Circulation. 2002;106:I-189-I-192
doi: 10.1161/01.cir.0000032913.33237.f5
This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Rothenburger, M.
Right arrow Articles by Schmid, C.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Rothenburger, M.
Right arrow Articles by Schmid, C.

(Circulation. 2002;106:I-189.)
© 2002 American Heart Association, Inc.


Thoracic Transplantation and Mechanical Support for Congestive Heart Failure

Treatment of Thrombus Formation Associated With the MicroMed DeBakey VAD Using Recombinant Tissue Plasminogen Activator

Markus Rothenburger, MD; Markus J. Wilhelm, MD; Dieter Hammel, MD; Christoph Schmidt, MD; Tonny D. T. Tjan, MD; Dirk Böcker, MD; Hans H. Scheld, MD; Christof Schmid, MD

From the Departments of Thoracic and Cardiovascular Surgery (M.R., M.J.W., D.H., T.D.T.T., H.H.S., C. Schmid), Anesthesiology and Operative Intensive Care Medicine (C. Schmidt), and epartment of Cardiology and Angiology (D.B.), University-Hospital Muenster, Germany.

Correspondence to Markus Rothenburger, MD, Department of Thoracic and Cardiovascular Surgery, University Hospital, Albert-Schweitzer Str. 33, 48129 Muenster, Germany. E-mail markus.rothenburger{at}thgms.uni-muenster.de

Abstract

Background The latest generation of left ventricular assist devices consists of nonpulsatile impeller pumps. In these small pumps, thrombus formation inside the device does not lead to thromboembolic end-organ dysfunction but may dramatically impair pump flow. We report on our experience with thrombus-related pump dysfunctions of the MicroMed DeBakey left ventricular assist device and its treatment.

Methods Eight of 22 patients with a MicroMed DeBakey VAD presented with a critically reduced pump flow. In 7 cases, an increased power demand indicative of progressive thrombus formation associated with the device was evident, whereas 1 case presented with thrombus formation within the inflow conduit associated with a very low power demand. Brief spontaneously resolving pump stops had been noted in 6 patients. All 8 patients were treated with 100 mg of recombinant tissue plasminogen activator (rt-PA), administered via an IV line.

Results Rt-PA lysis led to an increase of pump flow along with a reduction of power demand within a short time in all patients. No severe bleeding complications occurred. However, 4 patients experienced transient epistaxis. All patients could be discharged from intensive care immediately after discontinuation of thrombolytic therapy.

Conclusion Rt-PA lysis is a very effective tool for thrombus-related pump dysfunction in patients with impeller pumps, which renders emergency surgical exchange unnecessary in most cases.


Key Words: MicroMed DeBakey LVAD • thrombus formation • thrombolytic therapy • recombinant tissue plasminogen activator (rt-PA)

Left ventricular assist devices (LVAD) are commonly used as a bridge to heart transplantation in end-stage heart failure patients. Most of the LVADs are based on a pulsatile blood flow, such as the Novacor N100 (Baxter, Oakland, Calif) or the HeartMate system (Thoratec, Pleasanton, Calif).1,2 It has been reported that up to 47% of all patients with a pulsatile LVAD develop thromboembolic complications during support.1,3,4 The MicroMed DeBakey VAD (MicroMed Technology, Houston, Tex) has been introduced recently and generates continuous blood flow by using an impeller pump.5,6 In these patients, the thromboembolic hazard seems to be much lower, but thrombus formation inside the left ventricle may lead to malfunction of the device, causing a high risk of fatal outcome. In such a situation, restoring adequate flow within a very short period is necessary. One possibility is surgical device exchange, but this is associated with all the problems of an urgent redo procedure. The aim of our study was to investigate thrombolysis as an alternative treatment strategy.

Methods

Patients
Twenty-two patients, 20 men and 2 women, were enrolled in the study. The age ranged from 17 to 63 years, with a mean of 43.7±14.3 years. In 8 patients, thrombus formation developed during long-term mechanical support with the DeBakey VAD. Demographic data and patient characteristics are described in table 1.


View this table:
[in this window]
[in a new window]
 
TABLE 1. Patient Characteristics and Clinical Data of the Patients With Thrombus Formation Associated With the MicroMed DeBakey VAD

Device
The MicroMed DeBakey LVAD achieves axial blood flow by a valveless rotary blood pump (7.500 to 12.500 rpm) with inflow from the left ventricular apex and an outflow graft into the ascending aorta.5

Anticoagulation Management
In all patients, the same anticoagulation protocol was applied, mainly based on heparin. Before implantation, a partial thromboplastin time of 40 to 60 seconds was maintained. During cardiopulmonary bypass, full heparinization was achieved routinely. Postoperatively, intravenous heparin administration was continued, the goal being a partial thromboplastin time of 60 to 100 seconds. After cessation of drainage losses and hemodynamic stabilization, all patients were switched to oral anticoagulation with an international normalized ratio (INR) between 2.5 and 3.5, and aspirin (300 mg) and dipyramidole (75 mg) were added.

Device Control
Device function was supervised by an external controller module with audible and visible alarm signals and a laptop-based Clinical Data Acquisition System. As recently described, this system stores pump-operating data and displays pump and physiologic information like flow rate (L/min), power demand (watts), pump speed (rpm), and current (amps).5

Diagnosis of Device-Related Thrombus Formation
Diagnosis of thrombus formation was mainly established by indirect evidence. In our early experience with the MicroMed DeBakey VAD, 2 patients underwent emergency device exchange, not thrombolytic therapy, because of markedly reduced pump flow and increased power demand. Recurrent pump stops had occurred before surgical treatment. The examination of the explanted devices demonstrated thrombotic material inside the MicroMed DeBakey VAD (Figures 1, 2). Because of this experience, diagnosis of thrombus formation inside or dislodgement into the device was made from characteristic changes of flow rate and power demand of the device; ie, decrease of flow rate and increase of power demand markedly differing from the patient’s baseline were considered indicative of thrombus formation. Transesophageal echocardiography always failed to visualize a thrombus adhering to the inflow conduit. An angiography was considered too dangerous in an acute low output situation. In 1 patient with an unusual low power demand and low flow, we suspected that a major thrombus formation within the inflow conduit was tremendously hindering pump inflow.



View larger version (139K):
[in this window]
[in a new window]
 
Figure 1. Thrombus formation in the inflow conduit of a MicroMed DeBakey VAD.



View larger version (50K):
[in this window]
[in a new window]
 
Figure 2. Thrombus formation around the impeller of an exchanged MicroMed DeBakey VAD.

Treatment With rt-PA
Thrombolytic treatment was standardized with an IV infusion of 100 mg recombinant tissue plasminogen activator (rt-PA). Initially, 50 mg of rt-PA was given as a bolus, followed by another 50 mg administered over an interval of 1 to 2 hours. Since partial thromboplastin time (PTT) frequently increases during rt-PA lysis, the IV heparin dose was lowered and coagulation parameters monitored hourly. Following rt-PA lysis, PTT was augmented from 60 seconds to 100 seconds, and antiplatelet therapy was intensified.

Results

Device-Related Complications
Eight of the 22 patients with a MicroMed DeBakey VAD developed 10 episodes suspicious of thrombus formation inside the device and 1 episode of suspected involvement of the inflow conduit (Table 2). Six patients had a history of recurrent pump stops.


View this table:
[in this window]
[in a new window]
 
TABLE 2. Effect of rt-PA Lysis on Device Function and Complications

Effect of Thrombolytic Therapy With rt-PA
Table 2 depicts the effects of thrombolytic therapy on device function as well as the associated adverse effects. Immediately after termination of thrombolytic therapy, usually within 3 hours, the aspired increase of flow rate and decrease of power demand could be noted (Table 2). All patients survived thrombolytic therapy. No severe bleeding complications were observed, and there was no evidence of intracranial hemorrhage. No device exchange was necessary after thrombolytic therapy with rt-PA. However, 4 patients experienced epistaxis for 3 days, and 1 patient needed a nasal tamponade for 2 days.

Repeat Thrombolysis and Outcome
One patient underwent 2 and another patient 3 thrombolytic therapies with rt-PA during long-term mechanical support. In all instances, device malfunction resolved completely and without severe complications, similar to previous treatments. Six patients could be forwarded to successful heart transplantation. Two patients died during LVAD support because of multiorgan failure after 258 and 128 days on support (142 and 75 days after thrombolytic therapy, respectively).

Discussion

The underlying cause of thrombus formation despite effective anticoagulation is still unclear. At explantation, we always observed some thrombus formation inside the inflow conduit and sometimes also within the pump. There seems to be no relationship with heating or an association with the bearings. Because the company presumed a problem with the inner surface of the inflow conduit, the latter has been coated with heparin. With these heparin-coated devices we have not experienced that problem. It remains to be seen whether this modification really prevents thrombus formation. Nonetheless, a thrombus inside the left ventricle may dislodge into the pump and lead to a pump dysfunction. A suboptimal implantation technique as a potential cause for increased thrombus formation reflecting a learning-curve effect can be excluded, since the surgical technique is more or less identical to the one for the first-generation devices such as the Novacor.7 After more than 6 months of LVAD support, cerebral embolization was evident in 2 patients and was unrelated to thrombolysis. Liver or spleen infarction never occurred. Even if it were proven, we think that a thrombus drawn into the pump is split into minuscule pieces, unless a pump stop occurs, which may explain the low rate of thromboembolism. It may well be that silent thromboembolism occurs, but there is no clinical evidence.

Thrombolytic therapy using rt-PA is well known for several indications like stroke, pulmonary embolism, left or right ventricular thrombi, or acute myocardial infarction.811 The only case in the literature about lysis in a patient with a left ventricular assist device has been reported from Kasirajan et al.12 They performed an intraarterial urokinase lysis after a cerebral thromboembolic complication in a patient with a Novacor LVAD. This case demonstrated for the first time the possibility of local lysis in patients with ongoing LVAD support.

Noon et al reported increased power rates and reduced pump flows following cardiac catheterization after placement of a DeBakey VAD using highly viscous radiopaque contrast agents.5 They proposed that thrombolytic therapy be performed if a thrombus inside the pump is present but never actually performed thrombolysis.5

We performed thrombolytic therapy with 100 mg of rt-PA in our patients with suspected thrombus formation or thrombus dislodgement into the MicroMed DeBakey VAD. All patients recovered from device malfunction within 3 hours after lysis and without severe bleeding complications. We were well aware of the risk of intracranial bleeding; however, emergency surgical device exchange was considered a greater hazard.

Wieselthaler et al reported their single center experience on thrombus formation in 1 of 6 patients.6 They surgically treated a patient who had a pump stop by using ligation of the outflow graft to prevent deleterious regurgitant flow. Because of mild recovery of ventricular pump function, the patient survived the waiting period to heart transplantation, which was performed 2 days later. Because most patients on mechanical support are unable to survive ligation of the outflow graft, this procedure cannot be considered an appropriate alternative to thrombolytic therapy.

In conclusion, thrombolytic therapy with rt-PA is an effective and easily applicable treatment for thrombus formation inside the MicroMed DeBakey VAD and has an acceptable risk for the patients.

References

1. Vetter HO, Kaulbach HG, Schmitz C, et al. Experience with the Novacor left ventricular assist system as a bridge to cardiac transplantation including the new wearable system. J Thorac Cardiovasc Surg. 1995; 109: 74–80.[Abstract/Free Full Text]

2. McCarthy PM, Smedira NO, Vargo RL, et al. One hundred patients with the HeartMate left ventricular assist device: evolving concepts and technology. J Thorac Cardiovasc Surg. 1998; 115: 904–912.[Abstract/Free Full Text]

3. Schmid C, Weyand M, Nabavi DG, et al. Cerebral and systemic embolization during left ventricular support with the Novacor N100 device. Ann Thorac Surg. 1998; 65: 1703–1710.[Abstract/Free Full Text]

4. Himmelreich G, Ullmann H, Riess H, et al. Pathophysiologic role of contact activation in bleeding followed by thromboembolic complications after implantation of a ventricular assist device. ASAIO J. 1995; 41: M790–M794.[Medline] [Order article via Infotrieve]

5. Noon GP, Morley DL, Irwin S, et al. Clinical experience with the MicroMed DeBakey ventricular assist device. Ann Thorac Surg. 2001; 71: S133–S138.[Abstract/Free Full Text]

6. Wieselthaler GM, Schima H, Lassnigg AM, et al. Lessons learned from the first clinical implants of the DeBakey ventricular assist device axial pump: a single center report. Ann Thorac Surg. 2001; 71: S139–S143.[Abstract/Free Full Text]

7. Scheld HH, Hammel D, Schmid C, et al. Beating herat implantation of a wearable NOVACOR left-ventricular asssist device. Thorac Cardiovasc Surg. 1996; 44: 62–66.[Medline] [Order article via Infotrieve]

8. Chalela JA, Katzan I, Liebeskind DS, et al. Safety of intra-arterial thrombolysis in the postoperative period. Stroke. 2001; 32: 1365–1369.[Abstract/Free Full Text]

9. Chartier L, Bera J, Delmonez M, et al. Free-floating thrombi in the right heart. Diagnosis, management, and prognostic indexes in 38 consecutive patients. Circulation. 1999; 99: 2779–2783.[Abstract/Free Full Text]

10. Rester BT, Warnock JL, Patel PB, et al. Lysis of a left ventricular thrombus with recombinant tissue plaminogen activator. Chest. 2001; 120: 681–683.[Abstract/Free Full Text]

11. The GUSTO Angiographic Investigators. The comparative effects of tissue plasminogen activator, streptokinase, or both on coronary patency, ventricular function and survival after myocardial infarction. N Engl J Med. 1993; 329: 1615–1622.[Abstract/Free Full Text]

12. Kasirajan V, Smedira NG, Perl J, et al. Cerebral embolism associated with left ventricular assist device support and successful therapy with intraarterial urokinase. Ann Thorac Surg. 1999; 67: 1148–1150.[Abstract/Free Full Text]





This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Rothenburger, M.
Right arrow Articles by Schmid, C.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Rothenburger, M.
Right arrow Articles by Schmid, C.