Circulation. 2003;108:1355-1361
Published online before print August 25, 2003, doi: 10.1161/01.CIR.0000087603.97036.C2
This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
108/11/1355    most recent
01.CIR.0000087603.97036.C2v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Beerbaum, P.
Right arrow Articles by Meyer, H.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Beerbaum, P.
Right arrow Articles by Meyer, H.
Related Collections
Right arrow Cardiovascular imaging agents/Techniques
Right arrow Imaging
Right arrow CT and MRI
Right arrow Pediatric and congenital heart disease, including cardiovascular surgery

(Circulation. 2003;108:1355.)
© 2003 American Heart Association, Inc.


Clinical Investigation and Reports

Rapid Left-to-Right Shunt Quantification in Children by Phase-Contrast Magnetic Resonance Imaging Combined With Sensitivity Encoding (SENSE)

Philipp Beerbaum, MD; Hermann Körperich, PhD; Jürgen Gieseke, MSc; Peter Barth, MSc; Matthias Peuster, MD; Hans Meyer, MD

From Klinik für Angeborene Herzfehler, Herz- und Diabeteszentrum, Nordrhein-Westfalen, Ruhr-Universität Bochum, Germany, and Philips Medical Systems (J.G.), Best, The Netherlands.

Correspondence to Philipp Beerbaum, MD, Herz- und Diabeteszentrum, Ruhr-Universität Bochum, Georgstrasse 11, D-32545 Bad Oeynhausen, Germany. E-mail pbeerbaum{at}hdz-nrw.de

Received January 27, 2003; revision received June 30, 2003; accepted July 2, 2003.


*    Abstract
up arrowTop
*Abstract
down arrowIntroduction
down arrowMethods
down arrowResults
down arrowDiscussion
down arrowConclusions
down arrowReferences
 
Background— Parallel imaging by sensitivity encoding (SENSE) may considerably reduce scan time in MRI. For rapid flow quantification in children with congenital heart disease, we evaluated phase-contrast MRI (PC-MRI) techniques combined with SENSE.

Methods and Results— In 22 pediatric patients (mean age, 7.2±6.2 years) with cardiac left-to-right shunt, blood flow rate in the pulmonary artery (Qp) and ascending aorta (Qs) and flow ratio Qp/Qs were determined by PC-MRI with SENSE reduction-factor 2 and 3 (SF-2 and SF-3). Additionally, we used PC-MRI with higher spatial in-plane resolution (1.6x2.1 versus 2.3x3.1 mm) with and without SF-3. Results were compared with a recently validated standard PC-MRI protocol and tested in vitro using a pulsatile flow phantom. Reduction of signal averages from 2 to 1 and application of SENSE accelerated flow measurements by a factor of 3.5 (5.2) using PC-MRI with SF-2 (SF-3) compared with standard PC-MRI. For blood flow rate through the pulmonary artery and aorta, as well as for the Qp/Qs ratio we found negligible differences of ±3%, lower limits of agreement (mean±2 SD) of -7% to -18%, and upper limits of agreement (mean±2 SD) of +3 to +24%, demonstrating good agreement with standard PC-MRI. Mean Qp/Qs ratio by standard PC-MRI was 1.69±0.45 (range, 1.27 to 2.79). Interobserver variability was low, and high accuracy was confirmed in vitro for all protocols.

Conclusions— PC-MRI for flow quantitation may be combined with SENSE to achieve a substantive reduction of scanning time. In children with left-to-right shunt, Qp/Qs quantification is possible by PC-MRI+SF-3 in <60 seconds. Use of higher in-plane resolution did not improve measurement results.


Key Words: heart defects, congenital • pediatrics • magnetic resonance imaging • shunts • imaging


*    Introduction
up arrowTop
up arrowAbstract
*Introduction
down arrowMethods
down arrowResults
down arrowDiscussion
down arrowConclusions
down arrowReferences
 
In pediatric patients, noninvasive assessment or exclusion of cardiac left-to-right shunts by use of phase-contrast cine MRI (PC-MRI) has been demonstrated to be reliable and clinically useful.1–3 Thus, hemodynamic information may be added to a morphological cardiac MR investigation. In the clinical setting, however, repeated flow measurements3 at multiple sites4 may be required. Because each measurement usually takes 2.5 to 5 minutes with standard PC-MRI protocols2,3 depending on heart rate and imaging parameter settings, a thorough hemodynamic MR investigation may be time consuming. Additionally, variations of heart rate5 and respiration rate6 may introduce artifacts. Therefore, a shorter acquisition time would be desirable. The sensitivity-encoding (SENSE) technique uses different single-coil sensitivities of a multi-element phased-array surface coil to spatially encode the MR signal independently from conventional gradient encoding. Thus, Fourier imaging may be accelerated by undersampling of lines in k-space. The resulting foldover is unfolded mathematically using the encoding information available from the surface-coil sensitivities.7 The purpose of this study was to evaluate the reliability of rapid PC-MRI techniques at different in-plane resolutions using the SENSE approach. In pediatric patients with cardiac left-to-right shunt, measurements in the ascending aorta and pulmonary artery were compared with a conventional retrospectively gated PC-MRI sequence that was recently validated in children.3


*    Methods
up arrowTop
up arrowAbstract
up arrowIntroduction
*Methods
down arrowResults
down arrowDiscussion
down arrowConclusions
down arrowReferences
 
Study Population
From July 2001 to February 2002, we enrolled into the study 25 consecutive children with atrial or ventricular level left-to-right shunt clinically referred for additional diagnostic evaluation by MRI. Of these, 3 patients had no shunt as revealed by MRI and were excluded from analysis. The remaining 22 children (mean age, 7.2±6.2 years; median age, 4.6 years; range, 1 to 17 years; 12 girls) formed the study population. Twelve had secundum atrial septal defect (ASD II), 2 had partial anomalous pulmonary venous return (PAPVR), 1 had ASD-II with PAPVR, 3 had superior sinus-venosus defect with PAPVR, 1 had primum ASD, and 3 had ventricular septal defect. All patients were in sinus rhythm. The study was approved by the institutional review committee and informed, written consent was obtained from parents or caretakers.

Study Design
Each patient underwent MRI to measure through-plane flow in the ascending aorta and pulmonary artery (Figure 1). A standard PC-MRI measurement (reference PC-MRI, Table 1) was repeated twice in each location to serve as a clinical reference method3 and to determine repeatability (ie, precision, Figure 2). The mean value of both measurements was compared with 4 research PC-MRI pulse sequences (see below, Table 1) that were not repeated for the sake of total imaging time. In 6 randomly selected children, reference and research PC-MRI flow measurements were independently reanalyzed by 2 experienced observers to determine interobserver variability. Sedation was performed with midazolam and thiopental intravenously when necessary. Blood pressure, oxygen saturation, heart rate, and respiratory rate were monitored continuously.



View larger version (118K):
[in this window]
[in a new window]
 
Figure 1. PC-MRI for determination of flow rate in the pulmonary artery. Top, Plan scans (A, sagittal view; B, transverse view) for flow quantification in the pulmonary artery (PA, arrows, C and D) used segmented k-space turbo field-echo (2 shots, repetition time 9.5 ms, echo time 2.4 ms, flip angle 20 degrees, field-of-view 250 mm2, matrix 128x256, 6-mm slice thickness). Middle, Standard phase-contrast MRI (without SENSE), used as reference. In phase images (D), flow information is encoded through-plane for each pixel in the cross-sectional area (arrow, D). C, Corresponding magnitude image displaying anatomical information. Bottom, Magnitude (E) and phase images (F) by PC-MRI with SENSE reduction-factor 3 in the same patient (Table 1).


View this table:
[in this window]
[in a new window]
 
TABLE 1. Scan Protocols for Reference PC-MRI and Research PC-MRI With SENSE*



View larger version (24K):
[in this window]
[in a new window]
 
Figure 2. Flow-vs-time curves. Mean flow volumes at multiple instants (20 phases) during an average cardiac cycle yield a flow-vs-time curve. The area under the curve represents the stroke volumes of the pulmonary artery (74.4 and 74.9 mL, respectively; mean 74.6 mL) and the ascending aorta (38.7 and 41.4 mL, respectively; mean 40.1 mL). Two flow measurements were obtained with the reference PC-MRI technique in each vessel to assess repeatability (precision); both flow curves are given in this figure. For each SENSE or high-resolution scan, one measurement was obtained (not shown). The shunt is the difference between pulmonary and aortic flow (ie, 34.5 mL per cardiac cycle).

MRI Technique
All examinations were performed on a 1.5 Tesla whole-body MR scanner (Philips, ACS-NT, maximum gradient performance, 30-mT/m amplitude; slew rate, 150 T/m per s). After the survey scans, a SENSE reference scan was performed, as follows: fast field-echo, coronal orientation, repetition time 8.0 ms, echo time 0.6 ms, flip angle 7 degrees, voxel size 16.5x16.5x18 mm, 8 signal averages, water-fat-shift 0.06 pixels, free-breathing, and scan duration <1 minute. The conventional standard PC-MRI pulse sequence for both in vivo and in vitro flow measurements (reference PC-MRI) was based on a protocol recently validated in children with left-to-right shunt3 (Table 1) and served as reference method in our study. The 4 research PC-MRI protocols included a combination of standard PC-MRI with SENSE reduction-factor (SF) of 2 (PC-MRI+SF-2), SF of 3 (PC-MRI+SF-3), and a PC-MRI protocol with higher spatial resolution with SF of 3 (PC-MRIHR+SF-3) and without SENSE (PC-MRIHR). In-plane resolution was {approx}2.3x3.1 mm2 for PC-MRI with or without SENSE and 1.6x2.1 mm2 for PC-MRIHR with or without SENSE, respectively. In PC-MRI techniques using SENSE, no rectangular field of view (FOV) was used to avoid backfolding because of the SENSE reconstruction algorithm.7 All PC-MRI measurements were obtained under free-breathing conditions and used retrospective vector-ECG8 gating to include end-diastole flow, repetition time 15 ms, {approx}20 to 25 reconstructed heart phases, slice thickness 6 mm, flip-angle 30 degrees, and velocity-encoded value 200 cm/s. The body coil was used for signal transmission, and a 5-element phased-array surface (cardiac) coil was used for signal detection, with 3 elements placed in the back and 2 on the chest (patients positioned supine and head-first in the scanner). All flow measurements used a phase-correction algorithm provided by the manufacturer.

In-Vitro Validation
A polyvinylchloride tube with an inner diameter of 12.7 mm and a wall thickness of 2.4 mm was embedded in a glass tube (diameter 100 mm, length 340 mm) filled with copper sulfate solution (3 mmol). To approximate blood viscosity,9 the flow-phantom was filled with a mixture of distilled water and glycerol (60/40 vol/vol). The flow-phantom was placed parallel to the bore in the isocenter of the magnet. The cardiac surface coil was used for signal acquisition. The tubing was connected to a 60-mL artificial left ventricular system (Medos HIA VAD-System II) to deliver pulsatile flow. Flow pump and MR scanner were simultaneously triggered by an external ECG simulator. With a constant heart rate of 84 bpm, 6 different flow rates (ie, 497, 854, 1127, 1401, 1827, and 2219 mL/min) were achieved by appropriate adjustment of the systolic and diastolic pressures of the ventricular assist device. Flow rate was determined 3 times before and after the MR experiments by stopwatch/graduated cylinder to obtain the reference (true) values. All in vitro MR flow measurements were performed through-plane with velocity-encoding value of 150 cm/s and were repeated 3 times.

MR Image Analysis
Data analysis was performed offline on a computer workstation using a computer algorithm for semiautomatic vessel border detection developed by one of the authors (P.B.)3 to accelerate image analysis and optimize measurement reproducibility.10 In all standard and research PC-MRI flow measurements of 1 subject (5 years, ASD), we determined the signal-to-noise ratio (SNR) in magnitude images11 and the velocity-to-noise ratio (VNR) in phase images12 in the pulmonary artery and aorta (maximum flow, systole). The formula used to calculate SNR was as follows: equation


and the formula for VNR was as follows: equation


where SImean is signal intensity, SDnoise is standard deviation of noise, Venc is the velocity encoding value, and SDv is the standard deviation of velocity noise.

Statistical Analysis
All data are expressed as mean±SD. In vitro results were analyzed by 2-variable linear regression analysis. The analysis of Bland and Altman13 was used to determine PC-MRI interobserver variability and to evaluate the agreement between the different PC-MRI techniques and the mean of both reference PC-MRI measurements for flow rate in the pulmonary artery (Qp) and aorta (Qs) and flow ratio Qp/Qs (Figure 3).



View larger version (15K):
[in this window]
[in a new window]
 
Figure 3. Comparison of standard PC-MRI with PC-MRI+SF-2. A, Plot of aortic stroke volumes by standard (reference) PC-MRI vs aortic stroke volumes by PC-MRI+SF-2. B, Bland-Altman analysis of agreement for blood flow rate in the aorta. Plot of difference against mean for aortic stroke volumes by standard (reference) PC-MRI vs aortic stroke volumes by PC-MRI+SF-2, data log-transformed. Upper and lower limits of agreement, mean±2 SD.


*    Results
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
*Results
down arrowDiscussion
down arrowConclusions
down arrowReferences
 
In Vivo Results
PC-MRI studies were completed within 15 to 20 minutes. Sedation for MRI was well tolerated. The mean heart rate was 101±20/min. Scan time was reduced to 28% with PC-MRI+SF-2 and to {approx}19% with PC-MRI+SF-3 compared with the standard PC-MRI protocol and to 18% with PC-MRIHR+SF-3 compared with PC-MRIHR without SENSE, respectively (Table 1). Thus, scanning was accelerated by a factor of 3.5, 5.2, and 5.5 for PC-MRI+SF2, PC-MRI+SF-3, and PC-MRIHR+SF-3, respectively. This was achieved by reduction of signal averages and by application of SENSE (Table 1).

Comparison of Reference With Research PC-MRI in Children
In our pediatric study population with cardiac left-to-right shunt, the mean Qp/Qs was 1.69 (SD±0.45), as assessed by reference PC-MRI. The Bland-Altman analysis was applied to the log-transformed data,13 because differences increased linearly with mean stroke volumes and Qp/Qs values. Estimation of precision of the limits of agreement (defined as mean±2 SD) was based on calculation of 95% CI. A mean value of 1.0 after anti-log transformation (dimensionless ratio) is expected in the case of no difference between 2 tested methods. All calculated values are detailed in Table 2 for pulmonary artery (Qp), ascending aorta (Qs), and the flow ratio Qp/Qs. For through-plane flow measurements in the pulmonary artery (Figure 1) and ascending aorta and for the Qp/Qs ratio, we found a negligible difference of ±3% (mean, 0.97 to 1.03) between reference PC-MRI and all research PC-MRI techniques. Upper limits of agreement were in the range of 1.03 to 1.16, and lower limits of agreement were between 0.86 and 0.93, respectively. Thus, reference and research PC-MRI sequences assessing cardiac stroke volumes and Qp/Qs may differ by 3% to 16% above and 8% to 14% below in 95% of the cases, demonstrating good agreement between reference and research PC-MRI. With PC-MRI+SF-3 for assessing Qp/Qs, we observed slightly more scatter (Table 2), with upper and lower limits of agreement of 1.24 and 0.82 (95% CI, 1.14 to 1.35 and 0.75 to 0.89, respectively), but agreement was excellent (mean 1.01). Only little scatter was present with high-resolution PC-MRI+SF-3, where upper and lower limits of agreement were 1.12 and 0.89 (95% CI, 1.06 to 1.17 and 0.84 to 0.93, respectively).


View this table:
[in this window]
[in a new window]
 
TABLE 2. Children With Left-to-Right Shunt: Comparison of Standard PC-MRI (Reference) with Research PC-MRI Using Bland-Altman Analysis of Agreement (Data Log-Transformed*)

Repeatability and Interobserver Variability in Children
Repeatability (precision or absence of random error) of the 2 reference PC-MRI flow measurements was excellent for the pulmonary artery (aorta), with a mean deviation of 2.3±1.7% (1.9±2.1%). Reevaluation of standard and research PC-MRI flow measurements by 2 observers (Bland-Altman analysis, data log-transformed) revealed a negligible mean difference of 0.99 to 1.01, with limits of agreement ranging from 0.93 to 1.10 (mean±2 SD), demonstrating low interobserver variability.

Accuracy and Precision In Vitro
Table 3 summarizes the in vitro flow measurements by timed collection (stopwatch/graduated cylinder=true values) and by reference PC-MRI, PC-MRI+SF-2, PC-MRI+SF-3, and PC-MRIHR without SENSE and PC-MRIHR+SF-3. The flow values in Table 3 represent mean values of repeated measurements, which had a maximum deviation of ±2%, indicating high repeatability in vitro (ie, measurement precision) for all PC-MRI techniques. Results of linear regression analysis also presented in Table 3 demonstrate good correlation between true values and reference PC-MRI, PC-MRI with SF-2, and PC-MRI with SF-3, indicating high accuracy. However, we observed a small but statistically significant underestimation of flow rate using PC-MRIHR, both with and without SENSE.


View this table:
[in this window]
[in a new window]
 
TABLE 3. Accuracy In Vitro: Reference vs Research PC-MRI Protocols

Signal-to-Noise Ratio and Velocity-to-Noise Ratio
SNR and VNR values11,12 are expressed as percentages of the standard (reference) PC-MRI protocol with cardiac surface coil for signal acquisition (100%). In the aorta (pulmonary artery), SNR was reduced to 47% (43%) with PC-MRI+SF-2, to 26% (26%) with PC-MRI+SF-3, and to 20% (15%) with PC-MRIHR+SF-3. When the standard PC-MRI was used with body coil instead of cardiac coil, SNR was reduced to 29% (37%).

In contrast, in the aorta (pulmonary artery), VNR was only reduced to 95% (88%) with PC-MRI+SF-2, to 95% (82%) with PC-MRI+SF-3, and to 88% (74%) with PC-MRIHR+SF-3. When the standard PC-MRI was used with body coil instead of cardiac coil, VNR was reduced to 95% (96%).


*    Discussion
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
up arrowResults
*Discussion
down arrowConclusions
down arrowReferences
 
In the SENSE approach,7 an array of multiple, simultaneously operated receiver coils for signal acquisition (eg, cardiac 5-element phased-array surface coil) is used to spatially encode the MR signal as a function of the sensitivity of each coil element. This allows to reduce the number of phase-encoding steps by increasing the distance of readout lines in k-space (ie, parallel imaging) and thus accelerate acquisition, but at the expense of a lower SNR.7 Therefore, combination with SENSE is only recommended when SNR is not of critical importance for diagnostic image quality.

This study demonstrates that standard PC-MRI under free-breathing conditions may be favorably combined with SENSE. Quantitative flow measurements can be accelerated by a factor of 3.5 to 5 by reduction of signal averages (from 2 to 1) and application of SENSE. The corresponding SNR reduction was no limiting factor. Therefore, in children with congenital heart disease, left-to right shunt may be quantified reliably in less than 1 minute. The considerable time benefit is likely to be generally available in hemodynamic evaluation of congenital heart disease. Other possible applications include quantification of residual pulmonary valve regurgitation after surgery for Tetralogy of Fallot14 and the quantification of flows in the venae cavae and branch pulmonary arteries after Fontan palliation in patients with single-ventricle physiology.4 Thus, either an examination can be shortened (reduced duration of sedation/anesthesia in children or increased patient throughput) or more comprehensive flow information may be collected in a given amount of time (ie, multiple4 or repeated3 measurements to improve reliability). Additionally, in pediatric patients with higher heart rates, breath-hold scans during anesthesia have now become possible using a SENSE reduction factor of 3 (Table 1). This may additionally improve measurement accuracy, because artifacts from respiratory motion can thus be avoided.6 Although reliable, high in-plane resolution was not advantageous in our children but may be adequate for infants with smaller blood vessels to avoid possible flow overestimation resulting from partial volume effects.15

This 3- to 5-fold increase in imaging speed is achieved without changing imaging parameters designed to optimize measurement accuracy and precision.10,16 No drawbacks in measurement accuracy and precision were observed in vivo. This finding is surprising, because both reduction of numbers of signal averages and combination with SENSE result in lower SNR. This relates not only to SNR penalty bounded by a square-root function to the number of samples acquired but also contains noise from sensitivity maps as propagated in the SENSE reconstruction process,7 known to increase with higher SENSE-reduction factors. Our SNR estimations in PC-MRI magnitude images demonstrate considerable SNR improvement in vivo when a cardiac multi-element surface coil is used for signal acquisition instead of the body coil.3 This gain in signal can be invested to reduce scanning time. As apparent in Figure 1, SNR was sufficiently high even in PC-MRI+SF-3. Furthermore, our estimation of VNR in the phase images12 revealed no relevant decrease of VNR in research PC-MRI with reduced signal averages and SENSE compared with reference PC-MRI, although a decrease in SNR in the magnitude images11 was present. Useful phase-angle information, however, can still be extracted when SNR in magnitude images is reduced, because phase-angle images have an inherently better dynamic range compared with magnitude images reconstructed from the same raw data.17

Methodological Issues: PC-MRI With SENSE
A 1-minute scan is needed at the beginning to detect surface-coil sensitivities for all SENSE scans. In our experience, no breath-hold is required in children. To avoid severe deterioration of image quality from backfolding artifacts (attributable to the SENSE reconstruction process), no additional foldover should occur during planning of the desired FOV. Therefore, we chose a relatively large FOV of 380 mm to enable the use of a standardized PC-MRI+SENSE protocol at different body sizes or slice angulations with respect to geometrical properties and gradient performance. As a result, however, the impact of SENSE reconstruction may have been variable, depending on angulation and body size as related to the FOV. For instance, in an axial plane and a small child, the actual SENSE rate is lower than nominal because FOV exceeds the extent of the chest, and therefore only limited backfolding occurs that needs unfolding using the SENSE technique. On the other hand, in a larger subject with double-oblique slice orientation, this FOV may be even borderline small, so that compromises in spatial resolution and slice angulation may occasionally become necessary, especially the pulmonary artery. An anterior-posterior phase-encoding direction is most favorable when using SENSE. If necessary to achieve this direction at double-oblique slice angulation, some misalignment between flow-encoding gradient and blood flow direction may be allowed, with only minimal bias.16

Methodological Issues: Reference PC-MRI
Although there is no gold standard available for quantitative flow measurements in pediatric patients, PC-MRI protocols comparable to our reference method have been validated in vivo and in vitro.1–3,10,16 Some changes in the reference PC-MRI protocol compared with the recently validated method3 need consideration. Use of the cardiac surface coil (increased SNR, see the Results section) allowed to reduce signal averages in the reference PC-MRI protocol from 4 to 2, thus limiting overall scan time. Moreover, vector-ECG8 instead of conventional 3-lead-ECG allowed more reliable ECG-monitoring and hence assignment of heart phases. Therefore, it seems reasonable to assume that measurement conditions of the reference PC-MRI sequence were improved, as reflected by the high precision of flow measurements.

In Vitro Results
The flow phantom experiments were designed to test the same PC-MRI protocols as were used in vivo and demonstrated high accuracy at lower in-plane resolution. Apparently, no major artifact was introduced when using PC-MRI with SENSE in this model. The FOV, however, was too large to ensure that 2- and 3-fold undersampling would lead to actual foldover in the single coil data. Thus, no additional conclusions regarding the impact of SENSE acquisition and reconstruction on phase encoding of flow volume should be drawn. An adopted FOV with reduced matrix for a similar voxel size, however, would have been different from the PC-MRI protocols used in vivo.

High-resolution PC-MRI both with and without SENSE slightly underestimated flow volumes compared with values from timed collection (Table 3). Because stronger gradients are needed to achieve the higher in-plane resolution, we speculate that artificial effects from residual eddy currents and concomitant fields may be involved.18,19 Any possible explanation, however, suffers from the fact that this underestimation was not obvious in vivo (Table 2).

Limitations
No conclusions from our data are applicable to children with arrhythmia, irregular breathing pattern, or valvular stenosis with turbulent blood flow. Only relative changes should be emphasized in the SNR and VNR measurements, because absolute values may be influenced by background filtering in the reconstructed images, reducing the noise component.


*    Conclusions
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
up arrowResults
up arrowDiscussion
*Conclusions
down arrowReferences
 
PC-MRI for flow quantitation may be combined with SENSE to achieve a substantive reduction of scanning time. SNR reduction was not a limiting factor. In children with left-to-right shunt, Qp/Qs quantification is possible by PC-MRI+SF-3 in <60 seconds. Use of higher in-plane resolution did not improve measurement results but may be adequate for infants with smaller blood vessels. The method has potential for rapid quantification of hemodynamics in congenital heart disease.


*    Acknowledgments
 
The study was supported in part by Philips Medical Systems, Best, the Netherlands. We thank Kerstin Berkemeier, RN, for data postprocessing and Hermann Esdorn, MD, and Andreas Peterschröder, MD, for data acquisition.


*    Footnotes
 
Presented at the 37th Annual Meeting of the Association for European Pediatric Cardiology, Porto, Portugal, May 16, 2002.


*    References
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
up arrowResults
up arrowDiscussion
up arrowConclusions
*References
 
1. Sieverding L, Jung WI, Klose U. Noninvasive blood flow measurement and quantification of shunt volume by cine magnetic resonance in congenital heart disease: preliminary results. Pediatr Radiol. 1992; 22: 48–54.[CrossRef][Medline] [Order article via Infotrieve]

2. Powell AJ, Maier SE, Chung T. Phase-velocity cine magnetic resonance imaging measurement of pulsatile blood flow in children and young adults: in vitro and in vivo validation. Pediatr Cardiol. 2000; 21: 104–110.[CrossRef][Medline] [Order article via Infotrieve]

3. Beerbaum P, Korperich H, Barth P. Noninvasive quantification of left-to-right shunt in pediatric patients: phase-contrast cine magnetic resonance imaging compared with invasive oximetry. Circulation. 2001; 103: 2476–2482.[Abstract/Free Full Text]

4. Houlind K, Stenbog EV, Sorensen KE. Pulmonary and caval flow dynamics after total cavopulmonary connection. Heart. 1999; 81: 67–72.[Abstract/Free Full Text]

5. Frayne R, Rutt BK. Frequency response to retrospectively gated phase-contrast MR imaging: effect of interpolation. J Magn Reson Imaging. 1993; 3: 907–917.[Medline] [Order article via Infotrieve]

6. Wolf RL, Hangiandreou NJ, Felmlee JP. Error in MR volumetric flow measurements due to ordered phase encoding in the presence of flow varying with respiration. Magn Reson Med. 1995; 34: 470–475.[Medline] [Order article via Infotrieve]

7. Pruessmann KP, Weiger M, Scheidegger MB, et al. SENSE: sensitivity encoding for fast MRI. Magn Reson Med. 1999; 42: 952–962.[CrossRef][Medline] [Order article via Infotrieve]

8. Fischer SE, Wickline SA, Lorenz CH. Novel real-time R-wave detection algorithm based on the vectorcardiogram for accurate gated magnetic resonance acquisitions. Magn Reson Med. 1999; 42: 361–370.[CrossRef][Medline] [Order article via Infotrieve]

9. Oshinski J, Ku D, Pettigrew R. Turbulent fluctuation velocity: the most significant determinant of signal loss in stenotic vessels. Magn Reson Med. 1995; 33: 193–199.[Medline] [Order article via Infotrieve]

10. Bakker CJ, Hoogeveen RM, Viergever MA. Construction of a protocol for measuring blood flow by two-dimensional phase-contrast MRA. J Magn Reson Imaging. 1999; 9: 119–127.[CrossRef][Medline] [Order article via Infotrieve]

11. Wood ML. Fourier imaging. In: Stark DD, Bradley WG, eds. Magnetic Resonance Imaging. St Louis: Mosby Year Book; 1992: 21–67.

12. Pelc NJ, Herfkens RJ, Shimakawa A. Phase contrast cine magnetic resonance imaging. Magn Reson Q. 1991; 7: 229–254.[Medline] [Order article via Infotrieve]

13. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986; 1: 307–310.[CrossRef][Medline] [Order article via Infotrieve]

14. Helbing WA, de Roos A. Clinical applications of cardiac magnetic resonance imaging after repair of tetralogy of Fallot. Pediatr Cardiol. 2000; 21: 70–79.[CrossRef][Medline] [Order article via Infotrieve]

15. Tang C, Blatter DD, Parker DL. Accuracy of phase-contrast flow measurements in the presence of partial-volume effects. J Magn Reson Imaging. 1993; 3: 377–385.[Medline] [Order article via Infotrieve]

16. Greil G, Geva T, Maier SE. Effect of acquisition parameters on the accuracy of velocity encoded cine magnetic resonance imaging blood flow measurements. J Magn Reson Imaging. 2002; 15: 47–54.[CrossRef][Medline] [Order article via Infotrieve]

17. Conturo TE, Smith GD. Signal-to-noise in phase angle reconstruction: dynamic range extension using phase reference offsets. Magn Reson Med. 1990; 15: 420–437.[Medline] [Order article via Infotrieve]

18. Bernstein MA, Zhou XJ, Polzin JA. Concomitant gradient terms in phase contrast MR: analysis and correction. Magn Reson Med. 1998; 39: 300–308.[Medline] [Order article via Infotrieve]

19. Kozerke S. Assessment of Heart Valve Function by Motion Adapted Magnetic Resonance Imaging [dissertation]. Zurich, Switzerland: Swiss Federal Institute of Technology Zurich; 2000.




This article has been cited by other articles:


Home page
RadiologyHome page
K. Muellerleile, L. Baholli, M. Groth, A. A. Barmeyer, K. Koopmann, R. Ventura, R. Koester, G. Adam, S. Willems, and G. K. Lund
Interventricular Mechanical Dyssynchrony: Quantification with Velocity-encoded MR Imaging
Radiology, November 1, 2009; 253(2): 364 - 371.
[Abstract] [Full Text] [PDF]


Home page
J. Thorac. Cardiovasc. Surg.Home page
K. K. Whitehead, K. S. Sundareswaran, W. J. Parks, M. A. Harris, A. P. Yoganathan, and M. A. Fogel
Blood flow distribution in a large series of patients having the Fontan operation: A cardiac magnetic resonance velocity mapping study
J. Thorac. Cardiovasc. Surg., July 1, 2009; 138(1): 96 - 102.
[Abstract] [Full Text] [PDF]


Home page
Br. J. Radiol.Home page
K DEBL, B DJAVIDANI, S BUCHNER, N HEINICKE, F POSCHENRIEDER, S FEUERBACH, G RIEGGER, and A LUCHNER
Quantification of left-to-right shunting in adult congenital heart disease: phase-contrast cine MRI compared with invasive oximetry
Br. J. Radiol., May 1, 2009; 82(977): 386 - 391.
[Abstract] [Full Text] [PDF]


Home page
Circ Cardiovasc ImagingHome page
L. E.J. Thomson, A. L. Crowley, J. F. Heitner, P. J. Cawley, J. W. Weinsaft, H. W. Kim, M. Parker, R. M. Judd, J. K. Harrison, and R. J. Kim
Direct En Face Imaging of Secundum Atrial Septal Defects by Velocity-Encoded Cardiovascular Magnetic Resonance in Patients Evaluated for Possible Transcatheter Closure
Circ Cardiovasc Imaging, July 1, 2008; 1(1): 31 - 40.
[Abstract] [Full Text] [PDF]


Home page
Eur Heart J SupplHome page
A. Vonk-Noordegraaf, J.-W. Lankhaar, M. J.W. Gotte, J. T. Marcus, P. E. Postmus, and N. Westerhof
Magnetic resonance and nuclear imaging of the right ventricle in pulmonary arterial hypertension
Eur. Heart J. Suppl., December 1, 2007; 9(suppl_H): H29 - H34.
[Abstract] [Full Text] [PDF]


Home page
RadioGraphicsHome page
J. F. Glockner, H. H. Hu, D. W. Stanley, L. Angelos, and K. King
Parallel MR Imaging: A User's Guide
RadioGraphics, September 1, 2005; 25(5): 1279 - 1297.
[Abstract] [Full Text] [PDF]


Home page
Am. J. Physiol. Heart Circ. Physiol.Home page
V. Muthurangu, D. Atkinson, M. Sermesant, M. E. Miquel, S. Hegde, R. Johnson, R. Andriantsimiavona, A. M. Taylor, E. Baker, R. Tulloh, et al.
Measurement of total pulmonary arterial compliance using invasive pressure monitoring and MR flow quantification during MR-guided cardiac catheterization
Am J Physiol Heart Circ Physiol, September 1, 2005; 289(3): H1301 - H1306.
[Abstract] [Full Text] [PDF]


Home page
HeartHome page
T Kuehne, S Yilmaz, I Schulze-Neick, E Wellnhofer, P Ewert, E Nagel, and P Lange
Magnetic resonance imaging guided catheterisation for assessment of pulmonary vascular resistance: in vivo validation and clinical application in patients with pulmonary hypertension
Heart, August 1, 2005; 91(8): 1064 - 1069.
[Abstract] [Full Text] [PDF]


Home page
CirculationHome page
T. Kuehne, S. Yilmaz, P. Steendijk, P. Moore, M. Groenink, M. Saaed, O. Weber, C. B. Higgins, P. Ewert, E. Fleck, et al.
Magnetic Resonance Imaging Analysis of Right Ventricular Pressure-Volume Loops: In Vivo Validation and Clinical Application in Patients With Pulmonary Hypertension
Circulation, October 5, 2004; 110(14): 2010 - 2016.
[Abstract] [Full Text] [PDF]


Home page
CirculationHome page
T. S. Sorensen, H. Korperich, G. F. Greil, J. Eichhorn, P. Barth, H. Meyer, E. M. Pedersen, and P. Beerbaum
Operator-Independent Isotropic Three-Dimensional Magnetic Resonance Imaging for Morphology in Congenital Heart Disease: A Validation Study
Circulation, July 13, 2004; 110(2): 163 - 169.
[Abstract] [Full Text] [PDF]


Home page
CirculationHome page
H. Korperich, J. Gieseke, P. Barth, R. Hoogeveen, H. Esdorn, A. Peterschroder, H. Meyer, and P. Beerbaum
Flow Volume and Shunt Quantification in Pediatric Congenital Heart Disease by Real-Time Magnetic Resonance Velocity Mapping: A Validation Study
Circulation, April 27, 2004; 109(16): 1987 - 1993.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
108/11/1355    most recent
01.CIR.0000087603.97036.C2v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Beerbaum, P.
Right arrow Articles by Meyer, H.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Beerbaum, P.
Right arrow Articles by Meyer, H.
Related Collections
Right arrow Cardiovascular imaging agents/Techniques
Right arrow Imaging
Right arrow CT and MRI
Right arrow Pediatric and congenital heart disease, including cardiovascular surgery