Donate Help Contact The AHA Sign In Home
American Heart Association
Circulation
Search: search_blue_button Advanced Search
Circulation. 2004;109:1735-1740
Published online before print April 5, 2004, doi: 10.1161/01.CIR.0000124724.14491.6F
This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
109/14/1735    most recent
01.CIR.0000124724.14491.6Fv1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow Request Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Rogers, C.
Right arrow Articles by Macon, N. D.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Rogers, C.
Right arrow Articles by Macon, N. D.
Right arrowPubmed/NCBI databases
*Compound via MeSH
*Substance via MeSH
Medline Plus Health Information
*Coronary Artery Bypass Surgery
*Coronary Artery Disease
Related Collections
Right arrow Catheter-based coronary and valvular interventions: other
Right arrow Catheter-based coronary interventions: stents
Right arrow CV surgery: coronary artery disease
Right arrow Acute myocardial infarction
Right arrow By-pass procedures

(Circulation. 2004;109:1735-1740.)
© 2004 American Heart Association, Inc.


Clinical Investigation and Reports

Embolic Protection With Filtering or Occlusion Balloons During Saphenous Vein Graft Stenting Retrieves Identical Volumes and Sizes of Particulate Debris

Campbell Rogers, MD; Rany Huynh, BS; Philip A. Seifert, MS; Bernard Chevalier, MD; Joachim Schofer, MD; Elazer R. Edelman, MD, PhD; Gregory Toegel, BS; Arun Kuchela, MD; Amy Woupio, BS; Richard E. Kuntz, MD; Natalie D. Macon, PhD

From the Cardiovascular Division (Cardiac Catheterization Laboratory and Coronary Care Unit), Brigham and Women’s Hospital, Harvard Medical School (C.R., E.R.E., A.K., R.E.K.), Boston, Mass; Medtronic, Inc (R.H., A.W., N.D.M.), Danvers, Mass; Harvard-M.I.T. Division of Health Sciences and Technology, Massachusetts Institute of Technology (C.R., P.A.S., E.R.E., G.T.), Cambridge, Mass; Center Cardiologique du Nord (B.C.), Saint Denis, France; and Center of Cardiology and Vascular Intervention (J.S.), Hamburg, Germany.

Correspondence to Campbell Rogers, MD, Cardiovascular Division, Brigham and Women’s Hospital, 75 Francis St, Boston, MA 02115. E-mail crogers{at}partners.org

Received January 6, 2003; de novo received October 21, 2003; revision received January 9, 2004; accepted January 23, 2004.


*    Abstract
up arrowTop
*Abstract
down arrowIntroduction
down arrowMethods
down arrowResults
down arrowDiscussion
down arrowConclusions
down arrowReferences
 
Background— Distal embolization of plaque particulate liberated during stenting may cause periprocedural complications. The number, size, and volume of debris released during stenting, however, have not been quantified, rendering embolic protection approaches empiric. We used a novel method of microparticle size assessment to measure volume and characterize individual sizes of particles captured by the PercuSurge GuardWire balloon or a vascular filter during saphenous vein graft stenting.

Methods and Results— Braided nitinol filters (average distal pore size 100 µm) were used in 47 saphenous vein grafts in 44 patients. The PercuSurge GuardWire was used in 17 saphenous vein grafts in 16 patients. Particulate debris was subjected to microparticle size analysis (RapidVue, Beckman Coulter). All samples contained particulate debris. For both filter and GuardWire populations, most particles were <100 µm in longest dimension (87% and 90% of particles, respectively), and the distribution of particle sizes was identical. Total embolic load per lesion for both filters and GuardWire aspirates was also similar: median embolic load per filter was 16 mm3 (range 2 to 84 mm3). Median embolic load per GuardWire was also 16 mm3 (range 7 to 42 mm3). Histopathologic analysis demonstrated that most samples contained plaque elements and platelet-rich thrombus.

Conclusions— During saphenous vein graft interventions, particulate retrieved with a vascular filtering device or an occlusion balloon was similar in amount and character. This supports the notion that unless soluble mediators play an important role in adverse acute clinical events after stenting, the clinical efficacy of filtering devices may be equal to that of occlusion devices.


Key Words: bypass • embolism • myocardial infarction • stents


*    Introduction
up arrowTop
up arrowAbstract
*Introduction
down arrowMethods
down arrowResults
down arrowDiscussion
down arrowConclusions
down arrowReferences
 
A major complication of percutaneous interventions in degenerated saphenous vein grafts (SVGs) is distal embolization of plaque constituents that results in no-reflow, myocardial infarction, and stroke.1–3 Pharmacological approaches such as glycoprotein IIb/IIIa inhibition or direct stenting have not been shown to reduce complications resulting from distal embolization in SVG stenting.4–7 Distal protection devices8–17 have been developed, intended to trap plaque-derived emboli during interventions and prevent embolization into the distal vascular bed. There has been substantial debate surrounding the size and volume of particulate debris and the clinical relevance of large and small particles, as well as of soluble mediators generated during angioplasty and stenting. The first distal protection device evaluated clinically after SVG stenting, the PercuSurge GuardWire (Medtronic Inc), despite complete occlusion of the vessel during intervention with temporary cessation of flow, was shown to reduce the incidence of no-reflow by 67% and of in-hospital and 30-day major adverse cardiac events by 42%.18 By virtue of complete occlusion and poststenting complete graft evacuation, the PercuSurge device theoretically prevents embolization of all particles, no matter how small, as well as of soluble mediators released during stenting.

It has not been established that the efficacy of occlusion systems can be matched by newer approaches to embolic protection such as filtering devices, which do not occlude anterograde flow but rather strain the effluent so as to prevent particulate embolization. Potential advantages of filtering include perfusion during intervention and shorter procedure times. Potential disadvantages include incomplete protection from emboli or failure to protect against soluble mediators. Both occlusion balloons and filters have the potential to inflict damage to the vessel wall during deployment, use, or retrieval. The recently published FIRE trial (FilterWire EX Randomized Evaluation) showed parity in clinical outcomes after SVG stenting in patients randomly assigned to embolic protection with the FilterWire (Boston Scientific) or to the GuardWire (Medtronic).19 However, rigorous particulate analysis was not performed in that trial, and it is not known whether there may be benefits of an occlusion balloon system (eg, more complete particulate retrieval and prevention of soluble mediators released during intervention) that are being offset by requisite procedural ischemia provoked by occlusion.

We used a filter constructed of braided nitinol with a distal pore size of 100 µm and the GuardWire to quantify and characterize the volume and individual sizes of particles captured during SVG stenting, postulating that particulate retrieval with the filter would be less efficient than with the GuardWire. Using novel methods of microparticulate quantification, we found that most particles released during SVG stenting were small (<100 µm) and that, contrary to our hypothesis, the vascular filter trapped even very small particles in equal amounts to the GuardWire.

These clinical data support the potential role for vascular filters in the efficient prevention of distal embolization during SVG interventions. Although only direct comparisons of clinical end points can prove equivalence, our data suggest that filters may be as effective as occlusion balloons. Furthermore, the abundance of small particles retrieved from SVG interventions will need to be compared with particulate debris retrieved from other vascular beds, eg, carotid, coronary, and renal, to optimize embolic protection device design and use in these locations.


*    Methods
up arrowTop
up arrowAbstract
up arrowIntroduction
*Methods
down arrowResults
down arrowDiscussion
down arrowConclusions
down arrowReferences
 
Clinical Samples
Forty-seven filters from 44 patients entered in the Safety of Embolic Capture Utilizing the Medtronic AVE Distal Protection Device in Revascularization of SVGs (SECURE) study16 were submitted for analysis. Patients undergoing multiple SVG interventions had a new filter used for each SVG. The Medtronic filter (Medtronic/AVE) is a nitinol mesh basket attached to an 0.014-in guidewire. Initially constrained within a delivery sheath, once passed through the lesion into the distal SVG, the filter passively expands as the sheath is withdrawn. After stenting, the sheath is readvanced, and the filter is mechanically collapsed and withdrawn. After retrieval, each filter was immediately placed in 10% buffered formalin.

GuardWire samples were obtained from 16 consecutive patients undergoing protected stenting of 17 SVGs at Brigham and Women’s Hospital. Forty milliliters of blood aspirated via the Export catheter was passed through the Becton Dickinson Cell Strainer 40-µm mesh filter, which was then placed into 10% formalin. For the single patient undergoing multiple-SVG intervention, the aspirate from each SVG was analyzed separately.

Particulate Isolation
Medtronic Filter
All material in the collection vial was analyzed, including debris in the filter and loose in the formalin. Filters (Figure 1) were placed in 0.01 mol/L PBS, filter wires were cut, and particulate was displaced into surrounding PBS via gentle agitation. Loose particles in formalin were collected with 2-µm filter paper and combined with particulate isolated from the filter, then placed in PBS. Control samples consisted of unused filters (n =6) processed in the same manner as the clinically used filters.



View larger version (123K):
[in this window]
[in a new window]
 
Figure 1. Nitinol filter after use in SVG stenting procedure. Particulate is visible throughout filter cavity. After liberation from filter (inset), particles are freely suspended before size analysis.

GuardWire Aspirate
All material aspirated during intervention on a given lesion was fixed in 10% buffered formalin and analyzed.

Particulate Analysis
The Coulter RapidVue apparatus was calibrated with National Institute of Standards and Technology-certified spheres (200±4 µm; Duke Scientific Co) before automated sample analysis. Aggregate particle volume measurements were validated in a separate experiment with poly(vinyl alcohol) (PVA) particulate (71% PVA 50 and 29% PVA 100, Cook Inc) to verify accuracy. Each sample was transferred to the RapidVue, and PBS was added to bring the total sample volume to 75 mL. Samples received in volumes >75 mL were divided into 2 parts and run separately, and the results were summed.

The closed-circuit RapidVue system circulated the particles and captured digital images at an automatically variable frame rate optimized for volume quantification as all particles passed through a field of view over 100 seconds (frame rates: GuardWire 5.30 to 26.19 frames/s; filter 6.84 to 26.67 frames/s; control 25 to 28.05 frames/s). Frame rate was inversely proportional to particulate quantity. These optimized parameters were used for all samples to allow direct comparison of results.

Samples were analyzed 3 times each to ensure repeatability, then retained for histological analysis. The results of the 3 runs per sample were averaged to obtain volume and size distribution data. Particle sizes were calculated with the least bounding rectangular (LBR) model, which gave the major and minor axes (Figure 2). Particle sizes were collated by major-axis measurement with categories previously appearing in the literature (<56 µm, 56 to 96 µm, 96 to 213 µm, 213 to 763 µm, and >763 µm).20 Total volume of particulate for each sample was calculated as the LBR-derived volume output from RapidVue after 100 seconds (as determined by total volume calibration with PVA particulate) minus the volume calculated after 100 seconds for 6 control unused filters.



View larger version (52K):
[in this window]
[in a new window]
 
Figure 2. Particle size distribution plots for particulate retrieved during SVG stenting via GuardWire (A) or nitinol filter (B). x-axis units are length (µm); y-axis units are percent of particles. Upper plots are width of LBR; lower plots are length of LBR.

Histological Analysis
After particulate analysis, specimens were transferred into 70% ethanol, filtered into biopsy bags (Thermo Shandon), processed by a routine biopsy processing schedule, and embedded in paraffin. Serial 5-µm sections step-cut at 50-µm levels were collected and stained with Mayer’s hematoxylin and eosin-Y. Representative levels in particulate specimens were also stained with Carstair’s fibrin stain and with picrosirius red for collagen. Immunohistochemical staining was performed to label CD68 (clone: PG-M1, Dako Co) for monocyte/macrophages and smooth muscle cell {alpha}-actin (clone: 1A4, Dako Co) for smooth muscle and fibromuscular cells with the EnVision+ detection system (Dako Co). Stained sections were viewed and qualitatively scored blindly, and scored sections were sorted on the basis of the presence of tissue components into 3 groups: (1) thrombus with plaque elements (containing white thrombus, foam cells, and fibrous plaque elements), (2) thrombus only (containing either white or red-cell and white thrombus), and (3) no visible material.

Statistical Methods
Data are presented as median and range values. Comparisons between filter and GuardWire volumes and between particle size distributions for filter and GuardWire samples were performed with nonparametric statistics (2-tailed Mann-Whitney U test) rather than parametric statistics because the data were likely not normally distributed. P<0.05 was considered significant. To estimate the statistical power of this study, the differences detectable with 80% power between the 2 samples were estimated for volume and particle dimensions (major and minor axis) with parametric statistics (mean and SD estimates). Because the data are slightly right-skewed for all measurements and deviate somewhat from normal distribution, these power estimations are not exact.


*    Results
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
*Results
down arrowDiscussion
down arrowConclusions
down arrowReferences
 
Calibration With Microspheres
NIST-certified beads (200±4 µm) were placed in a 75-mL volume of PBS for RapidVue analysis. The beads were analyzed 4 times. The RapidVue-calculated bead size was compared with the NIST-certified size range of the beads, and the results (198.2 to 200 µm) were within the certified range.

Total Volume Calibration With PVA
Samples with known volume (5, 10, 30, 50, 70, or 100 mm3) and size distribution (71% PVA 50 and 29% PVA 100) of PVA particulate were measured in the RapidVue. Seven samples of each volume were used, and each sample was run 3 times. The RapidVue volume calculations (5 mm3: 6.79±1.2 mm3; 10 mm3: 12.80±1.08 mm3; 30 mm3: 29.05±2.26 mm3; 50 mm3: 53.03±9.86 mm3; 70 mm3: 64.88±5.95 mm3; and 100 mm3: 92±7 mm3) were comparable to the known volumes after 100 seconds.

Particle Size Distribution
Mean and median particle sizes of debris captured with filters or with GuardWire were nearly identical (filter: mean 62.9±13.2 µm [median 59.4 µm] in major axis, mean 34.7±4.8 µm [median 34.8 µm] in minor axis; GuardWire: mean 59.8±13.4 µm [median 55.4 µm] in major axis, mean 38.2±7.6 µm [median 34.8 µm] in minor axis, P=0.28). Particle size distributions were also nearly identical for filter and GuardWire samples (<56 µm in longest dimension: 70.5±6.7% [median 70.9%] of particles from each filter, 64.7±13.0% [median 68.2%] of particles from each GuardWire sample; and 56 to 96 µm in longest dimension: 16.8±2.7% [median 16.6%] for filters and 25.3±9.2% [median 24.9%] for GuardWire samples; Figure 3). The largest particles retrieved were 2148 µm with the GuardWire and 2980 µm with the filter in the longest dimension.



View larger version (40K):
[in this window]
[in a new window]
 
Figure 3. Size distribution of particulate retrieved from SVG stenting procedures via nitinol filter (solid bars) or GuardWire (open bars). Size partitions were chosen to match those reported previously for GuardWire embolic protection. No difference was observed in particulate size distribution between filter and GuardWire. Despite nominal distal pore size of 100 µm, filters trapped emboli far smaller than 100 µm in longest dimension.

Total Volume Distribution
Total embolic load per lesion for filters and GuardWire samples was similar (GuardWire: mean 17±9 mm3 [median 16 mm3, range 7 to 42 mm3]; Filter: mean 23±19 mm3 [median 16 mm3, range 2 to 83 mm3]). A box plot comparison (Figure 4) demonstrated identical median values and a larger range for the filter that correlated with its larger sample (GuardWire n =17, filter n =47).



View larger version (8K):
[in this window]
[in a new window]
 
Figure 4. Total volume by device showing similar expected values of retrieved debris. Median values (horizontal bar within box) were identical (16 mm3), and wider range of interquartile difference (width of 25th and 75th percentile boundaries, shown as lower and upper box edges) for filter correlated with its larger sample (GuardWire n =17; filter n =47). Whiskers above and below boxes extend to furthest points within 1.5 interquartile range above and below box edges, and open circles mark outlier values.

Statistical Power to Detect Differences
Given the sample size of each group and variance estimates observed, the present study had 80% power to detect a difference between the 2 groups of 16.8 mm3 in volume, 6.1 µm in the mean minimum-axis dimension, and 10.5 µm in the mean major-axis dimension.

Histological Evaluation
The histological assessment of particulate tissue obtained from 47 filters revealed that the braided filter contained thrombus together with plaque elements in 18 samples (38%) and thrombus alone in 23 samples (49%). Among the 17 GuardWire aspirate samples, thrombus together with plaque elements was seen in 5 samples (29%) and platelet-fibrin rich thrombus in 8 samples (47%).

No tissue was seen in 6 filter samples (13%) and 4 GuardWire samples (23%), likely related to loss of particles during processing for histological analysis after particle size analysis, because inspection of particulate after RapidVue analysis showed gross visualization of particulate from all filters and GuardWire aspirates. Similar rates of loss during handling have been reported previously.7


*    Discussion
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
up arrowResults
*Discussion
down arrowConclusions
down arrowReferences
 
Acute myocardial infarction occurs after SVG stenting in 15% to 28% of cases,13,21–23 often preceded by reduced intraprocedure flow despite adequate mechanical treatment of stenoses (no-reflow). Late morbidity and mortality are higher in patients who have a myocardial infarction after SVG stenting than in those who do not.24 Etiologic possibilities include soluble mediators released from red blood cells, platelets, or leukocytes during stenting, acutely dislodged plaque elements, and delayed embolization of plaque or thrombotic elements. Pharmacological strategies proposed to prevent or treat periprocedural no-reflow have included calcium channel blockers, nitrates, adenosine, thrombolytics, and antiplatelet agents. A registry reported from Brigham and Women’s hospital25 demonstrated that despite improving angiographic flow, antiplatelet and vasodilatory agents were not associated with improved clinical outcomes. Mechanical strategies proposed to prevent no-reflow have included direct stenting and the use of high surface-coverage stents or covered stent grafts to trap debris against the vessel wall.

Distal balloon occlusion during stenting with subsequent aspiration of dislodged debris is the only approach to date shown to limit adverse clinical events and angiographic no-reflow after SVG stenting. In the SAFER trial (Saphenous vein graft Angioplasty Free of Emboli Randomized trial), patients undergoing SVG stenting and randomly assigned to distal embolic protection with the PercuSurge GuardWire system had a 42% reduction in risk of periprocedural adverse cardiac events, primarily myocardial infarction.18 Despite the marked efficacy of occlusive distal protection, however, its widespread adoption has been limited. Clinical tolerance of occlusion is not universal, fluoroscopic visualization of stenoses before and after stenting is limited, and in vascular beds other than SVGs, the inability to protect proximal side branches from embolization limits the efficacy of distal balloon occlusion. Vascular filtering devices may offer superiority over occlusion devices in ease of use, patient tolerance, visualization, and proximal side-branch protection. On the other hand, filters may not be able to trap small fragments and certainly cannot prevent soluble mediators from reaching the distal vascular bed.

The randomized FIRE trial showed similar clinical efficacies of balloon occlusion and filter protection with the FilterWire (Boston Scientific) during SVG stenting19 (major adverse cardiac events occurred in 9.9% of FilterWire patients and 11.6% of GuardWire patients; P=0.0008 for noninferiority of the FilterWire). Importantly, that study did not examine mechanistic underpinnings of device efficacy, capture efficiency, or adverse events. For example, one might view the clinical equivalence demonstrated in FIRE, which used a filter with a nominal pore size of 110 µm, as an indication that particles smaller than 110 µm are unimportant in the genesis of adverse events after SVG stenting. However, without examining the particulate retrieved via filters or occlusion systems, such a conclusion would be speculative.

Previous analyses of embolic particulate liberated by arterial stenting have used scanning electron or light microscopy20 to measure manually the size of a subset of retrieved particulate but have not quantified retrieved particulate volume for relative comparison between occlusive and filtering distal protection techniques. Grube et al20 reported that >80% of particles retrieved with the GuardWire from SVG stenting were <96 µm in largest dimension, which lends strength to the notion that filters may afford less complete protection, because smaller particles might pass through unimpeded.

We constructed the present data analysis to conform to the size categories reported by Grube et al,20 and our data from GuardWire cases are virtually identical to theirs, which validates our measurement technique. The present study applied a highly sensitive and reproducible microparticle size analysis system to debris retrieved from SVG stenting procedures with GuardWire occlusion or with a vascular filtering device (average distal pore size 100 µm). We found that a vascular filter with a nominal distal pore size of 100 µm captured particles far smaller than 100 µm and retrieved the same total volume of particulate as a device that achieved complete occlusion. One possible explanation is that deposition of aggregate debris, platelets, or fibrin on the filter pores reduces the functional orifice size. Another is that debris adheres to the surface of the filter and is thereby prevented from distal embolization.

The connection between retrieved particulate and the prevention of embolism-related adverse clinical events (no-reflow and myocardial infarction) remains to be proven. In the present study, there were no in-hospital deaths or Q-wave myocardial infarctions in either group. No-reflow occurred in 5.3% and non–Q-wave myocardial infarction during hospitalization in 7.0%16 of filter-treated patients. Non–Q-wave myocardial infarction occurred in 6.0% of the 16 patients treated with the GuardWire. These incidence rates compare favorably with the 30-day event rates among patients who received distal embolic protection in the FIRE and SAFER trials.

Most explorations of distal embolic protection have involved degenerated SVGs. Native coronary occlusions in the setting of acute myocardial infarction and carotid26 and renal artery stenoses are also potential sites where distal embolization during intervention exacts a clinical price. Debris released in these settings may differ from SVG debris in quantity, histopathologic composition, and total volume. Particulate analysis will be an important element in determining optimal device selection and design for each clinical setting.

Study Limitations
The present data are nonrandomized, derived from 2 sets of patients undergoing SVG stenting. Therefore, the embolic loads liberated during stenting may have differed, although there were no selection criteria for filter patients that would make them particularly different from GuardWire patients. Specimen handling also varied slightly between groups: the filter group had the filter collapsed and withdrawn at the end of the procedure, whereas the GuardWire group had the sample aspirated via the Export catheter, then sieved with a 40-µm mesh. This difference could alter particulate size via compression or fragmentation of particles, although the data from the present study indicating that particulate volume and size distribution were essentially identical between devices are reassuring. Similarly, particles may have been liberated before or during deployment, may have escaped capture, and thus may not be analyzed. Finally, the lower limit of detection for the RapidVue is 10 µm. Fragments smaller than 10 µm might be captured more reliably by one system or the other, although a clinical effect of such erythrocyte-sized particles is unlikely.


*    Conclusions
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
up arrowResults
up arrowDiscussion
*Conclusions
down arrowReferences
 
Liberation of atherosclerotic plaque elements during vascular interventions may be universal and unavoidable and may be an important cause of periprocedural adverse events. Novel approaches to embolic protection, specifically vascular filtering rather than occlusion, have been hampered by the notion that some fragments will escape capture and cause adverse events. The present data suggest that filtering may be as efficient as balloon occlusion at particle capture, despite pore sizes larger than the vast majority of particles. This observation may help justify the application of existing technologies and direct future development optimizing safety and efficacy in diverse settings where procedurally induced plaque embolization occurs.


*    Acknowledgments
 
We are grateful to Dr Donald Baim and David Sauvageau for insightful review of the manuscript.


*    References
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
up arrowResults
up arrowDiscussion
up arrowConclusions
*References
 

  1. Erbel R, Heusch G. Coronary microembolization: its role in acute coronary syndromes and interventions. Herz. 1999; 24: 558–575.[Medline] [Order article via Infotrieve]
  2. Piana RN, Paik GY, Moscucci M, et al. Incidence and treatment of "no-reflow" after percutaneous coronary intervention. Circulation. 1994; 89: 2514–2518.[Abstract/Free Full Text]
  3. Smith SC Jr, Dove JT, Jacobs AK, et al. ACC/AHA guidelines for percutaneous coronary intervention: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Committee to Revise the 1993 Guidelines for Percutaneous Transluminal Coronary Angioplasty). J Am Coll Cardiol. 2001; 37: 2215–2239.[Free Full Text]
  4. Roffi M, Mukherjee D, Chew DP, et al. Lack of benefit from intravenous platelet glycoprotein IIb/IIIa receptor inhibition as adjunctive treatment for percutaneous interventions of aortocoronary bypass grafts. Circulation. 2002; 106: 3063–3067.[Abstract/Free Full Text]
  5. Ellis SG, Lincoff AM, Miller D, et al. Reduction in complications of angioplasty with abciximab occurs largely independently of baseline lesion morphology. J Am Coll Cardiol. 1998; 32: 1619–1623.[Abstract/Free Full Text]
  6. Mathew V, Grill DE, Scott CG, et al. The influence of abciximab use on clinical outcome after aortocoronary vein graft interventions. J Am Coll Cardiol. 1999; 34: 1163–1169.[Abstract/Free Full Text]
  7. Roffi M, Mukherjee D, Chew DP, et al. Lack of benefit from intravenous platelet GP IIb/IIIa receptor inhibition as adjunctive treatment for percutaneous interventions of aortocoronary bypass grafts. Circulation. 2002; 106: 3063–3067.[Abstract/Free Full Text]
  8. Webb J, Carere RG, Virmani R, et al. Retrieval and analysis of particulate debris after saphenous vein graft intervention. J Am Coll Cardiol. 1999; 34: 468–475.[Abstract/Free Full Text]
  9. Oesterle SN, Hayase M, Baim DS, et al. An embolization containment device. Cathet Cardiovasc Intervent. 1999; 47: 243–250.[CrossRef][Medline] [Order article via Infotrieve]
  10. Carlino M, Di Gregorio J, Di Mario C, et al. Prevention of distal embolization during saphenous vein graft lesion angioplasty. Circulation. 1999; 99: 3221–3223.[Abstract/Free Full Text]
  11. Henry M, Amor M, Henry I, et al. Carotid stenting with cerebral protection: first clinical experience using the PercuSurge GuardWire system. J Endovasc Surg. 1999; 6: 321–331.[CrossRef][Medline] [Order article via Infotrieve]
  12. Muller R, Gerckens U, Staberock M, et al. Clinical experiences with the PercuSurge GuardWire: a new system for prevention of peripheral embolisms in catheter interventions on degenerated coronary artery venous bypasses [in German]. Z Kardiol. 2000; 89: 316–322.[CrossRef][Medline] [Order article via Infotrieve]
  13. Ohki T, Roubin GS, Veith FJ, et al. Efficacy of a filter device in the prevention of embolic events during carotid angioplasty and stenting: an ex vivo analysis. J Vasc Surg. 1999; 30: 1034–1044.[CrossRef][Medline] [Order article via Infotrieve]
  14. Topol EJ, Yadav JS. Recognition of the importance of embolization in atherosclerotic vascular disease. Circulation. 2000; 101: 570–580.[Free Full Text]
  15. Al-Mubarek N, Columbo A, Gaines PA, et al. Multicenter evaluation of carotid artery stenting with a filter protection system. J Am Coll Cardiol. 2002; 38: 841–846.
  16. Schluter M, Chevalier B, Seth A, et al. Saphenous vein graft stenting using a novel filter device for distal protection. Am J Cardiol. 2003; 91: 736–739.[CrossRef][Medline] [Order article via Infotrieve]
  17. Reimers B, Corvaja N, Moshiri S, et al. Cerebral protection with filter devices during carotid artery stenting. Circulation. 2001; 104: 12–15.[Abstract/Free Full Text]
  18. Baim DS, Wahr D, George B, et al. Randomized trial of a distal embolic protection device during percutaneous intervention of saphenous vein aorto-coronary bypass grafts. Circulation. 2002; 105: 1285–1290.[Abstract/Free Full Text]
  19. Stone GW, Rogers C, Hermiller J, et al. Randomized comparison of distal protection with a filter-based catheter and a balloon occlusion and aspiration system during percutaneous intervention of diseased saphenous vein aorto-coronary bypass grafts. Circulation. 2003; 108: 548–553.[Abstract/Free Full Text]
  20. Grube E, Schofer J, Webb J, et al, for the Saphenous Vein Graft Angioplasty Free of Emboli (SAFE) Trial Study Group. Evaluation of a balloon occlusion and aspiration system for protection from distal embolization during stenting in saphenous vein grafts. Am J Cardiol. 2002; 89: 941–945.[CrossRef][Medline] [Order article via Infotrieve]
  21. Bourassa MG, Enjalbert M, Campeau L, et al. Progression of atherosclerosis in coronary artery bypass grafts: ten years later. Am J Cardiol. 1984; 53: 102C–107C.[CrossRef][Medline] [Order article via Infotrieve]
  22. Holmes DR Jr, Topol EJ, Califf RM, et al. A multicenter randomized trial of coronary angioplasty versus directional atherectomy for patients with saphenous vein bypass graft lesions. Circulation. 1995; 91: 1966–1974.[Abstract/Free Full Text]
  23. Bhargava B, Kornowski R, Mehran R, et al. Procedural results and intermediate clinical outcomes after multiple saphenous vein graft stenting. J Am Coll Cardiol. 2000; 35: 389–397.[Abstract/Free Full Text]
  24. Hong MK, Mehran R, Dangas G, et al. Creatine kinase-MB enzyme elevation following successful saphenous vein graft intervention is associated with late mortality. Circulation. 1999; 100: 2400–2405.[Abstract/Free Full Text]
  25. Resnic FS, Wainstein M, Lee MKY, et al. No-reflow is an independent predictor of death and myocardial infarction after percutaneous coronary intervention. Am Heart J. 2003; 145; 42–46.[Medline] [Order article via Infotrieve]
  26. Qureshi AI, Luft AR, Janardhan V, et al. Identification of patients at risk for periprocedural neurological deficits associated with carotid angioplasty and stenting. Stroke. 2000; 31: 376–382.[Abstract/Free Full Text]



This article has been cited by other articles:


Home page
J Am Coll Cardiol IntvHome page
D. J. Kereiakes, M. A. Turco, J. Breall, N. Z. Farhat, R. L. Feldman, B. McLaurin, J. J. Popma, L. Mauri, P. Zimetbaum, J. Massaro, et al.
A Novel Filter-Based Distal Embolic Protection Device for Percutaneous Intervention of Saphenous Vein Graft Lesions: Results of the AMEthyst Randomized Controlled Trial
J. Am. Coll. Cardiol. Intv., June 1, 2008; 1(3): 248 - 257.
[Abstract] [Full Text] [PDF]


Home page
CirculationHome page
J. H. Rogers and J. R. Laird
Overview of New Technologies for Lower Extremity Revascularization
Circulation, October 30, 2007; 116(18): 2072 - 2085.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
P. K. Shah
Distal Embolization After Percutaneous Coronary Interventions: Prediction, Prevention, and Relevance
J. Am. Coll. Cardiol., October 23, 2007; 50(17): 1647 - 1648.
[Full Text] [PDF]


Home page
J Am Coll CardiolHome page
T. Kawamoto, H. Okura, Y. Koyama, I. Toda, H. Taguchi, K. Tamita, A. Yamamuro, Y. Yoshimura, Y. Neishi, E. Toyota, et al.
The Relationship Between Coronary Plaque Characteristics and Small Embolic Particles During Coronary Stent Implantation
J. Am. Coll. Cardiol., October 23, 2007; 50(17): 1635 - 1640.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
L. Mauri, D. Cox, J. Hermiller, J. Massaro, J. Wahr, S. W. Tay, M. Jonas, J. J. Popma, J. Pavliska, D. Wahr, et al.
The PROXIMAL Trial: Proximal Protection During Saphenous Vein Graft Intervention Using the Proxis Embolic Protection System: A Randomized, Prospective, Multicenter Clinical Trial
J. Am. Coll. Cardiol., October 9, 2007; 50(15): 1442 - 1449.
[Abstract] [Full Text] [PDF]


Home page
CirculationHome page
L. Mauri, C. Rogers, and D. S. Baim
Devices for Distal Protection During Percutaneous Coronary Revascularization
Circulation, June 6, 2006; 113(22): 2651 - 2656.
[Full Text] [PDF]


Home page
Eur Heart JHome page
U. Limbruno and R. De Caterina
EMERALD, AIMI, and PROMISE: is there still a potential for embolic protection in primary PCI?
Eur. Heart J., May 2, 2006; 27(10): 1139 - 1145.
[Abstract] [Full Text] [PDF]


Home page
Am. J. Neuroradiol.Home page
M.D. Hill, W. Morrish, G. Soulez, A. Nevelsteen, G. Maleux, C. Rogers, K.E. Hauptmann, A. Bonafe, R. Beyar, L. Gruberg, et al.
Multicenter evaluation of a self-expanding carotid stent system with distal protection in the treatment of carotid stenosis.
AJNR Am. J. Neuroradiol., April 1, 2006; 27(4): 759 - 765.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
K. Leineweber, D. Bose, M. Vogelsang, M. Haude, R. Erbel, and G. Heusch
Intense Vasoconstriction in Response to Aspirate From Stented Saphenous Vein Aortocoronary Bypass Grafts
J. Am. Coll. Cardiol., March 7, 2006; 47(5): 981 - 986.
[Abstract] [Full Text] [PDF]


Home page
Eur Heart JHome page
J. Herrmann
Peri-procedural myocardial injury: 2005 update
Eur. Heart J., December 1, 2005; 26(23): 2493 - 2519.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
R. Zahn, T. Ischinger, B. Mark, S. Gass, U. Zeymer, W. Schmalz, K. Haerten, K. E. Hauptmann, E.-R. von Leitner, W. Kasper, et al.
Embolic Protection Devices for Carotid Artery Stenting: Is There a Difference Between Filter and Distal Occlusive Devices?
J. Am. Coll. Cardiol., June 7, 2005; 45(11): 1769 - 1774.
[Abstract] [Full Text] [PDF]


Home page
JAMAHome page
A. Schomig and A. Kastrati
Distal Embolic Protection in Patients With Acute Myocardial Infarction: Attractive Concept But No Evidence of Benefit
JAMA, March 2, 2005; 293(9): 1116 - 1118.
[Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
109/14/1735    most recent
01.CIR.0000124724.14491.6Fv1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow Request Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Rogers, C.
Right arrow Articles by Macon, N. D.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Rogers, C.
Right arrow Articles by Macon, N. D.
Right arrowPubmed/NCBI databases
*Compound via MeSH
*Substance via MeSH
Medline Plus Health Information
*Coronary Artery Bypass Surgery
*Coronary Artery Disease
Related Collections
Right arrow Catheter-based coronary and valvular interventions: other
Right arrow Catheter-based coronary interventions: stents
Right arrow CV surgery: coronary artery disease
Right arrow Acute myocardial infarction
Right arrow By-pass procedures