(Circulation. 2000;102:2024.)
© 2000 American Heart Association, Inc.
Brief Rapid Communications |
From the Department of Cardiology, Kerckhoff Heart Center, Bad Nauheim (S.B., R.A., M.R., C.W.H.); the Department of Cardiology (R.K., T.M.) and Institute of Mathematics and Computer Science in Medicine (J.B.), University Hospital Eppendorf, Hamburg; and the Department of Cardiology, University Hospital Frankfurt (M.E., D.H.W., W.A.-S., A.M.Z.), Germany.
Correspondence to Christian W. Hamm, MD, Kerckhoff Heart Center, Benekestrasse 2-8, 61231 Bad Nauheim, Germany. E-mail christian.hamm{at}kerckoff.med.uni-giessen.de
| Abstract |
|---|
|
|
|---|
Methods and ResultsIn a multicenter study, we followed up 109 consecutive patients (mean age 66±8 years, 12% female) who received polytetrafluoroethylene (PTFE) membranecovered stents for 125 de novo stenoses in vein grafts 11±5 years after bypass surgery. Stent deployment was successful in all but 1 patient; 1 patient suffered from subacute stent thrombosis. Six-month cardiac mortality was 7% (8 patients), 3 patients (3%) underwent repeat bypass surgery, and 9 patients (8%) required target-lesion PTCA. Repeat angiography revealed vessel occlusions in 9% and in-stent restenosis in 8% of patients by the end of follow-up.
ConclusionsMembrane-covered stents appear to be a safe and efficient treatment strategy associated with a low incidence of restenosis and target-vessel revascularization. Compared with previous studies, the investigated device is not associated with an increase in mortality or late vessel occlusions.
Key Words: grafting stents thrombosis restenosis polytetrafluoroethylene
| Introduction |
|---|
|
|
|---|
PTCA of venous bypass grafts is associated with a high risk of distal thrombus embolization and peri-interventional myocardial infarction.3 Moreover, the long-term results of venous grafts after PTCA are still unsatisfactory, considering a restenosis rate >45%.4 Remarkably, coronary stents, despite their proven benefit in native arteries, did not substantially reduce the incidence of restenosis in vein grafts.4 5
The investigated stent is covered by an expandable
polytetrafluoroethylene (PTFE) membrane
(Figure 1
), which was suggested to
diminish peri-interventional thrombus embolization and long-term
luminal renarrowing. We report here the first major cohort of patients
who received this device for de novo lesions in aortocoronary
vein grafts.
|
| Methods |
|---|
|
|
|---|
Major coronary risk factors were diabetes mellitus (n=30; 28%), hyperlipoproteinemia (n=87; 80%), hypertension (n=72; 66%), and current or recently stopped smoking (n=49; 45%). Seventy percent (n=76) had previous myocardial infarctions; repeat bypass surgery had been performed in 8 patients (7%). The patients presented with stable (n=76, 70%) as well as unstable (n=25, 23%) angina; 8 patients (7%) underwent stent implantation within primary PTCA. The left ventricular ejection fraction was reduced to 35% to 25% in 20 and <25% in 7 patients.
In all, 125 de novo lesions in venous bypass grafts were treated 11±5 years (range 1 to 21 years) after bypass surgery. Lesions were localized in vein grafts to the left anterior descending coronary artery (n=52), the circumflex artery (n=47), and the right coronary artery (n=26). The mean lesion length was 12±7 mm, ranging between 3 and 37 mm. In 21 lesions, thrombotic debris was visualized angiographically before angioplasty; 8 vessels were occluded.
Stent Implantation Procedure
A total of 125 stents (Jomed Coronary Stent Graft,
Sitomed) between 9 and 26 mm (mean 17±8 mm) in length were
implanted. The stents were mounted on semicompliant balloons ranging
between 2.5 and 5.0 mm in diameter (mean 3.6±0.5 mm) and
were deployed with a mean inflation pressure of 16±3 atm (range 10 to
22 atm). Thirty-one percent of the study population received
glycoprotein (GP) IIb/IIIa receptor antagonists
peri-interventionally (abciximab, n=26; tirofiban, n=8). Postprocedural
medication included aspirin 100 mg/d and ticlopidine 500 mg/d.
Ticlopidine was stopped 1 month after the procedure in the first 25
patients; the remaining patients received either ticlopidine 500 mg/d
or clopidogrel 75 mg/d for 10 weeks.
QCA Analysis
Angiograms were obtained in 2 orthogonal projections.
Quantitative analyses were performed offline at a core
laboratory as described elsewhere.7 In-stent
restenosis was defined as a reduction of 50% of the luminal
diameter within 5 mm proximal and distal to the stent. In-stent
restenosis was categorized as focal if its length was 10
mm and diffuse if it was >10 mm.
Definitions
Procedural success was defined as residual stenosis
<50% of the reference diameter. Q-wave myocardial infarction was
defined by new pathological Q waves in
2 contiguous leads and/or
elevation in creatine kinase (CK) or its MB isoenzyme to
3 times the
upper limit of normal. The latter was 80 U/L in all institutions; an
increase of the CK-MB >10% of the CK was considered pathological.
Cardiac death was defined as death that was proven or suspected to be
due to myocardial infarction, malignant arrhythmias, or
congestive heart failure. Major adverse cardiac events were defined as
death, acute myocardial infarction, bypass surgery, and
target-lesion PTCA.
| Results |
|---|
|
|
|---|
Six-Month Follow-Up
Clinical follow-up was obtained for all patients after a median of
5.5 months (range 6 to 47 weeks). Major adverse cardiac events are
displayed in Table 1
. Eight
patients died of definite or suspected cardiac causes within a median
of 98 days (range 8 to 161 days) after implantation of stents with a
mean length of 16±7 mm: 3 patients suffered fatal myocardial
infarctions; 2 of them were located in the target territory, with
angiographically proven in-stent vessel occlusion in 1 patient on day
124 after the procedure. Three additional patients died suddenly, 1
with an ejection fraction of 17% and another after proven patency of
the target vessel. Two patients died of unknown causes. Nonfatal
myocardial infarctions unrelated to the target lesion were observed in
3 patients; 14 patients underwent nontarget-lesion PTCA.
|
Quantitative Coronary Angiography
Follow-up angiograms were available for 64 patients (59%) and 75
lesions (60%). Quantitative angiographic measurements are summarized
in Table 2
; the cumulative distribution
functions of the minimal luminal diameter before treatment, after stent
deployment, and at follow-up are displayed in Figure 2
.
|
|
Five patients (8%) had in-stent restenoses, which were categorized as diffuse in 2 patients. Four of the 5 patients underwent repeat PTCA. In 6 patients (9%), follow-up angiography revealed vessel occlusions; 2 lesions were recanalized by enrollment. Five patients had not received GP IIb/IIIa receptor antagonists peri-interventionally, and 4 of the 6 patients received ticlopidine for 10 weeks after the intervention.
| Discussion |
|---|
|
|
|---|
The in-hospital event rates for death and myocardial infarction have been reported to be between 6% and 17%.4 8 A rate of postinterventional myocardial infarctions <1% and a 6% incidence of minor CK elevations in this trial demonstrate a marked reduction compared with the 16% incidence of procedure-related Q-wave and nonQ-wave myocardial infarctions after vein graft stenting reported recently,9 suggesting a major reduction of thrombotic embolization by the PTFE membrane. To entrap debris at the lesion site, low inflation pressures during predilatation and consecutive high-pressure expansion of the stent (>14 atm) were demonstrated to be most effective.
A 11% rate of target-lesion revascularization during follow-up including 3% of patients undergoing bypass surgery represents a major decline compared with the 17% and 7% incidences in target-lesion revascularization and bypass surgery in the SAVED trial.4 Furthermore, the 17% rate of in-stent restenosis and vessel occlusion reflects a substantial reduction compared with in-stent restenosis in vein grafts reported to exceed 35%.4 8 10
The 6 observed vessel occlusions have to be analyzed carefully: Insufficient stent expansion may have caused 1 subacute stent thrombosis. Second, endothelialization of the PTFE membrane is known to be critical, as described previously in animal models investigating PTFE membranes.11 Vessel occlusion is a common limitation of percutaneous interventions in vein grafts, however: Implantation of conventional stents is associated with graft occlusion >10%.4 Moreover, balloon and laser angioplasty are complicated by long-term vessel occlusions in 13% and 24%, respectively,12 13 whereas recanalization per se bears an increased risk for vessel occlusion, suggesting that late vessel occlusion after interventions in vein grafts is lesion siterelated rather than device-related.
Thus, we question whether the 8% mortality within the first 6 months inevitably reflects late in-stent thrombosis. The study population a priori has to be considered at high risk for cardiac death: The large number of acute coronary syndromes treated, the significantly impaired left ventricular function in nearly one third of the patients, and the treatment of predominantly degenerated vein grafts all are independent predictors of cardiac mortality.14 Peri-interventional ischemia, an additional independent predictor of late mortality,15 might be further reduced by the use of aspiration systems16 as well as by routine administration of GP IIb/IIIa receptor antagonists.
Conclusions
Although the study lacks a randomized design, these data strongly
suggest membrane-covered stents to be a superior treatment strategy for
lesions in aortocoronary venous bypass grafts.
| Acknowledgments |
|---|
Received July 28, 2000; revision received August 25, 2000; accepted August 31, 2000.
| References |
|---|
|
|
|---|
2. Harris WO, Mock MB, Orszulak TA, et al. Use of coronary artery bypass surgical procedure and coronary angioplasty in treatment of coronary artery disease: changes during a 10-year period at the Mayo Clinic Rochester. Mayo Clin Proc. 1996;71:927935.[Abstract]
3.
Lefkovits J, Holmes DR, Califf RM, et al. Predictors
and sequelae of distal embolization during saphenous vein graft
intervention from the CAVEAT-II Trial. Circulation. 1995;92:734740.
4.
Savage MP, Douglas JS, Fischman DL, et al. Stent
placement compared with balloon angioplasty for obstructed
coronary bypass grafts. N Engl J Med. 1997;337:740747.
5.
Kornowski R, Mehran R, Satler LF, et al. Procedural
results and late clinical outcomes following multivessel
coronary stenting. J Am Coll Cardiol. 1999;33:420426.
6. Elsner M, Auch-Schwelk W, Britten M, et al. Coronary stent grafts covered by a polytetrafluoroethylene membrane. Am J Cardiol. 1999;84:335338.[Medline] [Order article via Infotrieve]
7. Savage MP, Fischman DL, Schatz RA, et al. Long-term angiographic and clinical outcome after implantation of a balloon-expandable stent in the native coronary circulation. J Am Coll Cardiol. 1994;24:12071212.[Abstract]
8. Wong SC, Popma JJ, Kent KM, et al. Clinical experience with stent implantation in the treatment of saphenous vein graft lesions. J Intervent Cardiol. 1994;7:565573.[Medline] [Order article via Infotrieve]
9.
Bhargava B, Kornowski R, Mehran R, et al. Procedural
results and intermediate clinical outcomes after multiple saphenous
vein graft stenting. J Am Coll Cardiol. 2000;35:389397.
10. Waksman R, Weintraub WS, Ghazzal Z, et al. Short- and long-term outcome of narrowed saphenous vein bypass graft: a comparison of Palmaz-Schatz stent, directional coronary atherectomy and balloon angioplasty. Am Heart J. 1997;134:274281.[Medline] [Order article via Infotrieve]
11. Dolmatch BL, Tio FO, Li XD, et al. Patency and tissue response related to two types of polytetrafluoroethylene-covered stents in the dog. J Vasc Intervent Radiol. 1996;7:641649.[Medline] [Order article via Infotrieve]
12.
Van Belle E, Abolmaali K, Bauters C, et al.
Restenosis, late vessel occlusion and left
ventricular function six months after balloon angioplasty
in diabetic patients. J Am Coll Cardiol. 1999;34:476485.
13.
Strauss BH, Natarajan MK, Batchelor WB, et al. Early
and late quantitative angiographic results of vein graft lesions
treated by excimer laser with adjunctive balloon angioplasty.
Circulation. 1995;92:348356.
14. Holmes DR Jr, Kip KE, Kelsey SF, et al. Cause of death analysis in the NHLBI PTCA Registry: results and considerations for evaluating long-term survival after coronary interventions. J Am Coll Cardiol. 1997;30:881887.[Abstract]
15.
Kong TQ, Davidson CJ, Meyers SN, et al. Prognostic
implication of creatine kinase elevation following elective
coronary artery interventions. JAMA. 1997;277:461466.
16.
Carlino M, De Gregorio J, Di Mario C, et al. Prevention
of distal embolization during saphenous vein graft lesion angioplasty:
experience with a new temporary occlusion and aspiration system.
Circulation. 1999;99:32213223.
This article has been cited by other articles:
![]() |
V. Schachinger, C. W. Hamm, T. Munzel, M. Haude, S. Baldus, E. Grube, T. Bonzel, T. Konorza, R. Koster, R. Arnold, et al. A randomized trial of polytetrafluoroethylene-membrane-covered stents compared with conventional stents in aortocoronary saphenous vein grafts J. Am. Coll. Cardiol., October 15, 2003; 42(8): 1360 - 1369. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Stankovic, A. Colombo, P. Presbitero, F. van den Branden, L. Inglese, C. Cernigliaro, L. Niccoli, A. L. Bartorelli, P. Rubartelli, N. Reifart, et al. Randomized Evaluation of Polytetrafluoroethylene-Covered Stent in Saphenous Vein Grafts: The Randomized Evaluation of polytetrafluoroethylene COVERed stent in Saphenous vein grafts (RECOVERS) Trial Circulation, July 8, 2003; 108(1): 37 - 42. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Colombo, G. Stankovic, and J. W. Moses Selection of coronary stents J. Am. Coll. Cardiol., September 18, 2002; 40(6): 1021 - 1033. [Abstract] [Full Text] [PDF] |
||||
![]() |
Covered Stents for Vein-Graft Lesions Journal Watch Cardiology, December 15, 2000; 2000(1215): 5 - 5. [Full Text] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2000 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |