(Circulation. 2001;104:1320.)
© 2001 American Heart Association, Inc.
Current Perspective |
From the Cardiovascular Research Foundation, New York, NY.
Correspondence to Gary S. Mintz, MD, 611 Pennsylvania Ave, SE, #386, Washington, DC 20003. E-mail gsmintz{at}tctmd.com
Key Words: ultrasonics stents restenosis remodeling radioisotopes
| Introduction |
|---|
|
|
|---|
Most brachytherapy studies have included IVUS analyses, but to various degrees and with various methodologies. In some studieseg, the Washington Radiation for In-stent Restenosis (WRIST) trialsserial (postirradiation and follow-up) IVUS was performed in the majority of patients. In other trials, IVUS was usually a site-specific substudy in small numbers of patients (Table 1). These studies assessed only the short-term (typically 6 to 9 months) effects of brachytherapy.
|
| Methodological Considerations |
|---|
|
|
|---|
Understanding lesion effects is more straightforward than understanding edge effects. (1) Edge analysis was not performed in all studies (Table 1). (2) Some studies combined proximal and distal edges and analyzed them together; others analyzed proximal and distal edges separately. (3) Edge analysis requires a larger sample size because of the wider range of possible responses. (4) It may be hard to determine the relationship between the source, the injured segment, and the IVUS images to assess geographical miss. Angiographic guidance may work in stented lesions; even in stented lesions (which are often radiolucent), however, meticulous comparison of IVUS and angiographic studies is necessary. (5) Some studies reported volumetric (or mean planar) analysis; others reported CSA analysis millimeter by millimeter over a specified length. Because edge effects can be very focal and change over short distances, it is preferable to analyze studies millimeter by millimeter over the length of the segment. This requires very careful "registration" of serial (postirradiation and follow-up) images. (6) Most studies analyze 5-mm-long edge segments; yet, the "edge effect" may be more than 5 mm from the edge of the stent or lesion. To properly understand the incidence and mechanism of edge effects, protocols must be designed to image the entire irradiated (not just stented) length plus at least 5 to 10 mm proximal and distal to the irradiated segment.
| Radiation in Nonstented Lesions |
|---|
|
|
|---|
-radiation and balloon angioplasty. These studies, however, did not collect serial IVUS data. The data on ß-irradiation of nonstented lesions are limited to 2 published reports from the Beta Energy Restenosis Trial (BERT). The first study showed no change in lesion external elastic membrane (EEM), lumen, or plaque-plus-media (P&M) CSA from after irradiation to follow-up.3 In this report, only 1 image slice per lesion was studied, the location of this image slice was not specified, and edges were not analyzed. The second study was an extensive and detailed serial (ECG-gated) volumetric analysis of 30-mm-long lesions and contiguous 5-mm-long edges.4 At the lesion, there was significant positive remodeling (an increase in EEM volume) that was roughly equal to the increase in P&M volume; as a result, lumen volume remained unchanged. These "beneficial" effects on remodeling, however, did not extend to the edges, where there was an increase in P&M but no positive remodeling, resulting in lumen loss (Table 2). The different conclusions in these 2 reports may be related to the significant differences in methodology.
|
In another report from BERT, the irradiated segment (but not the edges) was analyzed in 2-mm-long subsegments.5 There were 3 independent predictors of the absolute P&M volume at follow-up: plaque volume after irradiation, minimum effective dose delivered to 90% of the adventitia, and the type of plaque (hard versus soft). There was wide variability, however, in the calculated minimum effective dose delivered to 90% of adventitia: 26.2% of the subsegments received <4 Gy, 32.5% received 4.0 to 5.9 Gy, 22.8% received 6.0 to 7.9 Gy, 13.6% received 8.0 to 9.9 Gy, and 4.9% received
10 Gy. The authors concluded that the minimum effective dose was 4 Gy, because subsegments that received >4 Gy demonstrated a significantly smaller increase in P&M volume than those that received <4 Gy (P<0.01). This analysis, however, assumed that adjacent subsegments behaved independently with regard to
EEM,
P&M, and
lumen, an assumption that may not be valid.
| Radiation Followed by Stenting |
|---|
|
|
|---|
The Proliferation Reduction with Vascular Energy Trial (PREVENT) was a trial of intracoronary radiation with 32P, a ß-emitting source, in patients with ISR (n=25), restenosis after PTCA (n=7), and de novo lesions treated with stents (n=64) or balloon angioplasty (n=9).79 Patients were randomized to placebo (n=25) or to 1 of 3 dose schedules. The IVUS analysis was limited primarily to the patients who received stents. Within the stented segment, there was an increase in IH volume in irradiated patients that was less than in placebo patients. An analysis of 40 stent edges in 23 patients showed a decrease in EEM (but no change in plaque), resulting in a decrease in lumen dimensions.9 There was geographic miss in 30 of 40 edges; in 77% of these 30 edges, negative remodeling was the dominant cause of edge lumen loss. Conversely, in the 10 edges with adequate radiation coverage, negative remodeling was seen in only 20%, and the main mechanism of edge lumen loss was IH (Table 2).
ISR Treated With -Irradiation
|
|---|
|
|
|---|
|
The WRIST Trial specifically addressed treatment of ISR.11 The dose prescription was 15 Gy at 2 mm from the source. At follow-up, the minimal lumen CSA was larger in the 192Ir group than the placebo group (P<0.0001, Table 3). There was a smaller increase in IH volume (4±33 versus 65±65 mm3, P<0.0001) and mean IH CSA (P<0.0001, Table 3) in the 192Ir group than placebo. Of note, more than half of the lesions had a reduction in IH between radiation and follow-up.
Edge Effects in WRIST
There were no significant overall differences in
EEM,
lumen, or
P&M in 192Ir versus placebo patients. Distal edge recurrence was noted, however, in 8
-irradiated patients. These patients were compared with 21 irradiated patients with no recurrence.12 Lesions with distal edge recurrence had (1) a greater decrease in distal lumen CSA (-3.0±1.2 versus -0.7±1.0 mm2, P=0.0002), (2) no change in distal EEM CSA (-0.2±0.5 mm2) versus an increase in distal EEM CSA of 1.0±0.9 mm2 in nonedge recurrence lesions (P=0.0047), and (3) a greater increase in distal P&M CSA (2.9±1.2 versus 1.7±0.6 mm2, P=0.0103). In addition, lesions with distal edge recurrence had a significant increase in mean intrastent IH, but the decrease in lumen CSA and the increase in IH CSA were more pronounced closer to the distal edge of the stent. These findings suggested that edge recurrence after
-irradiation was part of diffuse treatment failure.
In 6 of 8 distal edge recurrence lesions, there was evidence of geographical miss (versus 4 of 21 nonrecurring lesions, P=0.0046).12 Lesions with geographical miss had a greater decrease in lumen CSA because there was no change in EEM CSA (versus an increase in EEM CSA in lesions with adequate coverage); the increase in
P&M CSA was similar in both groups. Thus, as in BERT and PREVENT, edge lumen loss appeared to be related to remodeling, which appeared to be related to inadequate coverage of injured edge segments.
Impact of Brachytherapy on Reference Segments
One proposed solution to edge effects was to radiate longer segments. To determine the short-term safety of this approach, the effect of brachytherapy on irradiated but uninjured reference segments was analyzed.13 There was an increase in P&M CSA in both the 192Ir and placebo patients. In the 192Ir group, however, there was an increase in EEM CSA, whereas in the placebo group, there was a decrease in EEM CSA. Consequently, there was no change in lumen CSA in the 192Ir group, whereas in the placebo group, there was a decrease in lumen area. Thus, as in other studies noted above, brachytherapy promoted positive remodeling to prevent lumen loss despite an increase in plaque mass.
Effect of Lesion Length on Effectiveness of Brachytherapy
"Long WRIST" was a double-blind, placebo-controlled trial of brachytherapy in long, diffuse native artery ISR (length, 36 to 80 mm). The dose prescription was identical to that in WRIST. In placebo patients, the increases in IH CSA (Table 3) and volumes (120±81 versus 35±76 mm3) were significantly greater in irradiated patients (P<0.0001 for both comparisons).14
In an attempt to understand the impact of lesion length on treatment of ISR, irradiated Long WRIST lesions were compared with irradiated native artery WRIST lesions.15 Postintervention stent areas were larger in WRIST (8.9±2.5 mm2) than in Long WRIST (7.6±2.5 mm2, P=0.0274). In particular, minimum stent CSA in Long WRIST measured only 5.2±1.0 mm2. Serial IVUS documented reduced efficacy of brachytherapy in long, diffuse ISR lesions. There was (1) a significant decrease in mean lumen CSA and a significant increase in mean IH CSA in Long WRIST but not in WRIST patients; (2) a much greater increase in maximum IH CSA in Long WRIST versus WRIST; (3) a ratio of maximum to minimum follow-up IH CSA of 7.7±13.6 in Long WRIST versus 2.9±1.3 in WRIST, indicating a greater heterogeneity in neointimal recurrence in Long WRIST; and (4) a follow-up minimum lumen area of 2.9±1.0 mm2 in Long WRIST versus 4.2±2.0 mm2 in WRIST, P=0.0015.
The maximum and minimum distances from the IVUS catheter to the EEM were then measured as an index of the source-to-target distance. The maximum source-to-target distance was greater in Long WRIST than in WRIST (4.0±0.7 versus 2.8±0.4 mm, P<0.0001), and it correlated directly with ISR length (r=0.547, P<0.0001). Importantly,
minimum lumen CSA,
maximum IH CSA, and the ratio of maximum to minimum follow-up IH CSA all correlated with the maximum source-to-target distance. This indicated that the greater heterogeneity in the neointimal response in Long WRIST lesions was related to a greater variability in lesion geometry and source eccentricity.
This suggested 2 possible solutions to the long lesion problem: improving the dose homogeneity via centering (see below) or increasing the dose prescription. To test the second solution, a group of 25 patients with long, diffuse ISR lesions were then treated with a higher dose (18 Gy at 2 mm from the source). At follow-up, the minimum lumen CSA was larger in the high-dose group (4.0±1.4 versus 2.9±1.0 mm2, P=0.0009, Table 3).
Gamma-1 Trial
The Gamma-1 Trial was a multicenter study of
-irradiation treatment of ISR. The dose prescription was similar to that in SCRIPPS. The IVUS substudy was performed in 4 centers. At follow-up, the increase in IH volume (28±37 versus 50±40 mm3, P=0.0352) and the increase in IH CSA (Table 3) were less in the 192Ir versus placebo patients. Follow-up minimum lumen area was larger and the area stenosis (33±32% versus 55±38%, P=0.0124) smaller in the 192Ir patients. The lack of an edge effect that was observed in WRIST was substantiated in Gamma-1.16
IVUS Dosimetry
Two studies (SCRIPPS and Gamma-1) used IVUS dosimetry; the others did not. The results of fixed dosimetry were similar to those of IVUS dosimetry, which brings into question the need for IVUS guidance for dosimetric calculations.
Centering
Although uniform dosing of the adventitia may be desirable, it probably is not possible because it requires the source to be centered within the artery. Two parameters determine centering within the artery: (1) centering the source within the lumen and (2) P&M eccentricity. These 2 factors may be additive (to exaggerate dose heterogeneity) or they may cancel each other. There is a greater possibility that these 2 factors are additive in longer lesions, as was suggested by the comparison of Long WRIST with WRIST lesions.15 Calculations indicate that centering a source within the lumen (eliminating the first of these 2 factors) should improve dose homogeneity for both
- and ß-emitters, but especially for ß-emitters.17,18
| ß-Irradiation Treatment of ISR |
|---|
|
|
|---|
Stents and Radiation Therapy (START)20 was a randomized trial comparing the same brachytherapy system as used in BERT (hydraulic delivery of 90Sr/90Y seeds) with placebo in the treatment of ISR. The findings were similar to the
-emitter studies, with no change in the IH CSA in the treated group and a significant increase in the IH CSA in the placebo group (Table 3).
| Radiation-Emitting Stents |
|---|
|
|
|---|
Candy-Wrapper Effect
The Isostent experience illustrates the importance of the analysis plan in understanding edge effects. When analyzed millimeter by millimeter, there was a profound, focal, and sharply demarcated lumen loss peaking
1 mm outside the edge of the ß-emitting stent resulting from a focal increase in neointimal tissue.21,22,24 Conversely, when 5-mm-long segments proximal and distal to the stent edges were analyzed as a volume, the conclusion was quite different: (1) a decrease in lumen volume, (2) a decrease in EEM volume, but (3) no significant change in P&M volume.6
Two modifications were tried. Creation of cold ends (25-mm stent length with 15-mm long, 3- to 24-µCi, ß-emitting centers and 5-mm-long nonemitting ends) broadened the peak of the edge effect, shifted the peak IH response within the body of the stents, and caused more in-stent neointima.25 Conversely, hot ends appeared to attenuate the edge effect but were associated with 6 cases of late stent malapposition. Although there was no significant increase in mean EEM CSA in the "hot-ends cases," late stent malapposition was associated with positive remodeling, emphasizing the importance of looking at individual cases, not just at the entire cohort. One previous study showed no increase in EEM within the length of the 32P ß-emitting stent in either the low-activity (0.75 to 1.5 µCi) or high-activity (6.0 to 12.0 µCi) group.6 The activity of the 18-mm-long hot-ends Isostent, however, was
8 µCi within the 14-mm body (0.57 µCi/mm) and
10 µCi at the two 2-mm ends combined (2.6 µCi/mm); the total activity (
18 µCi) was more than that of the previously noted high-activity group.
Late Catch-Up
In a recent report, between 6 and 12 months there was a significant increase in IH volume, from 18.2±12.6 to 27.9±12.0 mm3 (P=0.001), mainly in the mid and distal portions of the stent. There was no change in edge lumen volume during this time period.26 Although the volume of neointimal tissue was less at 12 months than in nonbrachytherapy studies, such as the Heparin Infusion Prior to Stenting trial (40.5 mm3),27 the late increase is disturbing.
| IVUS Complications of Brachytherapy |
|---|
|
|
|---|
A second potential complication is late stentvessel wall malapposition, an effect of exaggerated positive remodeling in the absence of neointimal hyperplasia.3032 In Gamma-1, late malapposition was seen in only 1 patient; that patient was in the placebo group.16,31 In START, there was 1 documented case in the irradiated group and none in the placebo group. In hot-ends Isostent, there were 6 cases. Positive remodeling and neointima inhibition, the combination that appears to lead to late malapposition, are dose related.5
A third potential complication is the "black hole" phenomenon: echolucent neointimal tissue.33,34 This has been seen after both
- and ß-emitters, after ß-emitting stents, in patients with and without stents, and in patients with ISR. Directional atherectomy has been performed in 4 patients; specimens showed large myxoid areas with interspersed smooth muscle cells that were scattered in an extracellular matrix containing proteoglycan. The clinical implications of this finding, however, are not known.
As noted, none of these observations are unique to brachytherapy; however, they occur with increased frequency in irradiated patients. Despite the suggestion that these complications may contribute to late thrombosis, none of these patients developed late thrombosis, and no patient with late thrombosis had IVUS imaging. Thus, it is unclear whether any of these observations should be called a true complication. It is important, however, that these terms be defined carefully and used consistently. For example, the term "black hole"echolucent neointimal tissueis sometimes confused with the gaps between stent and vessel wall, ie, late malapposition.
| What Have We Learned? |
|---|
|
|
|---|
The mechanisms of brachytherapy in PTCA, stenting, and treatment of ISR are dose related: (1) decreased neointimal hyperplasia, (2) positive remodeling in PTCA-treated lesions, and (3) either no effect or positive remodeling along irradiated edges and uninjured reference segments. At the edges, where the dose is less, and especially in the presence of geographical miss, there is a combination of increased neointimal hyperplasia and either absence of positive remodeling or frank negative remodeling that contributes to lumen loss.
The findings in ß-emitting stents are idiosyncratic and very different. There is a dose-related decrease in intrastent neointima. The candy-wrapper effect is primarily a focal exaggeration of IH accumulation at the edge of the stent. The late catch-up in neointimal proliferation needs to be studied in other settings.
Unusual IVUS observations, such as unhealed dissections, late malapposition (also a manifestation of radiation-induced positive remodeling), and the black hole, are worrisome, but their exact clinical consequence is unknown.
All of these findings are lessons to be remembered in the new era of drug-eluting stents.
| References |
|---|
|
|
|---|
2.
Hoffmann R, Mintz GS, Dussaillant GR, et al. Patterns and mechanisms of in-stent restenosis: a serial intravascular ultrasound study. Circulation. 1996; 94: 12471254.
3.
Meerkin D, Tardif J-C, Crocker IR, et al. Effects of intracoronary ß-radiation therapy after coronary angioplasty: an intravascular ultrasound study. Circulation. 1999; 99: 16601666.
4.
Sabate M, Serruys PW, van der Giessen WJ, et al. Geometric vascular remodeling after balloon angioplasty and ß-radiation therapy: a three-dimensional intravascular ultrasound study. Circulation. 1999; 100: 11821188.
5.
Sabate M, Marijnissen JP, Carlier SG, et al. Residual plaque burden, delivered dose, and tissue composition predict 6-month outcome after balloon angioplasty and beta-radiation therapy. Circulation. 2000; 101: 24722477.
6.
Kay IP, Sabate M, Costa MA, et al. Positive geometric vascular remodeling is seen after catheter-based radiation followed by conventional stent implantation but not after radioactive stent implantation. Circulation. 2000; 102: 14341439.
7.
Raizner AE, Oesterle SN, Waksman R, et al. Inhibition of restenosis with ß-emitting radiotherapy: report of the Proliferation Reduction with Vascular Energy Trial (PREVENT). Circulation. 2000; 102: 951958.
8. Okura H, Lee DP, Handen CE, et al. Contribution of vessel remodeling to "edge effect" following intracoronary ß-irradiation: a serial volumetric intravascular ultrasound study. Circulation. 1999; 100 (suppl I): I-511.
9. Lee DP, Okura H, Handen CE, et al. Reference segment and target lesion effects of 32P radiation: intravascular ultrasound results of the PREVENT Trial. Circulation. 1999; 100 (suppl I): I-517.
10.
Teirstein PS, Massullo V, Jani S, et al. Catheter-based radiotherapy to inhibit restenosis after coronary stenting. N Engl J Med. 1997; 336: 16971703.
11.
Waksman R, White RL, Chan RC, et al. Intracoronary gamma radiation therapy after angioplasty inhibits recurrence in patients with in-stent restenosis. Circulation. 2000; 101: 21652171.
12. Ahmed JM, Mintz GS, Waksman R, et al. Serial intravascular ultrasound analysis of edge recurrence after intracoronary gamma radiation treatment of native artery in-stent restenosis lesions. Am J Cardiol. 2001; 87: 11451149.[Medline] [Order article via Infotrieve]
13.
Ahmed JM, Mintz GS, Waksman R, et al. Safety of intracoronary gamma radiation on uninjured reference segments during the first six months after treatment of in-stent restenosis: a serial intravascular ultrasound study. Circulation. 2000; 101: 22272230.
14.
Ahmed JM, Mintz GS, Leiboff B, et al. A serial intravascular ultrasound study to assess the efficacy of intracoronary
-irradiation in preventing recurrent diffuse in-stent restenosis in long lesions. Circulation. 2000; 102 (suppl II): II-692.
15.
Ahmed JM, Mintz GS, Waksman R, et al. Serial intravascular ultrasound analysis of the impact of lesion length on the efficacy of intracoronary
-irradiation for preventing recurrent in-stent restenosis. Circulation. 2001; 103: 188191.
16.
Mintz GS, Weissman NJ, Teirstein PS, et al. Effect of intracoronary gamma-radiation therapy on in-stent restenosis: an intravascular ultrasound analysis from the Gamma-1 Study. Circulation. 2000; 102: 29152918.
17. Armin A-Z, Russo RJ, Bhargava V, et al. Centering of the radiation source in the artery lumen results in crucial improvement of radiation dose heterogeneity in saphenous vein grafts: observations from the SCRIPPS Trial. Circulation. 1998; 98 (suppl I): I-779.
18. Bhargava B, Mintz GS, Leiboff BO, et al. Cross-sectional measurements after treatment of in-stent restenosis: an intravascular ultrasound guide for intracoronary radiation dosimetry. J Am Coll Cardiol. 1999; 33: 53A.
19. Bhargava B, Mintz GS, Mehran R, et al. Serial volumetric intravascular ultrasound analysis of the efficacy of beta irradiation in preventing recurrent in-stent restenosis. Am J Cardiol. 2000; 85: 651653.[Medline] [Order article via Infotrieve]
20. Takagi A, Morino Y, Fox T, et al. Efficacy of intracoronary ß-irradiation for the treatment on in-stent restenosis: volumetric analysis by intravascular ultrasound. Circulation. 2000; 102 (suppl II): II-422.
21.
Albiero R, Adamian M, Kobayashi N, et al. Short- and intermediate-term results of 32P radioactive ß-emitting stent implantation in patients with coronary artery disease: the Milan Dose-Response Study. Circulation. 2000; 101: 1826.
22. Weissman NJ, Titana A, Canos DA, et al. Comparison of IVUS findings of the ß-emitting Isostent versus a non-radiation control group from the HIPS trial. Circulation. 2000; 102 (suppl II): II-691.
23. Weissman NJ, Albiero R, Di Mario C, et al. Final IVUS results from the Isostent BXI multicenter dose response study. Circulation. 2000; 102 (suppl II): II-567.
24.
Albiero R, Nishida T, Adamian M, et al. Edge restenosis after implantation of high activity 32P radioactive ß-emitting stents. Circulation. 2000; 101: 24542457.
25. Weissman NJ, Albiero R, De Bruyne B, et al. Final IVUS results from the multicenter "cold ends" Isostent Study. Circulation. 2000; 102 (suppl II): II-568.
26.
Kay IP, Wardeh AJ, Kozuma K, et al. Radioactive stents delay but do not prevent in-stent neointimal hyperplasia. Circulation. 2001; 103: 1417.
27. Wilensky RL, Tanguay JF, Ito S, et al. Heparin Infusion Prior to Stenting (HIPS) trial: final results of a prospective, randomized, controlled trial evaluating the effects of local vascular delivery on intimal hyperplasia. Am Heart J. 2000; 139: 10611070.[Medline] [Order article via Infotrieve]
28.
Kay IP, Sabate M, Van Langenhove G, et al. Outcome from balloon induced coronary artery dissection after intracoronary beta radiation. Heart. 2000; 83: 332337.
29.
Meerkin D, Tardif JC, Bertrand OF, et al. The effects of intracoronary brachytherapy on the natural history of postangioplasty dissections. J Am Coll Cardiol. 2000; 36: 5964.
30. Okura H, Limpijankit T, Lee DP, et al. Late stent malapposition following intracoronary radiation: a serial intravascular ultrasound study. J Am Coll Cardiol. 2001; 37: 57A.
31. Kozuma K, Costa MA, Sabate M, et al. Late stent malapposition occurring after intracoronary beta-irradiation detected by intravascular ultrasound. J Invasive Cardiol. 1999; 11: 651655.[Medline] [Order article via Infotrieve]
32. Mintz GS, Weissman NJ, Pappas C, Waksman R. Positive remodeling, regression of in-stent neointimal hyperplasia, and late stent malapposition in the absence of brachytherapy. Circulation. 2000; 102: E11.
33.
Castagna MT, Mintz GS, Weissman NJ, et al. "Black hole": echolucent restenosis tissue following brachytherapy. Circulation. 2001; 103: 778.
34. Kay IP, Ligthart JM, Virmani R, et al. The black hole: a new IVUS observation after intracoronary radiation. Circulation. 2000; 102 (suppl II): II-568.
This article has been cited by other articles:
![]() |
S. Cook, P. Wenaweser, M. Togni, M. Billinger, C. Morger, C. Seiler, R. Vogel, O. Hess, B. Meier, and S. Windecker Incomplete Stent Apposition and Very Late Stent Thrombosis After Drug-Eluting Stent Implantation Circulation, May 8, 2007; 115(18): 2426 - 2434. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. J. Beller, J. Kosse, T. Radovits, R. Krempien, M.-L. Gross, I. Berger, S. Hagl, and G. Szabo Adjunct brachytherapy: a new concept to prevent intimal hyperplasia after surgical endarterectomy? Eur. J. Cardiothorac. Surg., March 1, 2006; 29(3): 334 - 342. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Ako, Y. Morino, Y. Honda, A. Hassan, S. Sonoda, P. G. Yock, M. B. Leon, J. W. Moses, H. N. Bonneau, and P. J. Fitzgerald Late Incomplete Stent Apposition After Sirolimus-Eluting Stent Implantation: A Serial Intravascular Ultrasound Analysis J. Am. Coll. Cardiol., September 20, 2005; 46(6): 1002 - 1005. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. W. Serruys, W. Wijns, G. Sianos, I. de Scheerder, P. A. van den Heuvel, W. Rutsch, H. D. Glogar, C. Macaya, P. H. Materne, S. Veldhof, et al. Direct stenting versus direct stenting followed by centered beta-radiation with intravascular ultrasound-guided dosimetry and long-term anti-platelet treatment: Results of a randomized trial: Beta-radiation Investigation with Direct stenting and Galileo in Europe (BRIDGE) J. Am. Coll. Cardiol., August 4, 2004; 44(3): 528 - 537. [Abstract] [Full Text] [PDF] |
||||
![]() |
M.-K. Hong, G. S. Mintz, C. W. Lee, Y.-H. Kim, S.-W. Lee, J.-M. Song, K.-H. Han, D.-H. Kang, J.-K. Song, J.-J. Kim, et al. Incidence, Mechanism, Predictors, and Long-Term Prognosis of Late Stent Malapposition After Bare-Metal Stent Implantation Circulation, February 24, 2004; 109(7): 881 - 886. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. S. Mintz, V. M. Shah, and N. J. Weissman Regional Remodeling as the Cause of Late Stent Malapposition Circulation, June 3, 2003; 107(21): 2660 - 2663. [Abstract] [Full Text] [PDF] |
||||
![]() |
M.-K. Hong, G. S. Mintz, C. W. Lee, J.-M. Song, K.-H. Han, D.-H. Kang, J.-K. Song, J.-J. Kim, N. J. Weissman, N. E. Fearnot, et al. Paclitaxel Coating Reduces In-Stent Intimal Hyperplasia in Human Coronary Arteries: A Serial Volumetric Intravascular Ultrasound Analysis From the ASian Paclitaxel-Eluting Stent Clinical Trial (ASPECT) Circulation, February 4, 2003; 107(4): 517 - 520. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Spanos, G. Stankovic, J. Tobis, and A. Colombo The challenge of in-stent restenosis: insights from intravascular ultrasound Eur. Heart J., January 2, 2003; 24(2): 138 - 150. [Full Text] [PDF] |
||||
![]() |
B. Syeda, P. Siostrzonek, R. Schmid, P. Wexberg, C. Kirisits, S. Denk, G. Beran, A. Khorsand, I. Lang, B. Pokrajac, et al. Geographical miss during intracoronary irradiation: impact on restenosis and determination of required safety margin length J. Am. Coll. Cardiol., October 2, 2002; 40(7): 1225 - 1231. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. M. Shah, G. S. Mintz, S. Apple, and N. J. Weissman Background Incidence of Late Malapposition After Bare-Metal Stent Implantation Circulation, October 1, 2002; 106(14): 1753 - 1755. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Seabra-Gomes Intracoronary brachytherapy for restenosis: an efficient technique in the struggle for survival? Eur. Heart J., September 1, 2002; 23(17): 1319 - 1321. [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2001 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |