(Circulation. 2003;108:1990.)
© 2003 American Heart Association, Inc.
Basic Science Reports |
From the Cardiovascular Institute (F.G.S.G.), Mt View, Calif; the Division of Cardiovascular Surgery (J.I.F., T.J.F.), Stanford University Medical Center, Stanford, Calif; Evalve, Inc (J.K.), Redwood City, Calif; the Division of Cardiology (E.F.), University of California-San Francisco Medical Center, San Francisco, Calif; the Department of Cardiothoracic Surgery (M.C.O.), Columbia University, New York, NY; the Division of Cardiology (T.F.), Evanston Northwestern Healthcare, Evanston, Ill; and the Division of Cardiology (P.C.B.), Emory University, Atlanta, Ga.
Correspondence to Frederick St. Goar, MD, Cardiovascular Institute, 2660 Grant Rd, Mt View, CA 94040. E-mail fstgoar{at}aol.com
Received July 30, 2003; revision received August 21, 2003; accepted August 22, 2003.
| Abstract |
|---|
|
|
|---|
Methods and Results Adult pigs (n=14) were anesthetized, and left thoracotomy was performed for epicardial echo imaging. Using femoral vein access, a steerable guide catheter was placed transseptally into the left atrium. An implantable clip designed to grasp and approximate the middle scallops of the anterior and posterior mitral leaflets was introduced through the guide catheter. The clip was opened in the left atrium, advanced through the mitral orifice, and retracted to grasp the leaflet edges. When a functional double-orifice valve was confirmed by echo, the clip was closed to coapt the leaflets and detached from the delivery catheter. Before final clip detachment, echo demonstrated a double orifice in all 14 animals. In 2 studies, the clip released from the anterior mitral leaflet. Retrospective analysis of echo images indicated an incomplete grasp of the anterior leaflet. Immediate postmortem examination revealed that the clip successfully approximated the middle scallops of the anterior and posterior leaflets in all 12 double-orifice studies.
Conclusions This study demonstrates for the first time that an endovascular system can be successfully used to perform the edge-to-edge repair technique in a nondiseased porcine model. This technique is potentially applicable as a percutaneous catheterization laboratory procedure for the treatment of mitral regurgitation in humans.
Key Words: mitral valve regurgitation catheterization
| Introduction |
|---|
|
|
|---|
|
| Methods |
|---|
|
|
|---|
|
Evalve Clip
The Evalve Clip is a 2-armed, soft tissue approximation device that when closed, has an outside diameter of 15 F, and when open, has a grasping span of 20 mm. It is designed to vertically coapt up to 8 mm of leaflet height specifically to replicate the Alfieri surgical approach with regard to width and length of tissue apposition. It is constructed of implant-grade metals and covered with polyester (Figure 2). The clip is designed to promote leaflet-to-leaflet healing around and into the device to maintain a point of permanent leaflet approximation.
Investigational Protocol
Fourteen consecutive studies in pigs ranging in weight from 101 to 135 kg are reported. A committee on animal research approved the protocol, and animals received humane care in compliance with the National Institutes of Health Guide for the Care and Use of Laboratory Animals.
After premedication with ketamine and atropine, endotracheal intubation was performed and general anesthesia achieved with isoflurane. The animal was positioned on the angiography suite table in the right lateral decubitus position. An 18-gauge angiocatheter was placed in the left femoral artery for continuous hemodynamic monitoring. The right femoral vein was exposed by surgical cutdown. Left thoracotomy was performed, the left fifth rib removed, and the pericardium retracted to allow for direct epicardial echocardiographic imaging. A baseline echo study was performed (5- to 7-MHz probe) with collection of 2-dimensional and Doppler data in the standard views, including the short- and long-axis and reverse apical 2- and 4-chamber views imaged through the dome of the atrium. Heparin was administered to maintain an activated clotting time of >600 seconds. A catheter was passed into the right atrium through a longitudinal incision in the femoral vein, and transseptal access was obtained using standard techniques. The steerable guide catheter and dilator were advanced over a soft-tipped 0.35 guidewire, and the tip of the guide was placed 2 cm into the left atrium using a combination of fluoroscopic and echocardiographic guidance. The dilator and wire were subsequently removed. The steerable guide catheter was then positioned so that the distal tip was centered over the mitral valve line of coaptation, with 20 to 24 mm of height above the plane of the mitral annulus. The clip delivery catheter was then introduced through the guide catheter.
Evalve Clip Delivery Protocol
The clip was advanced out the end of the guide catheter under fluoroscopic guidance. In a long-axis echo view, the distal tip of the clip was positioned
1 cm above the mitral valve and the arms opened with the handle control mechanisms. The clip was then rotated in the short-axis view until the arms were perpendicular to the line of leaflet coaptation and the tip of the clip was above the desired point of coaptation of the middle scallops of the anterior and posterior leaflets (Figure 2). With the delivery catheter aligned with the long axis of the ventricle, the clip was advanced without rotation through the mitral orifice to a level just below the free edges of the leaflets. Leaflets were grasped by clip retraction against the leaflets timed to occur during systole. A thorough 2-dimensional and color Doppler echocardiographic evaluation was performed to confirm success of the grasp, including location of the clip, depth of tissue insertion within the clip arms, and absence of mitral insufficiency. If the result was deemed suboptimal, the clip was opened, advanced into the ventricle, and additional grasps were attempted. If major repositioning was indicated, the clip arms were inverted and the clip retracted into the atrium, where manipulations and adjustments could be performed distant from the chordae tendinea.
Once a successful grasp was confirmed, the clip arms were closed under fluoroscopic guidance. Further echocardiographic evaluation was performed to confirm successful creation of a functional double orifice. Until the final detachment steps, tissue could be released and the clip removed.
Evaluation Protocol
After the guide and delivery catheters were withdrawn, a detailed 2-dimensional and Doppler echocardiographic examination was performed. The animal was then euthanized according to the protocol. At postmortem, the left atrium was opened and the mitral valve inspected for evaluation of clip placement. The atrial and ventricular chambers and the inferior vena cava were exposed to look for catheter- or clip-induced damage. The valve was subsequently excised for a more detailed inspection of clip placement and any possible trauma.
| Results |
|---|
|
|
|---|
After the clip was closed, echocardiography confirmed a functional double orifice without evidence of MR in all 14 pigs. After completely detaching the clip from the delivery catheter, a detailed echocardiographic study confirmed a functional double-orifice valve in 12 animals (Figures 3 and 4
). Left ventricular fractional shortening remained normal, and there was no Doppler evidence of mitral stenosis or regurgitation. No hemodynamic instability was encountered during clip deployment in any of the studies.
|
|
In 2 animals, once the clip was released from the delivery catheter, it was noted that the clip was no longer attached to the anterior leaflet. On retrospective analysis of the fluoroscopic and echocardiographic images, it was observed that in both studies, the clip arms had not been fully opened to the correct position before grasping, and the partial grasp could be appreciated echocardiographically before detachment.
Duration of the studies, once transseptal access had been obtained, ranged from 45 to 90 minutes. Subsequent postmortem examination of the 12 pigs with a double orifice showed the desired central placement of the clip perpendicular to the line of leaflet coaptation. No significant inferior vena cava, right atrial, atrial septal, left atrial, leaflet, left ventricle, or chordal trauma was observed in any animal. This included the studies that required inversion of the clip for repositioning as well as the 2 in which the clip did not retain one leaflet. The clip successfully approximated the middle scallops of the anterior and posterior leaflets and was deployed perpendicular to the line of coaptation, creating a functional double-orifice valve.
| Discussion |
|---|
|
|
|---|
Improved postoperative functional status and overall freedom from reoperation for surgical edge-to-edge repair has been reported to be 95% at 6 years.10 This durability is not surprising because the mathematically modeled stresses and net force on the point of mitral valve coaptation are minimal.13 Systolic forces applied to the valve by the left ventricle at the coaptation site are not only parallel to the annulus, but the force on each leaflet is symmetric and opposite the force on the opposing leaflet. This favors minimal systolic stress where coaptation occurs.14 The greatest stress on the leading edge of the leaflets and point of apposition occurs in diastole when the transvalvular gradient is typically only several millimeters of mercury. The surgical edge-to-edge repair in clinical series has not resulted in mitral stenosis.15
The simplicity and proven efficacy of edge-to-edge repair make it attractive for an endovascular method of repairing either the structurally or functionally deficient mitral valve. The present study demonstrates that a mitral valve edge-to-edge repair can be successfully performed on the beating heart in an animal model without cardiopulmonary bypass using a catheter-based endovascular system. The clip was successfully deployed with successful approximation of the middle scallops of the anterior and posterior leaflets in the majority (12 of 14; 86%) of study animals. Retrospective analysis of the 2 cases in which the anterior leaflet was not well grasped revealed correctable deficiencies in technique, including improper clip arm opening and improper analysis of echocardiographic images. The porcine model provides a challenging model for demonstrating functionality of the endovascular CVRS for several reasons. The pig mitral valve leaflets are thinner than human leaflets, providing a potentially more rigorous test of the atraumatic features of the devices and procedure. In addition, the porcine interatrial septum is a diminutive structure, and the pigs left atrium provides limited space to manipulate the catheters, especially in comparison to the dilated left atrium associated with human MR. Through experience, it became evident that to perform a successful grasp in this model, the distal tip of the guide catheter needed to be located
2 cm above the plane of the annulus. The manipulations required to establish this height included crossing the septum at least 15 mm above the annular plane. It was also learned that to predictably grasp the leaflets, the clip arms should be open to at least 120°, and the center axis of the clip-delivery catheter needed to be positioned relatively perpendicular to the valve plane and in line with the long axis of the ventricle. Significant deviation in an anterior, posterior, medial, or lateral direction produced a suboptimal grasp. This deviation was appreciated with echocardiography and, given the design features of the catheter and the clip that allow for repositioning, easily corrected.
The present study is limited to the short-term evaluation of the endovascular edge-to-edge valve repair system. Long-term animal data are necessary to assess long-term device performance. Extended animal data (up to 12 months) are being analyzed. The clip system has also been evaluated in an animal model of MR for a design conformation and a better appreciation of its applicability and effectiveness. Although extrapolation to human pathology is difficult, it is possible that the clip system could have limited value in subsets of patients with extreme pathology including rheumatic disease, a ruptured papillary muscle, or extreme mitral annular dilatation.
This study introduces the first technology for endovascular edge-to-edge mitral valve repair. The safety and efficacy of the described system in humans experiencing MR remain to be proven; however, the success of placing the clip in a short-term animal model provides encouraging support for a human clinical trial. As a percutaneous intervention in the catheterization laboratory, the technique offers the opportunity for substantial reduction in the morbidity and trauma associated with mitral valve surgery. The technology could also provide a therapeutic alternative for patients with significant MR and advanced left ventricular dysfunction who would not survive the stress of a surgical intervention.
| Acknowledgments |
|---|
| Footnotes |
|---|
Drs St. Goar, Fann, Oz, Fogarty, and Block have equity in and/or a consulting relationship with Evalve, Inc. Dr Komtebedde is an employee of Evalve, Inc.
| References |
|---|
|
|
|---|
2. Lawrie GM. Mitral valve repair vs replacement: current recommendations and long-term results. Cardiol Clin. 1998; 16: 437448.[CrossRef][Medline] [Order article via Infotrieve]
3. Enriquez-Sarano M, Schaff HV, Orszulak TA. Valve repair improves the outcome of surgery for mitral regurgitation: a multivariate analysis. Circulation. 1995; 91: 10221028.
4. The Society of Thoracic Surgeons. U.S. Cardiac Surgery Database. Mitral valve repair and replacement patients: incidence of complications summary. 1997. Available at: http://www.sts.org/file/STSNationalDataBase. Accessed July 2003.
5. Fann JI, Ingels NB, Miller C. Pathophysiology of mitral valve disease and operative indications. In: Edmunds LH, ed. Cardiac Surgery in the Adult. New York: McGraw-Hill; 1997: 959987.
6. Galloway AC, Grossi EA, Bizekis CE, et al. Evolving techniques for mitral valve reconstruction. Ann Surg. 2002; 3: 288294.
7. Alfieri O, Maisano F, DeBonis M, et al. The edge-to-edge technique in mitral valve repair: a simple solution for complex problems. J Thorac Cardiovasc Surg. 2001; 122: 674681.
8. Maisano F, Schreuder JJ, Oppizzi M, et al. The double orifice technique as a standardized approach to treat mitral regurgitation due to severe myxomatous disease: surgical technique. Eur J Cardiothorac Surg. 2000; 17: 201215.
9. Umana JP, Salehizadeh B, DeRose JJ, et al. Bow-tie mitral valve repair: an adjuvant technique for ischemic mitral regurgitation. Ann Thorac Surg. 1998; 66: 16401646.
10. Maisano F, Torracca L, Oppizzi M, et al. The edge-to-edge technique: a simplified method to correct mitral insufficiency. Eur J Cardiothorac Surg. 1998; 13: 240245.
11. Umana JP, DeRose J, Choudhri A, et al. Bow-tie mitral valve repair successfully addresses subvalvular dysfunction in ischemic mitral regurgitation. Surg Forum. 1997; 48: 279280.
12. Timek TA, Nielsen SL, Liang D, et al. Edge-to-edge mitral repair: gradients and three- dimensional annular dynamics in vivo during inotropic stimulation. Eur J Cardiothorac Surg. 2001; 19: 431437.
13. Arts T, Meerbaum S, Reneman R, et al. Stresses in the closed mitral valve: a model study. J Biomech. 1983; 16: 539527.[CrossRef][Medline] [Order article via Infotrieve]
14. Votta E, Maisano F, Soncini M, et al. 3-D computational analysis of the stress distribution on the leaflets after edge-to-edge repair of mitral regurgitation. J Heart Valve Dis. 2002; 11: 810822.[Medline] [Order article via Infotrieve]
15. Agricola E, Maisano F, Oppizzi M, et al. Mitral valve reserve in double-orifice technique: an exercise echocardiographic study. J Heart Valve Dis. 2002; 11: 637643.[Medline] [Order article via Infotrieve]
This article has been cited by other articles:
![]() |
M.J. Swaans, B.J.L. Van den Branden, J.A.S. Van der Heyden, M.C. Post, B.J.W.M. Rensing, F.D. Eefting, H.W.M. Plokker, and W. Jaarsma Three-dimensional transoesophageal echocardiography in a patient undergoing percutaneous mitral valve repair using the edge-to-edge clip technique Eur J Echocardiogr, December 1, 2009; 10(8): 982 - 983. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Feldman, S. Kar, M. Rinaldi, P. Fail, J. Hermiller, R. Smalling, P. L. Whitlow, W. Gray, R. Low, H. C. Herrmann, et al. Percutaneous mitral repair with the MitraClip system: safety and midterm durability in the initial EVEREST (Endovascular Valve Edge-to-Edge REpair Study) Cohort. J. Am. Coll. Cardiol., August 18, 2009; 54(8): 686 - 694. [Abstract] [Full Text] [PDF] |
||||
![]() |
J.-H. Kim, O. Kocaturk, C. Ozturk, A. Z. Faranesh, M. Sonmez, S. Sampath, C. E. Saikus, A. H. Kim, V. K. Raman, J. A. Derbyshire, et al. Mitral cerclage annuloplasty, a novel transcatheter treatment for secondary mitral valve regurgitation: initial results in swine. J. Am. Coll. Cardiol., August 11, 2009; 54(7): 638 - 651. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Piazza, A. Asgar, R. Ibrahim, and R. Bonan Transcatheter Mitral and Pulmonary Valve Therapy J. Am. Coll. Cardiol., May 19, 2009; 53(20): 1837 - 1851. [Abstract] [Full Text] [PDF] |
||||
![]() |
R D Christofferson, S R Kapadia, V Rajagopal, and E M Tuzcu Emerging transcatheter therapies for aortic and mitral disease Heart, January 15, 2009; 95(2): 148 - 155. [Full Text] [PDF] |
||||
![]() |
A. Plass, I. Valenta, O. Gaemperli, P. Kaufmann, H. Alkadhi, G. Zund, J. Grunenfelder, and M. Genoni Assessment of coronary sinus anatomy between normal and insufficient mitral valves by multi-slice computertomography for mitral annuloplasty device implantation Eur. J. Cardiothorac. Surg., April 1, 2008; 33(4): 583 - 589. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. Davidson and D. S. Baim Percutaneous Catheter-Based Mitral Valve Repair Card. Surg. Adult, January 1, 2008; 3(2008): 1101 - 1108. [Full Text] |
||||
![]() |
T. Feldman and M. B. Leon Prospects for Percutaneous Valve Therapies Circulation, December 11, 2007; 116(24): 2866 - 2877. [Full Text] [PDF] |
||||
![]() |
T. A. Timek, S. L. Nielsen, D. T. Lai, D. Liang, G. T. Daughters, N. B. Ingels Jr, and D. C. Miller Effect of Chronotropy and Inotropy on Stitch Tension in the Edge-to-Edge Mitral Repair Circulation, September 11, 2007; 116(11_suppl): I-276 - I-281. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Coats and P. Bonhoeffer NEW PERCUTANEOUS TREATMENTS FOR VALVE DISEASE Heart, May 1, 2007; 93(5): 639 - 644. [Full Text] [PDF] |
||||
![]() |
J. H. Rogers, J. A. Macoviak, D. A. Rahdert, P. A. Takeda, I. F. Palacios, and R. I. Low Percutaneous Septal Sinus Shortening: A Novel Procedure for the Treatment of Functional Mitral Regurgitation Circulation, May 16, 2006; 113(19): 2329 - 2334. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. K. Mishra, S. Mittal, P. Jaguri, and N. Trehan Coapsys Mitral Annuloplasty for Chronic Functional Ischemic Mitral Regurgitation: 1-Year Results Ann. Thorac. Surg., January 1, 2006; 81(1): 42 - 46. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Feldman, H. S. Wasserman, H. C. Herrmann, W. Gray, P. C. Block, P. Whitlow, F. St. Goar, L. Rodriguez, F. Silvestry, A. Schwartz, et al. Percutaneous Mitral Valve Repair Using the Edge-to-Edge Technique: Six-Month Results of the EVEREST Phase I Clinical Trial J. Am. Coll. Cardiol., December 6, 2005; 46(11): 2134 - 2140. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. C. Dang, M. S. Aboodi, T. Sakaguchi, H. S. Wasserman, M. Argenziano, D. M. Cosgrove, T. K. Rosengart, T. Feldman, P. C. Block, and M. C. Oz Surgical Revision After Percutaneous Mitral Valve Repair With a Clip: Initial Multicenter Experience Ann. Thorac. Surg., December 1, 2005; 80(6): 2338 - 2342. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Boudjemline, G. Agnoletti, D. Bonnet, L. Behr, N. Borenstein, D. Sidi, and P. Bonhoeffer Steps Toward the Percutaneous Replacement of Atrioventricular Valves: An Experimental Study J. Am. Coll. Cardiol., July 19, 2005; 46(2): 360 - 365. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. C. Block Percutaneous Mitral Valve Repair: Are They Changing the Guard? Circulation, May 3, 2005; 111(17): 2154 - 2156. [Full Text] [PDF] |
||||
![]() |
J. E. Sousa, M. A. Costa, E. M. Tuzcu, J. S. Yadav, and S. Ellis New Frontiers in Interventional Cardiology Circulation, February 8, 2005; 111(5): 671 - 681. [Full Text] [PDF] |
||||
![]() |
S. H. Rahimtoola The year in valvular heart disease J. Am. Coll. Cardiol., January 4, 2005; 45(1): 111 - 122. [Full Text] [PDF] |
||||
![]() |
G. Lutter, R. Ardehali, J. Cremer, and P. Bonhoeffer Percutaneous Valve Replacement: Current State and Future Prospects Ann. Thorac. Surg., December 1, 2004; 78(6): 2199 - 2206. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. V. Maniu, J. B. Patel, D. G. Reuter, D. M. Meyer, W. D. Edwards, C. S. Rihal, and M. M. Redfield Acute and chronic reduction of functional mitral regurgitation in experimental heart failure by percutaneous mitral annuloplasty J. Am. Coll. Cardiol., October 19, 2004; 44(8): 1652 - 1661. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Fukamachi, Z. B. Popovic, M. Inoue, K. Doi, S. Schenk, Y. Ootaki, M. W. Kopcak Jr., and P. M. McCarthy Changes in mitral annular and left ventricular dimensions and left ventricular pressure-volume relations after off-pump treatment of mitral regurgitation with the Coapsys device Eur. J. Cardiothorac. Surg., March 1, 2004; 25(3): 352 - 357. [Abstract] [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2003 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |