(Circulation. 2000;102:2823.)
© 2000 American Heart Association, Inc.
Clinical Investigation and Reports |
From the Department of Internal Medicine II (S.A., D.R., T.G., W.G.D., W.M.), Institute of Medical Physics (S.U., M.K., W.A.K.), and Department of Diagnostic Radiology (U.B., W.B.), University of Erlangen-Nürnberg, Germany.
Correspondence to PD Dr Stephan Achenbach, Medizinische Klinik II mit Poliklinik, Universität Erlangen-Nürnberg, Östliche Stadtmauerstr 29, D-91054 Erlangen, Germany. E-mail stephan.achenbach{at}rzmail.uni-erlangen.de
| Abstract |
|---|
|
|
|---|
Methods and ResultsTwenty-five patients in sinus rhythm (significant coronary artery stenoses ruled out by invasive angiography) were studied with a multislice spiral CT (Siemens SOMATOM Volume Zoom). In inspiration (mean breathhold, 37 seconds), a volume data set of the heart was acquired (intravenous contrast agent; 4x1-mm slice thickness; 500-ms rotation; table feed, 1.5 mm/360°). Simultaneous recording of the ECG permitted retrospective reconstruction of contiguous cross sections in intervals of 1 mm at any desired interval of the cardiac cycle. The mean duration of the image reconstruction window was 185 ms. Next to 3-dimensional reconstructions of the heart and coronary arteries, multiplanar reconstructions were rendered to determine the visualized length of the coronary arteries, the contrast-to-noise ratio, and the correlation of coronary artery diameters to quantitative coronary angiography.
ConclusionsThe coronary arteries could be visualized over long segments (left main, 9±4 mm; left anterior descending, 112±34 mm; left circumflex, 80±29 mm; right coronary artery, 116±33 mm). On average, 78±16% of these distances were visualized free of motion artifacts. The mean contrast-to-noise ratio was 9.3±3.3. Coronary artery diameters in multislice spiral CT showed close correlation to quantitative coronary angiography (CT, 3.3±1.0 mm; angiography, 3.2±0.9 mm; mean difference, 0.38 mm; r=0.86). Contrast-enhanced multislice spiral CT permits visualization of the coronary artery lumen. Further studies are necessary to determine whether image quality is sufficient to reliably detect coronary artery stenoses.
Key Words: imaging coronary disease tomography
| Introduction |
|---|
|
|
|---|
| Methods |
|---|
|
|
|---|
50% diameter
reduction) had been ruled out. All patients had sinus rhythm (heart
rate, 51 to 115 bpm; mean, 77 bpm) and were in a clinically stable
condition. The mean weight of the patients was 80.2 kg (range, 56 to
115 kg). All patients gave written informed consent, and the study
protocol was approved by the institutional review board.
Multislice Spiral CT
Multislice spiral CT (SOMATOM Volume Zoom, Siemens
AG) was performed 1 to 3 days after invasive angiography. Medications
were maintained between the invasive angiogram and the multislice
spiral CT study. All image acquisitions were performed in inspiratory
breathhold preceded by mild hyperventilation. First, a noncontrast
localization scan was performed that yielded an anteroposterior view of
the chest; it was used to position the imaging volume for coronary
artery imaging, which extended from the carina to
1 cm below the
diaphragmal face of the heart. In a second step, a bolus of 20 mL of
contrast agent (Iohexol, Ultravist350, Schering AG) was injected
intravenously at 4 mL/s via an 18-gauge catheter placed in the cubital
vein. After a delay of 10 seconds, a sequence of 10 axial images at the
level of the carina was acquired with an interval of 2 seconds between
subsequent images. From the time interval between contrast agent
injection and acquisition of the image with peak attenuation in the
aortic root, the contrast agent transit time was determined. In a third
step, the volume data set for coronary artery visualization was
acquired in spiral mode, with simultaneous acquisition of 4 parallel
slices with 1.0-mm collimation each. The gantry rotation time was 500
ms, and the patient table was continuously advanced at 3.0 mm/s. The
tube current was 150 mA. During the scan, the patients ECG was
digitized and continuously recorded.
Contrast agent (160 mL) was injected into the cubital vein at 4 mL/s, and the CT scan was initiated with a delay according to the previously determined contrast agent transit time. Using the PC program WinDose (Wellhöfer Dosimetrie/Scanditronix Medical), we estimated the effective radiation dose of the complete investigation to be between 3.9 mSv (men) and 5.8 mSv (women).
Image Reconstruction
The CT raw data were transferred to an offline,
PC-based workstation. Next to ungated image reconstruction with a
conventional spiral CT algorithm (180° Linear Interpolation;
180°LI),29
cross-sectional images with a slice thickness of 1.2 to 1.4 mm were
reconstructed in 1-mm intervals with the new algorithm
180°MCI.27 In
short, the algorithm collects spiral data from successive heart beats
by applying a filter function not only in the z
direction (along the patients longitudinal axis) but also in the
heart phase. The planar data set acquired by this means then undergoes
the normal, filtered backprojection process for image reconstruction in
CT. The width of the image reconstruction window is expressed through
the full-width tenth-maximum of the phase contribution
profile,27 which
strongly depends on the patients heart rate and was between 125 and
243 ms in our study (mean, 185±41 ms; see
Figure 1
). The field of view was 180 mm, with an image
matrix of 512x512 pixels. For each patient, 10 data sets were created.
For the first data set, the image acquisition window was centered
around the peak of the R wave. For the second data set, the center was
positioned at 10% of the R-R interval, and for the subsequent data
sets, the center of the image acquisition window was moved toward the
end of the cardiac cycle in increments of 10%. Because reconstruction
of one cross-sectional image required
2 seconds, 30 to 45 minutes
was necessary to create all 10 data sets. All image data sets were
inspected concerning the presence of motion artifacts, and for each
coronary artery, the data set that contained the fewest motion
artifacts was chosen for further evaluation.
|
Data Evaluation
The reconstructed images were transferred to a
commercially available image processing workstation (NetraMD, ScImage)
and evaluated by a single blinded observer. As measures of image
quality, the visualized length of each coronary artery, the
contrast-to-noise ratio (CNR), and the correlation of coronary artery
diameters in multislice spiral CT and in quantitative x-ray coronary
angiography were determined. For each artery (left main, left anterior
descending, left circumflex, and right coronary artery), a curved
multiplanar reconstruction was rendered as previously
described.30 In the
multiplanar reconstructions, the visualized length of each coronary
artery was measured. In addition, the length of the vessel segments
that were visually judged to be free of motion artifacts was
determined. The CNR was determined for each coronary artery by
measuring the mean CT density within the proximal artery lumen and the
mean CT density in a region of interest in the connective tissue
immediately next to the coronary artery and dividing the respective
difference by the image noise (SD of CT density in a region of interest
placed in the aortic root). Finally, the diameters of the left main and
proximal left anterior descending, left circumflex, and right coronary
artery were determined immediately before or after bifurcation points
that could be identified in both multislice spiral CT and invasive
coronary angiography. Diameters were measured by use of the principle
of "full-width
half-maximum":31
The multiplanar reconstructions were displayed with a window of 1
Hounsfield unit and a level that corresponded to one half of the peak
CT density measured in the proximal coronary artery segment. The length
of the shortest possible line that could be drawn across the coronary
artery lumen was considered to represent the coronary artery
diameter.
In addition, 3-dimensional reconstructions were created with the "volume rendering technique" after manual segmentation of the data sets to remove structures that would obstruct the view onto the coronary arteries, such as the chest wall, pulmonary trunk, and atrial appendages.
Quantitative Coronary Angiography
The x-ray coronary angiograms of all patients were
evaluated by means of QCA. The tip of the angiography catheter was used
for calibration, and the vessel diameters of the coronary segments
corresponding to those used for determination of the artery diameter in
multislice spiral CT were measured in 2 orthogonal planes with
automated contour detection. The mean of the 2 obtained values was
considered to represent the actual vessel diameter at that site.
Diameters were measured at the same time instant in the cardiac cycle
as in multislice spiral CT.
| Results |
|---|
|
|
|---|
Using retrospective ECG gating with the image reconstruction
algorithm 180°MCI, we could suppress cardiac motion artifacts in most
images (see
Figure 2
). The cardiac phase that yielded optimal image
quality varied strongly from patient to patient and between different
coronary arteries. For the left main and left anterior descending
coronary arteries, optimal image quality was usually achieved if the
data acquisition window was centered at 70% or 80% of the cardiac
cycle, whereas for the left circumflex and right coronary arteries, the
most frequent "optimal" position of the data acquisition window was
at 50% of the cardiac cycle (see
the
Table
). In 3-dimensional reconstructions, it was
possible to visualize the anatomy of the heart and coronary arteries
(see
Figures 3
and 4
). The mean visualized length of the left main
coronary artery was 9±4 mm (range, 4 to 17 mm). The mean visualized
length of the left anterior descending coronary was 112±34 mm (range,
52 to 155 mm); left circumflex coronary artery, 80±29 mm (range, 22 to
139 mm); and right coronary artery, 116±33 mm (range, 45 to 176 mm).
Artifacts caused by coronary artery motion were not always fully
suppressed and usually affected the mid right coronary artery and
distal segments of the left anterior descending coronary artery and
left circumflex coronary artery (see
Figure 5
). Although the left main coronary artery was free
of visible motion artifact in all cases, the mean proportion of the
left anterior descending coronary artery that was visualized free of
motion artifact was 79±16% (range, 55% to 100%). The mean
proportion of the left circumflex coronary artery that was visualized
free of motion artifact was 79±17% (range, 41% to 100%), and on
average, the right coronary artery could be visualized free of motion
artifact over 75±14% (range, 54% to 100%) of its entire
length.
|
|
|
|
|
The mean CNR determined for the proximal coronary arteries was 9.3±3.3 (left main, 8.9±4.1; left anterior descending, 9.9±3.4; left circumflex, 9.3±2.7; and right coronary artery, 9.3±2.8).
Coronary artery diameters were compared between multislice
spiral CT and QCA at 100 points (see
Figure 6
). The mean coronary artery diameter in multislice
spiral CT was 3.3±1.0 mm (range, 1.5 to 7.6 mm); the mean diameter in
QCA was 3.2±0.9 mm (range, 1.6 to 6.2 mm). The mean absolute
difference between the CT and angiographic measurement was 0.38±0.35
mm (range, 0.0 to 1.4 mm). There was no significant difference between
diameters in QCA and multislice spiral CT. The slope of the regression
line was 0.87, the intercept was 0.32, and the correlation coefficient
was 0.86.
|
| Discussion |
|---|
|
|
|---|
One major drawback of retrospectively ECG-gated spiral
CT is the fact that radiation is continuously applied, whereas image
reconstruction is restricted to data acquired during only a fraction of
the cardiac cycle. It is doubtful whether prospective triggering
algorithmsyet to be developedwould lead to a substantial reduction
in radiation exposure, because the optimal positioning of the image
reconstruction window used varies from patient to patient and between
the different coronary arteries. Based on our data, only a relatively
small portion of the cardiac cycle (
20%) could potentially be
excluded from data acquisition. Also, coronary artery visualization by
spiral CT requires intravenous injection of iodinated contrast agent,
which may lead to side effects and necessitates perfect timing of image
acquisition to contrast injection. Minor reductions in the amount of
contrast injection could be achieved by exactly adapting the amount of
contrast used to the duration of data acquisition, which depends on the
craniocaudal distance of the volume data set, instead of using a
standard dose of 160 mL as we did in our study protocol. Another
limitation is the relatively long breathhold. At the cost of increased
slice thickness, however, the speed of table movement could be
increased, leading to a shorter overall scan duration required to cover
the volume of the heart. Further investigations are necessary to
determine the optimal balance between breathhold duration and slice
thickness, which translates into image resolution along the patients
longitudinal axis.
In conclusion, multislice spiral CT with retrospective ECG gating is an interesting new approach for noninvasive coronary artery imaging. It could be demonstrated that the coronary arteries can be visualized with high spatial resolution and good overall image quality, although artifacts caused by motion, especially of the right coronary artery, can so far not be sufficiently eliminated. In the future, the accuracy of this approach concerning the detection of coronary artery stenoses will have to be determined, and the method needs to be compared against other evolving and established techniques for coronary artery visualization, such as MRI and EBT.
Received June 9, 2000; revision received July 20, 2000; accepted July 21, 2000.
| References |
|---|
|
|
|---|
2. Davidson CJ, Fishman RF, Bonow RO. Cardiac catheterization. In: Braunwald E, ed. Heart Disease: A Textbook of Cardiovascular Medicine. Philadelphia, Pa: WB Saunders; 1997;177203.
3.
Manning
WJ, Li W, Boyle NG, et al. Fat-suppressed breath-hold magnetic
resonance coronary angiography. Circulation. 1993;87:94104.
4.
Pennell
DJ, Keegan J, Firmin DN, et al. Magnetic resonance imaging of coronary
arteries: technique and preliminary results. Br Heart
J. 1993;70:315326.
5.
Duerinckx
AJ, Urman MK. Two-dimensional coronary MR angiography: analysis of
initial clinical results. Radiology. 1994;193:731738.
6. Hofman MBM, Paschal CB, Li D, et al. MRI of coronary arteries: 2D breath-hold vs 3D respiratory-gated acquisition. J Comput Assist Tomogr. 1995;19:5662.[Medline] [Order article via Infotrieve]
7.
Wielopolski
PA, van Geuns RJM, de Feyter PJ, et al. Breath-hold coronary MR
angiography with volume targeted imaging. Radiology. 1998;209:209219.
8.
Stuber
M, Botnar RM, Danias PG, et al. Double-oblique free-breathing high
resolution three-dimensional coronary magnetic resonance coronary
angiography. J Am Coll Cardiol. 1999;34:524531.
9.
van
Geuns RJM, de Bruin HG, Rensing BJWM, et al. Magnetic resonance imaging
of the coronary arteries: clinical results from three-dimensional
evaluation of a respiratory gated technique. Heart. 1999;82:515519.
10. Botnar RM, Stuber M, Kissinger KV, et al. Free-breathing 3D coronary MRA: the impact of "isotropic" image resolution. J Magn Reson Imaging. 2000;11:389393.[Medline] [Order article via Infotrieve]
11.
Moshage
W, Achenbach S, Seese B, et al. Coronary artery stenoses:
three-dimensional imaging with electrocardiographically triggered,
contrast agent-enhanced, electron-beam CT. Radiology. 1995;196:707714.
12.
Chernoff
DM, Ritchie CJ, Higgins CB. Evaluation of electron beam CT coronary
angiography in healthy subjects. Am J Roentgenol. 1997;169:9399.
13. Nakanishi T, Ito K, Imazu M, et al. Evaluation of coronary artery stenoses using electron-beam CT and multiplanar reformation. J Comput Assist Tomogr. 1997;21:121127.[Medline] [Order article via Infotrieve]
14.
Schmermund
A, Rensing BJ, Sheedy PF, et al. Intravenous electron-beam computed
tomographic coronary angiography for segmental analysis of coronary
artery stenoses. J Am Coll Cardiol. 1998;31:15471554.
15.
Reddy
G, Chernoff DM, Adams JR, et al. Coronary artery stenoses: assessment
with contrast-enhanced electron-beam CT and axial reconstructions.
Radiology. 1998;208:167172.
16.
Rensing
BJ, Bongaerts A, van Geuns RJ, et al. Intravenous coronary angiography
by electron beam computed tomography: a clinical evaluation.
Circulation. 1998;98:25092512.
17.
Achenbach
S, Moshage W, Ropers D, et al. Value of electron-beam computed
tomography for the detection of high-grade coronary artery stenoses and
occlusions. N Engl J Med. 1998;339:19641971.
18. Budoff MJ, Oudiz RJ, Zalace CP, et al. Intravenous three-dimensional coronary angiography using contrast-enhanced electron beam computed tomography. Am J Cardiol. 1999;83:840845.[Medline] [Order article via Infotrieve]
19.
Daniel
WG, Döhring W, Stender H-S, et al. Value and limitations of computed
tomography in assessing aortocoronary bypass graft patency.
Circulation. 1983;67:983987.
20. Tello R, Costello P, Ecker C, et al. Spiral CT evaluation of coronary artery bypass graft patency. J Comput Assist Tomogr. 1993;17:253259.[Medline] [Order article via Infotrieve]
21. Steiner P, Phillips F, Stiel G, et al. The value of spiral computed tomography in the imaging of aortocoronary bypass vessels. Rofo Fortschr Geb Rontgenstr Nuklearmed. 1994;160:2329.
22. Kachelrieß M, Kalender WA. ECG-correlated image reconstruction from subsecond spiral computed tomography scans of the heart. Med Phys. 1998;25:24172431.[Medline] [Order article via Infotrieve]
23.
Woodhouse
CE, Janowitz WR, Viamonte M. Coronary arteries: retrospective cardiac
gating technique to reduce cardiac motion artifact at spiral CT.
Radiology. 1997;204:566569.
24. Becker CR, Knez A, Jakobs TF, et al. Detection and quantification of coronary artery calcification with electron-beam and conventional CT. Eur Radiol. 1999;9:620624.[Medline] [Order article via Infotrieve]
25.
Mochizuki
T, Murase K, Koyama Y, et al. Images in cardiovascular medicine: LAD
stenosis detected by subsecond spiral CT. Circulation. 1999;99:1523.
26. Kachelrieß M, Ulzheimer S, Kalender WA. ECG-correlated imaging of the heart with subsecond multislice spiral CT. IEEE Trans Med Imaging 2000. In press.
27. Kachelrieß M, Ulzheimer S, Kalender WA. ECG-correlated image reconstruction from subsecond multislice spiral CT scans of the heart. Med Phys. 2000;27:1881-1902.[Medline] [Order article via Infotrieve]
28. Ohnesorge B, Flohr T, Becker C, et al. Cardiac imaging with rapid, retrospectively ECG synchronized multilevel spiral CT. Radiologe. 2000;40:111117.[Medline] [Order article via Infotrieve]
29.
Kalender
W, Seissler W, Klotz E, et al. Spiral volumetric CT with
single-breath-hold technique, continuous transport and continuous
scanner rotation. Radiology. 1990;176:181183.
30.
Achenbach
S, Moshage W, Ropers D, et al. Curved multiplanar reconstructions for
the evaluation of contrast-enhanced electron beam CT of the coronary
arteries. Am J Roentgenol. 1998;170:895899.
31. Rumberger JA. Principles and practice of cardiac edge-detection and border definition using ultrafast computed tomography. In: Stanford W, Rumberger JA, eds. Ultrafast Computed Tomography: Principles and Practice. Mount Kisco, NY: Futura Publishing Co; 1992:2536.
This article has been cited by other articles:
![]() |
S. Motoyama, M. Sarai, H. Harigaya, H. Anno, K. Inoue, T. Hara, H. Naruse, J. Ishii, H. Hishida, N. D. Wong, et al. Computed tomographic angiography characteristics of atherosclerotic plaques subsequently resulting in acute coronary syndrome. J. Am. Coll. Cardiol., June 30, 2009; 54(1): 49 - 57. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Donnino, J. E. Jacobs, J. V. Doshi, E. M. Hecht, D. C. Kim, J. S. Babb, and M. B. Srichai Dual-Source Versus Single-Source Cardiac CT Angiography: Comparison of Diagnostic Image Quality Am. J. Roentgenol., April 1, 2009; 192(4): 1051 - 1056. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. M. Miller, C. E. Rochitte, M. Dewey, A. Arbab-Zadeh, H. Niinuma, I. Gottlieb, N. Paul, M. E. Clouse, E. P. Shapiro, J. Hoe, et al. Diagnostic Performance of Coronary Angiography by 64-Row CT N. Engl. J. Med., November 27, 2008; 359(22): 2324 - 2336. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Nagao, H. Matsuoka, H. Kawakami, H. Higashino, T. Mochizuki, K. Murase, and M. Uemura Quantification of Myocardial Perfusion by Contrast-Enhanced 64-MDCT: Characterization of Ischemic Myocardium Am. J. Roentgenol., July 1, 2008; 191(1): 19 - 25. [Abstract] [Full Text] [PDF] |
||||
![]() |
X. Fei, X. Du, Q. Yang, Y. Shen, P. Li, J. Liao, and K. Li 64-MDCT Coronary Angiography: Phantom Study of Effects of Vascular Attenuation on Detection of Coronary Stenosis Am. J. Roentgenol., July 1, 2008; 191(1): 43 - 49. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Voros Can Computed Tomography Angiography of the Coronary Arteries Characterize Atherosclerotic Plaque Composition?: Is the CAT (Scan) Out of the Bag? J. Am. Coll. Cardiol. Intv., April 1, 2008; 1(2): 183 - 185. [Full Text] [PDF] |
||||
![]() |
C. L. Hollingsworth, T. T. Yoshizumi, D. P. Frush, F. P. Chan, G. Toncheva, G. Nguyen, C. R. Lowry, and L. M. Hurwitz Pediatric Cardiac-Gated CT Angiography: Assessment of Radiation Dose Am. J. Roentgenol., July 1, 2007; 189(1): 12 - 18. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Ropers, F.-K. Pohle, A. Kuettner, T. Pflederer, K. Anders, W. G. Daniel, W. Bautz, U. Baum, and S. Achenbach Diagnostic Accuracy of Noninvasive Coronary Angiography in Patients After Bypass Surgery Using 64-Slice Spiral Computed Tomography With 330-ms Gantry Rotation Circulation, November 28, 2006; 114(22): 2334 - 2341. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Achenbach Computed Tomography Coronary Angiography J. Am. Coll. Cardiol., November 21, 2006; 48(10): 1919 - 1928. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. Budoff, S. Achenbach, R. S. Blumenthal, J. J. Carr, J. G. Goldin, P. Greenland, A. D. Guerci, J. A.C. Lima, D. J. Rader, G. D. Rubin, et al. Assessment of Coronary Artery Disease by Cardiac Computed Tomography: A Scientific Statement From the American Heart Association Committee on Cardiovascular Imaging and Intervention, Council on Cardiovascular Radiology and Intervention, and Committee on Cardiac Imaging, Council on Clinical Cardiology Circulation, October 17, 2006; 114(16): 1761 - 1791. [Full Text] [PDF] |
||||
![]() |
J. R. Crouse III Thematic review series: Patient-Oriented Research. Imaging atherosclerosis: state of the art J. Lipid Res., August 1, 2006; 47(8): 1677 - 1699. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Nikolaou, A. Knez, C. Rist, B. J. Wintersperger, A. Leber, T. Johnson, M. F. Reiser, and C. R. Becker Accuracy of 64-MDCT in the diagnosis of ischemic heart disease. Am. J. Roentgenol., July 1, 2006; 187(1): 111 - 117. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Nikolaou, C. Rist, B. J. Wintersperger, T. F. Jakobs, R. van Gessel, M. A. Kirchin, A. Knez, F. von Ziegler, M. F. Reiser, and C. R. Becker Clinical value of MDCT in the diagnosis of coronary artery disease in patients with a low pretest likelihood of significant disease. Am. J. Roentgenol., June 1, 2006; 186(6): 1659 - 1668. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Zanzonico, L. N. Rothenberg, and H. W. Strauss Radiation Exposure of Computed Tomography and Direct Intracoronary Angiography: Risk Has its Reward J. Am. Coll. Cardiol., May 2, 2006; 47(9): 1846 - 1849. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. R. Coles, M. A. Smail, I. S. Negus, P. Wilde, M. Oberhoff, K. R. Karsch, and A. Baumbach Comparison of Radiation Doses From Multislice Computed Tomography Coronary Angiography and Conventional Diagnostic Angiography J. Am. Coll. Cardiol., May 2, 2006; 47(9): 1840 - 1845. [Abstract] [Full Text] [PDF] |
||||
![]() |
C Peebles Computed tomographic coronary angiography: how many slices do you need? Heart, May 1, 2006; 92(5): 582 - 584. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. Hoffmann, F. Moselewski, K. Nieman, I.-K. Jang, M. Ferencik, A. M. Rahman, R. C. Cury, S. Abbara, H. Joneidi-Jafari, S. Achenbach, et al. Noninvasive Assessment of Plaque Morphology and Composition in Culprit and Stable Lesions in Acute Coronary Syndrome and Stable Lesions in Stable Angina by Multidetector Computed Tomography J. Am. Coll. Cardiol., April 18, 2006; 47(8): 1655 - 1662. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. A.S. Cordeiro and J. A.C. Lima Atherosclerotic plaque characterization by multidetector row computed tomography angiography. J. Am. Coll. Cardiol., April 18, 2006; 47(8 Suppl): C40 - C47. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Jahnke, I. Paetsch, S. Achenbach, B. Schnackenburg, R. Gebker, E. Fleck, and E. Nagel Coronary MR Imaging: Breath-hold Capability and Patterns, Coronary Artery Rest Periods, and {beta}-Blocker Use Radiology, April 1, 2006; 239(1): 71 - 78. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Hausleiter, T. Meyer, M. Hadamitzky, E. Huber, M. Zankl, S. Martinoff, A. Kastrati, and A. Schomig Radiation Dose Estimates From Cardiac Multislice Computed Tomography in Daily Practice: Impact of Different Scanning Protocols on Effective Dose Estimates Circulation, March 14, 2006; 113(10): 1305 - 1310. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Alkadhi, S. Wildermuth, D. A. Bettex, A. Plass, B. Baumert, S. Leschka, L. M. Desbiolles, B. Marincek, and T. Boehm Mitral Regurgitation: Quantification with 16-Detector Row CT--Initial Experience Radiology, December 21, 2005; (2005) 2381042216. [Abstract] [Full Text] |
||||
![]() |
G. Kohl The Evolution and State-of-the-Art Principles of Multislice Computed Tomography Proceedings of the ATS, December 1, 2005; 2(6): 470 - 476. [Abstract] [Full Text] [PDF] |
||||
![]() |
N E Manghat, G J Morgan-Hughes, A J Marshall, and C A Roobottom Multidetector row computed tomography: imaging congenital coronary artery anomalies in adults Heart, December 1, 2005; 91(12): 1515 - 1522. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Achenbach, D. Ropers, F.-K. Pohle, D. Raaz, J. von Erffa, A. Yilmaz, G. Muschiol, and W. G. Daniel Detection of coronary artery stenoses using multi-detector CT with 16x0.75 collimation and 375 ms rotation Eur. Heart J., October 1, 2005; 26(19): 1978 - 1986. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Kaiser, J. Bremerich, S. Haller, H. P. Brunner-La Rocca, G. Bongartz, M. Pfisterer, and P. Buser Limited diagnostic yield of non-invasive coronary angiography by 16-slice multi-detector spiral computed tomography in routine patients referred for evaluation of coronary artery disease Eur. Heart J., October 1, 2005; 26(19): 1987 - 1992. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Alkadhi, D. Bettex, S. Wildermuth, B. Baumert, A. Plass, J. Grunenfelder, L. Desbiolles, B. Marincek, and T. Boehm Dynamic Cine Imaging of the Mitral Valve with 16-MDCT: A Feasibility Study Am. J. Roentgenol., September 1, 2005; 185(3): 636 - 646. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Hacker, T. Jakobs, F. Matthiesen, C. Vollmar, K. Nikolaou, C. Becker, A. Knez, T. Pfluger, M. Reiser, K. Hahn, et al. Comparison of Spiral Multidetector CT Angiography and Myocardial Perfusion Imaging in the Noninvasive Detection of Functionally Relevant Coronary Artery Lesions: First Clinical Experiences J. Nucl. Med., August 1, 2005; 46(8): 1294 - 1300. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Leschka, H. Alkadhi, A. Plass, L. Desbiolles, J. Grunenfelder, B. Marincek, and S. Wildermuth Accuracy of MSCT coronary angiography with 64-slice technology: first experience Eur. Heart J., August 1, 2005; 26(15): 1482 - 1487. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. H. Kim, J. Hur, S. J. Kim, H. S. Kim, B. W. Choi, K. O. Choe, Y. W. Yoon, and H. M. Kwon Two-Phase Reconstruction for the Assessment of Left Ventricular Volume and Function Using Retrospective ECG-Gated MDCT: Comparison with Echocardiography Am. J. Roentgenol., August 1, 2005; 185(2): 319 - 325. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Namdar, T. F. Hany, P. Koepfli, P. T. Siegrist, C. Burger, C. A. Wyss, T. F. Luscher, G. K. von Schulthess, and P. A. Kaufmann Integrated PET/CT for the Assessment of Coronary Artery Disease: A Feasibility Study J. Nucl. Med., June 1, 2005; 46(6): 930 - 935. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Koyama, H. Matsuoka, T. Mochizuki, H. Higashino, H. Kawakami, S. Nakata, J. Aono, T. Ito, M. Naka, Y. Ohashi, et al. Assessment of Reperfused Acute Myocardial Infarction with Two-Phase Contrast-enhanced Helical CT: Prediction of Left Ventricular Function and Wall Thickness Radiology, June 1, 2005; 235(3): 804 - 811. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. G. Flohr, S. Schaller, K. Stierstorfer, H. Bruder, B. M. Ohnesorge, and U. J. Schoepf Multi-Detector Row CT Systems and Image-Reconstruction Techniques Radiology, June 1, 2005; 235(3): 756 - 773. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. P. Lawler, H. K. Pannu, and E. K. Fishman MDCT Evaluation of the Coronary Arteries, 2004: How We Do It--Data Acquisition, Postprocessing, Display, and Interpretation Am. J. Roentgenol., May 1, 2005; 184(5): 1402 - 1412. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Marano, M. L. Storto, B. Merlino, N. Maddestra, G. Di Giammarco, and L. Bonomo A Pictorial Review of Coronary Artery Bypass Grafts at Multidetector Row CT Chest, April 1, 2005; 127(4): 1371 - 1377. [Abstract] [Full Text] [PDF] |
||||
![]() |
G J Morgan-Hughes, C A Roobottom, P E Owens, and A J Marshall Highly accurate coronary angiography with submillimetre, 16 slice computed tomography Heart, March 1, 2005; 91(3): 308 - 313. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Schlosser, K. Pagonidis, C. U. Herborn, P. Hunold, K.-U. Waltering, T. C. Lauenstein, and J. Barkhausen Assessment of Left Ventricular Parameters Using 16-MDCT and New Software for Endocardial and Epicardial Border Delineation Am. J. Roentgenol., March 1, 2005; 184(3): 765 - 773. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Nakanishi, Y. Kayashima, R. Inoue, K. Sumii, and Y. Gomyo Pitfalls in 16-Detector Row CT of the Coronary Arteries RadioGraphics, March 1, 2005; 25(2): 425 - 438. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. S. M. Chan, I. Y. F. Chan, K. H. Fung, G. Lee, K. L. Tsui, and T. C. Leung Demonstration of Complex Coronary-Pulmonary Artery Fistula by MDCT and Correlation with Coronary Angiography Am. J. Roentgenol., March 1, 2005; 184(3_supp): S28 - S32. [Full Text] [PDF] |
||||
![]() |
B. L. Gerber, E. Coche, A. Pasquet, E. Ketelslegers, D. Vancraeynest, C. Grandin, B. E. Van Beers, and J.-L. J. Vanoverschelde Coronary Artery Stenosis: Direct Comparison of Four-Section Multi-Detector Row CT and 3D Navigator MR Imaging for Detection--Initial Results Radiology, January 1, 2005; 234(1): 98 - 108. [Abstract] [Full Text] [PDF] |
||||
![]() |
S Schroeder, A Kuettner, T Wojak, J Janzen, M Heuschmid, T Athanasiou, T Beck, C Burgstahler, C Herdeg, C D Claussen, et al. Non-invasive evaluation of atherosclerosis with contrast enhanced 16 slice spiral computed tomography: results of ex vivo investigations Heart, December 1, 2004; 90(12): 1471 - 1475. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Cademartiri, K. Nieman, A. van der Lugt, R. H. Raaijmakers, N. Mollet, P. M. T. Pattynama, P. J. de Feyter, and G. P. Krestin Intravenous Contrast Material Administration at 16-Detector Row Helical CT Coronary Angiography: Test Bolus versus Bolus-tracking Technique Radiology, December 1, 2004; 233(3): 817 - 823. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. Hoffmann, F. Moselewski, R. C. Cury, M. Ferencik, I.-k. Jang, L. J. Diaz, S. Abbara, T. J. Brady, and S. Achenbach Predictive Value of 16-Slice Multidetector Spiral Computed Tomography to Detect Significant Obstructive Coronary Artery Disease in Patients at High Risk for Coronary Artery Disease: Patient- Versus Segment-Based Analysis Circulation, October 26, 2004; 110(17): 2638 - 2643. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Kuettner, T. Trabold, S. Schroeder, A. Feyer, T. Beck, A. Brueckner, M. Heuschmid, C. Burgstahler, A. F. Kopp, and C. D. Claussen Noninvasive detection of coronary lesions using 16-detector multislice spiral computed tomography technology: Initial clinical results J. Am. Coll. Cardiol., September 15, 2004; 44(6): 1230 - 1237. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. F. Viles-Gonzalez, M. Poon, J. Sanz, T. Rius, K. Nikolaou, Z. A. Fayad, V. Fuster, and J. J. Badimon In Vivo 16-Slice, Multidetector-Row Computed Tomography for the Assessment of Experimental Atherosclerosis: Comparison With Magnetic Resonance Imaging and Histopathology Circulation, September 14, 2004; 110(11): 1467 - 1472. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Martuscelli, A. Romagnoli, A. D'Eliseo, C. Razzini, M. Tomassini, M. Sperandio, G. Simonetti, and F. Romeo Accuracy of thin-slice computed tomography in the detection of coronary stenoses Eur. Heart J., June 2, 2004; 25(12): 1043 - 1048. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Desjardins and E. A. Kazerooni ECG-Gated Cardiac CT Am. J. Roentgenol., April 1, 2004; 182(4): 993 - 1010. [Full Text] [PDF] |
||||
![]() |
A. Kuettner, A. F. Kopp, S. Schroeder, T. Rieger, J. Brunn, C. Meisner, M. Heuschmid, T. Trabold, C. Burgstahler, J. Martensen, et al. Diagnostic accuracy of multidetector computed tomography coronary angiography in patients with angiographically proven coronary artery disease J. Am. Coll. Cardiol., March 3, 2004; 43(5): 831 - 839. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Achenbach Detection of coronary stenoses by multidetector computed tomography: it's all about resolution J. Am. Coll. Cardiol., March 3, 2004; 43(5): 840 - 841. [Full Text] [PDF] |
||||
![]() |
K. U. Juergens, M. Grude, D. Maintz, E. M. Fallenberg, T. Wichter, W. Heindel, and R. Fischbach Multi-Detector Row CT of Left Ventricular Function with Dedicated Analysis Software versus MR Imaging: Initial Experience Radiology, February 1, 2004; 230(2): 403 - 410. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. Budoff, S. Achenbach, and A. Duerinckx Clinical utility of computed tomography and magnetic resonance techniques for noninvasive coronary angiography J. Am. Coll. Cardiol., December 3, 2003; 42(11): 1867 - 1878. [Abstract] [Full Text] [PDF] |
||||
![]() |
G J Morgan-Hughes, P E Owens, C A Roobottom, and A J Marshall Three dimensional volume quantification of aortic valve calcification using multislice computed tomography Heart, October 1, 2003; 89(10): 1191 - 1194. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. K. Pannu, T. G. Flohr, F. M. Corl, and E. K. Fishman Current Concepts in Multi-Detector Row CT Evaluation of the Coronary Arteries: Principles, Techniques, and Anatomy RadioGraphics, October 1, 2003; 23(90001): S111 - 125. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. J. Morgan-Hughes, P. E. Owens, A. J. Marshall, and C. A. Roobottom Thoracic Aorta at Multi-Detector Row CT: Motion Artifact with Various Reconstruction Windows Radiology, August 1, 2003; 228(2): 583 - 588. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Hejmadi and D. J. Sahn What is the most effective method of detecting anomalous coronary origin in symptomatic patients? J. Am. Coll. Cardiol., July 2, 2003; 42(1): 155 - 157. [Full Text] [PDF] |
||||
![]() |
A W Leber, A Knez, C Becker, A Becker, C White, C Thilo, M Reiser, R Haberl, and G Steinbeck Non-invasive intravenous coronary angiography using electron beam tomography and multislice computed tomography Heart, June 1, 2003; 89(6): 633 - 639. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. L. Morin, T. C. Gerber, and C. H. McCollough Radiation Dose in Computed Tomography of the Heart Circulation, February 18, 2003; 107(6): 917 - 922. [Full Text] [PDF] |
||||
![]() |
P. Hunold, F. M. Vogt, A. Schmermund, J. F. Debatin, G. Kerkhoff, T. Budde, R. Erbel, K. Ewen, and J. Barkhausen Radiation Exposure during Cardiac CT: Effective Doses at Multi-Detector Row CT and Electron-Beam CT Radiology, January 1, 2003; 226(1): 145 - 152. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Saito, M. Saito, S. Komatu, and K. Ohtomo Real-Time Four-dimensional Imaging of the Heart with Multi-Detector Row CT RadioGraphics, January 1, 2003; 23(1): e8 - e8. [Abstract] [Full Text] |
||||
![]() |
K Nieman, B J Rensing, R-J M van Geuns, J Vos, P M T Pattynama, G P Krestin, P W Serruys, and P J de Feyter Non-invasive coronary angiography with multislice spiral computed tomography: impact of heart rate Heart, December 1, 2002; 88(5): 470 - 474. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. C. Gerber, R. S. Kuzo, R. E. Safford, D. Ropers, K. Pohle, R. Maeffert, M. Regenfus, W. Moshage, W. G. Daniel, S. Achenbach, et al. Computed Tomographic Imaging of Anomalous Coronary Arteries * Response Circulation, October 8, 2002; 106 (15): e67 - e67. [Full Text] [PDF] |
||||
![]() |
T. Giesler, U. Baum, D. Ropers, S. Ulzheimer, E. Wenkel, M. Mennicke, W. Bautz, W. A. Kalender, W. G. Daniel, and S. Achenbach Noninvasive Visualization of Coronary Arteries Using Contrast-Enhanced Multidetector CT: Influence of Heart Rate on Image Quality and Stenosis Detection Am. J. Roentgenol., October 1, 2002; 179(4): 911 - 916. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. K. Willmann, D. Weishaupt, M. Lachat, R. Kobza, J. E. Roos, B. Seifert, T. F. Luscher, B. Marincek, and P. R. Hilfiker Electrocardiographically Gated Multi-Detector Row CT for Assessment of Valvular Morphology and Calcification in Aortic Stenosis Radiology, October 1, 2002; 225(1): 120 - 128. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Tello, G. G. Hartnell, P. Costello, and C. P. Ecker Coronary Artery Bypass Graft Flow: Qualitative Evaluation with Cine Single-Detector Row CT and Comparison with Findings at Angiography Radiology, September 1, 2002; 224(3): 913 - 918. [Abstract] [Full Text] [PDF] |
||||
![]() |
P J de Feyter and K Nieman New coronary imaging techniques: what to expect? Heart, March 1, 2002; 87(3): 195 - 197. [Full Text] [PDF] |
||||
![]() |
S Schroeder, A F Kopp, B Ohnesorge, H Loke-Gie, A Kuettner, A Baumbach, C Herdeg, C D Claussen, and K R Karsch Virtual coronary angioscopy using multislice computed tomography Heart, March 1, 2002; 87(3): 205 - 209. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. C. Gerber, R. S. Kuzo, N. Karstaedt, G. E. Lane, R. L. Morin, P. F Sheedy II, R. E. Safford, J. L. Blackshear, and J. H. Pietan Current Results and New Developments of Coronary Angiography With Use of Contrast-Enhanced Computed Tomography of the Heart Mayo Clin. Proc., January 1, 2002; 77(1): 55 - 71. [Abstract] [PDF] |
||||
![]() |
S. Achenbach, T. Giesler, D. Ropers, S. Ulzheimer, H. Derlien, C. Schulte, E. Wenkel, W. Moshage, W. Bautz, W. G. Daniel, et al. Detection of Coronary Artery Stenoses by Contrast-Enhanced, Retrospectively Electrocardiographically-Gated, Multislice Spiral Computed Tomography Circulation, May 29, 2001; 103(21): 2535 - 2538. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|