(Circulation. 2001;103:2882.)
© 2001 American Heart Association, Inc.
Brief Rapid Communications |
From the Division of Cardiovascular Diseases and Internal Medicine (M.B., K.T., D.J., J.B.S.) and the Department of Diagnostic Radiology (J.F.B.), Mayo Clinic, Rochester, Minn.
| Abstract |
|---|
|
|
|---|
Methods and ResultsWe collected a series of 6 to 11 apical echocardiographic tomograms, depending on heart rate, in 11 patients. There was good correlation, low variability, and low bias between rapid 3D echocardiography and electron-beam computed tomography for measuring left ventricular end-diastolic volume (r=0.96; standard error of the estimate, 21.34 mL; bias, -4.93 mL) and left ventricular end-systolic volume (r=0.96; standard error of the estimate, 14.78 mL; bias, -6.97 mL).
ConclusionsThe rapid-acquisition 3D echocardiography extends the use of a multiplane, internally rotating handheld transducer so that it becomes a precise and clinically feasible tool for assessing left ventricular volumes and function. A rapid-image acquisition time of 6 s would allow repeated image collection during the course of a clinical echocardiographic examination. Additional work must address rapid and automated data processing.
Key Words: echocardiography ventricles tomography
| Introduction |
|---|
|
|
|---|
Using the 180° continuous-rotation mode of the multiplane transducer, we developed a rapid 3D image-acquisition technique that addresses the issue of clinical practicality of 3D echocardiography.
The purpose of this study was to assess the clinical feasibility and the precision and accuracy of the new rapid, 3D echocardiographic acquisition technique for measuring LV end-diastolic (LVEDV) and end-systolic (LVESV) volumes. Electron-beam computed tomography (EBCT) was used as the reference standard.
| Methods |
|---|
|
|
|---|
3D Echocardiographic Study
We used a HP 2500 system
(Hewlett-Packard) fitted with a prototype
5-MHz multiplane transthoracic transducer. An internal
crystal array continuously rotates and collects images over 180°
about its imaging axis. The desired diastolic and
systolic tomograms are selected during subsequent off-line data
processing on the basis of a synchronously acquired
electrocardiographic signal. The number of tomograms obtained from the
6-s 180° rotation cycle depends on the patients heart rate. For
example, with a rate of 80 beats/min, 8 tomograms are obtained. The
end-diastolic (temporally related to the R wave on the ECG)
and end-systolic (smallest chamber volume during the cardiac
cycle) phases were identified at each cardiac cycle. Images were
digitized, and LV endocardial borders were traced manually, including
the LV outflow tract up to the aortic valve. The mitral valve plane was
traced as the straight line between the boundaries of the mitral
annulus. Two independent observers (K.T. and D.J.), who were blinded to
the LV volumes obtained with the reference EBCT measurements, performed
endocardial diastolic and systolic tracings, which
took about 20 minutes per left ventricle. The LVEDV and LVESV were
measured from computer-generated LV cavity casts, which were
reconstructed from the interactive tracings using the Sun
SPARC station 20 and a custom software algorithm; this took
another 15 minutes.
EBCT Study
Each EBCT (Imatron C-100) study was performed within
48 hours of the echocardiographic examination. Each
subject was positioned in the scanner to obtain parallel tomographic
images in the short axis (transverse cardiac) from the LV apex to the
base of the right ventricular outflow tract. During
imaging, an intravenous infusion of nonionic contrast
medium was delivered at 3 mL/s for 20 s. Eight to 12 parallel
tomographic scans were obtained from the LV apex through the base,
depending on the overall long-axis dimensions of the ventricle. The
endocardial borders were traced manually, and the LVEDV and LVESV were
obtained using a disk summation method.
Statistical Evaluation
To assess precision, LV volumes estimated from rapid
3D echocardiography by 2 independent observers were
averaged and compared with those measured with EBCT by linear
regression. Interobserver variability was expressed as the coefficient
of variation between the 2 observers. To determine whether the
difference in the values between the 2 methods was statistically
significant, a paired t test
was performed; the level of significance was set to
P<0.05. The accuracy of the
rapid 3D echocardiography with respect to the EBCT
measurements was examined by a limits-of-agreement analysis.
The bias was expressed as the mean difference between the 2 methods,
and the limits of agreement as 2 SDs of the difference of the 2
methods.
| Results |
|---|
|
|
|---|
|
The linear regression and agreement plots of corresponding
LVEDV and LVESV values estimated from the 3D
echocardiographic data by 2 independent observers are
shown in Figure 2
. These results showed excellent correlation and
only a small bias. In the estimation of the LVEDV and LVESV values,
interobserver variability was 6.11% and 9.14%,
respectively.
|
| Discussion |
|---|
|
|
|---|
Advantages of Rapid 3D Echocardiography
The technique allows the rapid collection of data
during 6 s of suspended respiration, which makes the technique
feasible in most clinical scenarios. The prototype transducer is
capable of all current Doppler and harmonic imaging modalities.
Rotational geometry has been applied successfully to 3D
reconstruction,8 9 10
and the approach discussed herein represents a practical
extension of these efforts. An important characteristic of this
geometry is that the images are equispaced and particularly useful for
reconstructing the whole left ventricle. In addition, this system could
sample sufficient 4D data for a dynamic analysis of LV function
and shape.
Limitations
The manual delineation and off-line data processing
were time-consuming; however, this was not an issue from the viewpoint
of our experimental objectives and did not prolong the time needed to
examine a patient. The present implementation of the method is
limited to patients with a heart rate
60 beats/min to collect a
minimum of 6 tomograms. Multiplane transducers equipped with an
adjustable rate of rotation would overcome this limitation.
The 5-MHz frequency of the transducer was not optimal for adult echocardiography but was dictated by the initial prototype design using transesophageal transducer mechanics. Consequently, limited signal penetration led to the exclusion of 2 patient data sets from analysis. A production system would certainly use a rotating transducer with lower frequency for transthoracic clinical applications.
| Conclusions |
|---|
|
|
|---|
| Acknowledgments |
|---|
| Footnotes |
|---|
Received March 16, 2001; revision received April 23, 2001; accepted April 30, 2001.
| References |
|---|
|
|
|---|
2. Takuma S, Swas DR, Fard A, et al. Real-time, 3-dimensional echocardiography acquires all standard 2-dimensional images from 2 volume sets: a clinical demonstration in 45 patients. J Am Soc Echocardiogr. 1999;12:16.[Medline] [Order article via Infotrieve]
3. Canals R, Lamarque G, Chatain P. Volumetric ultrasound system for left ventricle motion imaging. IEEE Trans Ultrason Ferroelectr Freq Control. 1999;46:15271538.[Medline] [Order article via Infotrieve]
4. Nelson TR, Pretorius DH. Three-dimensional ultrasound imaging. Ultrasound Med Biol. 1998;24:12431270.[Medline] [Order article via Infotrieve]
5. Gilja OH, Hausken T, Olafsson S, et al. In vitro evaluation of three-dimensional ultrasonography based on magnetic scanhead tracking. Ultrasound Med Biol. 1998;24:11611167.[Medline] [Order article via Infotrieve]
6. King DL, Harrison MR, King DL Jr, et al. Improved reproducibility of left atrial and left ventricular measurements by guided three-dimensional echocardiography. J Am Coll Cardiol. 1992;20:12381245.[Abstract]
7. Papavassiliou D, Doelling NR, Bowman MK, et al. Initial experience with an internally rotating transthoracic three-dimensional echocardiographic probe and image acquisition on a conventional echocardiogram machine. Echocardiography. 1998;15:369376.[Medline] [Order article via Infotrieve]
8. Belohlavek M, Foley DA, Gerber TC, et al. Three- and four-dimensional cardiovascular ultrasound imaging: a new era for echocardiography. Mayo Clin Proc. 1993;68:221240.[Medline] [Order article via Infotrieve]
9. Roelandt J, Salustri A, Mumm B, et al. Precordial three-dimensional echocardiography with a rotational imaging probe: methods and initial clinical experience. Echocardiography. 1995;12:243252.[Medline] [Order article via Infotrieve]
10. Yao J, Cao Q-L, Pandian NG, et al. Multiplane transthoracic echocardiography: Image orientation, anatomic correlation, and clinical experience with a prototype phased array multiplane surface probe. Echocardiography. 1997;14:559577.[Medline] [Order article via Infotrieve]
This article has been cited by other articles:
![]() |
N.A. M. Estes III Sudden Cardiac Arrest From Primary Electrical Diseases: Provoking Concealed Arrhythmogenic Syndromes Circulation, October 11, 2005; 112(15): 2220 - 2221. [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2001 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |