| Abstract |
|---|
|
|
|---|
Methods and Results A newly designed metallic stent contains honeycombed strut elements with inlaid stacked layers of drug and polymer. In vitro studies evaluated recipes for loading paclitaxel to establish the parameters for controlling drug release. Manipulation of the layers of biodegradable polymer and drug allowed varying of the initial 24-hour burst release of paclitaxel from 69% to 8.6% (P<0.0001). Late release of drug could be adjusted dependently or independently of early burst release. A biphasic release profile was created by the addition of blank layers of polymer within the stack. In the 30-day porcine coronary model (n=17 pigs), there was a 70% reduction in late loss (0.3±0.5 versus 1.0±0.5 mm, P=0.04), a 28% increase in luminal volume (132±12 versus 103±21 mm3, P=0.02), and a 50% decrease in histological neointimal area (2.0±0.5 versus 4.0±1.6 mm2; P<0.001) compared with bare metal controls. Temporal and regional variations in vascular healing were seen histologically.
Conclusions Layered polymer/drug inlay stent technology permits flexible and controllable pharmacokinetic profiles. Programmable, complex chemotherapy using this approach may be feasible for the treatment of cardiovascular disease.
Key Words: pharmacokinetics stents angioplasty restenosis
| Introduction |
|---|
|
|
|---|
| Methods |
|---|
|
|
|---|
|
Polymer/Drug Loading
Polylactide coglycolide (PLGA) with the lactic acid to glycolic acid ratios of 50:50 and 85:15 (Birmingham Polymers) were dissolved in dimethyl sulfoxide (DMSO) to make concentrations of 5% for the PLGA 85:15 and 10% for the PLGA 50:50. Appropriate amounts of paclitaxel (Samyang Genex Co) were added to the polymer solutions to make final concentrations of 1% or 3%.
Polymer/drug solutions were shot into the holes from a piezoelectric nozzle with a 40-µm orifice positioned by a micromanipulator. The solvent evaporated in 30 minutes to leave a PLGA/paclitaxel layer inside the holes. The loading process was repeated on the same holes with as many as 22 layers.
For in vitro kinetic studies, individual layers were varied to create different stacking arrangements according to goals of the specific experiment. In addition to layers containing PLGA/paclitaxel, layer types without drug were added, including blank PLGA 85:15 to form a slower eroding film at the bottom of the holes and blank PLGA 50:50 to retard the release of paclitaxel. Processing variables, such as polymer molecular weight, viscosity, number of layers, layer thickness, drying time, temperature, humidity, and solvent type were found to influence kinetics and thus were tightly controlled.
For animal experiments, the 392 holes on the struts bounded by the ductile hinges were filled with polymer/drug, and the remaining 196 holes near the bridges were left empty to evaluate the spatial effects of paclitaxel release. As a starting point for future comparisons, paclitaxel kinetics were chosen to simulate first-order release kinetics, as seen with coated stents. These stents were layered with stacks of 12 loadings of PLGA 50:50 (molecular weight 40 to 70 kDa) containing 1% paclitaxel to a total dose of 95 µg of paclitaxel per stent. Approximately 60% of the drug was released by 24 hours and 100% released by 15 to 20 days. Stents were mounted on 3/20 mm PTCA balloons and sterilized with ethylene oxide.
In Vitro Kinetics
Individual stents (n=52, 3 of each stacking type) were immersed in agitated 2-mL aliquots of 2 mol/L N,N-diethylnicotinamide (DENA) for serial 1- to 3-day periods, for a total duration of 20 days. The DENA was found to improve paclitaxel solubility from 0.45 µg/mL in PBS to 980 µg/mL (>2000-fold increase), approximating an infinite sink without requiring simulated flow conditions. The aliquots were injected into a Symmetry C18 column (Waters Corporation). The HPLC system was optimized to detect paclitaxel and to remove the interfering peaks from the DENA control. Stents were examined by scanning electron microscopy (SEM) at 0, 2, 6, 14, 22, and 24 days to verify extent of polymer erosion.
Porcine Stent Restenosis Model
Animal experiments were approved by the Institutional Animal Care and Use Committee and conformed to AHA guidelines. Juvenile farm pigs (S & S Farms, Ranchita, Calif; n=17) underwent placement of single coronary stents in the mid right coronary artery as previously described.4 The pigs received 75 mg clopidogrel and 325 mg aspirin the day before and for 7 and 30 days thereafter, respectively. Stents with PLGA 95 µg paclitaxel (n=6) were compared with bare metal stents without holes (n=7). To evaluate polymer toxicity, stents containing PLGA polymer without drug (n=4) were similarly implanted. At 1 month, quantitative coronary angiography (QCA) and intravascular ultrasound (IVUS) were performed, followed by euthanasia under general anesthesia. The coronary arteries were perfusion-fixed and placed in formalin.
Data Acquisition and Analysis
QCA parameters included proximal and distal reference, minimal lumen diameter (MLD), and late loss. Eleven IVUS frames (7 within the stent and 2 in the 6-mm segments both proximal and distal to the stent) were selected at 2.5- to 3-mm intervals for planimetry. Measurements included lumen and intimal volumes by Simpsons rule.
Tissue blocks were embedded in methyl methacrylate and ground into 3 cross-sections at predetermined points to avoid sampling bias. The 2 sections cut across drug-containing struts were analyzed separately from a middle section cut across the sinusoidal bridges. Stained Paragon sections were examined by an experienced pathologist (M.C.F.) using a prospectively defined grading scale.4 Histomorphometric analysis of the tissue sections was performed with a Nikon E400 microscope and ImagePro software for measurement of the stent and lumen areas and mean neointimal thickness. Injury score was measured as described by Schwartz et al.5
Statistical Analysis
All quantitative measurements and histological parameters were obtained by blinded observations. Data are summarized as mean±SD, unless otherwise specified. Comparisons for interval data were by unpaired Students t test. Comparisons for ordinal data were by Mann-Whitney U test.
| Results |
|---|
|
|
|---|
Figure 2 graphs cumulative paclitaxel release and daily release rates from 4 in vitro experiments. Figure 2A shows stents with 10 stacked layers containing different concentrations of paclitaxel. The top 3 curves show that drug release profiles are parallel and proportional to the total drug loaded. After 1 week, another slight increase in release rate was observed, possibly because of the enlarging surface area of PLGA. The bottom curve resulted from 3 additional topcoats of PLGA 50:50 devoid of drug but the same total amount of drug as in the 1% paclitaxel curve immediately above it. Top coating reduced the first 24-hour burst release from 24% to 14.7% of the total dose loaded (P<0.0005). In Figure 2B, blank topcoats also reduced initial 24-hour burst release (26.9% to 16.8%, P<0.0002). The exaggerated change in slope after day 10 resulted from placing blank layers of PLGA 50:50 in the middle of the stack, thus creating a biphasic drug release. Figure 2C shows that initial and late release rates can be varied in an inversely linked fashion by adding top and bottom blank layers. Figure 2D uses recipes analogous to Figures 2B and 2C to show that initial and late release rates can be varied independently. The 24-hour burst release was reduced to a minimum of 8.6% (P<0.0001) by top coating (50:50) and bottom barrier layer (85:15). The late release of paclitaxel was affected by varying the concentration of drug, number of layers, and positioning of the various layers within the stack.
|
Effects of Paclitaxel Release In Vivo
All animals survived for 1 month. Table 1 summarizes QCA, IVUS, and histomorphometric parameters. Baseline QCA indices did not differ between the drug delivery and the control metal stents. At follow-up, lumen dimensions were significantly improved in the paclitaxel group, with the MLD increasing by 50% (P<0.001) and late loss decreasing by 70% (P=0.04).
|
Follow-up IVUS pullback showed that the paclitaxel group had a 28% larger luminal volume (P=0.02) and a trend toward 64% less neointimal volume (P=0.08), with no evidence of worsening in the edge regions proximal and distal to the stent (Figure 3).
|
Histomorphometry revealed that injury score and stent area were similar in both groups whether at the strut or bridge regions of the stent. Lumen area, intimal area, percent area stenosis, and mean intimal thickness were significantly improved in the paclitaxel group in the strut/drug-containing sections. Similar but nonsignificant trends toward improvement were seen at the sinusoidal bridges of the paclitaxel-containing stents.
Figures 4 and 5 show histological sections, and Table 2 summarizes the scored histological features from strut and bridge locations. Bare metal controls were similar at the struts or the bridges, with a thickened neointima composed of morphological smooth muscle cells and extracellular matrix, mild scores for neointimal fibrin, and adventitial fibrosis. A neoendothelium consisting of a coherent layer of flat, elongated cells was seen, and there was minimal inflammation, foreign body reaction, or medial necrosis. The paclitaxel group had a very different and spatially specific histological appearance. Sections containing paclitaxel-loaded struts had voids where the polymer occupied the holes and consistently had grade 3 fibrin deposition surrounding the struts. There was partial healing with smooth muscle cells and extracellular matrix present between the struts. There was significantly more inflammation, foreign body reaction, medial necrosis, and adventitial fibrosis than in bare metal controls. Sections taken from the bridge locations showed that the paclitaxel group did not differ from bare metal controls, apart from a minimal increase in intimal inflammation. Both groups showed complete healing with intimal smooth muscle cells and matrix, a neoendothelial layer, and minimal medial necrosis. Sections containing stents with polymer and no drug were identical to bare metal controls, except for the persistent voids where the polymer had been in the strut holes.
|
|
|
| Discussion |
|---|
|
|
|---|
Our study reports preliminary experience with a novel design drug-delivery stent loaded with paclitaxel as a model lipophilic agent. The Conor MedStent design differs from conventional drug delivery stents in that the stresses and strains of stent expansion are concentrated in small, specially contoured features called ductile hinges, which absorb all expansion forces. It is therefore possible to create holes or wells in the struts without sacrificing strength, scaffolding, or flexibility. The holes are filled with therapeutic drugs in a layered configuration, creating hundreds of individually adjustable drug reservoirs. Compared with conventional stents with a 5- to 10-µm coating, the MedStents polymer/drug volume is enlarged by 4- to 16-fold, polymer thickness is increased 14- to 28-fold, and polymer surface area is reduced by 64% to 81%. These features permit expanded drug-loading capacity, the ability to deliver multiple drugs in a specified sequence, and the ability to program the timing and spatial distribution patterns of drug release.
In vitro studies demonstrated that drug-release kinetics are flexible and programmable. Reproducibility of the kinetic profiles was suggested in several instances by the small variability between samples but not quantifiable because of the small number of samples per group. Drug release was controlled so that a wide range of dosages was deliverable without affecting the general shape of the kinetic release profile and thus facilitating dose-response studies. Initial 24-hour burst release was selectively variable from >60% to <10% of the total dose. Burst rates were controllable, with minimal alteration of chronic release rates. A two-step early and late pulsatile release was demonstrated, as was independent control over the magnitude of the early and late phases.
Studies in the 30-day porcine coronary stent model showed that paclitaxel delivered with a first-order kinetic profile from two thirds of the holes was biologically active, inhibiting neointimal thickening and delaying healing in close proximity to the drug-containing cavities. When stents with PLGA 50:50 but without drug were implanted, there was no demonstrable toxicity.
These preliminary studies have several limitations that should be considered with respect to making inferences about future clinical applications. Only a single drug was used to evaluate control over pharmacokinetics. Other agents were not tested, because the emphasis of this study was on stent mechanics and release pharmacokinetics. Paclitaxel was chosen as the model small molecule lipophilic agent because of a well-known mechanism of action and previously documented arterial wall responses.911 Only one dose of paclitaxel and one kinetic profile were used in the pilot animal study, because this was designed to be illustrative of biological effect using the new device rather than being exhaustive with respect to quantitation or to compare with conventional surface-coated drug-eluting stents. Although the numbers of animals used were small, the results were consistent and corroborated by independent measurement methods, thus permitting valid statistical inferences. The finding of incomplete healing (persistent fibrin and inflammation) suggests that the present dose may have been too large or delivered too rapidly or the duration of follow-up too short.12,13
These limitations not withstanding, our observations were consistent with previous porcine coronary studies delivering paclitaxel from surface-coated stents showing incomplete vascular healing at 1 month.911 By programming the Conor stent not to have polymer/drug in proximity to the sinusoidal bridges, we were able to demonstrate localized control over the spatial distribution of drug effects. We found islands of healed vessel wall with less reduction in neointimal thickening localized to the bridge sections and interstrut regions. These findings infer regional dose effects and raise the question that localized healed regions may be a scaffold to promote late complete vessel wall healing. This localized dose effect is consistent with the known partitioning behavior of paclitaxel, which is associated with large spatial concentration variations, resulting in concentrations that may be either toxic or inadequate.2,3
Additional work with the MedStent has included loading oligonucleotides. The system of filling individual holes one layer at a time is now fully computer automated, enhancing speed and scalability. Within any given hole, the stacking of inlaid polymer/drug can be varied to create a variety of controllable pharmacokinetic profiles. Figure 6 illustrates some of the potential programmability features of the MedStent. In case 1, and as demonstrated in vitro, a more slowly degrading barrier layer has been placed at the lumen-side surface of the hole, thus preferentially causing more rapid egress of drug in the direction of the wall. A topcoat containing no drug reduces burst release. Case 2 is an example showing how multiagent chemotherapy might be delivered to the peristrut vascular wall with early release of drug A and delayed release of drug B. Case 3 shows a method to deliver multitarget chemotherapy, with drug A localized to the stented wall and drug B eluted into the lumen for simultaneous treatment of vulnerable plaque or diffuse disease.
|
That preferential vectorial delivery in the direction of the wall or lumen may be achievable is suggested by three observations. First, the major mechanism of drug release is polymer erosion rather then diffusion. This is almost certainly the case, because release profiles are controllable on the basis of concentration of drug in sequential layers. More water-soluble agents may have different release kinetics governed by diffusion as well as bulk erosion. Next, the specific use of slower degrading polymer on luminal side with stacked layers of faster degrading PLGA on top reduced initial burst rates and delayed release rates in kinetic curves to the extent expected if all drug could only elute toward the side. Finally, serial SEMs in vitro confirm that the slower-degrading PLGA barrier on the lumen side does not erode through until 16 to 20 days. Although there is no present data showing an advantage of multiagent chemotherapy for preventing restenosis, such treatment may be considered in light of early reports, with single-agent paclitaxel and sirolimus having restenosis particularly at the edges.
Finally, our studies suggest that present methods of local drug delivery from vascular stents can be viewed in a perspective analogous to the progress in developing systemically administered drugs. In the past, systemic medication therapy was limited by the fixed pharmacokinetics of specific drugs. Combining drugs with controlled-release methodology to create flexible hybrid delivery platforms, thereby decoupling chemical and kinetic properties, has revolutionized systemic therapy. Preliminary experience with a second-generation programmable controlled-release stent technology suggests that similar approaches may be feasible for the interventional treatment of cardiovascular disease.
Received August 8, 2002; revision received October 18, 2002; accepted October 20, 2002.
| References |
|---|
|
|
|---|
2. Hwang C-W, Wu D, Edelman ER. Physiological transport forces govern drug distribution for stent-based delivery. Circulation. 2001; 104: 600605.
3. Creel CJ, Lovich MA, Edelman ER. Arterial paclitaxel distribution and deposition. Circ Res. 2000; 86: 879884.
4. Makkar R, Whiting J, Li A, et al. Effects of ß-emitting 188Re balloon in stented porcine coronary arteries- an angiographic, intravascular ultrasound and histomorphometric study. Circulation. 2000; 102: 31173123.
5. Schwartz RS, Huber KC, Murphy JG, et al. Restenosis and the proportional neointimal response to coronary artery injury: results in a porcine model. J Am Coll Cardiol. 1992; 19: 267274.[Abstract]
6. Morice M-C, Serruys PW, Sousa JE, et al. A randomized comparison of a sirolimus-eluting stent with a standard stent for coronary revascularization. N Engl J Med. 2002; 346: 17731780.
7. Fox R. American Heart Association 2001 scientific sessions: late breaking science-drug eluting stents. Circulation. 2001; 104: E9052.
8. Crank J. Diffusion in a plane sheet. In: The Mathematics of Diffusion. 2nd ed. London: Oxford University Press; 1975: 4468.
9. Drachman DE, Eldelman ER, Seifert P, et al. Neointimal thickening after stent delivery of paclitaxel: change in composition and arrest of growth over six months. J Am Coll Cardiol. 2000; 36: 23252332.
10. Heldman AW, Cheng LW, Jenkins GM, et al. Paclitaxel stent coating inhibits neointimal hyperplasia at 4 weeks in a porcine model of coronary restenosis. Circulation. 2001; 103: 22892295.
11. Farb A, Heller PF, Shroff S, et al. Pathological analysis of local drug delivery of paclitaxel via a polymer-coated stent. Circulation. 2001; 104: 473479.
12. Costa MA, Sabate M, van der Giessen WJ, et al. Late coronary occlusion after intracoronary brachytherapy. Circulation. 1999; 100: 789792.
13. Liistro F, Stankovic G, Di Mario C, et al. First clinical experience with a paclitaxel derivative-eluting polymer stent system implantation for in-stent restenosis: immediate and long-term clinical and angiographic outcome. Circulation. 2002; 105: 18831886.
This article has been cited by other articles:
![]() |
J. M. Wilson and J. T. Willerson Myocardial Revascularization with Percutaneous Devices Card. Surg. Adult, January 1, 2008; 3(2008): 573 - 598. [Full Text] |
||||
![]() |
P. W. Serruys Fourth Annual American College of Cardiology International Lecture: A Journey in the Interventional Field J. Am. Coll. Cardiol., May 2, 2006; 47(9): 1754 - 1768. [Full Text] [PDF] |
||||
![]() |
S. H. Hofma, W. J. van der Giessen, B. M. van Dalen, P. A. Lemos, E. P. McFadden, G. Sianos, J. M.R. Ligthart, D. van Essen, P. J. de Feyter, and P. W. Serruys Indication of long-term endothelial dysfunction after sirolimus-eluting stent implantation Eur. Heart J., January 2, 2006; 27(2): 166 - 170. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. W. Serruys, G. Sianos, A. Abizaid, J. Aoki, P. den Heijer, H. Bonnier, P. Smits, D. McClean, S. Verheye, J. Belardi, et al. The Effect of Variable Dose and Release Kinetics on Neointimal Hyperplasia Using a Novel Paclitaxel-Eluting Stent Platform: The Paclitaxel In-Stent Controlled Elution Study (PISCES) J. Am. Coll. Cardiol., July 19, 2005; 46(2): 253 - 260. [Abstract] [Full Text] [PDF] |
||||
![]() |
H.-J. Priebe Perioperative myocardial infarction--aetiology and prevention Br. J. Anaesth., July 1, 2005; 95(1): 3 - 19. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. A. Costa and D. I. Simon Molecular Basis of Restenosis and Drug-Eluting Stents Circulation, May 3, 2005; 111(17): 2257 - 2273. [Full Text] [PDF] |
||||
![]() |
R. Wessely, J. Hausleiter, C. Michaelis, B. Jaschke, M. Vogeser, S. Milz, B. Behnisch, T. Schratzenstaller, M. Renke-Gluszko, M. Stover, et al. Inhibition of Neointima Formation by a Novel Drug-Eluting Stent System That Allows for Dose-Adjustable, Multiple, and On-Site Stent Coating Arterioscler. Thromb. Vasc. Biol., April 1, 2005; 25(4): 748 - 753. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. E. Sousa, M. A. Costa, E. M. Tuzcu, J. S. Yadav, and S. Ellis New Frontiers in Interventional Cardiology Circulation, February 8, 2005; 111(5): 671 - 681. [Full Text] [PDF] |
||||
![]() |
P. H Chong and J. W. Cheng Early Experiences and Clinical Implications of Drug-Eluting Stents: Part 1 Ann. Pharmacother., April 1, 2004; 38(4): 661 - 669. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Auer, R. Berent, T. Weber, and B. Eber Risk of noncardiac surgery in the months following placement of a drug-eluting coronary stent J. Am. Coll. Cardiol., February 18, 2004; 43(4): 713 - 713. [Full Text] [PDF] |
||||
![]() |
D. B. Buxton, S. C. Lee, S. A. Wickline, M. Ferrari, and for the Working Group Members Recommendations of the National Heart, Lung, and Blood Institute Nanotechnology Working Group Circulation, December 2, 2003; 108(22): 2737 - 2742. [Abstract] [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2003 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |