Sonodynamic Therapy Decreased Neointimal Hyperplasia After Stenting in the Rabbit Iliac Artery
Koh Arakawa, MD;
Kousukue Hagisawa, MD;
Hiroyuki Kusano, MD;
Satoru Yoneyama, MD;
Akira Kurita, MD;
Tsunenori Arai, PhD;
Makoto Kikuchi, PhD;
Isao Sakata, PhD;
Shin-ichirou Umenura, PhD;
Fumitaka Ohsuzu, MD
From Internal Medicine I (K.A., K.H., H.K., S.Y., F.O.), Medical Engineering (T.A., M.K.), and Biomedical Engineering (A.K.), National Defense Medical College, Saitama, Japan; Photochemical Co, Ltd (I.S.), Okayama, Japan; and Hitachi Research Center (S.U.), Tokyo, Japan.
Correspondence to Koh Arakawa, MD, Department Medicine I, National Defense Medical College, 3-2, Namiki, Tokorozawa, Saitama 359-8513, Japan. E-mail karakawa{at}me.ndmc.ac.jp
 |
Abstract
|
|---|
Background In-stent restenosis remains a pivotal problem
after coronary and peripheral stenting. Sonodynamic therapy
inhibits tumor growth by means of cytotoxicity after the activation
of sonochemical sensitizers by ultrasound. PAD-S31 is known
to be a water-soluble, chlorin-derivative sonochemical sensitizer.
We assessed the efficacy of sonodynamic therapy using this sensitizer
on neointimal hyperplasia in a rabbit stent model.
Methods and Results Stents were implanted in the iliac arteries of 16 rabbits. A total of 32 stented arteries were randomized to sonodynamic therapy, control, ultrasound exposure, and PAD-S31 groups. One hour after the intravenous administration of PAD-S31 (25 mg/kg body weight), ultrasound energy (1 MHz, 0.3 W/cm2) was delivered transdermally to the sonodynamic therapy group. At 28 days, all stent sites were analyzed morphometrically. The size of the intimal cross-sectional area was smaller in the sonodynamic therapy group than in the control, ultrasound, and PAD-S31 groups (0.31±0.07 versus 1.38±0.47, 1.66±0.71, and 1.61±0.42 mm2, respectively; P<0.05). The ratio of the intimal and medial cross-sectional area was smaller in the sonodynamic therapy group than in the control, ultrasound, and PAD-S31 groups (0.71±0.22 versus 2.53±1.39, 2.48±0.60, and 3.45±1.42 mm2; P<0.05).
Conclusions Sonodynamic therapy with PAD-S31 is considered to be a feasible treatment modality for noninvasively inhibiting neointimal hyperplasia in a rabbit iliac stent model.
Key Words: ultrasonics stents restenosis sonodynamic therapy
 |
Introduction
|
|---|
Use of stent implantation has improved the restenosis rate after
balloon angioplasty. In-stent restenosis (ISR), however, remains
a pivotal limitation after coronary and peripheral stenting.
1,2 Both neointimal hyperplasia and matrix production, which are
initiated by activated smooth muscle in response to vessel wall
injury, have been shown to play an important role in ISR.
3,4
A water-soluble chlorin derivative, 13,17-bis[1-carboxypropionyl] carbamoylethyl-3-ethenyl-8-ethoxyiminoethylidene-7-hydroxy-2,7,12,18-tetramethyl porphyrin sodium (PAD-S31), which is the analogue of a new photosensitizer (ATX-S10), has the same biological effect as that of ATX-S10 for tumor treatment,5 and it also is known as a sonochemical sensitizer.6 A new treatment modality called sonodynamic therapy (SDT) has been developed, in which the ultrasound-induced cytotoxicity of sonochemical sensitizers inhibits tumor growth. The administration of PAD-S31 followed by ultrasound exposure has been shown to inhibit tumor growth both in vitro and in vivo. As a result, the sonodynamic effect of PAD-S31 on ISR was investigated in a rabbit model.
 |
Methods
|
|---|
Animal Preparation and Stent Implantation
Japanese white rabbits (3 kg weight; n=16, Tla, Tokyo) were
fed a 1% cholesterol diet containing 100 mg of aspirin per 100
g of diet 1 week before stent implantation. The diet was continued
for 4 weeks after stenting. The rabbits were preanesthetized
with xylazine (3 mg/kg) and ketamine (20 mg/kg) intramuscularly
and maintained with pentobarbital intravenously. After cutting
the skin, the femoral artery was exposed. A 5F sheath was inserted
into the femoral artery. A 9-mm stent (Wiktor) was remounted
on a balloon catheter measuring 3 mm in diameter and 20 mm in
length when inflated. After administering 200 U/kg of heparin,
the stent was delivered into the common iliac artery with 6
atmospheres of inflated pressure. A total of 32 stents were
implanted, consisting of 8 stents for the control group, 8 for
the ultrasound exposure group, 8 for the PAD-S31 group, and
8 for the group that received ultrasonic exposure with PAD-S31
(SDT). All rabbits were killed at 28 days after stent implantation.
Sonodynamic Therapy
An ultrasonic transducer measuring 24 mm in diameter (Hitachi Central Research Laboratory) was operated at 1 MHz. The ultrasonic intensity output from the transducer was calibrated against the drive voltage in water in a progressive wave mode. A water bag was placed on the body and the transducer was put on the water bag above the stent under fluoroscopic guidance. The transducer was driven at a voltage level corresponding to the calibrated ultrasonic output intensity of 0.3 W/cm2. The ultrasonic energy was delivered transdermally through the water bag to the stented artery for 15 minutes. As a sonodynamically active agent, 25 mg/kg of PAD-S31 (Photochemical Co, Ltd) was administered intravenously 1 hour before ultrasonic exposure, on the basis of the finding of a previous study7 that showed the maximum tissue concentration to be achieved at 1 hour after the intravenous administration in normal tissue. In 2 rabbits, a thermistor probe (Needle probe, T/T-T30, Omega) was placed in the subcutaneous tissue between the stent and the ultrasound transducer, and the tissue temperature was measured with an analyzing recorder (AR1100A, Yokogawa).
Pathological Studies
The stented arteries were excised and fixed with 10% formalin and embedded in polyester resin (Rigolac 2004 and 70F, Showa Highpolymer Co, Ltd). Each block of the stented artery was divided into 3 parts: both ends and a middle part. Each part was cut with a crystal cutter at 3- to 500-µm widths and then ground (Speed Lap ML-521-d and HT, Maruto Instrument Co, Ltd) to a thickness of 80 µm. Each section was stained with van Gieson. The narrowest part of the stented artery was evaluated as a representative part. The lumen cross-sectional area (CSA), vessel CSA, intimal CSA, and injury score8 were all measured by a pathologist (H.K.) who did not know the treatment category. The percent area stenosis was determined as (vessel CSA-lumen CSA)/vessel CSA.
Statistical Analysis
The data are presented as the mean±SD. Any differences between the individual groups were tested by 1-way ANOVA, and the significance of differences in unpaired measurement was assessed by the Bonferroni t test. P<0.05 was considered to be statistically significant.
 |
Results
|
|---|
All rabbits tolerated the SDT treatment well and no skin erosion
or skeletal muscle damage at the treated site was observed during
the 28-day follow-up. Morphometrical analysis was done 28 days
after stenting. Representative cross-sections for the control,
ultrasound exposure, PAD-S31, and SDT groups are shown in
Figure 1.
The arteries treated with SDT showed mild neointimal hyperplasia,
whereas the control, ultrasound exposure, and PAD-S31 arteries
all showed moderately increased neointimal hyperplasia. No significant
difference was seen in the injury scores among the control,
ultrasound, PAD-S31, or SDT groups (1.8±0.4, 1.6±0.4,
1.7±0.2, and 1.7±0.4). The lumen CSA (
Figure 2)
was significantly larger in the SDT group (ANOVA,
P<0.01)
than in the control group (2.75±0.74 versus 1.43±0.39
mm
2;
P<0.05), but not larger than in the ultrasound and PAD-S31
groups (2.07±0.82 and 2.45±0.45 mm
2). The intimal
CSA was significantly smaller in the SDT group (ANOVA,
P<0.001)
than in the control, ultrasound, and PAD-S31 groups (0.31±0.07
versus 1.38±0.47, 1.66±0.71, and 1.61±0.42
mm
2;
P<0.05). The ratio of intimal and medial CSA was also
smaller in the SDT group (ANOVA,
P<0.001) than in the control,
ultrasound, and PAD-S31 groups (0.71±0.22 versus 2.53±1.39,
2.48±0.60, and 3.45±1.42 mm
2;
P<0.05). In addition,
no differences were seen regarding the medial CSA in any of
the 4 groups. The percent area stenosis was smaller in the SDT
group (ANOVA,
P<0.001) than in the control, ultrasound, and
PAD-S31 groups (23±7% versus 58±11%, 53±14%,
and 65±9%;
P<0.05). The histological findings of the
arteries treated with SDT seemed to be essentially normal. The
tissue temperature during ultrasound exposure was measured in
2 rabbits undergoing SDT. The mean change in the temperature
was +1.4±1.3°C.

View larger version (116K):
[in this window]
[in a new window]
|
Figure 1. Histological cross-sections of a stented segment in the rabbit iliac artery. Neointimal growth occurred in the control (A), ultrasound exposure (B), and PAD-S31 (C) groups. The degree of neointimal growth was less in the SDT group (D).
|
|

View larger version (21K):
[in this window]
[in a new window]
|
Figure 2. The lumen, intimal, medial, and vessel CSA, ratio of the intimal and medial CSA, and percent area stenosis at 28 days for the control (Cont), ultrasound exposure (US), PAD-S31 (PAD), and sonodynamic therapy (SDT) groups. * P<0.05 versus Cont, P<0.05 versus US, P<0.05 versus PAD.
|
|
 |
Discussion
|
|---|
In the present study, we demonstrated that SDT with PAD-S31
reduced the degree of intimal hyperplasia after stent implantation
in the rabbit iliac artery, by the intravenous administration
of sonochemical sensitizer and transdermal ultrasound exposure.
Mechanism for the activation of a sensitizer by ultrasound is
still not well understood. There are 2 possibilities,
5,9,10 namely: (1) Ultrasound energy itself activates a sonochemical
sensitizer, or (2) acoustic cavitation induces light emission
(sonoluminescence), which thus activates the sonochemical sensitizer.
The spectrum of sonoluminescence
11 is broad in wavelength, from
200 nm to 700 nm, with the most intense emission at

300 to 400
nm. The most intense absorbance of PAD-S31 is

400 nm in wavelength,
which is

4 times greater than that at a second peak absorbance
of 670 nm. A shorter wavelength of light has less transmission
in tissue depth than a longer wavelength has. Therefore, despite
the low activation efficiency of 670 nm, coherent light, a longer
wavelength of light,
12 must be used to clinically activate the
sensitizer in photodynamic therapy (PDT). On the other hand,
ultrasound has an appropriate tissue penetration, no matter
which mechanism activated the sensitizer.
PDT is an intriguing new approach for inhibiting injury-induced intimal hyperplasia.11 An activated sensitizer induced cell depletion, while inhibiting both cell growth and cytokine release and thereby inhibiting the development of intimal hyperplasia.1315 Ultrasonically activated PAD-S31 reduced the degree of intimal hyperplasia without medial mass reduction. Interestingly, the cell nuclei counts in the media did not seem to decrease and no acellular media was observed in the SDT group.
Recently, sonotherapy16 was introduced to prevent ISR both experimentally and clinically. In our study, SDT inhibited neointimal hyperplasia after stenting more effectively than ultrasound alone. Ultrasound itself had no effect on the inhibition of neointimal hyperplasia because the ultrasound power we used in this study was insufficient for sonotherapy. However, the ultrasound power was strong enough to have a sonodynamic effect on PAD-S31. The different ultrasound used in our study (operating mode of continuous wave versus pulse, frequency of 1 MHz versus 700 KHz, exposure time of 15 minutes versus 2 minutes) may also have played a role in the better results of SDT.
Limitations
The number of rabbits studied in each group was relatively small and the follow-up period was only 28 days. We studied only one condition of ultrasound (1 MHz, 0.3 W/cm2) and sonochemical sensitizer (25 mg/kg of PAD-S31). This dose6 was the same as that used for tumor treatment with SDT in an animal model. This treatment has the potential to be clinically adopted; however, further studies elucidating the dose-response of the ultrasound output and sonochemical sensitizer and the long-term effect are still called for.
Conclusions
These preliminary data suggest that SDT with PAD-S31 be considered a feasible treatment modality for noninvasively inhibiting neointimal hyperplasia in a rabbit iliac stent model. However, the exact mechanism of SDT on the vessel wall and the optimal conditions for using the ultrasound output and sonochemical sensitizer still remain to be elucidated in future studies.
Received September 10, 2001;
revision received November 15, 2001;
accepted November 15, 2001.
 |
References
|
|---|
-
Serruys PW, Luijten HE, Beatt KJ, et al. Incidence of restenosis after successful coronary angioplasty: time-related phenomenon: quantitative angiographic study in 342 consecutive patients at 1, 2, 3, and 4 months. Circulation. 1988; 77: 361371.[Abstract/Free Full Text]
-
Fischman DL, Leon M, Baim DS, et al. A randomized comparison of coronary-stent placement and balloon angioplasty in the treatment of coronary artery disease: stent restenosis study investigators. N Engl J Med. 1994; 331: 496501.[Abstract/Free Full Text]
-
Miller M, Kuntz R, Friedrich S, et al. Frequency and consequences of intimal hyperplasia in specimens retrieved by directional atherectomy of native primary coronary stenosis and subsequent restenosis. Am J Cardiol. 1993; 71: 652658.[CrossRef][Medline]
[Order article via Infotrieve]
-
Simons M, Leclerc G, Safian R, et al. Relation between activated smooth muscle cells in coronary-artery lesions and restenosis after atherectomy. N Engl J Med. 1993; 328: 608613.[Abstract/Free Full Text]
-
Mori M, Sakata I, Hirano T, et al. Photodynamic therapy for experimental tumors using ATX-S10(Na), a hydrophilic chlorin photosensitizer, and diode laser. Jpn J Cancer Res. 2000; 91: 753759.[Medline]
[Order article via Infotrieve]
-
Yumita N, Nishigaki R, Sakata I, et al. Sonodynamically induced antitumor effect of 4-formyloximethylidene-3-hydroxy-2-vinyl-deuteerio-porphynyl(IX)-6,7-diaspartic acid (ATX-S10). Jpn J Cancer Res. 2000; 91: 255260.[Medline]
[Order article via Infotrieve]
-
Monden S. Intracranial and systemic distribution of new photosensitizer ATX-S10 in malignant brain tumor model. J Hiroshima Med Ass. 1995; 43: 7785.
-
Schwartz RS, Huber KC, Murphy JG, et al. Restenosis and proportional neointimal response to coronary artery injury: results in a porcine model. J Am Coll Cardiol. 1992; 19: 267274.[Abstract]
-
Umemura S, Yumita N, Nishigaki R, et al. Mechanism of cell damage by ultrasound in combination with hematoporphyrin. Jpn J Cancer Res. 1990; 81: 962966.[Medline]
[Order article via Infotrieve]
-
Misik V, Riesz P. Free radical intermediated in sonodynamic therapy. Ann N Y Acad Sci. 2000; 899: 335348.[CrossRef][Medline]
[Order article via Infotrieve]
-
Hiller R, Putterman SJ, Barber BP. Spectrum of synchronous picosecond sonoluminescence. Phys Rev Lett. 1992; 69: 11821184.[CrossRef][Medline]
[Order article via Infotrieve]
-
Rockson SG, Lorenz DP, Cheong WF. Photoangioplasty: an emerging clinical cardiovascular role for photodynamic therapy. Circulation. 2000; 102: 591596.[Abstract/Free Full Text]
-
LaMauraglia G, Adili F, Karp S, et al. Photodynamic therapy inactivates extracellular matrix-basic fibroblast growth factor: insights to its effect on the vascular wall. J Vasc Surg. 1997; 25: 10441052.[CrossRef][Medline]
[Order article via Infotrieve]
-
Statius van Eps RG, Mark LL, Schiereck J, et al. Photodynamic therapy inhibits the injury-induced fibrotic response of vascular smooth muscle cells. Eur J Vasc Endovasc Surg. 1999; 18: 417423.[CrossRef][Medline]
[Order article via Infotrieve]
-
Adili F, Statius van Eps RG, Karp S, et al. Differential modulation of vascular endothelial and smooth muscle cell function by photodynamic therapy of extracellular matrix: novel insights into radical-mediated prevention of intimal hyperplasia. J Vasc Surg. 1996; 23: 698705.[CrossRef][Medline]
[Order article via Infotrieve]
-
Fitzgerald PJ, Takagi A, Moore P, et al. Intravascular sonotherapy decreases neointimal hyperplasia after stent implantation in swine. Circulation. 2001; 103: 18281831.[Abstract/Free Full Text]
This article has been cited by other articles:

|
 |

|
 |
 
A. N Cheema, N. Nili, C. W Li, H. A Whittingham, J. Linde, R. J van Suylen, M. R Eskandarian, A. P Wong, B. Qiang, J.-F. Tanguay, et al.
Effects of intravascular cryotherapy on vessel wall repair in a balloon-injured rabbit iliac artery model
Cardiovasc Res,
July 1, 2003;
59(1):
222 - 233.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
C J McMahon, C E Mullins, and H G El Said
Intrastent sonotherapy in pulmonary vein restenosis: a new treatment for a recalcitrant problem
Heart,
February 1, 2003;
89(2):
e6 - 6.
[Abstract]
[Full Text]
[PDF]
|
 |
|