(Circulation. 2002;106:2026.)
© 2002 American Heart Association, Inc.
Current Perspective |
From The Zena and Michael A. Wiener Cardiovascular Institute (Z.A.F., V.F.), New York, NY; the Department of Radiology (Z.A.F.), Mount Sinai School of Medicine, New York, NY; and the Department of Clinical Radiology (K.N., C.B.), Klinikum Grosshadern, University of Munich, Munich, Germany.
Correspondence to Zahi Fayad, PhD, Imaging Science Laboratories, Mount Sinai School of Medicine, Box 1030, New York, NY 10029. E-mail zahi.fayad{at}mssm.edu
Key Words: imaging plaque angiography magnetic resonance imaging computed tomography
| Introduction |
|---|
|
|
|---|
Disruption-prone plaques in the coronary arteries, the so-called "vulnerable plaques," tend to have a thin fibrous cap (cap thickness
65 to 150 µm) and a large lipid core (American Heart Association [AHA] plaque type IV-Va). Acute coronary syndromes often result from disruption of a modestly stenotic vulnerable plaque, not visible by x-ray angiography, which results in a thrombotic complication (AHA plaque type VI). During its evolution, a type Va plaque may also become fibrotic (AHA plaque type Vc) or calcified (AHA plaque type Vb).2,3 In contrast to coronary artery vulnerable plaques characterized by high lipid content and a thin fibrous cap, high-risk plaques of the carotid arteries tend to be fibrotic and severely stenotic.3
Imaging of Atherothrombotic Disease
Because there is striking heterogeneity in the composition of human atherothrombotic plaques, even within the same individual, reliable noninvasive imaging tools that can detect early atherothrombotic disease in the various regions and characterize the composition of the plaques are clinically desirable.3 Such imaging tools would improve our understanding of the pathophysiological mechanisms underlying atherothrombotic processes and allow us to better risk-stratify the disease. Additionally, such tools may permit optimal tailoring of treatment and allow direct monitoring of the vascular response.
Presently, a number of imaging modalities are employed to study atherosclerosis; most identify luminal diameter or stenosis, wall thickness, and plaque volume.3 Two noninvasive imaging modalities, computed tomography and MRI, have been introduced to the study of atherothrombosis. They allow identification of flow-limiting coronary stenoses, calcified plaques, imaging of the atherothrombotic lesions directly, measurement of atherosclerotic burden, and characterization of the plaque components.3 Together, by revealing the degree of stenosis and the plaque composition, they provide information that may predict cardiovascular risk, facilitate further study of atherothrombosis progression and its response to therapy, and provide for assessment of subclinical disease.
| Computed Tomography |
|---|
|
|
|---|
Electron-Beam CT
The necessity for very short image acquisition times to virtually freeze cardiac motion urged the development of a cardiac-dedicated CT system in 1982 on the basis of a nonmechanical movement of the x-ray source and fixed detector arrays. The electron-beam CT (EBCT) uses a single, curved anode with 4 tungsten targets underneath the patient, and a focused electron beam that is rapidly swept across these targets to produce an x-ray fan beam detected by 2 detector rows above the patient (Table 1).
|
Multidetector-Row CT
Mechanical multidetector-row CT (MDCT) systems were introduced in 1998, and allow for scanning with one x-ray tube and 4 detector rows in a single gantry rotating twice per second around the patient. Continuous gantry rotation and table movement causes the projection data to be obtained along a spiral or helical path (Table 1).
Presently, the spatial resolution of MDCT is higher (0.6x 0.6x1.0 mm3) than that of EBCT (0.8x 0.8x2.5 mm3). However, continuous x-ray radiation during systole and diastole with MDCT means significantly higher exposure (9 mSv) than with EBCT (1.5 mSv). A technique designed to reduce the x-ray tube current of spiral MDCT during the cardiac cycle may lower the radiation exposure by 48% for males and 45% for females.4
| Applications |
|---|
|
|
|---|
EBCT for Coronary Calcium
Since the early 1990s, EBCT has been considered the "gold standard" for the assessment of calcified plaques.5,6 ECG-triggered acquisition of the entire heart is performed with 3-mm contiguous slices progressing from the trunk of the pulmonary artery to the base of the heart during a single breathhold (
30 to 40 s).
According to the American College of Cardiology/American Heart Association consensus document on coronary calcium,7 the clinical indications that may be considered are the detection of coronary calcium in patients with atypical chest pain, as well as the quantification and follow-up of the coronary calcium plaque load in asymptomatic patients with cardiovascular risk factors.8
Coronary artery calcium is assessed through the measurement of the number of pixels in the CT image with a density
130 Hounsfield units (HU). The total calcium score quantification scheme proposed by Agatston et al6 is the most widely used. Some limitations of the total calcium score include the low and variable reproducibility (14% to 51%).9 To overcome these limitations, the calcium volumetric score (CVS) was introduced.10 With the use of CVS, Callister et al8 illustrated the effect of lipid-lowering therapy on coronary calcification. However, there is no evidence that changes of coronary calcification correspond to changes in the cardiovascular events risk.7
A high coronary calcium score (CS) is a sensitive but not a specific marker for obstructive CAD.7 In many patients, coronary calcium is found even in the absence of any clinical coronary symptoms. Therefore, it has been suggested that coronary calcium may represent a preclinical stage of CAD. Whether coronary calcium is independent and superior to conventional risk factors as a predictor for future cardiac events is not yet well established.11 Future randomized large prospective studies (eg, Prospective Army Coronary Calcium study12) may provide the definitive evidence of the predictive value of CS.
Multidetector-Row CT for Coronary Calcium
MDCT with retrospective ECG triggering acquires 3-mm slices with 1.5-mm overlap in a 15-s breathhold time (see Data Supplement figure). The CVS from MDCT correlates well with EBCT data in the same patient group.13 Compared with EBCT, MDCT demonstrated higher spatial resolution and signal-to-noise ratio; however, the longer exposure time (250 ms with MDCT versus 100 ms with EBCT), leads to motion artifacts.
The high spatial resolution and soft tissue delineation possible with contrast-enhanced (see later) multidetector-row coronary CT angiography (MD-CTA) may provide some information on noncalcified coronary artery lesions and noninvasively image the coronary artery wall.14 In a study by Schröder et al,15 soft, intermediate, and calcified plaques in the coronary arteries as defined by IVUS had a unique density, detected by contrast-enhanced MD-CTA (14±26 HU for soft plaques, 91±21 HU for intermediate plaques, and 419±194 HU for calcified plaques).15 In a preliminary study of heart specimen, we found that lipid-rich plaques had low density (50±12 HU), whereas fibrous plaques presented with higher density (89±31 HU) (Figures 1 and 2). 16 Similar findings have been reported in carotid artery plaques.17
|
|
Further studies are essential to establish whether detection with contrast-enhanced MDCT of coronary lesions with noncalcified components identifies vulnerable plaques or patients that are at risk for developing coronary thrombosis and cardiac events.
Coronary CT Angiography
Electron-Beam CT Angiography
Intravenous contrast was applied with electron-beam CT angiography (EB-CTA) for the assessment and quantification of coronary stenosis. EB-CTA is performed with 1.5- to 3-mm slices with a 1-mm slice overlap resulting in an
40-s breathhold time. Homogenous enhancement of the coronary arteries is obtained with the intravenous administration of 120 mL of contrast media (400 mg iodine) injected at a flow rate of 3 mL/s. Initial results from the current studies are shown in Table 2.
|
It has also been shown18 that extensive calcifications may interfere with the detection of coronary artery stenoses, leading to false-negative results compared with selective coronary angiography. Small vessel diameters (<1.5 mm) were found to lead to false-positive stenosis findings.
Multidetector-Row CT Angiography
MD-CT angiography with retrospective ECG-gating is performed with 1.3-mm slices with 0.6-mm overlap (Figures 3 and 4). Similar to EB-CTA, after intravenous injection (3 mL/s) of a nonionic contrast medium (120 mL, 300 mg iodine) into the antecubital vein, the entire heart is scanned within a single breathhold time of
40 s. The exposure time of 250 ms per slice requires lowering the heart rate to
60 bpm (eg, with the use of oral or intravenous ß-blockers) to avoid cardiac motion image artifacts.19 However, in some patients, heart rates
60 bpm could not always be achieved at the expense of poor coronary artery visibility.20
|
|
Two of the studies21,22 excluded
30% of coronary segments mainly because of image degradation from cardiac motion. The image artifacts and the number of excluded coronary segments may be significantly reduced (to
10%) by the use of ß-blocker and careful patient selection.23 This allowed the imaging of the entire coronary tree within one breathhold.23 A summary of some of the coronary CTA studies by EBCT and MDCT is shown in Table 2.
Possible Future Improvements
EBCT
Single-slice axial EBCT imaging with the use of nonspiral scanning diminishes the volume coverage of a single breathhold ECG-gated acquisition. Multiple stationary detectors may be needed for greater volume coverage and higher spatial resolution. New reconstruction algorithms may reduce the cone beaminduced artifacts24 and allow ECG-gated reconstruction from nonspiral continuous acquisition.
Multidetector-Row CT
The next generation MDCT scanner will allow for faster gantry rotation and simultaneous acquisition of >4 slices (eg, 16 slices). The breathhold time may decrease to
20 s, thus reducing the necessary contrast media (eg, 60 mL) for sufficient enhancement of the coronary arteries. The temporal and spatial resolution may most likely increase to 200 ms and 0.8-mm slice thickness, respectively. These enhancements may help in the detection and grading of stenoses of the coronary arteries even in the presence of coronary calcium, and in quantification of calcified as well noncalcified lesions.
| Magnetic Resonance Imaging |
|---|
|
|
|---|
Multicontrast MR Plaque Imaging
High-resolution MR has emerged as the potential leading noninvasive in vivo imaging modality for atherosclerotic plaque characterization. MR differentiates plaque components on the basis of biophysical and biochemical parameters such as chemical composition and concentration, water content, physical state, molecular motion, or diffusion. MR provides imaging without ionizing radiation and can be repeated over time. In vivo MR plaque imaging and characterization have been performed utilizing a multicontrast approach with high-resolution black blood spin echo and fast spin echo (FSE) based MR sequences. The signal from the blood flow is rendered black through preparatory pulses (eg, radiofrequency spatial saturation or inversion recovery pulses) to better image the adjacent vessel wall.26 However, bright blood imaging (ie, 3D fast time of flight) can be employed in assessing fibrous cap thickness and morphological integrity of the carotid artery plaques.27 This sequence enhances the signal from flowing blood and a mixture of T2* and proton density contrast weighting highlights the fibrous cap.
Atherosclerotic plaque characterization by MR is generally based on the signal intensities and morphological appearance of the plaque on T1-weighted, proton densityweighted, and T2-weighted images as previously validated (see references in recent reviews by Fayad et al3 and Yuan et al28).
MR Coronary Angiography
Coronary artery MR angiography (CMRA) is one of the most challenging areas of cardiovascular MR because of the size and topology of the arteries, as well as cardiac and respiratory motion. Since the publication of the first clinical results in 1993,29 CMRA has undergone numerous technical improvements and innovations.30 Three major groups or "generations" of coronary MRA techniques can be discerned. From all techniques available, 2D breathhold CMRA scans,31 3D retrospective respiratory navigator-gated coronary MRA,3239 and breathhold 3D CMRA40,41 have been evaluated clinically (Table 2 and Figure 5).
|
No specific MR techniques have emerged that can provide sensitivity and specificity for coronary lesion detection similar to those of traditional contrast coronary x-ray angiography. Moreover, compared with x-ray angiography, each of the techniques mentioned above suffer from limitations in temporal and spatial resolution, achieving spatial resolutions of
1 mm3 voxel size and temporal resolutions of 100 ms acquisition time per heartbeat, at best. Another limitation of CMRA, compared with x-ray angiography and CTA, is the necessity of data acquisition over several heartbeats, which can lead to motion artifacts. A more detailed review of the current status of CMRA application can be found in Duerinckx.30
| Applications |
|---|
|
|
|---|
In vivo, black-blood MR atherosclerotic plaque characterization of the human aorta has been reported recently by Fayad et al,26 who assessed thoracic aorta plaque composition and size with the use of T1-weighted, T2-weighted, and proton densityweighted images. The acquired images had a resolution of 0.8x0.8x5.0 mm3 with a phased-array coil. Rapid high-resolution imaging was performed with an FSE sequence in conjunction with velocity-selective flow suppression preparatory pulses. Matched cross-sectional aortic imaging with MR and transesophageal echocardiography showed a strong correlation for plaque composition and mean maximum plaque thickness. An asymptomatic patient with complex carotid, aortic arch, and descending aorta plaque as detected by MR is shown in Figure 6. Contrast-enhanced MRA with the use of gadolinium-based contrast agents may provide additional information for plaque characterization by identifying neovascularization in the atherosclerotic plaque and potentially improve the differentiation between necrotic core and fibrous tissue.46 Furthermore, other nonspecific and specific contrast agents may facilitate accurate plaque constituent characterization and the identification of specific molecular and biological activity.47,48
|
Coronary Plaques With MRI
Preliminary studies in a pig model showed that the difficulties of coronary wall imaging are the result of a combination of cardiac and respiratory motion artifacts, nonlinear course, small size, and location.49 Fayad et al extended the black-blood MR methods used in the human carotid artery and aorta to the imaging of the coronary arterial lumen and wall.43 The method was validated in swine coronary lesions induced by balloon angioplasty.49
High-resolution black-blood MR of both normal and atherosclerotic human coronary arteries was performed for direct assessment of coronary wall thickness and the visualization of focal atherosclerotic plaque in the wall. The difference in maximum wall thickness between the normal subjects and patients (
40% stenosis) was statistically significant. Figure 7 shows in vivo MR coronary plaque images from 3 patients. The coronary MR plaque imaging study by Fayad et al43 was performed during breathholding to minimize respiratory motion with a resolution of 0.46x 0.46x2.0 mm3. To alleviate the need for the patient to hold his or her breath, Botnar et al50 have combined the black-blood FSE method and a real-time navigator for respiratory gating and real-time slice position correction. A near isotropic spatial resolution (0.7x0.7x1 mm3) was achieved with the use of a 2D local inversion and black-blood preparatory pulses.50 This method provided a quick way to image a long segment of the coronary artery wall and may be useful for rapid coronary plaque burden measurements. Future studies need to address these possibilities.
|
In Vivo Monitoring of Progression and Regression of Plaques With MRI
As shown in animal experimental studies,51,52 MR is a powerful tool to serially and noninvasively investigate the progression and regression of atherosclerotic lesions in vivo. In asymptomatic, untreated, hypercholesterolemic patients with carotid and aortic atherosclerosis, Corti et al53 have shown that MR can be used to measure the efficacy of lipid-lowering therapy (statins). Atherosclerotic plaques were assessed with MR at different times after lipid-lowering therapy. Significant regression of atherosclerotic lesions was observed. Despite the early and expected hypolipidemic effect of the statins, changes in the vessel wall were observed for 12 months. As with previous experimental studies, there was a decrease in the vessel wall area and no alteration in the lumen area after 12 months.51,52 Recently, a case-controlled study demonstrated substantially reduced carotid plaque lipid content (but no change in overall lesion area) in patients treated for 10 years with an aggressive lipid-lowering regimen compared with untreated controls.54
Coronary MR Angiography
Magnetic resonance offers several advantages for coronary imaging.30 MR does not use ionizing radiation and does not necessarily require the injection of a contrast agent. Several MR techniques have been proposed for the detection of coronary stenosis with MR, but no large population study has been presented.
A large multicenter study that uses a standardized 3D prospective respiratory navigator approach has been presented.39 The study population consisted of 109 subjects from 7 US and European centers who were scheduled to undergo their first elective x-ray coronary angiography for suspected CAD. The mean CMRA time was 74 minutes. The evaluation was limited to the proximal 3 to 5 cm of the major coronary arteries. Of the 759 coronary segments available for analysis, 636 (84%) could be interpretable by CMRA. In these segments (
50% stenosis), 78 of 94 stenoses (83%) were detected by CMRA. Overall, CMRA has an accuracy rate of 72% (95% confidence interval, 63% to 81%) in diagnosing CAD. From this study we can reach the following 3 conclusions: (1) If the goal is to identify those without any coronary disease and save them from coronary x-ray angiography, CMRA will make x-ray angiography unnecessary in 20% (however, 18% of this group (4% of total) will have their single-vessel disease undiagnosed) (2) if the goal is to identify those without left main or 3-vessel disease and save them from coronary x-ray angiography, CMRA will make x-ray angiography unnecessary in 69% of patients (in a similar cohort); and (3) most patients with isolated distal coronary disease will be missed by CMRA. It is necessary to define the groups of patients who may benefit from this CMRA technique and to evaluate thoroughly the new and evolving imaging techniques in the appropriate clinical settings.55
Today, established clinical applications of CMRA include the evaluation of the patency of coronary artery bypass grafts and the imaging of anomalous coronary arteries.30
Possible Future Improvements
MR Plaque Imaging
Thinner slices, such as those obtained with 3D acquisition techniques, could further improve artery wall imaging.50 Additional MR techniques, such as water diffusion weighting,56 magnetization transfer weighting,28 steady-state free precession (SSFP) sequences,57 contrast enhancement,46 and molecular imaging47,48 may provide complementary structural information and allow more detailed plaque characterization. New and improved blood suppression methods50 are necessary for accurate plaque imaging, especially in the carotid artery bifurcation.
MR Coronary Angiography
Various new CMRA techniques await further clinical trials to determine their effectiveness. For example, Stuber et al58 combined a dual-inversion 3D FSE imaging sequence with real-time navigator technology for high-resolution, free-breathing black-blood CMRA. Deshpande et al59 evaluated the effectiveness of an extracellular contrast agent with the use of breathhold segmented echoplanar imaging and found substantial improvement in the delineation of the coronary arteries. Segmented 3D SSFP was also described as a promising technique with substantially increased signal-to-noise ratio and contrast-to-noise ratio for coronary artery imaging compared with a conventional gradient-echo technique with the same imaging time.60 New intravascular contrast agents may provide the long-awaited boost for reliable magnetic resonance coronary angiography. Initial studies in subjects have shown promising results.61
CT and MR Imaging Combination
Cross-sectional modalities MRI and CT offer more information than just displays of the patient coronary vessel lumen. The advantage of MDCT is the potentially complete assessment of the entire coronary artery tree within a very short scan time, and MRI offers excellent soft tissue contrast. However, because of the limited scan range and long examination time for MRI, MDCT may first be used to localize suspicious CAD lesions in the coronary arteries. With the knowledge of the problematic site in the coronary arteries from MDCT, this may be followed by MRI for further lesion assessment.
Improved CT and MR imaging of the coronaries may in the future allow for better noninvasive evaluation of atherosclerotic plaques. For example, a pilot study is being conducted (Figure 8) for the characterization of plaques with the use of CT and MR in high-risk asymptomatic patients. This study involves CT, CS measurements, contrast-enhanced CT for plaque burden and stenosis assessment, and multicontrast MR plaque characterization for a detailed analysis and follow-up of the composition of coronary atherosclerotic lesions. The flow chart in Figure 8 is not a clinical recommendation or guideline for practice but serves for a better understanding of the utility of the noninvasive combined CT and MR imaging modalities for future therapeutic strategies. Further studies are needed to test the concepts we are presenting here.
|
| Conclusions |
|---|
|
|
|---|
Finally, imaging may address the high-risk plaque, as described in this paper, but it does not take into account the blood hypercoagulable state or markers of inflammation. Therefore, one of the ultimate goals for the clinicians is the identification of the high-risk patient through a combination of strategies such as assessment of conventional risk factors, blood markers, and imaging.
6269
|
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
2. Stary HC. The development of calcium deposits in atherosclerotic lesions and their persistence after lipid regression. Am J Cardiol. 2001; 88: 16E19E.[Medline] [Order article via Infotrieve]
3. Fayad ZA, Fuster V. Clinical imaging of the high-risk or vulnerable atherosclerotic plaque. Circ Res. 2001; 89: 305316.
4. Jakobs TF, Becker CR, Ohnesorge B, et al. Multislice helical CT of the heart with retrospective ECG gating: reduction of radiation exposure by ECG-controlled tube current modulation. Eur Radiol. 2002; 12: 10811086.[CrossRef][Medline] [Order article via Infotrieve]
5. Schmermund A, Erbel R. Unstable coronary plaque and its relation to coronary calcium. Circulation. 2001; 104: 16821687.
6. Agatston AS, Janowitz WR, Hildner FJ, et al. Quantification of coronary artery calcium using ultrafast computed tomography. J Am Coll Cardiol. 1990; 15: 827832.[Abstract]
7. ORourke RA, Brundage BH, Froelicher VF, et al. American College of Cardiology/American Heart Association Expert Consensus document on electron-beam computed tomography for the diagnosis and prognosis of coronary artery disease. Circulation. 2000; 102: 126140.
8. Callister TQ, Raggi P, Cooil B, et al. Effect of HMG-CoA reductase inhibitors on coronary artery disease as assessed by electron-beam computed tomography. N Engl J Med. 1998; 339: 19721978.
9. Devries S, Wolfkiel C, Shah V, et al. Reproducibility of the measurement of coronary calcium with ultrafast computed tomography. Am J Cardiol. 1995; 75: 973975.[CrossRef][Medline] [Order article via Infotrieve]
10. Callister TQ, Cooil B, Raya SP, et al. Coronary artery disease: improved reproducibility of calcium scoring with an electron-beam CT volumetric method. Radiology. 1998; 208: 807814.
11. Wayhs R, Zelinger A, Raggi P. High coronary artery calcium scores pose an extremely elevated risk for hard events. J Am Coll Cardiol. 2002; 39: 225230.
12. OMalley PG, Taylor AJ, Gibbons RV, et al. Rationale and design of the Prospective Army Coronary Calcium (PACC) Study: utility of electron beam computed tomography as a screening test for coronary artery disease and as an intervention for risk factor modification among young, asymptomatic, active-duty United States Army Personnel. Am Heart J. 1999; 137: 932941.[CrossRef][Medline] [Order article via Infotrieve]
13. Becker CR, Kleffel T, Crispin A, et al. Coronary artery calcium measurement: agreement of multirow detector and electron beam CT. AJR Am J Roentgenol. 2001; 176: 12951298.
14. Becker CR, Ohnesorge BM, Schoepf UJ, et al. Current development of cardiac imaging with multidetector-row CT. Eur J Radiol. 2000; 36: 97103.[CrossRef][Medline] [Order article via Infotrieve]
15. Schroeder S, Kopp AF, Baumbach A, et al. Noninvasive detection and evaluation of atherosclerotic coronary plaques with multislice computed tomography. J Am Coll Cardiol. 2001; 37: 14301435.
16. Nikolaou K, Becker CR, Muders M, et al. High-resolution magnetic resonance and multi-slice CT imaging of coronary artery plaques in human ex vivo coronary arteries. Radiology. 2001; 221: 503.
17. Estes JM, Quist WC, Lo Gerfo FW, et al. Noninvasive characterization of plaque morphology using helical computed tomography. J Cardiovasc Surg (Torino). 1998; 39: 527534.[Medline] [Order article via Infotrieve]
18. Schmermund A, Rensing BJ, Sheedy PF, et al. Intravenous electron-beam computed tomographic coronary angiography for segmental analysis of coronary artery stenoses. J Am Coll Cardiol. 1998; 31: 15471554.
19. Hong C, Becker CR, Huber A, et al. ECG-gated, retrospectively reconstructed multidetector-row CT coronary angiography: effect of varying trigger delay on image quality. Radiology. 2001; 220: 712717.
20. Vogl TJ, Abolmaali ND, Diebold T, et al. Techniques for the detection of coronary atherosclerosis: multi-detector row CT coronary angiography. Radiology. 2002; 223: 212220.
21. Achenbach S, Giesler T, Ropers D, et al. Detection of coronary artery stenoses by contrast-enhanced, retrospectively electrocardiographically-gated, multislice spiral computed tomography. Circulation. 2001; 103: 25352538.
22. Niemann K, Oudkerk M, Rensing BJ, et al. Coronary Angiography with Multi-Slice Computed Tomography. Lancet. 2001; 357: 599603.[CrossRef][Medline] [Order article via Infotrieve]
23. Knez A, Becker CR, Leber A, et al. Usefulness of multislice spiral computed tomography angiography for determination of coronary artery stenoses. Am J Cardiol. 2001; 88: 11911194.[CrossRef][Medline] [Order article via Infotrieve]
24. Ohnesorge B, Flohr T, Klingenbeck-Regn K. Efficient object scatter correction algorithm for third and fourth generation CT scanners. Eur Radiol. 1999; 9: 563569.[CrossRef][Medline] [Order article via Infotrieve]
25. Fayad ZA, Worthley SG, Helft G, et al. Magnetic resonance imaging of the vasculature system.In: Fuster V, Alexander RW, ORourke RA, eds. Hursts The Heart. New York, NY: McGraw-Hill; 2001: 609627.
26. Fayad ZA, Nahar T, Fallon JT, et al. In vivo MR evaluation of atherosclerotic plaques in the human thoracic aorta: a comparison with TEE. Circulation. 2000; 101: 25032509.
27. Hatsukami TS, Ross R, Polissar NL, et al. Visualization of fibrous cap thickness and rupture in human atherosclerotic carotid plaque In vivo with high-resolution magnetic resonance imaging. Circulation. 2000; 102: 959964.
28. Yuan C, Mitsumori LM, Beach KW, et al. carotid atherosclerotic plaque: noninvasive MR characterization and identification of vulnerable lesions. Radiology. 2001; 221: 285299.
29. Manning WJ, Li W, Edelman RR. A preliminary report comparing magnetic resonance coronary angiography with conventional angiography. N Engl J Med. 1993; 328: 828832.
30. Duerinckx A, ed. Coronary Magnetic Resonance Angiography. New York, NY: Springer-Verlag; 2002.
31. Post JC, van Rossum AC, Hofman MBM, et al. Clinical utility of two-dimensional magnetic resonance angiography in detecting coronary artery disease. Eur Heart J. 1997; 18: 426433.
32. Post JC, van Rossum AC, Hofman MBM, et al. Three-dimensional respiratory-gated MR angiography of coronary arteries: comparison with conventional coronary angiography. Am J Roentgenol. 1996; 166: 13991404.
33. Müller MF, Fleisch M, Kroeker R, et al. Proximal coronary artery stenosis: three-dimensional MRI with fat saturation and navigator echo. J Magn Reson Imaging. 1997; 7: 644651.[Medline] [Order article via Infotrieve]
34. van Geuns RJ, de Bruin HG, Rensing BJ, et al. Magnetic resonance imaging of the coronary arteries: clinical results from three dimensional evaluation of a respiratory gated technique. Heart. 1999; 82: 515519.
35. Sandstede JJ, Pabst T, Beer M, et al. Three-dimensional MR coronary angiography using the navigator technique compared with conventional coronary angiography. AJR Am J Roentgenol. 1999; 172: 135139.
36. Sardanelli F, Molinari G, Zandrino F, et al. Three-dimensional, navigator-echo MR coronary angiography in detecting stenoses of the major epicardial vessels, with conventional coronary angiography as the standard of reference. Radiology. 2000; 214: 808814.
37. Nikolaou K, Huber A, Knez A, et al. Navigator echo-based respiratory gating for 3D-MR coronary angiography: Reduction of scan time using a slice-interpolation technique. J Comput Assist Tomogr. 2001; 25: 378387.[CrossRef][Medline] [Order article via Infotrieve]
38. Huber A, Nikolaou K, Gonschior P, et al. Navigator echo-based respiratory gating for three-dimensional MR coronary angiography: results from healthy volunteers and patients with proximal coronary artery stenoses. Am J Roentgenol. 1999; 173: 95101.
39. Kim WY, Danias PG, Stuber M, et al. Coronary magnetic resonance angiography for the detection of coronary stenoses. N Engl J Med. 2001; 345: 18631869.
40. Regenfus M, Ropers D, Achenbach S, et al. Noninvasive detection of coronary artery stenosis using contrast-enhanced three-dimensional breath-hold magnetic resonance coronary angiography. J Am Coll Cardiol. 2000; 36: 4450.
41. van Geuns RJ, Wielopolski PA, de Bruin HG, et al. MR coronary angiography with breath-hold targeted volumes: preliminary clinical results. Radiology. 2000; 217: 270277.
42. Toussaint JF, LaMuraglia GM, Southern JF, et al. Magnetic resonance images lipid, fibrous, calcified, hemorrhagic, and thrombotic components of human atherosclerosis in vivo. Circulation. 1996; 94: 932938.
43. Fayad ZA, Fuster V, Fallon JT, et al. Noninvasive in vivo human coronary artery lumen and wall imaging using black-blood magnetic resonance imaging. Circulation. 2000; 102: 506510.
44. Yuan C, Mitsumori LM, Ferguson MS, et al. In vivo accuracy of multispectral magnetic resonance imaging for identifying lipid-rich necrotic cores and intraplaque hemorrhage in advanced human carotid plaques. Circulation. 2001; 104: 20512056.
45. Yuan C, Beach KW, Smith LH Jr, et al. Measurement of atherosclerotic carotid plaque size in vivo using high-resolution magnetic resonance imaging. Circulation. 1998; 98: 26662671.
46. Yuan C, Kerwin WS, Ferguson MS, et al. Contrast-enhanced high-resolution MRI for atherosclerotic carotid artery tissue characterization. J Magn Reson Imaging. 2002; 15: 6267.[CrossRef][Medline] [Order article via Infotrieve]
47. Ruehm SG, Corot C, Vogt P, et al. Magnetic resonance imaging of atherosclerotic plaque with ultrasmall superparamagnetic particles of iron oxide in hyperlipidemic rabbits. Circulation. 2001; 103: 415422.
48. Flacke S, Fischer S, Scott MJ, et al. Novel MRI contrast agent for molecular imaging of fibrin: implications for detecting vulnerable plaques. Circulation. 2001; 104: 12801285.
49. Worthley SG, Helft G, Fuster V, et al. Noninvasive in vivo magnetic resonance imaging of experimental coronary artery lesions in a porcine model. Circulation. 2000; 101: 29562961.
50. Botnar RM, Kim WY, Bornert P, et al. 3D coronary vessel wall imaging utilizing a local inversion technique with spiral image acquisition. Magn Reson Med. 2001; 46: 848854.[CrossRef][Medline] [Order article via Infotrieve]
51. McConnell MV, Aikawa M, Maier SE, et al. MRI of rabbit atherosclerosis in response to dietary cholesterol lowering. Arterioscler Thromb Vasc Biol. 1999; 19: 19561959.
52. Helft G, Worthley SG, Fuster V, et al. Progression and regression of atherosclerotic lesions: monitoring with serial noninvasive magnetic resonance imaging. Circulation. 2002; 105: 993998.
53. Corti R, Fayad ZA, Fuster V, et al. Effects of lipid-lowering by simvastatin on human atherosclerotic lesions: a longitudinal study by high-resolution, noninvasive magnetic resonance imaging. Circulation. 2001; 104: 249252.
54. Zhao XQ, Yuan C, Hatsukami TS, et al. Effects of prolonged intensive lipid-lowering therapy on the characteristics of carotid atherosclerotic plaques in vivo by MRI: a case-control study. Arterioscler Thromb Vasc Biol. 2001; 21: 16231629.
55. Achenbach S, Daniel WG. Noninvasive coronary angiography: an acceptable alternative? N Engl J Med. 2001; 345: 19091910.
56. Toussaint JF, Southern JF, Fuster V, et al. Water diffusion properties of human atherosclerosis and thrombosis measured by pulse field gradient nuclear magnetic resonance. Arterioscler Thromb Vasc Biol. 1997; 17: 542546.
57. Coombs BD, Rapp JH, Ursell PC, et al. Structure of plaque at carotid bifurcation: high-resolution MRI with histological correlation. Stroke. 2001; 32: 25162521.
58. Stuber M, Botnar RM, Kissinger KV, et al. Free-breathing black-blood coronary MR angiography: initial results. Radiology. 2001; 219: 278283.
59. Deshpande VS, Wielopolski PA, Shea SM, et al. Coronary artery imaging using contrast-enhanced 3D segmented EPI. J Magn Reson Imaging. 2001; 13: 676681.[CrossRef][Medline] [Order article via Infotrieve]
60. Deshpande VS, Shea SM, Laub G, et al. 3D magnetization-prepared true-FISP: a new technique for imaging coronary arteries. Magn Reson Med. 2001; 46: 494502.[CrossRef][Medline] [Order article via Infotrieve]
61. Klein C, Nagel E, Schnackenburg B, et al. The intravascular contrast agent Clariscan (NC 100150 injection) for 3D MR coronary angiography in patients with coronary artery disease. Magma. 2000; 11: 6567.[CrossRef][Medline] [Order article via Infotrieve]
62. Nakanishi T, Ito K, Imazu M, et al. Evaluation of coronary artery stenoses using electron-beam CT and multiplanar reformation. J Comput Assist Tomogr. 1997; 21: 121127.[CrossRef][Medline] [Order article via Infotrieve]
63. Achenbach S, Ropers D, Regenfus M, et al. Contrast enhanced electron beam computed tomography to analyse the coronary arteries in patients after acute myocardial infarction. Heart. 2000; 84: 489493.
64. Reddy GP, Chernoff DM, Adams JR, et al. Coronary artery stenoses: assessment with contrast-enhanced electron- beam CT and axial reconstructions. Radiology. 1998; 208: 167172.
65. Budoff MJ, Oudiz RJ, Zalace CP, et al. Intravenous three-dimensional coronary angiography using contrast enhanced electron beam computed tomography. Am J Cardiol. 1999; 83: 840845.[CrossRef][Medline] [Order article via Infotrieve]
66. Achenbach S, Moshage W, Ropers D, et al. Value of electron-beam computed tomography for the noninvasive detection of high-grade coronary-artery stenoses and occlusions. N Engl J Med. 1998; 339: 19641971.
67. Kessler W, Achenbach S, Moshage W, et al. Usefulness of respiratory gated magnetic resonance coronary angiography in assessing narrowings > or=50% in diameter in native coronary arteries and in aortocoronary bypass conduits. Am J Cardiol. 1997; 80: 989993.[CrossRef][Medline] [Order article via Infotrieve]
68. Gonschior P, Pragst I, Valassis G, et al. High-resolution MR angiography: results in diseased arteries. J Invasive Cardiol. 2001; 13: 151157.[Medline] [Order article via Infotrieve]
69. Lethimonnier F, Furber A, Morel O, et al. Three-dimensional coronary artery MR imaging using prospective real-time respiratory navigator and linear phase shift processing: comparison with conventional coronary angiography. Magn Reson Imaging. 1999; 17: 11111120.[CrossRef][Medline] [Order article via Infotrieve]
This article has been cited by other articles:
![]() |
H. Brodoefel, C. Burgstahler, A. Sabir, C.-S. Yam, F. Khosa, C. D. Claussen, and M. E. Clouse Coronary Plaque Quantification by Voxel Analysis: Dual-Source MDCT Angiography Versus Intravascular Sonography Am. J. Roentgenol., March 1, 2009; 192(3): W84 - W89. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Ohayon, G. Finet, A. M. Gharib, D. A. Herzka, P. Tracqui, J. Heroux, G. Rioufol, M. S. Kotys, A. Elagha, and R. I. Pettigrew Necrotic core thickness and positive arterial remodeling index: emergent biomechanical factors for evaluating the risk of plaque rupture Am J Physiol Heart Circ Physiol, August 1, 2008; 295(2): H717 - H727. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Shinohara, T. Yamashita, H. Tawa, M. Takeda, N. Sasaki, T. Takaya, R. Toh, A. Takeuchi, T. Ohigashi, K. Shinohara, et al. Atherosclerotic plaque imaging using phase-contrast X-ray computed tomography Am J Physiol Heart Circ Physiol, February 1, 2008; 294(2): H1094 - H1100. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Ohayon, O. Dubreuil, P. Tracqui, S. Le Floc'h, G. Rioufol, L. Chalabreysse, F. Thivolet, R. I. Pettigrew, and G. Finet Influence of residual stress/strain on the biomechanical stability of vulnerable coronary plaques: potential impact for evaluating the risk of plaque rupture Am J Physiol Heart Circ Physiol, September 1, 2007; 293(3): H1987 - H1996. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Abbate, G. G.L. Biondi-Zoccai, P. Agostoni, M. J. Lipinski, and G. W. Vetrovec Recurrent angina after coronary revascularization: a clinical challenge Eur. Heart J., May 1, 2007; 28(9): 1057 - 1065. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. S. Gidding, C. A. McMahan, H. C. McGill, L. A. Colangelo, P. J. Schreiner, O. D. Williams, and K. Liu Prediction of Coronary Artery Calcium in Young Adults Using the Pathobiological Determinants of Atherosclerosis in Youth (PDAY) Risk Score: The CARDIA Study Arch Intern Med, November 27, 2006; 166(21): 2341 - 2347. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. R. Elmaleh, A. J. Fischman, A. Tawakol, A. Zhu, T. M. Shoup, U. Hoffmann, A.-L. Brownell, and P. C. Zamecnik Detection of inflamed atherosclerotic lesions with diadenosine-5',5'''-P1,P4-tetraphosphate (Ap4A) and positron-emission tomography PNAS, October 24, 2006; 103(43): 15992 - 15996. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Nikolaou, C. Rist, B. J. Wintersperger, T. F. Jakobs, R. van Gessel, M. A. Kirchin, A. Knez, F. von Ziegler, M. F. Reiser, and C. R. Becker Clinical value of MDCT in the diagnosis of coronary artery disease in patients with a low pretest likelihood of significant disease. Am. J. Roentgenol., June 1, 2006; 186(6): 1659 - 1668. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Dey, T. Callister, P. Slomka, F. Aboul-Enein, H. Nishina, X. Kang, H. Gransar, N. D. Wong, R. Miranda-Peats, S. Hayes, et al. Computer-Aided Detection and Evaluation of Lipid-Rich Plaque on Noncontrast Cardiac CT Am. J. Roentgenol., June 1, 2006; 186(6_Supplement_2): S407 - S413. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Achenbach and W. G. Daniel Reply J. Am. Coll. Cardiol., February 21, 2006; 47(4): 891 - 892. [Full Text] [PDF] |
||||
![]() |
P. Raggi, A. Taylor, Z. Fayad, D. O'Leary, S. Nissen, D. Rader, and L. J. Shaw Atherosclerotic Plaque Imaging: Contemporary Role in Preventive Cardiology Arch Intern Med, November 14, 2005; 165(20): 2345 - 2353. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Fuster, Z. A. Fayad, P. R. Moreno, M. Poon, R. Corti, and J. J. Badimon Atherothrombosis and High-Risk Plaque: Part II: Approaches by Noninvasive Computed Tomographic/Magnetic Resonance Imaging J. Am. Coll. Cardiol., October 4, 2005; 46(7): 1209 - 1218. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Fuster and R. J. Kim Frontiers in Cardiovascular Magnetic Resonance Circulation, July 5, 2005; 112(1): 135 - 144. [Full Text] [PDF] |
||||
![]() |
B. Bertrand, F. Esteve, H. Elleaume, C. Nemoz, S. Fiedler, A. Bravin, G. Berruyer, T. Brochard, M. Renier, J. Machecourt, et al. Comparison of synchrotron radiation angiography with conventional angiography for the diagnosis of in-stent restenosis after percutaneous transluminal coronary angioplasty Eur. Heart J., July 1, 2005; 26(13): 1284 - 1291. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. A. McMahan, S. S. Gidding, Z. A. Fayad, A. W. Zieske, G. T. Malcom, R. E. Tracy, J. P. Strong, H. C. McGill Jr, and for the Pathobiological Determinants of Atheroscle Risk Scores Predict Atherosclerotic Lesions in Young People Arch Intern Med, April 25, 2005; 165(8): 883 - 890. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. E. Sousa, M. A. Costa, E. M. Tuzcu, J. S. Yadav, and S. Ellis New Frontiers in Interventional Cardiology Circulation, February 8, 2005; 111(5): 671 - 681. [Full Text] [PDF] |
||||
![]() |
S. Paul, T. Mihaljevic, M. Leacche, M. Landzberg, C. Y. Ho, G. J. Blake, and J. G. Byrne Postinfarction Ventricular Septal Defect With Pseudoaneurysm Repair After Failed Percutaneous Closure Ann. Thorac. Surg., February 1, 2005; 79(2): 701 - 703. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Glaser, F. Selzer, D. P. Faxon, W. K. Laskey, H. A. Cohen, J. Slater, K. M. Detre, and R. L. Wilensky Clinical Progression of Incidental, Asymptomatic Lesions Discovered During Culprit Vessel Coronary Intervention Circulation, January 18, 2005; 111(2): 143 - 149. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. R. Moreno and V. Fuster The year in atherothrombosis J. Am. Coll. Cardiol., December 7, 2004; 44(11): 2099 - 2110. [Full Text] [PDF] |
||||
![]() |
J. F. Viles-Gonzalez, M. Poon, J. Sanz, T. Rius, K. Nikolaou, Z. A. Fayad, V. Fuster, and J. J. Badimon In Vivo 16-Slice, Multidetector-Row Computed Tomography for the Assessment of Experimental Atherosclerosis: Comparison With Magnetic Resonance Imaging and Histopathology Circulation, September 14, 2004; 110(11): 1467 - 1472. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Schoenhagen, S. S. Halliburton, A. E. Stillman, S. A. Kuzmiak, S. E. Nissen, E. M. Tuzcu, and R. D. White Noninvasive Imaging of Coronary Arteries: Current and Future Role of Multi-Detector Row CT Radiology, July 1, 2004; 232(1): 7 - 17. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. J. Schoepf, C. R. Becker, B. M. Ohnesorge, and E. K. Yucel CT of Coronary Artery Disease Radiology, July 1, 2004; 232(1): 18 - 37. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Sirol, V. V. Itskovich, V. Mani, J. G. S. Aguinaldo, J. T. Fallon, B. Misselwitz, H.-J. Weinmann, V. Fuster, J.-F. Toussaint, and Z. A. Fayad Lipid-Rich Atherosclerotic Plaques Detected by Gadofluorine-Enhanced In Vivo Magnetic Resonance Imaging Circulation, June 15, 2004; 109(23): 2890 - 2896. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. A. Wickline Plaque characterization: surrogate markers or the real thing? J. Am. Coll. Cardiol., April 7, 2004; 43(7): 1185 - 1187. [Full Text] [PDF] |
||||
![]() |
S. Achenbach, D. Ropers, U. Hoffmann, B. MacNeill, U. Baum, K. Pohle, T. J. Brady, E. Pomerantsev, J. Ludwig, F. A. Flachskampf, et al. assessment of coronary remodeling in stenotic and nonstenotic coronary atherosclerotic lesions by multidetector spiral computed tomography J. Am. Coll. Cardiol., March 3, 2004; 43(5): 842 - 847. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. H. K. Hoffmann, H. Shi, F. T. Schmid, H. Gelman, H.-J. Brambs, and A. J. Aschoff Noninvasive Coronary Imaging with MDCT in Comparison to Invasive Conventional Coronary Angiography: A Fast-Developing Technology Am. J. Roentgenol., March 1, 2004; 182(3): 601 - 608. [Full Text] [PDF] |
||||
![]() |
S. Achenbach, F. Moselewski, D. Ropers, M. Ferencik, U. Hoffmann, B. MacNeill, K. Pohle, U. Baum, K. Anders, I.-k. Jang, et al. Detection of Calcified and Noncalcified Coronary Atherosclerotic Plaque by Contrast-Enhanced, Submillimeter Multidetector Spiral Computed Tomography: A Segment-Based Comparison With Intravascular Ultrasound Circulation, January 6, 2004; 109(1): 14 - 17. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. C. McGill Jr and C. A. McMahan Starting Earlier to Prevent Heart Disease JAMA, November 5, 2003; 290(17): 2320 - 2322. [Full Text] [PDF] |
||||
![]() |
P. M. Winter, A. M. Morawski, S. D. Caruthers, R. W. Fuhrhop, H. Zhang, T. A. Williams, J. S. Allen, E. K. Lacy, J. D. Robertson, G. M. Lanza, et al. Molecular Imaging of Angiogenesis in Early-Stage Atherosclerosis With {alpha}v{beta}3-Integrin-Targeted Nanoparticles Circulation, November 4, 2003; 108(18): 2270 - 2274. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. K. Pannu, T. G. Flohr, F. M. Corl, and E. K. Fishman Current Concepts in Multi-Detector Row CT Evaluation of the Coronary Arteries: Principles, Techniques, and Anatomy RadioGraphics, October 1, 2003; 23(90001): S111 - 125. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. F. Fearon, M. Nakamura, D. P. Lee, M. Rezaee, R. H. Vagelos, S. A. Hunt, P. J. Fitzgerald, P. G. Yock, and A. C. Yeung Simultaneous Assessment of Fractional and Coronary Flow Reserves in Cardiac Transplant Recipients: Physiologic Investigation for Transplant Arteriopathy (PITA Study) Circulation, September 30, 2003; 108(13): 1605 - 1610. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. M. Tuzcu and P. Schoenhagen Acute coronary syndromes, plaque vulnerability,and carotid artery disease: The changing role ofatherosclerosis imaging J. Am. Coll. Cardiol., September 17, 2003; 42(6): 1033 - 1036. [Full Text] [PDF] |
||||
![]() |
D. B. Mark, L. J. Shaw, M. S. Lauer, P. G. O'Malley, and P. Heidenreich Task force #5--is atherosclerosis imaging cost effective? J. Am. Coll. Cardiol., June 4, 2003; 41(11): 1906 - 1917. [Full Text] [PDF] |
||||
![]() |
A. Maseri and V. Fuster Is There a Vulnerable Plaque? Circulation, April 29, 2003; 107(16): 2068 - 2071. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2002 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |