(Circulation. 2002;105:1887.)
© 2002 American Heart Association, Inc.
Brief Rapid Communications |
From the Cardiovascular Center Bethanien (H.S., T.T., D.F., U.K., D.S.), Frankfurt; Germany; University of California (M.D.L., T.N.), San Francisco; Appriva Medical (M.D., T.N.), Sunnyvale, Calif; the Department of Cardiology (H.O.), University Hospital Bonn, Bonn, Germany; Centro Cardiologico Monzino Istituto di Ricerca e Cura a Carattere Scientifico (A.B., P.D.B.), Milano, Italy; Swedish Medical Center (M.R.), Seattle, Wash; Hospital S Raffaele (C.D.), Milano, Italy; Emory University Hospital (P.B.), Atlanta, Ga; and St Lukes Hospital (P.K.), Kansas City, Mo.
Correspondence to Horst Sievert, MD, Cardiovascular Center Bethanien, Im Prüfling 23, 60389 Frankfurt, Germany. E-mail horst.sievert@ dgn.de
| Abstract |
|---|
|
|
|---|
Methods and Results We evaluated the feasibility and safety of implanting a novel device for percutaneous left atrial appendage transcatheter occlusion (PLAATO). LAA occlusion using the PLAATO system was attempted in 15 patients with chronic AF at high risk for stroke, who are poor candidates for long-term warfarin therapy. The implant consists of a self-expanding nitinol cage covered with a polymeric membrane (ePTFE). The LAA was successfully occluded in 15/15 patients (100%). Angiography and transesophageal echocardiography (TEE) during the procedure showed that the device was well-seated in all patients and that there was no evidence of perforation, device embolization, or interference with surrounding structures. In 1 patient, the first procedure was complicated by a hemopericardium, which occurred during LAA access. A second attempt 30 days later was successful with no untoward sequela. No other complications occurred. At 1-month follow-up, chest fluoroscopy and TEE revealed continued stable implant position with smooth atrial-facing surface and no evidence of thrombus.
Conclusions Thus, transcatheter closure of the LAA is feasible in humans. This novel implant technology may be appropriate for patients with AF who are not suitable candidates for anticoagulation therapy. Further trials are needed to show the long-term safety and its efficacy in reducing stroke.
Key Words: atrial flutter embolism stroke thrombus atrium
| Introduction |
|---|
|
|
|---|
A new device has been developed that allows percutaneous left atrial appendage transcatheter occlusion (PLAATO) via transseptal catheterization. Studies in dogs demonstrated the ability of the device to seal the LAA.5 We report our initial experience in humans.
| Methods |
|---|
|
|
|---|
PLAATO Device and Procedure
The PLAATO system (Appriva Medical), consists of an implant and a delivery catheter. The implant (Figure 1) is a self-expanding nitinol cage (range of diameters from 15 to 32 mm) covered with an occlusive expanded polytetrafluoroethylene (ePTFE) membrane, which is laminated directly to the frame structure so that the perimeter has intimate contact with the inner wall of the appendage. The purpose of the membrane is both to occlude the orifice of the left atrial appendage and to allow tissue incorporation into the device. Small anchors along the struts and passing through the occlusive membrane assist with device anchoring and encourage healing response. The device is delivered through a custom 14Fr transseptal sheath curved to point at the LAA.
|
All procedures were performed under TEE guidance. After transseptal access, angiography was performed to assess the size and shape of the appendage (Figure 2a). An initial device was chosen that had a diameter 20% to 40% larger than the diameter of the LAA ostium. Patients were given heparin (after transseptal puncture) in order to keep activated clotting time (ACT) >250 seconds. The delivery catheter and transseptal sheath were withdrawn, revealing the implant, which was allowed to expand, filling the appendage. Radiocontrast was injected both distal (through a special lumen in the device; Figure 2b) and proximal (ie, in the left atrium [LA]) to assess for leaks and positioning. If sealing was not adequate, the device was collapsed, repositioned, and re-expanded, or completely removed while transseptal access was maintained and replaced with a device of different diameter. Until final release, devices were completely retrievable. Final seal was evaluated via contrast fluoroscopy in the LA once the device had been completely released (Figure 2c).
|
Follow-Up Evaluation
All patients were placed on aspirin, 300 mg per day indefinitely, and clopidogrel, 75 mg per day for 6 months after the procedure. Patients were assessed with sequential chest fluoroscopy, transthoracic echocardiography, and clinical examination. TEE with color Doppler was performed 1 month after the procedure.
| Results |
|---|
|
|
|---|
The LAA was successfully occluded in 15/15 patients. The mean orifice diameter of the LAA was 20.1±4.3 mm and the median size of the final implanted device was 26 mm (range 18 to 32 mm). Mean procedure time from initial groin puncture to removal of all sheaths was 92.7±43.3 minutes.
In 1 patient, the initial procedure was complicated by hemopericardium associated with LAA access before device implantation and diagnosed with intraprocedure TEE. The procedure was halted and pericardiocentesis performed without sequelae. The appendage was successfully occluded 4 weeks later.
The implant was removed and exchanged for a new size in patient Nos. 3, 4, 5, and 8 with no problems. The series of device changes in these 4 patients was 26 to 29 to 18 mm, 29 to 32 mm, 29 to 26 to 23 to 32 mm, and 32 to 32 mm. In 11/15 patients, the first device selected was implanted without need for exchange. At 1-month follow-up in all patients, chest X-ray and TEE revealed stable implant position, no evidence of migration, erosion, or encroachment on surrounding structures, and smooth healing on the LA-facing surface with no thrombi and no remaining atrial shunt after transseptal puncture. There have been no late complications or embolic events during follow-up.
| Discussion |
|---|
|
|
|---|
Despite clear guidelines, warfarin is not used or used improperly on a large scale.1114 In perhaps the largest and most influential survey, Stafford and Singer14,15 found that in only about a third of outpatients was anticoagulation used.
There is an extensive literature supporting the notion that the LAA is the source of cardioemboli in over 90% of patients with AF, at least in those without rheumatic valve disease.1 It is logical therefore to hypothesize that closing or obliterating the LAA would be an effective way to prevent most cardioemboli in patients with AF. Techniques for surgical obliteration have been described.16,17 Nonetheless, surgical or thoracoscopic LAA closure, other than as an adjunctive procedure as recommended by ACC guidelines18 in patients undergoing mitral valve surgery, has not been enthusiastically accepted because of its invasive nature.
PLAATO provides for a less invasive, percutaneous approach to closing the LAA. Previous animal studies of the device with follow-up of up to 1 year have demonstrated occlusion of the LAA with complete healing, absence of erosions, new thrombus formation on the device, or interference with atrial function.19
In this initial cohort of 15 patients, occlusion of the LAA was successful in all, as proven by LA angiography. There have been no complications associated with the device, either acutely during the implantation procedure or during follow-up. The only complication during the study was hemopericardium in the first patient attempted, which was not device-related, resulted from LAA access, and should be easily avoided with increased experience. The procedure was successful in a second attempt in this patient.
All patients have done well in follow-up. One theoretical concern is the development of new thrombi on the implant. However, the use of ePTFE on the implant surface results in particularly benign healing.20 Histological examination in dogs undergoing PLAATO reveal partial endothelialization at 1 month, which is complete by 2 to 3 months. In our patients, TEE at 1 month has shown the surface to be completely smooth and free of mobile thrombi.
Conclusion
This initial study supports the concept that implantation of a mechanical device to occlude the LAA can be done safely and with relative ease. Further studies are needed before PLAATO can be recommended for the large number of patients with AF at risk for stroke. These studies will need to confirm reduction in stroke incidence and longer term safety. The attraction of the procedure, however, resides in its directly addressing the mechanism of stroke in AF and its operational simplicity. It may become a valuable alternative for those patients with chronic nonrheumatic AF in whom standard anticoagulation therapy is contraindicated or poorly tolerated.
Note Added in Proof
Since acceptance of this manuscript, 16 additional patients have been treated. In 1 patient, a hemopericardium occurred during a successful implantation without sequelae. No other complication occurred.
| Footnotes |
|---|
Received January 28, 2002; revision received March 4, 2002; accepted March 4, 2002.
| References |
|---|
|
|
|---|
2. Manning WJ, Waksmonski CA, Haering JM. Sensitivity and specificity of transesophageal echo for left atrial thrombi: a prospective, consecutive surgical study. Circulation. 1994; 90: 1202a.Abstract.
3. Aberg H. Atrial fibrillation, I: a study of atrial thrombosis and systemic embolism in a necropsy material. Acta Med Scand. 1969; 185: 373379.
4. Tsai LM, Chen JH, Lin LJ, et al. Role of transesophageal echocardiography in detecting left atrial thrombus and spontaneous echo contrast in patients with mitral valve disease or non-rheumatic atrial fibrillation. J Formos Med Assoc. 1990; 89: 270274.
5. Nakai T, Gerstenfeld EP, Lesh MD, et al. Assessment of transcatheter left atrial appendage occlusion device for preventing thromboembolism. Am J Cardiol. 2001; 88 (suppl 5A): 119G.Abstract.
6. Wolf PA, Dawber TR, Thomas HE, et al. Epidemiologic assessment of chronic atrial fibrillation and risk of stroke: the Framingham Study. Neurology. 1978; 28: 973977.
7. Gage BF, Waterman AD, Shannon W, et al. Validation of clinical classification schemes for preventing stroke. JAMA. 2001; 285: 28642870.
8. Wolf PA, Benjamin EJ, Belanger AJ, et al. Secular trends in the prevalence of atrial fibrillation: the Framingham Study. Am Heart J. 1996; 131: 790795.
9. Kannel WB, Wolf PA, Benjamin EJ, et al. Prevalence, incidence, prognosis, and predisposing conditions for atrial fibrillation: population-based estimates. Am J Cardiol. 1998; 82: 2N9N.
10. Levine MN, Raskob G, Landefeld S, et al. Hemorrhagic complications of anticoagulant treatment. Chest. 2001; 119 (suppl 1): 108S121S.
11. Munschauer FE, Priore RL, Hens M, et al. Thromboembolism prophylaxis in chronic atrial fibrillation: practice patterns in community and tertiary-care hospitals. Stroke. 1997; 28: 7276.
12. Lawson F, McAlister F, Ackman M, et al. The utilization of antithrombotic prophylaxis for atrial fibrillation in a geriatric rehabilitation hospital. J Am Geriatr Soc. 1996; 44: 708711.
13. Schlicht JR, Davis RC, Naqi K, et al. Physician practices regarding anticoagulation and cardioversion of atrial fibrillation. Arch Intern Med. 1996; 156: 290294.
14. Stafford R, Singer D. National patterns of warfarin use in atrial fibrillation. Arch Intern Med. 1996; 156: 25372541.
15. Stafford RS, Singer DE. Recent national patterns of warfarin use in atrial fibrillation. Circulation. 1998; 97: 12311233.
16. Odell JA, Blackshear JL, Davies E, et al. Thoracoscopic obliteration of the left atrial appendage: potential for stroke reduction? Ann Thorac Surg. 1996; 61: 565569.
17. Johnson WD, Ganjoo AK, Stone CD, et al. The left atrial appendage: our most lethal human attachment! Surgical implications. Eur J Cardiothorac Surg. 2000; 17: 718722.
18. Bonow RO, Carabello B, de Leon AC Jr, et al. ACC/AHA guidelines for the management of patients with valvular heart disease: a report of the American College of Cardiology/American Heart Association Task Force on practice guidelines (committee on management of patients with valvular heart disease). J Am Coll Cardiol. 1998; 32: 14861582.
19. Nakai T, Gerstenfeld EP, Stillson C, et al. Intracardiacechocardiography guided implantation of a novel left atrial appendage occlusion device for preventing thromboembolism. PACE. 2001; 24: 639.Abstract.
20. Virmani R, Kolodgie FD, Dake MD, et al. Histopathologic evaluation of an expanded polytetrafluoroethylene-nitinol stent endoprosthesis in canine iliofemoral arteries. J Vasc Interv Radiol. 1999; 10: 445456.
This article has been cited by other articles:
![]() |
D. R. Holmes Jr and R. S. Schwartz Left Atrial Appendage Occlusion Eliminates the Need for Warfarin Circulation, November 10, 2009; 120(19): 1919 - 1926. [Full Text] [PDF] |
||||
![]() |
P. C. Block, S. Burstein, P. N. Casale, P. H. Kramer, P. Teirstein, D. O. Williams, and M. Reisman Percutaneous Left Atrial Appendage Occlusion for Patients in Atrial Fibrillation Suboptimal for Warfarin Therapy: 5-Year Results of the PLAATO (Percutaneous Left Atrial Appendage Transcatheter Occlusion) Study J. Am. Coll. Cardiol. Intv., July 1, 2009; 2(7): 594 - 600. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Sievert and Y. L. Bayard Percutaneous Closure of the Left Atrial Appendage: A Major Step Forward J. Am. Coll. Cardiol. Intv., July 1, 2009; 2(7): 601 - 602. [Full Text] [PDF] |
||||
![]() |
H. Fumoto, A. M. Gillinov, Y. Ootaki, M. Akiyama, D. Saeed, T. Horai, C. Ootaki, D. G. Vince, Z. B. Popovic, R. Dessoffy, et al. A novel device for left atrial appendage exclusion: the third-generation atrial exclusion device. J. Thorac. Cardiovasc. Surg., October 1, 2008; 136(4): 1019 - 1027. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. L Bayard, S. H Ostermayer, and H. Sievert Alternatives to warfarin in atrial fibrillation: drugs and devices Heart, September 1, 2008; 94(9): 1113 - 1116. [Full Text] [PDF] |
||||
![]() |
P Su, K P McCarthy, and S Y Ho Occluding the left atrial appendage: anatomical considerations Heart, September 1, 2008; 94(9): 1166 - 1170. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. M. Gillinov Advances in Surgical Treatment of Atrial Fibrillation Stroke, February 1, 2007; 38(2): 618 - 623. [Abstract] [Full Text] [PDF] |
||||
![]() |
O. Onalan and E. Crystal Left Atrial Appendage Exclusion for Stroke Prevention in Patients With Nonrheumatic Atrial Fibrillation Stroke, February 1, 2007; 38(2): 624 - 630. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Kamohara, K. Fukamachi, Y. Ootaki, M. Akiyama, F. Cingoz, C. Ootaki, D. G. Vince, Z. B. Popovic, M. W. Kopcak Jr, R. Dessoffy, et al. Evaluation of a novel device for left atrial appendage exclusion: The second-generation atrial exclusion device. J. Thorac. Cardiovasc. Surg., August 1, 2006; 132(2): 340 - 346. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. M. Hylek, J. D'Antonio, C. Evans-Molina, C. Shea, L. E. Henault, and S. Regan Translating the Results of Randomized Trials into Clinical Practice: The Challenge of Warfarin Candidacy Among Hospitalized Elderly Patients With Atrial Fibrillation Stroke, April 1, 2006; 37(4): 1075 - 1080. [Abstract] [Full Text] [PDF] |
||||
![]() |
O. K. Mohrs, R. Schraeder, S. E. Petersen, D. Scherer, B. Nowak, H.-U. Kauczor, and T. Voigtlaender Percutaneous Left Atrial Appendage Transcatheter Occlusion (PLAATO): Planning and Follow-Up Using Contrast-Enhanced MRI Am. J. Roentgenol., February 1, 2006; 186(2): 361 - 364. [Full Text] [PDF] |
||||
![]() |
K. Kamohara, K. Fukamachi, Y. Ootaki, M. Akiyama, F. Zahr, M. W. Kopcak Jr, R. Dessoffy, Z. B. Popovic, M. Daimon, D. M. Cosgrove, et al. A novel device for left atrial appendage exclusion J. Thorac. Cardiovasc. Surg., December 1, 2005; 130(6): 1639 - 1644. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. H. Ostermayer, M. Reisman, P. H. Kramer, R. V. Matthews, W. A. Gray, P. C. Block, H. Omran, A. L. Bartorelli, P. Della Bella, C. Di Mario, et al. Percutaneous Left Atrial Appendage Transcatheter Occlusion (PLAATO System) to Prevent Stroke in High-Risk Patients With Non-Rheumatic Atrial Fibrillation: Results From the International Multi-Center Feasibility Trials J. Am. Coll. Cardiol., July 5, 2005; 46(1): 9 - 14. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Schneider, J. Finsterer, and C. Stollberger Effects of percutaneous left atrial appendage transcatheter occlusion (PLAATO) on left atrial structure and function J. Am. Coll. Cardiol., February 15, 2005; 45(4): 634 - 635. [Full Text] [PDF] |
||||
![]() |
V. Thijs More bad news about atrial fibrillation Eur. Heart J., October 1, 2004; 25(19): 1670 - 1671. [Full Text] [PDF] |
||||
![]() |
I. R. Hanna, P. Kolm, R. Martin, M. Reisman, W. Gray, and P. C. Block Left atrial structure and function after percutaneous left atrial appendage transcatheter occlusion (PLAATO): Six-month echocardiographic follow-up J. Am. Coll. Cardiol., May 19, 2004; 43(10): 1868 - 1872. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. A. Quader, P. M. McCarthy, A. M. Gillinov, J. M. Alster, D. M. Cosgrove III, B. W. Lytle, and E. H. Blackstone Does preoperative atrial fibrillation reduce survival after coronary artery bypass grafting? Ann. Thorac. Surg., May 1, 2004; 77(5): 1514 - 1524. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. S. Kaufman and A. L. Waldo The impact of asymptomatic atrial fibrillation J. Am. Coll. Cardiol., January 7, 2004; 43(1): 53 - 54. [Full Text] [PDF] |
||||
![]() |
C. Stollberger, B. Schneider, and J. Finsterer Elimination of the Left Atrial Appendage To Prevent Stroke or Embolism?: Anatomic, Physiologic, and Pathophysiologic Considerations Chest, December 1, 2003; 124(6): 2356 - 2362. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. W Troughton, C. R Asher, and A. L Klein The role of echocardiography in atrial fibrillation and cardioversion Heart, December 1, 2003; 89(12): 1447 - 1454. [Full Text] [PDF] |
||||
![]() |
D. E. Singer A 60-Year-Old Woman With Atrial Fibrillation JAMA, October 22, 2003; 290(16): 2182 - 2189. [Full Text] [PDF] |
||||
![]() |
J. L. Blackshear, W. D. Johnson, J. A. Odell, V. S. Baker, M. Howard, L. Pearce, C. Stone, D. L. Packer, and H. V. Schaff Thoracoscopic extracardiac obliteration of the left atrial appendage for stroke risk reduction in atrial fibrillation J. Am. Coll. Cardiol., October 1, 2003; 42(7): 1249 - 1252. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. n. Garcia-Fernandez, E. Perez-David, J. Quiles, J. Peralta, I. Garcia-Rojas, J. Bermejo, M. Moreno, and J. Silva Role of left atrial appendageobliteration in stroke reductionin patients with mitral valve prosthesis: A transesophageal echocardiographic study J. Am. Coll. Cardiol., October 1, 2003; 42(7): 1253 - 1258. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. L. Halperin and M. Gomberg-Maitland Obliteration of the leftatrial appendage forprevention of thromboembolism J. Am. Coll. Cardiol., October 1, 2003; 42(7): 1259 - 1261. [Full Text] [PDF] |
||||
![]() |
G.Q. Villani, M. Piepoli, P.E. Villani, and A. Capucci Anticoagulation in atrial fibrillation: what is certain and what is to come Eur. Heart J. Suppl., September 1, 2003; 5(suppl_H): H45 - H50. [Abstract] [PDF] |
||||
![]() |
C. Stollberger, B. Schneider, J. Finsterer, H. Sievert, T. Trepels, D. Fleschenberg, U. Krumsdorf, D. Scherer, M. D. Lesh, T. Nakai, et al. Is Percutaneous Left Atrial Appendage Transcatheter Occlusion an Alternative to Oral Anticoagulation in Patients With Atrial Fibrillation? * Response Circulation, January 7, 2003; 107 (1): e11 - e12. [Full Text] [PDF] |
||||
![]() |
G. Bonanomi, M. A. Zenati, and D. Schwartzman Left atrial appendectomy and maze J. Am. Coll. Cardiol., January 1, 2003; 41(1): 170 - 170. [Full Text] [PDF] |
||||
![]() |
Percutaneous Left Atrial Appendage Occlusion Journal Watch Cardiology, June 7, 2002; 2002(607): 4 - 4. [Full Text] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2002 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |