(Circulation. 2002;105:157.)
© 2002 American Heart Association, Inc.
Clinical Investigation and Reports |
From the Cardiology Department, Kings College Hospital, London, UK.
Correspondence to Dr J.E. Hancock, Cardiology Department, Wythenshawe Hospital, Southmoor Rd, Manchester M23 9LT, UK. E-mail janehancock{at}hotmail.com
| Abstract |
|---|
|
|
|---|
Methods and Results One hundred thirteen patients with AMI had peak-to-peak cyclic IBS measured in the myocardial territory supplied by their IRA as well as a remote territory with normal function from the parasternal long- or short-axis view. This analysis took 5 to 10 minutes. Wall motion score index was assessed, and coronary angiography, to determine patency of the IRA, was performed in all patients. Cyclic IBS in the IRA territory was much lower in segments supplied by an occluded IRA (3.3 versus 4.6 dB, P<0.00001). Using a difference in cyclic IBS between infarcted and normal segments of 15% (or 1.5 dB) as a cutoff, the sensitivity, specificity, positive and negative predictive values to determine an occluded IRA were 92%, 75%, 81%, and 89%, respectively.
Conclusions The difference in cyclic IBS between IRA and remote normal segments, which can be analyzed rapidly, can be used to predict patency of the IRA in patients with AMI. This provides a noninvasive method to determine those patients who may require urgent invasive investigation.
Key Words: reperfusion thrombolysis myocardial infarction ultrasonics
| Introduction |
|---|
|
|
|---|
The ideal technique to evaluate reperfusion at a microvascular level would be one that is easy to perform, can be done at the bedside, can be analyzed quickly, and is sufficiently accurate to allow stratification of patients with AMI into those who have an occluded IRA (or possibly a patent IRA but no reflow) and who require cardiac catheter facilities urgently and those who have a patent IRA (or occluded IRA but collaterals) and will require cardiac catheterization at an interval, if at all. To date, this has not been achieved.
Of the echocardiographic techniques presently available for this assessment, the most promising are intravenous MCE and ultrasonic tissue characterization with integrated backscatter (IBS). So far, intravenous MCE has not been thoroughly quantified in humans, and several technical limitations remain. Additionally, a few patients may have contraindications to ultrasound contrast agents.
IBS has been more thoroughly investigated. Several studies exist in the setting of AMI35 that have shown that the cardiac cycledependent variation in this parameter is blunted in regions of myocardium supplied by the IRA compared with adjacent normal myocardial segments. There is recovery of cyclic IBS in the IRA territories supplied by a patent IRA but not in those with an occluded IRA at 7 days in the absence of any change in wall motion, suggesting that this may predict viability.3 Takiuchi et al6 confirmed this, demonstrating that in patients who had an anterior myocardial infarction treated by coronary angioplasty, an increase in cyclic IBS in the IRA territory 3 days after the procedure was not associated with any change in wall motion score, but these patients had a significant improvement in their wall motion score at 21 days. This suggests that cyclic IBS has the potential to detect viable myocardium and predict functional recovery.2,6,7 Animal and in vitro studies support this.811
However, most studies to date have involved small patient numbers. There are also some limitations to the use of cyclic IBS acutely. Although the data are easy to collect, analysis is more difficult. Myocardial fiber orientation and depth of field significantly affect the magnitude of the IBS.12,13 In the short-axis view, the anterior septum has the highest levels, where the beam is perpendicular to the myocardial fibers. The lateral wall and inferior septum have the lowest levels, where the fiber orientation is parallel to the insonating ultrasound beam. The posterior wall has IBS levels that are lower than the anterior septum because of the increased depth. Thus, the effect of anisotropy must be corrected for. Furthermore, the phase of the peak of cyclic IBS is affected both by orientation of fibers and by damaged myocardium. Therefore, many authors believe that it is necessary to make a correction for phase of the cyclic IBS.14,15 Even in subjects with normal myocardium, there are some variations. Studies have shown wide variation in both the magnitude and phase of cyclic IBS between myocardial segments in healthy subjects.16 So, although measurement of cyclic IBS is useful, the above constraints have meant that data have had to be analyzed and manipulated offline and limited its clinical application in the acute setting.
In the present study, we have measured differences in the magnitude of the cyclic variation in IBS without additional manipulation in patients in the setting of AMI to reduce the analysis time to 5 to 10 minutes. These data are compared with coronary angiographic findings to elucidate whether we could differentiate between patients with patent and occluded IRAs using this method.
| Methods |
|---|
|
|
|---|
2 mm in
2 contiguous chest leads or
1 mm in
2 contiguous limb leads. All patients were treated with thrombolysis or primary angioplasty (18 patients). Patients who were thrombolyzed subsequently had coronary angiography so that the coronary anatomy and, in particular, information on the patency of the infarct-related artery (IRA) was known for all patients. The study had local ethical approval.
|
Data Acquisition
Cyclic IBS was measured in the territory of the IRA as well as a remote territory with normal function. Additionally, images were acquired for analysis of wall motion. This was done before coronary angiography in all cases, but as near as possible to 90 minutes after thrombolysis in the patients who were thrombolyzed. All 4 standard, echocardiographic views (parasternal long-axis, parasternal short-axis, apical 4-chamber, and apical 2-chamber) were imaged using an Agilent Technologies Sonos 5500 or 2500 ultrasound system. Acoustic densitometry (AD) with IBS software was available on both systems. With the ultrasound machine operating in the AD acquisition mode, real-time IBS images were acquired for 60 consecutive frames at a rate of 30 frames per second (2 to 3 cardiac cycles), displayed, stored to magneto-optical disk, and subsequently analyzed. System settings were optimized by adjusting the depth, total gain, and time-gain compensators so that all segments of myocardium were clearly visualized. Imaging was performed using either a 2.5-MHz or an S4 (2- to 4-MHz) transducer.
Data Analysis
IBS data were analyzed offline for each myocardial region for each subject using the AD analysis package available on the SONOS 5500 imaging system. An elliptical region of interest of 31x31 mm or 41x41 mm was placed in the midmyocardial region in each segment analyzed, positioned to avoid endocardium and epicardium, and tracked manually on a frame-by-frame basis for the entire loop of 60 frames. No attempt was made to differentiate between subendocardial and other myocardial regions within the segment.11,17 For each segment, the analysis package provides a mean IBS value in decibels for the entire loop as well as a mean value for each frame. This allows assessment of the magnitude of the cyclic variation, which we have taken as the difference between the average peak and average nadir values of IBS.14 To avoid the effects of anisotropy as much as possible and to reduce phase correction, only the parasternal views were used for analysis, and region of interests were placed in the anterior septum and posterior, or inferior, wall as perpendicular as possible to the ultrasound beam and directly opposite each other whenever possible.
Wall motion analysis was also performed in each subject for all visible segments in each echocardiographic view. This was done using the standard 16-segment model, as defined by the American Society of Echocardiography,18 using a visual scoring system where 1 indicates normal or hyperdynamic; 2, hypokinetic; 3, akinetic; and 4, dyskinetic. This analysis was performed by 2 independent observers without knowledge of the coronary angiographic data.
Statistics
Difference in mean values between infarcted and normal myocardial segments in each imaging mode were analyzed using a Students t test. Interobserver reliability was compared using ANOVA. ROC curves were used to determine the optimal cutoff value of difference in cyclic IBS between IRA and normal segments both in decibels and as a percentage for the determination of patency of the IRA.19 The study was powered to detect a 1-dB difference in cyclic IBS at the 0.05 level. Results were considered significant when the P<0.05.
| Results |
|---|
|
|
|---|
|
The percentage difference in cyclic IBS between infarcted (IRA) and normal myocardial segments was also significantly different depending on the patency of the IRA, being 9±12% in patients with a patent IRA and 27±10% in patients with an occluded IRA (P<0.00001) (Figure 2).
|
The sensitivity, specificity, positive predictive value, and negative predictive value of various differences in cyclic IBS between infarcted and normal myocardial segments is shown in Table 2 for percentage differences and in Table 3 for absolute differences in decibels. The magnitude of cyclic variation in IBS in Figure 1 and Table 3 has been presented in the familiar units of decibels. It must be noted, however, that the ROC analysis results were obtained by first transforming the measurement data from the nonlinear, logarithmic (decibel) domain to the linear domain using the equation Vd/Vs=10PPI/20, where Vd/Vs is the ratio of the average peak voltage at end diastole to the nadir of the voltage at end systole and PPI is the magnitude of the cyclic variation in decibels. The results were then transformed back to the log (decibel) domain by using the inverse transformation PPI=20xlog(Vd/Vs). The ROC curves suggest that the optimal cutoff point to predict occlusion of the IRA is 15% or 1.5 dB (Figure 3).
|
|
|
Wall motion score index (WMSI) for patients with an occluded IRA was 1.84 and 1.67 for patients with a patent IRA (P<0.001). However, the ROC curve for this data did not show a clear cutoff (Figure 3), so this could not be used accurately to predict an occluded IRA.
Interobserver reliability was 0.86 for cyclic IBS measurements in myocardial segments supplied by the IRA and 0.8 for remote segments with normal function. Intraobserver reliability was 0.95 for cyclic IBS measurements in myocardial segments supplied by the IRA and 0.93 for remote segments with normal function.
| Discussion |
|---|
|
|
|---|
In the present study, we have demonstrated that average peak-to-average nadir values of cyclic IBS can be used to determine the patency of the artery supplying an infarcted myocardial segment. Acquisition of the data and measurement of these values, without additional manipulation (eg, phase correction) can be achieved within 10 minutes and often within 5 minutes. A difference in absolute cyclic IBS between the IRA segment and a remote segment with normal function of
1.5 dB can predict an occluded IRA with a sensitivity of 88%, a specificity of 77%, a positive predictive value of 81%, and a negative predictive value of 85%.
If these are converted to percentage values, a difference between an IRA segment and a remote segment of
15% can predict an occluded IRA with a sensitivity of 92%, a specificity of 75%, a positive predictive value of 81%, and a negative predictive value of 89%. This is comparable with other noninvasive techniques and at least as good as IBS with phase correction reported in previous studies.3,5 In these studies, analysis took a minimum 20 minutes and often 1 hour.3,5
WMSI was found to be significantly different between the 2 groups, being higher in patients with an occluded IRA. This may be because both hyperdynamic and normal myocardial segments were given a score of 1. However, it is clear from the ROC curve that it would not be possible to use WMSI to determine an occluded IRA. This is in keeping with other studies.3,8,17,21
Because it is likely that differences in cyclic IBS actually represent viability and regional contractile performance, in the present study some patients with a patent IRA would have been determined to have an occluded IRA by this method (false-positive), presumably because of the no reflow phenomenon. There were 6 patients who would have fallen into this category. Two of these patients died, and 2 more had no recovery of left ventricular function on long-term follow-up. The remaining 3 did not have follow-up studies. Similarly, some patients with an occluded IRA will have been determined to have a patent IRA (false-negative). This occurred with 5 of the 59 patients in the present study. One might speculate that this was because of collateral flow. However, this was not seen in these 5 patients. Indeed, only 2 patients had retrograde filling of the IRA at angiography. An occluded IRA was predicted by cyclic IBS in these 2 patients. An alternative explanation is that in these patients, the remote normal territory was not, in fact, normal, but supplied by an occluded or significantly stenosed artery. Of the 5 false-negative patients, 2 had single-vessel disease. The remaining 3 had 2-vessel disease, but the diseased vessel supplied the remote normal territory analyzed in only 2 of these. A separate analysis of the subgroup of patients with multivessel disease has not been performed. This is a potential limitation of the study.
Another potential limitation is that no attempt has been made to correct for phase, as many other authors have done. However, as stated previously, the main reason for recording only the average peak to average nadir value of the cyclic IBS was to reduce analysis time to make the technique more clinically useful in the acute situation. Most centers would be able to have an echo machine with IBS software and be in a position to record and analyze these data accurately after some training. However, not all would be in a position to have the necessary computer program or time to perform the more complex data manipulation required for phase correction. Our technique for analysis can be used in a clinical setting to guide management of patients, particularly to determine which patients have failed thrombolysis and require more invasive investigation.
Furthermore, in previous studies with phase correction, segments which were found to have a normalized time delay of >1.2 were considered to be asynchronous, so the peak-to-peak IBS level was multiplied by -1.6 This has the effect of making the cyclic IBS value from the IRA segment even smaller or negative, additionally increasing the difference in cyclic IBS between this segment and the remote segment with normal function. Some authors have reported similar peak-to-peak cyclic IBS measurements from abnormal and normal segments but found that these were completely out of phase with each other.6,10,22 This was not found in our study. Cyclic IBS from normal and abnormal segments was similar when the IRA was patent, so that there was less myocardial damage because reperfusion had been successful. The exception to this was in the 5 false-negative cases discussed above. In those cases, there was no significant difference in phase between the peak value in the IRA territory and remote segments.
Another potential limitation is the fact that only the parasternal views have been used to reduce the effects of phase difference and anisotropy as much as possible. In fact, many other authors have done this, few studying apical views.3,5,6 In all cases in the present study, the segments analyzed all included a segment with abnormal wall motion and a remote segment with normal function, which would have been at the same angle to the insonating beam as the abnormal (IRA) segment. This should reduce the effect of anisotropy as much as possible while still analyzing representative segments.
Another limitation of this study is that we have made no attempt to incorporate a comparison with subjects with normal myocardium. This might have affected our results. However, given the known differences in cyclic IBS in normal myocardium,16 it is likely that these results, in the setting of AMI, may have reached even greater significance.
| Conclusions |
|---|
|
|
|---|
| Acknowledgments |
|---|
|
|
|---|
Received October 8, 2001; revision received November 5, 2001; accepted November 5, 2001.
| References |
|---|
|
|
|---|
2.
Ito H, Tomooka T, Sakai N, et al. Lack of myocardial perfusion immediately after successful thrombolysis: a predictor of poor recovery of left ventricular function in anterior myocardial infarction. Circulation. 1993; 85: 16991705.
3.
Milunski MR, Mohr GA, Perez JE, et al. Ultrasonic tissue characterisation with integrated backscatter: acute myocardial, reperfusion and stunned myocardium in patients. Circulation. 1989; 80: 491503.
4. Saeian K, Rhyne TL, Sagar KB. Ultrasonic tissue characterisation for diagnosis of acute myocardial infarction in the coronary care unit. Am J Cardiol. 1994; 74: 12111215.[CrossRef][Medline] [Order article via Infotrieve]
5. Lin LC, Wu CC, Ho YL, et al. Ultrasonic tissue characterisation for coronary care unit patients with acute myocardial infarction. Ultrasound Med Biol. 1998; 24: 187196.[CrossRef][Medline] [Order article via Infotrieve]
6.
Takiuchi S, Ito H, Iwakura K, et al. Ultrasonic tissue characterisation predicts myocardial viability in early stage of reperfused acute myocardial infarction. Circulation. 1998; 97: 350362.
7. Lin LC, Wu CC, Ho YL, et al. Ultrasonic tissue characterisation in predicting residual ischemia and myocardial viability for patients with acute myocardial infarction. Ultrasound Med Biol. 1998; 24: 11071120.[CrossRef][Medline] [Order article via Infotrieve]
8. Barzilai B, Madaras EI, Sobel BE, et al. Effects of myocardial contraction on ultrasonic backscatter before and after ischemia. Am J Physiol. 1984; 247: H478H483.
9.
Glueck RM, Mottley JG, Miller JG, et al. Effects of coronary artery occlusion and reperfusion on cardiac cycle-dependent variation of myocardial ultrasonic backscatter. Circ Res. 1985; 56: 683689.
10.
Wickline SA, Thomas LJIII, Miller JG, et al. Sensitive detection of the effects of reperfusion on myocardium by ultrasonic tissue characterisation with integrated backscatter. Circulation. 1986; 74: 389400.
11.
Sagar KB, Rhyne TL, Warltier DC, et al. Intramyocardial variability in integrated backscatter: effects of coronary occlusion and reperfusion. Circulation. 1987; 75: 436442.
12. Finch-Johnston AE, Gussak HM, Mobley J, et al. Cyclic variation of integrated backscatter: dependence of time delay on the echocardiographic view used and the myocardial segment analyzed. J Am Soc Echo. 2000; 13: 917.[Medline] [Order article via Infotrieve]
13. Holland MR, Wilkenshoff UM, Finch Johnston AE, et al. Effects of myocardial fiber orientation in echocardiography: quantitative measurements and computer simulation of the regional dependence of backscattered ultrasound in the parasternal short axis view. J Am Soc Echo. 1998; 11: 929937.[CrossRef][Medline] [Order article via Infotrieve]
14. Finch-Johnston AE, Gussak HM, Mobley J, et al. Dependence of "apparent" magnitude on the time delay of cyclic variation in integrated backscatter. Ultrasound Med Biol. 1999; 25: 759762.[CrossRef][Medline] [Order article via Infotrieve]
15.
Fitzgerald PJ, McDaniel MD, Rolett EL, et al. Two dimensional ultrasonic tissue characterisation: backscatter power, endocardial wall motion and their phase relationship for normal, ischemic and infarcted myocardium. Circulation. 1987; 76: 850859.
16. Lange A, Moran CM, Palka P, et al. The variation of integrated backscatter in human hearts using different ultrasonic transthoracic views. J Am Soc Echo. 1995; 8: 830838.[CrossRef][Medline] [Order article via Infotrieve]
17.
Wickline SA, Thomas LJ III, Miller JG, et al. The dependence of myocardial ultrasonic integrated backscatter on contractile performance. Circulation. 1985; 72: 183192.
18. Schiller NB, Shah PM, Crawford M, et al. Recommendations for quantitation of the left ventricle by two dimensional echocardiography. J Am Soc Echocardiogr. 1989; 2: 358368.[Medline] [Order article via Infotrieve]
19. Wagner RF, Wear KA, Perez JE, et al. Quantitative assessment of myocardial ultrasound tissue characterisation through receiver operating characteristic analysis of Bayesian classifiers. J Am Coll Cardiol. 1995; 25: 17061711.[Abstract]
20.
Lythall DA, Gibson DG, Kushwaha SS, et al. Changes in myocardial echo amplitude during reversible ischaemia in humans. Br Heart J. 1992; 67: 368376.
21. Vered Z, Mohr GA, Barzilai B, et al. Ultrasound integrated backscatter tissue characterisation of remote myocardial infarction in human subjects. J Am Coll Cardiol. 1989; 89: 8491.
22.
Vitale DF, Bonow RO, Gerundo G, et al. Alterations in ultrasonic backscatter during exercise-induced myocardial ischemia in humans. Circulation. 1995; 92: 14521457.
This article has been cited by other articles:
![]() |
J.E Hancock, J.C Cooke, and M.J Monaghan Effect of harmonic imaging on the measurement of ultrasonic integrated backscatter and its interpretation in patients following myocardial infarction Eur J Echocardiogr, June 1, 2004; 5(3): 189 - 195. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Iwakura, H. Ito, S. Kawano, A. Okamura, K. Asano, T. Kuroda, K. Tanaka, T. Masuyama, M. Hori, and K. Fujii Detection of TIMI-3 Flow Before Mechanical Reperfusion With Ultrasonic Tissue Characterization in Patients With Anterior Wall Acute Myocardial Infarction Circulation, July 1, 2003; 107(25): 3159 - 3164. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2002 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |