From the Clinic of Diagnostic Radiology (A.B., E.S., F.F., R.W.G.) and the Clinic of Pediatric Cardiology (E.G.M.), University of Technology Aachen, Aachen, Germany; the Department of Pediatric Cardiology (R.G.), Kiel University, Kiel, Germany; Philips Research Laboratories (T.S.), Hamburg, Germany; and Philips Medical Systems (J.J.v.V.), Best, the Netherlands.
Correspondence to Arno Buecker, MD, Clinic of Diagnostic Radiology, Pauwelsstr. 30, D-52057 Aachen, Germany. E-mail buecker{at}rad.rwth-aachen.de
| Abstract |
|---|
|
|
|---|
Methods and Results Real-time MRI with radial or spiral k-space filling (15 frames per second) on an interventional 1.5-T high-field whole-body system was exploited to examine the feasibility of MR-guided closure of the PFO in 7 piglets weighing
14 kg. A specially designed prototype nonmagnetic closure device was introduced via the femoral vein. The short bore of the magnet and in-room monitors allowed for visualization and steering of the catheter with the loaded occluder. Catheterization of the left atrium and, finally, correct placement of the device was possible in all animals. Deployment of the device was depicted by real-time MR, and initial misplacement, which occurred in 2 animals, was easily detected and corrected.
Conclusions Real-time MR guidance of PFO closure, without the use of ionizing radiation, is feasible in an animal model.
Key Words: magnetic resonance imaging catheterization defects
| Introduction |
|---|
|
|
|---|
| Methods |
|---|
|
|
|---|
MR Scanner and Real-Time Imaging Sequence
The interventions were performed on a dedicated interventional 1.5-T ACS-NT MR scanner (Philips Medical Systems). X-ray control of the MR-guided interventions was performed by a specially shielded C-arm unit positioned in the same room with the MR scanner. LCD screens beside the magnet enabled direct control of the MR-guided intervention. A remote control was used to change the image plane, orientation, and angulation as required during the procedure. Specially designed real-time reconstruction hardware (Philips Research Laboratories) performed the real-time gridding interpolation of the raw data, consequently allowing standard fast Fourier reconstruction of the images also in real time.6 A 5-element synergy cardiac coil was used for signal acquisition.
Real-time MR image data acquisition was achieved by applying a gradient echo sequence with either radial (repetition time [TR] 2.5 ms, echo time [TE] 1.2 ms, flip angle 45°, field of view 300 mm2, matrix 128x256, slice thickness 10 mm, and 80 radials) or spiral (TR 30 ms, TE 4 ms, flip angle 40°, field of view 250 mm2, matrix 1282, slice thickness 10 mm, and 4 spirals) k-space filling. Additionally, the sliding window reconstruction technique7 was applied, yielding a frame rate of 15 images per second. This technique, also called view sharing or echo sharing, uses partially old and partially new data to calculate the next frame, yielding a higher frame rate compared with the acquisition time of 1 completely new MR image, which was 200 ms for radial scanning (5 frames per second) and 120 ms for spiral scanning (8.3 frames per second).
Atrial Septal Occluder and MR-Guided Intervention
Self-made nonmagnetic prototypes of atrial septal occluders made of nitinol were used because they were favorably depicted during in vitro experiments applying real-time MRI sequences.8 The occluder consists of 2 connected button halves, one or both of which can be applied with a membrane (Figure 1). The system runs over a 0.25-mm nitinol guidewire. In addition, it is fixed by one 0.15-mm nitinol guidewire, which allowed us to perform the "Minnesota wiggle" procedure9 and, in conjunction with a small nitinol tube (1 mm), to recapture the occluder into the delivery catheter. The whole system as described passes through a 2.7-mm (8F) delivery catheter.
|
A 3.3-mm (10F) introducer sheath was placed in the right femoral vein by the Seldinger technique. Next, the left atrium was catheterized through the foramen ovale. In the first animal, a 17.5-kg piglet, the foramen was not patent, and an artificial opening had to be created by use of a transjugular portosystemic shunt needle (Cook) and dilation with a 12-mm balloon catheter. In all other animals, it was easily possible to pass the atrial septum through the patent foramen ovale with a 1.6-mm (5F) multipurpose catheter and to secure the path with a 0.9-mm (0.035-in) nitinol guidewire. These procedures were performed under x-ray guidance in the first 4 animals and purely under MR guidance in the last 3 animals. The general imaging plane for the intervention was chosen on a parasagittal plane along the orientation of the catheter to contain the inferior vena cava and the right and the left atria (Figure 2). In the first 4 animals, the mounted occluder was advanced under MR guidance over a guidewire previously positioned under x-ray guidance into the left atrium. In the last 3 animals, catheterization of the left atrium was performed under MR guidance, steering the catheter with a loaded occluder through the foramen ovale (Figure 3, Movie I [representative image quality for all experiments]). All occluders were deployed and positioned over the foramen ovale solely under MR guidance (Figure 4, Movie I). The position of each device as determined by real-time MR was confirmed by x-ray. Correct positioning of the device was tested under MR by wiggling the still connected occluder. Free movement in the right or left atrium without movement of the septum was considered as misplacement, with both halves of the occluder either in the right or left atrium. This was proved by the same wiggle maneuver as performed under x-ray guidance (Figure 5). If misplacement occurred, the occluder was recaptured and replaced.
|
|
|
|
MR images of the occluder were obtained by applying a fast, navigator-gated, free-breathing, segmented k-space, cardiac-triggered, 3D, true fast imaging with steady state precession (FISP) sequence (TR 3.9 ms, TE 1.9 ms, flip angle 75°, resolution 1.2x1.2 mm2, slice thickness 1.5 mm, 16 slices, and acquisition time <2 minutes [triggering included]) (Figure 6). Finally, the animals were euthanized, and their hearts were removed and opened to macroscopically inspect the occluder position.
|
| Results |
|---|
|
|
|---|
| Discussion |
|---|
|
|
|---|
Correct occluder placement was determined by a wiggle procedure identical to that (Minnesota wiggle) carried out under x-ray fluoroscopy9 (Figure 5, Movie II). If misplacement occurred, the occluder was recaptured into the delivery catheter and reinserted successfully under MR guidance and confirmed by wiggle procedures performed under x-ray fluoroscopy and macroscopically in the ex vivo heart specimens
The tomographic nature of MR made it important to choose a suitable imaging plane for control of the procedure. Because we wanted to visualize the advancing catheter as well as the atrial septum, we planned an imaging plane including the inferior vena cava and both atria. Planning was performed on a parasagittal slice, as shown in Figure 2. The resulting para-axial image contained parts of all heart chambers and the aortic valve. Slight adjustments of this imaging plane were made during the procedure by moving the imaging plane parallel to the originally planned slice. Because of the interactive abilities of the interventional MR scanner, this was possible as required, without any significant time delay. The fully anesthetized animals did not inadvertently move, but the interactive abilities can be expected to be sufficient for quick slice repositioning in a clinical setting, where patient motion might occur.
The present study demonstrates the general feasibility of artifact-free real-time MRI of the heart, enabling MR-guided placement of an atrial septal occluder. Neither cardiac nor respiratory triggering nor breath-holding was necessary to achieve high-contrast images of the heart chambers. The occluder was depicted passively by susceptibility artifacts, which are due to the different magnetization of the metal nitinol compared with tissue.15 An advantage of MR compared with x-ray fluoroscopy is the lack of exposure of the patient and technician to ionizing radiation. Furthermore, the depiction of the heart anatomy shows the relation of the occluder to the atrial septum, pulmonary veins, and atrioventricular valves. The value of MR compared with echocardiography, which already has been shown to be applicable to the procedure in clinical studies,4 remains to be examined. The high costs of an MR scanner, the ability to deal with complications such as loss of the occluder, and use of a cardioverter close to the MR scanner have to be considered. Furthermore, the use of metallic nitinol instruments, which can at least in theory act as antennas and consequently heat up during the MR experiment,16 has to be thoroughly examined in septal occluder systems before clinical trials can be undertaken. Nonetheless, the experiments show the general feasibility of real-time MR guidance for cardiac interventions with a sufficient imaging speed and contrast.
In conclusion, high quality real-time MRI of the heart allows MR guidance of transcatheter patent foramen ovale occlusion in an animal model.
| Footnotes |
|---|
Received March 20, 2002; revision received May 1, 2002; accepted May 2, 2002.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
D. S. Baim and R. Y. Kwong Is Magnetic Resonance Image Guidance the Key to Opening Chronic Total Occlusions? Circulation, February 28, 2006; 113(8): 1053 - 1055. [Full Text] [PDF] |
||||
![]() |
A. N. Raval, P. V. Karmarkar, M. A. Guttman, C. Ozturk, S. Sampath, R. DeSilva, R. J. Aviles, M. Xu, V. J. Wright, W. H. Schenke, et al. Real-Time Magnetic Resonance Imaging-Guided Endovascular Recanalization of Chronic Total Arterial Occlusion in a Swine Model Circulation, February 28, 2006; 113(8): 1101 - 1107. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. J. Lederman Cardiovascular Interventional Magnetic Resonance Imaging Circulation, November 8, 2005; 112(19): 3009 - 3017. [Full Text] [PDF] |
||||
![]() |
A. N. Raval, J. D. Telep, M. A. Guttman, C. Ozturk, M. Jones, R. B. Thompson, V. J. Wright, W. H. Schenke, R. DeSilva, R. J. Aviles, et al. Real-Time Magnetic Resonance Imaging-Guided Stenting of Aortic Coarctation With Commercially Available Catheter Devices in Swine Circulation, August 2, 2005; 112(5): 699 - 706. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. K. Raman, P. V. Karmarkar, M. A. Guttman, A. J. Dick, D. C. Peters, C. Ozturk, B. S.S. Pessanha, R. B. Thompson, A. N. Raval, R. DeSilva, et al. Real-Time Magnetic Resonance-Guided Endovascular Repair of Experimental Abdominal Aortic Aneurysm in Swine J. Am. Coll. Cardiol., June 21, 2005; 45(12): 2069 - 2077. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. A. Krombach, J. G. Pfeffer, S. Kinzel, M. Katoh, R. W. Gunther, and A. Buecker MR-guided Percutaneous Intramyocardial Injection with an MR-compatible Catheter: Feasibility and Changes in T1 Values after Injection of Extracellular Contrast Medium in Pigs Radiology, May 1, 2005; 235(2): 487 - 494. [Abstract] [Full Text] [PDF] |
||||
![]() |
R Corti, J Badimon, G Mizsei, F Macaluso, M Lee, P Licato, J F Viles-Gonzalez, V Fuster, and W Sherman Real time magnetic resonance guided endomyocardial local delivery Heart, March 1, 2005; 91(3): 348 - 353. [Abstract] [Full Text] [PDF] |
||||
![]() |
I M Barbash, J Leor, M S Feinberg, A Tessone, S Aboulafia-Etzion, A Orenstein, J Ruiz-Cabello, J S Cohen, and Y Mardor Interventional magnetic resonance imaging for guiding gene and cell transfer in the heart Heart, January 1, 2004; 90(1): 87 - 91. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. J. Dick, M. A. Guttman, V. K. Raman, D. C. Peters, B. S.S. Pessanha, J. M. Hill, S. Smith, G. Scott, E. R. McVeigh, and R. J. Lederman Magnetic Resonance Fluoroscopy Allows Targeted Delivery of Mesenchymal Stem Cells to Infarct Borders in Swine Circulation, December 9, 2003; 108(23): 2899 - 2904. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Schalla, M. Saeed, C. B. Higgins, A. Martin, O. Weber, and P. Moore Magnetic Resonance-Guided Cardiac Catheterization in a Swine Model of Atrial Septal Defect Circulation, October 14, 2003; 108(15): 1865 - 1870. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Haage, W. Piroth, G. Krombach, S. Karaagac, T. Schaffter, R. W. Gunther, and A. Bucker Pulmonary Embolism: Comparison of Angiography with Spiral Computed Tomography, Magnetic Resonance Angiography, and Real-Time Magnetic Resonance Imaging Am. J. Respir. Crit. Care Med., March 1, 2003; 167(5): 729 - 734. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2002 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |