Department of Pediatrics, Division of Cardiology (S.T., G.W.), Department of Anesthesia and Critical Care Medicine (S.T., L.M.M., C.D.K., J.MS, R.I.G., S.C.N.), Department of Surgery, Division of Cardiothoracic Surgery (T.L.S.), and Department of Respiratory Care (S.M.D., R.I.G.), The Childrens Hospital of Philadelphia and the University of Pennsylvania School of Medicine, Philadelphia, and the Department of Anesthesia (C.R.), Childrens Hospital Medical Center and the University of Washington School of Medicine, Seattle.
Correspondence to Sarah Tabbutt, MD, PhD, The Cardiac Center, The Childrens Hospital of Philadelphia, 34th St and Civic Center Boulevard, Philadelphia, PA 19104. E-mail tabbutt{at}email.chop.edu
| Abstract |
|---|
|
|
|---|
Methods and Results Ten anesthetized and paralyzed preoperative infants with HLHS were evaluated in a prospective, randomized, crossover trial comparing hypoxia (17% FIO2) with hypercarbia (2.7% FICO2). Each patient was treated in a random order (10 minutes per condition) with a recovery period (15 to 20 minutes) in room air. Arterial (SaO2) and superior vena caval (SvO2) co-oximetry and cerebral oxygen saturation (ScO2) measurements were made at the end of each condition and recovery period. ScO2 was measured by near infrared spectroscopy. Hypoxia significantly decreased both SaO2 (-5.2±1.1%, P=0.0014) and SvO2 (-5.6±1.7%, P=0.009) compared with baseline, but arteriovenous oxygen saturation (AVO2) difference (SaO2-SvO2) and ScO2 remained unchanged. Hypercarbia decreased SaO2 (-2.6±0.6%, P=0.002) compared with baseline but increased both ScO2 (9.6±1.8%, P=0.0001) and SvO2 (6±2.2%, P=0.022) and narrowed the AVO2 difference (-8.5±2.3%, P=0.005). Both hypoxia and hypercarbia decreased the balance between pulmonary and systemic blood flow (Qp:Qs) compared with baseline.
Conclusions In preoperative infants with HLHS, under conditions of anesthesia and paralysis, although Qp:Qs falls in both conditions, oxygen delivery is unchanged during hypoxia and increased during hypercarbia. These data cannot differentiate cerebral from systemic oxygen delivery.
Key Words: heart defects, congenital hypoxia cardiac output blood flow physiology
| Introduction |
|---|
|
|
|---|
Our goal was to determine in a prospective, randomized, crossover study the impact of hypoxia (17% FIO2) versus hypercarbia (2.7% FICO2) on oxygen delivery in preoperative neonates with HLHS under conditions of anesthesia, paralysis, and fixed minute ventilation.
| Methods |
|---|
|
|
|---|
Inspired Gas Mixtures
All patients were ventilated in a pressure-regulated volume control mode with a Servo 300 Ventilator (Siemenes-Elema). Hypercarbic gas mixture was administered by using carbon dioxide metered into the ventilator outflow port. The flow was adjusted until the PICO2 was 20 mm Hg (equivalent to a FICO2 of 2.7%) as directly measured by a capnometer in the inspiratory limb of the ventilator circuit. Hypoxic gas mixture was administered by using nitrogen metered into the ventilator O2 inflow port. The flow was adjusted until an FIO2 of 17% was directly measured in the ventilator circuit by using an oxygen analyzer. This configuration enabled maintenance of a constant minute ventilation throughout the study. Actual tidal volume was determined by subtracting the compressible volume lost in the ventilator circuit from the exhaled tidal volume measured on the ventilator. The ventilator measured minute ventilation.
Near-Infrared Spectroscopy
Near-infrared spectroscopy (NIRS) is an optical technology that has been shown in children to measure a mixed cerebral vascular oxygen saturation, ScO2 (including capillaries, venules, and arterioles), with the use of phase-shifted infrared light where oxygenated and deoxygenated hemoglobin have distinct absorption spectra.14 The near-infrared spectrophotometer used in the present study is a frequency-domain prototype cerebral oximeter (PMD 4002, NIM Inc). NIRS is noninvasive, with optical fibers contained in a foam rubber probe that is applied to the infants forehead below the hairline with hook-and-loop fastener (Velcro) straps. Light is generated by 3 class I laser diodes emitting at 754 nm, 785 nm, and 816 nm. Continuous ScO2 was computer-recorded and downloaded to a disk for storage. NIRS data were available for 8 of the 10 patients.
Vascular Access and Patient Monitoring
All patients had an indwelling umbilical or peripheral arterial catheter for continuous blood pressure monitoring and obtaining arterial blood gas and co-oximetry (SaO2) measurements. Infants with the intracardiac anatomy of mitral stenosis/aortic stenosis have the potential to generate antegrade flow in the ascending aorta with higher oxygen content than the pulmonary artery and ductal blood. Patients with aortic stenosis had their arterial catheters in the umbilical artery (n=3) or posterior tibial artery (n=1) and, therefore, demonstrated no difference in baseline SaO2 compared with patients with aortic atresia. All patients had an indwelling SVC catheter for obtaining mixed venous co-oximetry measurements (SvO2). All patients had continuous end-tidal CO2, transcutaneous oxygen saturation, heart rate, and temperature monitoring.
Study Protocol
Patients were stabilized at a baseline of 21% FIO2, with a goal PaCO2 of 40 mm Hg. Gas mixtures were administered in randomized order. Each condition (hypoxia or hypercarbia) was delivered for 10 minutes, followed by a baseline recovery of 15 to 20 minutes in 21% FIO2. The entire protocol took 65 minutes. Simultaneous measurements were made at the end of each condition and at the end of each baseline or recovery period, including heart rate, temperature, blood pressure, transcutaneous oxygen saturation, arterial blood gas and co-oximetry (SaO2), NIRS (ScO2), and SVC co-oximetry (SvO2).
Concurrent Therapies
No changes were made in prostaglandin E1 or inotropic infusions during the study period. No boluses of calcium, bicarbonate, or volume (crystalloid or blood products) were administered during the study period. All patients were anesthetized with a fentanyl bolus (median 20, range 10 to 20 µg/kg) and infusion (median 2.5, range 0 to 3 µg · kg-1 · h-1), and paralyzed with a pancuronium bolus (0.2 mg/kg for all patients).
Qp:Qs Calculations
The Qp:Qs ratio can be measured by applying the Fick principle to infants with HLHS: Qp:Qs=(SaO2-SvO2)/(SpvO2-SaO2), where SaO2, SvO2, and SpvO2 are the arterial, mixed systemic venous, and pulmonary venous oxygen saturations, respectively. SaO2 was directly measured. The SVC saturation was used to represent SvO2. The SpvO2 was calculated by using the alveolar gas equation: PAO2=(PATM-PWATER)xFIO2-PCO2/RQ, where PAO2 is the alveolar partial pressure of oxygen (mm Hg), PATM is the atmospheric pressure (760 mm Hg at sea level), PWATER is the water vapor pressure (47 mm Hg at 37°C), FIO2 is the fraction of inspired oxygen, PCO2 is the arterial partial pressure of carbon dioxide, and RQ is the respiratory quotient (0.8). FIO2 and PCO2 were directly measured (see Table 3). Equations are as follows: for baseline, PAO2=713x0.21-38/0.8=102.2 mm Hg; for hypoxia, PAO2=713x 0.17-36/0.8=76.2 mm Hg; and for hypercarbia, PAO2=713x 0.21-54/0.8=82.2 mm Hg.
|
No patient had demonstrable pulmonary pathology by chest radiograph, and we assumed that the pulmonary vein oxygen saturation (SpvO2) was equal to the alveolar oxygen saturation (SAO2). The Hill equation describes the oxyhemoglobin dissociation curve, SpvO2= PAO2n÷(PAO2n+P50n), with use of a Hill coefficient (n) of 3, and if 80% hemoglobin F (P50F=18) and 20% hemoglobin A (P50A=27) are assumed,15 then SpvO2=0.8[PAO23÷(PAO23+ 183)]+ 0.2[PAO23÷(PAO23+273)].
Therefore, SpvO2=0.990 in room air (FIO2=0.21), SpvO2=0.982 during hypoxia (FIO2=0.17), and SpvO2=0.985 during hypercarbia (FICO2=0.027). Because of the potential error with a very narrow SpvO2-SaO2, a maximum Qp:Qs value of 5 was assigned if the calculated Qp:Qs was
5.
Oxygen Delivery Calculations
Oxygen delivery (DO2) is defined as the product of arterial oxygen content (CaO2) and cardiac output. Oxygen consumption (
O2) is defined as the product of the oxygen extracted (CaO2-CVO2), where CVO2 is the mixed venous oxygen content, and the cardiac output. Oxygen content is defined as (PO2x0.003)+(1.34xHbxSO2). Under conditions of normal or low FIO2, the quantity of dissolved oxygen (PO2x0.003) is negligible, and oxygen delivery can be defined as follows: DO2=
O2xSaO2/ (SaO2-SvO2).
If it is assumed that (under the study conditions of anesthesia and paralysis) oxygen consumption remains unchanged, then oxygen delivery should be proportional to SaO2/(SaO2-SvO2). For patients in which SaO2 remains constant, trends in oxygen delivery will reflect SvO2.
Statistical Analysis
Paired 2-tailed t test analysis was used to evaluate the significance of changes in measured variables between each condition and the average of the precondition and postcondition baseline or recovery period. Data with nonnormal distribution was analyzed by the nonparametric Wilcoxon signed rank test. Statistical analyses were performed by using SigmaStat software (Jandel Scientific Software, SSPS Science).
| Results |
|---|
|
|
|---|
|
Two inspired gas conditions, hypoxia (17% FIO2) and hypercarbia (2.7% FICO2, equivalent to 20 mm Hg PICO2), were applied to each patient for 10 minutes each, in randomized order. Each condition was preceded and followed by a 15- to 20-minute baseline or recovery period in 21% FIO2. Measurements at the end of each condition were compared with the average of the precondition and postcondition baseline by paired 2-tailed t test. All patients were anesthetized, paralyzed, and maintained with fixed minute ventilation throughout the study period. During the study, there was no significant change in temperature (36.6±0.2°C before study, 36.4±0.3°C after study; P=0.35) or hemoglobin (14.3± 0.6 g/dL before study, 13.7± 0.7 g/dL after study; P=0.52). All patients completed the study protocol and remained hemodynamically stable throughout.
Ventilation Parameters
All patients were ventilated with a Servo 300 ventilator in a pressure-regulated volume-control mode with the following set parameters: tidal volume of 8.9±0.7 cm3/kg, mandatory breath rate of 19±0.5 cycles per minute, and positive end-expiratory pressure of 0.5±0.5 cm H2O. Measured parameters included peak inspiratory pressure (21±0.9 cm H2O) and minute ventilation that did not significantly differ between condition and baseline (hypoxia 1.52±0.14 L/min, baseline 1.32±0.08 L/min [P=0.25]; hypercarbia 1.37±0.08 L/min, baseline 1.32±0.08 L/min [P=0.06]).
Hemodynamic Data
Table 2 shows the transcutaneous oxygen saturation, heart rate, and systolic and diastolic arterial blood pressures. There was a significant decrease in transcutaneous oxygen saturation for both hypoxia and hypercarbia compared with baseline. There was a significant increase in both systolic and diastolic blood pressure with hypercarbia but not with hypoxia. There was no significant change in heart rate with either hypoxia or hypercarbia.
|
Arterial Blood Gas Data
Table 3 shows the arterial pH, oxygen tension (PO2), and carbon dioxide tension (PCO2) for hypoxia and hypercarbia compared with baseline. Hypoxia resulted in a significant decrease in PO2, a small but significant increase in pH, and an insignificant decrease in PCO2. Hypercarbia did not affect PO2 but significantly decreased pH and increased PCO2.
Co-Oximetry and NIRS Data
Table 4 shows the arterial (SaO2) and mixed venous (SvO2) co-oximetry and the arteriovenous saturation (AVO2) difference data. SaO2 co-oximetry was directly measured from an umbilical or peripheral artery catheter, SvO2 co-oximetry was directly measured from a superior vena cava catheter, and AVO2 difference is (SaO2-SvO2). Figure 1 shows the difference between hypoxia or hypercarbia and baseline for SaO2, SvO2, AVO2 difference, and ScO2, with significance determined by paired t test. Hypoxia significantly decreased both SaO2 (-5.2±1.1%, P=0.0014) and SvO2 (-5.6±1.7%, P=0.009), but AVO2 difference (0.44±1.4, P=0.76) and ScO2 (-0.4±1.5%, P=0.8) remained unchanged. Hypercarbia decreased arterial saturation SaO2 (-2.6±0.6%, P=0.002) but increased both ScO2 (9.6±1.8%, P=0.0001) and SvO2 (6±2.2%, P=0.022) and narrowed the AVO2 difference (-8.5±2.3%, P=0.005).
|
|
Qp:Qs Calculations
The Qp:Qs ratio was determined by (SaO2-SvO2)÷ (SpvO2-SaO2), where SaO2 and SvO2 were directly measured, and the pulmonary vein oxygen saturation (SpvO2) was assumed as 99% for baseline, 98.2% for hypoxia, and 98.5% for hypercarbia (see Methods). Qp:Qs decreased for both hypoxia (2.55±0.48 versus 3.36±0.46, P=0.056) and hypercarbia (2.19±0.55 versus 3.11±0.45, P=0.026) compared with baseline (Figure 2).
|
Oxygen Delivery Calculations
Figure 3 shows that with hypoxia there was an insignificant decrease in DO2 compared with baseline (3.62±0.50 versus 3.73±0.49, P=0.70); however, hypercarbia demonstrated a significant increase in DO2 compared with baseline (6.14±1.43 versus 3.64±0.45, P=0.004).
|
| Discussion |
|---|
|
|
|---|
Several animal studies have evaluated inspired gas mixtures after cardiopulmonary bypass. Immediately after surgical creation of a single-ventricle model in newborn piglets, increased FICO216 or decreased FIO217 resulted in increased PVR and decreased Qp:Qs. Reddy et al18 report the only prebypass data by creating a single-ventricle model (proximal Damus-Kaye-Stancel anastomosis, placement of a 5-mm aortopulmonary shunt, and ligation of the distal main pulmonary artery) in near-term fetal lambs.18 At 2 to 3 days after delivery and 30 minutes after ligation of the patent ductus arteriosus, the lambs demonstrated increased PVR (48%, P=0.0003; 35%, P<0.0001) and decreased Qp:Qs (-32%, P<0.001; -27%, P=0.001) after 5 minutes of profound alveolar hypoxia (10% FIO2) or after 10 minutes of hypercarbia (35 mm Hg PICO2), respectively.
We report the first human study to evaluate the effects of inspired gas mixtures on oxygen delivery. Our goal was to determine in a prospective, randomized, crossover study, the impact of hypoxia versus hypercarbia on oxygen delivery in preoperative infants with HLHS. In an effort to minimize patient variables, we evaluated only patients with stable hemodynamics and an absence of pulmonary pathology by chest radiograph. Additionally, all patients were under identical conditions of anesthesia, paralysis, and fixed minute ventilation. Patients (n=10) were exposed in a randomized order to 2 inspired gas conditions, hypoxia (17% FIO2) and hypercarbia (2.7% FICO2), for 10 minutes each. Each condition was preceded and followed by a 15- to 20-minute baseline or recovery period in 21% FIO2. Throughout the study, patients were successfully maintained without a significant change in temperature, hemoglobin, or minute ventilation.
Barnea et al19 eloquently described a mathematical model of oxygen delivery (DO2) in newborns with single-ventricle physiology.19 In the unique situation of a ductal-dependent (preoperative) or shunted (postoperative) patient with single-ventricle physiology, the DO2 is dependent not only on the total cardiac output but also on the balance between pulmonary and systemic blood flow (Qp:Qs). Accurate measurements of cardiac output and Qp:Qs have been reported only from animal studies with flow probes on the pulmonary artery and aorta.1618 In these studies, a close correlation has been shown between systemic mixed venous saturation (SvO2) and both DO2 and Qp:Qs, with a maximum SvO2 at Qp:Qs of 1:1.18 In the absence of a pulmonary venous catheter, calculations of Qp:Qs require an estimation of SpvO2. The errors introduced by estimations of SpvO2 are minimized at higher values of SpvO2.19 Calculations of SpvO2 (see Methods) corrected for fetal hemoglobin but did not account for small but measurable changes in pH. Based on calculated estimations of SpvO2 (baseline 99%, hypoxia 98.2%, and hypercarbia 98.5%), a small, but significant, decrease in Qp:Qs was found for both hypoxia and hypercarbia.
Total cardiac output can be augmented by elevated arterial PCO2, with or without increases in heart rate.20 This effect is thought to be secondary to sympathetic stimulation and can be blunted by effective anesthesia. Although total cardiac output could not be measured in the present study, under conditions of hypercarbia, there was a significant increase in blood pressure without a change in heart rate compared with baseline. These changes were not observed with hypoxia.
Animal and clinical data suggest that SvO2,3,21,22 particularly in conjunction with the AVO2 difference,23 is a reliable estimate of systemic DO2 in infants with HLHS. Others have argued that SaO2÷(SaO2-SvO2) provides the best estimate of DO2.19,24 By use of SvO2 as an estimate of DO2, hypoxia significantly decreases DO2, whereas hypercarbia significantly increases DO2. However, by use of SaO2÷(SaO2-SvO2) as an estimate of DO2, the decrease in DO2 with hypoxia becomes insignificant, whereas the increase in DO2 with hypercarbia remains significant. Whichever approach one favors, the AVO2 difference remains unchanged with hypoxia, indicating that systemic oxygen delivery continues to exceed oxygen demand. However, with hypercarbia, the AVO2 difference narrows significantly, consistent with an increase in the oxygen delivery-to-consumption ratio.
NIRS measures a mixed cerebral vascular (capillaries, arterioles, and venules) oxygen saturation, ScO2, in the neocortex, reflecting oxygen extraction and delivery. ScO2 increases with decreased cerebral metabolic rate (eg, hypothermia) or with increased oxygen delivery (eg, increased cerebral blood flow, higher SaO2).14 Hypercarbia significantly increased ScO2, whereas hypoxia had no effect on ScO2. If no difference in cerebral oxygen extraction between hypoxia and hypercarbia is assumed, the increase in ScO2 with hypercarbia may reflect a combination of increased SaO2 and increased cerebral blood flow. In summary, whereas both hypoxia and hypercarbia lowered Qp:Qs, it is possible that total cardiac output is decreased by hypoxia and increased by hypercarbia.
Unlike patients without intracardiac shunting, it is not possible to accurately measure a true mixed systemic venous saturation in a newborn with HLHS. Current standard clinical practice is the use of superior vena caval co-oximetry to represent the SvO2.3,23,25 Therefore, the present study is limited by the inability to differentiate between cerebral and systemic oxygen delivery. Increased PaCO2 and decreased pH can increase cerebral blood flow. Thus, although hypercarbia clearly increases DO2, we cannot determine whether this represents isolated improved cerebral DO2 or a combination of improved cerebral and systemic DO2. However, indirect evidence, including increased systemic arterial blood pressure and decreased Qp:Qs, would imply some improvement in systemic cardiac output.
The results of the present study apply for patients under anesthesia, paralysis, and controlled minute ventilation. This is clinically applicable to those infants with HLHS who have presented in shock or those with a functionally unrestricted atrial septum and hemodynamically compromising pulmonary overcirculation. For these infants who most often require sedation, paralysis, and controlled minute ventilation, the present study suggests that hypercarbia is the preferred therapy. The present study also supports the use of hypercarbia during the induction and maintenance of general anesthesia before cardiopulmonary bypass irrespective of the surgical approach (stage 1 reconstruction or primary cardiac transplantation). However, the majority of preoperative infants with HLHS do not require controlled ventilation. Further studies are important to determine whether there is a role for manipulation of inspired gases in these patients, to assess what optimal inspired gas mixture may minimize the risks of pulmonary overcirculation (eg, necrotizing enterocolitis), and to maximize neurological outcome.
| Acknowledgments |
|---|
| References |
|---|
|
|
|---|
2.
Hraska V, Nosal M, Sykora P, et al. Results of modified Norwoods operation for hypoplastic left heart syndrome. Eur J Cardiothorac Surg. 2000; 18: 214219.
3.
Tweddell JS, Hoffman GM, Fedderly RT, et al. Patients at risk for low systemic oxygen delivery after the Norwood procedure. Ann Thorac Surg. 2000; 69: 18931899.
4.
Mahle WT, Clancy RR, Moss EM, et al. Neurodevelopmental outcome and lifestyle assessment in school-aged and adolescent children with hypoplastic left heart syndrome. Pediatrics. 2000; 105: 10821089.
5. Barnea O, Austin EH, Richman B, et al. Balancing the circulation: theoretic optimization of pulmonary/systemic flow ratio in hypoplastic left heart syndrome. J Am Coll Cardiol. 1994; 24: 13761381.[Abstract]
6.
McElhinney DB, Hedrick HL, Bush DM, et al. Necrotizing enterocolitis in neonates with congenital heart disease: risk factors and outcomes. Pediatrics. 2000; 106: 10801087.
7. Shime N, Hashimoto S, Hiramatsu N, et al. Hypoxic gas therapy using nitrogen in the preoperative management of neonates with hypoplastic left heart syndrome. Pediatr Crit Care Med. 2000; 1: 3841.[Medline] [Order article via Infotrieve]
8.
Day RW, Barton AP, Pysher TJ. Pulmonary vascular resistance of children treated with nitrogen during early infancy. Ann Thorac Surg. 1998; 65: 14001404.
9. Jobes DR, Nicolson SC, Steven JM, et al. Carbon dioxide prevents pulmonary overcirculation in hypoplastic left heart syndrome. Ann Thorac Surg. 1992; 54: 150151.[Abstract]
10. Gullquist SD, Schmitz ML, Hannon GD, et al. Carbon dioxide in the inspired gas improves early postoperative survival in neonates with congenital heart disease following stage 1 palliation (Norwood). Circulation. 1992; 86 (suppl I): I-360. Abstract.
11. Chang AC, Zucker HA, Hickey PR, et al. Pulmonary vascular resistance in infants after cardiac surgery: role of carbon dioxide and hydrogen ion. Crit Care Med. 1995; 23: 568575.[Medline] [Order article via Infotrieve]
12.
Wessel DL. Commentary. simple gases and complex single ventricles. J Thorac Cardiovasc Surg. 1996; 112: 655657.
13. Ramamoorthy C, Tabbutt S, Kurth CD, et al. Effects of inspired 17% O2 or 3% CO2 on cerebral O2 saturation in neonates with hypoplastic left heart syndrome. Anesth Analg. 2000; 90: 26. Abstract.
14. Watzman HM, Kurth CD, Montenegro LM, et al. Arterial and venous contributions to near-infrared cerebral oxygenation. Anesthesiology. 2000; 93: 947953.[Medline] [Order article via Infotrieve]
15. Schieber RA. Cardiovascular physiology in infants and children.In: Motoyama EK, Davis PJ, eds. Anesthesia for Infants and Children. St. Louis, Mo: CV Mosby Co; 1990: 79.
16. Mora GA, Pizarro C, Jacobs ML, et al. Experimental model of single ventricle, influence of carbon dioxide on pulmonary vascular dynamics. Circulation. 1994; 90 (suppl II): II-43II-46.
17.
Riordan CJ, Randsbaek F, Storey JH, et al. Effects of oxygen, positive end-expiratory pressure, and carbon dioxide on oxygen delivery in an animal model of the univentricular heart. J Thorac Cardiovasc Surg. 1996; 112: 644654.
18.
Reddy VM, Liddicoat JR, Finemann JR, et al. Fetal model of single ventricle physiology: hemodynamic effects of oxygen, nitric oxide, and hypoxia in the early postnatal period. J Thorac Cardiovasc Surg. 1996; 112: 437449.
19.
Barnea O, Santamore WP, Rossi A, et al. Estimation of oxygen delivery in newborns with a univentricular circulation. Circulation. 1998; 98: 14071413.
20. Benumof JL. Respiratory physiology and respiratory function during anesthesia. In: Miller RD, ed. Anesthesia. 3rd ed. New York, NY: Churchill Livingstone; 1990: 544545.
21.
Riordan CJ, Locher JP, Santamore WP, et al. Monitoring systemic venous oxygen saturation in the hypoplastic left heart syndrome. Ann Thorac Surg. 1997; 63: 835837.
22. Rossi AF, Sommer RJ, Lotvin A, et al. Usefulness of intermittent monitoring of mixed venous oxygen saturation after stage 1 palliation for hypoplastic left heart syndrome. Am J Cardiol. 1994; 73: 11181123.[Medline] [Order article via Infotrieve]
23.
Hoffman GM, Ghanayem NS, Kampine JM, et al. Venous saturation and the anaerobic threshold in neonates after the Norwood procedure for hypoplastic left heart syndrome. Ann Thorac Surg. 2000; 70: 15151521.
24. Butheitel G, Scharf J, Hofbeck M, et al. Estimation of cardiac index by means of the arterial and the mixed venous oxygen content and pulmonary oxygen uptake determination in the early post-operative period following surgery of congenital heart disease. Intensive Care Med. 1994; 20: 500503.[Medline] [Order article via Infotrieve]
25. Lock JE. Hemodynamic evaluation of congenital heart disease.In: Lock JE, Keane JF, Fellows KE, eds. Diagnostic and Interventional Catheterization in Congenital Heart Disease. Boston, Mass: Martinus Nijhoff Publishing; 1987: 56.
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2001 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |