(Circulation. 2000;102:813.)
© 2000 American Heart Association, Inc.
Basic Science Reports |
From the Service de Cardiologie Pédiatrique, Hôpital Necker Enfants Malades, Paris, France.
Correspondence to Dr Philipp Bonhoeffer, Service de Cardiologie Pédiatrique, Hôpital Necker Enfants Malades, 149, rue de Sèvres, 75743 Paris Cedex 15, France. E-mail philipp.bonhoeffer{at}nck.ap-hop-paris.fr
| Abstract |
|---|
|
|
|---|
Methods and ResultsA fresh bovine jugular vein containing a native valve was sutured into a vascular stent and then cross-linked with a 0.6% glutaraldehyde solution for 36 hours. After being hand-crimped onto a balloon catheter, the device was inserted percutaneously according to standard stent-placing techniques. The valved stent was finally deployed in the position of the native pulmonary valve of the lamb. Hemodynamic evaluation was carried out before and 2 months after implantation. Anatomic evaluation was finally performed. Percutaneous pulmonary valve replacement was successful in 5 lambs. No complications were noted. Early and late angiographic and hemodynamic studies confirmed a good position of the stents with a competent valve at the end of the protocol. One stent was slightly stenotic, with macroscopically visible calcifications.
ConclusionsNonsurgical implantation of pulmonary valves is possible in the lamb. This new technique is similar to standard stent implantation. Thus, it should be feasible in humans, in whom it will lead to a significant reduction of reoperations in patients in need of pulmonary valve replacement.
Key Words: : stents catheterization prosthesis valves
| Introduction |
|---|
|
|
|---|
Surgical indications for pulmonary valve replacement in these patients are delicate. Cardiopulmonary bypass, in fact, aggravates right ventricular dysfunction, and clinical benefit for the patient solely through the reduction of the pulmonary regurgitant volume cannot be guaranteed.
We set out to develop a nonsurgical technique for placement of a biological valve in the pulmonary position through a percutaneous approach. Here, we report our initial experience of percutaneous pulmonary valve implantation in lambs.
| Methods |
|---|
|
|
|---|
Expandable Vascular Platinum Stent
The radially expandable vascular stent includes several
cylindrical fine wire sections that are interconnected into a single
tubular structure. The wires are made of a soft and highly malleable
alloy consisting of
90% platinum and 10% iridium.
Biological Valve
A jugular vein was harvested from a fresh bovine cadaver, and a
section of the vein containing a native valve was selected for
preparation. The valves were bicuspid or tricuspid, with extremely thin
leaflets. Thereafter, the venous wall of the harvested section was
trimmed to remove extraneous material and to reduce profile. However,
sufficient material had to remain to allow safe attachment of the
valved venous segment to the stent.
Valved Stent
Before the biological valve was attached, the intravascular
stent was preexpanded to a radial diameter of 18 mm. The
biological valve was then sutured to the interior of the stent with a
7-0 polypropylene thread. The tubular portion of the jugular vein
spanned the entire axial length of the stent (Figure 1
).
|
In Vitro Testing
Fluid passing into the valved stent in the appropriate
direction filled the pouch-like valvar sinuses defined by each of the
leaflets. In this way, the valve leaflets coapted inside the segment of
the vein, thereby forming an effective valve closure. Fluid flow in the
opposite direction was not restricted (Figure 2
). The valvar competence was tested
before and after crimping and reexpansion by the balloon catheter.
|
Sterilization of Valved Stent
The valved stent was finally cross-linked with a buffered saline
solution containing 0.6% glutaraldehyde for 36 hours
at 4°C. After fixation, it was transferred to a 60% ethanol solution
for storage. Before implantation, the valve assembly was immersed 3
times and carefully soaked in a physiological
saline solution to remove the ethanol.
Preparation of Animals
Eleven lambs 2 months old and weighing 16 to 18 kg
underwent catheterization for transcatheter
pulmonary valve replacement. All lambs were treated according
to the European regulations for animal experimentation.3
Anesthesia was induced with 10 mg/kg of thiopental
sodium and maintained with halothane in mechanically ventilated
lambs. Cardiac and respiratory functions were monitored throughout the
procedure. The right internal jugular vein was prepared for
catheterization. Heparin (100 IU/kg) and penicillin
were administered prophylactically. After the procedure,
the jugular vein was preserved to allow for repeat
catheterizations. Anticoagulant therapy was continued
with low-molecular-weight heparin once a day and substituted 48
hours after stent implantation by low-dose aspirin.
Implantation and Deployment Strategy
Through the internal jugular approach, a guidewire was
positioned in the pulmonary artery. The valved stent was then
hand-crimped onto the inflatable portion of an 18-, 20-, or 22-mm
balloon catheter. The size of the balloon was dependent on the size of
the pulmonary artery measured by angiography. To limit the risk
of dislodgment or slipping of the balloon, the assembly was
front-loaded in a 16F Mullins long sheath (Arrow), as shown in Figure
1. After the loading onto the previously positioned guidewire,
the whole system was advanced and percutaneously
implanted. The position of the valved stent was easily tracked
fluoroscopically because of the highly radiopaque material of the
stent. The valved stent was then balloon-expanded and deployed in the
native pulmonary valve of the lamb to impinge upon the function
of this native valve and to fix the device on the pulmonary
wall. The balloon was subsequently deflated, and the catheter was
removed, leaving the replacement valve assembly in the desired
position.
Cardiac Catheterization, Angiographic Studies, and
Hemodynamic Evaluation
The right ventricular and pulmonary
arterial pressures were measured before and after the
implantation and before the animal was euthanized.
Angiography was performed before the procedure to locate the precise position of the native pulmonary valve, after the procedure to confirm the appropriate position of the stent and to verify the function of the implanted valve, and at the end of the protocol.
Graft Retrieval
All grafts were explanted 2 months after implantation. Before
harvesting, heparin (100 IU/kg IV) was administered. The valved stent
was rinsed with a saline solution to remove excess intraluminal blood.
All grafts were inspected macroscopically, and in vitro testing of
valvar competence was repeated.
| Results |
|---|
|
|
|---|
In the 7 lambs that received implants, 5 stents were in the desired position, impinging on the function of the native valve of the lamb. In the remaining 2, the stent was implanted proximal to the valve in one and distal in the other.
No complications were noted during the procedure or the follow-up. There were no infectious processes. The mild fever that occurred in 2 animals disappeared within 48 hours without any antibiotic treatment other than the penicillin prophylaxis. No early or late migrations were observed.
The pulmonary systolic, diastolic, and mean
pressures before and after implantation and before retrieval are shown
in Figure 3
. Six of the 7 valved stents
were angiographically competent just after the procedure. The early
evaluation of the seventh revealed a small insufficiency, probably due
to the catheter in the pulmonary valve during the angiographic
evaluation. One valved stent was mildly stenotic, with a
15 mm Hg pressure gradient between the right ventricle and the
pulmonary artery.
|
In the 5 animals with a stent in the correct position, macroscopic findings at autopsy showed that the native pulmonary valve was nonfunctional because it was stuck between the pulmonary wall and the stent. The extremities of the stent were embedded into the pulmonary wall and the right ventricular outflow, respectively. Normal function of the native pulmonary valve was found in the animal with proximal placement of the stent. The valve inside the stent was completely covered by pannus ingrowth. The function of this valve could be restored in vitro after removal of the pannus. Similarly, the stent placed distal to the native valve interfered only partially with this valve. The valve in the stent showed one normally functioning leaflet, with the other being covered by pannus. The in vitro function of this valve could also be restored after removal of this fibrous tissue.
Among the 5 well-placed stents, 4 revealed transparent, mobile, and
competent valves (Figure 4
). One stent
was slightly stenotic and showed macroscopically visible
calcifications.
|
| Discussion |
|---|
|
|
|---|
Surgical pulmonary valve replacement is usually done with either homograft material from the human or the pulmonary valve from a pig.8 In addition, the bovine pericardium has been widely used to construct pulmonary valves, and more recently, the native bovine jugular valve has been used as a valved conduit with very promising results.9
We had the idea of fixing a bovine jugular valve in an expandable stent designed for percutaneous placement. We sutured a valved venous segment into a stent. Thereafter, we verified the function of the resulting valved stent after reducing it in diameter and reexpanding it by a balloon catheter in our bench testing. The ability to reduce the external wall of the vein allowed us to reduce the diameter of the stent to a very low profile, without interfering with the function of the valve.
Recent improvements in interventional catheterization and stent technology have made percutaneous stent placement a routine procedure in pediatric as well as adult cardiology and radiology. More than 20 stent designs have been developed to fit the multiple indications.10 Stents are used extensively in many cardiovascular applications, including stent placement in prosthetic conduits.11 More recently, covered stents have enlarged the armamentarium of stent designs.12 The outside layer on the covered stents led to a considerable increase in the diameter of these devices and to the use of big introducer sheaths. Therefore, experience with 16F introducers, as required for the placement of a valve on an 18- to 22-mm balloon, has already been acquired in other applications.
We succeeded in deploying 5 valved stents precisely in the native pulmonary valve of the lambs. In another 2 animals, stent delivery was possible, but the position of the stent was suboptimal, allowing for a residual function of the native pulmonary valve of the animal. In the remaining 4 lambs, we failed to implant the stent because of the narrow angle between the tricuspid valve and the right ventricular outflow. Unfortunately, the femoral vein of the lambs is of small size because of their small hind limbs. This forced us to use the jugular venous approach. However, the large size of the femoral vein in the human should allow for a straighter catheter course, which should reduce this technical difficulty.
The study was designed to continue for 2 months, considering the somatic growth of the lambs. In this period, all the lambs were completely asymptomatic and nearly doubled their body weight. At the end of the protocol, hemodynamic evaluation showed normal pulmonary pressure in all lambs in which stents were implanted. In the 5 lambs with correct position of the stent in particular, diastolic pressures in the pulmonary artery were as high as before implantation when the normal native valves were functioning. This confirmed the adequate diastolic valvar function throughout the time of the study. Moreover, only 1 animal had a mild systolic pressure gradient of 15 mm Hg through the stent between the right ventricle and the pulmonary artery.
Pulmonary regurgitation occurred in 1 lamb, but angiographs were obtained with a catheter that passed through the valve. This made the interpretation of the pulmonary regurgitation difficult. After euthanasia, the in vitro testing confirmed a transparent, highly mobile, and well-functioning valve in 4 of the 5 lambs in which the stent position was accurate. The valve that was mildly stenotic showed macroscopic alterations with evidence of calcification. Interestingly, the 2 valved stents that were in suboptimal position had clear dysfunction of the valve. In 1, the function of the native valve of the lamb was entirely preserved, whereas the valve in the stent was embedded in fibrous tissue inside the stent. The stent, which only partially impeded the function of the native valve, showed 1 normally functioning leaflet, with the other being covered by the same fibrous reaction. After explantation, the valvar function of both stents could be restored for the bench testing by the simple removal of the fibrous tissue.
The degeneration of bioprosthetic valves has been widely studied in recent years.13 Various etiological factors are involved that influence the durability of the valve to different degrees. The actuarial functional life of small valves (<25 mm) is shorter than that of larger ones. The role of ventricular systolic pressure in accelerating the degenerative process of bioprosthetic valves is also widely reported.14 After 2 months of growth, our lambs essentially doubled their body weight. Therefore, the implanted stent inflated to 18 mm could have become stenotic, leading to an increased ventricular systolic pressure. Furthermore, the chemical treatment of the valves for sterilization obviously plays a major role in the durability of the valves and their tendency to calcify. The presence of degenerative processes and early calcification in 1 of the valves is likely to be due to the suboptimal sterilization process we used in our protocol.13 15
As far as degeneration of the valve is concerned, the inconvenience is the same between the surgical and the interventional approaches. The specific risks of the valved stent placement (dislodgment, balloon rupture, etc) are similar to the ones described in the literature for normal stent placement.16 Therefore, they are likely to be very acceptable and less important than the damage from cardiopulmonary bypass associated with the risks of reintervention. A not fully expanded stent implanted in a child can potentially be redilated to adapt its size to the somatic growth without interfering with valvar function. We verified this with in vitro testing. Indeed, we expanded our valved stents to different sizes from 10 to 28 mm; the valves were completely competent and functioning well at all sizes. The use of bigger jugular veins allowed bigger diameters if needed. Moreover, in the lamb with the 15 mm Hg systolic pressure gradient between the right ventricle and the pulmonary artery, we dilated the stenosis effectively without interfering with the closure of the valve, as confirmed by the hemodynamic profile of the diastolic pressure.
Degenerated and stenotic valves in conduits, or better, in valved stents, potentially allow for a second valved stent implantation without the need for surgery. We explanted the valved stent at the end of the study and showed that surgical stent removal is possible and therefore, surgical options are not hampered by even multiple stent implantations. The morbidity of reoperation might also decrease after improvement of the right ventricular function by preliminary valved stent implantation.
In conclusion, the future opportunity to implant pulmonary valves percutaneously in humans will lead to a significant reduction of reoperations in patients suffering from pulmonary valvar incompetence in need of a valve replacement. Although the implantation of such a valve is technically difficult in the lamb, application in humans should not fundamentally differ from routine stent implantation. The function of the implanted heterografts appears to be identical to bioprosthetic valves implanted surgically. Therefore, we think that percutaneous pulmonary valve implantation in humans should be possible in the very near future.
| Acknowledgments |
|---|
Received February 14, 2000; revision received March 16, 2000; accepted March 16, 2000.
| References |
|---|
|
|
|---|
2. Bove EL, Byrum CJ, Thomas FD, et al. The influence of pulmonary insufficiency on ventricular function following repair of tetralogy of Fallot: evaluation using radionuclide ventriculography. J Thorac Cardiovasc Surg. 1983;85:691696.[Abstract]
3. Convention européenne sur la protection des animaux vertébrés utilisés à des fins expérimentales ou à dautres fins scientifiques. Journal Officiel des Communautés Européennes. L222/29 à L222/37. August 24, 1999.
4. Shimazaki Y, Blackstone EH, Kirklin JW. The natural history of isolated congenital pulmonary valve incompetence: surgical implications. J Thorac Cardiovasc Surg. 1984;32:257259.
5. Bove EL, Kavey RE, Byrum CJ, et al. Improved right ventricular function following late pulmonary valve replacement for residual pulmonary insufficiency or stenosis. J Thorac Cardiovasc Surg. 1985;90:5055.[Abstract]
6. Connelly MS, Webb GD, Somerville J, et al. Canadian Consensus Conference on Adult Congenital Heart Disease 1996. Can J Cardiol. 1998;14:395452.[Medline] [Order article via Infotrieve]
7. Zhao HX, Miller DC, Reitz BA, et al. Surgical repair of tetralogy of Fallot: long-term follow-up with particular emphasis on late death and reoperation. J Thorac Cardiovasc Surg. 1985;89:204220.[Abstract]
8. Kawachi Y, Masuda M, Tominaga R, et al. Comparative study between St. Jude Medical and bioprosthetic valves in the right side of the heart. Jpn Circ J. 1991;55:553562.[Medline] [Order article via Infotrieve]
9. Brawn WJ. The use of a glutaraldehyde-preserved ovine pulmonary valve, as a pulmonary valve substitute in infants. Semin Thorac Cardiovasc Surg. 1995;7:154156.[Medline] [Order article via Infotrieve]
10. Oesterle SN, Whitbourn R, Fitzgerald PJ, et al, for the Stanford Stent Summit faculty. The stent decade: 1987 to 1997. Am Heart J. 1998;136:578599.[Medline] [Order article via Infotrieve]
11. Saliba Z, Bonhoeffer P, Aggoun Y, et al. Traitement des obstructions des prothèses tubulaires par implantation percutanée de stents. Arch Mal Coeur Vaiss. 1999;92:591596.[Medline] [Order article via Infotrieve]
12.
Kato N, Dake MD, Miller DC, et al. Traumatic thoracic
aortic aneurysm: treatment with endovascular stent-grafts.
Radiology. 1997;205:657662.
13.
Rao KP, Shanthi C. Reduction of calcification by
various treatments in cardiac valves. J Biomater Appl. 1999;13:238268.
14. Ilbawi MN, Idriss FS, DeLeon SY, et al. Long-term results of porcine valve insertion for pulmonary regurgitation following repair of tetralogy of Fallot. Ann Thorac Surg. 1986;41:478482.[Abstract]
15.
Grabenwoger M, Sider J, Fitzal F, et al. Impact of
glutaraldehyde on calcification of pericardial
bioprosthetic heart valve material. Ann Thorac Surg. 1996;62:772777.
16.
Shaffer KM, Mullins CE, Grifka RG, et al. Intravascular
stents in congenital heart disease: short- and long-term results from a
large single-center experience. J Am Coll Cardiol. 1998;31:661667.
This article has been cited by other articles:
![]() |
G.-J. Zong, Y. Bai, H.-B. Jiang, W.-P. Li, H. Wu, X.-X. Zhao, and Y.-W. Qin Use of a novel valve stent for transcatheter pulmonary valve replacement: An animal study. J. Thorac. Cardiovasc. Surg., June 1, 2009; 137(6): 1363 - 1369. [Abstract] [Full Text] [PDF] |
||||
![]() |
Z. Astroulakis and J. M. Hill Does 'SWITCHing' from heparin to bivalirudin monotherapy result in a bleeding benefit? Eur. Heart J. Suppl., June 1, 2009; 11(suppl_C): C13 - C18. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Piazza, A. Asgar, R. Ibrahim, and R. Bonan Transcatheter Mitral and Pulmonary Valve Therapy J. Am. Coll. Cardiol., May 19, 2009; 53(20): 1837 - 1851. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Suzuki, A. C. Yeung, and F. Ikeno The Pre-Clinical Animal Model in the Translational Research of Interventional Cardiology J. Am. Coll. Cardiol. Intv., May 1, 2009; 2(5): 373 - 383. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. H. Huber, M. Hurni, V. Tsang, and L. K. von Segesser Valved stents for transapical pulmonary valve replacement J. Thorac. Cardiovasc. Surg., April 1, 2009; 137(4): 914 - 918. [Abstract] [Full Text] [PDF] |
||||
![]() |
G.-W. Meng, J.-Y. Zhou, Y. Tang, Z.-K. Ye, Y. Zhang, G.-M. Liu, and S.-S. Hu Off-Pump Pulmonary Valve Implantation of a Valved Stent With an Anchoring Mechanism. Ann. Thorac. Surg., February 1, 2009; 87(2): 597 - 601. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. T.L. Chiam and C. E. Ruiz Percutaneous Transcatheter Aortic Valve Implantation: Assessing Results, Judging Outcomes, and Planning Trials: The Interventionalist Perspective J. Am. Coll. Cardiol. Intv., August 1, 2008; 1(4): 341 - 350. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Grube, G. Schuler, L. Buellesfeld, U. Gerckens, A. Linke, P. Wenaweser, B. Sauren, F.-W. Mohr, T. Walther, B. Zickmann, et al. Percutaneous Aortic Valve Replacement for Severe Aortic Stenosis in High-Risk Patients Using the Second- and Current Third-Generation Self-Expanding CoreValve Prosthesis: Device Success and 30-Day Clinical Outcome J. Am. Coll. Cardiol., July 3, 2007; 50(1): 69 - 76. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Nordmeyer, S. Khambadkone, L. Coats, S. Schievano, P. Lurz, G. Parenzan, A. M. Taylor, J. E. Lock, and P. Bonhoeffer Risk Stratification, Systematic Classification, and Anticipatory Management Strategies for Stent Fracture After Percutaneous Pulmonary Valve Implantation Circulation, March 20, 2007; 115(11): 1392 - 1397. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. Antunes Off-pump aortic valve replacement with catheter-mounted valved stents.: Is the future already here? Eur. J. Cardiothorac. Surg., January 1, 2007; 31(1): 1 - 3. [Full Text] [PDF] |
||||
![]() |
E. Grube, J. C. Laborde, U. Gerckens, T. Felderhoff, B. Sauren, L. Buellesfeld, R. Mueller, M. Menichelli, T. Schmidt, B. Zickmann, et al. Percutaneous Implantation of the CoreValve Self-Expanding Valve Prosthesis in High-Risk Patients With Aortic Valve Disease: The Siegburg First-in-Man Study Circulation, October 10, 2006; 114(15): 1616 - 1624. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. L.S. Morales, B. E. Braud, K. S. Gunter, K. E. Carberry, K. A. Arrington, J. S. Heinle, E. D. McKenzie, and C. D. Fraser Jr Encouraging results for the Contegra conduit in the problematic right ventricle-to-pulmonary artery connection. J. Thorac. Cardiovasc. Surg., September 1, 2006; 132(3): 665 - 671. [Abstract] [Full Text] [PDF] |
||||
![]() |
U.A. Stock, I. Degenkolbe, T. Attmann, K. Schenke-Layland, S. Freitag, and G. Lutter Prevention of device-related tissue damage during percutaneous deployment of tissue-engineered heart valves J. Thorac. Cardiovasc. Surg., June 1, 2006; 131(6): 1323 - 1330. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. O. Shebani, S. McGuirk, M. Baghai, J. Stickley, J. V. De Giovanni, F. A. Bu'Lock, D. J. Barron, and W. J. Brawn Right ventricular outflow tract reconstruction using Contegra(R) valved conduit: natural history and conduit performance under pressure Eur. J. Cardiothorac. Surg., March 1, 2006; 29(3): 397 - 405. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. G. Pennington The Impact of New Technology on Cardiothoracic Surgical Practice Ann. Thorac. Surg., January 1, 2006; 81(1): 10 - 18. [Full Text] [PDF] |
||||
![]() |
Y. Boudjemline, E. Pineau, N. Borenstein, L. Behr, and P. Bonhoeffer New insights in minimally invasive valve replacement: description of a cooperative approach for the off-pump replacement of mitral valves Eur. Heart J., October 1, 2005; 26(19): 2013 - 2017. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Attmann, T. Jahnke, R. Quaden, A. Boening, S. Muller-Hulsbeck, J. Cremer, and G. Lutter Advances in Experimental Percutaneous Pulmonary Valve Replacement Ann. Thorac. Surg., September 1, 2005; 80(3): 969 - 975. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Khambadkone, L. Coats, A. Taylor, Y. Boudjemline, G. Derrick, V. Tsang, J. Cooper, V. Muthurangu, S. R. Hegde, R. S. Razavi, et al. Percutaneous Pulmonary Valve Implantation in Humans: Results in 59 Consecutive Patients Circulation, August 23, 2005; 112(8): 1189 - 1197. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Ma, P. Tozzi, C. H. Huber, S. Taub, G. Gerelle, and L. K. von Segesser Double-crowned valved stents for off-pump mitral valve replacement Eur. J. Cardiothorac. Surg., August 1, 2005; 28(2): 194 - 198. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Boudjemline, G. Agnoletti, D. Bonnet, L. Behr, N. Borenstein, D. Sidi, and P. Bonhoeffer Steps Toward the Percutaneous Replacement of Atrioventricular Valves: An Experimental Study J. Am. Coll. Cardiol., July 19, 2005; 46(2): 360 - 365. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. A. Vassiliades Jr, P. C. Block, L. H. Cohn, D. H. Adams, J. S. Borer, T. Feldman, D. R. Holmes, W. K. Laskey, B. W. Lytle, M. J. Mack, et al. The Clinical Development of Percutaneous Heart Valve Technology: A Position Statement of the Society of Thoracic Surgeons (STS), the American Association for Thoracic Surgery (AATS), and the Society for Cardiovascular Angiography and Interventions (SCAI) Endorsed by the American College of Cardiology Foundation (ACCF) and the American Heart Association (AHA) J. Am. Coll. Cardiol., May 3, 2005; 45(9): 1554 - 1560. [Full Text] [PDF] |
||||
![]() |
American College of Cardiology Foundation (ACCF) a, T. A. Vassiliades Jr, P. C. Block, L. H. Cohn, D. H. Adams, J. S. Borer, T. Feldman, D. R. Holmes, W. K. Laskey, B. W. Lytle, et al. The clinical development of percutaneous heart valve technology: A position statement of the Society of Thoracic Surgeons (STS), the American Association for Thoracic Surgery (AATS), and the Society for Cardiovascular Angiography and Interventions (SCAI) J. Thorac. Cardiovasc. Surg., May 1, 2005; 129(5): 970 - 976. [Full Text] [PDF] |
||||
![]() |
B. Lytle and M. Mack The Future of Cardiac Surgery: The Times, They Are a Changin' Ann. Thorac. Surg., May 1, 2005; 79(5): 1470 - 1472. [Full Text] [PDF] |
||||
![]() |
T. A. Vassiliades Jr, P. C. Block, L. H. Cohn, D. H. Adams, J. S. Borer, T. Feldman, D. R. Holmes, W. K. Laskey, B. W. Lytle, M. J. Mack, et al. The Clinical Development of Percutaneous Heart Valve Technology: A Position Statement of The Society of Thoracic Surgeons (STS), the American Association for Thoracic Surgery (AATS), and the Society for Cardiovascular Angiography and Interventions (SCAI) Ann. Thorac. Surg., May 1, 2005; 79(5): 1812 - 1818. [Full Text] [PDF] |
||||
![]() |
M. Pawelec-Wojtalik, L. K. von Segesser, M. Liang, and D. Bukowska Closure of left ventricle perforation with the use of muscular VSD occluder Eur. J. Cardiothorac. Surg., April 1, 2005; 27(4): 714 - 716. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Boudjemline, S. Schievano, C. Bonnet, L. Coats, G. Agnoletti, S. Khambadkone, D. Bonnet, J. Deanfield, D. Sidi, and P. Bonhoeffer Off-pump replacement of the pulmonary valve in large right ventricular outflow tracts: A hybrid approach J. Thorac. Cardiovasc. Surg., April 1, 2005; 129(4): 831 - 837. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Meyns, L. Van Garsse, D. Boshoff, B. Eyskens, L. Mertens, M. Gewillig, S. Fieuws, E. Verbeken, and W. Daenen The Contegra conduit in the right ventricular outflow tract induces supravalvular stenosis J. Thorac. Cardiovasc. Surg., December 1, 2004; 128(6): 834 - 840. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Lutter, R. Ardehali, J. Cremer, and P. Bonhoeffer Percutaneous Valve Replacement: Current State and Future Prospects Ann. Thorac. Surg., December 1, 2004; 78(6): 2199 - 2206. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Pawelec-Wojtalik, P. Antosik, G. Wasiatycz, and M. Wojtalik Use of muscular VSD Amplatzer occluder for closing right ventricular free wall perforation after hybrid procedure Eur. J. Cardiothorac. Surg., November 1, 2004; 26(5): 1044 - 1046. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. D. Fish Percutaneous Heart Valve Replacement: Enthusiasm Tempered Circulation, October 5, 2004; 110(14): 1876 - 1878. [Full Text] [PDF] |
||||
![]() |
A. Frigiola, A.N. Redington, S. Cullen, and M. Vogel Pulmonary Regurgitation Is an Important Determinant of Right Ventricular Contractile Dysfunction in Patients With Surgically Repaired Tetralogy of Fallot Circulation, September 14, 2004; 110(11_suppl_1): II-153 - II-157. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Vahanian and I. F. Palacios Percutaneous Approaches to Valvular Disease Circulation, April 6, 2004; 109(13): 1572 - 1579. [Full Text] [PDF] |
||||
![]() |
Y. Boudjemline, G. Agnoletti, D. Bonnet, D. Sidi, and P. Bonhoeffer Percutaneous pulmonary valve replacement in a large right ventricular outflow tract: An experimental study J. Am. Coll. Cardiol., March 17, 2004; 43(6): 1082 - 1087. [Abstract] [Full Text] [PDF] |
||||
![]() |
Z. M. Hijazi Transcatheter valve replacement: a new era of percutaneous cardiac intervention begins J. Am. Coll. Cardiol., March 17, 2004; 43(6): 1088 - 1089. [Full Text] [PDF] |
||||
![]() |
A. Cribier, H. Eltchaninoff, C. Tron, F. Bauer, C. Agatiello, L. Sebagh, A. Bash, D. Nusimovici, P. Y. Litzler, J.-P. Bessou, et al. Early experience with percutaneous transcatheter implantation of heart valve prosthesis for the treatment of end-stage inoperable patients with calcific aortic stenosis J. Am. Coll. Cardiol., February 18, 2004; 43(4): 698 - 703. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Q. Zhou, A. F. Corno, C. H. Huber, P. Tozzi, and L. K. von Segesser Self-expandable valved stent of large size: off-bypass implantation in pulmonary position Eur. J. Cardiothorac. Surg., August 1, 2003; 24(2): 212 - 216. [Abstract] [Full Text] [PDF] |
||||
![]() |
I-S. Kang, A. N. Redington, L. N. Benson, C. Macgowan, E. R. Valsangiacomo, K. Roman, C. J. Kellenberger, and S.-J. Yoo Differential Regurgitation in Branch Pulmonary Arteries After Repair of Tetralogy of Fallot: A Phase-Contrast Cine Magnetic Resonance Study Circulation, June 17, 2003; 107(23): 2938 - 2943. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. F. Corno, J. Zhou, P. Tozzi, and L. K. von Segesser Off-bypass implantation of a self-expandable valved stent between inferior vena cava and right atrium Interactive CardioVascular and Thoracic Surgery, June 1, 2003; 2(2): 166 - 169. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Kuehne, M. Saeed, C. B. Higgins, K. Gleason, G. A. Krombach, O. M. Weber, A. J. Martin, D. Turner, D. Teitel, and P. Moore Endovascular Stents in Pulmonary Valve and Artery in Swine: Feasibility Study of MR Imaging-guided Deployment and Postinterventional Assessment Radiology, February 1, 2003; 226(2): 475 - 481. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Cribier, H. Eltchaninoff, A. Bash, N. Borenstein, C. Tron, F. Bauer, G. Derumeaux, F. Anselme, F. Laborde, and M. B. Leon Percutaneous Transcatheter Implantation of an Aortic Valve Prosthesis for Calcific Aortic Stenosis: First Human Case Description Circulation, December 10, 2002; 106(24): 3006 - 3008. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Herijgers, S. Ozaki, E. Verbeken, A. Van Lommel, B. Meuris, E. Lesaffre, W. Daenen, and W. Flameng Valved jugular vein segments for right ventricular outflow tract reconstruction in young sheep J. Thorac. Cardiovasc. Surg., October 1, 2002; 124(4): 798 - 805. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. K. von Segesser 'Surgical technique versus technology for surgery': a plea for an open mind towards new technology Interactive CardioVascular and Thoracic Surgery, September 1, 2002; 1(1): 1 - 3. [Full Text] [PDF] |
||||
![]() |
Y. Boudjemline and P. Bonhoeffer Percutaneous implantation of a valve in the descending aorta in lambs Eur. Heart J., July 1, 2002; 23(13): 1045 - 1049. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Bonhoeffer, Y. Boudjemline, S. A. Qureshi, J. Le Bidois, L. Iserin, P. Acar, J. Merckx, J. Kachaner, and D. Sidi Percutaneous insertion of the pulmonary valve J. Am. Coll. Cardiol., May 15, 2002; 39(10): 1664 - 1669. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Lutter, D. Kuklinski, G. Berg, P. von Samson, J. Martin, M. Handke, P. Uhrmeister, and F. Beyersdorf Percutaneous aortic valve replacement: An experimental study. I. Studies on implantation J. Thorac. Cardiovasc. Surg., April 1, 2002; 123(4): 768 - 776. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Boudjemline and P. Bonhoeffer Steps Toward Percutaneous Aortic Valve Replacement Circulation, February 12, 2002; 105(6): 775 - 778. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. BOUDJEMLINE and P. BONHOEFFER Percutaneous aortic valve replacement: will we get there? Heart, December 1, 2001; 86(6): 705a - 706. [Full Text] |
||||
![]() |
T. Kuehne, M. Saeed, G. Reddy, H. Akbari, K. Gleason, D. Turner, D. Teitel, P. Moore, and C. B. Higgins Sequential Magnetic Resonance Monitoring of Pulmonary Flow With Endovascular Stents Placed Across the Pulmonary Valve in Growing Swine Circulation, November 6, 2001; 104(19): 2363 - 2368. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2000 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |