(Circulation. 2000;102:I-95.)
© 2000 American Heart Association, Inc.
ECC Guidelines |
| Oxygenation Devices |
|---|
|
|
|---|
In patients with acute MI, supplemental oxygen reduces both the magnitude and the extent of ST-segment changes on the ECG. We recommend oxygen administered at 4 L/min by nasal cannula for the first 2 to 3 hours for all patients with suspected acute coronary syndromes (Class IIa). The use of oxygen beyond 3 to 6 hours is indicated for patients with continuing or recurrent ischemia, complicated infarcts with congestive heart failure, or arrhythmia until hypoxemia has resolved and the patient is clinically stable.
| Ventilatory Devices |
|---|
|
|
|---|
An adequate seal is best achieved with a mouth-to-mask device when the
rescuer is positioned at the top of the patients head (Figure 1
). The rescuer ventilates the victim by
sealing his or her lips around the coupling adapter of the mask. Use
both hands to hold the mask securely in position and maintain airway
patency with head tilt. Manikin practice with masks should be required
of all personnel who are likely to use a mask for mouth-to-mask
ventilation.
|
Bag-Valve Devices
Bag-valve devices consist of a bag (self-inflating) and a valve
(nonrebreathing). They may be used with a mask, a tracheal tube, or
other alternative airway adjuncts. Most commercially available adult
bag-mask units have a volume of approximately 1600 mL. This volume is
much greater than currently recommended tidal volumes for CPR (10
mL/kg, 700 to 1000 mL). When the airway is unsecured (as with a mask
versus a tracheal tube), the possibility of overventilation with
gastric inflation, subsequent regurgitation, and
aspiration becomes a significant concern. In several studies many
rescuers were able to deliver adequate tidal volumes (6 to 7 mL/kg,
approximately 500 mL) with a bag-valve and mask to unintubated manikins
(Class IIa).2
To optimize bag-valve and mask performance, one rescuer must be positioned at the top of the victims head. Generally an oral airway should be inserted (see below) and, if possible, the head elevated if no concern for neck injury exists. While the head is maintained in position, deliver the selected tidal volume (preferably 6 to 7 mL/kg) over 2 seconds.3 Slow, gentle ventilation minimizes risk of gastric inflation. A bag-valve device may be used with any airway adjunct, such as tracheal tube, laryngeal mask airway, or esophageal-tracheal Combitube. Proper use of these combinations requires training, practice, and demonstrated proficiency.
A satisfactory bag-valve unit should have (1) a self-refilling bag, (2)
a nonjam valve system allowing for a maximum oxygen inlet flow of 30
L/min, (3) a nopop-off valve, (4) standard 15-/22-mm fittings, (5) a
system for delivering high concentrations of oxygen through an
ancillary oxygen reservoir, (6) a true nonrebreathing valve, and (7)
the capability to perform satisfactorily under all common environmental
conditions and extremes of temperature (Figure 2
).
|
| Automatic Transport Ventilators |
|---|
|
|
|---|
A number of ATVs are commercially available.5 6 7 8 9 Studies comparing ATVs with self-inflating bag-ventilation devices during intrahospital transport show that both devices can maintain a satisfactory minute ventilation and appropriate arterial blood gas exchange.10 11 12 13 Bag-ventilation devices are accurate only when tidal volume and minute ventilation are constantly monitored, an impractical approach in prehospital care.14 Although not as accurate, ATVs remain effective without measures of tidal volume and minute ventilation.
Studies have also revealed that ATVs are as effective as other devices used in prehospital care in intubated patients.5 15 In addition, studies on mechanical models and in animals demonstrate the superiority of ATVs for ventilating unintubated patients in respiratory arrest.16 Further studies evaluating the use of these devices are warranted. At the present time, ATVs are considered to provide advantages over alternative methods of ventilation:
Studies have observed improved lung inflation or absent gastric inflation when ATVs were compared with other devices, including mouth-to-mask, bag-mask, and manually triggered devices.12 13 This is due to the lower inspiratory flow and longer inspiratory times (2 seconds) provided by ATVs.
Disadvantages of ATVs include the need for an oxygen source and electric power. In addition, some ATVs may be inappropriate for children under 5 years of age. Because most ATVs require either an oxygen source or an electric power source, a self-inflating bag-valve device or a simple mask should always be available in case the oxygen source is depleted, or no oxygen or electric power source is available, or the ventilator malfunctions.
ATVs used for prehospital care should be simple and time- or volume-cycled. Avoid pressure-cycled devices.12 Delivered tidal volumes should be relatively unaffected by changes in lung-thorax impedance (<10% change).7 Operational gas consumption should be <5 L/min. ATVs should have the following minimum features:
4 kg), compact, rugged design, with a carrying or
mounting bracket A demand-flow valve may be incorporated into the ATV to reduce the work of breathing should spontaneous breathing return. The valve should be able to deliver a peak inspiratory flow rate of at least 120 L/min (2 L/s) on demand. The pressure to trigger spontaneous flow should not exceed -1 to -2 cm H2O.
Some ATVs allow higher preselected ventilator breathing rates. During CPR use caution in selecting ventilation rates more frequent than 10 per minute in adults or 20 per minute in children, because adequate time for exhalation is necessary to prevent air trapping and a positive end-expiratory pressure (auto-PEEP) effect. Auto-PEEP may reduce forward blood flow (ie, effective cardiac output) because pulmonary perfusion pressures are very low during CPR. Pulmonary capillary flow is easily impeded by high alveolar pressure. An appropriate exhalation time to maintain a 1:2 inhalation-to-exhalation time (I:E) ratio is necessary to minimize air trapping.17
Additional desirable features include a pressure manometer, provision for PEEP in more sophisticated ventilators, at least two controls (one for rate and one for tidal volume), and alarms to indicate depletion of oxygen cylinders, ventilator disconnect, or low battery.
Directors of prehospital or transport programs need to ensure that ATVs are used only by personnel who have received adequate ATV training. Monitoring of use and complication rates is essential to ensure safe and effective use of ATVs.
Oxygen-Powered, Manually Triggered Devices
Oxygen-powered, manually triggered devices have been used in
prehospital care for >25 years despite a paucity of high-level
scientific evidence supporting their use. When used with a face mask,
high inspiratory flow and pressure may cause massive gastric inflation.
When used with other airway adjuncts, high flow and pressure may cause
barotrauma. In an effort to limit the damage caused by these devices,
in 1986 a recommendation was made to limit flow to 40 L/min.
Oxygen-powered manually triggered devices are not recommended at this
time (Class Indeterminate). Further in vivo studies are needed to
compare their efficacy with that of bag-valve devices and ATVs.
| Airway Adjuncts |
|---|
|
|
|---|
Nasopharyngeal Airways
Nasopharyngeal airways are especially useful in patients with
trismus, biting, clenched jaws, or maxillofacial injuries, which
prevent placement of an oral airway (Class IIa). They should be used
with caution in patients with suspected fracture at the base of the
skull. In patients who are not deeply unconscious, nasopharyngeal
airways are better tolerated than oropharyngeal ones. Insertion can
cause damage to the nasal mucosa, resulting in bleeding. If the tube is
too long, it may stimulate the laryngeal or glossopharyngeal reflexes
to produce laryngospasm, retching, or vomiting. As with all adjunctive
equipment, safe use of nasopharyngeal airways requires adequate
training, practice, and retraining.
Alternative Airways
In some communities tracheal intubation is not permitted, or
patients are so few that practitioners obtain little
experience. Alternative airways that require blind passage of the
device into the airway may be simpler to master than passage of a
tracheal tube under direct vision. Alternative airways include the
laryngeal mask airway (LMA), the esophageal-tracheal Combitube (ETC),
and the pharyngotracheal lumen airway (PTL). When used by adequately
trained healthcare providers, the LMA and the ETC provide superior
ventilation compared with face masks in patients in cardiac arrest
(Class IIa). To achieve good outcomes with these devices, healthcare
providers must maintain a high level of knowledge and skills through
frequent practice and field use.
Esophageal-Tracheal Combitube
The ETC is an invasive double-lumen airway with 2 inflatable
balloon cuffs that is inserted without visualization of the vocal
cords. Assessment of the location of the distal orifice18
is then made, and the patient is ventilated through the appropriate
opening. One lumen contains ventilating side holes at the
hypopharyngeal level and is closed at the distal end; the other lumen
has a distal open end with a cuff similar to a tracheal tube. When
inflated the large pharyngeal balloon fills the space between the base
of the tongue and the soft palate, anchoring the ETC into position, and
isolates the oropharynx from the hypopharynx. The tube most commonly
finds its way into the esophagus because of the stiffness and curve of
the tube and the shape and structure of the pharynx. The tube is
advanced until the patients teeth lie between 2 marks printed on the
tube. The pharyngeal and distal balloons are then inflated, thus
isolating the oropharynx above the upper balloon and the esophagus (or
trachea) below the lower balloon.
The advantages of the ETC over the face mask are similar to those of the tracheal tube over the face mask: isolation of the airway, reduction in the risk of aspiration, and more reliable ventilation. The advantages of the ETC over the tracheal tube relate chiefly to ease of training and maintenance of placement skills, because laryngoscopy and visualization of the vocal cords are not necessary for insertion of the ETC. Ventilation and oxygenation with the ETC compare favorably with those achieved with the tracheal tube. Successful insertion rates with the ETC range from 69% to 100%.19 20 21 22 23 24 25 Because successful insertion is not ensured, providers should have a strategy for airway management when they are unable to ventilate with their first-choice adjunct. Fatal complications with the ETC may occur if the position of the distal lumen of the ETC in the esophagus or the trachea is identified incorrectly. In one EMS system a retrospective review reported that the incorrect port was used for ventilation in 3.5% of cases.20 For this reason use the ETC in conjunction with an end-tidal CO2 or esophageal detector device.26 27
Another possible complication from the ETC is esophageal trauma.28 Eight cases of subcutaneous emphysema were retrieved from a retrospective review of 1139 patients resuscitated with the ETC by emergency medical technicians. Four patients underwent autopsy, and 2 were found to have esophageal lacerations.29 To optimize insertion rates and to minimize complications, providers should receive adequate initial training in use of the ETC and should practice with the device regularly. To ensure optimal outcomes, we also highly recommend that EMS and other healthcare providers monitor their success rates and the occurrence of complications.
Laryngeal Mask Airway
The LMA is an adjunctive airway device composed of a tube with a
cuffed mask-like projection at the distal end (see Figure 3A
).30 The LMA is introduced
into the pharynx (Figure 3B
) and advanced until resistance is
felt as the distal portion of the tube locates in the hypopharynx. The
cuff is then inflated, which seals the larynx, leaving the distal
opening of the tube just above the glottis, providing a clear, secure
airway (Figure 3C
and 3D
).
|
The LMA provides a more secure and reliable means of ventilation than the face mask.31 32 Although the LMA does not ensure absolute protection against aspiration, studies have shown that regurgitation is less likely with the LMA than with the bag-mask device and that aspiration is uncommon.33 34 In comparison with the tracheal tube, the LMA provides equivalent ventilation. Training in the placement and use of an LMA is simpler than tracheal intubation because laryngoscopy and visualization of the vocal cords are unnecessary for insertion of the LMA.35 36 37 The LMA may have advantages over the tracheal tube when access to the patient is limited,38 there is a possibility of unstable neck injury,39 or appropriate positioning of the patient for tracheal intubation is impossible.
Studies have examined the use of the LMA by nurses, respiratory therapists, and EMS personnel, many of whom had not previously used either an LMA or a tracheal tube. Successful insertion rates with the LMA range from 64% to 100%.25 34 40 41 42 43 44
Even when the LMA can be inserted, studies report that a small proportion of patients cannot be ventilated with the LMA. Because insertion and ventilation are not ensured, it is important for providers to have an alternative strategy for management of the airway. Providers should receive adequate initial training in the use of the LMA and should practice with the device regularly to optimize insertion rates and to minimize complications. To ensure optimal outcomes we also highly recommend that EMS and other healthcare providers monitor their success rates and the occurrence of complications.
Transtracheal (Translaryngeal) Catheter Ventilation
In those rare cases when airway obstruction is not relieved by any
of the methods described above, additional procedures are necessary.
These include transtracheal catheter or translaryngeal ventilation.
Only specially trained and experienced personnel should attempt such
procedures.
Pharyngotracheal Lumen Airway
The PTL is a double-lumen tube similar in structure and
function to the ETC.45 The tube is inserted blindly into
the pharynx, ending in either the esophagus or the trachea. Assessment
of its location is then made, and the patient is ventilated through the
proper lumen. In the only published study since 1992 in which use of
the PTL was examined, the tube performed well but was preferred less
than the ETC.21 The PTL is not currently in wide use
(Class Indeterminate).
Cuffed Oropharyngeal Airway
The cuffed oropharyngeal airway (COPA) was first described
in 1992.46 Although designed originally for spontaneous
ventilation in anesthetized subjects, it may represent
a useful adjunct during resuscitation. The device is a modified
oropharyngeal airway with a distal inflatable cuff and proximal
standard 15-mm connector to which a self-inflating bag can be attached.
Recent data suggests that the COPA is relatively easy to use and may
offer an effective method of providing an adequate airway during
resuscitation by personnel not trained in more advanced
techniques.47
| Tracheal Intubation |
|---|
|
|
|---|
Tracheal intubation should be preceded by preoxygenation of the patient. If the patient is ventilating spontaneously, preoxygenation is achieved by providing 3 minutes of high-flow oxygen. If spontaneous ventilation is insufficient, assist ventilation with a bag-mask device.54
Currently the tracheal tube is considered the ventilation adjunct of choice because it keeps the airway patent, permits suctioning of airway secretions, ensures delivery of a high concentration of oxygen, provides a route for the administration of certain drugs, facilitates delivery of a selected tidal volume, and protects the airway from aspiration of gastric contents or blood and mucus from the oropharynx.55 Repeated safe and effective placement of the tracheal tube, over the wide range of patient and environmental conditions encountered in resuscitation, requires considerable skill and experience. Unless initial training is sufficient and ongoing practice and experience are adequate, fatal complications may result.
Multiple or unsuccessful intubation attempts may affect outcome from cardiac arrest adversely. Rates for failure to intubate are as high as 50% in EMS systems with a low patient volume and providers who do not perform intubation frequently.56 57 When tracheal intubation is attempted by providers with insufficient skill, the following complications may be seen: trauma to the oropharynx, ventilation withheld for unacceptably long periods, delayed or withheld chest compressions, esophageal or bronchial intubation, failure to secure the tube, and failure to recognize misplacement of the tube. Therefore, inexperienced providers should use only those airway management devices for which they have been adequately trained. Those who perform tracheal intubation require either frequent experience or frequent retraining. EMS systems should keep a record for each provider, documenting the number of intubations performed and success rates and complications (Class IIa).
Indications for tracheal intubation include (1) inability of the rescuer to ventilate the unconscious patient with less invasive methods and (2) absence of protective reflexes (coma or cardiac arrest).
During the process of tracheal intubation, the maximum interruption to ventilation should be 30 seconds. If more than 1 attempt at intubation is required, adequate ventilation and oxygenation must be provided between attempts. If the patient has a perfusing rhythm, use pulse oximetry and ECG monitoring continuously during intubation attempts.
Whenever possible a second rescuer should apply cricoid pressure during tracheal intubation in adults to protect against regurgitation of gastric contents and to help ensure tube placement in the tracheal orifice. Apply pressure with the thumb and index fingers to the right and left anterolateral aspects of the cricoid cartilage just lateral to the midline. Avoid overzealous pressure; it will occlude the airway and impair tracheal intubation.58 59 Maintain cricoid pressure until the cuff of the tracheal tube is inflated.60 61 The BURP (Backward, Upward, Rightward Pressure) technique may be useful in bringing the vocal cords into the field of vision of the intubator.
Tracheal tubes should be available in a variety of sizes. They should have standard 15-/22-mm connectors and should have high-volume, low-pressure cuffs suitable for adults and older children. The size of tracheal tube required typically is 8 mm for average adult women and 8 mm for average adult men. Because of the variation in size of adults, a range of tubes should be available.
A stylet should be available and may be used to assist with tracheal tube insertion by providing some stiffness to the tube and by allowing the direction of the tube to be controlled better during manipulation. When used, the stylet should not extend beyond the distal end of the tube.
Another excellent device to assist with placement of the tracheal tube is the gum elastic bougie. Because of its size and flexibility, the bougie is easier to place in the trachea than a tracheal tube. Once the bougie is placed in the trachea, the tracheal tube is passed over the bougie and into position in the trachea.62 63 64 65 66
Difficulties in achieving tracheal intubation usually occur because of inability to bring the vocal cords into view through the laryngoscope. Visualization is best accomplished by flexing the neck and extending the head at the atlanto-occipital joint (the "sniffing position"). Once the vocal cords are seen, the tube should be placed so that the cuff is just beyond the cords. In the average adult this position usually results in the tube lying at a depth marked on the side of the tube between the 19- and 23-cm marks at the front teeth. The cuff is then inflated with just enough air to occlude the airway (usually 10 mL). An adequate seal is confirmed by listening over the larynx while ventilation is continued and air is added to the cuff. While a normal tidal volume is delivered, air is added to the cuff just until the audible air leak around the tube disappears.
Immediately after insertion of the tracheal tube, confirm placement by auscultating over the epigastrium, the midaxillary, and the anterior chest line on the right and left sides of the chest. Even when the tracheal tube is seen to pass through the vocal cords and is verified in the trachea by auscultation, make secondary confirmation of placement with an end-tidal CO2 or esophageal detection device (Class IIa).67 Extensive data shows that clinical signs of proper tube placement (such as condensation in the tube, auscultation over the lungs and abdomen, and chest rise) are not always reliable indicators of correct tube placement.68 69 70 71 72
To protect against unrecognized esophageal intubation, confirmation of tube placement by an expired CO2 or esophageal detection device is necessary. In the out-of-hospital setting, unrecognized misplacement of the tracheal tube has been reported in as many as 17% of patients.73 Once the tube is placed, especially out of hospital, the location of the tracheal tube must be monitored closely.
The esophageal detector device depends on the ability to aspirate air from the lower airways through a tracheal tube placed in the cartilage-supported rigid trachea. When the tube is in the esophagus, air cannot be aspirated because the esophagus collapses when aspiration is attempted. The esophageal detector device is generally reliable in patients with both a perfusing and a nonperfusing rhythm (Class IIa),74 75 76 77 78 79 80 but it may yield misleading results in patients with morbid obesity, late pregnancy, or status asthmaticus or when there are copious tracheal secretions.78 81 82 With some of these conditions, the trachea tends to collapse. In the case of status asthmaticus, the airway secretions or the small-airway obstruction that characterizes severe asthma blocks air aspiration from the lower airways.
How to Confirm Accurate Placement of Tracheal Tube: Primary
Confirmation
Confirm tube placement immediately, assessing the first breath
delivered by the bag-valve unit.
How to Confirm Accurate Placement of Tracheal Tube: Secondary
Confirmation
A variety of electronic and mechanical devices, ranging from
simple and inexpensive to complex and costly, are available for both
in-hospital and out-of-hospital use. These include several models of
end-tidal CO2 detectors (qualitative,
quantitative, and continuous) and a variety of esophageal detector
devices.
End-Tidal CO2 Detectors
Several commercial devices measure the concentration
of exhaled CO2 from the lungs. The presence of
exhaled CO2 indicates proper tracheal tube
placement. A lack of CO2 on the detector
generally means that the tube is in the esophagus, particularly in
patients with spontaneous circulation.
False-positive readings* (tube is really in the trachea; device indicates in the esophagus; leads to unnecessary tube removal) may occur because CO2 delivery is low in cardiac arrest patients with low blood flow to the lungs or in patients with a large amount of dead space (eg, significant pulmonary embolus). False-negative readings (tube is really in the esophagus; device indicates in the trachea) have been reported from patients who had ingested carbonated liquids before the arrest.
Continuous (usually quantitative as well) end-tidal CO2 monitors can confirm successful tracheal tube placement within seconds of an intubation attempt. These monitors can also detect subsequent tracheal tube dislodgment, an event that is more likely to occur during out-of-hospital transportation of a patient.
Esophageal Detector Devices
These devices create a suction force at the tracheal end of
the tracheal tube, either from pulling back the plunger on a large
syringe or compressing a flexible bulb. If the tube is in the
esophagus, the suction will pull the esophageal mucosa against the
distal end of the detector and prevent movement of the device plunger
or reexpansion of the suction bulb.
Expired CO2 detectors are very reliable in patients with perfusing rhythms and are recommended to confirm tube position in these patients (Class IIa).83 84 During cardiac arrest, pulmonary blood flow may be so low that there is insufficient expired CO2, so a correctly placed tracheal tube is not identified by the expired CO2 detector. When expired CO2 is detected in cardiac arrest, it is a reliable indicator of tube position, but when it is absent, we recommend adding a second method of confirming tracheal tube placement, such as the esophageal detector device (Class IIb).85 86 87 88 89 A variety of electronic as well as simple, inexpensive, colorimetric detectors that detect exhaled CO2 are available for both in-hospital and out-of-hospital use.
After you confirm the position of the tube in the trachea, careful auscultation is needed to avoid inadvertent right main bronchial intubation. Once you achieve the correct positioning of the tube, record the depth of the tube as marked at the front teeth and secure the tube. Once the tube is secured, place an oropharyngeal airway or bite block. The respiratory rate during cardiac or respiratory arrest when the patient has been intubated should be 12 to 15 breaths per minute (1 breath every 4 to 5 seconds). Once a tracheal tube is in place, ventilation need not be synchronized with chest compressions. Once spontaneous circulation is restored after cardiac arrest, continue to provide 12 to 15 breaths per minute.
After tube confirmation and fixation, obtain a chest x-ray to confirm proper position of the end of the tracheal tube above the carina.
In patients with severe obstructive pulmonary disease with increased resistance to exhalation, care should be taken not to induce air trapping, which may result in auto-PEEP. In patients with hypovolemia this may cause a profound reduction in blood pressure. Lower respiratory rates (6 to 8 per minute) should be used, allowing more time for complete exhalation of gas.
| Suction Devices |
|---|
|
|
|---|
The installed suction unit should be powerful enough to provide an airflow of >40 L/min at the end of the delivery tube and a vacuum of >300 mm Hg when the tube is clamped. The amount of suction should be adjustable for use in children and intubated patients. Hand-powered suction units lack the problems associated with electric pumps and have had considerable anecdotal clinical success, although no formal evaluations have been published.
An additional set of rigid pharyngeal suction tips (tonsil suction tips) and sterile, curved tracheal suction catheters of various sizes should be available. For tracheal suction, a Y-piece or T-piece or a lateral opening should lie between the suction tube and the source of the on-off suction control. The suction yoke, collection bottle, water for rinsing, and suction tube should be readily accessible to the attendant in charge of the airway. Suction apparatus must be designed for easy cleaning and subsequent decontamination.
| Airway Summary |
|---|
|
|
|---|
|
| Footnotes |
|---|
Circulation. 2000;102(suppl I):I-95I-104.
| References |
|---|
|
|
|---|
2.
Melker RJ, Banner MJ. Ventilation during CPR:
two-rescuer standards reappraised. Ann Emerg Med. 1985;14:397402.
3.
Dorges V, Ocker H, Hagelberg S, Wenzel V, Idris
AH, Schmucker P. Smaller tidal volumes with room-air are not sufficient
to ensure adequate oxygenation during bag-valve-mask
ventilation. Resuscitation. 2000;44:3741.
4.
Harber T, Lucas BG. An evaluation of some
mechanical resuscitators for use in the ambulance service. Ann R
Coll Surg Engl. 1980;62:291203.
5.
Branson RD, McGough EK. Transport ventilators.
Probl Crit Care. 1990;4:254274.
6.
Nolan JP, Baskett PJF. Gas-powered and portable
ventilators: an evaluation of six models. Prehospital Disaster
Med. 1992;7:2534.
7.
McGough EK, Banner MJ, Melker RJ. Variations in
tidal volume with portable transport ventilators. Respir
Care. 1992;37:233239.
8.
Johannigman JA, Branson RD, Davis K, Hurst JM.
Techniques of emergency ventilation: a model to evaluate tidal volume,
airway pressure, and gastric insufflation. Trauma. 1991;31:9398.
9.
Davis K, Johannigman JA, Johnson RC, Branson RD.
Lung compliance following cardiac arrest. Acad Emerg Med.. 1995;2:874878.
10.
Braman SS, Dunn SM, Amico CA, Millman RP.
Complications of intrahospital transport in critically ill patients.
Ann Intern Med. 1987;107:469473.
11.
Gervais HW, Eberle B, Konietzke D, Hennes HJ,
Dick W. Comparison of blood gases of ventilated patients during
transport. Crit Care Med. 1987;15:761763.
12.
Johannigman JA, Branson RD, Johnson DJ, Davis K
Jr, Hurst JM. Out-of-hospital ventilation: bag-valve device vs
transport ventilator. Acad Emerg Med. 1995;2:719724.
13.
Updike G, Mosesso VN Jr, Auble TE, Delgado E.
Comparison of bag-valve-mask, manually triggered ventilator, and
automated ventilator devices used while ventilating a nonintubated
manikin model. Prehosp Emerg Care.. 1998;2:5255.
14.
Weg JG, Haas CF. Safe intrahospital transport of
critically ill ventilator-dependent patients. Chest. 1989;96:631635.
15.
Hurst JM, Davis K Jr, Branson RD, Johannigman JA.
Comparison of blood gases during transport using two methods of
ventilatory support. J Trauma. 1989;29:16371640.
16.
Fuerst RS, Banner MJ, Melker RJ. Gastric
inflation in the unintubated patient: a comparison of ventilating
devices. Ann Emerg Med. 1992;21:636.
17.
Pepe PE, Marini JJ. Occult positive
end-expiratory pressure in mechanically ventilated patients with
airflow obstruction: the auto-PEEP effect. Am Rev Respir
Dis. 1982;126:166170.
18.
Frass M, Frenzer R, Rauscha F, Weber H, Packer R,
Leithner C. Evaluation of esophageal tracheal Combitube in
cardiopulmonary resuscitation. Crit Care Med.. 1986;15:609611.
19.
Atherton GL, Johnson JC. Ability of paramedics to
use the Combitube in prehospital cardiac arrest. Ann Emerg
Med. 1993;22:12631268.
20.
Lefrancois D. Use of the esophageal-tracheal
Combitube (ETC) in prehospital cardiorespiratory arrest (CRA) in an
EMT-D level EMS system. Resuscitation.. 1998;37:544.
Abstract.
21.
Rumball CJ, Macdonald D. The PTL, Combitube,
laryngeal mask, and oral airway: a randomized prehospital comparative
study of ventilatory device effectiveness and cost-effectiveness in 470
cases of cardiorespiratory arrest. Prehosp Emerg Care.. 1997;1:110.
22.
Staudinger T, Brugger S, Watschinger B. Emergency
intubation with the Combitube: comparison with the endotracheal tube.
Ann Emerg Med.. 1993;22:15731575.
23.
Staudinger T, Brugger S, Roggla M. Comparison of
the Combitube with the endotracheal tube in cardiopulmonary
resuscitation in the prehospital phase. Wien Klin
Wochenschr.. 1994;106:412415.
24.
Frass M, Frenzer R, Rauscha F. Ventilation with
esophageal tracheal Combitube in cardiopulmonary resuscitation:
promptness and effectiveness. Chest.. 1988;93:781784.
25.
Tanigawa K, Shigematsu A. A choice of airway
device for 12,020 cases of non-traumatic cardiac arrest in Japan.
Prehosp Emerg Care. 1998;2:96100.
26.
Butler BD, Little T, Drtil S. Combined use of the
esophageal-tracheal Combitube with a colorimetric
carbon dioxide detector for emergency intubation/ventilation.
Clin Monit.. 1995;11:311316.
27.
Wafai Y, Salem MR, Baraka A. Effectiveness of the
self-inflating bulb for verification of proper placement of the
esophageal tracheal Combitube. Anesth Analg.. 1995;80:122126.
28.
Klein H, Williamson M, Sue-Ling HM. Esophageal
rupture associated with the use of the Combitube. Anesth
Analg.. 1997;85:937939.
29.
Vezina D, Lessard MR, Bussieres J.
Complications associated with the use of the esophageal tracheal
Combitube. Can J Anaesth. 1998;45:7680.
30.
Brain AI. The laryngeal mask: a new concept in
airway management. Br J Anaesth.. 1983;55:801805.
31.
Martin PD, Cyna AM, Hunter WAH. Training nursing
staff in airway management for resuscitation: a clinical comparison of
the facemask and laryngeal mask. Anaesthesia.. 1993;48:3337.
32.
Alexander R, Hodgson P, Lomax D. A comparison of
laryngeal mask airway and Guedel airway, bag and facemask for manual
ventilation. Anaesthesia.. 1993;48:231234.
33.
Stone BJ, Chantler PJ, Basket PJ. The incidence
of regurgitation during cardiopulmonary
resuscitation: a comparison between the bag valve mask and laryngeal
mask airway. Resuscitation. 1998;38:36.
34.
Baskett PJF. The use of the laryngeal mask airway
by nurses during cardiopulmonary resuscitation: results of a
multicentre trial. Anaesthesia. 1994;49:37.
35.
Pennant JH, Walker MB. Comparison of the
endotracheal tube and laryngeal mask in airway management by
paramedical personnel. Anesth Analg. 1992;74:531534.
36.
Reinhart DJ. Laryngeal mask airway (LMA) vs
endotracheal tube by paramedics and respiratory therapists. Ann
Emerg Med.. 1994;24:260263.
37.
Davies PRF, Tighe SQ, Greenslade GL. Laryngeal
mask and tracheal tube insertion by unskilled personnel.
Lancet.. 1990;336:977979.
38.
Greene MK, Radon R, Hinchley G. The laryngeal
mask airway: two cases of prehospital care. Anaesthesia.. 1992;47:686689.
39.
Pennant H, Pace NA, Gajraj NM. Role of the
laryngeal mask airway in the immobile cervical spine. J Clin
Anesth. 1993;5:226230.
40.
Kokkinis K. The use of the laryngeal mask in CPR.
Resuscitation.. 1994;27:912.
41.
Kokkinis K, Leach A, Alexander CA, Stone B. The
laryngeal mask in cardiopulmonary resuscitation in a district
general hospital: a preliminary communication. Multicentre Trial
(Co-ordinator Baskett PJF). Resuscitation.. 1993;25:245248.
42.
Samarkandi AH, Seraj MA, el Dawlatly A. The role
of the laryngeal mask airway in cardiopulmonary resuscitation.
Resuscitation. 1994;28:103106.
43.
Verghese C, Prior Willeard PFS, Baskett PJF.
Immediate management of the airway during cardiopulmonary
resuscitation in a hospital without a resident anaesthesiologist.
Eur J Emerg Med.. 1994;1:123125.
44.
Grantham H, Phillips G, Gilligan JE. The
laryngeal mask in prehospital emergency care. Emerg Med. 1994;6:193197.
45.
Niemann JT, Rosborough JP, Myers R, Scarberry EN.
The pharyngeo-tracheal lumen airway: preliminary investigation of a new
adjunct. Ann Emerg Med.. 1984;13:591596.
46.
Asai T, Koga K, Stacey MR. Use of the cuffed
oropharyngeal airway after difficult ventilation through a facemask.
Anaesthesia.. 1997;52:12361237.
47.
Rees SG, Gabbott DA. Use of the cuffed
oropharyngeal airway for manual ventilation by nonanaesthetists.
Anaesthesia.. 1999;54:10891093.
48.
Bowman FP, Menegazzi JJ, Check BD, Duckett TM.
Lower esophageal sphincter pressure during prolonged cardiac arrest and
resuscitation. Ann Emerg Med.. 1995;26:216219.
49.
Ruben H, Knudsen EJ, Carugati G. Gastric
inflation in relation to airway pressure. Acta Anaesthesiol
Scand.. 1961;5:107114.
50.
Weiler N, Heinrichs W, Dick W. Assessment of
pulmonary mechanics and gastric inflation pressure during mask
ventilation. Prehospital Disaster Med.. 1995;10:101105.
51.
Ruben H, Johansen SH. Gastric inflation due to
artificial respiration. Acta Anaesthesiol Scand.. 1970;14:281285.
52.
Wenzel V, Idris AH, Banner MJ. Respiratory system
compliance decreases after cardiopulmonary resuscitation and
stomach inflation: impact of large and small tidal volumes on
calculated peak airway pressure. Resuscitation. 1998;38:113118.
53.
Berg MD, Idris AH, Berg RA. Severe ventilatory
compromise due to gastric distension during cardiopulmonary
resuscitation. Resuscitation.. 1998;36:7173.
54.
McGowan P, Skinner A.
Preoxygenation: the importance of a good facemask seal.
Br J Anaesth.. 1995;75:777778.
55.
Pepe PE, Copass MK, Joyce TH. Prehospital
endotracheal intubation: rationale for training emergency medical
personnel. Ann Emerg Med. 1985;14:10851092.
56.
Bradley JS, Billows GL, Olinger ML, Boha SP,
Cordell WH, Nelson DR. Prehospital oral endotracheal intubation by
rural basic emergency medical technicians. Ann Emerg Med. 1998;32:2632.
57.
Sayre MR, Sakles JC, Mistler AF, Evans JL, Kramer
AT, Panicioli AM. Field trial of endotracheal intubation by basic EMTs.
Ann Emerg Med. 1998;31:228233.
58.
Hartsilver EL, Vanner RG. Airway obstruction with
cricoid pressure. Anaesthesia.. 2000;55:208211.
59.
MacG Palmer JH, Ball DR. The effect of cricoid
pressure on the cricoid cartilage and vocal cords: an endoscopic study
in anaesthetised patients. Anaesthesia. 2000;55:263268.
60.
Sellick BA. Cricoid pressure to control
regurgitation of stomach contents during induction of
anaesthesia. Lancet. 1961;2:24042406.
61.
Wraight WJ, Chamney AR, Howells TH. The
determination of an effective cricoid pressure. Anaesthesia. 1983;38:461466.
62.
Nolan JP, Wilson ME. Orotracheal intubation in
patients with potential cervical spine injuries: an indication for the
gum elastic bougie. Anaesthesia. 1993;48:630633.
63.
Viswanathan S, Campbell C, Wood DG. The Eschmann
tracheal tube introducer (gum elastic bougie). Anesthesiol
Rev. 1992;19:2934.
64.
Kidd JF, Dyson A, Latto IP. Successful difficult
intubation use of the gum elastic bougie. Anaesthesia.. 1988;43:437438.
65.
Nolan JP, Wilson ME. An evaluation of the gum
elastic bougie: intubation times and incidence of sore throat.
Anaesthesia.. 1992;47:878881.
66.
Dogra S, Falconer R, Latto IP. Successful
difficult intubation: tracheal tube placement over a gum elastic
bougie. Anaesthesia.. 1990;45:774776.
67.
White SJ, Slovis CM. Inadvertent
esophageal intubation in the field: reliance on a fools "gold
standard." Acad Emerg Med.. 1997;4:8991.
68.
Charters P. Normal chest expansion with
esophageal placement of a tracheal tube. Anaesthesia.. 1989;44:365.
69.
Andersen KH, Hald A. Assessing the position of
the tracheal tube: the reliability of different methods.
Anaesthesia. 1989;44:984985.
70.
Vaghadia H, Jenkins LC, Ford RW. Comparison of
end-tidal carbon dioxide, oxygen saturation and clinical signs for the
detection of esophageal intubation. Can J Anaesth. 1989;36:560564.
71.
Kelly JJ. Use of tube condensation as an
indicator of endotracheal tube placement. Ann Emerg Med.. 1998;31:575578.
72.
Anderson KH, Schultz-Lebahn T. Esophageal
intubation can be undetected by auscultation of the chest. Acta
Anaesthesiol Scand.. 1994;38:580582.
73.
Katz SH, Falk JL. Misplaced endotracheal tubes by
paramedics in an urban emergency medical services system. Acad
Emerg Med.. 1998;5:429.
74.
Zaleski L, Abello D, Gold MI. The esophageal
detector device: does it work? Anesthesiology. 1993;79:244247.
75.
Williams KN, Nunn JF. The oesophageal detector
device: a prospective trial on 100 patients. Anaesthesia. 1989;44:412414.
76.
Wee MY, Walker AK. The oesophageal detector
device: an assessment with uncuffed tubes in children.
Anaesthesia. 1991;46:869871.
77.
Marley CD Jr, Eitel DR, Anderson TE. Evaluation
of a prototype esophageal detection device. Acad Emerg Med. 1995;2:503507.
78.
Baraka A, Khoury PJ, Siddik SS. Efficacy of the
self-inflating bulb in differentiating esophageal from tracheal
intubation in the parturient undergoing cesarean section. Anesth
Analg. 1997;84:533537.
79.
Salem MR, Wafai Y, Joseph NJ. Efficacy of the
self-inflating bulb in detecting esophageal intubation: does the
presence of a nasogastric tube or cuff deflation make a difference?
Anesthesiology. 1994;80:4248.
80.
Kasper CL, Deem S. The self-inflating bulb to
detect esophageal intubation during emergency airway management.
Anesthesiology. 1998;88:898902.
81.
Lang DJ, Wafai Y, Salem MR. Efficacy of the
self-inflating bulb in confirming tracheal intubation in the morbidly
obese. Anesthesiology. 1996;85:246253.
82.
Davis DP, Stephen KA, Vilke GM. Inaccuracy in
endotracheal tube verification using a Toomey syringe. Emerg
Med. 1999;17:3538.
83.
MacLeod BA, Heller MB, Gerard J, Yealy DM,
Menegazzi JJ. Verification of endotracheal tube placement with
colorimetric end-tidal CO2
detection. Ann Emerg Med.. 1991;20:267270.
84.
Chow LH, Lui PW, Cheung EL, Jong HR, Yang TC,
Chen YC. Verification of endotracheal tube misplacement with the
colorimetric carbon dioxide detector during
anesthesia. Chung Hua I Hsueh Tsa Chih. 1993;51:415418.
85.
Hayden SR, Sciammarella J, Viccellio P.
Colorimetric end-tidal CO2
detector for verification of endotracheal tube placement in
out-of-hospital cardiac arrest. Acad Emerg Med. 1995;2:499502.
86.
Anton WR, Gordon RW, Jordan TM. A disposable
end-tidal CO2 detector to verify endotracheal
intubation. Ann Emerg Med. 1991;20:271275.
87.
Bozeman WP, Hexter D, Liang HK, Kelen GD.
Esophageal detector device versus detection of end-tidal carbon dioxide
level in emergency intubation. Ann Emerg Med.. 1996;27:595599.
88.
Schaller RJ, Huff JS, Zahn A. Comparison of a
colorimetric end-tidal CO2
detector and an esophageal aspiration device for verifying endotracheal
tube placement in the prehospital setting: a six-month experience.
Prehospital Disaster Med.. 1997;12:5763.
89.
Jenkins WA, Verdile VP, Paris PM. The syringe
aspiration device technique to verify endotracheal tube position.
Am J Emerg Med.. 1994;12:413416.
This article has been cited by other articles:
![]()
![]()

![]()
![]()
![]()
C. S.H. Ng, S. Wan, A. P.C. Yim, and A. A. Arifi
Pulmonary Dysfunction After Cardiac Surgery*
Chest,
April 1, 2002;
121(4):
1269 - 1277.
[Abstract]
[Full Text]
[PDF]
![]()
![]()
![]()

![]()
![]()
![]()
M. A. Chaney
Corticosteroids and Cardiopulmonary Bypass : A Review of Clinical Investigations
Chest,
March 1, 2002;
121(3):
921 - 931.
[Abstract]
[Full Text]
[PDF]
![]()
![]()
![]()

![]()
![]()
![]()
P. Menasche
The systemic factor: the comparative roles of cardiopulmonary bypass and off-pump surgery in the genesis of patient injury during and following cardiac surgery
Ann. Thorac. Surg.,
December 1, 2001;
72(6):
S2260 - 2265.
[Abstract]
[Full Text]
[PDF]
![]()
This Article ![]()
![]()
Extract
![]()
Alert me when this article is cited
![]()
Alert me if a correction is posted
![]()
Services ![]()
![]()
Email this article to a friend
![]()
Similar articles in this journal
![]()
Similar articles in PubMed
![]()
Alert me to new issues of the journal
![]()
Download to citation manager
![]()
Request Permissions ![]()
Citing Articles ![]()
![]()
Citing Articles via HighWire
![]()
Citing Articles via Google Scholar
![]()
Google Scholar ![]()
![]()
Search for Related Content
![]()
PubMed ![]()
![]()
PubMed Citation
![]()
Related Collections ![]()
![]()
AHA Statements and Guidelines
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2000 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |