(Circulation. 2003;108:192.)
© 2003 American Heart Association, Inc.
Clinical Investigation and Reports |
From II Medical Clinic, University Hospital, Mainz, Germany.
Correspondence to PD Dr med Ewald Himmrich, II. Medical Clinic, University Hospital, Langenbeckstraße 1, D-55131 Mainz, Germany. E-mail himmrich{at}2-med.klinik.uni-mainz.de
Received December 13, 2002; revision received April 1, 2003; accepted April 9, 2003.
| Abstract |
|---|
|
|
|---|
Methods and Results All consecutive patients presenting exclusively with pacemaker-induced tachycardias (PITs) were included in a prospective study using a crossover protocol. Patients were randomized to either group 1 (augmentation of the baseline frequency of the pacemaker to 60 bpm) or group 2 (pacemaker turned off) and were followed up for 1 year and then crossed over to the other programming, looking for reoccurrence of PIT. Of 150 consecutive patients, 39 (26%) had PIT, 13 of them exclusively (8.6%). Forty of 1063 analyzed tachyarrhythmias of all the patients were PIT (3%). Before inclusion in the study, the patients had 2.7±0.9 PITs in 11±6.5 months with their pacemakers programmed empirically at 42.3 bpm. During the study phase, no PIT occurred while the pacemaker was turned off, whereas programming to 60 bpm led to the recurrence of PIT in 5 of 6 patients (1.4±0.6 per patient). At the end of the study, 9 patients underwent a prolonged follow-up with their pacemakers turned off, resulting in spontaneous episodes of ventricular tachycardia/fibrillation in 5 patients, but PITs were no longer observed.
Conclusions This crossover protocol proves the potential proarrhythmic effect of pacemaker stimulation in implantable cardioverter-defibrillator patients. Resulting PITs led to clinical symptoms and antitachycardia therapy by the implantable cardioverter-defibrillator. Thus, in patients presenting with PIT but without a pacemaker indication, the pacemaker feature should be turned off, or, alternatively, the longest possible escape interval should be programmed.
Key Words: electrical stimulation tachyarrhythmias pacing cardioverter-defibrillator
| Introduction |
|---|
|
|
|---|
Using a crossover procedure, we identified the first case of a PIT at our hospital in February 1993 and subsequently developed the defining criteria for pacemaker-induced VT/VF as well as the following study design.
The aim of this randomized, crossover study, therefore, was to prove a proarrhythmic effect of right ventricular VVI-backup pacing in patients with an ICD.
| Methods |
|---|
|
|
|---|
Other criteria for enrollment were as follows: (1) documentation of VT/VF in stored electrograms showing the start of the tachycardia, (2) faultless functioning of pacemaker feature and ICD (especially no sensing problems at follow-up), and (3) written informed consent. Exclusion criteria were age <18 years, a recognized pacemaker indication according to American Heart Association/American College of Cardiology guidelines, a myocardial infarction or coronary revascularization (bypass surgery or angioplasty) in the course of the study, or progression to NYHA stage IV heart failure. For the analysis, nonsustained VT/VF episodes and inappropriate shock discharges were not taken into consideration.
Pacemaker and ICD Programming
The pulse amplitude for pacemaker stimulation was set at double the pacing threshold with a pulse width of 0.4 ms. The sensitivity was usually programmed at nominal values (0.28 mV, Guidant and 0.3 mV, Medtronic). The tachycardia detection rate was programmed depending on the patients needs (mean, 156±19 bpm). Duration of detection was set at 1 second for the VF zone and 5 seconds for the VT zone. In the VT zone, burst, scan, and ramp therapy were all admissible.
Study Protocol
The baseline frequency of the pacemaker was initially set at an empirical value of 40 to 50 bpm. After a reproducible PIT phenomenon had been documented (
2 episodes per patient), patients were randomly assigned to 1 of 2 study groups: in group 1, the baseline frequency of the pacemaker was elevated to 60 bpm; in group 2, the pacemakers were switched off. After 12 months, the baseline frequency was reprogrammed in a crossover procedure (Figure 1), and the patients were followed for another 12 months. The end point was defined as the occurrence of a PIT.
|
During the follow-up, all patients were seen in our outpatient clinic every 3 to 4 months for physical check-ups, ICD interrogation, and a function test of the device. The latter evaluated battery status, pacing threshold, pacing and defibrillation lead impedance, and if applicable, frequency of ventricular pacing since the last visit. Intracardial and body surface ECGs were recorded simultaneously to document the morphology of spontaneous and effectively paced ventricular rhythms. Episode data, including event counters and stored electrograms, were interrogated, documented on printout and a patients disk, and subsequently deleted. Each patient underwent annual echocardiography. At the end of the study, patients were followed up for another period of 24 months looking for episodes of spontaneous VT/VF.
Randomized reprogramming of the pacemaker baseline frequency for PIT prevention was approved by the Ethics Commission of Medical Council in Rheinland-Pfalz, Germany.
Implanted Electrodes and ICDs
Patients enrolled in the study received transvenously implanted electrodes from Medtronic or Endotak electrodes from Guidant allowing either unipolar or integrated bipolar pacing. Single-chamber ICDs from Guidant and Medtronic were implanted.
Definition of PIT
An episode of VT was considered to be pacemaker induced if the onset was after a single visible and effective pacemaker stimulus. The stimulus was considered effective if the morphology of the intracardial QRS complex was identical to the morphology of QRS complexes proven to be of effective ventricular stimulation through cross-checking with the surface ECG during previous follow-up visits. In addition, the stimulated ventricular complex was required to show a change of morphology on the intracardial ECG compared with the morphology of the tachycardia and of normal rhythm (sinus rhythm or atrial fibrillation) (Figures 2 and 3
).
|
|
If the pacemaker impulse was not visible (Figure 3), the pause before the onset of tachycardia was required to correspond with the programmed escape interval. Furthermore, the QRS complex after the pause had to show a morphology identical to that of QRS complexes of proven effective ventricular stimulation.
Episodes were not defined as PIT if >1 consecutive effective ventricular pacemaker stimulus preceded the onset of the tachycardia.
Statistical Analysis
The statistical analysis was performed using McNemars test and Students t test.
| Results |
|---|
|
|
|---|
|
Incidence of Ventricular Stimulation
Before randomization, ventricular stimulation was
1% (maximum, 5%) in 7 of 9 patients and <1% in 2 patients with ICDs from Guidant. The average number of ventricular stimuli in patients with devices from Medtronic was 36 117±38 277 per month. Hence, (sporadic) ventricular stimulation in ICD patients without an indication for a pacemaker occurs frequently.
Incidence of PIT Episodes
Using the criteria outlined above, we identified a total of 39 patients (26%) of 150 of our overall patient cohort with appropriate therapy delivery showing a PIT phenomenon. In 26 of these patients, PIT (n=67) occurred in addition to spontaneous episodes of VT/VF (n=158). The 13 patients enrolled in this study (8.6% of 150) showed only PIT; episodes of spontaneous VT/VF were not observed. Of the 1063 analyzed episodes of sustained VT/VF, 40 episodes (3%) were pacemaker-induced.
During the screening period of 11±6.5 months before randomization, 40 episodes of VT/VF occurred in our study group (mean, 2.7±0.92 PITs per patient; minimum 2, maximum 5 episodes) with the baseline frequency of the pacemaker set empirically at 40 bpm in 10 patients and at 50 bpm in 3.
Incidence of PIT After Pacemaker Reprogramming
According to randomization (Figure 1), the baseline frequency of the pacemaker of the 6 patients in group 1 was increased from 42.3±4.4 to 60 bpm. During the following 12 months, 5 of 6 patients continued to experience appropriate therapy delivery. Without exception, all treated tachyarrhythmias were PIT episodes resulting in an average of 1.4±0.6 PITs per patient (minimum 1, maximum 3). In 1 patient, no episodes of VT/VF were seen. This patient never exhibited a ventricular arrhythmia in the total period of observation spanning 36 months after conclusion of the crossover phase. Thus, in this case, it cannot be proved whether the absence of VT/VF was caused solely by increasing the baseline frequency.
No episodes of VT/VF were recorded in group 2 patients, ie, with the bradycardia therapy deactivated.
Five patients in group 1 and 3 patients in group 2 underwent the crossover procedure. One patient in group 1 died of progressive heart failure before crossover. Two patients in group 2 withdrew their consent because of concern about the possibility of shock delivery (hence, pacemakers remained off), another patient was lost to follow-up 11 months after randomization because of a change of address, and 1 patient died after valve surgery 13 months after study enrollment.
None of the 5 patients in group 1 had episodes of sustained ventricular arrhythmia after crossover (pacemaker turned off). In contrast, after crossover (pacemaker baseline frequency reprogrammed at 60 bpm), each of the 3 remaining patients in group 2 showed tachyarrhythmias that were exclusively PIT. During this period, an average of 1.3±0.4 (minimum 1, maximum 2) episodes per patient was observed. In these cases, the pacemaker was turned off for the additional observation period after crossover.
In summary, the incidence of PIT at a pacemaker baseline frequency of 42.3 bpm was 100% (enrollment criterion); at a baseline frequency of 60 bpm, it was reduced to 89% (P>0.05), and the incidence went down to 0% when the pacemaker was switched off (P<0.001 compared with initial pacemaker programming as well as baseline frequency of 60 bpm).
During follow-up, none of the patients with their pacemakers deactivated suffered a syncope that was not a result of a tachyarrhythmia.
Characteristics of PIT Episodes
In all 40 PIT episodes, a pause of
1200 ms was observed at the onset of tachycardia; in 26 cases (65%) after
1 PVC, in 8 cases during bradycardia in atrial fibrillation, and in 6 cases during sinus bradycardia. Of the PIT episodes, 80% (32 episodes) were polymorphic VT or VF, and 20% displayed a monomorphic VT.
Characteristics of Spontaneous Episodes After Conclusion of the Study
After conclusion of the crossover phase, the pacemakers of 8 of the remaining 9 patients were deactivated, and the patients were followed up for another 24 months to identify cases of spontaneous ventricular arrhythmia. In 5 of the patients (55.6%), a total of 26 spontaneous VT/VF episodes were seen by the end of the prolonged follow-up. Of the 26 episodes, 19 (73%) were monomorphic VT; the other tachyarrhythmias were episodes of polymorphic VT or VF. Only 2 of these cases (7.7%) had a pause of >1000 ms at the onset of the tachycardia: 1 case of bradycardia in atrial fibrillation and 1 case of sinus bradycardia. In 14 episodes (53.8%), the tachycardia began suddenly, ie, without a preceding pause (Figure 3C), and in the other 10 cases (38.5%), either a short-long-short or a short-short-(short)-long-short sequence was seen at onset. The average "long interval" in these episodes was 833±131 ms (minimum 520, maximum 961 ms). In 4 patients, no spontaneous VT/VF episodes were seen.
Comparison of Spontaneous Tachycardia and PIT
The intraindividual analysis of spontaneous tachycardia and PIT in 5 patients (Table 2) showed a more frequent occurrence of polymorphic VT or VF in PIT (84%) compared with spontaneous VT/VF (27%; P<0.05). A significantly higher frequency was also shown for the occurrence of a pause after an ectopic ventricular beat, couplets and triplets in cases of PIT.
|
| Discussion |
|---|
|
|
|---|
For the first time, our randomized crossover study with highly selected patients exhibiting only PIT could show that sustained ventricular arrhythmias no longer occurred after the pacemaker feature was deactivated. Reactivating the feature led to a resurgence of exclusively pacemaker-induced tachyarrhythmias. Therefore, it can be asserted that the stimulated impulse was actually responsible for the induction of the VT/VF in these cases and that the arrhythmia did not occur coincidentally with the ventricular stimulation. In our view, this clearly proves a proarrhythmic effect of antibradycardia backup pacing.
The pathophysiology of PIT is still unclear, and it will not be clarified through this study. However, it appears obvious that a pause followed by a ventricular stimulus is to be held responsible. Earlier studies have shown that a sudden pause and a short-long-short sequence cause dispersion of the ventricular refractory period, which is regarded as a prerequisite for ventricular arrhythmia.1115 If the pacemaker feature is switched on, an effective ventricular stimulus after the pause will lead to an ectopic ventricular excitation. This may intensify the ventricular electrical instability to such an extent that an arrhythmia ensues. With a deactivated pacemaker, however, a supraventricular rhythm after the pause would lead to a physiological ventricular depolarization so that the electrical stability is restored.
Nevertheless, it must be assumed that a pause with a consecutive ventricular stimulation is not the sole cause of PIT. The large number of ventricular stimuli per year, as high as 350 000, is several times higher than the PIT incidence of 2 to 5 episodes per year as documented in our study. Therefore, it seems that PIT, like spontaneous VT/VF, can occur only in a setting of additional factors promoting electrical instability (eg, ischemia, electrolyte imbalance, influence of the autonomic nervous system). A pause occurring at this very moment with a consecutive ventricular stimulus could destabilize the ventricle to such an extent that a VT results that would not have occurred spontaneously. The more often a ventricle is paced, the higher is the chance of accidental coincidence of electrical instability and proarrhythmic ventricular stimulation.
Our investigation shows that the incidence of PIT was similar at escape intervals of 1000 and 1500 ms. Thus, the duration of the pause seems to be of minor relevance for the occurrence of PIT (Figures 1 and 2
).
On the basis of our results and the following reasoning, one may presume that it is not the pause but rather the ventricular stimulation that is the responsible proarrhythmic factor. First, a pause >1000 ms at the start of tachycardia was rarely documented during spontaneous VT/VF episodes (7.7%; Table 2). Second, it is very likely that pauses >1000 ms also occurred after the pacemaker feature was switched off. This, however, did not lead to VT/VF during crossover. It is therefore doubtful that a long pause plays a crucial role in the origin of the PIT phenomenon. In an attempt to answer that question, a prospective randomized study (Prevent Study) is currently under way testing special algorithms such as frequency smoothing, which shortens the pause after an ectopic beat or after a sudden drop in frequency.
Practical Suggestions
Recipients of ICDs with documented PIT should have the pacemaker feature switched off when possible (not postshock stimulation). As an alternative, the baseline frequency could be programmed as low as possible or the hysteresis may be programmed at 30 bpm where applicable. Frequency-smoothing algorithms should be programmed only in patients with syncope and a documented sudden drop in frequency.
In patients with implantable pacemakers, the rate of sudden cardiac death is 23% of all deaths, which is much higher than that in comparable cohorts without pacemakers.8 Several pacemaker recipients have structural heart disease, and PITs may occur that lead to lethal tachyarrhythmias in these patients. Therefore, with the results of the MADIT II study in mind, we recommend that for patients with paroxysmal bradycardia and structural heart disease with depressed left ventricular function, the indication for the implantation of an ICD instead of a pacemaker should be discussed.16
Limitations
The study described was performed as a prospective and randomized crossover study but was not designed as a multicenter investigation. Primarily endocardial electrodes that allow for integrated bipolar stimulation were used in this study. Exclusive bipolar stimulation could have altered the incidence of PIT. Pauses not associated with the start of an episode of tachycardia did not appear on stored ICD electrograms. Therefore, the assumption that pauses of duration >1000 to 1500 ms continued to occur after the pacemaker was switched off is speculation. Future investigations may avoid this disadvantage by using subthreshold ventricular stimulation (pseudopacing) and hence prove the existence of pauses.
In our investigation, statistical significance was not achieved for the reduction of the PIT incidence by increment of the pacemaker baseline frequency. The number of patients in the group may not be sufficient to allow for a clear statistical analysis.
This study could not analyze the relationship between the incidence of ventricular stimulation and the incidence of PIT on the basis of characteristics of counteralgorithms (Guidant, Medtronic).
Why PITs occur only in some patients and what the risk factors might be for pacemaker-induced arrhythmias remains unclear.
| References |
|---|
|
|
|---|
2. Khastgir T, Aarons D, Veltri E. Sudden bradyarrhythmic death in patients with the implantable cardioverter-defibrillator: report of two cases. Pacing Clin Electrophysiol. 1991; 14: 395398.[CrossRef][Medline] [Order article via Infotrieve]
3. Reiffel JA, Cormilas J, Zimmermann JM, et al. Drug-device interactions: clinical considerations. Pacing Clin Electrophysiol. 1985; 8: 368373.
4. Roelke M, ONunain S, Osswald S, et al. Ventricular pacing induced ventricular tachycardia in patients with implantable cardioverter defibrillators. Pacing Clin Electrophysiol. 1995; 18: 486491.[CrossRef][Medline] [Order article via Infotrieve]
5. Vlay LC, Clay SC. Pacing induced ventricular fibrillation in internal cardioverter defibrillator patients: a new form of proarrhythmia. Pacing Clin Electrophysiol. 1997; 20: 132133.[CrossRef][Medline] [Order article via Infotrieve]
6. Goldman DS, Levine PA. Pacemaker-mediated polymorphic ventricular tachycardia. Pacing Clin Electrophysiol. 1998; 21: 19931995.[CrossRef][Medline] [Order article via Infotrieve]
7. Karbenn U, Borggrefe M, Breithardt G. Pacemaker-induced ventricular tachycardia in normally functioning ventricular demand pacemakers. Am J Cardiol. 1989; 63: 120122.[CrossRef][Medline] [Order article via Infotrieve]
8. Zehender M, Buchner C, Meinertz T, et al. Prevalence, circumstances, mechanisms, and risk stratification of sudden cardiac death in unipolar single-chamber ventricular pacing. Circulation. 1992; 85: 596605.
9. Callans DJ, Hook BG, Marchlinski FE. Paced beats following single nonsensed complexes in a codependent cardioverter defibrillator and bradycardia pacing system: potential for ventricular tachycardia induction. Pacing Clin Electrophysiol. 1991; 14: 12811287.[CrossRef][Medline] [Order article via Infotrieve]
10. Lefroy DC, Crake T, Davies DW. Ventricular tachycardia: an unusual pacemaker-mediated tachycardia. Br Heart J. 1994; 71: 481483.
11. Cranefeld PF, Aronson RS. Torsade de pointes and other pause-induced ventricular tachycardia: the short-long-short sequence and early after depolarisation. Pacing Clin Electrophysiol. 1988; 11: 670678.[CrossRef][Medline] [Order article via Infotrieve]
12. Denker S, Lehmann M, Mahmud R, et al. Facilitation of ventricular tachycardia induction with abrupt changes in ventricular cycle length. Am J Cardiol. 1984; 53: 508515.[CrossRef][Medline] [Order article via Infotrieve]
13. Gomes JA, Alexopoulus D, Winters SL, et al. The role of silent ischemia, the arrhythmic substrate and the short-long-short sequence in the genesis of sudden cardiac death. J Am Coll Cardiol. 1989; 14: 16181625.[Abstract]
14. Kay GN, Plumb VJ, Areiniegas JG, et al. Torsade de pointes: the long-short initiating sequence and other clinical features observations in 32 patients. J Am Coll Cardiol. 1983; 2: 806817.[Abstract]
15. Luck JC, Minor ST, Mann DE, et al. Effect of bradycardia on dispersion of ventricular refractoriness. Am J Cardiol. 1985; 55: 10091014.[CrossRef][Medline] [Order article via Infotrieve]
16. Moss AJ, Zareba W, Hall J, et al. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med. 2002; 346: 877883.
This article has been cited by other articles:
![]() |
J. Osorio, D. J. Dosdall, R. P. Robichaux Jr, P. B. Tabereaux, and R. E. Ideker In a Swine Model, Chest Compressions Cause Ventricular Capture and, By Means of a Long-Short Sequence, Ventricular Fibrillation Circ Arrhythm Electrophysiol, October 1, 2008; 1(4): 282 - 289. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Tung, P. Zimetbaum, and M. E. Josephson A Critical Appraisal of Implantable Cardioverter-Defibrillator Therapy for the Prevention of Sudden Cardiac Death J. Am. Coll. Cardiol., September 30, 2008; 52(14): 1111 - 1121. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Ghosh, E. K. Rhee, J. N. Avari, P. K. Woodard, and Y. Rudy Cardiac Memory in Patients With Wolff-Parkinson-White Syndrome: Noninvasive Imaging of Activation and Repolarization Before and After Catheter Ablation Circulation, August 26, 2008; 118(9): 907 - 915. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Gardiwal, H. Yu, H. Oswald, U. Luesebrink, A. Ludwig, A. M. Pichlmaier, H. Drexler, and G. Klein Right ventricular pacing is an independent predictor for ventricular tachycardia/ventricular fibrillation occurrence and heart failure events in patients with an implantable cardioverter-defibrillator Europace, March 1, 2008; 10(3): 358 - 363. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. O. Sweeney, L. L. Ruetz, P. Belk, T. J. Mullen, J. W. Johnson, and T. Sheldon Bradycardia Pacing-Induced Short-Long-Short Sequences at the Onset of Ventricular Tachyarrhythmias: A Possible Mechanism of Proarrhythmia? J. Am. Coll. Cardiol., August 14, 2007; 50(7): 614 - 622. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Nagele, S. Hashagen, M. Azizi, S. Behrens, and M.A. Castel Analysis of terminal arrhythmias stored in the memory of pacemakers from patients dying suddenly Europace, June 1, 2007; 9(6): 380 - 384. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Arzt and T. D. Bradley Treatment of Sleep Apnea in Heart Failure Am. J. Respir. Crit. Care Med., June 15, 2006; 173(12): 1300 - 1308. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Biffi, G. Boriani, M. Bertini, P. Silvestri, C. Martignani, and A. Branzi Pacing with capture verification in candidates for resynchronisation therapy: A feasibility study Europace, January 1, 2005; 7(3): 255 - 265. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2003 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |