(Circulation. 2000;102:I-343.)
© 2000 American Heart Association, Inc.
ECC Guidelines |
| Major Guidelines Changes |
|---|
|
|
|---|
At the Guidelines 2000 Conference, we made the following recommendations:
Temperature
Oxygenation and Ventilation
Chest Compressions
Medications, Volume Expansion, and Vascular Access
Ethics
| Introduction |
|---|
|
|
|---|
Approximately 5% to 10% of the newly born population require some degree of active resuscitation at birth (eg, stimulation to breathe),1 and approximately 1% to 10% born in the hospital are reported to require assisted ventilation.2 More than 5 million neonatal deaths occur worldwide each year. It has been estimated that birth asphyxia accounts for 19% of these deaths, suggesting that the outcome might be improved for more than 1 million infants per year through implementation of simple resuscitative techniques.3 Although the need for resuscitation of the newly born infant often can be predicted, such circumstances may arise suddenly and may occur in facilities that do not routinely provide neonatal intensive care. Thus, it is essential that the knowledge and skills required for resuscitation be taught to all providers of neonatal care.
With adequate anticipation, it is possible to optimize the delivery setting with appropriately prepared equipment and trained personnel who are capable of functioning as a team during neonatal resuscitation. At least 1 person skilled in initiating neonatal resuscitation should be present at every delivery. An additional skilled person capable of performing a complete resuscitation should be immediately available.
Neonatal resuscitation can be divided into 4 categories of action:
Tracheal intubation may be required during any of these steps. All newly born infants require rapid assessment, including examination for the presence of meconium in the amniotic fluid or on the skin; evaluation of breathing, muscle tone, and color; and classification of gestational age as term or preterm. Newly born infants with a normal rapid assessment require only routine care (warmth, clearing the airway, drying). All others receive the initial steps, including warmth, clearing the airway, drying, positioning, stimulation to initiate or improve respirations, and oxygen as necessary.
Subsequent evaluation and intervention are based on a triad of characteristics: (1) respirations, (2) heart rate, and (3) color. Most newly born infants require only the basic steps, but for those who require further intervention, the most crucial action is establishment of adequate ventilation. Only a very small percentage will need chest compressions and medications.4
Certain special circumstances have unique implications for resuscitation of the newly born infant. Care of the infant after resuscitation includes not only supportive care but also ongoing monitoring and appropriate diagnostic evaluation. In certain clinical circumstances, noninitiation or discontinuation of resuscitation in the delivery room may be appropriate. Finally, it is important to document resuscitation interventions and responses in order to understand an individual infants pathophysiology as well as to improve resuscitation performance and study resuscitation outcomes.5 6 7 8
| Background |
|---|
|
|
|---|
Using questions and controversies identified during the ILCOR process, the Neonatal Resuscitation Program Steering Committee (AAP), the Pediatric Working Group (ILCOR), and the Pediatric Resuscitation Subcommittee of the Emergency Cardiovascular Care Committee (AHA) carried out further evidence evaluation. At the Evidence Evaluation Conference and International Guidelines 2000 Conference on CPR and ECC, these groups and panels of international experts and participants developed additional recommendations. The International Guidelines 2000 recommendations form the basis of this document.
Definition of "Newly Born," "Neonate," and
"Infant"
Although the guidelines for neonatal resuscitation focus on newly
born infants, most of the principles are applicable throughout the
neonatal period and early infancy. The term "newly born" refers
specifically to the infant in the first minutes to hours after
birth. The term "neonate" is generally defined as an infant
during the first 28 days of life. Infancy includes the neonatal
period and extends through 12 months of age.
Unique Physiology of the Newly Born
The transition from fetal to extrauterine life is characterized by
a series of unique physiological events: the lungs
change from fluid-filled to air-filled, pulmonary blood flow
increases dramatically, and intracardiac and extracardiac shunts
(foramen ovale and ductus arteriosus) initially reverse direction and
subsequently close. Such physiological
considerations affect resuscitative interventions in the newly
born.
For initial lung expansion, fluid-filled alveoli may require higher ventilation pressures than are commonly used in rescue breathing during infancy.18 19 Physical expansion of the lungs, with establishment of functional residual capacity and increase in alveolar oxygen tension, both mediate the critical decrease in pulmonary vascular resistance and result in an increase in pulmonary blood flow after birth. Failure to normalize pulmonary vascular resistance may result in persistence of right-to-left intracardiac and extracardiac shunts (persistent pulmonary hypertension). Failure to adequately expand alveolar spaces may result in intrapulmonary shunting of blood with resultant hypoxemia. In addition to disordered cardiopulmonary transition, disruption of the fetoplacental circulation also may render the newly born at risk for resuscitation because of acute blood loss.
Developmental considerations at various gestational ages also influence pulmonary pathology and resuscitation physiology in the newly born. Surfactant deficiency in the premature infant alters lung compliance and resistance.20 Meconium passed into the amniotic fluid may be aspirated, leading to airway obstruction. Complications of meconium aspiration are particularly likely in infants small for gestational age and those born post term or with significant perinatal compromise.21
Although certain physiological features are unique to the newly born, others pertain to infants throughout the neonatal period and into the first months of life. Severe illness due to a wide variety of conditions continues to manifest as disturbances in respiratory function (cyanosis, apnea, respiratory failure). Convalescing preterm infants with chronic lung disease often require significant ventilatory support regardless of the etiology of their need for resuscitation. Persistent pulmonary hypertension, persistent patency of the ductus arteriosus, and intracardiac shunts may produce symptoms during the neonatal period or even into infancy. Thus, many of the considerations and interventions that apply to the newly born may remain important for days, weeks, or months after birth.
The point at which neonatal resuscitation guidelines should be replaced by pediatric resuscitation protocols varies for individual patients. Objective data is lacking on optimal compression-ventilation ratios by age and disease state. However, infants with acute or chronic lung disease may benefit from a lower compression-ventilation ratio well into infancy. For these infants, continued use of some aspects of the neonatal guidelines is reasonable. Conversely, a neonate with a cardiac arrhythmia resulting in poor perfusion requires use of protocols more fully detailed in pediatric advanced life support. Factors of age, pathophysiology, and caregiver training should be evaluated for each patient and the most appropriate resuscitation routines and care setting identified.
| Anticipation of Resuscitation Need |
|---|
|
|
|---|
Communication
Appropriate preparation for an anticipated high-risk delivery
requires communication between the person(s) caring for the mother and
those responsible for resuscitation of the newly born. Communication
among caregivers should include details of antepartum and intrapartum
maternal medical conditions and treatment as well as specific
indicators of fetal condition (fetal heart rate monitoring, lung
maturity, ultrasonography). Table 1
lists
examples of the antepartum and intrapartum circumstances that place the
newly born infant at risk.
|
| Preparation for Delivery |
|---|
|
|
|---|
Equipment
Although the need for resuscitation at birth often can be
predicted by risk factors, for many infants resuscitation cannot be
anticipated.22 Therefore, a clean and warm environment
with a complete inventory of resuscitation equipment and drugs should
be maintained at hand and in fully operational condition wherever
deliveries occur. Table 2
presents a
list of suggested neonatal supplies, medications, and equipment.
|
Standard precautions should be followed carefully in delivery areas, where exposure to blood and body fluids is likely. All fluids from patients should be treated as potentially infectious. Personnel should wear gloves and other appropriate protective barriers when handling newly born infants or contaminated equipment. Techniques involving mouth suction by the healthcare provider should not be used.
| Evaluation |
|---|
|
|
|---|
Response to Extrauterine Environment
Most newly born infants will respond to the stimulation of the
extrauterine environment with strong inspiratory efforts, a vigorous
cry, and movement of all extremities. If these responses are intact,
color improves steadily from cyanotic or dusky to pink, and heart rate
can be assumed to be adequate. The infant who responds vigorously to
the extrauterine environment and who is term can remain with the mother
to receive routine care (warmth, clearing the airway, drying).
Indications for further assessment under a radiant warmer and possible
intervention include
Further assessment of the newly born infant is based on the triad of respiration, heart rate, and color.
Respiration
After initial respiratory efforts, the newly born infant should be
able to establish regular respirations sufficient to improve color and
maintain a heart rate >100 bpm. Gasping and apnea are signs that
indicate the need for assisted ventilation.23
Heart Rate
Heart rate is determined by listening to the precordium
with a stethoscope or feeling pulsations at the base of the umbilical
cord. Central and peripheral pulses in the neck and
extremities are often difficult to feel in infants,24 25
but the umbilical pulse is readily accessible in the newly born and
permits assessment of heart rate without interruption of ventilation
for auscultation. If pulsations cannot be felt at the base of the cord,
auscultation of the precordium should be performed. Heart rate
should be consistently >100 bpm in an uncompromised newly born
infant. An increasing or decreasing heart rate also can provide
evidence of improvement or deterioration.
Color
An uncompromised newly born infant will be able to maintain a pink
color of the mucous membranes without supplemental oxygen. Central
cyanosis is determined by examining the face, trunk, and mucous
membranes. Acrocyanosis is usually a normal finding at birth and is not
a reliable indicator of hypoxemia, but it may indicate other
conditions, such as cold stress. Pallor may be a sign of decreased
cardiac output, severe anemia, hypovolemia, hypothermia, or
acidosis.
| Techniques of Resuscitation |
|---|
|
|
|---|
|
Basic Steps
Warmth
Preventing heat loss in the newly born is vital because cold
stress can increase oxygen consumption and impede effective
resuscitation.26 27 Hyperthermia should be avoided,
however, because it is associated with perinatal respiratory
depression28 29 (Class III, level of evidence [LOE] 3).
Whenever possible, deliver the infant in a warm, draft-free area.
Placing the infant under a radiant warmer, rapidly drying the skin,
removing wet linen immediately, and wrapping the infant in prewarmed
blankets will reduce heat loss. Another strategy for reducing heat loss
is placing the dried infant skin-to-skin on the mothers chest or
abdomen to use her body as a heat source.
Recent animal and human studies have suggested that selective (cerebral) hypothermia of the asphyxiated infant may protect against brain injury.30 31 32 Although this is a promising area of research, we cannot recommend routine implementation until appropriate controlled studies in humans have been performed (Class Indeterminate, LOE 2).
Clearing the Airway
The infants airway is cleared by positioning of the infant and
removal of secretions if needed.
Positioning
The newly born infant should be placed supine or lying on its
side, with the head in a neutral or slightly extended position. If
respiratory efforts are present but not producing effective tidal
ventilation, often the airway is obstructed; immediate efforts must be
made to correct overextension or flexion or to remove secretions. A
blanket or towel placed under the shoulders may be helpful in
maintaining proper head position.
Suctioning
If time permits, the person assisting delivery of the infant
should suction the infants nose and mouth with a bulb syringe after
delivery of the shoulders but before delivery of the chest. Healthy,
vigorous, newly born infants generally do not require suctioning after
delivery.33 Secretions may be wiped from the nose and
mouth with gauze or a towel. If suctioning is necessary, clear
secretions first from the mouth and then the nose with a bulb syringe
or suction catheter (8F or 10F). Aggressive pharyngeal suction can
cause laryngeal spasm and vagal bradycardia34 and delay
the onset of spontaneous breathing. In the absence of meconium or
blood, limit mechanical suction with a catheter in depth and duration.
Negative pressure of the suction apparatus should not
exceed 100 mm Hg (13.3 kPa or 136 cm H2O).
If copious secretions are present, the infants head may be turned
to the side, and suctioning may help clear the airway.
Clearing the Airway of Meconium
Approximately 12% of deliveries are complicated by the presence
of meconium in the amniotic fluid.35 When the amniotic
fluid is meconium-stained, suction the mouth, pharynx, and nose as soon
as the head is delivered (intrapartum suctioning) regardless of whether
the meconium is thin or thick.36 Either a large-bore
suction catheter (12F to 14F) or bulb syringe can be
used.37 Thorough suctioning of the nose, mouth, and
posterior pharynx before delivery of the body appears to decrease the
risk of meconium-aspiration syndrome.36 Nevertheless, a
significant number (20% to 30%) of meconium-stained infants will have
meconium in the trachea despite such suctioning and in the absence of
spontaneous respirations.38 39 This suggests the
occurrence of in utero aspiration and the need for tracheal suctioning
after delivery in depressed infants.
If the fluid contains meconium and the infant has absent or depressed respirations, decreased muscle tone, or heart rate <100 bpm, perform direct laryngoscopy immediately after birth for suctioning of residual meconium from the hypopharynx (under direct vision) and intubation/suction of the trachea.40 41 There is evidence that tracheal suctioning of the vigorous infant with meconium-stained fluid does not improve outcome and may cause complications (Class I, LOE 1).42 43 Warmth can be provided by a radiant heater; however, drying and stimulation generally should be delayed in such infants. Accomplish tracheal suctioning by applying suction directly to a tracheal tube as it is withdrawn from the airway. Repeat intubation and suctioning until little additional meconium is recovered or until the heart rate indicates that resuscitation must proceed without delay. If the infants heart rate or respiration is severely depressed, it may be necessary to institute positive-pressure ventilation despite the presence of some meconium in the airway. Suction catheters inserted through the tracheal tube may be too small to accomplish initial removal of particulate meconium; subsequent use of suction catheters inserted through a tracheal tube may be adequate to continue removal of meconium. Delay gastric suctioning to prevent aspiration of swallowed meconium until initial resuscitation is complete. Meconium-stained infants who develop apnea or respiratory distress should receive tracheal suctioning before positive-pressure ventilation, even if they are initially vigorous.
Tactile Stimulation
Drying and suctioning produce enough stimulation to initiate
effective respirations in most newly born infants. If an infant fails
to establish spontaneous and effective respirations after drying with a
towel or gentle rubbing of the back, flicking the soles of the feet may
initiate spontaneous respirations. Avoid more vigorous methods of
stimulation. Tactile stimulation may initiate spontaneous respirations
in newly born infants who are experiencing primary apnea. If these
efforts do not result in prompt onset of effective ventilation,
discontinue them because the infant is in secondary apnea and
positive-pressure ventilation will be required.23
Oxygen Administration
Hypoxia is nearly always present in a newly born
infant who requires resuscitation. Therefore, if cyanosis, bradycardia,
or other signs of distress are noted in a breathing newborn during
stabilization, administration of 100% oxygen is indicated while
determining the need for additional intervention. Free-flow oxygen can
be delivered through a face mask and flow-inflating bag, an oxygen
mask, or a hand cupped around oxygen tubing. The oxygen source should
deliver at least 5 L/min, and the oxygen should be held close to the
face to maximize the inhaled concentration. Many self-inflating bags
will not passively deliver sufficient oxygen flow (ie, when not being
squeezed). The goal of supplemental oxygen use should be normoxia;
sufficient oxygen should be administered to achieve pink color in the
mucous membranes. If cyanosis returns when supplemental oxygen is
withdrawn, post-resuscitation care should include monitoring of
administered oxygen concentration and arterial oxygen
saturation.
Ventilation
Most newly born infants who require positive-pressure ventilation
can be adequately ventilated with a bag and mask. Indications for
positive-pressure ventilation include apnea or gasping respirations,
heart rate <100 bpm, and persistent central cyanosis despite 100%
oxygen.
Although the pressure required for establishment of air breathing is variable and unpredictable, higher inflation pressures (30 to 40 cm H2O or higher) and longer inflation times may be required for the first several breaths than for subsequent breaths.18 19 Visible chest expansion is a more reliable sign of appropriate inflation pressures than any specific manometer reading. The assisted ventilation rate should be 40 to 60 breaths per minute (30 breaths per minute when chest compressions are also being delivered). Signs of adequate ventilation include bilateral expansion of the lungs, as assessed by chest wall movement and breath sounds, and improvement in heart rate and color. If ventilation is inadequate, check the seal between mask and face, clear any airway obstruction (adjust head position, clear secretions, open the infants mouth), and finally increase inflation pressure. Prolonged bag-mask ventilation may produce gastric inflation; this should be relieved by insertion of an 8F orogastric tube that is aspirated with a syringe and left open to air. If such maneuvers do not result in adequate ventilation, endotracheal intubation should follow.
After 30 seconds of adequate ventilation with 100% oxygen,
spontaneous breathing and heart rate should be checked. If spontaneous
respirations are present and the heart rate is
100 bpm,
positive-pressure ventilation may be gradually reduced and
discontinued. Gentle tactile stimulation may help maintain and improve
spontaneous respirations while free-flow oxygen is administered. If
spontaneous respirations are inadequate or if heart rate remains below
100 bpm, assisted ventilation must continue with bag and mask or
tracheal tube. If the heart rate is <60 bpm, continue assisted
ventilation, begin chest compressions, and consider endotracheal
intubation.
The key to successful neonatal resuscitation is establishment of adequate ventilation. Reversal of hypoxia, acidosis, and bradycardia depends on adequate inflation of fluid-filled lungs with air or oxygen.44 45 Although 100% oxygen has been used traditionally for rapid reversal of hypoxia, there is biochemical evidence and preliminary clinical evidence to argue for resuscitation with lower oxygen concentrations.46 47 48 Current clinical data, however, is insufficient to justify adopting this as routine practice. If assisted ventilation is required, deliver 100% oxygen by positive-pressure ventilation. If supplemental oxygen is unavailable, initiate resuscitation of the newly born infant with positive-pressure ventilation and room air (Class Indeterminate, LOE 2).
Ventilation Bags
Resuscitation bags used for neonates should be no larger than 750
mL; larger bag volumes make it difficult to judge delivery of the small
tidal volumes (5 to 8 mL/kg) that newly born infants require. Bags for
neonatal resuscitation can be either self-inflating or
flow-inflating.
Self-Inflating Bags
The self-inflating bag refills independently of gas flow
because of the recoil of the bag. To permit rapid reinflation, most
bags of this type have an intake valve at one end that pulls in room
air, diluting the oxygen flowing into the bag at a fixed rate. Delivery
of high concentrations of oxygen (90% to 100%) with a self-inflating
bag requires an attached oxygen reservoir.
To maintain inflation pressure for at least 1 second, a minimum bag volume of 450 to 500 mL may be necessary. If the device contains a pressure-release valve, it should release at approximately 30 to 35 cm H2O pressure and should have an override feature to permit delivery of higher pressures if necessary to achieve good chest expansion. Self-inflating bags that are not pressure limited or that have a device to bypass the pressure-release valve should be equipped with an in-line manometer. Do not use self-inflating bags to deliver oxygen passively through the mask because the flow of oxygen is unreliable unless the bag is being squeezed.
Flow-Inflating Bags
The flow-inflating (anesthesia) bag inflates only when
compressed gas is flowing into it and the patient outlet is at least
partially occluded. Proper use requires adjustment of the flow of gas
into the gas inlet, adjustment of the flow of gas out through the
flow-control valve, and creation of a tight seal between the mask and
face. Because a flow-inflating bag is capable of delivering very high
pressures, a manometer should be connected to the bag to monitor peak
and end-expiratory pressures. More training is required for proper use
of the flow-inflating bag than the self-inflating bag,49
but the flow-inflating bag can provide a greater range of peak
inspiratory pressures and more reliable control of oxygen
concentration. High concentrations of oxygen may be delivered passively
through the mask of a flow-inflating bag.
Face Masks
Masks should be of appropriate size to seal around the mouth and
nose but not cover the eyes or overlap the chin. A range of sizes
should be available. A round mask can seal effectively on the face of a
small infant; anatomically shaped masks better fit the contours of a
large term infants face. Masks should be designed to have low dead
space (<5 mL). A mask with a cushioned rim is preferable to one
without because the cushioned rim facilitates creation of a tight seal
without exerting excessive pressure on the face.50
Laryngeal Mask Airway Ventilation
Masks that fit over the laryngeal inlet have been shown to be
effective for ventilating newly born full-term infants.51
There is limited data on the use of these devices in small preterm
infants,52 however, and their use in the setting of
meconium-stained amniotic fluid has not been studied. The laryngeal
mask airway, when used by appropriately trained providers, may be an
effective alternative for establishing an airway in resuscitation of
the newly born infant, especially in the case of ineffective bag-mask
ventilation or failed endotracheal intubation (Class Indeterminate, LOE
5). However, we cannot recommend routine use of the laryngeal mask
airway at this time, and the device cannot replace endotracheal
intubation for meconium suctioning.
Endotracheal Intubation
Endotracheal intubation may be indicated at several points during
neonatal resuscitation:
The timing of endotracheal intubation may also depend on the skill and experience of the resuscitator.
Keep the supplies and equipment for endotracheal intubation together
and readily available in each delivery room, nursery, and Emergency
Department. Preferred tracheal tubes have a uniform diameter (without a
shoulder) and have a natural curve, a radiopaque indicator line, and
markings to indicate the appropriate depth of insertion. If a stylet is
used, it must not protrude beyond the tip of the tube. Table 3
provides a guideline for selection of
tracheal tube sizes and depths of insertion. Positioning the vocal cord
guide (a line proximal to the tip of the tube) at the level of the
vocal cords should position the tip of the tube above the carina.
Proper depth of insertion can also be estimated by calculating the
depth at the lips according to the following formula:
weight in kilograms + 6 cm=insertion depth at lip in cm
|
Perform endotracheal intubation orally, using a laryngoscope with a straight blade (size 0 for premature infants, size 1 for term infants). Insert the tip of the laryngoscope into the vallecula or onto the epiglottis and elevate gently to reveal the vocal cords. Cricoid pressure may be helpful. Insert the tube to an appropriate depth through the vocal cords as indicated by the vocal cord guide line and check its position by the centimeter marking at the upper lip. Record and maintain this depth of insertion. Variation in head position will alter the depth of insertion and may predispose to unintentional extubation or endobronchial intubation.53 54
After endotracheal intubation, confirm the position of the tube by the following:
An exhaled-CO2 monitor may be used to verify tracheal tube placement.55 These devices are associated with some false-negative but few false-positive results.56 Monitoring of exhaled CO2 can be useful in the secondary confirmation of tracheal intubation in the newly born, particularly when clinical assessment is equivocal (Class Indeterminate, LOE 5). Data about sensitivity and specificity of exhaled CO2 detectors in reflecting tracheal tube position is limited in newly born infants. Extrapolation of data from other age groups is problematic because conditions common to the newborn period, including inadequate pulmonary expansion, decreased pulmonary blood flow, and small tidal volumes, may influence the interpretation of the exhaled CO2 concentration.
Chest Compressions
Asphyxia causes peripheral vasoconstriction, tissue
hypoxia, acidosis, poor myocardial
contractility, bradycardia, and eventually cardiac
arrest. Establishment of adequate ventilation and
oxygenation will restore vital signs in the vast
majority of newly born infants. In deciding when to initiate chest
compressions, consider the heart rate, the change of heart rate, and
the time elapsed after initiation of resuscitative measures. Because
chest compressions may diminish the effectiveness of ventilation, do
not initiate them until lung inflation and ventilation have been
established.
The general indication for initiation of chest compressions is a heart rate <60 bpm despite adequate ventilation with 100% oxygen for 30 seconds. Although it has been common practice to give compressions if the heart rate is 60 to 80 bpm and the heart rate is not rising, ventilation should be the priority in resuscitation of the newly born. Provision of chest compressions is likely to compete with provision of effective ventilation. Because no scientific data suggests an evidence-based resolution, the ILCOR Working Group recommends that compressions be initiated for a heart rate of <60 bpm based on construct validity (ease of teaching and skill retention).
Compression Technique
Compressions should be delivered on the lower third of the
sternum.57 58 Acceptable techniques are (1) 2 thumbs on
the sternum, superimposed or adjacent to each other according to the
size of the infant, with fingers encircling the chest and supporting
the back (the 2 thumbencircling hands technique), and (2) 2 fingers
placed on the sternum at right angles to the chest with the other hand
supporting the back.59 60 61 Data suggests that the 2
thumbencircling hands technique may offer some advantages in
generating peak systolic and coronary perfusion
pressure and that providers prefer this technique to the 2-finger
technique.59 60 61 62 63 For this reason, we prefer the 2
thumbencircling hands technique for healthcare providers performing
chest compressions in newly born infants and older infants whose size
permits its use (Class IIb, LOE 5).
Consensus of the ILCOR Working Group supports a relative rather than absolute depth of compression (ie, compress to approximately one third of the anterior-posterior diameter of the chest) to generate a palpable pulse. The pediatric basic life support guidelines recommend a relative compression depth of one third to one half of the anterior-posterior dimension of the chest. In the absence of specific data about ideal compression depth, these guidelines recommend compression to approximately one third the depth of the chest, but the compression depth must be adequate to produce a palpable pulse. Deliver compressions smoothly. A compression to relaxation ratio with a slightly shorter compression than relaxation phase offers theoretical advantages for blood flow in the very young infant.64 Keep the thumbs or fingers on the sternum during the relaxation phase.
Coordinate compressions and ventilations to avoid
simultaneous delivery.65 There should be a 3:1
ratio of compressions to ventilations, with 90 compressions and 30
breaths to achieve approximately 120 events per minute. Thus, each
event will be allotted approximately 1/2 second, with exhalation
occurring during the first compression following each ventilation.
Reassess the heart rate approximately every 30 seconds. Continue chest
compressions until the spontaneous heart rate is
60 bpm.
| Medications |
|---|
|
|
|---|
Medications and Volume Expansion
Epinephrine
Administration of epinephrine is indicated when the heart
rate remains <60 bpm after a minimum of 30 seconds of adequate
ventilation and chest compressions (Class I). Epinephrine is
particularly indicated in the presence of asystole.
Epinephrine has both
- and ß-adrenergicstimulating
properties; however, in cardiac arrest,
-adrenergicmediated
vasoconstriction may be the more important action.67
Vasoconstriction elevates the perfusion pressure during chest
compression, enhancing delivery of oxygen to the heart and
brain.68 Epinephrine also enhances the contractile
state of the heart, stimulates spontaneous contractions, and increases
heart rate.
The recommended intravenous or endotracheal dose is 0.1 to 0.3 mL/kg of a 1:10 000 solution (0.01 to 0.03 mg/kg), repeated every 3 to 5 minutes as indicated. The data regarding effects of high-dose epinephrine for resuscitation of newly born infants is inadequate to support routine use of higher doses of epinephrine (Class Indeterminate, LOE 4). Higher doses have been associated with exaggerated hypertension but lower cardiac output in animals.69 70 The sequence of hypotension followed by hypertension likely increases the risk of intracranial hemorrhage, especially in preterm infants.71
Volume Expanders
Volume expanders may be necessary to resuscitate a newly born
infant who is hypovolemic. Suspect hypovolemia in any infant who fails
to respond to resuscitation. Consider volume expansion when there has
been suspected blood loss or the infant appears to be in shock (pale,
poor perfusion, weak pulse) and has not responded adequately to other
resuscitative measures (Class I). The fluid of choice for volume
expansion is an isotonic crystalloid solution such as normal saline or
Ringers lactate (Class IIb, LOE 7). Administration of O-negative red
blood cells may be indicated for replacement of large-volume blood loss
(Class IIb, LOE 7). Albumin-containing solutions are less
frequently used for initial volume expansion because of limited
availability, risk of infectious disease, and an observed association
with increased mortality.72
The initial dose of volume expander is 10 mL/kg given by slow intravenous push over 5 to 10 minutes. The dose may be repeated after further clinical assessment and observation of response. Higher bolus volumes have been recommended for resuscitation of older infants. However, volume overload or complications such as intracranial hemorrhage may result from inappropriate intravascular volume expansion in asphyxiated newly born infants as well as in preterm infants.73 74
Bicarbonate
There is insufficient data to recommend routine use of bicarbonate
in resuscitation of the newly born. In fact, the hyperosmolarity and
CO2-generating properties of sodium bicarbonate
may be detrimental to myocardial or cerebral
function.75 76 77 Use of sodium bicarbonate is discouraged
during brief CPR. If it is used during prolonged arrests unresponsive
to other therapy, it should be given only after establishment of
adequate ventilation and circulation.78 Later use of
bicarbonate for treatment of persistent metabolic acidosis
or hyperkalemia should be directed by
arterial blood gas levels or serum chemistries, among other
evaluations. A dose of 1 to 2 mEq/kg of a 0.5 mEq/mL solution may be
given by slow intravenous push (over at least 2 minutes)
after adequate ventilation and perfusion have been established.
Naloxone
Naloxone hydrochloride is a narcotic antagonist
without respiratory-depressant activity. It is specifically indicated
for reversal of respiratory depression in a newly born infant whose
mother received narcotics within 4 hours of delivery. Always establish
and maintain adequate ventilation before administration of naloxone. Do
not administer naloxone to newly born infants whose mothers are
suspected of having recently abused narcotic drugs because it may
precipitate abrupt withdrawal signs in such infants.
The recommended dose of naloxone is 0.1 mg/kg of a 0.4 mg/mL or 1.0 mg/mL solution given intravenously, endotracheally, orif perfusion is adequateintramuscularly or subcutaneously. Because the duration of action of narcotics may exceed that of naloxone, continued monitoring of respiratory function is essential, and repeated naloxone doses may be necessary to prevent recurrent apnea.
Routes of Medication Administration
The tracheal route is generally the most rapidly accessible route
for drug administration during resuscitation. It may be used for
administration of epinephrine and naloxone, but it should not
be used during resuscitation for administration of caustic agents such
as sodium bicarbonate. The tracheal route of administration may result
in a more variable response to epinephrine than the
intravenous route79 80 81 ; however, neonatal
data is insufficient to recommend a higher dose of epinephrine
for tracheal administration.
Attempt to establish intravenous access in neonates who fail to respond to tracheally administered epinephrine. The umbilical vein is the most rapidly accessible venous route; it may be used for epinephrine or naloxone administration as well as for administration of volume expanders and bicarbonate. Insert a 3.5F or 5F radiopaque catheter so that the tip is just below skin level and a free flow of blood returns on aspiration. Deep insertion poses the risk of infusion of hypertonic and vasoactive medications into the liver. Take care to avoid introduction of air emboli into the umbilical vein.
Peripheral sites for venous access (scalp or peripheral vein) may be adequate but are usually more difficult to cannulate. Naloxone may be given intramuscularly or subcutaneously but only after effective assisted ventilation has been established and only if the infants peripheral circulation is adequate. We do not recommend administration of resuscitation drugs through the umbilical artery because the artery is often not rapidly accessible and complications may result if vasoactive or hypertonic drugs (eg, epinephrine or bicarbonate) are given by this route.
Intraosseous lines are not commonly used in newly born infants because the umbilical vein is more accessible, the small bones are fragile, and the intraosseous space is small in a premature infant. Intraosseous access has been shown to be useful in the neonate and older infant when vascular access is difficult to achieve.82 Intraosseous access can be used as an alternative route for medication/volume expansion if umbilical or other direct venous access is not readily attainable (Class IIb, LOE 5).
| Special Resuscitation Circumstances |
|---|
|
|
|---|
|
Prematurity
The incidence of perinatal depression is markedly increased among
preterm neonates because of the complications associated with preterm
labor and the physiological immaturity and lability
of the preterm infant.83 Diminished lung compliance,
respiratory musculature, and respiratory drive may contribute to the
need for assisted ventilation.
Some experts recommend early elective intubation of extremely preterm infants (eg, <28 weeks of gestation) to help establish an air-fluid interface,84 while others recommend that this be accomplished with oxygen administration via mask or nasal prongs.85 Many infants younger than 30 to 31 weeks will undergo intubation for surfactant administration after the initial stages of resuscitation have been successful.86
A number of factors can complicate resuscitation of the premature infant. Because premature infants have low body fat and a high ratio of surface area to body mass, they are also more difficult to keep warm. Their immature brains and the presence of a fragile germinal matrix predispose them to development of intracranial hemorrhage after episodes of hypoxia or rapid changes in vascular pressure and osmolarity.77 87 88 For this reason, avoid rapid boluses of volume expanders or hyperosmolar solutions.
Multiple Births
Multiple births are more frequently associated with a need for
resuscitation because of abnormalities of placentation, compromise of
cord blood flow, or mechanical complications during delivery.
Monozygotic multiple fetuses may also have abnormalities of blood
volume resulting from interfetal vascular anastomoses.
| Postresuscitation Issues |
|---|
|
|
|---|
Documentation of Resuscitation
Thorough documentation of assessments and resuscitative actions is
essential for good clinical care, for communication, and for
medicolegal concerns. The Apgar scores quantify and summarize the
response of the newly born infant to the extrauterine environment and
to resuscitation (Table 5
).89 90 Assign Apgar scores
at 1 and 5 minutes after birth and then sequentially every 5 minutes
until vital signs have stabilized. The Apgar scores should not dictate
appropriate resuscitative actions, nor should interventions for
depressed infants be delayed until the 1-minute assessment. Complete
documentation must also include a narrative description of
interventions performed and their timing.
|
Continuing Care of the Family
When time permits, the team responsible for care of the newly born
should introduce themselves to the mother and family before delivery.
They should outline the proposed plan of care and solicit the familys
questions. Especially in cases of potentially lethal fetal
malformations or extreme prematurity, the family should be asked to
articulate their beliefs and desires about the extent of resuscitation,
and the team should outline its planned approach (see below).
After delivery the mother continues to be a patient herself, with physical and emotional needs. The team caring for the newly born infant should inform the parents of the infants condition at the earliest opportunity. If resuscitation is necessary, inform the parents of the procedures undertaken and their indications. Encourage the parents to ask questions, and answer their questions as frankly and honestly as possible. Make every effort to enable the parents to have contact with the newly born infant.
| Ethics |
|---|
|
|
|---|
Noninitiation of Resuscitation
The delivery of extremely immature infants and infants with severe
congenital anomalies raises questions about initiation of
resuscitation.91 92 93 Noninitiation of resuscitation in the
delivery room is appropriate for infants with confirmed gestation <23
weeks or birth weight <400 g, anencephaly, or confirmed trisomy 13 or
18. Current data suggests that resuscitation of these newly born
infants is very unlikely to result in survival or survival without
severe disability94 (Class IIb, LOE 5).95
However, antenatal information may be incomplete or unreliable. In
cases of uncertain prognosis, including uncertain gestational age,
resuscitation options include a trial of therapy and noninitiation or
discontinuation of resuscitation after assessment of the infant. In
such cases, initiation of resuscitation at delivery does not mandate
continued support.
Noninitiation of support and later withdrawal of support are generally considered to be ethically equivalent; however, the latter approach allows time to gather more complete clinical information and to provide counseling to the family. Ongoing evaluation and discussion with the parents and the healthcare team should guide continuation versus withdrawal of support. In general, there is no advantage to delayed, graded, or partial support; if the infant survives, outcome may be worsened as a result of this approach.
Discontinuation of Resuscitation
Discontinuation of resuscitative efforts may be appropriate if
resuscitation of an infant with cardiorespiratory arrest does not
result in spontaneous circulation in 15 minutes. Resuscitation of newly
born infants after 10 minutes of asystole is very unlikely to result in
survival or survival without severe disability (Class IIb, LOE
5).96 97 98 99 We recommend local discussions to formulate
guidelines consistent with local resources and outcome
data.
| Footnotes |
|---|
| References |
|---|
|
|
|---|
2.
Palme-Kilander C. Methods of resuscitation in
low-Apgar-score newborn infants: a national survey. Acta
Paediatr. 1992;81:739744.
3.
World Health Report. Geneva, Switzerland:
World Health Organization; 1995.
4.
Perlman JM, Risser R. Cardiopulmonary
resuscitation in the delivery room: associated clinical events.
Arch Pediatr Adolesc Med. 1995;149:2025.
5.
Cummins RO, Chamberlain DA, Abramson NS, et al.
Recommended guidelines for uniform reporting of data from
out-of-hospital cardiac arrest: the Utstein style.
Circulation. 1991;84:960975.
6.
Cummins RO, Chamberlain DA, Hazinski MF, et al.
Recommended guidelines for reviewing, reporting, and conducting
research on in-hospital resuscitation: the in-hospital "Utstein
style." Circulation. 1997;95:22132239.
7.
Zaritsky A, Nadkarni V, Hazinski MF, Foltin G, Quan L,
Wright J, Fiser D, Zideman D, OMalley P, Chameides L, Writing Group.
Recommended guidelines for uniform reporting of pediatric advanced life
support: the pediatric Utstein style: a statement for healthcare
professionals from a task force of the American Academy of Pediatrics,
the American Heart Association, and the European Resuscitation Council.
Circulation. 1995;92:20062020.
8.
Idris AH, Becker LB, Ornato JP, Hedges JR, Bircher NG,
Chandra NC, Cummins RO, Dick W, Ebmeyer U, Halperin HR, Hazinski MF,
Kerber RE, Kern KB, Safar P, Steen PA, Swindle MM, Tsitlik JE, von
Planta I, von Planta M, Wears RL, Weil MH, Writing Group. Utstein-style
guidelines for uniform reporting of laboratory CPR research: a
statement for healthcare professionals from a task force of the
American Heart Association, the American College of Emergency
Physicians, the American College of Cardiology, the
European Resuscitation Council, the Heart and Stroke Foundation of
Canada, the Institute of Critical Care Medicine, the Safar Center for
Resuscitation Research, and the Society for Academic Emergency
Medicine. Circulation. 1996;94:23242336.
9.
Nadkarni V, Hazinski MF, Zideman D, Kattwinkel J, Quan
L, Bingham R, Zaritsky A, Bland J, Kramer E, Tiballs J. Paediatric life
support: an advisory statement by the Paediatric Life Support Working
Group of the International Liaison Committee on Resuscitation.
Resuscitation. 1997;34:115127.
10.
Kattwinkel J, Niermeyer S, Nadkarni V, Tibballs J,
Phillips B, Zideman D, Van Reempts P, Osmond M. ILCOR advisory
statement: resuscitation of the newly born infant: an advisory
statement from the pediatric working group of the International Liaison
Committee on Resuscitation. Circulation. 1999;99:19271938.
11.
Guidelines for cardiopulmonary resuscitation
and emergency cardiac care: Emergency Cardiac Care Committee and
Subcommittees, American Heart Association, part V: pediatric basic life
support [see comments]. JAMA. 1992;268:22512261.
12.
Bloom RS, Cropley C, AHA/AAP Neonatal Resuscitation
Program Steering Committee, American Heart Association. American
Academy of Pediatrics. Textbook of Neonatal Resuscitation/Ronald S.
Bloom, Catherine Cropley, and the AHA/AAP Neonatal Resuscitation
Program Steering Committee [Rev. ed.];1 v. (various pagings): ill.;
28 cm. Elk Grove Village, IL: American Academy of Pediatrics: American
Heart Association; 1994.
13.
Kloeck WGJ, Kramer E. Resuscitation Council of Southern
Africa: new recommendations for BLS in adults, children and infants.
Trauma Emerg Med. 1997;14:1332.
14.
Advanced Life Support Committee of the Australian
Resuscitation Council. Paediatric advanced life support: Australian
Resuscitation Council guidelines: Advanced Life Support Committee of
the Australian Resuscitation Council. Med J Aust. 1996;165:199201, 204206.
15.
European Resuscitation Council. Pediatric basic life
support: to be read in conjunction with the International Liaison
Committee on Resuscitation Pediatric Working Group Advisory Statement
(April 1997). Resuscitation. 1998;37:97100.
16.
European Resuscitation Council. Pediatric advanced life
support: to be read in conjunction with the International Liaison
Committee on Resuscitation Pediatric Working Group Advisory Statement
(April 1997). Resuscitation. 1998;37:101102.
17.
European Resuscitation Council. Recommendations on
resuscitation of babies at birth: to be read in conjunction with the
International Liaison Committee on Resuscitation Pediatric Working
Group Advisory Statement (April 1997). Resuscitation. 1998;37:103110.
18.
Vyas H, Milner AD, Hopkin IE, Boon AW. Physiologic
responses to prolonged slow-rise inflation in the resuscitation of the
asphyxiated newborn infant. J Pediatr.. 1981;99:635639.
19.
Vyas H, Field D, Milner AD, Hopkin IE. Determinants of
the first inspiratory volume and functional residual capacity at birth.
Pediatr Pulmonol. 1986;2:189193.
20.
Jobe A. The respiratory system. In: Fanaroff AA, Martin
RJ, et al, eds. Neonatal Perinatal Medicine. St Louis, Mo:
CV Mosby; 1997:9911018.
21.
Gregory GA, Gooding CA, Phibbs RH, Tooley WH. Meconium
aspiration in infants: a prospective study. J Pediatr. 1974;85:848852.
22.
Peliowski A, Finer NN. Birth asphyxia in the term
infant. In: Sinclair JC, Bracken MB, et al, eds. Effective Care
of the Newborn Infant. Oxford, UK: Oxford University Press;
1992:249273.
23.
Dawes GF. Fetal and Neonatal Physiology: A
Comparative Study of the Changes at Birth. Chicago,
Ill: Year Book Medical Publishers; 1968:149151.
24.
Whitelaw CC, Goldsmith LJ. Comparison of two
techniques for determining the presence of a pulse in an infant
[letter]. Acad Emerg Med. 1997;4:153154.
25.
Theophilopoulos DT, Burchfield DJ. Accuracy of
different methods for heart rate determination during simulated
neonatal resuscitations. J Perinatol. 1998;18:6567.
26.
Gandy GM, Adamson SK Jr, Cunningham N, Silverman WA,
James LS. Thermal environment and acid-base homeostasis in human
infants during the first few hours of life. J Clin
Invest. 1964;43:751758.
27.
Dahm LS, James LS. Newborn temperature and calculated
heat loss in the delivery room. Pediatrics. 1972;49:504513.
28.
Perlman JM. Maternal fever and neonatal depression:
preliminary observations. Clin Pediatr. 1999;38:287291.
29.
Lieberman E, Lang J, Richardson DK, Frigoletto FD,
Heffner LJ, Cohen A. Intrapartum maternal fever and neonatal outcome.
Pediatrics. 2000;105:813.
30.
Vannucci RC, Perlman JM. Interventions for perinatal
hypoxic-ischemic encephalopathy [see comments].
Pediatrics. 1997;100:10041014.
31.
Edwards AD, Wyatt JS, Thoreson M. Treatment of
hypoxic-ischemic brain damage by moderate hypothermia.
Arch Dis Child Fetal Neonatal Ed. 1998;78:F85F88.
32.
Gunn AJ, Gluckman PD, Gunn TR. Selective head cooling
in newborn infants after perinatal asphyxia: a safety study [see
comments]. Pediatrics. 1998;102:885892.
33.
Estol PC, Piriz H, Basalo S, Simini F, Grela C.
Oro-naso-pharyngeal suction at birth: effects on respiratory adaptation
of normal term vaginally born infants. J Perinatal Med. 1992;20:297305.
34.
Cordero L Jr, Hon EH. Neonatal bradycardia following
nasopharyngeal stimulation. J Pediatr. 1971;78:78:441447.
35.
Wiswell TE, Tuggle JM, Turner BS. Meconium aspiration
syndrome: have we made a difference? [see comments].
Pediatrics. 1990;85:85:715721.
36.
Carson BS, Losey RW, Bowes WA Jr, Simmons MA. Combined
obstetric and pediatric approach to prevent meconium aspiration
syndrome. Am J Obstet Gynecol. 1976;126:126:712715.
37.
Locus P, Yeomans E, Crosby U. Efficacy of bulb versus
DeLee suction at deliveries complicated by meconium stained amniotic
fluid [see comments]. Am J Perinatol. 1990;7:8791.
38.
Rossi EM, Philipson EH, Williams TG, Kalhan SC.
Meconium aspiration syndrome: intrapartum and neonatal attributes [see
comments]. Am J Obstet Gynecol. 1989;161:11061110.
39.
Falciglia HS. Failure to prevent meconium aspiration
syndrome. Obstet Gynecol. 1988;71:71:349353.
40.
Greenough A. Meconium aspiration syndrome: prevention
and treatment. Early Hum Dev. 1995;40:955981.
41.
Wiswell TE, Bent RC. Meconium staining and the meconium
aspiration syndrome: unresolved issues. Pediatr Clin North
Am. 1993;40:955981.
42.
Wiswell TE. Meconium in the Delivery Room Trial Group:
delivery room management of the apparently vigorous meconium-stained
neonate: results of the multicenter collaborative trial.
Pediatrics. 2000;105:17.
43.
Linder N, Aranda JV, Tsur M, et al. Need for
endotracheal intubation and suction in meconium-stained neonates.
J Pediatr. 1988;112:613615.
44.
de Burgh Daly M, Angell-James JE, Elsner R. Role of
carotid-body chemoreceptors and their reflex interactions in
bradycardia and cardiac arrest. Lancet. 1979;1:764767.
45.
de Burgh Daly M. Interactions between respiration and
circulation. In: Cherniack NS, Widdicombe JG, eds. Handbook of
Physiology, Section 3, The Respiratory System. Bethesda, Md:
American Physiological Society; 1986:529595.
46.
Rootwelt T, Odden J, Hall C, Ganes T, Saugstad OD.
Cerebral blood flow and evoked potentials during
reoxygenation with 21 or 100% O2
in newborn pigs. J Appl Physiol. 1993;75:20542060.
47.
Ramji S, Ahuja S, Thirupuram S, Rootwelt T, Rooth G,
Saugstad OD. Resuscitation of asphyxic newborn infants with room air or
100% oxygen. Pediatr Res. 1993;34:809812.
48.
Saugstad OD, Rootwellt T, Aalen O. Resuscitation of
asphyxiated newborn infants with room air or oxygen: an international
controlled trial: the Resair 2 Study. Pediatrics. 1998;102:e1.
49.
Kanter RK. Evaluation of mask-bag ventilation in
resuscitation of infants. Am J Dis Child. 1987;141:761763.
50.
Palme C, Nystrom B, Tunell R. An evaluation of the
efficiency of face masks in the resuscitation of newborn infants.
Lancet. 1985;1:207210.
51.
Paterson SJ, Byrne PJ, Molesky MG, Seal RF, Finucane
BT. Neonatal resuscitation using the laryngeal mask airway [see
comments]. Anesthesiology. 1994;80:12481253, discussion
27A.
52.
Gandini D, Brimacombe JR. Neonatal resuscitation with
the laryngeal mask airway in normal and low birth weight infants.
Anesth Analg. 1999;89:642643.
53.
Todres ID, deBros F, Kramer SS, Moylan FM, Shannon DC.
Endotracheal tube displacement in the newborn infant. J
Pediatr. 1976;89:126127.
54.
Rotschild A, Chitayat D, Puterman ML, Phang MS, Ling E,
Baldwin V. Optimal positioning of endotracheal tubes for ventilation of
preterm infants. Am J Dis Child.. 1991;145:10071012.
55.
Aziz HF, Martin JB, Moore JJ. The pediatric end-tidal
carbon dioxide detector role in endotracheal intubation in newborns.
J Perinatol. 1999;19:110113.
56.
Bhende MS, Thompson AE, Orr RA. Utility of an end-tidal
carbon dioxide detector during stabilization and transport of
critically ill children. Pediatrics. 1992;89:10421044.
57.
Orlowski JP. Optimum position for external cardiac
compression in infants and young children. Ann Emerg Med. 1986;15:667673.
58.
Phillips GW, Zideman DA. Relation of infant heart to
sternum: its significance in cardiopulmonary resuscitation.
Lancet. 1986;1:1:10241025.
59.
Thaler MM, Stobie GHC. An improved
technique of external cardiac compression in infants and young
children. N Engl J Med. 1963;269:606610.
60.
David R. Closed chest cardiac massage in the newborn
infant. Pediatrics. 1988;81:552554.
61.
Todres ID, Rogers MC. Methods of external cardiac
massage in the newborn infant. J Pediatr. 1975;86:781782.
62.
Menegazzi JJ, Auble TE, Nicklas KA, Hosack GM, Rack L,
Goode JS. Two-thumb versus two-finger chest compression during CPR in a
swine infant model of cardiac arrest [see comments]. Ann Emerg
Med. 1993;22:240243.
63.
Houri PK, Frank LR, Menegazzi JJ, Taylor R. A
randomized, controlled trial of two-thumb vs two-finger chest
compression in a swine infant model of cardiac arrest [see comment].
Prehosp Emerg Care. 1997;1:6567.
64.
Dean JM, Koehler RC, Schleien CL, Berkowitz I, Michael
JR, Atchison D, Rogers MC, Traystman RJ. Age-related effects of
compression rate and duration in cardiopulmonary resuscitation.
J Appl Physiol. 1990;68:554560.
65.
Berkowitz ID, Chantarojanasiri T, Koehler RC, Schleien
CL, Dean JM, Michael JR, Rogers MC, Traystman RJ. Blood flow during
cardiopulmonary resuscitation with simultaneous
compression and ventilation in infant pigs. Pediatr Res. 1989;26:558564.
66.
Burchfield DJ. Medication use in neonatal
resuscitation. Clin Perinatol. 1999;26:683691.
67.
Zaritsky A, Chernow B. Use of
catecholamines in pediatrics. J Pediatr. 1984;105:341350.
68.
Berkowitz ID, Gervais H, Schleien CL, Koehler RC, Dean
JM, Traystman RJ. Epinephrine dosage effects on cerebral and
myocardial blood flow in an infant swine model of
cardiopulmonary resuscitation. Anesthesiology. 1991;75:10411050.
69.
Berg RA, Otto CW, Kern KB, Hilwig RW, Sanders AB, Henry
CP, Ewy GA. A randomized, blinded trial of high-dose
epinephrine versus standard-dose epinephrine in a swine
model of pediatric asphyxial cardiac arrest. Crit Care Med. 1996;24:16951700.
70.
Burchfield DJ, Preziosi MP, Lucas VW, Fan J. Effect of
graded doses of epinephrine during asphyxia-induced bradycardia
in newborn lambs. Resuscitation. 1993;25:235244.
71.
Pasternak JF, Groothus DR, Fischer JM, Fischer DP.
Regional cerebral blood flow in the beagle puppy model of neonatal
intraventricular hemorrhage: studies during
systemic hypertension. Neurology. 1983;33:559566.
72.
Cochrane Injuries Group Albumin Reviewers.
Human albumin administration in critically ill patients:
systematic review of randomised controlled trials. BMJ. 1998;317:235240.
73.
Usher R, Lind J. Blood volume of the newborn premature
infant. Acta Paediatr Scand. 1965;54:419431.
74.
Funato M, Tamai H, Noma K, et al. Clinical events in
association with timing of intraventricular
hemorrhage in preterm infants. J Pediatr. 1992;121:614619.
75.
Kette F, Weil MH, von Planta M, Gazmuri RJ, Rackow EC.
Buffer agents do not reverse intramyocardial acidosis during cardiac
resuscitation. Circulation. 1990;81:16601666.
76.
Kette F, Weil MH, Gazmuri RJ. Buffer solutions may
compromise cardiac resuscitation by reducing coronary perfusion
pressure [published correction appears in JAMA. 1991;266:3286] [see comments]. JAMA. 1991;266:21212126.
77.
Papile LA, Burstein J, Burstein R, Koffler H, Koops B.
Relationship of intravenous sodium bicarbonate infusions
and cerebral intraventricular hemorrhage.
J Pediatr. 1978;93:834836.
78.
Hein HA. The use of sodium bicarbonate in neonatal
resuscitation: help or harm? Pediatrics. 1993;91:496497.
79.
Lindemann R. Resuscitation of the newborn: endotracheal
administration of epinephrine. Acta Paediatr Scand. 1984;73:210212.
80.
Lucas VW, Preziosi MP, Burchfield DJ.
Epinephrine absorption following endotracheal administration:
effects of hypoxia-induced low pulmonary blood flow.
Resuscitation. 1994;27:3134.
81.
Mullett CJ, Kong JQ, Romano JT, Polak MJ. Age-related
changes in pulmonary venous epinephrine concentration
and pulmonary vascular response after intratracheal
epinephrine. Pediatr Res. 1992;31:458461.
82.
Ellemunter H, Simma B, Trawoger R, Maurer H.
Intraosseous lines in preterm and full term neonates. Arch Dis
Child Fetal Neonatal Ed. 1999;80:F7475.
83.
MacDonald HM, Mulligan JC, Allen AC, Taylor PM.
Neonatal asphyxia, I: relationship of obstetric and neonatal
complications to neonatal mortality in 38,405 consecutive deliveries.
J Pediatr. 1980;96:898902.
84.
Poets CF, Sens B. Changes in intubation rates and
outcome of very low birth weight infants: a population study.
Pediatrics. 1996;98:2427.
85.
Avery ME, Tooley WH, Keller JB, et al. Is chronic lung
disease in low birth weight infants preventable? a survey of eight
centers. Pediatrics. 1987;79:2630.
86.
Kattwinkel J. Surfactant: evolving issues.
Clin Perinatol. 1998;25:1732.
87.
Simmons MA, Adcock EW III, Bard H, Battaglia FC.
Hypernatremia and intracranial hemorrhage in neonates.
N Engl J Med. 1974;291:610.
88.
Hambleton G, Wigglesworth JS. Origin of
intraventricular haemorrhage in the preterm
infant. Arch Dis Child. 1976;51:651659.
89.
Apgar V, James LS. Further observations of the newborn
scoring system. Am J Dis Child. 1962;104:419428.
90.
Chamberlain G, Banks J. Assessment of the Apgar score.
Lancet. 1974;2:12251228.
91.
Byrne PJ, Tyebkhan JM, Laing LM. Ethical
decision-making and neonatal resuscitation. Semin Perinatol. 1994;18:3641.
92.
Davies JM, Reynolds BM. The ethics of
cardiopulmonary resuscitation, I: background to decision
making. Arch Dis Child. 1992;67:14981501.
93.
Landwirth J. Ethical issues in pediatric and neonatal
resuscitation. Ann Emerg Med. 1993;22:502507.
94.
Tyson JE, Younes N, Verter J, Wright LL. Viability,
morbidity, and resource use among newborns of 501- to 800-g birth
weight: National Institute of Child Health and Human Development
Neonatal Research Network. JAMA. 1996;276:16451651.
95.
Finer NN, Horbar JD, Carpenter JH.
Cardiopulmonary resuscitation in the very low birth weight
infant: the Vermont Oxford Network experience. Pediatrics. 1999;104:428434.
96.
Davis DJ. How aggressive should delivery room
cardiopulmonary resuscitation be for extremely low birth weight
neonates? [see comments]. Pediatrics. 1993;92:447450.
97.
Jain L, Ferre C, Vidyasagar D, Nath S, Sheftel D.
Cardiopulmonary resuscitation of apparently stillborn infants:
survival and long-term outcome [see comments]. J
Pediatr. 1991;118:778782.
98.
Yeo CL, Tudehope DI. Outcome of resuscitated apparently
stillborn infants: a ten year review. J Paediatr Child
Health. 1994;30:129133.
99.
Casalaz DM, Marlow N, Speidel BD. Outcome of
resuscitation following unexpected apparent stillbirth. Arch Dis
Child Fetal Neonatal Ed. 1998;78:F112F115.
This article has been cited by other articles:
![]()
![]()

![]()
![]()
![]()
R. D. White
2005 American Heart Association Guidelines for Cardiopulmonary Resuscitation: Physiologic and Educational Rationale for Changes
Mayo Clin. Proc.,
June 1, 2006;
81(6):
736 - 740.
[Full Text]
[PDF]
![]()
![]()
![]()

![]()
![]()
![]()
The International Liaison Committee on Resuscitati
The International Liaison Committee on Resuscitation (ILCOR) Consensus on Science With Treatment Recommendations for Pediatric and Neonatal Patients: Neonatal Resuscitation
Pediatrics,
May 1, 2006;
117(5):
e978 - e988.
[Full Text]
[PDF]
![]()
![]()
![]()

![]()
![]()
![]()
M. F. Hazinski, V. M. Nadkarni, R. W. Hickey, R. O'Connor, L. B. Becker, and A. Zaritsky
Major Changes in the 2005 AHA Guidelines for CPR and ECC: Reaching the Tipping Point for Change
Circulation,
December 13, 2005;
112(24_suppl):
IV-206 - IV-211.
[Full Text]
[PDF]
![]()
![]()
![]()

![]()
![]()
![]()
Part 7: Neonatal Resuscitation
Circulation,
November 29, 2005;
112(22_suppl):
III-91 - III-99.
[Full Text]
[PDF]
![]()
This Article ![]()
![]()
Extract
![]()
Alert me when this article is cited
![]()
Alert me if a correction is posted
![]()
Services ![]()
![]()
Email this article to a friend
![]()
Similar articles in this journal
![]()
Similar articles in PubMed
![]()
Alert me to new issues of the journal
![]()
Download to citation manager
![]()
Request Permissions ![]()
Citing Articles ![]()
![]()
Citing Articles via HighWire
![]()
Citing Articles via Google Scholar
![]()
Google Scholar ![]()
![]()
Search for Related Content
![]()
PubMed ![]()
![]()
PubMed Citation
![]()
Related Collections ![]()
![]()
AHA Statements and Guidelines
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2000 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |