(Circulation. 2004;109:3068-3072.)
© 2004 American Heart Association, Inc.
Mini-Review: Expert Opinions |
From the Cardiology Section, Dartmouth Medical School, Dartmouth-Hitchcock Medical Center, Lebanon, NH (A.V.K., M.S.); Department of Medicine, Brigham and Womens Hospital, Boston, Mass (D.S.B., R.E.K.); Department of Radiology, Massachusetts General Hospital, Boston, Mass (J.J.S.); Devices Sector, Medical Devices Agency-United Kingdom (D.J.); Department of Surgery, Stanford University, Stanford, Calif (T.J.F.); and Lenox Hill Heart and Vascular Institute, New York, NY (M.B.L.). Dr Feigal is the former Director of the Center for Devices and Radiologic Health, Food and Drug Administration-USA.
Correspondence to Aaron V. Kaplan, MD, Section of Cardiology, Dartmouth-Hitchcock Medical Center, One Medical Center Dr, Lebanon, NH 03756-0001. E-mail aaron.v.kaplan{at}hitchcock.org
| Introduction |
|---|
|
|
|---|
| Invention |
|---|
|
|
|---|
A small percentage of device ideas are conceived in academic medical centers using federal or other grant funding. Few academic centers have the intrinsic capabilities to develop the device beyond the early prototype stage. Intellectual property is typically out-licensed to an existing company or start-up for further development. To facilitate the development of ideas that emerge out of federally funded programs, Congress passed the Bayh-Dole Act that assigns all intellectual property rights to the academic medical center.
| Initial Clinical Testing |
|---|
|
|
|---|
Clinical testing of an unapproved significant-risk medical device requires FDA approval in the form of an Investigational Device Exemption (IDE). The IDE application provides information to the FDA on device design and qualification, as well as on the study protocol. The FDA is mandated to respond to the IDE application within 30 days.1 An IDE may also be required for studies in which an approved device is used for a purpose distinct from its approved indication.2 This is typically the case when a trial is sponsored by a company for the purpose of expanding the indication of a device or making significant changes in the instructions for use.
| Regulatory Approval Process (United States) |
|---|
|
|
|---|
Class I devices (lowest risk) are subject to general controls, which are published standards pertaining to labeling, manufacturing, post-market surveillance, and reporting. Devices are placed into class I when there is reasonable assurance that general controls alone are adequate to assure safety and effectiveness. The general controls that typically apply to class I devices include prohibitions against adulteration and misbranding, requirements for establishing registration and device listing, adverse event reporting, and good manufacturing practices. Furthermore, remedies including seizure, injunction, criminal prosecution, civil penalties, and recall authority are provided to FDA. Formal FDA review is not required for most class I devices before their market introduction.3
Class II devices are those higher-risk devices for which general controls alone have been found to be insufficient to provide reasonable assurance of safety and effectiveness, but for which there is adequate information available to establish special controls. Special controls may include performance standards, design controls, and post-market surveillance programs. In addition, most class II devices require FDA clearance of a premarket notification application (PMA or 510[k]) before the device may be marketed. In the 510(k) application, the medical device manufacturer must provide data to demonstrate that the new device is "substantially equivalent" to a legally marketed device. Although substantial equivalence can usually be demonstrated on the basis of bench and animal testing alone, approximately 10% of 510(k) applications include clinical data.4
Class III devices, such as heart valves, pacemakers/implantable cardioverter-defibrillators, and coronary stents, are judged to pose the highest potential risk. These devices are either life-sustaining/supporting, of substantial importance in preventing impairment of human health, or present a high risk of illness or injury. Consequently, general and special controls alone are inadequate to provide reasonable assurance of safety and effectiveness. Most class III devices require FDA approval of a PMA before they can be legally marketed. Approval of the PMA generally requires clinical data demonstrating reasonable assurance that the device is safe and effective in the target population.5
The Human Device Exemption (HDE) is a new pathway to allow for commercialization of class III devices designed to address small markets, ie, diseases or conditions that affect fewer than 4000 patients in the United States each year. Approval of an HDE requires demonstration that the device is safe and the probable benefits outweigh the probable risks. Although the process may require smaller clinical trials, an HDE device must continue to operate under local IRB approval at each participating institution and must continue to collect case report forms akin to an ongoing clinical trial. The PMA process typically involves a series of studies starting with first clinical use and culminating in a multicenter, prospective randomized control trial (pivotal trial). The complexity and extent of the clinical testing program is dictated by the nature of the device and its proposed use. The clinical study program is developed by the company in conjunction with clinician investigators, all in close collaboration with FDA/CDRH.6
The first and arguably most important step in this process is the pre-IDE meeting, in which the company, often accompanied by the lead clinical investigator(s), meets with FDA/CDRH to present data about the device, its clinical development program, and its intended use after approval. The FDA/CDRH staff reviews existing bench and animal data (as well as any outside-the-United States clinical data) and makes informal non-binding suggestions regarding the need (if any) for additional pre-clinical data (bench and animal), as well as the study design. The sponsor then submits an IDE application to FDA/CDRH for formal review.
Clinical development of a new class III device is typically divided into pilot and pivotal trial phases. The purpose of the pilot phase (starting with first clinical use) is to establish safety and to assist in design of the pivotal trial. Pilot-phase testing is typically limited to fewer than 100 patients treated at a few centers. The purpose of the pivotal trial is to generate data that define patient populations in which use of the device is safe and effective. The dialogue initiated during the pre-IDE meeting continues and intensifies between FDA/CDRH and the company over the specifics of the pivotal trial and includes the patient population, the control group against which the new device will be evaluated, and the primary and secondary end points of the evaluation. For first-in-class devices, eg, drug-eluting stents, where there are few data regarding short- or long-term outcomes, FDA/CDRH requires prospective randomized controlled studies. Though high profile, devices that require randomized data for approval are the exception rather than the rule. The vast majority of device clinical trials are case series that carefully document product performance. Still more products are approved as "tools."7
Most devices currently in testing are similar to well-characterized approved devices, eg, next generation bare metal stent, angioplasty balloon, etc. When FDA/CDRH has substantial data on the device class metrics, comparisons may be made to historical data or objective performance criteria. When few data on existing standards are available, the FDA typically requires randomized rather than single-arm studies, in which the new device is compared against concurrent controls treated with current best medical practice. That comparison may be powered to show that the new treatment is superior to prior approaches, or that it is non-inferior (equivalent or better) compared with a previously approved device in a new area. The issues relating to the type of trial are discussed below.
The specifics regarding study design may have profound impact on the time and cost of bringing a new device to market. Though the primary mission of the FDA/CDRH is to ensure safety and effectiveness of commercially available devices, when exerting regulatory oversight the agency must balance its primary mission with the costs of introducing new technologies to the clinical marketplace. This has been codified by the FDA Modernization Act and the FDA Modernization Act-II, which require the agency to pursue the "least burdensome means" available to establish device safety and efficacy.8,9 Although surrogate end points, eg, angiographic restenosis rather than recurrent cardiac events, may be allowed as secondary end points, the primary end point of a pivotal trial for a first-in-class class III device is usually a clinical end point (or a well-established surrogate such as infarct size in a myocardial infarction treatment device). Depending on the end points negotiated, such pivotal trials may require enrollment of 1000 or more patients at 30 to 50 sites over a period of 1 to 2 years, with appropriate follow-up frequently to 1 year after treatment. The trial must be conducted according to good clinical practices standard, with the approval of the local IRB at each participating center.
| Clinical Site (IRB, Contract, Conflict of Interest) |
|---|
|
|
|---|
Every clinical site is federally mandated to have an IRB responsible to ensure the protection of the rights, safety, and welfare of research subjects. Regulation of the IRB review of protocols involving medical devices is under the purview of the FDA. The Office of Protection From Research Risks (OPRR) is responsible for oversight regarding all human research and is in direct communication with the FDA/CDRH. Studies involving human subjects that do not involve products regulated by the FDA fall under the direct purview of the OPRR. Both the FDA and the OPRR are in the Department of Health and Human Services. Each IRB must meet standards for the composition, leadership, and processes set forth by that department. IRBs are subject to periodic audits by the FDA to ensure that records and procedures are in compliance with regulations.9,10
Working with the sponsor, the principal investigator prepares an application to the IRB at his/her institution that includes the consent form describing in lay language the device, the proposed clinical study, the inclusion and exclusion criteria for the trial, and a draft consent form describing the risks and benefits of participation in the study. The IRB then formally reviews the application and frequently requests changes, particularly to the informed consent. The IRB process typically requires approximately 3 months, but at times can take considerably longer.
The company must also negotiate agreements with each clinical site addressing the many issues associated with the clinical trial. In addition to the study costs/reimbursement (per-patient enrolled and overhead), these agreements typically include indemnification and the assignment of ownership rights of new discoveries (intellectual property) made in the course of the study. The resources required at each center to perform the high quality research necessary for a PMA protocol are formidable. In addition to resources required to maintain an excellent clinical program (physician, nursing, and technical staff coupled with state of the art facility), the study center must dedicate additional resource, primarily research nurses to perform high-quality clinical research. The role of the research nurse is critical to all phases of the trial, including general study management, IRB process, patient recruitment, and accurate completion of case report forms. The sponsor is responsible for accrediting each clinical site to document that the necessary resources are in place to fulfill the demands set forward in the protocol. The clinical research staff do not provide clinical care and are thus not part of the hospitals salaried clinical nursing staff. The per-patient costs include all additional charges the patient incurs for the study, including procedures (therapeutic and diagnostic), clinic visits, and diagnostic studies, as well as research nurses salaries. It is estimated that it takes 3 to 6 months to formally recruit each clinical site.
Device development from the earliest stages requires active involvement of practicing clinicians. Clinician/inventors are frequently involved in creation of the device concept and are often integral members of the design team performing the majority of the early animal studies. Through this involvement, the clinician/inventor obtains intimate knowledge of device performance and failure modes. Safety concerns during first clinical use and pilot phase mandate participation by these clinician/inventors. The clinician/inventors frequently take leadership roles and have equity positions in the company developing the device. These interests present important conflicts of interest which must be addressed to ensure patient safety, data integrity, and public trust in the process. Many institutions have set up formal processes to address these conflicts of interest, potentially adding more time to the institutional recruitment process.
Pivotal studies required for a PMA application are typically large multicenter randomized trials and often represent the largest commercial risk and expense in the device development process. In addition to obtaining an IDE from the FDA and formally recruiting clinical sites, the sponsor must also put into place an extensive infrastructure that typically includes engaging a contract research organization (CRO), core laboratories, formation of a data safety monitoring board (DSMB), and an executive committee. The CRO provides the infrastructure required to recruit, qualify, and audit sites. Core laboratories evaluate primary data, eg, angiography, ECG, and ultrasound results, in a uniform and blinded manner. An executive committee is typically composed of the clinician investigators, company representative, CRO, and core laboratories. The DSMB, composed of a group senior clinical investigators and statisticians with no other involvement in the study, periodically reviews trial data at specified intervals. The DSMB has the mandate to stop or modify a study, eg, discontinue randomization to one of the groups if complications associated with the study device are in excess of anticipated rates, if differences between study groups reach statistical significance, or if further patient enrollment will not impact study outcome.
| Regulatory Approval Process (European Union) |
|---|
|
|
|---|
The European Union system relies heavily on notified bodies (NBs), which are independent commercial organizations to implement regulatory control over medical devices. NBs have the ability to issue the CE mark, the official marking required for certain medical devices. NBs are designated, monitored, and audited by the relevant member states via the national competent authorities. Many functions performed by the FDA/CDRH within the United States are performed by NBs, including medical device certification, device type designation, assessment and verification of quality systems, and review of design dossiers for high-risk devices.11,12 Currently, there are more than 50 active NBs within Europe. A company is free to choose any notified body designated to cover the particular class of device under review. After approval, post-market surveillance functions are the responsibility of the member state via the competent authority. NBs typically function in a closed manner, providing little visibility on criteria required for approval. This dynamic allows for a high degree of variation as well as competition among NBs. As a result, NBs are perceived by industry to be less bureaucratic organizations that can respond more quickly and efficiently than the FDA. These potential benefits may be offset by a system that is intrinsically more fragmented and highly variable and has resulted in the approval and continued marketing of devices, eg, abdominal aorta stent grafts, in Europe that failed efficacy trials in the United States.
Criteria for approval of high-risk devices are different in the European Union. To receive approval to market a class III high-risk (and some class II) device in the United States, the manufacturer must demonstrate the device to be reasonably safe and effective, which typically requires a prospective, randomized controlled clinical trial. To receive approval to market a device in the European Union, the manufacturer must demonstrate that the device is safe and that it performs in a manner consistent with the manufacturers intended use.12 This difference has a profound impact on the size and scope of the clinical studies for regulatory approval. This significant difference is illustrated by examining the introduction of distal protection systems. The GuardWire developed by PercuSurge, Inc (later acquired by Medtronic) is a specialized coronary guidewire with an elastomeric balloon mounted at the tip. During an angioplasty/stent procedure, the operator crosses the lesion with the GuardWire and inflates the balloon. Stent placement is then performed, after which a specialized catheter is used to evacuate any arterial debris that may have become dislodged during the procedure. The GuardWire balloon is then deflated. Demonstration of safety and performance, ie, ability to aspirate material during the stenting procedure, was demonstrated in a 22-patient single-arm study.13 In contrast, in the United States, this device was designated class II (requiring 501[k] clearance and clinical data). To satisfy US criteria for clearance, the standard of safety and effectiveness required in this case was defined as the ability to reduce complications associated with stenting of saphenous vein grafts. To meet this criterion, an 800-patient multicenter trial randomized trial comparing distal protection to usual care (no protection) was performed. At 30 days, a 42% relative reduction in major adverse cardiac events (primarily myocardial infarction) was observed.14 A trial of this type is estimated to cost US$10 to $12 million and may take 24 months to perform.
| Conclusion |
|---|
|
|
|---|
Taken together, these factors account for the 1- to 3-year delays in the introduction of new device technologies into general clinical practice within the United States as compared with Europe. Each system has strengths and weaknesses that must be evaluated within the context of different health delivery systems. Through better understanding of these systems, will we be able to recommend modifications and improvements toward improving speed and efficiency (shorter delay to US versus European testing and ultimate clinical approval), without compromising the basic demonstration of safety and efficacy that remains the US regulatory mandate.
| Acknowledgments |
|---|
| Footnotes |
|---|
This article is the result in part from Sessions at the Dartmouth Drug and Device Development Symposium (3DI) held at the Woodstock Inn, Woodstock, Vt, October 30 and 31, 2003.
| References |
|---|
|
|
|---|
812.30 (2003).2. Chai JY. Medical device regulation in the United States and the European Union: a comparative study. Food and Drug Law Journal. 2000; 55: 5780.[Medline] [Order article via Infotrieve]
5. Medical Device Amendments of 1976, Pub L No 94-295 (1976).
6. 61 Federal Register 33233 (1996) (codified at 21 CFR
814.102).
7. Feigal DW, Gardner SN, McClellan M. Ensuring safe and effective medical devices. N Engl J Med. 2003; 348: 191192.
9. Food and Drug Administration Modernization Act of 1997, Pub L No 105-115 (1997).
11. US Food and Drug Administration web site. Available at: http://www.fda.gov. Accessed May 28, 2004.
12. Council Directive 98/97/European Economic Community. Annex I.A. 35; 1998. O.J. (L331) 15.6.
13. Webb JG, Carere RG, Virmani R, et al. Retrieval and analysis of particulate debris following saphenous vein graft intervention. J Am Coll Cardiol. 1999; 34: 468475.
14. Baim DS, Wahr D, George B, et al. Randomized trial of a distal embolic protection device during percutaneous intervention of saphenous vein aorto-coronary bypass grafts. Circulation. 2002; 105: 12851290.
This article has been cited by other articles:
![]() |
D. S. Baim, R. Mehran, D. J. Kereiakes, T. P. Gross, M. Simons, D. Malenka, and A. V. Kaplan Postmarket Surveillance for Drug-Eluting Coronary Stents: A Comprehensive Approach Circulation, February 14, 2006; 113(6): 891 - 897. [Full Text] [PDF] |
||||
![]() |
A. V. Kaplan, E. D. Harvey, R. E. Kuntz, H. Shiran, J. F. Robb, and P. Fitzgerald Humanitarian Use Devices/Humanitarian Device Exemptions in Cardiovascular Medicine Circulation, November 1, 2005; 112(18): 2883 - 2886. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Somberg, W. Laskey, D. Morrison, J. Weinberger, D. J. Kereiakes, A. V. Kaplan, and R. Mehran Letter Regarding Articles by Kereiakes and Willerson, Mehran et al, and Kaplan et al, "Mini-Review: Expert Opinion" * Response Circulation, March 29, 2005; 111(12): e164 - e165. [Full Text] [PDF] |
||||
![]() |
D. J. Kereiakes and J. T. Willerson Medical Technology Development and Approval: The Future Is Now Circulation, June 29, 2004; 109(25): 3078 - 3080. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2004 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |