(Circulation. 2001;104:1023.)
© 2001 American Heart Association, Inc.
Clinical Investigation and Reports |
From the Mayo Clinic, Rochester, Minn (P.A.F., S.C.H.); the University of Maastricht, Maastricht, the Netherlands (B.D.); and Medtronic, Inc, Minneapolis, Minn (E.N.W., H.A.X., R.M., M.S.S.).
Correspondence to Paul A. Friedman, MD, Department of Internal Medicine and Cardiovascular Diseases, Mayo Clinic, 200 First St SW, Rochester, MN 55905. E-mail pfriedman@ mayo.edu
| Abstract |
|---|
|
|
|---|
Methods and Results Patients with a standard indication for the implantation of an ICD and 2 episodes of atrial tachyarrhythmias in the preceding year received a dual-chamber ICD (Medtronic 7250 Jewel AF) that uses pacing and shock therapies for prevention and/or termination of atrial tachyarrhythmias. In a multicenter trial, patients were randomized to 3-month periods with atrial therapies "on" or "off" and subsequently crossed over. Analysis was performed on the 52 of 269 patients who had episodes of atrial tachyarrhythmia and had
30 days of follow-up with atrial therapies on and off. The atrial therapies resulted in a reduction of atrial tachyarrhythmia burden from a mean of 58.5 to 7.8 h/mo. A paired analysis (Wilcoxon signed-rank test) showed that the median difference in burden (1.1 h/mo) was highly significant (P=0.007). When the subgroup of 41 patients treated only with atrial pacing therapies was analyzed, the reduction in burden persisted (P=0.01).
Conclusions In this study, patients with a standard ICD indication and atrial tachyarrhythmias had a significant reduction in atrial tachyarrhythmia burden with use of atrial pacing and shock therapies.
Key Words: tachyarrhythmias atrium atrial flutter fibrillation defibrillation pacing
| Introduction |
|---|
|
|
|---|
25% at the time of the implantation.1 Furthermore, ICD patients will develop a first episode of paroxysmal AF after the implantation at an annual rate of
2.6%.1 Management of atrial tachyarrhythmias in patients with ICDs is important because of the potential for increased morbidity and mortality and the cost of care. The Antiarrhythmics Versus Implantable Defibrillators trial showed that AF was an independent predictor of mortality in ICD patients,2 and other studies have shown that AF is associated with increased mortality in patients with depressed ventricular function.3 In addition, atrial tachyarrhythmias may be symptomatic and may precipitate symptoms in patients with congestive heart failure.4 Sustained atrial tachyarrhythmias may also increase the need for anticoagulation. Antiarrhythmic drugs have been the mainstay of therapy for the management of patients with AF.5 Drug therapy is often unsatisfactory, however, because of limited efficacy, risk of proarrhythmia, and undesirable side effects, particularly in patients with depressed ventricular function. In patients who must already be exposed to the procedural risk of ICD placement for ventricular arrhythmias, a dual-chamber ICD that offers therapies for both atrial and ventricular tachyarrhythmias may represent a significant advance in arrhythmia management.
A new dual-chamber ICD has been introduced (Medtronic 7250 Jewel AF) that automatically detects and treats episodes of AF, atrial tachycardia (AT), ventricular fibrillation, and ventricular tachycardia and provides antibradycardia pacing. This device also uses separate algorithms for the prevention of atrial tachyarrhythmias. The primary objective of this study was to evaluate the effect of the combination of prevention and termination therapies on the percentage of time an ICD patient spends in atrial tachyarrhythmia (AT/AF burden). A second objective was to determine the effect of these atrial therapies on the frequency of AT/AF episodes.
| Methods |
|---|
|
|
|---|
1 episode. Patients with chronic fibrillation (unable to sustain sinus rhythm) were excluded. All patients gave written informed consent according to a protocol approved by the Human Subjects Committee of the institution at which the devices were implanted. The ICD pulse generator and electrodes were inserted through a single left pectoral incision by standard methods. A 2-lead system was implanted in 84.6% of patients, a 3-lead system in 14.6%, and a 4-lead system in 0.8%. A step-up atrial defibrillation threshold protocol was recommended but not required.
Device Characteristics
A detailed description of the detection and termination algorithms was reported previously.6 The device discriminates AT from AF on the basis of 2 programmable detection zones, which may overlap (Figure 1). If the median atrial cycle length is in the overlap zone, the rhythm is classified as AT if it is regular and AF if it is irregular. Outside of the overlap zone, the rhythm is classified as AF if the median cycle length is in the AF zone and AT if it is in the AT zone. Programming of the AT/AF detection intervals was left to the discretion of the physician.
|
The 2 prevention algorithms and 4 types of termination therapies that are available in the device are listed in Table 1. For AT episodes, the first 3 programmable therapies are pacing and the last 3 are shocks. The first 2 therapies may be programmed to either a burst+ or ramp, but the third therapy can only be programmed to a 50-Hz burSt. For AF episodes, the first programmable therapy is pacing (50-Hz burst) followed by a maximum of 5 shock therapies. For both AT and AF, any of the 6 possible therapies can be programmed to be skipped. There is a programmable delay (0 to 24 hours) between AT/AF detection and the onset of therapy, programmed separately for pacing (nominal 1 minute) and shocks (nominal 30 minutes). The nominal programming of the AT pacing therapies is 6 sequences of ramp (8 pulses), followed by 6 sequences of burst+ (15 pulses), followed by 5 sequences of 50-Hz burst (1 second each). The nominal programming of pacing for AF is 10 sequences of 50-Hz burst (1 second each). Programming of all atrial pacing variables, therapy sequences, and atrial shocks was left to the discretion of the treating physician.
|
If atrial shocks were enabled, it was recommended that the first shock for AF be delivered at twice the defibrillation threshold. All atrial shocks were synchronized to the QRS as sensed by the right ventricular lead and were withheld if the R-R interval preceding the synchronized QRS complex was less than a programmable value (nominally, 500 ms) to avoid shock delivery during ventricular repolarization.
Study Design
Ventricular therapies were enabled in all patients. The effect of atrial therapies on AT/AF burden was assessed by use of a single-blind, randomized, crossover design in which atrial prevention and termination therapies were jointly programmed "on" or "off" for 3 months, and then each patient was crossed over to the opposite arm for an additional 3 months. Patients were followed up with routine device interrogations at 1, 3, and 6 months after implantation, with additional visits as clinically indicated. Changes in antiarrhythmic drugs during the study period were discouraged, and use of anticoagulants was left to the physicians discretion.
Data Analysis
All data beyond the randomization period (ie, the initial 6 months of follow-up) were censored. To be included in the data analysis, a patient was required to have
1 AT/AF episode during the first 6 months of follow-up and
30 days of follow-up in each of the on and off periods. If patients crossed over early or late, then unequal follow-up times were used in the 2 arms of the study. Patients were included only if their devices were programmed to deliver both prevention therapies and
1 termination therapy (antitachycardia pacing, 50 Hz, or a shock) in the on arm, and no prevention and termination therapies in the off arm (ie, programmed as per protocol). A total of 52 patients fulfilled these criteria and are the focus of this report. Rhythms are reported on the basis of device classification. The device-based definition (AT or AF) at the time of first therapy delivery for a given episode was used to classify the episodes, regardless of rhythm transitions (between AT and AF) thereafter. An atrial therapy was classified as successful if 5 consecutive beats of sinus (or atrial paced) rhythm were detected by the device before arrhythmia redetection. If a shock was delivered at any time during an episode, pacing was considered unsuccessful.
Atrial tachyarrhythmia burden was defined as the total duration of all spontaneous AT/AF episodes divided by the corresponding follow-up time. Atrial tachyarrhythmia frequency was defined as the number of spontaneous AT/AF episodes divided by the time the patient was in sinus rhythm (ie, follow-up time minus time in AT/AF). The time in sinus rhythm was used to calculate AT/AF frequency because the prevention algorithms are operative and can reduce the frequency of AT/AF only during sinus rhythm. The follow-up time between 2 consecutive device interrogations was adjusted (reduced) to account for lost episodes. Episodes were lost when the number of episodes exceeded the device memory capabilities (30 treated and 128 untreated). The follow-up time was reduced because of lost episodes in the on period in 1 patient and in the off period in 6 patients. Paired comparisons of burden and frequency between the off and on periods were performed with a Wilcoxon signed-rank test because the data did not follow a normal distribution. An analysis was performed to determine whether there was a significant order effect, ie, treating patients in an on-then-off sequence versus an off-then-on sequence. Because no such effect was seen, data from the 2 randomized groups were combined.
| Results |
|---|
|
|
|---|
|
Of the 158 patients with AF/AT episodes during the course of the randomized study, 52 were included in this analysis. Reasons for exclusion included lack of crossover (n=47) or <30 days of follow-up in each arm (n=7) and nonprotocol programming of atrial therapies (eg, atrial prevention on with therapies off or other deviation, n= 52).
For the 52 patients with episodes who were properly followed up and programmed, the mean durations of the on and off arms were 98±22 and 89±30 days, respectively. In these 52 patients, the programming for the on period included
1 atrial pacing therapy, with a mean time to initiation of 4 minutes in 44 patients (median, 1 minute; range, 0 to 60 minutes), and
1 atrial shock therapy, with a mean time to shock of 104 minutes in 32 patients (median, 30 minutes; range, 1 to 1440 minutes). Atrial tachyarrhythmia burden was significantly reduced with the use of prevention and termination therapies. Individual patient results are shown in Figure 2A. During the off period, the mean burden was 58.5 h/mo (median, 2.82 h/mo; range, 0 to 470 h/mo), versus 7.8 h/mo (median, 0.63 h/mo; range, 0 to 66 h/mo) with therapies on. This reflects an 87% reduction in mean burden. The median off-on difference was 1.1 h/mo (P=0.007).
|
Review of the individual patient responses in Figure 2 shows that there were 10 major responders (burden reduction >100 h/mo), who accounted for the majority of the mean burden reduction. All 10 major responders had
1 episode longer than 48 hours during the off period versus only 2 of the remaining 42 patients (P= 0.001). Responders were also found to have higher ejection fractions (46±18% versus 31±14%, P=0.02). No differences were found in the other patient baseline characteristics, use of antiarrhythmia medications, pacing success rate, AT/AF ratio, and average atrial arrhythmia cycle length.
There was a trend toward a reduction in the number of AT/AF episodes per month of sinus rhythm. The mean frequency was 21.6/mo (median, 1.35/mo; range, 0 to 176/mo) with therapies off, versus 6.9/mo (median, 1.27/mo; range, 0 to 52/mo) with therapies on (P=0.07). Individual patient results are shown in Figure 2B. There was no correlation between changes in burden and changes in frequency (Figure 3). Patients with few long episodes had the largest burden reduction, whereas patients with many short episodes had the largest frequency reduction. Consequently, most of the patients with large burden reduction had small changes in frequency, and most of the patients with large frequency reduction had small changes in burden. There were 332 AT/AF episodes that were treated in 34 patients. In the 211 episodes treated as AT (64%), pacing therapies successfully terminated 62%. In the 121 episodes treated as AF (36%), pacing therapies successfully terminated 24%. The combined pacing success rate for all AT/AF episodes was 49%. The average number of pacing sequences delivered for successful termination was 4.2 with ramp and/or burst+ combined and 6.6 with ramp, burst+, and/or 50 Hz combined. Shock therapies successfully terminated 18 of 23 AT/AF episodes (78%). None of the atrial shock therapies resulted in ventricular proarrhythmia. In addition, 21 shock therapies were delivered in 4 patients because of inappropriate ventricular tachycardia/ventricular fibrillation detection.
|
A subanalysis of burden and frequency was performed in 41 patients in whom no shocks were delivered (ie, only pacing therapies used). The AT/AF burden reduction persisted in this subgroup treated only with pacing therapies (Figure 4A). During the period with therapies off, the mean AT/AF burden was 53.6 h/mo (median, 2.7 h/mo; range, 0 to 470 h/mo) versus 6.2 h/mo (median, 0.1 h/mo; range, 0 to 66 h/mo) with therapies on. This reflects an 89% reduction in mean burden. The median off-on difference in burden was 0.54 h/mo (P=0.01). The corresponding number of AT/AF episodes per month of sinus rhythm is shown in Figure 4B. The mean number of episodes was 22.3/mo (median, 1.1/mo; range, 0 to 176/mo) with therapies off versus 6.5/mo (median, 1.0/mo; range, 0 to 52/mo) with therapies on (P=0.16).
|
| Discussion |
|---|
|
|
|---|
Pacing Therapies for the Termination of Atrial Tachyarrhythmias
The patient discomfort associated with defibrillation has spurred interest in the development of less painful methods to terminate AF. Antitachycardia pacing has been highly effective in isthmus-dependent (typical) atrial flutter, which is a highly organized rhythm with a large excitable gap.9,10 Rate-adaptive antitachycardia pacing has not been effective in disorganized rhythms, such as atypical flutter or AF.11 The demonstration that AF has an excitable gap and that high-rate overdrive pacing can capture up to a 4-cm region of the atrium in a fibrillating animal model, however, has suggested a rationale for the use of this form of pacing for AF termination.12
Clinical trials of high-frequency (50-Hz) burst pacing in short cycle length or long-standing AF have been disappointing.13 In atypical flutter, however, which may be a precursor to AF, the results have been more promising, with short-term success rates of up to 65%.11 In the present study, although the documented clinical arrhythmia was AF in 81% of the patients, 64% of spontaneous episodes were initially treated as an AT, suggesting that many episodes of clinical fibrillation begin with a more organized rhythm that subsequently degenerates. The presence of an indwelling device capable of prompt arrhythmia detection may permit termination with pacing techniques during the early, more vulnerable phase of arrhythmogenesis.
Presumably, the use of pacing therapies reduced the need for shocks, because pacing therapies were effective in terminating 49% of all AT/AF episodes. Because the programming of shocks was left to the discretion of the physicians, only 11 of the 52 patients included in the analysis received shocks. Of the 41 patients who received only pacing therapies, the median reduction in burden with therapies on was 0.54 h/mo (P=0.01). This is an important finding and suggests that a device with pacing and shock therapies may have much more clinical utility than a shock-only device.
Role of an Atrial Defibrillator in Patients With Structural Heart Disease
Previous experience with implantable atrial defibrillators has been limited to patients with structurally normal hearts.14 The role for this therapy may be particularly compelling, however, in patients with structural heart disease. Patients with depressed ventricular function and AF have an increased mortality compared with those with normal sinus rhythm independent of the ventricular heart rate, raising the possibility that atrial rhythm control may introduce a mortality benefit.3 In addition, patients with an indication for a ventricular defibrillator must be exposed to the risk of device implantation, so that a device capable of treating both atrial and ventricular arrhythmias may add therapeutic benefit with little or no additional risk. Although maintenance of sinus rhythm is more difficult in the setting of cardiac structural abnormalities,15 we found that atrial tachyarrhythmia burden was significantly reduced in this population. Reduction in AT/AF burden may improve quality of life, reduce tachycardia-induced cardiomyopathy, and/or reduce the number of hospitalizations. The clinical benefits of lowering the AT/AF burden in these patients were not evaluated in this study, however, and will need to be investigated in a prospective trial.
Limitations
Patients with no AT/AF episodes during the 6 months of follow-up (n=111) or with episodes whose device programming was not in compliance with the study protocol (n=106) were excluded from this analysis. We felt justified in excluding patients without episodes, because any antiarrhythmic therapy would be ineffective if it were delivered to patients without arrhythmias. The lack of adherence to the study protocol may have biased the patients inclusion and thus affected study outcome. The study duration covered only a period early after device implantation.
Use of antiarrhythmic drugs (amiodarone, sotalol, or class I drugs) was stable in 69% of the 52 patients during the 6 months of follow-up. Antiarrhythmic drug changes may have modified AT/AF burden and/or frequency in some patients.
Because both prevention and termination therapies were used and because the type of atrial pacing was selected by the physician, we were unable to determine the relative impact of each of these elements independently on burden reduction.
Conclusions
The use of pacing and shock therapies for atrial tachyarrhythmia prevention and termination reduced arrhythmia burden in this study. Atrial arrhythmia burden is reduced because of the presence of shorter and possibly fewer episodes. These therapies may be useful in managing patients with a ventricular ICD indication and atrial tachyarrhythmias if the clinical value of AT/AF burden reduction can be demonstrated.
| Footnotes |
|---|
Received March 9, 2001; revision received June 22, 2001; accepted June 22, 2001.
| References |
|---|
|
|
|---|
2.
The Antiarrhythmics Versus Implantable Defibrillators (AVID) Investigators. A comparison of antiarrhythmic drug therapy with implantable defibrillators in patients resuscitated from near-fatal ventricular arrhythmias. N Engl J Med. 1997; 337: 15761584.
3.
Dries DL, Exner DV, Gersh BJ, et al. Atrial fibrillation is associated with an increased risk for mortality and heart failure progression in patients with asymptomatic and symptomatic left ventricular systolic dysfunction: a retrospective analysis of the SOLVD trialsStudies of Left Ventricular Dysfunction. J Am Coll Cardiol. 1998; 32: 695703.
4. Savelieva I, Camm A. Clinical relevance of silent atrial fibrillation: prevalence, prognosis, quality of life, and management. J Interv Card Electrophysiol. 2000; 4: 369382.[Medline] [Order article via Infotrieve]
5. Prystowsky E, Benson W, Fuster V, et al. Management of patients with atrial fibrillation: a statement for healthcare professionals from the subcommittee on electrocardiography and electrophysiology, American Heart Association. Circulation. 1996; 92: 12621277.
6.
Swerdlow C, Schoels W, Dijkman B, et al. Detection of atrial fibrillation and flutter by a dual-chamber implantable cardioverter-defibrillator. Circulation. 2000; 101: 878885.
7.
Clair W, Wilkinson W, McCarthy E, et al. Spontaneous occurrence of symptomatic paroxysmal atrial fibrillation and paroxysmal supraventricular tachycardia in untreated patients. Circulation. 1993; 87: 11141122.
8.
Rodriguez LM, Timmermans C, Wellens HJ. Are electrophysiological changes induced by longer lasting atrial fibrillation reversible? Observations using the atrial defibrillator. Circulation. 1999; 100: 113116.
9.
Friedman PA, Luria D, Fenton AM, et al. Global right atrial mapping of human atrial flutter: the presence of posteromedial (sinus venosa region) functional block and double potentials: a study in biplane fluoroscopy and intracardiac echocardiography. Circulation. 2000; 101: 15681577.
10. Waldo AL, Mackall JA, Biblo LA. Mechanisms and medical management of patients with atrial flutter. Cardiol Clin. 1997; 15: 661676.[Medline] [Order article via Infotrieve]
11. Giorgberidze I, Saksena S, Mongeon L, et al. Effects of high-frequency atrial pacing in atypical atrial flutter and atrial fibrillation. J Interv Card Electrophysiol. 1997; 1: 111123.[Medline] [Order article via Infotrieve]
12.
Kirchhof C, Chorro F, Scheffer G, et al. Regional entrainment of atrial fibrillation studied by high-resolution mapping in open-chest dogs. Circulation. 1993; 88: 736749.
13. Paladino W, Bahu M, Knight BP, et al. Failure of single- and multisite high-frequency atrial pacing to terminate atrial fibrillation. Am J Cardiol. 1997; 80: 226227.[Medline] [Order article via Infotrieve]
14.
Wellens HJ, Lau CP, Luderitz B, et al. Atrioverter: an implantable device for the treatment of atrial fibrillation. Circulation. 1998; 98: 16511656.
15. Arnar D, Danielsen R. Factors predicting maintenance of sinus rhythm after direct current cardioversion of atrial fibrillation and flutter: a reanalysis with recently acquired data. Cardiology. 1996; 87: 181188.[Medline] [Order article via Infotrieve]
This article has been cited by other articles:
![]() |
A. M. Gillis, M. Morck, D. V. Exner, R. S. Sheldon, H. J. Duff, B. L. Mitchell, and G. D. Wyse Impact of atrial antitachycardia pacing and atrial pace prevention therapies on atrial fibrillation burden over long-term follow-up Europace, August 1, 2009; 11(8): 1041 - 1047. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. P. Ricci, A. Quesada, J. Almendral, F. Arribas, C. Wolpert, P. Adragao, M. Zoni-Berisso, X. Navarro, T. DeSanto, A. Grammatico, et al. Dual-chamber implantable cardioverter defibrillators reduce clinical adverse events related to atrial fibrillation when compared with single-chamber defibrillators: a subanalysis of the DATAS trial Europace, May 1, 2009; 11(5): 587 - 593. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. J. Camm, P. Kirchhof, G. Y.H. Lip, I. Savelieva, and S. Ernst CHAPTER 29 Atrial Fibrillation ESC Textbook of Cardiovascular Medicine, January 1, 2009; 2(1): med-9780199566990-chapter - med-9780199566990-chapter. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. E. Epstein, J. P. DiMarco, K. A. Ellenbogen, N.A. M. Estes III, R. A. Freedman, L. S. Gettes, A. M. Gillinov, G. Gregoratos, S. C. Hammill, D. L. Hayes, et al. ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices) Developed in Collaboration With the American Association for Thoracic Surgery and Society of Thoracic Surgeons J. Am. Coll. Cardiol., May 27, 2008; 51(21): e1 - e62. [Full Text] [PDF] |
||||
![]() |
Writing Committee Members, A. E. Epstein, J. P. DiMarco, K. A. Ellenbogen, N.A. M. Estes III, R. A. Freedman, L. S. Gettes, A. M. Gillinov, G. Gregoratos, S. C. Hammill, et al. ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices): Developed in Collaboration With the American Association for Thoracic Surgery and Society of Thoracic Surgeons Circulation, May 27, 2008; 117(21): e350 - e408. [Full Text] [PDF] |
||||
![]() |
M. E.W. Hemels, J. H. Ruiter, G. P. Molhoek, N. J.G.M. Veeger, A. C.P. Wiesfeld, A. V. Ranchor, M. van Trigt, A. Pilmeyer, I. C. Van Gelder, and for The Features in AT500TM study; Chances for pat Right atrial preventive and antitachycardia pacing for prevention of paroxysmal atrial fibrillation in patients without bradycardia: a randomized study Europace, March 1, 2008; 10(3): 306 - 313. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Mont, R. Ruiz-Granell, J. G. Martinez, J. R. Carmona, M. Fidalgo, E. Cobo, M. Riera, X. Navarro, and on behalf of the Prevention or Termination (POT) S Impact of anti-tachycardia pacing on atrial fibrillation burden when added on top of preventive pacing algorithms: results of the prevention or termination (POT) trial Europace, January 1, 2008; 10(1): 28 - 34. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Gradaus, K. Seidl, T. Korte, E. Himmrich, H. Wieneke, A. Schuchert, W. Bauer, J. Gerss, C. G. Wollmann, M. Borggrefe, et al. Reduction of ventricular tachyarrhythmia by treatment of atrial fibrillation in ICD patients with dual-chamber implantable cardioverter/defibrillators capable of atrial therapy delivery: the REVERT-AF Study Europace, July 1, 2007; 9(7): 534 - 539. [Abstract] [Full Text] [PDF] |
||||
![]() |
Developed in Collaboration With the European Heart, D. P. Zipes, A. J. Camm, M. Borggrefe, A. E. Buxton, B. Chaitman, M. Fromer, G. Gregoratos, G. Klein, A. J. Moss, et al. ACC/AHA/ESC 2006 Guidelines for Management of Patients With Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death: A Report of the American College of Cardiology/American Heart Association Task Force and the European Society of Cardiology Committee for Practice Guidelines (Writing Committee to Develop Guidelines for Management of Patients With Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death) J. Am. Coll. Cardiol., September 5, 2006; 48(5): e247 - e346. [Full Text] [PDF] |
||||
![]() |
Writing Committee Members, D. P. Zipes, A. J. Camm, M. Borggrefe, A. E. Buxton, B. Chaitman, M. Fromer, G. Gregoratos, G. Klein, A. J. Moss, et al. ACC/AHA/ESC 2006 guidelines for management of patients with ventricular arrhythmias and the prevention of sudden cardiac death: A report of the American College of Cardiology/American Heart Association Task Force and the European Society of Cardiology Committee for Practice Guidelines (Writing Committee to Develop Guidelines for Management of Patients With Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death) Developed in collaboration with the European Heart Rhythm Association and the Heart Rhythm Society Europace, September 1, 2006; 8(9): 746 - 837. [Full Text] [PDF] |
||||
![]() |
M. Gulizia, S. Mangiameli, S. Orazi, G. Chiaranda, G. Boriani, G. Piccione, N. DiGiovanni, A. Colletti, C. Puntrello, G. Butera, et al. Randomized comparison between Ramp and Burst+ atrial antitachycardia pacing therapies in patients suffering from sinus node disease and atrial fibrillation and implanted with a DDDRP device Europace, July 1, 2006; 8(7): 465 - 473. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Drago, M. S. Silvetti, G. Grutter, and A. De Santis Long term management of atrial arrhythmias in young patients with sick sinus syndrome undergoing early operation to correct congenital heart disease Europace, July 1, 2006; 8(7): 488 - 494. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Schwartzman, M. N. Harvey, R. H. Hoyt, J. L. Koehler, M. R. Ujhelyi, and D. E. Euler Utility of adjunctive single oral bolus propafenone therapy in patients with atrial defibrillators. Europace, March 1, 2006; 8(3): 211 - 215. [Abstract] [Full Text] [PDF] |
||||
![]() |
Z. Goldberger and R. Lampert Implantable Cardioverter-Defibrillators: Expanding Indications and Technologies JAMA, February 15, 2006; 295(7): 809 - 818. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. E.W. Hemels, A. C.P. Wiesfeld, B. Inberg, P. F.H.M. Van Dessel, W. Nieuwland, E. S. Tan, H. Mulder, D. J. Van Veldhuisen, and I. C. Van Gelder Right atrial overdrive pacing for prevention of symptomatic refractory atrial fibrillation. Europace, February 1, 2006; 8(2): 107 - 112. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. P. Knight, B. J. Gersh, M. D. Carlson, P. A. Friedman, R. L. McNamara, S. A. Strickberger, H. F. Tse, A. L. Waldo, and for the AHA Writing Group Role of Permanent Pacing to Prevent Atrial Fibrillation: Science Advisory From the American Heart Association Council on Clinical Cardiology (Subcommittee on Electrocardiography and Arrhythmias) and the Quality of Care and Outcomes Research Interdisciplinary Working Group, in Collaboration With the Heart Rhythm Society Circulation, January 18, 2005; 111(2): 240 - 243. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Padeletti, M. Santini, G. Boriani, G. Botto, A. Capucci, M. Gulizia, R. Ricci, A. Spampinato, P. Pieragnoli, E. Warman, et al. Temporal variability of atrial tachyarrhythmia burden in bradycardia-tachycardia syndrome patients Eur. Heart J., January 2, 2005; 26(2): 165 - 172. [Abstract] [Full Text] [PDF] |
||||
![]() |
J.C. Geller, S. Reek, C. Timmermans, T. Kayser, H.-F. Tse, C. Wolpert, W. Jung, A.J. Camm, C.-P. Lau, H. J.J. Wellens, et al. Treatment of atrial fibrillation with an implantable atrial defibrillator -- long term results Eur. Heart J., December 1, 2003; 24(23): 2083 - 2089. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. A. Lee, R. Weachter, S. Pollak, M. S. Kremers, A. M. Naik, R. Silverman, J. Tuzi, W. Wang, L. J. Johnson, D. E. Euler, et al. The effect of atrial pacing therapies on atrial tachyarrhythmia burden and frequency: Results of a randomized trial in patients with bradycardia and atrial tachyarrhythmias J. Am. Coll. Cardiol., June 4, 2003; 41(11): 1926 - 1932. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. L. Page, T. W. Tilsch, S. J. Connolly, D. J. Schnell, S. R. Marcello, W. E. Wilkinson, E. L.C. Pritchett, and for the Azimilide Supraventricular Arrhythmia Prog Asymptomatic or "Silent" Atrial Fibrillation: Frequency in Untreated Patients and Patients Receiving Azimilide Circulation, March 4, 2003; 107(8): 1141 - 1145. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. F. Lozano, A. Vincent, J. Roda, M. Méndez, J. M. M. Ferrer, F. Andrade, J. J. Manzano, R. Ceres, J. Errejon, and J. Toquero Paroxysmal atrial fibrillation prevention by pacing in patients with pacemaker indication Europace, January 1, 2003; 5(3): 267 - 273. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Santini and R. Ricci Atrial fibrillation and heart failure Europace, January 1, 2003; 5(s1): S55 - S59. [Full Text] [PDF] |
||||
![]() |
C.-P. Lau Pacing for atrial fibrillation Heart, January 1, 2003; 89(1): 106 - 112. [Full Text] [PDF] |
||||
![]() |
T. H. Everett IV, J. G. Akar, L.-C. Kok, J. R. Moorman, and D. E. Haines Use of global atrial fibrillation organization to optimize the success of burst pace termination J. Am. Coll. Cardiol., November 20, 2002; 40(10): 1831 - 1840. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. M. Gillis, C. Unterberg-Buchwald, H. Schmidinger, S. Massimo, K. Wolfe, D. J. Kavaney, M. F. Otterness, S. H. Hohnloser, and GEM III AT Worldwide Investigators Safety and efficacy of advanced atrial pacing therapies for atrial tachyarrhythmias in patients with a new implantable dual chamber cardioverter-defibrillator J. Am. Coll. Cardiol., November 6, 2002; 40(9): 1653 - 1659. [Abstract] [Full Text] [PDF] |
||||
![]() |
M.J.P. Raatikainen and H.V. Huikuri Management of atrial fibrillation in patients with implantable cardioverter defibrillator. Do all need a dual chamber device? Eur. Heart J., September 2, 2002; 23(18): 1412 - 1414. [Full Text] [PDF] |
||||
![]() |
K. M. Stein, D. E. Euler, R. Mehra, K. Seidl, D. J. Slotwiner, S. Mittal, S. M. Markowitz, B. B. Lerman, and Jewel AF Worldwide Investigators Do atrial tachyarrhythmias beget ventricular tachyarrhythmias in defibrillator recipients? J. Am. Coll. Cardiol., July 17, 2002; 40(2): 335 - 340. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. M. Cooper, M. S. Katcher, and M. V. Orlov Implantable Devices for the Treatment of Atrial Fibrillation N. Engl. J. Med., June 27, 2002; 346(26): 2062 - 2068. [Full Text] [PDF] |
||||
![]() |
S.R. Raj and D.G. Wyse Rhythm control for paroxysmal atrial fibrillation after AV junction ablation: is it worth the effort? Eur. Heart J., June 1, 2002; 23(11): 845 - 848. [Full Text] [PDF] |
||||
![]() |
J. C. Daubert and P. Mabo Implantable devices to treat atrial fibrillation: real prospects or just new gimmicks? Europace, January 1, 2002; 4(2): 161 - 164. [PDF] |
||||
![]() |
W. Xu, H.-F. Tse, F. H.Y. Chan, P. C. W. Fung, K. L.-F. Lee, and C.-P. Lau New Bayesian Discriminator for Detection of Atrial Tachyarrhythmias Circulation, March 26, 2002; 105(12): 1472 - 1479. [Abstract] [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2001 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |