| Abstract |
|---|
|
|
|---|
Methods and
ResultsSerial volumetric intravascular
ultrasound analyses (poststent and follow-up) were performed in
55 lesions treated with a balloon-expandable stent (ACS MultiLink)
using standard stent deployment techniques. The vessel volume (VV),
lumen volume (LV), and volume bordered by the stent (SV) were measured
using Simpsons method. The volume of plaque and neointima
outside the stent (peri-stent volume, PSV) and volume of
neointima within the stent (intrastent volume) were also
measured. The change of each parameter during the follow-up
period (follow-up minus poststent) was calculated and then divided by
SV to normalize these values (designated as percent change [%
]).
As expected, %
PSV directly correlated with %
VV
(P<0.0001,
r=0.935), with no significant
SV. A highly significant inverse correlation was seen between
%
PSV and the percent change of intrastent volume
(P<0.0001,
r=0.517). Consequently, %
LV
significantly correlated with peri-stent remodeling, as measured by
%
VV (P<0.0001,
r=0.602).
ConclusionPositive remodeling of the vessel exterior to a coronary stent occurs to a variable degree after stent implantation. There is a distinct trade-off between positive remodeling and in-stent hyperplasia: in segments in which the degree of peri-stent remodeling is less, intrastent neointimal proliferation is greater and accompanied by more significant late lumen loss.
Key Words: coronary disease plaque remodeling stents restenosis
| Introduction |
|---|
|
|
|---|
| Methods |
|---|
|
|
|---|
IVUS Imaging
All patients were premedicated with aspirin and
received heparin (100 U/kg) before the procedures. Both poststent and
follow-up ultrasound studies were performed using a commercially
available imaging system with either a 3.2 F, 30 MHz or a 2.6 F, 40 MHz
mechanical ultrasound catheter (Boston Scientific Corporation). After
the intracoronary administration of 200 µg of
nitroglycerin, the imaging catheter was advanced distal
to the stented segment under fluoroscopic guidance. Using automated
pullback (0.5 mm/s), ultrasound images were obtained and
recorded on 0.5-inch. S-VHS videotape for off-line quantitative
analysis.
Quantitative IVUS Analysis
All measurements were performed by one individual who
was blinded to clinical and angiographic information. Three-dimensional
reconstruction of IVUS images was performed using a commercially
available quantitative analysis system, which runs on an Intel
Pentium-based PC system with Windows NT
(echoPlaque, Indec Systems, Inc). After digitization of IVUS
recordings at a frame rate of 15 images per second,
longitudinal views of the studied segments were automatically processed
by the system. Lumen, stent, and external elastic membrane areas were
manually traced at 16 frame intervals
(Figure 1
), and the interpolated measurements of the
remaining frames were automatically generated. Using Simpsons method,
lumen volume (LV), stent volume (SV), and vessel volume inside the
external elastic membrane (VV) were calculated, and the following
volumetric parameters were defined for analysis:
(1) intra-stent neointimal volume (ISV)=SV-LV and (2)
peri-stent plaque plus neointimal volume (PSV)=VV-SV. The
change of each parameter (
) was calculated as follow-up
minus poststent measurements. To normalize the data for vessel size,
these parameters were divided by the implanted SV in each
case (thus correcting for the fact that for the same percentage change
in neointima, the absolute volume change of each
parameter would be much greater in a large vessel than a
small vessel). Both the normalized (designated as percent change
[%
]) and absolute data are presented.
|
To assess the reproducibility of IVUS measurements, follow-up images of 10 cases were randomly selected and reanalyzed at least 4 weeks after the initial reading. The intraobserver correlation coefficients for VV and IV were 0.997 and 0.998, and the percent errors were 1.82±1.04% and 2.93±1.73%, respectively.
Statistical Analysis
Statistical analysis was performed using
StatView 4.5 (SAS Institute). Quantitative data,
presented as mean±SD, were compared using a 2-tailed, paired
Students t test or linear
regression analysis.
P<0.05 was considered
significant.
| Results |
|---|
|
|
|---|
%
LV correlated strongly with %
ISV by linear
regression analysis
(P<0.0001,
r=0.943), which is
consistent with the observation that intrastent
neointimal growth was the mechanism of LV reduction, as
reported previously.3 A
strong inverse correlation was seen between normalized %
ISV and
%
PSV (P<0.0001,
r=0.517;
Figure 2
). A significant correlation was also noted between
absolute
ISV and
PSV
(P<0.01,
r=0.346). As expected, %
VV
directly correlated with %
PSV
(P<0.0001,
r=0.935). Accordingly, %
LV
was inversely correlated with %
PSV
(P<0.0001,
r=0.505) and, therefore, it
correlated with %
VV
(P<0.0001,
r=0.602).
|
Regarding the potential effect of the stent implantation technique, there was only a weak positive correlation between the maximum inflation pressure and peri-stent positive remodeling (P<0.05, r=0.33). No significant correlation was observed between the maximum inflation pressure and neointimal growth inside the stent.
| Discussion |
|---|
|
|
|---|
Coronary artery remodeling is a major factor contributing to restenosis after balloon angioplasty and atherectomy.1 2 The existence of positive, peri-stent arterial remodeling has been suggested in a study of first-generation stents5 and in a trial of stenting plus brachytherapy6 However, whether or not there is an association between positive peri-stent remodeling and intrastent neointimal proliferation has remained controversial. In their recent study using second-generation stents, Kay et al6 suggested that positive remodeling occurs in radiated segments but not in normally stented segments. The present study, involving a single current-generation stent and complete volumetric IVUS analysis, showed a significant inverse correlation between peri-stent remodeling and intrastent neointimal proliferation in normal (nonirradiated) segments. This observation provides a new perspective on some other important issues in restenosis. The link between remodeling and in-stent restenosis is consistent with the fact that patients with impaired glucose tolerance, who are known to have a blunted adaptive remodeling response, show exaggerated intimal proliferation after stenting.7 8 9 The trade-off between positive remodeling and intimal proliferation could also help explain the beneficial effect of prestent debulking on restenosis for some patients. Previous studies have shown that a proportion of atherectomy cases have significant positive remodeling,2 perhaps associated with weakening or excision of external lamina and other support structures in the arterial wall. Finally, as Kay et al6 have suggested, the impact of brachytherapy on intimal proliferation after stenting may also be due in substantial part to the effects of this treatment on positive remodeling.
The results of the current study clearly demonstrate the relationship between neointimal growth and vessel remodeling after conventional stent implantation. An important implication of this study is that the search for treatments of in-stent restenosis should include approaches to facilitate positive remodeling in the stented segment.
Study Limitations
First, this study is based on a relatively small,
retrospectively selected patient population in whom a single stent was
used, raising the possibility of selection bias. Second, in general,
vessel dimensions beyond metallic stents may be obscured by struts and
thus may not reflect the true vessel size of the stented segment. In
this study, however, we included only the cases treated with a single
MultiLink stent, the design of which permits fairly even transmission
of the ultrasound beam around the stent struts. Furthermore, we
excluded the cases in which complete tracking of the peri-stent vessel
boundary was not available throughout the stented segment. These strict
criteria for study patients minimized the influence of stent struts on
the measurement of peri-stent vessel dimensions. Third, the current
study does not address the extent of vessel remodeling beyond the stent
edge to the adjacent reference segments. Also, the volumetric
analyses used in this study may not represent well
localized vessel responses in a specific subsegment. Therefore, further
study to focus on these issues is needed. Finally, there are several
inherent limitations in 3D analysis techniques, including the
effects of movement artifacts and distortion due to curvature of the
vessels.10
| Acknowledgments |
|---|
Received December 29, 2000; revision received February 28, 2001; accepted March 8, 2001.
| References |
|---|
|
|
|---|
2.
Lansky AJ, Mintz
GS, Popma JJ, et al. Remodeling after directional coronary
atherectomy (with and without adjunct percutaneous
transluminal coronary angioplasty): a serial angiographic and
intravascular ultrasound analysis from the Optimal Atherectomy
Restenosis Study. J Am Coll
Cardiol. 1998;32:329337.
3.
Hoffmann R, Mintz
GS, Dussaillant GR, et al. Patterns and mechanisms of in-stent
restenosis: a serial intravascular ultrasound study.
Circulation. 1996;94:12471254.
4.
Komatsu R, Ueda M,
Naruko T, et al. Neointimal tissue response at sites of
coronary stenting in humans: macroscopic,
histological, and immunohistochemical analyses.
Circulation. 1998;98:224233.
5. Hoffmann R, Mintz GS, Popma JJ, et al. Chronic arterial responses to stent implantation: a serial intravascular ultrasound analysis of Palmaz-Schatz stents in native coronary arteries. J Am Coll Cardiol. 1996;28:11341139.[Abstract]
6.
Kay IP, Sabate M,
Costa MA, et al. Positive geometric vascular remodeling is seen after
catheter-based radiation followed by conventional stent implantation
but not after radioactive stent implantation.
Circulation. 2000;102:14341439.
7.
Kornowski R, Mintz
GS, Kent KM, et al. Increased restenosis in diabetes mellitus
after coronary interventions is due to exaggerated intimal
hyperplasia: a serial intravascular ultrasound study.
Circulation. 1997;95:13661369.
8. Kornowski R, Mintz GS, Lansky AJ, et al. Paradoxic decreases in atherosclerotic plaque mass in insulin-treated diabetic patients. Am J Cardiol. 1998;81:12981304.[Medline] [Order article via Infotrieve]
9.
Takagi T, Yoshida
K, Akasaka T, et al. Hyperinsulinemia during oral
glucose tolerance test is associated with increased
neointimal tissue proliferation after coronary
stent implantation in nondiabetic patients: a serial intravascular
ultrasound study. J Am Coll
Cardiol. 2000;36:731738.
10.
de Vrey EA, Mintz
GS, von Birgelen C, et al. Serial volumetric (three-dimensional)
intravascular ultrasound analysis of restenosis after
directional coronary atherectomy.
J Am Coll Cardiol. 1998;32:18741880.
This article has been cited by other articles:
![]() |
H-J Kang, Y-S Kim, B-K Koo, K W Park, H-Y Lee, D-W Sohn, B-H Oh, Y-B Park, and H-S Kim Effects of stem cell therapy with G-CSF on coronary artery after drug-eluting stent implantation in patients with acute myocardial infarction Heart, May 1, 2008; 94(5): 604 - 609. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Okura, H. Taguchi, T. Kubo, I. Toda, M. Yoshiyama, J. Yoshikawa, and K. Yoshida Impact of arterial remodelling and plaque rupture on target and non-target lesion revascularisation after stent implantation in patients with acute coronary syndrome: an intravascular ultrasound study Heart, October 1, 2007; 93(10): 1219 - 1225. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Aoki, A. Colombo, D. Dudek, A. P. Banning, J. Drzewiecki, K. Zmudka, F. Schiele, M. E. Russell, J. Koglin, P. W. Serruys, et al. Peristent Remodeling and Neointimal Suppression 2 Years After Polymer-Based, Paclitaxel-Eluting Stent Implantation: Insights From Serial Intravascular Ultrasound Analysis in the TAXUS II Study Circulation, December 20, 2005; 112(25): 3876 - 3883. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Tanabe, P. W. Serruys, M. Degertekin, E. Grube, G. Guagliumi, W. Urbaszek, J. Bonnier, J.-M. Lablanche, T. Siminiak, J. Nordrehaug, et al. Incomplete Stent Apposition After Implantation of Paclitaxel-Eluting Stents or Bare Metal Stents: Insights From the Randomized TAXUS II Trial Circulation, February 22, 2005; 111(7): 900 - 905. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Tanabe, P. W. Serruys, M. Degertekin, G. Guagliumi, E. Grube, C. Chan, T. Munzel, J. Belardi, W. Ruzyllo, L. Bilodeau, et al. Chronic Arterial Responses to Polymer-Controlled Paclitaxel-Eluting Stents: Comparison With Bare Metal Stents by Serial Intravascular Ultrasound Analyses: Data From the Randomized TAXUS-II Trial Circulation, January 20, 2004; 109(2): 196 - 200. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. F. Fearon, M. Nakamura, D. P. Lee, M. Rezaee, R. H. Vagelos, S. A. Hunt, P. J. Fitzgerald, P. G. Yock, and A. C. Yeung Simultaneous Assessment of Fractional and Coronary Flow Reserves in Cardiac Transplant Recipients: Physiologic Investigation for Transplant Arteriopathy (PITA Study) Circulation, September 30, 2003; 108(13): 1605 - 1610. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. H. Stone, A. U. Coskun, S. Kinlay, M. E. Clark, M. Sonka, A. Wahle, O. J. Ilegbusi, Y. Yeghiazarians, J. J. Popma, J. Orav, et al. Effect of Endothelial Shear Stress on the Progression of Coronary Artery Disease, Vascular Remodeling, and In-Stent Restenosis in Humans: In Vivo 6-Month Follow-Up Study Circulation, July 29, 2003; 108(4): 438 - 444. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. S. Mintz, V. M. Shah, and N. J. Weissman Regional Remodeling as the Cause of Late Stent Malapposition Circulation, June 3, 2003; 107(21): 2660 - 2663. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Spanos, G. Stankovic, J. Tobis, and A. Colombo The challenge of in-stent restenosis: insights from intravascular ultrasound Eur. Heart J., January 2, 2003; 24(2): 138 - 150. [Full Text] [PDF] |
||||
![]() |
V. M. Shah, G. S. Mintz, S. Apple, and N. J. Weissman Background Incidence of Late Malapposition After Bare-Metal Stent Implantation Circulation, October 1, 2002; 106(14): 1753 - 1755. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2001 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |