(Circulation. 2002;106:I-33.)
© 2002 American Heart Association, Inc.
Surgery for Valvular Heart Disease |
From the Department of Cardiovascular Surgery (T.A.T., P.D., D.T.L., F.T., G.T.D., N.B.I, D.C.M.), Division of Cardiovascular Medicine (D.L., M.H.), Stanford University School of Medicine, Stanford, Calif., and Laboratory of Cardiovascular Physiology and Biophysics, Research Institute of the Palo Alto Medical Foundation, Palo Alto, Calif. (G.T.D., N.B.I.).
Correspondence to D. Craig Miller, MD, Falk Cardiovascular Research Center, Stanford University School of Medicine, Stanford, CA 94305-5407. E-mail dcm{at}stanford.edu
| Abstract |
|---|
|
|
|---|
Methods Radiopaque markers were placed on the LV, mitral annulus (MA), and leaflets in 13 sheep. Seven sheep served as controls, and 6 had a St. Jude Tailor partial flexible ring implanted (29 mm in 5, 31 mm in 1). After 8±1 day, the animals were studied with biplane videofluoroscopy and echocardiography before and during acute posterolateral LV ischemia (balloon occlusion of circumflex artery). Mitral annular area (MAA), septal-lateral annular diameter (SL), annular perimeters, and leaflet edge separation were calculated from 3-D marker coordinates.
Results The average degree of mitral regurgitation increased from 0.0±0.0 to 2.1±0.7 (P=0.0006) in the control group during acute ischemia but remained unchanged in the Tailor group (0.1±0.2 for both conditions). The change in MAA throughout the cardiac cycle before ischemia was 17±4% in control animals, but only 5±2% (P=0.0002) in the Tailor ring group. Unlike the control animals, there was no increase in MAA (5.4±0.8 and 5.5±0.7 cm2, respectively; p=NS) nor dilatation of the muscular annulus (6.2±0.3 and 6.2±0.4, respectively; p=NS) during ischemia with the Tailor ring. Mitral SL dimension increased slightly with ischemia (2.3±0.2 versus 2.2±0.2 cm, P=0.03). Although posterior leaflet motion was limited, as observed with complete rings, normal annular flexion was maintained with the Tailor ring before and during acute ischemia.
Conclusions The Tailor partial annuloplasty ring prevented acute IMR probably by limiting SL diameter dilatation during acute ischemia. In this animal model of acute IMR, a partial, flexible posterior annuloplasty ring is as effective as a complete ring.
Key Words: ischemia mitral valve regurgitation valvuloplasty
| Introduction |
|---|
|
|
|---|
Cosgrove and colleagues introduced a partial flexible annuloplasty band that offers annular size reduction yet appears to preserve normal annular dynamics.15,16 A similar partial annuloplasty ring, the Tailor partial flexible annuloplasty band, preserved normal annular flexion in experimental ovine experiments, although mitral annular area was fixed during the cardiac cycle.17
As annular dilatation and displacement primarily of the posterior annulus are associated with ischemic mitral regurgitation,1820 partial posterior ring annuloplasty may offer a simpler and more physiologic repair. Although partial ring annuloplasty has been used in patients with IMR,15,21 its efficacy in models of acute LV ischemia has not been assessed. Therefore, we investigated a partial posterior flexible annuloplasty ring in an acute model of LV ischemia in sheep. This current study extends the observations of our previous report17 to include the response to acute posterolateral LV ischemia.
| Methods |
|---|
|
|
|---|
|
Data Acquisition
The animals were studied during closed-chest conditions, sedated with ketamine and diazepam, intubated and mechanically ventilated (veterinary anesthesia ventilator 2000, Hallowell EMC) with 100% oxygen. Esmolol (20 to 50 µg/kg/min) and atropine sulfate (0.01 mg/kg/min) intravenous infusions were utilized to minimize reflex sympathetic and parasympathetic responses. Left circumflex coronary balloon occlusion (distal to the first obtuse marginal artery) to produce posterolateral LV ischemia was carried out.20 Simultaneous biplane videofluoroscopy and hemodynamic data recordings were obtained during steady-state conditions and over a range of LV filling volumes during abrupt preload reduction using vena caval occluders. Data recordings were acquired before and after balloon circumflex occlusion. Resultant IMR was graded subjectively by an experienced echocardiographer (D. Liang) according to the extent and width of the regurgitant jet and categorized as none (0), mild (+1), moderate (+2), moderate to severe (+3), or severe (+4).
All animals received humane care in compliance with the "Principles of Laboratory Animal Care" formulated by the National Society for Medical Research and the "Guide for Care and Use of Laboratory Animals" prepared by the National Academy of Sciences and published by the National Institutes of Health (DHEW NIHG publication 85 to 23, revised 1985). This study was approved by the Stanford Medical Center Laboratory Research Animal Review committee and conducted according to Stanford University policy.
A Philips Optimus 2000 biplane Lateral ARC 2/poly DIAGNOST C2 system (Philips Medical Systems, North America Company) was used to record videofluoroscopic data at 60 Hz with the image intensifiers in the 9-inch mode. Two-dimensional images from each of the 2 x-ray views (45° right anterior oblique and 45° left anterior oblique) were digitized and then merged to yield 3-D coordinates for each marker every 16.7 ms.23 LV pressure, aortic pressure, and ECG voltage were digitized and recorded simultaneously.
Data Analysis
End-systole was defined as the videofluoroscopic frame preceding maximum negative dP/dt (-dP/dtmax); end-diastole was defined as the videofluoroscopic frame containing the peak of the ECG R-wave. An instantaneous estimate of LV volume was calculated every 16.7 msec from the epicardial LV markers using a multiple tetrahedral model reconstructed from the marker coordinates.24 External LV pressure-volume stroke work (SW) was calculated as the integral of LV pressure (P) on volume (V) over a cardiac cycle for several beats at baseline and during caval occlusion as:
|
|
Preload recruitable stroke work (PRSW) was computed by linear regression of SW on EDV as:
|
|
Where MW and VW are the slope and volume axis intercept, respectively. Adequate volume occlusion for PRSW calculations was obtained in 6 animals for the control and 4 animals in the Tailor groups.
Mitral Annular Geometry
Mitral annular area, perimeters, and dimensions where computed from the 3-D marker coordinates without assuming circular or planar geometry. Mitral annular area was calculated by first determining the centroid of the annular markers and then dividing the annulus into 8 triangular areas which were summed to yield the total annular area. The septal-lateral (SL) diameter of the annulus was calculated as the distance in 3-D space between the 2 markers placed in the middle of the septal and lateral mitral annulus, respectively (#1 and #5, Figure 2). The commissure-commissure (CC) diameter was calculated as the distance between the 2 annular commissural markers (#3 and #7, Figure 2). The muscular annular perimeter was computed as the sum of the 6 contiguous mitral annular segment lengths enclosed between markers #2 and #8. The fibrous annular perimeter was computed from the sum of segment lengths enclosed between markers #1 and #2, and markers #1 and #8 with markers #2 and #8 representing the right and left fibrous trigones, respectively. Flexion of the anterior annulus was expressed as the change in elevation during systole of the "saddle horn" marker (#1) above the plane fitted to the posterior annular markers using least-squares estimation.
|
Mitral Leaflet Dynamics
To assess leaflet mobility throughout the cardiac cycle, angular position of the anterior leaflet edge was calculated as the angle (
AML) between the line from the anterior leaflet edge marker to the "saddle horn" and the septal-lateral annular diameter. Posterior leaflet edge angular position (
PML) was calculated in similar fashion. Leaflet excursion was calculated from diastolic maximum to systolic minimum angle. Anterior and posterior leaflet lengths were measured as the sum of distances between adjacent markers placed on the central meridian of each leaflet. Leaflet edge separation was determined as the distance in 3-D space between the 2 edge markers on the anterior and posterior mitral leaflets with valve closure defined as the time at which this distance reached its minimum plateau. Because of loss of the marker on the anterior leaflet edge in some animals, leaflet separation distance was calculated in 5 control and 4 Tailor animals.
Statistical Analysis
All data are reported as mean plus or minus 1 standard deviation (±1 SD). Hemodynamic and marker-derived data from 2 to 3 consecutive steady state beats that were time-aligned at end-diastole, and data from these beats were averaged for each animal. The data were analyzed for 20 frames before and 20 frames after end-diastole. Data were compared using Students t test for paired comparisons.
| Results |
|---|
|
|
|---|
|
Mitral Annular Dynamics
Group mean data for mitral annular area and SL and CC diameters as well as lengths of the fibrous and muscular portions of the mitral annulus are tabulated in Table 2 and shown in Figures 3 and 4. During acute posterolateral ischemia in the control group, the mitral annulus (particularly in the SL diameter) dilated substantially, and the length of the muscular portion of the annulus increased significantly. Conversely, during circumflex occlusion in the Tailor group, annular area did not change, and there was no appreciable increase in either the SL diameter or the muscular annular perimeter. Mitral annular area changed only slightly during the cardiac cycle before and during ischemia in the animals fitted with the partial flexible annuloplasty ring. Although annular area remained fixed in the Tailor group, flexion of the septal annulus, as shown in Figure 5, was similar in both groups before ischemia (1.9±0.9 and 1.5±0.3 mm for control and Tailor, respectively; P=0.4) and did not change during circumflex occlusion.
|
|
|
|
Mitral Leaflet Dynamics
Group mean anterior and posterior leaflet edge angular positions during the cardiac cycle for both groups before ischemia are illustrated in Figure 6. Anterior leaflet excursion from minimum to maximum during the cardiac cycle was similar between the 2 groups (45±12 and 50±10 degrees for control and Tailor, respectively; P=0.5), posterior leaflet motion was severely limited in the Tailor group (38±3 versus 14±7 degrees; P=0.0001). Changes in leaflet excursion in both groups during ischemia along with valve closure times and lengths of the anterior and posterior leaflets at end-systole are summarized in Table 3.pick;0253t3;0;;;ZPICKFOOT;T3>During circumflex occlusion, posterior leaflet excursion decreased in control animals but did not change in the Tailor group. The observed mitral regurgitation in the control group was associated with delayed valve closure and early systolic leaflet "loitering" (Figure 7). In the Tailor group, although there was a trend toward later valve closure at baseline compared with the control group (P=0.054), delayed valve closure or leaflet "loitering" during acute posterolateral LV ischemia did not occur.
|
|
|
| Discussion |
|---|
|
|
|---|
Previous studies demonstrated that complete ring annuloplasty abolished acute IMR by reducing the annular septal-lateral (or, in clinical jargon, "antero-posterior") dimension and prevented mitral leaflet "loitering"11 and perturbations of leaflet geometry.27 The Tailor ring effectively fixed annular size throughout the cardiac cycle and did not permit dilatation of the muscular annulus during acute ischemia. Consequently, no IMR was observed, and the mitral septal-lateral annular dimension increased only slightly; which is the main direction of annular enlargement during acute posterolateral LV ischemia.18,20 Although the Tailor ring is a partial ring, annular dilatation was not observed with circumflex occlusion because the fibrous annulus did not change: The dynamics of the fibrous annulus did not change throughout the cardiac cycle in either group, nor did they change during ischemia. Therefore, perhaps all that is needed in annular reduction procedures aimed at correcting IMR is reduction of the muscular annulus; however, caution must be exercised because dilatation of the fibrous annulus has actually been observed in certain circumstances, for example, patients with28 and experimental models of dilated cardiomyopathy.29 Avoiding circumferential reduction of the mitral annulus may be related to preserved flexion of the anterior annulus that we recently reported in animals with a Tailor ring, as such was not seen with a complete flexible ring;17 maintained annular flexion may play a role in reducing the likelihood of systolic anterior motion of the anterior mitral leaflet which might cause postoperative LV outflow tract obstruction after ring annuloplasty.30 In the current experiment, annular flexion was also conserved with the Tailor ring during acute posterolateral LV ischemia.
Complete ring annuloplasty abolishes leaflet "loitering" and prevents delayed leaflet coaptation in models of acute ovine IMR.11 The current data suggest that the Tailor ring is equally effective in facilitating timely valve closure and preventing IMR during acute LV ischemia; however, annular reduction with the Tailor ring was also associated with restriction of posterior leaflet motion as found previously with complete rings.13,14 We believe that the observed "frozen" posterior mitral leaflet in the setting of mitral annuloplasty is related to reduction of annular area and altered geometric relationships between the posterior mitral annulus and the papillary muscle tips. As the reduction in annular area induced by the Tailor ring is comparable to that achieved with complete rings,12 this limited posterior leaflet motion may not be surprising. Additionally, "reefing" of the posterior leaflet by the annuloplasty band must also be considered and may be reflected in the baseline differences in posterior leaflet lengths between the 2 groups. Extension of the anterior leaflet during acute IMR has been described in control sheep during acute IMR in a prior report27 and attributed to papillary muscle "tethering". In animals with complete annuloplasty rings, systolic anterior leaflet length remained unchanged during acute ischemia, a finding corroborated by the current study. Thus, the Tailor partial flexible annuloplasty ring has essentially the same effects on mitral leaflet motion and geometry during acute LV ischemia in this model as do complete flexible and semi-rigid annuloplasty rings, except that annular flexion is preserved.
| Summary |
|---|
|
|
|---|
Study Limitations
The results of the current study must be interpreted in the context of several limitations. This experimental design does not imitate clinical practice where prior ischemia or infarction is present and then is treated surgically with placement of a ring- in this study, an annuloplasty ring was implanted in a normal heart before the ischemic insult and therefore represented a "prophylactic" surgical correction that is distinctly different than the clinical situation. It is also important to emphasize that these findings can only be interpreted in the setting of acute LV ischemia in a normal sheep heart and should not be extrapolated directly to patients with chronic IMR where LV dilatation, sub-valvular geometric perturbations induced by previous infarction and ischemia, and LV remodeling may play a predominant role. This animal model also has other limitations in that sheep have a less well defined posterior annulus than do humans and more atrial tissue above and below the line of leaflet insertion;31 however, annular dynamics are similar in both humans and sheep.22,32 While the myocardial marker method requires suturing small metal markers to numerous intra-cardiac structures, previous echocardiographic studies show that the markers do not interfere with normal mitral annular or leaflet motion as they are very small (aggregate mass=20±6 mg). Although there were baseline differences in animal size and some hemodynamic parameters between the 2 groups, we believe that these differences did not contribute in an important way to the observed differences in annular and leaflet dynamics between the control sheep and the ring animals.
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
2. Reul RM, Cohn LH. Mitral valve reconstruction for mitral insufficiency. Prog Cardiovasc Dis. 1997; 39: 567599.[CrossRef][Medline] [Order article via Infotrieve]
3. Yau TM E-GY, Armstrong S, Ivanov J, et al. Mitral valve repair and replacement for rheumatic disease. J Thorac Cardiovasc Surg. 2000;119:5360.
4. Rankin JS, Feneley MP, Hickey MS, et al. A clinical comparison of mitral valve repair versus valve replacement in ischemic mitral regurgitation. J Thorac Cardiovasc Surg. 1988; 95: 165177.[Abstract]
5. Replogle RL, Campbell CD. Surgery for mitral regurgitation associated with ischemic heart disease. Results and strategies. Circulation. 1989; 79: I122I25.
6. Dion R. Ischemic mitral regurgitation: when and how should it be corrected? J Heart Valve Dis. 1993; 2: 536543.[Medline] [Order article via Infotrieve]
7. David TE. Techniques and results of mitral valve repair for ischemic mitral regurgitation. J Card Surg. 1994; 9: 274277.[Medline] [Order article via Infotrieve]
8. Czer LS, Maurer G, Trento A, et al. Comparative efficacy of ring and suture annuloplasty for ischemic mitral regurgitation. Circulation. 1992; 86: II46I52.
9. Cohn LH, Rizzo RJ, Adams DH, et al. The effect of pathophysiology on the surgical treatment of ischemic mitral regurgitation: operative and late risks of repair versus replacement. Eur J Cardiothorac Surg. 1995; 9: 568754.[Abstract]
10. Grossi EA, Lapietra A, Galloway AC, et al. Late results of isolated mitral annuloplasty for ischemic mitral insufficiency. Circulation. 2000; 102: II-491 Abstract.
11. Timek T, Glasson JR, Dagum P, et al. Ring annuloplasty prevents delayed leaflet coaptation and mitral regurgitation during acute left ventricular ischemia. J Thorac Cardiovasc Surg. 2000; 119: 774783.
12. Glasson JR, Green GR, Nistal JF, et al. Mitral annular size and shape in sheep with annuloplasty rings. J Thorac Cardiovasc Surg. 1999; 117: 302309.
13. Green GR, Dagum P, Glasson JR, et al. Restricted posterior leaflet motion after mitral ring annuloplasty. Ann Thorac Surg. 1999; 68: 21002106.
14. van Rijk-Zwikker GL, Mast F, Schipperheyn JJ, et al. Comparison of rigid and flexible rings for annuloplasty of the porcine mitral valve. Circulation. 1990; 82: IV58IV64.
15. Cosgrove DM, Arcidi JM, Rodriguez L, et al. Initial experience with the Cosgrove-Edwards Annuloplasty System. Ann Thorac Surg. 1995; 60: 499503.
16. Gillinov AM, Cosgrove DM, Shiota T, et al. Cosgrove-Edwards Annuloplasty System: midterm results. Ann Thorac Surg. 2000; 69: 717721.
17. Dagum P, Timek T, Green GR, et al. Three-dimensional geometric comparison of partial and complete flexible mitral annuloplasty rings. J Thorac Cardiovasc Surg. 2001; 122: 665673.
18. Kay GL, Kay JH, Zubiate P, et al. Mitral valve repair for mitral regurgitation secondary to coronary artery disease. Circulation. 1986; 74: I8898.
19. Gorman JH, Gorman RC, Jackson BM, et al. Distortions of the mitral valve in acute ischemic mitral regurgitation. Ann Thorac Surg. 1997; 64: 10261031.
20. Glasson JR, Komeda M, Daughters GT, et al. Early systolic mitral leaflet "loitering" during acute ischemic mitral regurgitation. J Thorac Cardiovasc Surg. 1998; 116: 193205.
21. Camilleri LF, Miguel B, Bailly P, et al. Flexible posterior mitral annuloplasty: five-year clinical and Doppler echocardiographic results. Ann Thorac Surg. 1998; 66: 16921697.
22. Glasson JR, Komeda M, Daughters GT, et al. Most ovine mitral annular 3-D size reduction occurs before ventricular systole and is abolished with ventricular pacing. Circulation. 1997; 96: II-115II123.
23. Daughters GT, Sanders WJ, Miller DC, et al. A comparison of two analytical systems for 3-D reconstruction from biplane videoradiograms. IEEE Computers in Cardiology. 1989; 15: 7982.
24. Moon MR, DeAnda A, Daughters GT, et al. Experimental evaluation of different chordal preservation methods during mitral valve replacement. Ann Thorac Surg. 1994; 58: 931944.[Abstract]
25. Liel-Cohen N, Otsuji Y, Vlahakes GJ, et al. Functional ischemic mitral regurgitation can persist despite ring annuloplasty: mechanistic insights. Circulation. 1997;96:I-540. Abstract
26. DallAgata A, Taams MA, Fioretti PM, et al. Cosgrove-Edwards mitral ring dynamics measured with transesophageal three-dimensional echocardiography. Ann Thorac Surg. 1998; 65: 485490.
27. Lai DTM, Timek TA, Green GR, et al. The effects of ring annuloplasty on mitral leaflet geometry during acute left ventricular ischemia. J Thorac Cardiovasc Surg. 2000; 120: 966975.
28. Hueab AC, Jatene FB, Pomerantzeff PMA, et al. Mechanisms of mitral valve insufficiency and annular dilatation in dilated cardiomyopathy: anatomic study. 14th Annual Meeting of the European Association for Cardio-Thoracic Surgery October 711, 2000, Frankfurt, Germany. Abstract.
29. Timek TA, Dagum P, Lai DT, et al. 3-D Mitral annular dynamics in tachycardia induced cardiomyopathy. J Am Coll Cardiol. 2001;37:217A. Abstract
30. Dagum P, Green GR, Glasson JR, et al. Potential mechanism of left ventricular outflow tract obstruction after mitral ring annuloplasty. J Thorac Cardiovasc Surg. 1999; 117: 472480.
31. Walmsley R. Anatomy of human mitral valve in adult cadaver and comparative anatomy of the valve. Br Heart J. 1978; 40: 35166.
32. Ormiston JA, Shah PM, Tei C, et al. Size and motion of the mitral valve annulus in man. I. A two-dimensional echocardiographic method and findings in normal subjects. Circulation. 1981; 64: 113120.
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2002 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |