| Abstract |
|---|
|
|
|---|
Methods and Results The crews of 4 mobile intensive care units, staffed by an emergency physician and a paramedic, were trained to use the device. Fifty patients were randomized by sealed envelopes to either Lifestick (n=24) or conventional (n=26) resuscitation. No differences were found regarding demographic and logistical conditions between the groups. Nineteen of the patients (73%) with conventional resuscitation had ventricular fibrillation, 13 of whom survived to hospital admission (no survivals with other arrhythmias) and 7 were discharged. In contrast, in the Lifestick-CPR group, only 9 patients had ventricular fibrillation (38%; P=<0.02; OR, 2.5; 95% CI, 0.6 to 10.6). Four of these 9 patients and 5 of 15 patients with other arrhythmias survived to hospital admission, but none survived to hospital discharge. Autopsy in a subgroup of patients who died at the scene revealed less injuries with Lifestick than with conventional resuscitation.
Conclusion Lifestick resuscitation is feasible and safe and may be advantageous in patients with asystole or pulseless electric activity.
Key Words: cardiopulmonary resuscitation death, sudden heart-assist device
| Introduction |
|---|
|
|
|---|
Recently a new method has been developed that combines ACD-CPR and interposed abdominal compression techniques.6 This phased chest and abdominal compression-decompressionCPR (Lifestick-CPR) has been shown to result in increased coronary perfusion pressure and total cerebral blood flow, higher end-tidal carbon dioxide, and better survival in animals.6,7 In a series of patients with prolonged unsuccessful resuscitation efforts, the new technique has also led to a significant increase of coronary perfusion pressure.8 The aim of the present study was to determine feasibility, safety, and efficacy of phased chest and abdominal compression-decompression compared with conventional CPR in patients with nontraumatic out-of-hospital cardiac arrest.
| Methods |
|---|
|
|
|---|
In cases of a life-threatening situation (eg, cases of cardiac arrest), the nearest MICU is alarmed parallel to the paramedic ambulance (so-called rendezvous system). If the MICU next to the scene is not available during the daytime, a rescue helicopter (also staffed by an emergency physician and a paramedic) is alerted to guarantee an alarm-scene interval of
10 minutes for the emergency physician in
90% of alarms.
The Lifestick consists of a rigid frame attached to 2 adhesive pads (Figure 1). The smaller pad (20x17 cm) has to be placed on the mid-sternum, the larger pad (37x25 cm) on the epigastrium. The pads are fixed to the Lifestick before placement on the patient. For this study, the Lifestick resuscitator was used in a 15:2 compression-ventilation ratio, at 60 cycles per minute. A metronome-driven 240° thoracic-abdominal phase shift (waltz-timing) was applied, which has been shown to result in optimal hemodynamic response.9 The abdominal compression force was limited to 18 to 28 kg and controlled by a colored LED display on the top of the frame. The display for the chest forces can be switched from a low (28 to 45 kg) setting to a medium (41 to 63 kg) or a high (54 to 82 kg) setting to achieve the target compression depth of 4 to 5 cm.
|
The crews of 4 MICUs and the helicopter crew were trained in small groups in application of the Lifestick, by use of a CPR-training mannequin. The initial training was performed with help from and under supervision of the monitoring staff of the manufacturer (Datascope Cardiac Assist Division). Further crew training was continuously repeated under supervision of the emergency physician on duty.
In principal, CPR was performed according to the guidelines of the European Resuscitation Council,10,11 except that patients randomized to the treatment arm had Lifestick-CPR instead of manual CPR. Patients were allocated in a 1:1 ratio to Lifestick-CPR or conventional CPR, by numbered prerandomized sealed envelopes.
The inclusion criteria were age
18 years, normothermia, and nontraumatic eye-witnessed arrest of assumed cardiac etiology irrespective of underlying arrhythmia. The estimated time interval of arrest and start of randomized treatment had to be
15 minutes, and the duration of advanced cardiac life support (defibrillation and/or pharmacologic intervention)
5 minutes. Patients with ventricular fibrillation were included only if up to 3 consecutive countershocks did not result in return of spontaneous circulation (ROSC). Exclusion criteria were pregnancy, any thoracic or abdominal trauma/surgery appearing to be recent, known terminal or end-stage disease, or severe neurologic impairment.
Time course of resuscitation measures (eg, time point of intubation, pharmacotherapy) were taped on the voice recorder of the semiautomated defibrillators used in all resuscitation attempts (Lifepack 300, Physio-Control). These devices also simultaneously record the rhythm of the patient. Lifestick-CPR or conventional CPR was applied for
20 minutes after randomization if ROSC was not achieved earlier. For patients who could not be resuscitated, relatives were asked for consent to perform an autopsy. Autopsies were performed by one of us (T.R.), blinded to the applied CPR method. Special attention was given to possible CPR-related injuries.
Study endpoints were safety (adverse events and autopsy results), ROSC
5 minutes, hospital admission and 1-hour survival rates, the hospital discharge rate, and 6-month survival with neurologic assessment. The study protocol was approved by the Ethics Committee of the Universitätsklinikum Benjamin Franklin, Free University, Berlin, Germany.
Statistical Analysis
Values of continuous variables are given as median, maximum, and minimum. We used the
2 test with continuity correction for dichotomous variables and the Wilcoxon test for continuous variables for group comparisons. Statistical significance was considered at P<0.05.
| Results |
|---|
|
|
|---|
|
Table 1 gives the demographic data, response intervals, and information on the situation at the scene. The first arrhythmia registered was derived from the tape recordings of the semiautomated defibrillators. Although we found no differences within the 2 study arms in most of the variables, there were significant differences in the distribution of arrhythmias (P<0.02; OR, 4.5; 95% CI, 1.4 to 15) at first registration in favor of ventricular fibrillation in the conventional CPR group. A similar trend was observed at randomization (P=0.11; OR, 3.3; 95% CI, 0.9 to 11.4).
|
The response intervals correspond well with the general experience in the Berlin EMS. Other data reveal some minor differences compared with previous results from the Berlin EMS,12 which may all be explained by the fact that we included only patients with an eye-witnessed arrest. The rate of bystander CPR and the proportion of patients found in ventricular fibrillation was somewhat higher in our study than generally observed in Berlin. Also, the median age of patients in this study was lower than that usually seen in resuscitations in Berlin, where it was 69 years.12
Outcome
In a total of 22 of the 50 patients (44%), ROSC was observed (Table 2). With conventional CPR, 13 of 26 patients (50%) had ROSC compared with 9 of 24 (38%) with Lifestick-CPR (P=0.55; OR, 1.7; 95% CI, 0.5 to 5.2). If related to baseline arrhythmia, however, differences become more distinct. In patients with ventricular fibrillation, ROSC was achieved in 13 of 19 patients (68%) with standard CPR and in 44% (4 of 9) with the new technique (P=0.43; OR, 2.7; 95% CI, 0.5 to 13.9). In contrast, in patients with pulseless electric activity or asystole, ROSC was achieved in 5 of 15 patients (33%) with Lifestick-CPR but in none of the 7 patients with conventional CPR (P=0.23).
|
Twelve of the 26 patients (46%) in the standard CPR group survived to hospital admission and to
1 hour after admission. Seven patients of this group were discharged, and 6 survived for >6 months (P<0.05 compared with the Lifestick-CPR group). One of the patients remained in a coma in a vegetative state. He died after 12 weeks while being cared for in a nursing home. A second patient who was a long-term survivor had a moderate cerebral and overall disability; a third patient had a moderate cerebral disability but a good overall performance. The other 4 long-term survivors were without neurological deficit.
With Lifestick-CPR, 7 patients were admitted to a hospital (4 patients with ventricular fibrillation, 3 with other arrhythmia at baseline registration). All but one of these patients survived for >1 hour after admission, but none regained consciousness or survived to discharge. One patient with asystole at first arrhythmia registration died in irreversible cardiogenic shock shortly after admission to an intensive care unit.
Safety of the Device
There were no reports of severe injuries or resuscitation-related major complications during clinical course in survivors. In 9 of the 14 patients with conventional CPR and 7 of 17 patients with Lifestick-CPR who died at the scene, an autopsy was performed (Table 3). Sternal or rib fractures were found more frequently with conventional CPR, the difference being significant (P<0.05) with rib fractures. With conventional CPR, we observed 1 unusual fracture of the thoracal spine in an elderly female patient with severe osteoporosis. With Lifestick-CPR, minor abdominal lesions (insignificant peritoneal bleedings) were observed in 1 patient, and 1 patient had a minor hematoma in the hilar region.
|
Practicability of the Lifestick Device
The median time of interruption of cardiac massage necessary to fix the device was 20 seconds (range, 10 to 73 seconds). More than 30 seconds were needed in 4 patients. In one of the patients, fixation of the device was unsatisfactory because of hairy and wet skin, leading to a pause in the cardiac massage of 73 seconds. There were 9 reports of poor and unsatisfactory fixation of the adhesive pads to the skin, leading to instability of the device in 4 patients. No other difficulties with respect to application of the Lifestick were noticed. The ability to operate and deliver the compression force was judged to be good in 14 cases and fair in 10 cases on a 4-step scale from superior to poor. The use of the device was subjectively felt to not be more strenuous than conventional CPR and was well accepted.
| Discussion |
|---|
|
|
|---|
One major problem with CPR is the mechanical trauma, which in some cases may lead to life-threatening injuries. Depending on the definition and exactness of investigation, the rate of injuries is reported to be between 21% and >65%.14 In a series of 705 autopsies after conventional resuscitation, thoracic complications of different degrees were observed in a total of 42.7% of cases. Rib fractures were found in 31.6%, sternal fractures in 21.1%. In 18.3%, mediastinal hemorrhages were observed. Moreover, abdominal injuries were described in 30.8%, and pulmonary complications in 13%. In nearly 0.5% of the cases, the injuries were categorized as life-threatening.15 With ACD-CPR, the rate of severe thorax injuries was equal or somewhat higher than that of standard CPR.16 In 1 investigation in cadavers, sternal fractures were found in 11 of 37 cases and rib fractures in 27 of 37 cases at autopsy after ACD-CPR,17 predominantly in women. In our study, we generally found rib and sternal fractures in a high rate of autopsies (Table 3) with conventional CPR, which corresponds to literature,15,16,18 and a significantly lower incidence of fractures after Lifestick-CPR. Most importantly, there was no tendency of an increase of abdominal lesions with the new technique. Autopsy was, however, not performed in all nonsurvivors, and moreover, there was no systematic search for injuries in survivors, which may lead to underreporting.
On first glance, outcome with Lifestick-CPR seems to be unfavorable compared with that of standard CPR. Because of the small sample size, we cannot exclude the possibility of a type II error, ie, standard CPR is superior to Lifestick-CPR. There is, however, a surprisingly high proportion in the subgroup with asystole and pulseless electric activity who developed ROSC, a result similar to observations with interposed abdominal compression CPR.3 Finally, bias is a special problem in a small study with so many inherent factors that have a decisive influence on the prognosis of the victim of cardiac arrest.
The 2 most important factors with respect to prognosis are the cause of arrest (some of which are absolutely irreversible) and the arrhythmia first registered with the patient, which is in turn also related to the principal underlying cause of arrest. Therefore, when evaluating resuscitation results, autopsy diagnoses with respect to cause of death are of particular interest (Table 4). In fact, in our study we found a significantly unequal distribution of rhythm disturbances in favor of ventricular fibrillation in the standard CPR group, which has the best prognosis of all arrhythmias. As may be expected from this unequal distribution, autopsy revealed a cardiac cause of death (6 patients died from coronary heart disease, 1 patient from dilatative cardiomyopathy) in nearly all autopsies performed in patients with standard CPR. In contrast, in the group with Lifestick-CPR, consisting of 63% of patients with asystole or pulseless electric activity at first registration, only 2 patients died from coronary heart disease, whereas 4 patients had extracardial causes of arrest, which was absolutely irreversible in at least one of them.
|
With respect to practicability, the Lifestick was well accepted by the rescue personnel. In contrast to the Cardio-Pump,19 Lifestick users did not complain about exhaustion or have the impression of strenuous handling. There were also no reports of discomfort in the back bone or other complaints by the operators. There were a few reports on instability of the device because of unsatisfactory fixation of the pads to the skin, which lead in 1 case to an interruption of cardiac massage of 73 seconds. No problems with respect to defibrillation were reported. Defibrillation was exclusively performed via the adhesive pads used with semiautomated defibrillators.
The metronome giving the thoracic-abdominal phase-shift was found to be useful. In contrast to the abdominal force control, the chest force switch and control light was felt to be troublesome. The operators would have preferred a purely optical control of the necessary compression depth of 4 to 5 cm, because the force needed to achieve this depth has to be individualized depending on the elasticity of the thorax and may not easily be controlled by the 3 force settings employed by the device. This problem while using the device may also explain failure to continuously stay exactly with the 2:15 ratio of ventilation to compression/decompression cycles in 9 cases and the observation of temporary simultaneous compression and ventilation during resuscitation in 3 patients. Both handling errors may lead to a decreased coronary blood flow and consecutively worsen the chances for successful resuscitation.20
We conclude from our results that cardiopulmonary resuscitation by use of the Lifestick after a comprehensive and repeated training is feasible, safe, and easily accepted by users. There are less CPR-related injuries with Lifestick-CPR compared with standard CPR, and it may be advantageous for a subgroup of patients, eg, patients with asystole or pulseless electric activity. Clearly, this promising new technique deserves a large-scale clinical trial for a conclusive evaluation.
| Acknowledgments |
|---|
Received March 14, 2001; revision received May 22, 2001; accepted May 31, 2001.
| References |
|---|
|
|
|---|
2.
Halperin HR, Tsitlik JE, Gelfand M, et al. A preliminary study of cardiopulmonary resuscitation by circumferential compression of the chest with use of a pneumatic vest. N Engl J Med. 1993; 329: 762768.
3.
Sack JB, Kesselbrenner MB, Jarrad A. Interposed abdominal compression-cardiopulmonary resuscitation and resuscitation outcome during asystole and electromechanical dissociation. Circulation. 1992; 86: 16921700.
4.
Cohen TJ, Tucker KJ, Lurie KG, et al. Active compression-decompression: a new method of cardiopulmonary resuscitation. JAMA. 1992; 267: 29162923.
5.
Plaisance P, Lurie KG, Vicaut E, et al. A comparison of standard cardiopulmonary resuscitation and active compression-decompression resuscitation for out-of-hospital cardiac arrest. N Engl J Med. 1999; 341: 569575.
6.
Tang W, Weil MH, Shock R, et al. Phased chest and abdominal compression-decompression: a new option for cardiopulmonary resuscitation. Circulation. 1997; 95: 13351340.
7. Wenzel V, Lindner KH, Prengel AW, et al. Effect of phased chest and abdominal compression-decompression cardiopulmonary resuscitation on myocardial and cerebral blood flow in pigs. Crit Care Med. 2000; 28: 11071112.[Medline] [Order article via Infotrieve]
8. Sterz F, Behringer W, Berzianovich A, et al. Active compression-decompression of thorax and abdomen (Lifestick-CPR) in patients with cardiac arrest. Circulation. 1996 (Suppl I); 94: I-9.Abstract.
9. Beyar R, Kimmel E, Sideman S, et al. Effect of thoracic and abdominal pressure waves on blood flow in cardiopulmonary resuscitation. Int J Heat Mass Transfer. 1984; 27: 14731483.
10. Handley A, Becker Ö, Allen M, et al. Single rescuer adult basic life support: an advisory statement from the Basic Life Support Working Group of the International Liaison Committee on Resuscitation (ILCOR). Resuscitation. 1997; 34: 101108.[Medline] [Order article via Infotrieve]
11. Kloeck W, Cummins D, Chamberlain D, et al. The universal ALS algorithm: an advisory statement by the Advanced Life Support Working Group of the International Liaison Committee on Resuscitation. Resuscitation. 1997; 34: 109111.[Medline] [Order article via Infotrieve]
12. Arntz HR, Oeff M, Willich SN, et al. Establishment and results of an EMT-D program in a two-tiered physician-escorted rescue system: the experience in Berlin, Germany. Resuscitation. 1993; 26: 3946.[Medline] [Order article via Infotrieve]
13. Babbs CF, Weaver C, Ralston S, et al. Cardiac, thoracic, and abdominal pump mechanisms in cardiopulmonary resuscitation: studies in an electrical model of the circulation. Am J Emerg Med. 1984; 2: 299308.[Medline] [Order article via Infotrieve]
14. Sommers MS. Potential for injury: trauma after cardiopulmonary resuscitation. Heart Lung. 1991; 20: 287293.[Medline] [Order article via Infotrieve]
15.
Krischer JP, Fine EG, Davis JH, et al. Complications of cardiac resuscitation. Chest. 1987; 92: 287291.
16. Baubin M, Sumann G, Rabl W, et al. Increased frequency of thorax injuries with ACD-CPR. Resuscitation. 1999; 41: 3338.[Medline] [Order article via Infotrieve]
17. Baubin M, Rabl W, Pfeiffer KP, et al. Chest injuries after active compression-decompression cardiopulmonary resuscitation (ACD-CPR) in cadavers. Resuscitation. 1999; 43: 915.[Medline] [Order article via Infotrieve]
18. Kloss T, Puschel K, Wischhusen F, et al. Resuscitation injuries. Anaesth Intensivther Notfallmed. 1983; 18: 199203.
19. Shultz JJ, Mianulli MJ, Gisch TM, et al. Comparison of exertion required to perform standard and active compression-decompression cardiopulmonary resuscitation. Resuscitation. 1995; 29: 2331.[Medline] [Order article via Infotrieve]
20. Kern KB, Hilwig RW, Berg RA, et al. Optimizing ventilation with phased chest and abdominal compression-decompression (Lifestick) CPR. J Am Coll Cardiol. 1999; 33: 369.Abstract.
This article has been cited by other articles:
![]() |
M. P. Anstadt and J. E. Lowe Cardiopulmonary Resuscitation Card. Surg. Adult, January 1, 2008; 3(2008): 487 - 506. [Full Text] |
||||
![]() |
D. Muller, R. Agrawal, and H.-R. Arntz How Sudden Is Sudden Cardiac Death? Circulation, September 12, 2006; 114(11): 1146 - 1150. [Abstract] [Full Text] [PDF] |
||||
![]() |
Part 6: CPR Techniques and Devices Circulation, December 13, 2005; 112(24_suppl): IV-47 - IV-50. [Full Text] [PDF] |
||||
![]() |
Part 4: Advanced Life Support Circulation, November 29, 2005; 112(22_suppl): III-25 - III-54. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2001 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |