(Circulation. 2001;103:1793.)
© 2001 American Heart Association, Inc.
Basic Science Reports |
From the Krannert Institute of Cardiology (H.-P.S., A.T.N., D.H., K.L.M.); the PET Facility, Department of Radiology, Indiana University School of Medicine (G.D.H., W.L.W., C.R.A.); and the Richard L. Roudebush Veterans Administration Medical Center (K.L.M.), Indianapolis, Ind; and the Department of Neuropathology, University Hospital, Homburg/Saar, Germany (B.R.).
Correspondence to Hans-Peter Stoll, MD, Innere Medizin III, Kardiologie, Universitätskliniken des Saarlandes, 66421 Homburg/Saar, Germany. E-mail inhsto{at}med-rz.uni-saarland.de
| Abstract |
|---|
|
|
|---|
Methods and ResultsThe feasibility, efficacy, and safety of 68Ga liquid-filled balloon brachytherapy were investigated in the porcine coronary overstretch model. Four groups of 5 balloon-induced coronary lesions were irradiated with 8, 12, 16, and 24 Gy targeted to the adventitia. Ten unirradiated lesions served as controls. Segments treated with 16 or 24 Gy exhibited marked suppression of neointimal proliferation at 28-day follow-up, with quantitative parameters of intraluminal proliferation reduced to <20%. This beneficial effect was not compromised by untoward edge effects. Uninjured but irradiated vessels did not show histological signs of radiation damage. The 68Ga whole-body dose due to balloon rupture was estimated to be 5 rem/50 mCi treatment activity and compared favorably with that of 188Re (78 rem/50 mCi).
Conclusions68Ga positron radiation suppresses neointimal proliferation at doses of 16 and 24 Gy. This biological efficacy, coupled with the attractive safety profile, suggests the selection of 68Ga as an attractive isotope for liquid-filled balloon brachytherapy.
Key Words: angioplasty restenosis radioisotopes
| Introduction |
|---|
|
|
|---|
We hypothesized that the use of positron emitters with half-lives <2 hours might overcome the safety problems and performed initial phantom experiments and cell-culture studies (H.-P.S., unpublished data, 1998) to investigate the tissue penetration and antiproliferative efficacy of positron radiation emitted by 18F, 11C, 13N, 15O, and 68Ga. Generator-available 68Ga, with a half-life of only 68 minutes, appeared to be the most attractive candidate among these isotopes.
Consequently, the goal of this study was to assess the antirestenotic efficacy of 68Ga in a porcine coronary model and to compare the remaining risk of this approach with that of previously suggested rhenium isotopes.
| Methods |
|---|
|
|
|---|
15 minutes and was performed by a technician
of the cyclotron laboratory serving the clinical PET
facility. The concentrated generator eluate was filled into balloon catheters (ACS Lifestream RX 2.5/30 mm or RX 3.0/30 mm, Guidant Corp) that were used as intracoronary radiation sources providing a 30-mm-long dose field. The useful in-balloon volume was 0.15 to 0.22 mL, and the total filling volume, including dead space, was 0.35 to 0.44 mL. The Lifestream perfusion balloons were selected because their design allows maintenance of antegrade blood flow during radiation delivery. The balloon length of 30 mm was chosen to ensure that the treatment balloons exceeded the length of the injury (created with 20-mm-long balloons) on both sides by 5 mm.
Dosimetry
68Ga (half-life 68
minutes) deploys a therapeutic tissue dose through positrons of a
maximum energy of 2.9 MeV. Accurate dosage requires knowledge of the
dose deployed per disintegration of the isotope for a given balloon
geometry and a given distance from the balloon surface. These constants
were determined in a preceding series of phantom measurements using
radiosensitive film. On the basis of these values, a computer program
was written to facilitate intraprocedural dosimetry and to free the
operators attention for medical aspects. Once the initial in-balloon
activity was entered, this program continuously calculated the
remaining in-balloon activity. Before irradiation, the software
informed the operator about the expected dwell time. During
irradiation, the accumulated dose and remaining dwell time were
displayed, and most importantly, the program called for source removal
once the prescribed dose was reached.
Animal Experiments in the Porcine Model
Anesthesia
All experiments were conducted in accordance with the
guidelines for animal experiments set forth by Indiana University.
Seventeen juvenile farm pigs weighing 23 to 25 kg were premedicated
with intramuscular ketamine (33 mg/kg), acepromazine (0.25 mg/kg), and
atropine (0.5 mg). Anesthesia was induced by pentothal (25 mg/kg IV),
and animals were intubated and mechanically ventilated with 2.0 to 3.0
vol% isoflurane and 40 vol% oxygen. Animals were kept on aspirin (320
mg/d) throughout the entire study.
Coronary Imaging, Creation of a Lesion, and
Intracoronary Irradiation
The coronary system was imaged by standard
angiographic technique and a right carotid approach. Coronary artery
diameters were measured with the shape of the guiding catheter as
reference for calibration. Two coronary sites per pig, located in the
left anterior descending and the left circumflex arteries, were
overstretched with 20-mm-long standard balloons for angioplasty and a
1.3:1 overstretch ratio3 and
subsequently irradiated.
Fluid ports of treatment balloon catheters were connected to a 3-way stopcock, which in turn was hooked up to a large-volume indeflator (20 mL, Guidant Corp). Balloons were carefully evacuated with 3 strokes of the indeflator until a strong negative pressure was reached. A 2.0-mL syringe containing 0.5 mL of a radioactive 68Ga solution with air bubbles removed and the meniscus carefully advanced to the distal outlet of the Luer-lock connector was attached to the free arm of the 3-way stopcock. The lever was then switched to allow for free fluid communication between the balloon and the 68Ga syringe such that the radioactive fluid was immediately advanced into the balloon by negative pressure forces. This procedure was found to be particularly helpful for avoiding air-bubble formation, even in case of repeated inflations. The balloon was pressurized outside the animal to nominal pressure (8 bar) with air through the indeflator and inserted into a dose calibrator (Atomlab300, Biodex Medical Systems) for measurement of in-balloon activity, which was then entered into the dosimetry program. The treatment balloon was subsequently deflated without removal of the liquid radioactivity, advanced to the desired coronary treatment site, and reinflated to nominal pressure. Once the dosimetry program had called for source removal, the balloon was deflated and retracted, and the neck wound was closed. A prophylactic antibiotic dose of 1000 mg cephazolin IM was given, and the animal was allowed to recover. Catheters, syringes, and tubing that were in contact with the radioactive [68Ga]GaCl3 solution were stored in a lead-shielded radioactive waste container for decay and put into the regular waste after a 3-day decay period.
Treatment Groups
Ten overstretched but not irradiated segments served
as control (group I). A total of 20 segments were randomly selected to
receive doses of 8 (group II), 12 (group III), 16 (group IV), or 24 Gy
(group V). Thus, each treatment group contained 5 segments. All doses
were directed to a prescription point 1 mm beyond the intimal surface,
deep enough to target adventitial cell populations whose key role in
triggering the restenotic response to vessel injury was recently
postulated.4 5 6
Euthanasia and Sample Preparation
Twenty-eight days after irradiation, a preeuthanasia
angiography was performed, animals were euthanized, the hearts were
excised, and pressure fixation was performed with 10% buffered
formalin. The left anterior descending and left circumflex arteries
were sectioned into 3-mm-long segments, embedded in paraffin, and cut
into 4-µm-thick cross-sectional slices, which were stained for
microscopic evaluation with hematoxylin-eosin and Verhoeffvan Gieson
staining.
Sample Analysis
We analyzed the treated vessel segments under 25-fold
magnification (ocular x objective) with a light microscope (Leitz
Wetzlar) equipped with a video camera and an image digitizer. Luminal
circumference, luminal area, media fracture length, external elastic
lamina (EEL) circumference, EEL area, adventitial circumference, and
adventitial area were measured with the NIH image program. Neointimal
area (IA), IA/fracture length ratio (IA/FL), and maximal intimal
thickness (MIT) were used to characterize the restenotic
response.3 The parameters
characterizing each analyzed artery were calculated by averaging the
data from all segmental cross sections exhibiting an intimal lesion
with disruption of the internal elastic lamina
(IEL).
Operator Doses
Operator dose increments due to
68Ga positron annihilation radiation were
modeled assuming simplified radiation geometry. Radiation was taken to
originate from a point source 5 cm deep inside the chest; operator
distance from the patients skin was taken as 1 m, and complete
positron absorption within the patients body was assumed. Similar
calculations were performed for 188Re.
X-rayinduced neck-collar doses reported for cardiologists performing
a typical diagnostic and interventional
procedures7 were used for
comparison.
Another calculation was done to estimate operator and personnel exposure after a spill of 68Ga in the catheter laboratory. A loss of 10 mCi (20% of in-balloon activity) was taken, and it was assumed that 50% were successfully decontaminated. Positron radiation was further assumed to be completely absorbed by appropriate Plexiglas shielding, and the exposure rate due to 511 keV annihilation radiation was calculated at a distance of 30 cm from the source.
Statistics
Continuous parameters in
Tables 1
and 2
are given as mean±SEM.
Histomorphometric parameters for individual arteries were calculated by
averaging values from all cross sections of this artery that showed
signs of injury. Parameters given in
Table 2
were finally calculated for each individual dose
group by averaging results from individual arteries. Differences
between the dose groups were analyzed by 1-way ANOVA. Differences
between the histomorphometric measures were assessed by comparison of
each dose group versus control by
t test analysis. In case of
failure of the data to meet t
test premise criteria, data were compared by the Mann-Whitney rank sum
test. All statistical calculations were performed with the SigmaStat
software package (SPSS, Inc).
|
|
| Results |
|---|
|
|
|---|
Table 1
shows radiologically measured preoverstretch vessel
diameters and the selected balloon/artery ratios together with the
obtained fracture lengths to characterize the created lesions. None of
the input parameters exhibited a statistically significant difference
between the 5 groups.
Histological Results
Representative photomicrographs characterizing the
vessel segments treated with each of the different doses are given in
Figure 1
. Unirradiated control segments (A) were
characterized by transmedial tears filled out by cellular tissue
encroaching on the lumen. The EEL region surrounding the tear site
occupied an expanded area and contained an increased amount of
collagen, fragmented elastic fibers, and fibroblast-like cells.
Segments irradiated with 8 Gy (B) exhibited additional enlargement of
intraluminal proliferative tissue for any particular tear length. A
significant perivascular fibrosis was expressed over the lesion site,
including vasa vasorum and cardiac nerves (indicated by an arrow).
Segments irradiated with 12 Gy exhibited less neointimal tissue bulk
than seen in the 8-Gy group and a trend to less than in the control
group. The perivascular fibrosis was similar to that found in the 8-Gy
group. Segments irradiated with 16 and 24 Gy were markedly different,
characterized by a nearly complete absence of intraluminal
proliferation. There was neither a clear repair of transmural tears nor
luminal narrowing. The perivascular fibrosis remained similar to that
seen at lower doses, with similar amounts of collagen and elastic
fibers (D and E). Uninjured segments (F) treated with doses of 8 and 16
Gy did not show any evidence of radiation damage. In particular, there
was no focal necrosis, no inflammatory reaction, no augmented collagen
deposition, and no increased density of elastic fibers, nor was a
reduced cell count observed in the media. Taking those criteria
together, the uninjured segments were morphologically equal to normal
porcine coronary arteries.
|
One artery in the 24-Gy group showed a thin unresolved mural thrombus, and 1 artery in the 12-Gy group showed signs of an unresolved intramural hemorrhage. All irradiated segments were angiographically patent at 28 days.
Histomorphometric Results
Morphometric analysis of injured vessels revealed that
the neointimal response was proportional to the degree of injury in the
control, 8-Gy, and 12-Gy groups, as previously described. A marked and
statistically significant suppression of all parameters of neointimal
volume (IA, IA/FL, and MIT) down to 20% of control values was
observed, however, in the 16- and 24-Gy groups
(Table 2
,
Figure 2
), such that no relationship to the degree of injury
was detectable. Each vessel in these groups displayed multiple segments
with complete suppression of measurable neointima formation, whereas
other segments showed a small amount of neointimal tissue. No
stimulated neointimal response or less effective neointima suppression
on the edges of the lesions was observed in either high-dose group
histologically or in the preeuthanasia angiograms. The group receiving
12 Gy showed a nonsignificant decrease in the neointimal parameters to
75% of control. The group receiving 8 Gy was associated with a
trend toward increased neointimal formation.
|
Determination of arterial remodeling by analysis of EEL circumference at the injury sites did not reveal any significant differences between irradiated and nonirradiated arteries or between arteries that had received different radiation doses.
Operator Exposure
Because of our model calculations, operator exposure
rate increments due to brachytherapy were 0.6 mrem for
68Ga and 0.2 mrem for
188Re, assuming in-balloon activities of 50
mCi required to deploy a 24-Gy dose within 4 minutes. For comparison,
doses accruing because of x-ray exposure were reported to be 5 to 16
mrem for a typical procedure, including diagnostic coronary angiography
and single-vessel
intervention.7
The exposure rate for operator and personnel after a spill of 68Ga was estimated to be 3.0 mR · h-1 · mCi-1 at a distance of 30 cm from the radioactive spot once positron radiation was absorbed by appropriate shielding. Thus, an initial dose rate of 15 mR/h would result from a spill of 5 mCi. After 3.5 hours, the remaining dose rate would fall below the threshold of 2 mR/h because of the rapid isotope decay.
| Discussion |
|---|
|
|
|---|
17
hours.
Accidental radiointoxication of the patient may occur
because of a loss of liquid radioactivity into the circulation. The
hazard of such an event is comparatively small with
68Ga, because resulting whole-body and organ
doses are low.
Table 3
specifies such doses calculated from tabulated
doses per unit of activity estimated by the International Commission on
Radiological Protection (ICRP, publication No. 53). These data are
based on pharmacokinetic measurements in patients after systemic
delivery of various gallium
isotopes.8 Remarkably, the
effective 68Ga whole-body dose of 5.0 rem is
>15 times lower than the life-threatening dose of 78 rem caused by a
ruptured balloon loaded with the same 188Re
or 186Re
activity.9 The low
68Ga doses also compare favorably with
x-rayinduced patient doses accruing during an interventional
procedure7 and are far below
the threshold of 50 rem, above which physiological changes due to acute
radiation exposure have been described.
|
Accidental contamination of the laboratory is another important concern raised against the liquid-filled balloon technique. Because of current radiation safety regulations, an area contaminated with 68Ga would need to be decontaminated, covered with plastic foil, shielded with Plexiglas, and posted as a "radiation area" once the dose rate at 30 cm distance exceeded 2 mR/h. Our estimates for a realistic spill of 5 mCi indicate that such shielding and posting needed to be maintained for only 3.5 hours to ensure that personnel would actively keep the maximal possible distance away from the spill zone. Accordingly, we consider the low patient doses after balloon rupture and the moderate radiation safety measures after a spill to be the most pronounced advantages that make 68Ga safer than any other liquid radioisotope previously proposed for brachytherapy.
Another concern has been that the 511-keV radiation associated with positron annihilation might lead to high operator exposure rates. Estimated dose values indicate, however, that fluoroscopy and cineangiography will in fact cause the vast majority of the operator radiation burden during any application of this approach.
On the basis of the enhanced safety profile of
68Ga, our study focused on ascertaining the
antirestenosis efficacy of this isotope in the porcine model. Marked
suppression of neointima formation at doses of 16 and 24 Gy is the
major treatment effect of this approach. Thus, this study confirms that
exposure of vascular tissue to positron radiation indeed yields the
same favorable result already gained in earlier experimental studies
with
- and ß-radiation.
The histological results seen with the 8-Gy dose
highlight an important side effect of this (and other) brachytherapy
approaches: the possibility of stimulation of intraluminal
proliferation at low doses. We believe that (1) therapeutic exposures
should provide
16 Gy at 1-mm depth; (2) penetration of radiation into
the depth of the adventitia may be important not only for targeting
restenosis-triggering cell
populations4 5 6
but also for avoiding low doses in the immediate neighborhood of the
vessel; and (3) the dose field should safely overlap both ends of the
lesion in the longitudinal vessel axis to avoid the possibility that
submaximal doses hit marginal portions of the lesion. On the basis of
these considerations, we intentionally used irradiation balloons longer
than the overstretch injury (30 versus 20 mm) and have not observed
untoward edge effects. The rationale for this approach is based on
another important observation of this study, which is that uninjured
but irradiated vessels did not exhibit histological abnormalities or
additional intraluminal proliferation at doses of 8 or 16 Gy after 28
days
(Figure 1F
).
Perivascular fibrosis is observed in irradiated vessels at
28 days
(Figure 1
), characterized by formation of collagen and
elastic fibers. This fibrosis is consistently noted over the lesion
site, and its presence would suggest a chronic constrictive effect. No
such constriction was observed, however, either in these experiments or
in a larger cohort of juvenile swine investigated and characterized in
an earlier
investigation.3
In conclusion, liquid 68Ga contained within angioplasty balloons suppresses neointimal proliferation at doses of 16 and 24 Gy in the porcine model, and the severity of patient radiointoxication after potential intravascular balloon rupture is significantly reduced in comparison with earlier suggested ß-emitters. Thus, we recommend 68Ga as an isotope of attractive choice for further clinical investigation of liquid-filled balloon brachytherapy.