(Circulation. 2001;104:908.)
© 2001 American Heart Association, Inc.
Clinical Investigation and Reports |
From the Childrens Hospital/University of Colorado Health Sciences Center, Denver, Co (R.S., D.D.I., E.N., C.G.D., L.V.-C.), and the Department of Mechanical Engineering, University of Colorado, Boulder, Co (R.S., C.W., J.H.).
Correspondence to Robin Shandas, PhD, Cardiovascular Flow Research Laboratory, The Childrens Hospital, 1056 E. 19th Ave, B-100, Denver, CO 80219. E-mail shandas.robin{at}tchden.org
| Abstract |
|---|
|
|
|---|
Methods and Results We tested the hypothesis using mathematical modeling, in vitro experiments, and preliminary clinical studies. The mathematical model showed that pressure and velocity tracings are closely correlated in time and that 6 to 18 ms time resolution was needed to resolve propagation times within typical main pulmonary artery lengths (2 to 5 cm). The in vitro experiments demonstrated that it was feasible to use color M-mode to measure Velprop and that Velprop correlated well with downstream resistance [y=(-1.01x)+22.77; R=0.96]. The method was then evaluated on patients undergoing acute pulmonary reactivity testing (n=22 measurements). Good correlation between Velprop and PVR was found [y=(-1.71x)+26.0; R=0.90; SEE=2.41].
Conclusion This newly developed method promises to be useful in the noninvasive evaluation of adults and children with pulmonary hypertension.
Key Words: hypertension, pulmonary echocardiography heart defects, congenital
| Introduction |
|---|
|
|
|---|
Several noninvasive techniques using echocardiography and ultrasound Doppler have been proposed to evaluate pulmonary hypertension. These have focused on examining anatomy using echocardiography and estimating PA pressure by continuous-wave Doppler measurement of the tricuspid regurgitant jet velocity.610 Neither measurement has shown correlation to actual PVR, especially when PVR changes short-term. The use of Doppler methods to calculate pulmonary blood flow is also limited due to assumptions of a nondistendible arterial cross-section and a flat velocity profile.11,12
We hypothesized that short-term changes in PVR would cause concomitant changes in the speed of the pressure pulse propagation wave within the main PA and that, due to the close association between pressure and velocity fields, these changes would be reflected by changes in the centerline velocity trace along the PA. An estimate of PVR should therefore be possible by measuring the velocity propagation (Velprop) slope using ultrasound color M-mode imaging of flow along the centerline of the main PA. This hypothesis was tested using mathematical modeling, in vitro simulation, and clinically in the catheterization laboratory for 11 patients (total of 22 measurements) in whom PVR was altered in the short-term.
| Methods |
|---|
|
|
|---|
Routinely measuring pressure pulse propagation speed in the clinical situation is not realistic because this involves either mounting strain gauges on the arterial wall or invasive pressure measurements.14,15,17 However, changes in local pressure (and so local pressure gradient) due to the changes in the pulse propagation wave speed should also affect local velocities within the artery due to the close association between velocity and pressure fields.17,19,20 By extension, pressure pulse propagation should also cause an equivalent Velprop.
Velprop can be calculated from a color M-mode trace of flow down the centerline of the PA by tracking the location of fluid moving at one particular velocity. By manipulating the aliasing velocity on the color M-mode trace, the location of this isovelocity as a function of time and space can be visually tracked as the point of color change. Such a method should provide an easy yet quantitative means of estimating Velprop in the clinical situation.
Mathematical Analysis
The goals of the mathematical analysis were to (1) study the time-correlation between pressure and velocity traces within the tube and (2) quantify typical time lags seen in the pressure propagation wave as it travels down the artery. We focused this model on the main PA segment. The motion of the blood can be governed by the quasi one-dimensional unsteady equations of motion for incompressible and viscous fluid.21 Tube length was 16 cm; however, only a central section 5 cm in length was used for the numerical study. This was done to minimize end-effects while maintaining a length that resembled the in vitro model and was a reasonable approximation of the main PA. The governing equations of a deformable tube of varying cross-section may be expressed as the continuity equation:
|
|
|
|
|
|
In these equations, S indicates cross-sectional area; v, blood velocity averaged over the local cross-section; p, locally-averaged transmural pressure;
, blood density; t, time; x, axial coordinate;
, differential operator; µ, viscosity; and F, a viscous resistance factor.
We can assume a linear relationship between cross-sectional area and transmural pressure of all large vessels22 (diameter >1000 µm), and this relationship can be expressed as follows:
|
|
In this equation, S0 indicates the relaxed (diastolic) cross-sectional area; S, the cross-sectional area;
, the average compliance of the tube segment; and p, locally-averaged transmural pressure. The above equations were combined and solved using the MacCormack predictor-corrector method. Further details on this method are given elsewhere.23
In Vitro Studies
The in vitro studies were designed to (1) test the color M-mode technique as a means of calculating Velprop and (2) correlate Velprop to downstream PVR. As Figure 1 shows, the in vitro system consisted of a pulsatile pump (Harvard Apparatus) that was used to deliver pulsatile flow (33% glycerin:water mixture; viscosity, 3.5cP; density, 1.04 g/mL) at variable frequencies (60 to 100 bpm) and stroke volumes (20 to 100 cc/beat) into a compliant test section. Cornstarch (0.5% by weight) was also added for ultrasound reflectivity. The compliant test section, which served as the model PA, was custom made from silicone rubber using a dip process whereby thin layers (0.2 mm) were added to the tube. Tube diameter was 1.9 cm, and pulse pressure varied from 10 to 40 mm Hg. Total tube length was 16 cm but, as in the numerical studies, measurements were taken within a central section 5 cm in length. Instantaneous flow rate (Transonics Inc) and pressure (Millar Inc) were monitored. Downstream resistance varied from 0.5 to 16 Wood Units to simulate PVR changes.
|
Color M-Mode Studies
Ultrasound color M-mode Doppler recordings of flow within the model PA were performed using a VingMed System 800 (GE Medical Systems). The transducer was placed proximal to the PA, and the ultrasound beam was aimed parallel to the flow direction. Two-dimensional color Doppler flow imaging was used to align the color M-mode cursor directly along the centerline of the tube. Color M-mode tracings (10 beats) were obtained at various Nyquist limits (0.25 to 0.75 m/s) and digitally transferred to an analysis computer (EchoDisp, VingMed Sound Inc) for off-line analysis.
The Velprop calculation method involved tracking the location of fluid moving at one particular velocity as a function of time and distance down the PA. Such a measurement was facilitated by adjusting the aliasing velocity of the ultrasound system, because a clear color change occurred at the point where local velocity exceeded the aliasing limit. Baseline shifting was used on the off-line images to change the aliasing velocity until a clear delineation of isovelocity slope was seen. The slope of Velprop was then calculated from the slope using the following equation:
|
|
In this equation, d2 and d1 are the distance points corresponding to the beginning and end of the propagation slope, and t2 and t1 are the corresponding time points. Thus, Velprop has units of velocity (distance/time) and corresponds to the velocity at which the flow is propagating down the artery. The above method was used to calculate Velprop for both in vitro and clinical studies.
Preliminary Clinical Studies
All clinical studies were conducted with patient consent and institutional review board approval. During cardiac catheterization, color M-mode imaging was performed through a parasternal short-axis approach using the VingMed 800 system. Two-dimensional echo and color Doppler were used to align the color M-mode cursor parallel to the main flow direction within the PA. Data were obtained during an evaluation of reactivity under conditions of room air, O2, inhaled nitric oxide (20 ppm via pulsed nasal or mask delivery), or diltiazem infusion. Inhaled NO was given as previously described.3 Cardiac output was measured using the method of Fick while on room air and assumed oxygen consumption24 or by thermodilution in the absence of intra-cardiac shunts.3 Main PA pressure was measured using standard fluid-filled catheters. Pulmonary venous pressures were obtained using pulmonary wedge pressure. All hemodynamic data were recorded before and during treatment to calculate PVR.
Statistical Analysis
All values are given as mean±SD. Linear regression was used to examine whether Velprop correlated with measured PVR for both in vitro and clinical data. Intraobserver variability was calculated by comparing the differences in 5 consecutive measurements of Velprop performed by the main observer. A subset of the measurements (25%) was repeated by a second blinded observer to calculate interobserver variability. A value of 5% was used to designate statistical significance.
| Results |
|---|
|
|
|---|
|
In Vitro Studies
Figure 3 shows typical in vitro color M-mode tracings for PVR values of 4.08, 8.83, and 16.40 Wood Units. The aliasing velocity was baseline-shifted to produce a clear velocity propagation slope, which was measured to obtain Velprop. As PVR increases, the slope of the alias line decreases from 20.27 cm/s for PVR of 4.08 to 9.21 cm/s for a PVR of 16.40. Figure 4 shows color M-modemeasured Velprop versus PVR (Wood Units) for the in vitro experiments; these measurements showed excellent correlation, indicating that Velprop can be used to follow PVR.
|
|
Clinical Studies
A total of 11 patients were studied. Mean age was 7.0±6.3 years. In some cases, the patient underwent multiple challenges to evaluate pulmonary vascular reactivity. Each challenge produced a distinct PVR value resulting in a total of 22 separate conditions. The Table summarizes the clinical data.
|
Figure 5 shows color M-mode traces for a patient with pulmonary hypertension under normal (room air) conditions who was then treated with 100% O2, which caused a short-term decrease in PVR. The slope of the color M-mode alias increases perceptibly from the high to low PVR condition. Measuring Velprop shows a corresponding increase from 8.8 cm/s at a PVR of 8.4 Wood Units to 19.4 when PVR was decreased to 5.4 Wood Units.
|
Figure 6 shows Velprop versus PVR for all preliminary clinical data. Good correlation between the 2 variables was seen, again indicating the value of Velprop in following actual PVR. Intraobserver and interobserver variabilities were 8.2% and 9.3%, respectively. To assess the ability of this method to distinguish between high resistance and high pressure, Velprop was also regressed against mean PA pressure for patients with high right side pressure (mean PA pressure >45 mm Hg; n=7). In this limited group of patients, we found that Velprop correlates better to PVR [y=(-1.55x)+24.0; R=0.87] than to pressure [y=(-0.58x)+39.93; R=0.73].
|
| Discussion |
|---|
|
|
|---|
The quantitative assessment of PVR continues to be important in the evaluation and treatment of children with pulmonary hypertension. Hemodynamic measurements during cardiac catheterization, quantitative pulmonary wedge angiograms, and lung biopsies have been proposed as a progressive approach to determine surgical risk and outcome.2528 Findings from cardiac catheterization, including right atrial pressure and PVR, correlate with survival and outcome in primary pulmonary hypertension. However, invasive procedures such as catheterization to measure pressure and flow have been associated with increased patient risk.
It is therefore no surprise that echocardiographers continue to search for accurate, noninvasive means of quantifying PVR. Conventional echocardiography only provides structural information and so is limited to evaluating acute hemodynamic changes. Ultrasound Doppler techniques have fallen into either estimation of PA flow acceleration time or measuring the tricuspid regurgitant jet velocity and using the Bernoulli equation to estimate PA pressure.7,9 However, both methods have significant problems, including variation due to other factors such as heart rate, ventricular function, right ventricular preload, and afterload and assumptions inherent in the use of the Bernoulli equation. More importantly, Doppler velocity measurements provide only a measure of pulmonary pressures and not PVR.
Analogy of Proposed Method to Noninvasive Quantitation of Diastolic Function
The closest analogy to the clinical method proposed here is the noninvasive evaluation of diastolic dysfunction using ultrasound color M-mode Doppler interrogation of transmitral flow. It is useful to examine briefly the parallels in the underlying physics between these 2 techniques. In the presence of relaxation abnormalities and/or restrictive physiology, the "stiffer" ventricle is unable to relax actively in response to filling demands. This is analogous to a fluid system wherein the downstream chamber becomes stiffer and thus resists the flow to a higher degree. The flow propagation speed within the ventricle decreases under such conditions.2932 A similar decrease in flow propagation with increased downstream resistance was also found in our studies. This is not surprising because the increased right-sided afterload acts in a similar manner to decrease the flow propagation speed within the larger PAs.
Limitations
It is important to recognize several limitations in our study. The mathematical model was only a one-dimensional model representing centerline hemodynamics, and results from this model should be viewed as the theoretical ideal. We did not evaluate other variables such as cardiac output, tube compliance, or transient flow changes. These are obviously important clinical considerations and will be examined in upcoming studies. The clinical studies were intended to prove initial feasibility only, and further stratification of the data were not performed due to the small numbers present.
Clinical Application of the Method
This method may be an important additional tool in the overall evaluation of pediatric pulmonary hypertension, especially for patients awaiting surgery in whom PVR may predict improved long-term surgical outcomes.2 The technique may also provide information to complement that obtained from other novel methods currently being evaluated to assess right heart function, such as MRI and SPECT equilibrium radionuclide angiography. The method may also be useful when evaluating pulmonary reactivity, because a decrease in PVR should produce a corresponding increase in Velprop. On the basis of the preliminary results, a Velprop value >18 cm/s seems to indicate a PVR<6 Wood Units · m2. The Velprop technique may thereby reduce the number of catheterizations required to measure PVR. Further collection of clinical data to examine the utility of the method when both pressure and output are high but PVR is low, such as in left-to-right shunt lesions in a young patient, is needed.
| Acknowledgments |
|---|
Received December 6, 2000; revision received June 7, 2001; accepted June 21, 2001.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
C. E. Weinberg, J. R. Hertzberg, D. D. Ivy, K. S. Kirby, K. C. Chan, L. Valdes-Cruz, and R. Shandas Extraction of Pulmonary Vascular Compliance, Pulmonary Vascular Resistance, and Right Ventricular Work From Single-Pressure and Doppler Flow Measurements in Children With Pulmonary Hypertension: a New Method for Evaluating Reactivity: In Vitro and Clinical Studies Circulation, October 26, 2004; 110(17): 2609 - 2617. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Mukerjee, D. St. George, C. Knight, J. Davar, A. U. Wells, R. M. Du Bois, C. M. Black, and J. G. Coghlan Echocardiography and pulmonary function as screening tests for pulmonary arterial hypertension in systemic sclerosis Rheumatology, April 1, 2004; 43(4): 461 - 466. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. E. Abbas, F. D. Fortuin, N. B. Schiller, C. P. Appleton, C. A. Moreno, and S. J. Lester A simple method for noninvasive estimation of pulmonary vascular resistance J. Am. Coll. Cardiol., March 19, 2003; 41(6): 1021 - 1027. [Abstract] [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2001 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |