(Circulation. 2004;109:487-493.)
© 2004 American Heart Association, Inc.
Clinical Investigation and Reports |
From the Department of Cardiology, University Hospitals of Leicester (A.G., A.S.-L.), Glenfield Hospital, Leicester, UK; Department of Interventional Cardiology, University Hospitals Leuven (I.D.S.), Leuven, Belgium; Department of Interventional Cardiology, Centre Cardiologique du Nord (B.C.), Saint-Denis, France; University Hospital of Geneva (E.C.), Geneva, Switzerland; University Hospital Antwerp (C.V.), Edegem, Belgium; Department of Kardiologisches Institute of Bad Soden (N.R.), Bad Soden, Germany; A Z St Jan Hospital (L.M.), Brugges, Belgium; Clinique Cecil (J.-J.G.), Lausanne, Switzerland; The Johns Hopkins Medical Institutions (J.A.B., A.W.H.), Baltimore, Md; Methodist De Bakey Heart Center and Baylor College of Medicine (A.E.R.), Houston, Tex; and Cardiovascular Department (P.U.), La Tour Hospital, Geneva, Switzerland.
Correspondence to Anthony Gershlick, Department of Cardiology, University Hospitals of Leicester, Glenfield Hospital, Leicester, UK LE3 9QP.
Received September 26, 2002; de novo received August 11, 2003; revision received October 13, 2003; accepted October 24, 2003.
| Abstract |
|---|
|
|
|---|
Methods and Results On the basis of prior animal studies, the European evaLUation of the pacliTaxel Eluting Stent (ELUTES) pilot clinical trial (n=190) investigated the safety and efficacy of V-Flex Plus coronary stents (Cook Inc) coated with escalating doses of paclitaxel (0.2, 0.7, 1.4, and 2.7 µg/mm2 stent surface area) applied directly to the abluminal surface of the stent in de novo lesions compared with bare stent alone. The primary efficacy end point was angiographic percent diameter stenosis at 6 months. At angiographic follow-up, percent diameter stenosis was 33.9±26.7% in controls (n=34) and 14.2±16.6% in the 2.7-µg/mm2 group (n=31; P=0.006). Late loss decreased from 0.73±0.73 to 0.11±0.50 mm (P=0.002). Binary restenosis (
50% at follow-up) decreased from 20.6% to 3.2% (P=0.056), with no significant benefit from intermediate paclitaxel doses. Freedom from major adverse cardiac events in the highest (effective) dose group was 92%, 89%, and 86% at 1, 6, and 12 months, respectively (P=NS versus control). No late stent thromboses were seen in any treated group despite clopidogrel treatment for 3 months only.
Conclusions Paclitaxel applied directly to the abluminal surface of a bare metal coronary stent, at a dose density of 2.7 µg/mm2, reduced angiographic indicators of in-stent restenosis without short- or medium-term side effects.
Key Words: stents restenosis paclitaxel
| Introduction |
|---|
|
|
|---|
Prevention of recurrence may be possible by delivery of a drug via the stent to inhibit the proliferative process locally, thereby reducing systemic side effects. Studies in vitro and in animal models demonstrated that paclitaxel, which enhances polymerization of cellular microtubules and thus prevents mitosis and cell migration, reduced restenosis when delivered locally.57 Here, we report the European evaLUation of pacliTaxel Eluting Stent (ELUTES) trial, a multicenter, randomized, controlled, and triple-blinded clinical study designed to evaluate the efficacy (and safety) of a paclitaxel-eluting stent across a range of paclitaxel dose densities. The null hypothesis was that deployment of either bare metal stents or stents loaded with 4 doses of paclitaxel results in no difference in percent diameter stenosis at 6 months either between these doses or compared with controls.
| Methods |
|---|
|
|
|---|
Angioplasty and Stent Procedure
Angioplasty and stent placement were performed by the femoral approach and a monorail technique. After mandated predilation, the premounted stent was delivered at low-pressure inflation (maximum 8 atm) and postdilated if necessary to achieve a visual minimal residual stenosis <10%. If a second stent was required (eg, for an edge tear), an uncoated stent was used.
Antiplatelet Therapy
Adjunct medications included aspirin and procedural heparin according to operator practice. Antiplatelet therapy with aspirin plus clopidogrel was administered for 3 months. Glycoprotein IIb/IIIa receptor blocker use was operator dependent (administered in 58 of the 190 patients, with a uniform intergroup distribution).
Study End Points
The primary study end point was in-stent percent diameter stenosis (%DS) at 6 months, but late loss and binary in-stent restenosis rate were also reported. Secondary end points at 1, 6, and 12 months included hierarchically determined death, Q-wave MI, subacute stent thrombosis (>24 hours postprocedure), nonQ-wave MI, or CABG or PTCA required to treat symptomatic target-lesion restenosis.
Angiographic Analysis
Intracoronary nitrate was administered before image acquisition. Careful note was taken of procedural projection and table height to ensure identical follow-up image acquisition. Investigators reported lesion morphology, side-branch involvement, dissections, and TIMI (Thrombolysis In Myocardial Infarction) flow. A blinded independent angiographic core laboratory used the CAAS II system for standard quantitative coronary angiography analysis of periprocedural and follow-up angiograms. Analysis included measurement to 5 mm beyond the stent edges to account for any so-called edge effect.
Study Power and Statistical Analysis
To show reduction in 6-month angiographic %DS from 35% to 20% (with P<0.05 and 80% power), a minimum of 32 patients per group were required, with any 1 dose compared with control. Statistical analyses were prespecified and performed with Systat version 10 (SPSS). Continuous variables (%DS, minimal luminal diameter, and late loss) are reported as mean and SD. Standard ANOVA was used to analyze differences in continuous variables across the 5 groups. Multiple pairwise comparisons were performed between the groups with the Bonferroni correction to adjust for type I error. Dichotomous variables were reported in percent, and group-wise comparisons were performed with a Pearsons
2 test or Fisher exact test. All analysis was performed on an intention-to-treat basis that used all patients with available follow-up data.
| Results |
|---|
|
|
|---|
|
|
Clinical and Angiographic Follow-Up at 6 Months
The mean time for 6-month angiographic follow-up was 187±18 (range 113 to 259) days. At 6 months follow-up, 1 patient had died (2.7 µg/mm2 group), and 14 patients had refused angiographic follow-up, with relatively even distribution among groups (control, 3; 0.2 µg/mm2, 3; 0.7 µg/mm2, 4; 1.4 µg/mm2, 2; and 2.7 µg/mm2, 2; Figure 1).
|
Angiographic Outcome
Six-month quantitative coronary angiography demonstrated benefit in the highest-dose-density group. %DS was 14.2±16.6% in the 2.7 µg/mm2 group compared with 33.9±26.7% in controls (P=0.006), and late loss was 0.11±0.50 versus 0.73±0.73 mm (P=0.002; Table 2). The cumulative distribution for minimal luminal diameter for the 2.7 µg/mm2 dose is shown in Figure 2. Binary in-stent restenosis (diameter stenosis
50% at 6-month follow-up) was reduced from 20.6% in controls to 3.2% in the 2.7 µg/mm2 group (P=0.056). Although there were no significant differences between intermediate groups for the primary end point (%DS), multiple pairwise comparisons with Bonferroni correction demonstrated that with respect to minimum lumen diameter, the 2.7 µg/mm2 group was significantly different from the 0.2 µg/mm2 group (P=0.032) and from the 0 µg/mm2 (control) group (P=0.014). Additionally, the late loss index in both the 2.7 and 1.4 µg/mm2 treatment groups was significantly less than both the control group and the 0.2 µg/mm2 group (2.7 µg/mm2 versus control, P=0.022; 2.7 µg/mm2 versus 0.2 µg/mm2, P=0.032; 1.4 µg/mm2 versus control, P=0.016). A trend to benefit with escalating doses was observed (Table 2; Figure 3).
|
|
Safety and Efficacy Data
There were no 30-day major adverse coronary events (MACE) in intermediate-dose groups. In the 2.7 µg/mm2 group, there were 3 events: a nonQ-wave infarct (day 0), a stent thrombosis (day 8), and a death at day 10. There was 1 subacute stent thrombosis in the control group. Between 1 and 6 months, there were no additional deaths, bypass grafting, or Q-wave myocardial infarctions. One additional nonQ-wave myocardial infarction occurred in the 0.7 µg/mm2 dose group. There were 4 symptom-driven target-lesion revascularizations (TLRs) in the control group, 1 in the 0.2 µg/mm2 dose treatment group, 3 in the 1.4 µg/mm2 group, and 1 in the 2.7 µg/mm2 group who did not have binary in-stent restenosis but required the reintervention at 183 days for symptomatic persistent restenosis. The 1 patient in the 2.7 µg/mm2 dose treatment group who did have angiographic binary restenosis did not require reintervention. There were no further stent thromboses at 6 months.
One-Year Clinical Follow-Up
Between 6 and 12 months, there were no additional deaths, stent thromboses, or myocardial infarcts in any group. Bypass surgery was performed in 1 patient in each of the control and 0.7 µg/mm2 dose groups. One additional target lesion revascularization (PCI) was undertaken for persistent restenosis in the 2.7 µg/mm2 dose group and 1 for in-stent restenosis in the controls. Thus, at 12 months, there were 6 TLRs (16%) in controls and 2 (5%) in the 2.7 µg/mm2 group. Freedom from MACE was 82% in controls versus 86% in the highest-dose group (P=NS; Table 3). Despite the apparent higher event rate at lower concentrations of paclitaxel, all pairwise comparisons were nonsignificant.
|
| Discussion |
|---|
|
|
|---|
Paclitaxel is currently used primarily as systemic chemotherapy.15 It promotes stability of microtubules, and cell replication is inhibited in the G0/G1 and G2/M phases. Cell motility, shape, and transport between organelles may also be affected.
Preclinical data suggest paclitaxel has beneficial effects on the vessel response to injury. In animal models, late loss and intimal area are significantly reduced,6,7,1618 but there is also an unwanted dose-dependent inhibition of reendothelialization,19 and additional adverse effects have been shown at high doses.7 In 1 study, stents coated with a cross-linked biodegradable polymer and loaded with different doses of paclitaxel reduced mean intimal thickness by 49% and 36% at doses of 42.0 and 20.2 µg per stent at 28 days but were associated with persistent intimal fibrin deposition, intraintimal hemorrhage, and increased intimal and adventitial inflammation.6 We believed, therefore, that a dose-finding (efficacy and safety) human pilot study of paclitaxel-eluting stents was warranted, especially because available animal data failed to specify which dose may be beneficial. Further preclinical considerations included uncertainty about the impact of "release rates" in humans, not least because this trial involved the use of pure drug on the stent (ie, no polymer to modulate the release of the drug). As such, there was no release rate per se. Because the paclitaxel is in direct contact with the tissue, uptake is determined by the rate at which the drug crosses the cell membrane. In-house animal data showed that the amount of drug remaining on the stent was 72±7% at 4 hours, 53±10% at 4 days, and 31±13% at 14 days. Preclinical data indicated that immediate, 1-week, and 1 month estimation of systemic blood paclitaxel was less than the sensitivity of the assay (20 to 40 ng/mL) even at the highest total dose (60 µg).
The ELUTES trial was therefore primarily a dose-evaluation study. Although it was unclear from animal models which dose would be beneficial in humans, we wished to stay below the maximum doses tested by the National Institutes of Health group7 that caused fibrin deposition and hemorrhage in the vessel wall. One group loaded paclitaxel by dipping the stent in ethanol/paclitaxel and then air-drying it, resulting in crystalline paclitaxel being deposited in uneven clumps.7 Furthermore, the Palmaz-Schatz stent used in their studies was structurally different from ours. Not withstanding these differences, compared with the stent we planned to use in ELUTES, their top dose (total dose 187 µg) was considered potentially "toxic," and therefore the doses chosen for ELUTES were set at levels well below this, with a maximum total dose of 60 µg (2.7 µg/mm2). It might be argued that an abluminal dose of 60 µg is equivalent to 120 µg on both sides of the stent, and this is thus close to the dose loaded by Heldmans group. However, even taking account of the lipophilic nature of the drug, which encourages tissue deposition, we believed that a 60-µg abluminal dose was still well within the 90 µg per stent side dose used by Heldman.
Given the potential toxicity of paclitaxel, it is encouraging that ELUTES demonstrated clinical safety, with reendothelialization appearing to have occurred, illustrated by the lack of late subacute thrombosis despite only 3 months of clopidogrel. Although overall MACE rates were no different between the highest-dose group and controls and were very low in all groups, the event rates in the highest-dose group require comment. These all occurred within 10 days and were therefore unlikely to be due to any negative effect of paclitaxel on reendothelialization, which would not have been completed in the control group either at this time point. This is supported to some extent by the nature of the events. The nonQ-wave infarct occurred on day 1 after a procedural spiral dissection. The second patient had a trial stent placed in the distal right coronary artery and required a nontrial (8-mm) stent for a proximal edge tear. At day 8, the patient presented with chest pain, and angiography demonstrated total occlusion of the artery proximal to the nontrial stent. The third patient event was a protocol violation. Having sustained an acute myocardial infarct and received an investigational thrombolytic agent as part of another trial, the patient was subsequently inappropriately randomized to ELUTES, receiving a trial stent to a noninfarct-related artery. The patient died suddenly at day 10 in a nontrial hospital, and no postmortem examination was undertaken. We believe this was an arrhythmic death, but it could have been due to a subacute stent thrombosis. Despite the nature of the events, it is possible that they were caused by early paclitaxel toxicity on the vessel wall (independent of its effects on later reendothelialization), and it is for this reason that all events are included in the analysis.
The ELUTES trial was powered to show a significant benefit with 32 patients per group if the percent diameter stenosis was effectively reduced from 35% to 20%. The actual number of patients who reached angiographic assessment was 90% overall, although the number undergoing angiography in the highest-dose group was only 31; 2 patients refused, 2 reached early (<30 day) end points, and 2 were unevaluable (1 because of total occlusion of the vessel proximal to the study stent, reported at the time of procedure as being a suboptimal deployment, and the other because of worsening stenosis of an unstented lesion injured during the initial PCI). These angiographically unevaluable patients were included in the clinical end-point count; 1 had a subacute thrombosis and the other a TLR. The unevaluable control patient was not counted in the adverse report because this patient had no event. Therefore, for the highest-dose group, even with the patients with unevaluable angiograms at 6 months counted to allow for worst outcomes, there is no evidence that efficacy was associated with adverse MACE with this dose. Although not powered for clinical outcome, it is interesting that at 1 year, 2 clinically driven TLRs (PCI) were needed in the highest-dose group (5.4%), but 6 (5 PCIs and 1 CABG) were required in the control group (15.7%). The TLRs in the ELUTES highest-dose group were required for progression of disease outside and proximal to the stent (perhaps the consequence of using a 16-mm stent on a 20-mm balloon). All control-group TLRs were for diffuse restenosis within the stent itself.
The biological potency of the treatment was such that despite the planned angiographic numbers not being available in the highest-dose group, significant differences were reached, with a probability value of 0.006 (a 20% difference). Although angiographic loss to follow-up is always concerning, trial robustness in a blinded, randomized study such as ELUTES is supported by having independent safety and efficacy, data handling, and angiographic committees, and clinical follow-up was 100%. It is difficult to deal with missing angiographic quantitative data, but if all the unevaluable patients in the high-dose group were assumed to have developed binary restenosis (ie, diameter stenosis >50%) and all the unevaluable patients in the control group were assumed not to have been, then the unsubstantiated "worse case" for this trial would be 7 (18.9%) of 37 for the 2.7 µg/mm2 group and 11 (28.9%) of 38 for controls (compared with our reported 1/31 [3.2%] versus 7/34 [20.6%], respectively).
The decision not to undertake intravascular ultrasound can potentially be seen as a limitation. There was, however, no angiographic evidence of increase in minimum lumen diameter in contiguous reference-vessel segments, nor within the stent itself. Furthermore, in the Asian Paclitaxel-Eluting Stent Clinical Trial (ASPECT) trial,20 which used a similar maximal dose (3.1 µg/mm2), 6-month intravascular ultrasound studies showed only 1 malapposition, and intimal volume fell from 31 mm3 in the control group to 12 mm3 in the 3.1 µg/mm2 group.
Data from ELUTES showed that although the highest dose was effective, lower doses were not so, although a trend was seen for decreasing diameter stenosis and late loss with increasing intermediate doses. This suggests that dosing may be an important factor. The design of this study does not allow us to identify the therapeutic window of this platform in humans because the upper threshold has not been established. In the ASPECT20 trial, the overall dose delivered was 130 µg at the 3.1 µg/mm2 dose density compared with 60 µg at the 2.7 µg/mm2 dose density in ELUTES. The control and highest-dose diameter stenoses at follow-up were 33.9% and 14.1% (ELUTES) and 39% and 14% (ASPECT), showing similar biological responses despite different overall amounts of drug delivered (Figure 3) and indicating that dose density is more important than overall concentration or total dose.
The absence of polymer may or may not be an advantage. Although polymer may facilitate higher and more even drug delivery, the amount of drug delivered to the tissue may be governed more by the character of the tissue being targeted and the lipophilic nature of the agent. In this study, a "depot" of drug was sandwiched between stent and tissue and appeared to be effective on the basis of the various parameters of measured restenosis. Additionally, absence of polymer overcomes any concerns regarding its potential inflammatory or prothrombogenic effects.21
There have been concerns regarding paclitaxel inhibition of reendothelialization, with its associated excess thrombotic risk. In ELUTES, there was no significant difference in the MACE-free rate between groups, although the study was not powered to demonstrate clinical benefit. Importantly, despite no antiplatelet therapy between 3 and 6 months, no stent thromboses occurred, unlike the Study to COmpare REstenosis between QueST and QuaDS-QP2 (SCORE),22 which used a large dose of paclitaxel derivative (4000 µg) and was stopped because of excess stent thromboses.23
After completion of the ELUTES study, the RX Achieve Drug ELuting coronary stent systems In the treatment of patients with de noVo nativE coronaRy lesions (DELIVER I).24 trial was undertaken with the trial dose based on the ELUTES highest dose, and the results have been presented in abstract form. It compared a 3.0 µg/mm2 dose with controls in 1043 patients with target de novo lesions
25 mm. Although in-stent late loss was significantly less for the paclitaxel-loaded stent (0.81 versus 0.98 mm, P<0.003), this was clearly higher in the treated group than we have measured in ELUTES and was not low enough to drive a difference in angiographic restenosis (16.7% versus 22.4%, P=0.15). There were trends toward fewer TLRs (8.1% versus 11.3%) and repeat PCIs (7.0% versus 10.3%) in the paclitaxel group, but target-vessel failure, the primary end point, was not significantly different (11.9% versus 14.5%, P=0.13). The reasons for the success of ELUTES and failure of DELIVER are still being explored, but it has been suggested by the Principle Investigator ONeill that it may relate to nonuniform use of glycoprotein IIb/IIIa inhibitors across groups and the presence of longer lesions.24 Additionally, the use of a different stent or a change in the facility for loading could have led to an inability to accurately load the stent with the effective (ELUTES) dose. It would appear from the late loss that not enough drug may have been loaded.
Paclitaxel, considering the ELUTES and Treatment of de novo coronary disease using a single pAclitaXel elUting Stent (TAXUS) trials and despite the DELIVER I results, is the second agent shown to benefit the vessel-wall response to stent trauma, the first being sirolimus in the RAVEL (RAndomized study with the sirolimus-eluting Bx VElocity balloon-expandable stent in the treatment of patients with de novo coronary Lesions) trial. 25 Recent studies have tested both sirolimus (SIRIUS; a multicenter randomized double-blind study of the SIRolImUS-coated Bx Velocity stent in the treatment of patients with de novo coronary lesions) and paclitaxel (TAXUS II) in slightly more complex lesions, and efficacy has been reported recently, extending the biological robustness of these agents. However, drug-eluting stents with these or other agents still need to be studied in patients with lesions at highest risk of restenosis, and longer-term uncertainty about the outcome for any of these agents justifies extended follow-up. Furthermore, the long-term importance of the presence of polymer once the drug has eluted has yet to be evaluated in trials that have used this technology.
| Conclusions |
|---|
|
|
|---|
| Appendix |
|---|
|
|
|---|
Clinical Events Committee
Harvard Clinical Research Institute, Harvard University, Boston, Mass. Vice President of Research, Richard Kuntz, MD.
Data Safety Monitoring Board
Chairman, Willis A. Tacker, Jr, MD, PhD, Professor of Basic Medical Sciences, Purdue University, West Lafayette, Ind.
| Footnotes |
|---|
Dr Heldman received research support from Angiotech Pharmaceuticals. Dr Raizner received research support from Cook, Inc.
| References |
|---|
|
|
|---|
2. Fischman DL, Leon MB, Baim DS, et al. A randomised comparison of coronary stent placement and balloon angioplasty in the treatment of coronary artery disease: Stent restenosis study investigators. N Engl J Med. 1994; 331: 496501.
3. Van Belle E, Bauters C, Hubert E, et al. Restenosis rates in diabetics: a comparison of stenting and balloon angioplasty in native coronary vessels. Circulation. 1997; 96: 14541460.
4. Serruys PW, Unger F, van Hout BA, et al. The ARTS study (Arterial Revascularization Therapies Study). Semin Interv Cardiol. 1999; 4: 209219.[Medline] [Order article via Infotrieve]
5. Herdeg C, Oberhoff M, Baumbach A, et al. Local paclitaxel delivery for the prevention of restenosis: biological effects and efficacy in vivo. J Am Coll Cardiol. 2000; 35: 19691976.
6. Farb A, Heller PF, Shroff S, et al. Pathological analysis of local delivery of paclitaxel via a polymer-coated stent. Circulation. 2001; 104: 473479.
7. Heldman AW, Cheng L, Jenkins GM, et al. Paclitaxel stent coating inhibits neointimal hyperplasia at 4 weeks in a porcine model of coronary restenosis. Circulation. 2001; 103: 22892295.
8. Sriram V, Patterson C. Cell cycle in vasculoproliferative diseases: potential interventions and routes of delivery. Circulation. 2001; 103: 24142419.
9. Boehm M, Nabel E. Cell cycle and cell migration: new pieces to the puzzle. Circulation. 2001; 103: 28792881.
10. Rozenman Y, Mereuta A, Schechter D, et al. Long-term outcome of patients with very long stents for treatment of diffuse coronary disease. Am Heart J. 1999; 138: 441445.[CrossRef][Medline] [Order article via Infotrieve]
11. Al-Suwaidi J, Yeh W, Cohen HA, et al. Immediate and one-year outcome in patients with coronary bifurcation lesions in the modern era (NHLBI dynamic registry). Am J Cardiol. 2001; 87: 11391144.[CrossRef][Medline] [Order article via Infotrieve]
12. Serruys PW, Hamburger JN, Koolen JJ, et al. Total occlusion trial with angioplasty by using laser guidewire: the TOTAL trial. Eur Heart J. 2000; 21: 17971805.
13. Serruys PW, Emanuelsson H, van der Giessen W, et al. Heparin-coated Palmaz-Schatz stents in human coronary arteries: early outcome of the Benestent-II pilot study. Circulation. 1996; 93: 412422.
14. Elezi S, Kastrati A, Neumann FJ, et al. Vessel size and long-term outcome after coronary stent placement. Circulation. 1998; 98: 18751880.
15. Riondel J, Jacrot M, Picot F. Therapeutic response to taxol of six human tumors xenografted into nude mice. Cancer Chemother Pharmacol. 1986; 17: 137142.[CrossRef][Medline] [Order article via Infotrieve]
16. Sollott SJ, Cheng L, Pauly RR. Taxol inhibits neointimal smooth muscle cell accumulation after angioplasty in the rat. J Clin Investig. 1995; 95: 18691876.[Medline] [Order article via Infotrieve]
17. Drachman DE, Edelman ER, Kamath KR. Sustained stent-based delivery of paclitaxel arrests neointimal thickening and cell proliferation. Circulation. 1998; 98 (suppl I): I-740.
18. Hong MK, Kornowski R, Bramwell O, et al. Paclitaxel-coated Gianturco-Roubin II (GR II) stents reduces neointimal hyperplasia in a porcine coronary in stent restenosis model. Coron Artery Dis. 2001; 12: 513515.[CrossRef][Medline] [Order article via Infotrieve]
19. Axel DI, Kunert W, Goggelmann C. Paclitaxel inhibits arterial smooth muscle cell proliferation and migration in vitro and in vivo using local drug delivery. Circulation. 1997; 96: 636645.
20. Hong MK, Mintz GS, Lee CW, et al. Paclitaxel coating reduces in-stent intimal hyperplasia in human coronary arteries: a serial volumetric intra-vascular ultrasound analysis from the Asian Paclitaxel-Eluting Stent Clinical Trial (ASPECT). Circulation. 2003; 107: 517520.
21. van der Giessen WJ, Lincoff AM. Marked inflammatory sequelae to implantation of biodegradable and nonbiodegradable polymers in porcine coronary arteries. Circulation. 1996; 94: 16901697.
22. de la Fuente LM, Miano J, Mrad J, et al. Initial results of the Quanum drug eluting stent (QuaDS-QP-2) registry (BARDS) in human subjects. Cathet Cardiovasc Intervent. 2001; 53: 480488.[CrossRef][Medline] [Order article via Infotrieve]
23. Liistro F, Colombo A. Late acute thrombosis after paclitaxel eluting stent implantation. Heart. 2001; 86: 262265.
24. ONeill WW. The DELIVER trial: a randomized comparison of paclitaxel coated versus metallic stents for treatment of coronary lesions. Paper presented at the 2003 meeting of the American College of Cardiology in Chicago Ill, March 30, 2003.
25. Morice MC, Serruys PJ, de Sousa JE. A randomized comparison of a sirolimus-eluting stent with a standard stent for coronary revascularization. N Engl J Med. 2002; 346: 17731780.
This article has been cited by other articles:
![]() |
H. S. Hecht, S. Polena, V. Jelnin, M. Jimenez, T. Bhatti, M. Parikh, G. Panagopoulos, and G. Roubin Stent gap by 64-detector computed tomographic angiography relationship to in-stent restenosis, fracture, and overlap failure. J. Am. Coll. Cardiol., November 17, 2009; 54(21): 1949 - 1959. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Higo, Y. Ueda, J. Oyabu, K. Okada, M. Nishio, A. Hirata, K. Kashiwase, N. Ogasawara, S. Hirotani, and K. Kodama Atherosclerotic and thrombogenic neointima formed over sirolimus drug-eluting stent an angioscopic study. J. Am. Coll. Cardiol. Img., May 1, 2009; 2(5): 616 - 624. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. C. Philpott MD, D. A. Southern MSc, F. M. Clement PhD, P. D. Galbraith BN MSc, M. Traboulsi MD, M. L. Knudtson MD, W. A. Ghali MD, and for the APPROACH Investigators Long-term outcomes of patients receiving drug-eluting stents Can. Med. Assoc. J., January 20, 2009; 180(2): 167 - 174. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. G Raja and G. D Dreyfus Current Status of Off-pump Coronary Artery Bypass Surgery Asian Cardiovasc Thorac Ann, April 1, 2008; 16(2): 164 - 178. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. E. Kip, K. Hollabaugh, O. C. Marroquin, and D. O. Williams The Problem With Composite End Points in Cardiovascular Studies The Story of Major Adverse Cardiac Events and Percutaneous Coronary Intervention. J. Am. Coll. Cardiol., February 19, 2008; 51(7): 701 - 707. [Abstract] [Full Text] [PDF] |
||||
![]() |
American College of Cardiology/American Heart Asso, 2007 Writing Group to Review New Evidence and Upda, S. B. King III, S. C. Smith Jr, J. W. Hirshfeld Jr, A. K. Jacobs, D. A. Morrison, and D. O. Williams 2007 Focused Update of the ACC/AHA/SCAI 2005 Guideline Update for Percutaneous Coronary Intervention J. Am. Coll. Cardiol., January 15, 2008; 51(2): 172 - 209. [Full Text] [PDF] |
||||
![]() |
S. B. King III, S. C. Smith Jr, J. W. Hirshfeld Jr, A. K. Jacobs, D. A. Morrison, D. O. Williams, 2005 WRITING COMMITTEE MEMBERS, S. C. Smith Jr, T. E. Feldman, J. W. Hirshfeld Jr, et al. 2007 Focused Update of the ACC/AHA/SCAI 2005 Guideline Update for Percutaneous Coronary Intervention: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines: 2007 Writing Group to Review New Evidence and Update the ACC/AHA/SCAI 2005 Guideline Update for Percutaneous Coronary Intervention, Writing on Behalf of the 2005 Writing Committee Circulation, January 15, 2008; 117(2): 261 - 295. [Full Text] [PDF] |
||||
![]() |
R. Wessely, A. Kastrati, J. Mehilli, A. Dibra, J. Pache, and A. Schomig Randomized trial of rapamycin- and paclitaxel-eluting stents with identical biodegradable polymeric coating and design Eur. Heart J., November 2, 2007; 28(22): 2720 - 2725. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Daemen and P. W. Serruys Drug-Eluting Stent Update 2007: Part I: A Survey of Current and Future Generation Drug-Eluting Stents: Meaningful Advances or More of the Same? Circulation, July 17, 2007; 116(3): 316 - 328. [Full Text] [PDF] |
||||
![]() |
Abstracts Heart, June 1, 2007; 93(suppl_1): A5 - A105. [Full Text] [PDF] |
||||
![]() |
C Roiron, P Sanchez, A Bouzamondo, P Lechat, and G Montalescot Drug eluting stents: an updated meta-analysis of randomised controlled trials Heart, May 1, 2006; 92(5): 641 - 649. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. G. Raja Drug-Eluting Stents and the Future of Coronary Artery Bypass Surgery: Facts and Fiction Ann. Thorac. Surg., March 1, 2006; 81(3): 1162 - 1171. [Abstract] [Full Text] [PDF] |
||||
![]() |
C.-H. Lee, H.-C. Tan, and Y.-T. Lim Update on Drug-Eluting Stents for Prevention of Restenosis Asian Cardiovasc Thorac Ann, February 1, 2006; 14(1): 75 - 82. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Fisher Results of the Management of Atherothrombosis With Clopidogrel in High-Risk Patients Trial: Implications for the Neurologist Arch Neurol, January 1, 2006; 63(1): 20 - 24. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. Alvarez Jr and N. K. Kapur Drug Eluting Stent Technology: A Paradigm Shift in the Treatment and Prevention of Restenosis Journal of Pharmacy Practice, December 1, 2005; 18(6): 461 - 478. [Abstract] [PDF] |
||||
![]() |
K D Dawkins, T Gershlick, M de Belder, A Chauhan, G Venn, P Schofield, D Smith, J Watkins, H H Gray, and Joint Working Group on Percutaneous Coronary Inter Percutaneous coronary intervention: recommendations for good practice and training Heart, December 1, 2005; 91(suppl_6): vi1 - vi27. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. D. Dawkins, E. Grube, G. Guagliumi, A. P. Banning, K. Zmudka, A. Colombo, L. Thuesen, K. Hauptman, J. Marco, W. Wijns, et al. Clinical Efficacy of Polymer-Based Paclitaxel-Eluting Stents in the Treatment of Complex, Long Coronary Artery Lesions From a Multicenter, Randomized Trial: Support for the Use of Drug-Eluting Stents in Contemporary Clinical Practice Circulation, November 22, 2005; 112(21): 3306 - 3313. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. J. Nawarskas and L. A. Osborn Paclitaxel-eluting stents in coronary artery disease Am. J. Health Syst. Pharm., November 1, 2005; 62(21): 2241 - 2251. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. F. Granada, D. Ensenat, A. N. Keswani, G. L. Kaluza, A. E. Raizner, X.-m. Liu, K. J. Peyton, M. A. Azam, H. Wang, and W. Durante Single Perivascular Delivery of Mitomycin C Stimulates p21 Expression and Inhibits Neointima Formation in Rat Arteries Arterioscler Thromb Vasc Biol, November 1, 2005; 25(11): 2343 - 2348. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Mauri, E. J. Orav, S. C. Candia, D. E. Cutlip, and R. E. Kuntz Robustness of Late Lumen Loss in Discriminating Drug-Eluting Stents Across Variable Observational and Randomized Trials Circulation, November 1, 2005; 112(18): 2833 - 2839. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Windecker, A. Remondino, F. R. Eberli, P. Juni, L. Raber, P. Wenaweser, M. Togni, M. Billinger, D. Tuller, C. Seiler, et al. Sirolimus-Eluting and Paclitaxel-Eluting Stents for Coronary Revascularization N. Engl. J. Med., August 18, 2005; 353(7): 653 - 662. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. W. Serruys, G. Sianos, A. Abizaid, J. Aoki, P. den Heijer, H. Bonnier, P. Smits, D. McClean, S. Verheye, J. Belardi, et al. The Effect of Variable Dose and Release Kinetics on Neointimal Hyperplasia Using a Novel Paclitaxel-Eluting Stent Platform: The Paclitaxel In-Stent Controlled Elution Study (PISCES) J. Am. Coll. Cardiol., July 19, 2005; 46(2): 253 - 260. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. C. Henry, M. M. Bonar, P. N. Kearns, H. Cui, M. M. Mutchler, M. V. Martin, A. R. Orsini, H. L. Elford, C. A. Bush, J. L. Zweier, et al. Inhibition of Ribonucleotide Reductase Reduces Neointimal Formation following Balloon Injury J. Pharmacol. Exp. Ther., July 1, 2005; 314(1): 70 - 76. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Mauri, E. J. Orav, and R. E. Kuntz Late Loss in Lumen Diameter and Binary Restenosis for Drug-Eluting Stent Comparison Circulation, June 28, 2005; 111(25): 3435 - 3442. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Balakrishnan, A. R. Tzafriri, P. Seifert, A. Groothuis, C. Rogers, and E. R. Edelman Strut Position, Blood Flow, and Drug Deposition: Implications for Single and Overlapping Drug-Eluting Stents Circulation, June 7, 2005; 111(22): 2958 - 2965. [Abstract] [Full Text] [PDF] |
||||
![]() |
A H Gershlick Drug eluting stents in 2005 Heart, June 1, 2005; 91(suppl_3): iii24 - iii31. [Full Text] [PDF] |
||||
![]() |
M. A. Costa and D. I. Simon Molecular Basis of Restenosis and Drug-Eluting Stents Circulation, May 3, 2005; 111(17): 2257 - 2273. [Full Text] [PDF] |
||||
![]() |
A. J. Lansky, R. A. Costa, M. Mooney, M. G. Midei, H. K. Lui, W. Strickland, R. Mehran, M. B. Leon, M. E. Russell, S. G. Ellis, et al. Gender-based outcomes after paclitaxel-eluting stent implantation in patients with coronary artery disease J. Am. Coll. Cardiol., April 19, 2005; 45(8): 1180 - 1185. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. W. O'Neill, S. R. Dixon, and C. L. Grines The year in interventional cardiology J. Am. Coll. Cardiol., April 5, 2005; 45(7): 1117 - 1134. [Full Text] [PDF] |
||||
![]() |
A. A. Bavry, D. J. Kumbhani, T. J. Helton, and D. L. Bhatt What is the risk of stent thrombosis associated with the use of paclitaxel-eluting stents for percutaneous coronary intervention?: A meta-analysis J. Am. Coll. Cardiol., March 15, 2005; 45(6): 941 - 946. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Moreno, C. Fernandez, R. Hernandez, F. Alfonso, D. J. Angiolillo, M. Sabate, J. Escaned, C. Banuelos, A. Fernandez-Ortiz, and C. Macaya Drug-eluting stent thrombosis: Results from a pooled analysis including 10 randomized studies J. Am. Coll. Cardiol., March 15, 2005; 45(6): 954 - 959. [Abstract] [Full Text] [PDF] |
||||
![]() |
J.-J. Goy, J.-C. Stauffer, M. Siegenthaler, A. Benoit, and C. Seydoux A prospective randomized comparison between paclitaxel and sirolimus stents in the real world of interventional cardiology: The TAXi trial J. Am. Coll. Cardiol., January 18, 2005; 45(2): 308 - 311. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y.-M. Deng, B. J. Wu, P. K. Witting, and R. Stocker Probucol Protects Against Smooth Muscle Cell Proliferation by Upregulating Heme Oxygenase-1 Circulation, September 28, 2004; 110(13): 1855 - 1860. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. J. Lansky, R. A. Costa, G. S. Mintz, Y. Tsuchiya, M. Midei, D. A. Cox, C. O'Shaughnessy, R. A. Applegate, L. A. Cannon, M. Mooney, et al. Non-Polymer-Based Paclitaxel-Coated Coronary Stents for the Treatment of Patients With De Novo Coronary Lesions: Angiographic Follow-Up of the DELIVER Clinical Trial Circulation, April 27, 2004; 109(16): 1948 - 1954. [Abstract] [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2004 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |