(Circulation. 2002;105:2493.)
© 2002 American Heart Association, Inc.
Clinical Investigation and Reports |
From the Cardiology Clinic (C.H., S.A., A.M.) and Radiation Oncology (D.B., M.K.) at St Josef Hospital, Ruhr University, Bochum, Germany, and the Department of Pathology (P.G., K.M.M.) at Bergmannsheil-Kliniken, Ruhr-University, Bochum, Germany.
Correspondence to Dr Andreas Mügge, Cardiology Clinic, St Josef Hospital, Ruhr University, Gudrunstrasse 56, 44791 Bochum, Germany. E-mail andreas.muegge{at}ruhr-uni-bochum.de
| Abstract |
|---|
|
|
|---|
-emitters) are available in clinical routine. It was the purpose of this study to compare the radiation doses at the level of the adventitia in diseased and stented human coronary arteries for 192Ir and 90Sr/Y emitters in routine use. In contrast to previously published work, we performed dosimetry instead of calculating depth-dose distribution by use of the Monte Carlo system.
Methods and Results Postmortem calcified human coronary artery segments were stented and placed in an organ bath. Commercially available
-emitters (192Ir; Cordis Checkmate) and ß-emitters (90Sr/Y; Novoste Beta-Cath) were used. Relative dose distributions along the adventitia were measured by a specially designed scintillation detector system. Whereas dose perturbations caused by stents and calcified plaque were negligible for the 192Ir source, radiation from the beta source was significantly impaired (as much as 40%) at the level of the adventitia (3.0-mm vessel diameter). Dose perturbation was clearly dependent on the extent and severity of calcification, less affected by stent material.
Conclusions Dose perturbation caused by calcified plaque and metallic stents is significant for ß-sources. This dosimetric difference between ß- and
-emitters in diseased coronary arteries should be considered when calculating doses in intravascular brachytherapy.
Key Words: angioplasty restenosis stenosis stents coronary disease
| Introduction |
|---|
|
|
|---|
Various photon- or electron-emitting radioactive isotopes have been used or suggested for use in brachytherapy. The most frequently used
-emitters are irdium192 isotopes (192Ir) and the most frequently used ß-emitters are strontium/yttrium90 sources (90Sr/Y). Many theoretical questions have been raised regarding the effectiveness of both types of emitters in delivering the required dose of 14 to 20 Gy along the vessel wall (in particular the adventitia), such as the importance of accurate source centering10 and the dose perturbation caused by calcified plaque and/or metallic stents.11 In animal studies, 14 Gy radiation from a noncentred 90Sr/Y source was successful in preventing neointima formation only when radiation was performed before rather than after stenting. The authors strongly suggest that this effect is related to perturbation of the dose field by the metallic stent.12
Despite such theoretical considerations, the clinical studies suggest that both types,
-emitters as well as ß-emitters, are effective in reducing restenosis. However, these studies have not directly compared the efficiency of
- and ß-emitters. Thus, the question remains open whether one emitter offers more advantages than the other under specific clinical conditions, such as the presence of stent and/or calcification,13 long and complex lesions, or large vessel size. Accordingly, recent studies for ß-emitters point out that only a dose of at least 18 Gy (tissue depth 1 mm) appears to be effective for reducing restenosis in nonstented arteries and that the presence of stents may inhibit this beneficial effect.14
This study is intended to compare dose homogeneity between routinely used 192Ir and 90Sr/Y emitters in isolated human coronary arteries in the presence of calcified plaques and metallic stents. Since it is well accepted from previous work that metallic stents do perturb the dose of ß-emitters and low-energy photon emitters,11,15,16 we focus on the relative dose distribution along the adventitial surface under conditions similar to clinical circumstances. In contrast to previous work on a similar topic,11 direct measurements were performed instead of using a calculated Monte Carlo system.
| Methods |
|---|
|
|
|---|
|
For dosimetry, an Optidos scintillation system was used (PTW). The scintillator tip (1.0x1.0 mm) was placed directly on the surface (adventitia) of the vessel wall so that it could be moved stepwise (1.0-mm steps) along the entire longitudinal axis of the coronary segment by an automated micrometer device. In a typical experiment, the scintillator tip was located
1.2 to 1.6 mm above the 5F and 3.7F delivery catheters, respectively; once a distance was chosen, this distance was kept constant during a set of experiments (dosimetry for the
- and ß-sources). The Optidos System is designed for quality assurance measurements in vascular brachytherapy and fulfills the requirements recommended in the report of the American Association of Physicists in Medicine (AAPM) Task Group No. 60. This system determines, in catheter systems, the source strength in terms of absorbed dose rate to water at a distance of 2 mm from the catheter axis, as per AAPM report, or at different distances, as per suppliers instructions. In addition, dose distribution, depth-dose curve, and the longitudinal and rotational homogeneity of line sources can be assessed. During radiation, the scintillation light is guided by fiberoptic cable and converted into current by a microphotomultiplier. The dosimeter signal, being directly proportional to the absorbed dose, is integrated by the Optidos system for variable time intervals. Sensitivity is as low as 400 µGy/s, with <1% deviation. Usually, the time of radiation exposure was adjusted to 30 seconds at each point of measurement.
The US national standard laboratory NIST has calibrated the Optidos system for sources that emit electrons. The process of calibrating the detector for
-emitters is ongoing, and calibration factors are not available yet. Thus, an absolute dose rate was only calculated for the ß-source but not for the
-source. At the level of the adventitia in noncalcified segments, an absolute dose rate of on average 782±96 mGy/30 s was measured for the Novoste Beta Cath System, corresponding to a dose rate of 0.095 Gy/s at a depth of 2 mm from the center of the source train.
The ionizing radiation delivering systems used were the Novoste Beta-Cath System (40-mm length) and the Cordis Checkmate System (39-mm length). Source activities were 9.62 to 11.87 GBq and 1.92 GBq for the 192Ir and 90Sr/Y sources, respectively. The delivery catheters for Beta-Cath and Checkmate systems had an outer diameter of 5F and 3.7F, respectively.
After completing measurement, a small needle was used to pinpoint the beginning and end of the evaluated area along the vessel wall. The arterial segments were placed in 5% formalin. Radiography (with a Faxitron 8050 table-model x-ray unit from Rode & Schwarz) of the segments documented stent expansion and localized calcified plaque/deposits, in particular those related to the area analyzed for radiation perturbance. After radiography, each segment was dehydrated (ethanol was increased from 50% to 96% vol/vol), defatted, and embedded in methamethylacrylate plastic (Technovit 9100 from Hereus Kulzer GmbH). Parallel cross sections (1 mm) of the stented coronary artery segments were cut with a rotary cutter microtome (Accutom-5 from Struers). Cross sections were ground and polished (Phoenix 4000 from Jean Wirtz) to a thickness <8 µm and stained with a modified Elasticavan Gieson for light microscopic analysis, which provides clear delineation of calcified plaque. Artery segments were divided into three parts (
5 to 6 mm length each), and the extent of calcification was graded semiquantitatively into three classes: no calcification, minor calcification (thickness
0.4 mm), and moderate/severe calcification (thickness >0.4 mm). Care was taken to consider only such calcifications as were located between the scintillation detector and the radiation source. Morphometry revealed the following diameters: lumen, 3.1±0.1 mm; intima, 0.45±0.29 mm; and media, 0.13±0.07 mm. The thickness of calcified plaques/deposits was in the range 0.14 to 1.02 mm (median, 0.41 mm).
Statistics
Individual data from representative experiments are the mean of three individual measurements; otherwise, data are given as mean±SD. (Absolute dose rate is given in mGy/30 s.) If indicated, relative dose is normalized to 100% (in the absence of stented vessel wall). Dose perturbations are given as a percentage of dose reduction in the absence and presence of the stented coronary segment. Mean values were compared by unpaired Students t test; a value of P<0.05 was considered significant.
| Results |
|---|
|
|
|---|
-source, respectively. As shown previously,11 the decrease of radial dose is high for ß-sources and lower for
-sources.
Figure 2 shows a representative set of measurements in a severely calcified and stented coronary artery segment. Calcifications were documented by radiography and histology (Figure 2, A and B). Neither stent material (Palmaz-Schatz, 18 mm length) nor calcified deposits significantly perturbed the
-radiation; almost identical doses were measured in the absence and presence of the artery segment along the adventitial surface (Figure 2C). By contrast, ß-radiation showed significant dose inhomogeneity along the adventitial surface; dose perturbation up to 40±4% (n=3 measurements; as compared with the condition in the absence of the vessel) was noted in this particular coronary segment (Figure 2D).
|
The Tablesummarizes the measurements for all vessels. As expected, the
-radiation was not significantly affected by stents and calcifications irrespective of the thickness of calcified plaque/deposits. The radiation dose at the level of the adventitia from the ß-emitter was reduced by between 11±8% (no calcification, only stented vessel wall) and 32±18% (stent with moderate/severe calcification).
|
| Discussion |
|---|
|
|
|---|
-radiation and 90Sr/Y for ß-radiation.
On the basis of physical principles and theoretical calculation models (eg, the EGS4 Monte Carlo system),
-emitters may combine some advantages, in particular better penetration and no need for source centering in coronary arteries. In contrast, ß-sources offer advantages in terms of radiation protection and dose rate (which determines time of radiation exposure). Since most patients treated with brachytherapy have been stented in severely diseased coronary arteries, we were interested in finding out how stents, in combination with various degrees of calcified plaque, affect dosimetry at the level of the adventitia, that is, the most likely target of brachytherapy, for commercially available
- and ß-sources. This topic has been previously addressed by several investigators; however, most of them used Monte Carlo simulation techniques10,11,16,18,19 or measured the effect of metallic stents only.15
Our main findings are (1) dose perturbations caused by stent and calcified plaque were negligible for the 192Ir source; (2) stent material in combination with calcifications significantly impaired radial dose distribution for the 90Sr/Y source. This dose reduction was on average 11% in noncalcified stented vessels and mounted up to 40% in severely calcified, stented arteries. It appears that the degree of calcification has a greater impact on radial dose distribution for ß-particles than metallic stents themselves. Calcified (stenotic) plaques could influence dose homogeneity at the level of the adventitia, mainly by two mechanisms: First, by shielding effects of the plaque itself11,16; and second, de-centering of the delivery catheter within the vessel lumen. Both systems used in the present study were not centered. Our experimental setup plus other limitations of the measurements (limited control of distance between radiation source and vessel surface/tip of the scintillation system, limited precision of the dosimetry) do not allow us to differentiate between the two causes of dose homogeneity, shielding, and de-centering effects.
Our measurements are in accordance with previous calculated results, that is, that dose perturbations caused by the presence of metallic stents were found to be far more significant for the 90Sr/Y source than those for the 192Ir source.11 Amols and coworkers15 reported an average dose reduction of ß-emitters between 4% and 14% in the presence of various stents, for example,
11% at 0.125 mm radial distance from steel stents. This range of dose reduction is comparable to that observed in the present study in stented but noncalcified human artery segments. More importantly, we observed a major dose perturbation for ß-particles in severely calcified human arteries (dose reduction on average, 31.8%). Thus, we believe that the dosimetric differences between ß- and
-sources in diseased human coronary arteries should be considered for dose prescription in intravascular brachytherapy, for example, by increasing the dose for ß-sources in calcified stented arteries. Whether the considerations are clinically important is, however, still questionable. In at least one pilot study at a 6-month follow-up after ß-radiation therapy, 20 delivered dose as well as tissue composition have been shown to fundamentally influence volumetric outcome. Furthermore, it has been observed in a recent dose-finding study that ß-radiation therapy prevents restenosis in a dose-dependent manner and that only a dose as high as 18 Gy both reduced restenosis and induced luminal enlargement.14 These observations underline the importance of precise dose prescription.
In controlled and open clinical studies, both ß- and
-emitters were effective to prevent restenosis in patients with in-stent stenosis.21 It is interesting to note that the dose prescribed in these studies that used ß-emitters was up to 30% to 40% higher (
18 to 24 Gy at a depth of 2 mm) as compared with those that used
-emitters (
14 to 15 Gy).21 It is speculated that this difference in dose prescription between ß- and
-sources may account for the similarity in the clinical outcome to prevent or treat in-stent restenosis.
| Acknowledgments |
|---|
Received January 11, 2002; revision received March 22, 2002; accepted March 22, 2002.
| References |
|---|
|
|
|---|
2.
Serruys PW, de Jaegere P, Kiemeneij F, et al. A comparison of balloon-expandable-stent implantation with balloon angioplasty in patients with coronary artery disease. N Engl J Med. 1994; 331: 489495.
3.
Fishman DL, Leon MB, Baim DS, et al. A randomised comparison of coronary-stent placement and balloon angioplasty in the treatment of coronary artery disease. N Engl J Med. 1994; 331: 496501.
4.
Condado JA, Waksman R, Gurdiel O, et al. Long-term angiographic and clinical outcome after percutaneous transluminal coronary angioplasty and intracoronary radiation therapy in humans. Circulation. 1997; 96: 727732.
5.
Teirstein PS, Massullo V, Jani S, et al. A double-blinded randomised trial of catheter-based radiotherapy to inhibit restenosis following coronary stenting. N Engl J Med. 1997; 336: 16971703.
6.
King SB, Williams DO, Chougule P, et al. Endovascular ß-radiation to reduce restenosis after coronary balloon angioplasty results of the Beta Energy Restenosis Trial (BERT). Circulation. 1998; 97: 20252030.
7.
Raizner AE, Oesterle SN, Waksman R, et al. Inhibition of restenosis with ß-emitting radiotherapy. report of the Proliferation Reduction with Vascular Energy Trial (PREVENT). Circulation. 2000; 102: 951958.
8.
Waksman R, Ghargava B, White L, et al. Intracoronary ß-radiation therapy inhibits recurrence of in-stent restenosis. Circulation. 2000; 101: 18951898.
9.
Waksman R, White RL, Chan RC, et al. Intracoronary
-radiation therapy after angioplasty inhibits recurrence in patients with in-stent restenosis. Circulation. 2000; 101: 21652171.
10. Amols HI, Zaider M, Weinberger J, et al. Dosimetric considerations for catheter-based beta and gamma emitters in the therapy of neointimal hyperplasia in human coronary arteries. Int J Radiat Oncol Biol Phys. 1996; 36: 913921.[CrossRef][Medline] [Order article via Infotrieve]
11. Li XA, Wang R, Yu C, et al. Beta versus gamma for catheter-based intravascular brachytherapy: dosimetric perspectives in the presence of metallic stents and calcified plaques. Int J Radiat Oncol Biol Phys. 2000; 46: 10431049.[Medline] [Order article via Infotrieve]
12. Waksman R, Robinson KA, Crocker IR, et al. Intracoronary beta radiation before versus after stent implantation for inhibition of neointima formation in the porcine model. Circulation. 1996; 94 (suppl I): I-619. Abstract.
13. Weinberger J. Irradiation and stenting. Semin Intervent Cardiol. 1997; 2: 103108.[Medline] [Order article via Infotrieve]
14.
Verin V, Popowski Y, DeBruyne B, et al. Endoluminal beta-radiation therapy for the prevention of coronary stenosis after balloon angioplasty. N Engl J Med. 2001; 344: 243249.
15.
Amols HI, Trichter F, Weinberger J. Intracoronary radiation for prevention of restenosis: dose perturbations caused by stents. Circulation. 1998; 98: 20242029.
16. Nath R, Yue N. Shielding effects of metallic encapsulations and radiographic contrast agents for catheter-based intravascular brachytherapy. Cardiovasc Radiat Med. 2001; 2: 93103.[CrossRef][Medline] [Order article via Infotrieve]
17. Wilcox JN Waksman R. King SB, et al. The role of the adventitia in the arterial response to angioplasty: the effect of intravascular radiation. Int J Radiat Oncol Biol Phys. 1996; 36: 789796.[CrossRef][Medline] [Order article via Infotrieve]
18. Rahdert DA, Sweet WL, Tio FO, et al. Measurement of density and calcium in human atherosclerotic plaque and implications for arterial brachytherapy. Cardiovasc Radiat Med. 1999; 1: 358367.[CrossRef][Medline] [Order article via Infotrieve]
19. Ye SJ, Li XA, Zimmer JR, et al. Dosimetric perturbations of linear array of beta-emitter seeds and metallic stent in intravascular brachytherapy. Med Phys. 2000; 27: 374380.[CrossRef][Medline] [Order article via Infotrieve]
20.
Sabate M, Marijnissen JPA, Carlier SG, et al. Residual plaque burden, delivered dose, and tissue composition predict 6-month outcome after balloon angioplasty and ß-radiation therapy. Circulation. 2000; 101: 24722477.
21. Silber S. Intracoronary radiation therapy in controlled and open clinical studies with afterloading-systems and "hot balloons": analysis of 6.692 patients. Herz. 2002; 27: 3055.[CrossRef][Medline] [Order article via Infotrieve]
This article has been cited by other articles:
![]() |
P. G. Guerra, M. Talajic, B. Thibault, M. Dubuc, D. Roy, L. Macle, T. K. Leung, A. Arsenault, F. Harel, J. Gregoire, et al. {beta}-Radiation for the Creation of Linear Lesions in the Canine Atrium Circulation, August 24, 2004; 110(8): 911 - 914. [Abstract] [Full Text] [PDF] |
||||
![]() |
P H Grewe, T Deneke, C Hanefeld, and K-M Muller Human coronary morphology after {beta} radiation brachytherapy of in-stent restenosis Heart, June 1, 2004; 90(6): e32 - e32. [Abstract] [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2002 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |