(Circulation. 2003;107:785.)
© 2003 American Heart Association, Inc.
Basic Science Reports |
From the Department of Cardiothoracic Surgery, University of Tokyo, Tokyo, Japan.
Correspondence to Yoshihiro Suematsu, MD, Department of Cardiothoracic Surgery, University of Tokyo, 7-3-1 Hongo Bunkyo-ku, Tokyo, 113-8655, Japan. E-mail suematsu-THO{at}h.u-tokyo.ac.jp
| Abstract |
|---|
|
|
|---|
Methods and Results RT3DE was developed with prototype ultrasound equipment consisting of a high-speed 3D rendering unit with a frame rate of 5 to 10 frames/s. We also developed a prototype semiautomatic suture device and suture cutting system. In the experiment, 12 mongrel dogs were anesthetized, and after median sternotomy, the echo probe was applied directly to the surface of the right atrium. Three surgical maneuvers (balloon atrial septectomy, enlargement of the ASD, and ASD closure) were performed through the atrial port inserted into the right atrial appendage. The heart was then excised, and the area of the ASD measured by RT3DE was compared with its area measured directly. The ASD was successfully closed in all experimental animals except the first 2. Examination of the excised heart showed that none of the sutures were loose. The mean area of the ASD was 82.5±38.6 mm2 when measured by RT3DE and 81.6±38.2 mm2 when measured directly, and there was a significant correlation between the areas measured by RT3DE and those measured directly (echo measurements=1.007xdirect measurements+0.337; P<0.0001). A Bland-Altman analysis revealed close agreement between the results obtained by the 2 methods (7.807 mm2 upper and -6.024 mm2 lower limit of agreement).
Conclusions Introduction of RT3DE, a semiautomatic suture device, and a suture cutting system made beating ASD closure without cardiopulmonary bypass possible.
Key Words: cardiopulmonary bypass echocardiography heart septal defects
| Introduction |
|---|
|
|
|---|
3D echocardiography has been developed recently and offers clinicians and surgeons a new means of visualizing the heart. This new methodology makes it possible to obtain both qualitative and quantitative information on cardiac disorders.36 Schmidt et al7 demonstrated that 3D echocardiography allows rapid and accurate measurement of left ventricular volume and mass. Nevertheless, because it takes a long time to display the 3D images, 3D ultrasonography can generally only be used as a diagnostic method and not to assist treatment, except according to a few reports.8,9 Faster and more highly automated image processing was needed, and to meet this need, we developed real-time 3D echocardiography (RT3DE) and applied it to monitoring heart operations, especially ASD closure, as a means of assisting treatment. We assessed the feasibility of off-pump ASD closure during monitoring by RT3DE in an animal experiment and evaluated the reliability of the images obtained by RT3DE.
| Methods |
|---|
|
|
|---|
The 3D rendering unit contains our original technology, referred to as "volume mode imaging" (VMI).1012 VMI is based on a volume-rendering algorithm widely used for volume visualization in the computer graphics field.13,14 VMI is characterized by the ray for rendering coinciding with an ultrasonic beam and the viewpoint invariably being fixed on the ultrasonic transducer (Figure 1). This feature allows the rendering process to start as soon as the first echo is received, without the data-reconstruction process generally used in computer graphics. After all echo data have been obtained from an ultrasonic beam, the volume-rendering values obtained are plotted on a screen as pixel brightness. Scanning of ultrasonic beams electrically yields column elements of a 3D image, and mechanical scanning with the transducer gradually produces a 3D image on the screen. At an imaging rate of 10 frames/s, each 3D image is displayed on the monitor with a maximum delay of 130 ms after the first echo is received. A 3D image of ASD is obtained by the users manipulation of a region of interest.
|
Surgical Devices
In addition to developing the original RT3DE system, to make the surgical procedure easier, we compensated for the shortcomings of conventional surgical devices by developing a prototype semiautomatic suture device (Figure 2A) suitable for off-pump intracardiac surgery. A thread (4-0 Prolene) is attached to one end of a needle that has a conical point at the top end. The bottom end of the needle is fixed in a niche at the tip of the proximal jaw of the semiautomatic suture device by pulling down on the thread with one hand. The jaws close, and the needle is passed through the tissue to be sutured. When the jaw is slowly opened again, the tip of the needle is automatically grasped by the distal jaw, which has a resilient slit. When the device is gently lifted off the tissue, the thread should penetrate the tissue properly. After the device is extracted from the cardiac cavity with the needle, the same action is repeated. The device is 5 mm in diameter so that it can be inserted into the cardiac cavity through a 5-mm trocar.
|
We used a commercially available knot pusher (Olympus Co) for minimally invasive cardiac surgery to tie the thread and a suture-cutting device (Figure 2B), which we designed to make it easy to cut the thread. The suture-cutting device consists of a shaft with a cutting edge and a handle. The shaft in turn consists of an inner tube and an outer tube, and the outer diameter of the outer tube is 2.5 mm. The inner tube has a cutting edge. The handle is joined to the end of the shaft to turn the inner tube. Thread can be cut by inserting it between the inner and outer tubes.
Animal Study
The study protocol was approved by a review committee at the University of Tokyo. All animals received humane care in compliance with the Guide for the Care and Use of Laboratory Animals published by the US National Institutes of Health (NIH publication 85-23, 1985) and with the approval of the University of Tokyo Institutional Animal Care and Use Committee.
Twelve mongrel dogs (weight 10.5 to 24.3 kg) were anesthetized with Nembutal (30 mg/kg IV), and anesthesia was maintained with 2% isoflurane. Respiration was maintained with a volume-control respirator. The ECG was monitored continuously throughout the procedure. Arterial blood gases were determined every 30 minutes, and bicarbonate was added as needed to maintain a physiological pH between 7.35 and 7.45. Aortic blood pressure was monitored with a 5F micromanometer-tipped catheter (Millar Instruments) introduced through the left femoral artery.
A median sternotomy was performed. The right atrium was suspended upward with a few stay sutures, and the echo probe was applied directly to the surface of the right atrium. A purse-string suture of 4-0 polypropylene was placed around the right atrial appendage. Three surgical maneuvers were performed through a dilating tip trocar (Ethicon Inc) inserted through the purse-string suture after intravenous administration of heparin 100 U/kg (Figure 3). When the devices were inserted, blood was slowly drawn through the side hole of the trocar to prevent introduction of air into the cardiac cavity. The surgeon performed all of the surgical maneuvers while monitoring them by RT3DE. First, a balloon catheter was inserted, and after the balloon was used to perform atrial septectomy within the fossa ovalis, the atrial communication was enlarged with a Kerrison bone punch,15 and the area of the ASD on the real-time 3D images was calculated. Next, the suture device was used to carefully close the ASD with interrupted sutures. Each suture was tied with the knot pusher, and the thread was cut with the suture-cutting system. Finally, the heart was excised, and the success of these manipulations was confirmed. After the right atrium was incised, all sutures were cut and removed, and the ASD was reopened to determine its true area.
|
After the ASD was closed, the atrial shunt was graded by 2D epicardial echocardiography as described previously16 with the following scale: (1) none, no defect detected and no color Doppler disturbance on the right atrium side; (2) trivial, no defect detected and minimal color disturbance on the right atrium side (<1 mm wide at origin of color Doppler jet); (3) small, no defect detected and a 1- to 2-mm-wide color Doppler jet; (4) moderate, defect detected and >2-mm-wide color jet; and (5) large, defect detected and large and/or multiple color jets. Off-pump ASD closure was judged to have been successful when the atrial shunt was graded "none" or "trivial," all of the sutures were tightly knotted, no massive introduction of air into the cardiac cavity was observed during the procedure, and there were no hemodynamic deteriorations except the hemodynamic changes related to ASD creation or ASD closure.
Validation Study
Because images obtained by echocardiography sometimes overestimate or underestimate the actual dimensions, we evaluated the reliability of the echocardiographic images. An experienced echocardiographer who was unaware of the true area scanned the area of the ASD by RT3DE. The images were recorded on videotape, and the echocardiographic images recorded were measured later with a digitizing tablet. All ASDs in the actual specimens were then photographed with a digital camera, and after they were transferred to a personal computer, the area of ASD was measured planimetrically and calculated by an independent, blinded observer. The images obtained by echocardiography and by direct measurement were then compared.
Statistical Analysis
All values are expressed as mean±SD. Linear regression analysis was used to assess the correlation between the area of the ASD obtained by echocardiography and the area obtained by direct measurement. A Bland-Altman analysis17 and Pitmans test were also performed to assess agreement between the results obtained by the 2 methods. Differences were considered significant when confidence limits exceeded 95% (P<0.05).
| Results |
|---|
|
|
|---|
|
|
The Table summarizes the results of the experiment. An appropriate atrial communication was observed in all animals, and there was no introduction of air through the atrial port. Residual shunts were observed by 2D Doppler echocardiography after ASD closure in the first 2 cases. They were thought to be attributable to the small number of sutures compared with the size of the ASD, and we increased the number of sutures thereafter. As a result, there were no or only trivial residual shunts after ASD closure in all of the animals except the first 2. When the heart was excised and the right atrium opened, all sutures were confirmed by the independent observer to be tightly knotted. Examination of the excised heart showed that all sutures were located within a permissible range (mean distance from the rim of the ASD 2.09±0.29 mm [range 1.50 to 2.65 mm]/mean distance between sutures 2.42±0.49 mm [range 1.50 to 4.10 mm]), and the ASD was concluded to have been successfully closed in all animals except the first 2. The mean area of the ASD created was 82.5±38.6 mm2 (range 22.4 to 138.5 mm2) based on measurements of recorded RT3DE images made by an independent echocardiographer and 81.6±38.2 mm2 (range 23.6 to 137.1 mm2) based on direct measurements. The areas measured on RT3DE images were significantly correlated with the areas measured directly (echo measurements=1.007xdirect measurements+0.337; P<0.0001; n=12; Figure 6). Bland-Altman analysis showed a negligible difference of only 0.892±2.197 between the results obtained by the 2 methods of measurement. The upper and lower limits of agreement were 7.807 and -6.024 mm2, respectively. No significant correlation was observed between the differences across 2 measurements and their means (Pitmans test of difference in variance: r=0.120, P=0.711). These results indicated excellent agreement between the 2 methods.
|
|
| Discussion |
|---|
|
|
|---|
Although several devices have become available since the first successful attempt at per-catheter ASD closure2123, they have all been associated with occasional complications.24 The Amplatzer Septal Occluder has recently been developed as a new device,25 and although some favorable results compared with conventional surgical closure have been reported,2628 limitations remain, including the need for patient selection in terms of age, body weight, and type of ASD, as well as a lower success rate than that of surgical closure. If a surgical approach could be accomplished with an incision size the same as that of per-catheter techniques, a minimally invasive, off-pump surgical approach would offer significant advantage.29
2D echocardiography is widely used to examine patients with cardiovascular disease. Although a vast amount of clinically useful information can be obtained at a relatively low cost, the technique has significant limitations. Instead of providing precise quantitative information, geometric assumptions are made to calculate chamber volume, mass, and ejection fraction. Thus, many of the observations are subjective, which results in significant interobserver variation.3 To overcome such limitations, a variety of 3D echocardiographic systems have been developed during the past decade, including a system that reconstructs a series of 2D echocardiographic images.4 Nevertheless, the current system of 3D reconstruction from multiple cross-sectional echocardiographic scans is still cumbersome and time-consuming, and the reconstruction process requires a substantial learning curve.5 Imaging time is shorter with RT3DE than with 3D reconstruction techniques applied to 2D echocardiography, because all planes are imaged simultaneously. In addition, RT3DE eliminates the need for ECG and respiratory gating. This new methodology makes it possible to obtain qualitative and quantitative information on heart disorders.30 Acar et al recently reported 3D echocardiographic techniques to monitor aortic valve balloon dilatation8 and have imaged per-catheter ASD closure before and after implantation.9 In addition, Cao et al31 demonstrated the usefulness of 2D and 3D transesophageal echocardiography in patient selection and transcatheter closure of multiple ASDs. However, RT3DE has generally only been applied to diagnostic methods.6,7 Our system is applicable to treatment as a method of monitoring surgical closure of ASD, making it possible to perform an operation that was thought to be difficult.
Surgical procedures may not be impossible with 2D images alone, but spatial relationships cannot be grasped, and the surgical procedures cannot be performed precisely, which is unacceptable to any surgeon. The present study shows that all sutures were located within a permissible range and that none of the sutures were loose. Downing et al32 reported a successful suture rate of 75% and a mean distance from the ideal position at the midportion of the leaflet of 8.5±5.0 mm in experimental beating-heart mitral valve surgery monitored by 2D echocardiography. Apparently, this was not considered accurate and would not be acceptable in clinical situations because the target structures were small. Use of the RT3DE systems allowed the "surgeons view" to be obtained in real time, and the surgeon could perform 3 procedures in the same manner as endoscopic surgery or ordinary surgery.
A limitation of the present study is that the frame rate and spatial resolution of the RT3DE images still need to be improved, because the frame rate we used was only 5 to 10 frames/s. A new RT3DE system and the use of a 2D array probe may be one solution to this problem.33 Another limitation of this study is that we did not evaluate individual variability in measurements of the area of the ASD by RT3DE, because 3D images themselves depend greatly on an appropriate opacity line and echo gain, which were decided by the echocardiographer. Further assessment will be needed to elucidate individual variability. Finally, median sternotomy was required to perform all of the surgical procedures in the present study. If the current RT3DE system is used through a small skin incision, 2 different thoracotomies would be needed because of the large echo probe. The use of a new robotics system and technological development of the RT3DE system, including a transesophageal echo probe, will probably make this possible. We believe that the indications for this type of intervention will be extended to other intracardiac operations, such as closure of ventricular septum defects, tricuspid annuloplasty, and mitral valvuloplasty.
In summary, we successfully closed ASDs by RT3DE monitoring in an animal experiment, and examination of the excised heart showed that all sutures were located within a permissible range and that none of the sutures were loose. There were no changes in blood pressure or heart rate during the intracardiac manipulations.
| Acknowledgments |
|---|
Received August 21, 2002; revision received October 24, 2002; accepted October 28, 2002.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
N. V. Vasilyev, I. Melnychenko, K. Kitahori, F. P. Freudenthal, A. Phillips, R. Kozlik-Feldmann, I. S. Salgo, P. J. del Nido, and E. A. Bacha Beating-heart patch closure of muscular ventricular septal defects under real-time three-dimensional echocardiographic guidance: A preclinical study J. Thorac. Cardiovasc. Surg., March 1, 2008; 135(3): 603 - 609. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Handke, G. Heinrichs, U. Moser, F. Hirt, F. Margadant, F. Gattiker, C. Bode, and A. Geibel Transesophageal Real-Time Three-Dimensional Echocardiography: Methods and Initial In Vitro and Human In Vivo Studies J. Am. Coll. Cardiol., November 21, 2006; 48(10): 2070 - 2076. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. M. Lang, V. Mor-Avi, L. Sugeng, P. S. Nieman, and D. J. Sahn Three-Dimensional Echocardiography: The Benefits of the Additional Dimension J. Am. Coll. Cardiol., November 21, 2006; 48(10): 2053 - 2069. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Suematsu, B. N. Mora, T. Mihaljevic, and P. J. del Nido Totally Endoscopic Robotic-Assisted Repair of Patent Ductus Arteriosus and Vascular Ring in Children Ann. Thorac. Surg., December 1, 2005; 80(6): 2309 - 2313. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Suematsu, J. F. Martinez, B. K. Wolf, G. R. Marx, J. A. Stoll, P. E. DuPont, R. D. Howe, J. K. Triedman, and P. J. del Nido Three-dimensional echo-guided beating heart surgery without cardiopulmonary bypass: Atrial septal defect closure in a swine model J. Thorac. Cardiovasc. Surg., November 1, 2005; 130(5): 1348 - 1357. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Orihashi, T. Sueda, K. Okada, K. Imai, K. Ban, and M. Hamamoto Real-time three dimensional echo-guided closure of atrial septal defect: an experimental model Interactive CardioVascular and Thoracic Surgery, October 1, 2005; 4(5): 391 - 395. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Suematsu, G. R. Marx, J. A. Stoll, P. E. DuPont, R. O. Cleveland, R. D. Howe, J. K. Triedman, T. Mihaljevic, B. N. Mora, B. J. Savord, et al. Three-dimensional echocardiography-guided beating-heart surgery without cardiopulmonary bypass: A feasibility study J. Thorac. Cardiovasc. Surg., October 1, 2004; 128(4): 579 - 587. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. E. Weyman The year in echocardiography J. Am. Coll. Cardiol., January 7, 2004; 43(1): 140 - 148. [Full Text] [PDF] |
||||
![]() |
Z. Amin, D. A. Danford, J. Lof, K. F. Duncan, and S. Froemming Intraoperative device closure of perimembranous ventricular septal defects without cardiopulmonary bypass: Preliminary results with the perventricular technique J. Thorac. Cardiovasc. Surg., January 1, 2004; 127(1): 234 - 241. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|