovine pericardium, may be 1 factor, but usually both should provide comparable hemodynamic features. Differences definitively exist in the stent design. The M prosthesis is mounted on a stent originating from the Hancock II valve. The profile inherent to this stent may cause the comparably worse hemodynamic outcome.

All operations were performed using a similar surgical technique. Most importantly, horizontal mattress sutures were used with Teflon felts below the annulus, bringing all xenografts into a slightly supra-anular position. Thus no relevant hemodynamic differences should arise from the implantation technique.

In the future, use of supra annular prostheses may lead to further improvement in hemodynamic function after AVR. Upsizing of the prosthesis can be safely performed in most patients, leading to a larger effective orifice area. Further studies on this interesting aspect are necessary.

There are some limitations of the present study that need to be addressed. Sufficient numbers of implants to perform reliable statistical comparison were available in the 23- and 25-mm annulus groups only. However, the study results presented are based on sufficient data in those 2 groups. In-hospital follow-up was only complete in 77 of 92 patients. To achieve comparable results, we decided to include in-hospital echocardiographic follow-up for comparative analyses. The data presented were clearly sufficient to answer the most important question of this study and to define whether there was a hemodynamically different outcome between the different prostheses. Follow-up included >12 months on average, thus covering the period of most dramatic regression of left ventricular mass after AVR.

In summary, both xenografts provide acceptable hemodynamic function with significant left ventricular mass regression in all patients at follow-up. However, the hemodynamic profile was significantly better in the most frequently implanted aortic annulus diameter groups for the P aortic valve.


*    Footnotes
 
Presented at the 76th Scientific Sessions of the American Heart Association at Orlando, November 9–12, 2003.


*    References
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
up arrowResults
up arrowDiscussion
*References
 
1. Bonow RO, Carabello BA, deLeon AC, et al. Guidelines for the management of patients with valvular heart disease. Circulation. 1998; 98: 1949–1984.[Free Full Text]

2. Davis EA, Greene PS, Cameron DE, et al. Bioprosthetic versus mechanical prostheses for aortic valve replacement in the elderly. Circulation. 1996; 94: S121–S125.

3. Jones EL, Weintraub WS, Craver JM, et al. Interaction of age and coronary disease after valve replacement: implications for valve selection. Ann Thorac Surg. 1994; 58: 378–384.[Abstract]

4. Burr LH, Jamieson WR, Munro AI, et al. Porcine bioprostheses in the elderly: clinical performance by age groups and valve positions. Ann Thorac Surg. 1995; 60: S264–9.

5. Fann JI, Miller DC, Moore KA, et al. Twenty year clinical experience with porcine bioprostheses. Ann Thorac Surg. 1996; 62: 1301–1311.[Abstract/Free Full Text]

6. Jamieson WR, Burr LH, Munro AI, et al. Carpentier-Edwards standard porcine bioprosthesis: a 21-year experience. Ann Thorac Surg. 1998; 66: S40–S43.

7. Jamieson WR. Modern cardiac valve devices-bioprostheses and mechanical prostheses: State of the art. J Card Surg. 1993; 8: 89–98.[Medline] [Order article via Infotrieve]

8. Fann JI, Burdon TA. Are the indications for tissue valves different in 2001 and how do we communicate these changes to our cardiology colleagues? Curr Opin Cardiol. 2001; 16: 126–135.[CrossRef][Medline] [Order article via Infotrieve]

9. Walther T, Falk V, Langebartels G, et al. Prospectively randomized evaluation of stentless versus conventional biological aortic valves: Impact on early regression of left ventricular hypertrophy. Circulation. 1999; 100: II-6–II-10.

10. Autschbach R, Walther T, Falk V, et al. Prospectively randomized comparison of different mechanical aortic valves. Circulation. 2000; 102: III-1–III-4.

11. Bonchek LI, Burlingame MW, Vazales BE. Accuracy of sizers for aortic valve prostheses. J Thorac Cardiovasc Surg. 1987; 94: 632–634.[Abstract]

12. Cochran RP, Kunzelman KS. Discrepancies between labeled and actual dimensions of prosthetic valves and sizers. J Card Surg. 1996; 11: 318–324.[Medline] [Order article via Infotrieve]

13. Christakis GT, Buth KJ, Goldman BS, et al. Inaccurate and misleading valve sizing: a proposed standard for valve size nomenclature. Ann Thorac Surg. 1998; 66: 1198–1203.[Abstract/Free Full Text]

14. Bartels C, Leyh RG, Bechtel MJF, et al. Discrepancies between sizer and valve dimensions: implications for small aortic root. Ann Thorac Surg. 1998; 65: 1631–1633.[Abstract/Free Full Text]

15. Zuckerman BD, Sapirstein W. Prosthetic cardiac valve replacement. J Thorac Cardiovasc Surg. 2003; 125: 1563.[Free Full Text]

16. Walther T, Falk V, Weigl C, et al. Discrepancy of sizers for conventional and stentless aortic valve implants. J Heart Valve Dis. 1997; 6: 145–148.[Medline] [Order article via Infotrieve]

17. Blais C, Dumesnil JG, Baillot R, et al. Impact of prosthesis-patient mismatch on short-term mortality after aortic valve replacement. Circulation. 2003; 108: 983–988.[Abstract/Free Full Text]

18. Quintero LJ, Lohre JM, Hernandez N, et al. Evaluation of in vivo models for studying calcification behavior of commercially available bovine pericardium. J Heart Valve Dis. 1998; 7: 262–267.[Medline] [Order article via Infotrieve]

19. Walther T, Falk V, Autschbach R, et al. Comparison of different anticalcification treatments for stentless aortic valves. Ann Thorac Surg. 1998; 66: S249–54.

20. Fehske W. Praxis der konventionellen und farbkodierten Doppler-Echokardiographie, 1st ed. Toronto, Canada: Huber; 1988.

21. Devereux RB, Reichek N. Echocardiographic determination of left ventricular mass in man: anatomic validation of the method. Circulation. 1977; 55: 613–618.[Abstract/Free Full Text]

22. Edmunds LH, Clark RE, Cohn LH, et al. Guidelines for reporting morbidity and mortality after cardiac valvular operations. Ann Thorac Surg. 1996; 62: 932–935.[Abstract/Free Full Text]

23. Jamieson WRE, Janusz MT, MacNab J, et al. Hemodynamic comparison of second- and third- generation stented bioprostheses in aortic valve replacement. Ann Thorac Surg. 2001; 71: S282–S284.[Abstract/Free Full Text]

24. Dellgren G, David TE, Raanani E, et al. Late hemodynamic and clinical outcomes of aortic valve replacement with the Carpentier-Edwards Perimount pericardial bioprosthesis. J Thorac Cardiovasc Surg. 2002; 124: 146–154.[Abstract/Free Full Text]

25. Banbury MK, Cosgrove DM 3rd, Thomas JD et al. Hemodynamic stability during 17 years of the Carpentier-Edwards aortic pericardial bioprosthesis. Ann Thorac Surg. 2002; 73: 1460–1465.[Abstract/Free Full Text]

26. Thomson DJ, Jamieson WRE, Dumesnil JG, et al. Medtronic Mosaic porcine bioprosthesis: Midterm investigational trial results. Ann Thorac Surg. 2001; 71: S269–72.[Abstract/Free Full Text]

27. Gansera B, Botzenhardt F, Gunzinger R, et al. The Mosaic bioprosthesis in the aortic position: seven years’ results. J Heart Valve Dis. 2003; 12: 354–361.[Medline] [Order article via Infotrieve]





This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Walther, T.
Right arrow Articles by Mohr, F. W.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Walther, T.
Right arrow Articles by Mohr, F. W.
Related Collections
Right arrow CV surgery: valvular disease