(Circulation. 2003;107:3022.)
© 2003 American Heart Association, Inc.
Clinical Investigation and Reports |
From the Department of Internal Medicine II and Department of Nuclear Medicine (S.N.R.), University of Ulm, and Department of Nuclear Medicine (J.K.), University of Dresden, Germany.
Correspondence to PD Dr Martin Höher, FESC, Internal-Medicine-II, University of Ulm, Robert-Koch-Strasse-8, 89081 Ulm, Germany. E-mail martin.hoeher{at}medizin.uni-ulm.de
| Abstract |
|---|
|
|
|---|
Methods and Results After successful coronary angioplasty with or without stenting, 225 patients (71% de novo lesions) were randomly assigned to receive 22.5 Gy intravascular ß-irradiation in 0.5-mm tissue depth (n=113) or to receive no additional intervention (n=112). Clinical and procedural data did not differ between the groups except a higher rate of stenting in the control group (63%) compared with the rhenium-188 group (45%, P<0.02). After 6 months of follow-up, late loss was significantly lower in the irradiated group compared with the control group, both of the target lesion (0.11±0.54 versus 0.69±0.81 mm, P<0.0001) and of the total segment (0.22±0.67 versus 0.70±0.82 mm, P<0.0001). This was also evident in the subgroup of patients with de novo lesions and independent from stenting. Binary restenosis rates were significantly lower at the target lesion (6.3% versus 27.5%, P<0.0001) and of the total segment (12.6% versus 28.6%, P<0.007) after rhenium-188 brachytherapy compared with the control group. Target vessel revascularization rate was significantly lower in the rhenium-188 (6.3%) compared with the control group (19.8%, P=0.006).
Conclusions Intracoronary ß-brachytherapy with a rhenium-188 liquid-filled balloon is safe and efficiently reduces restenosis and revascularization rates after coronary angioplasty.
Key Words: radioisotopes angioplasty stents restenosis revascularization
| Introduction |
|---|
|
|
|---|
- or ß-emitters is now a proven concept for treatment of in-stent restenosis,36 but its effectiveness for de novo lesions has not been shown. With ß-irradiation, a dose-finding study in de novo lesions showed a significant dose-dependent reduction in loss of luminal diameter at 6 months of follow-up.7 Irradiation with a liquid rhenium-188filled balloon catheter provides a dose-dependent significant antiproliferative effect in vitro8 and in animal models.9,10 The proliferative response was almost completely abolished after balloon denudation and ß-radiation with 22.5 Gy in 0.5-mm tissue depth in hypercholesterolemic rabbits.10 In a clinical pilot study, we have shown the safety and feasibility of intracoronary ß-irradiation using the liquid rhenium-188filled balloon.11 After irradiation with 15 Gy in 0.5-mm tissue depth, we revealed a restenosis rate at the target lesion of 12% (n=3 of 26). However, the restenosis rate of the total segment was 46% because of edge stenoses and geographical miss. Other pilot trials with rhenium-188 balloon brachytherapy using 28 Gy (n=11)12 and 30 Gy (n=2613 and n=5014) in 0.5-mm tissue depth to treat in-stent restenosis revealed an angiographic restenosis rate of 10% to 18% at 6 months.
This EndoCoronary-Rhenium-Irradiation-Study (ECRIS) represents a randomized, controlled trial of ß-emitting radiation therapy in a broad spectrum of patients undergoing percutaneous coronary intervention. The objective of our trial was to evaluate the efficacy of ß-irradiation with a liquid rhenium-188filled balloon in patients with de novo or restenotic lesions. To overcome the limitations of our pilot trial,11 a longer irradiation segment exceeding the traumatized area by at least 5 mm at each side and a higher dose of 22.5 Gy in 0.5-mm tissue depth was selected.
| Methods |
|---|
|
|
|---|
Angioplasty and Irradiation Procedure
Before angioplasty, the patients received 10 000 U heparin, adjusted to the activated clotting time (>280 seconds). Provisional stenting was performed at the operators discretion. The irradiation procedure with the liquid rhenium-188filled balloon catheter has been described previously.11 Rhenium-188 was obtained from a tungsten-188/rhenium-188 radionuclide generator as a solution of sodium perrhenate in 0.9% chloride (Oak Ridge National Laboratory). The mean activity was 97±28 mCi/mL. The angioplasty balloon was replaced by a noncompliant balloon of the same diameter but at least 10 mm more in length (Tacker, Cordis Europe). To maintain maximum activity, rhenium-188 was not mixed with contrast agent. Exact positioning was accomplished by means of the proximal and distal radio-opaque marker of the irradiation balloon. The target lesion was covered, and the balloon-traumatized area at both sides was overlapped by the irradiation balloon by at least 5 mm. For angiographic analysis, the irradiated segment was identified by contrast injection with the deflated irradiation balloon in place and the same level of inspiration as with the preprocedural and postprocedural angiograms, giving the exact positions of the proximal and distal marker within the vessel. The maximum length of the irradiation balloon was 40 mm. In 4 patients this was not sufficient to cover the target irradiation zone, and the irradiation procedure was done in a sequential positioning technique with careful alignment of the balloon markers. For irradiation, the balloon was manually inflated. Because of the low viscosity of the liquid rhenium-188, manual inflation of the balloon is sufficient to achieve complete filling without bubbles.11 When the irradiation had to be fractionated, the balloon was deflated for about 2 to 5 minutes but left in place across the target lesion to avoid shifts of the balloon and to minimize radiation exposure to other areas. A team composed of 2 cardiologists and 1 nuclear physician performed the procedure. Acetylsalicylic acid 100 mg per day was continued throughout the study. Patients with stents additionally received ticlopidine 250 mg twice daily or clopidogrel 75 mg daily for 4 weeks. From August 1999 on, the combined antiplatelet therapy after irradiation plus stenting was prolonged to 6 months.
Quantitative Coronary Angiography
Coronary angiography of the target lesion before and after angioplasty and at follow-up was performed in the same projections after intracoronary application of glycerol-trinitrate. Angiographic measurements were done with the Pie Medical software 2.1 (Pie Medical Imaging). Quantitative coronary measurements were performed from the target lesion and from the total segment in the same 2 orthogonal views. Data from the target lesion represent the initial stenosis and the segment injured by angioplasty and irradiation. Data from the total segment include the target lesion and the edge zone. The edge zone was defined as the 6-mm segment proximal and the 6-mm segment distal to the target lesion. Restenosis was defined as diameter stenosis of greater than 50%. Restenoses within the edge zones were defined as edge stenoses. Acute gain was calculated as the difference in minimal lumen diameter (MLD) after and before the procedure. Late loss was calculated as the difference between the postprocedural MLD and at 6 months.
Statistical Analysis
Power analysis of ECRIS was done on the assumption of a restenosis rate after angioplasty of 35% and an aimed 50% restenosis reduction by coronary irradiation therapy, resulting in 105 irradiated and 105 control patients required to receive statistical significance (
=0.05, ß=0.2). Continuous variables are presented as mean±1 SD. For the comparison of continuous variables, 2-tailed t test was used. Discrete variables were expressed as counts and percentages compared by means of
2 analysis. Survival analysis was performed with the log-rank test. Statistical significance was set at the 5%
-error level (P<0.05).
| Results |
|---|
|
|
|---|
|
|
Irradiation Data
The length of the irradiated segment varied between 20 (n=2), 30 (n=66), 40 (n=41), 60 (n=2), 80 (n=1), and 90 mm (n=1), with a mean length of 35.0±9.2 mm. The diameter of the liquid rhenium-188filled balloon varied between 3.0 (n=72), 3.5 (n=30), and 4.0 mm (n=11), with a mean diameter of 3.2±0.3 mm. In all patients, the prescribed dose of 22.5 Gy was delivered. The mean irradiation time was 6.8±3.6 minutes. Irradiation was performed in 1 (n=31), 2 (n=34), 3 (n=22), 4 (n=12), 5 (n=8), 6 (n=2), and 7 fractions (n=4). There were no adverse effects of the irradiation procedure except anginal pain and ST-segment changes. No radiation leakage occurred. There were 4 protocol-violating procedures in which the balloon-traumatized zone of the vessel was clearly not covered by the irradiation balloon. The 4 patients were excluded from angiographic analysis per protocol but remained in clinical follow-up and in the angiographic analysis according to intention to treat.
Clinical Follow-Up
There was no in-hospital death. In the control group, 1 patient died in cardiogenic shock from myocardial infarction 116 days after the intervention, and 1 patient with history of myocardial infarction experienced an episode of nonischemic ventricular fibrillation 97 days after the interventional procedure. Twenty patients who were asymptomatic at 6 months declined to undergo follow-up angiography. Recatheterization was performed in 203 of 225 (90.2%) patients. In the irradiation group 7 of 113 (6.2%) patients suffered from acute myocardial infarction, occurring after 6, 8, 45, 59, 70, 77, and 139 days. Of the 7 patients, 4 had an angiographically documented acute vessel occlusion treated by angioplasty and 3 received thrombolysis without previous angiography in another hospital. Six of the 7 patients had a newly implanted stent. In the control group, 5 of 112 (4.5%) patients experienced an acute myocardial infarction, 3 within 24 hours after stent implantation and 2 in patients without a stent on days 77 and 90. Four patients had an angiographically documented occlusion of the target vessel and received angioplasty, and 1 patient underwent conservative treatment in another hospital. All myocardial infarctions occurred when the patients took only acetylsalicylic acid without additional ticlopidine or clopidogrel. There was no stent occlusion or myocardial infarction during prolonged combined antiplatelet therapy from August 1999. Because the lesion underneath an acute thrombotic occlusion cannot be measured by angiographic analysis and based on the different pathomechanism (thrombus versus intimal proliferation), the 12 patients with myocardial infarction were excluded from restenosis analysis per protocol but remained in analysis of MACE and angiographic analysis by intention to treat.
Angiographic Data and Revascularization
Ninety-five patients in the rhenium-188 group and 91 patients in the control group were included in the angiographic analysis per protocol. Table 3 lists the results of quantitative coronary measurements. There was no coronary aneurysm in both groups at 6 months of follow-up. The MLD before and after angioplasty did not differ between both groups. At 6 months, the irradiated group compared with the control group had a significantly lower late loss, a significantly higher MLD, and a significantly lower percent diameter stenosis both of the target lesion and the total segment. The binary restenosis rates of the target lesion and of the total segment were significantly lower in the rhenium-188irradiated group compared with the control group (Table 3). The cumulative MLD distribution curves of the control group and of the irradiated group illustrate the striking difference of MLD at follow-up (Figure 1). Edge stenoses occurred in 6 patients (6.3%) of the rhenium-188 group and in 1 (1.1%) patient of the control group (P=0.06). The revascularization rate of the target lesion was significantly lower in the irradiated group with 2.1% (n=2 of 95) compared with 18.7% (n=17 of 91) in the control group (P=0.0002). In addition, the revascularization rate of the total segment was significantly lower in the rhenium-188 group with 6.3% (n=6 of 95) compared with 19.8% (n=18 of 91) in the control group (P=0.006).
|
|
In the subgroup of patients with de novo lesions (n=131), rhenium-188 irradiation resulted in the same significant reduction of late loss and percent stenosis diameter as in the whole study group, both for the target lesion and the total segment (Table 3). The significantly lower late loss in the irradiated group was independent from stenting. In both subgroups of patients with stent implantation or balloon angioplasty alone, late loss of the target lesion and of the total segment was significantly lower after rhenium-188 irradiation (Figure 2).
|
Intention-to-Treat Analysis
Table 4 shows the angiographic data from intention-to-treat analysis, including all lesions with angiographic control (n=203) and the respective subgroup of de novo lesions (n=144). Intention-to-treat analysis also revealed a significant lower late loss, a lower percent diameter stenosis, and a significant higher MLD in the irradiated group compared with the control group. In accordance with the analysis per protocol, this effect was present at the target lesion and at the total segment, both with all patients and in the subgroup of patients with de novo lesions. Even in a worst case scenario, imputing a 0-mm MLD at follow-up for all 12 patients with myocardial infarction, the reduction of late loss of the total segment remained significant (Figure 2).
|
Major Adverse Cardiac Events
Table 5 lists MACEs during 30 days and 6 months of follow-up. MACEs were significantly lower in the irradiated compared with the control group for the subgroup (n=146) of patients included after August 1999 with prescribed combined antiplatelet therapy for 6 months after irradiation and stenting. Because late coronary occlusion may occur after 6 months, clinical follow-up was extended to 12 months. After 6 months there was 1 myocardial infarction from a nontarget vessel in each group and 1 late stent occlusion in the irradiated group. This stent occlusion occurred in a patient without any antiplatelet therapy 10 days after prostatectomy from cancer.
|
| Discussion |
|---|
|
|
|---|
Compared with other irradiation devices, the rhenium-188 liquid-filled balloon provides various advantages.15 The diameter and length of the irradiation balloon can be easily adapted to the individual vessel segment without need for intravascular ultrasound for dosimetry. Furthermore, the irradiation balloon is self-centering, providing a homogeneous dose delivery and a 100% success in dose delivery attributable to the use of a standard monorail balloon catheter technique. ß-radiation compared with
-radiation requires less radiation protection and shorter treatment times and is associated with a lower radiation exposure to the medical staff. The initial occurrence of myocardial infarctions attributable to delayed endothelialization, predominantly after stenting, could be solved by prolonging the combined antiplatelet therapy with acetylsalicylic acid and clopidogrel for up to 6 months.
De Novo Lesions
Gamma- and ß-irradiation has been shown to be effective for the treatment of in-stent restenosis.3,5,6 In de novo lesions, ß-irradiation using a strontium-90/yttrium-90 coil source has been shown to result in a dose-dependent decrease of the restenosis rates after 6 months from 28.6% to 15.0% after irradiation with 9, 12, 15, and 18 Gy at 1-mm tissue depth.7 With the rhenium-188filled balloon, total segment restenosis rate of de novo lesions was 12.5%, despite a 1.6-times lower radiation dose compared with the 18-Gy group from Verin et al7 (22.5 Gy in 0.5-mm tissue depth corresponds to 11.3 Gy in 1-mm depth). Different from previous randomized studies with ß-irradiation of de novo lesions, ECRIS revealed a significant irradiation efficacy both at the target lesion and the total segment. Compared with our pilot trial,11 edge stenoses decreased from 35% to 6%. The randomized BETACATH trial showed a significant reduction of the restenosis rate at the target lesion (22%) but an unfavorable restenosis rate of the total segment (39%). Similar results were achieved using a phosphor-32 wire source.16 In the PREVENT trial, a total of 105 patients with de novo (70%) and restenotic lesions were randomized to receive radiation with 0, 16, 20, or 24 Gy in 1-mm tissue depth. The irradiated patients had a restenosis rate at the target lesion of 8% and at the total segment of 22%.
Differences between BETACATH, PREVENT, and ECRIS include the dose, the length of the irradiated segment and of the edge zones, and the irradiation system. In BETACATH, a noncentered 30-mm radiation source train (strontium-90/yttrium-90) with an effective irradiation length of 25 mm was used. The dose was 28% lower than in ECRIS (16.1 compared with 22.5 Gy in 0.5-mm tissue depth for a 3.0-mm vessel). The length by which the irradiated segment exceeded the traumatized area at each site was only 0.5 to 2.5 mm in BETACATH, disregarding additional extension of the traumatized area by longitudinal shifts of the balloon, but was ≥5 mm in ECRIS. Especially in the stent group of BETACATH, there was almost no overlap of the irradiated zone, resulting in a 24% rate of edge stenoses. In PREVENT, a centered 27-mm radiation source and a dose 42% to 113% higher than in ECRIS was used. With an effective irradiation segment of 22 mm, the overlap of the irradiation zone over the traumatized area ranged between 0 and 2.5 mm at each site, resulting in 14% edge stenoses. Comparing ECRIS, BETACATH, and PREVENT, the length of the irradiated segment exceeding the full traumatized area and homogeneous dose delivery facilitated by a centered device seems to be critical. Whether compression of the vessel tissue during rhenium-188 balloon brachytherapy and the slightly reduced radiation shielding by intramural calcific deposits or stents struts attributable to the larger diameter of the rhenium-188 balloon source in contrast to a wire source are of additional advantage remains open.
Incorporating prior lesions learned from early brachytherapy trials regarding dosimetry and antiplatelet therapy, our data show that rhenium-188 ß-brachytherapy can be applied to a broad spectrum of patients undergoing coronary intervention, including de novo lesions, and could be an alternative to an expensive drug-eluting stent technology.
Limitations
The trial was not designed to study small arteries, because it was limited to vessels suitable for a 3-mm radiation balloon at least. Despite careful randomization and analysis blinded to the operators, stents were used more often in the control group compared with the irradiated group. However, rhenium-188 brachytherapy was effective in the subgroups with and without stenting, and stents were no longer associated with increased late thrombosis after prolonging antiplatelet therapy. Because stenting reduces restenosis rate, the lower stent rate of the irradiated group is even in favor of the control group.
De novo lesions were the largest subgroup (71% of the patients). However, the study was not powered for subgroup analysis. Whether the positive results of rhenium-188 brachytherapy can be expected in patients with diabetes, small vessels, or bypass grafts remains to be determined.
Conclusions
We conclude that intracoronary ß-brachytherapy with a rhenium-188 liquid-filled balloon is safe and efficiently reduces restenosis and revascularization rates after coronary angioplasty.
| Acknowledgments |
|---|
| Footnotes |
|---|
Received March 3, 2003; revision received March 27, 2003; accepted March 31, 2003.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
J H SHIN, S K LEE, H-Y. SONG, J-S KIM, H CHOE, E-H KIM, I J LEE, T-H KIM, E-Y KIM, C-W WOO, et al. The effects of 188rhenium-filled balloon dilation following bare stent placement in a rabbit oesophageal model Br. J. Radiol., May 1, 2008; 81(965): 413 - 421. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. Willhauck, B.-R. Sharif Samani, F.-J. Gildehaus, I. Wolf, R. Senekowitsch-Schmidtke, H.-J. Stark, B. Goke, J. C. Morris, and C. Spitzweg Application of 188Rhenium as an Alternative Radionuclide for Treatment of Prostate Cancer after Tumor-Specific Sodium Iodide Symporter Gene Expression J. Clin. Endocrinol. Metab., November 1, 2007; 92(11): 4451 - 4458. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Marx, J. Wohrle, T. Nusser, D. Walcher, A. Rinker, V. Hombach, W. Koenig, and M. Hoher Pioglitazone Reduces Neointima Volume After Coronary Stent Implantation: A Randomized, Placebo-Controlled, Double-Blind Trial in Nondiabetic Patients Circulation, November 1, 2005; 112(18): 2792 - 2798. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Sabate, G. Pimentel, C. Prieto, J. Corral, C. Banuelos, D. J. Angiolillo, F. Alfonso, R. Hernandez-Antolin, J. Escaned, P. Fantidis, et al. Intracoronary brachytherapy after stenting de novo lesions in diabetic patients: Results of a randomized intravascular ultrasound study J. Am. Coll. Cardiol., August 4, 2004; 44(3): 520 - 527. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Waksman Vascular brachytherapy and coronary stenting for de novo lesions: Love on the rocks J. Am. Coll. Cardiol., August 4, 2004; 44(3): 538 - 540. [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2003 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |