(Circulation. 2000;102:3028.)
© 2000 American Heart Association, Inc.
Brief Rapid Communication |
From Centro Cuore Columbus, Milan, Italy, and the Department of Cardiology, Heart Center, Siegburg, Germany (E.G.).
Correspondence to Antonio Colombo, MD, Centro Cuore Columbus, Via M. Buonarroti, 48, I-20145, Milan, Italy. E-mail columbus{at}micronet.it
| Abstract |
|---|
|
|
|---|
Methods and ResultsSince May 1997 (when polytetrafluoroethylene [PTFE]-covered stents were available at our institutions), 11 of the 12 consecutive patients who had coronary ruptures that were unsuccessfully sealed with prolonged balloon inflation and reversal of anticoagulation were treated with 12 PTFE-covered stents (PTFE group). The efficacy of the PTFE-covered stent was compared with that of noncovered stents, which were used to treat 17 perforations (non-PTFE group). One patient sustained a very distal perforation that was not suitable for covered stent sealing and underwent emergency surgery. All vessel ruptures treated with PTFE-covered stent implantation were successfully sealed. The time necessary to deploy the stent was 10±3 minutes (range, 4 to 15 minutes). All patients but one were discharged from the hospital and had an optimal early clinical outcome. One patient underwent emergency bypass surgery and died in the intensive care unit. The occurrence of cardiac tamponade and the necessity for emergency surgery was significantly lower in the PTFE group than in the non-PTFE group. At 14±4 months, the 10 discharged patients had not experienced any major adverse cardiac events.
ConclusionsThis preliminary study supports the utility of the PTFE-covered stent for the nonsurgical treatment of vessel ruptures.
Key Words: angioplasty coronary disease hemorrhage stents
| Introduction |
|---|
|
|
|---|
| Methods |
|---|
|
|
|---|
Coronary perforations were classified according to method of Ellis et al.1 Two-dimensional echocardiography and pulsed-wave Doppler were performed to assess the presence and severity of pericardial effusion.
Management of Vessel Perforation
A conventional strategy to treat perforation includes
prolonged perfusion balloon inflation and reverse-heparin
anticoagulation with protamine to reach an activated clotting time
<130 s; a platelet transfusion is also administered if abciximab was
given. If cardiac tamponade occurs, pericardiocentesis is
performed.
When the conventional strategy was not able to seal perforations, the eventual nonsurgical strategy was, in the non-PTFE group, noncovered stent implantation when the patient did not already have a stent at the perforation site and the procedure seemed feasible and, in the PTFE group, PTFE-covered stent implantation.
We used the JOSTENT stent (JOMED), a balloon-expandable, slotted-tube stent manufactured by sandwiching a layer of PTFE between 2 stents, as our PTFE-covered stent. The potential diameter of the stent is 2.5 to 5.0 mm. The stent is currently Prior to European Community approval (CE mark), but the protocol was approved by the local ethics committees. After stent implantation, angiographic optimization was performed with high-pressure balloon dilation to achieve <20% residual stenosis by visual estimate. In 6 patients (50%), the final stent apposition was evaluated by intravascular ultrasound (3.2-F monorail system with a 30-MHz transducer-tipped catheter; Ultracross Cardiovascular Imaging System, Boston Scientific Co.). Angiographic measurements were performed with an automated computer-based system (Cardiovascular Measurement System, MEDIS).
Monitoring After Successful Nonsurgical
Management
Antiplatelet therapy consisted of 325 mg/d aspirin
indefinitely and 250 mg of ticlopidine BID for 3 months. Serial
echocardiography was performed to detect pericardial effusion. Major
adverse cardiac events (MACEs), which included death, Q- and
nonQ-wave myocardial infarction, and coronary bypass surgery, were
assessed. NonQ-wave myocardial infarction was defined as chest pain
or ST-segment or T-wave abnormalities associated with an increase of
total creatine kinase
2 times above normal values, with a positive
creatine kinase-MB fraction without any new pathological Q waves.
Clinical follow-up was obtained by an interview or a telephone
conversation with the patient. Exercise tests were routinely performed
at 1 to 2 months after the procedure. Repeated angiography was
performed at 6 months after stent placement or earlier if clinically
indicated. Restenosis was defined as a stenosis
50% at the
angiographic follow-up.
Statistical Analysis
Differences between groups were assessed by
2 analysis for categorical variables and
Students t test for
continuous variables. P<0.05
was significant.
| Results |
|---|
|
|
|---|
|
|
|
Non-PTFE Group
To seal the coronary rupture, a noncovered stent
(Palmaz-Schatz) was implanted at the perforation
site in 5 of the 17 cases
(Tables 1
and 2
). This strategy was successful in only 2
cases (40%). In the remaining 15 cases, urgent bypass surgery was
necessary.
Clinical Outcome
The in-hospital outcomes of the patients in the PTFE
and non-PTFE groups are summarized in
Table 2
. The in-hospital MACE rate was significantly lower
in the PTFE group. In particular, PTFE-covered stent implantation
reduced the occurrence of cardiac tamponade and the necessity of bypass
surgery. Pericardial or mediastinal effusion without hemodynamic
impairment was identified by echocardiography in 5 PTFE patients and 17
non-PTFE patients (P<0.001).
At 14±4 months (range, 6 to 26 months) , the 10 PTFE patients
discharged from the hospital had not experienced any MACEs. Exercise
tests were performed by all 10 patients and were positive in 2 of them.
Angiographic follow-up was performed in 7 of the 9 eligible patients
(78%) at 6±2 months. Two patients had angiographic restenosis (29%);
one of them underwent repeat percutaneous transluminal coronary
angioplasty.
| Discussion |
|---|
|
|
|---|
Covered Stents
In the past few years, covered stents have become an
alternative to surgery when conservative approaches fail. Some case
reports have shown promising results with autologous venous-covered
stents.6 8 However,
this strategy is technically demanding and may be difficult to perform
in emergency situations like coronary rupture. Of note, the time
interval from vein harvest to covered stent deployment has been
reported as between 20 and 45
minutes.6 8 In
contrast, PTFE-covered stent implantation is easy, rapid, and does not
require extensive training.
In our study, PTFE-covered stents successfully sealed 91% of coronary perforations after other conservative approaches failed. Compared with the non-PTFE group, PTFE-covered stent significantly reduced the necessity for emergency surgery. Time to deploy the PTFE-covered stent and, subsequently, to seal the perforation was relatively short, and this may explain why cardiac tamponade occurred in only one patient. Furthermore, none of the 10 patients effectively treated with the PTFE-covered stent implantation experienced any MACEs at follow-up, and angiographic restenosis was 29% (2 of 7 patients). The lack of late thrombo-occlusive events9 in the PTFE group could be due to (1) the frequent use (50%) of intravascular ultrasound to assess proper stent implantation, (2) the routine use of final high pressure balloon inflation, (3) the prolonged (3 months) ticlopidine administration after the procedure, and (4) the fact that this event has a very low frequency of occurrence and it was evaluated in a small group.
Study Limitations
Because of the small sample size and the nonrandomized
nature of the study, no firm conclusions can be drawn about the
incidence of restenosis and late stent thrombosis. The limited
flexibility of this stent restricts its implantation to lesions
accessible with no or moderate tortuosity. Furthermore, this device
prevents any access to side
branchs.
| Acknowledgments |
|---|
Received August 3, 2000; revision received October 17, 2000; accepted October 18, 2000.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
J. M. Wilson and J. T. Willerson Myocardial Revascularization with Percutaneous Devices Card. Surg. Adult, January 1, 2008; 3(2008): 573 - 598. [Full Text] |
||||
![]() |
J.M. van Opstal, G. Geskes, F. Eerens, S. Aydin, and J. Vainer Venous graft perforation leading to cardiac tamponade Eur J Echocardiogr, January 1, 2008; 9(1): 162 - 162. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. A. Bittl Reducing the Risk of Emergency Bypass Surgery for Failed Percutaneous Coronary Interventions J. Am. Coll. Cardiol., December 6, 2005; 46(11): 2010 - 2012. [Full Text] [PDF] |
||||
![]() |
F Airoldi, C Di Mario, G Stankovic, C Briguori, M Carlino, A Chieffo, F Liistro, M Montorfano, P Pagnotta, V Spanos, et al. Clinical and angiographic outcome of directional atherectomy followed by stent implantation in de novo lesions located at the ostium of the left anterior descending coronary artery Heart, September 1, 2003; 89(9): 1050 - 1054. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Colombo, G. Stankovic, and J. W. Moses Selection of coronary stents J. Am. Coll. Cardiol., September 18, 2002; 40(6): 1021 - 1033. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Singh, H. H. Ting, P. B. Berger, K. N. Garratt, D. R. Holmes Jr, and B. J. Gersh Rationale for on-site cardiac surgery for primary angioplasty: a time for reappraisal J. Am. Coll. Cardiol., June 19, 2002; 39(12): 1881 - 1889. [Abstract] [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2000 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |