(Circulation. 2000;102:1407.)
© 2000 American Heart Association, Inc.
Clinical Investigation and Reports |
From Riverside-Methodist Hospital, Columbus, Ohio (E.G.D.); Academic Hospital Maastricht, the Netherlands (C.T.); Virginia Mason Medical Center, Seattle, Wash (C.F.); Iowa Heart Institute, Des Moines, Iowa (R.H.); Rhode Island Hospital, Providence, RI (R.L.); and InControl/Guidant, Inc, Redmond, Wash/St Paul, Minn (K.D., G.M.A.).
Correspondence to Emile Daoud, MD, MidOhio Cardiology Research, 3545 Olentangy River Rd, Room 325, Columbus, OH 43214. E-mail egd{at}mocc.cc
| Abstract |
|---|
|
|
|---|
Methods and ResultsAn atrial defibrillator was implanted in 105 patients (75 men; mean age, 59±12 years) with recurrent, symptomatic, drug-refractory atrial fibrillation. After successful 3-month testing, patients could transition to ambulatory delivery of shock therapy. Patients completed questionnaires regarding shock therapy discomfort and therapy satisfaction using a 10-point visual-analog scale (1 represented "not at all," 10 represented "extremely") after each treated episode of atrial fibrillation. During a mean follow-up of 11.7 months, 48 of 105 patients satisfied criteria for transition and received therapy for 275 episodes of atrial fibrillation. Overall shock therapy efficacy was 90% with 1.6±1.2 shocks delivered per episode (median, 1). Patients rated shock discomfort as 5.2±2.4 for successful therapy and 4.2±2.2 for unsuccessful therapy (P>0.05). The satisfaction score was higher for successful versus unsuccessful therapy (3.4±3.3 versus 8.7±1.3, P<0.05). There was no ventricular proarrhythmia observed throughout the course of this study.
ConclusionsAmbulatory use of an implantable atrial defibrillator can safely and successfully convert most episodes of atrial fibrillation, often requiring only a single shock. Successful therapy is associated with high satisfaction and only moderate discomfort.
Key Words: defibrillation shock cardioversion outpatient therapy
| Introduction |
|---|
|
|
|---|
| Methods |
|---|
|
|
|---|
|
Atrial Defibrillator
An atrial defibrillator (Metrix models 3000 and
3020, InControl/Guidant, Inc) was implanted in conjunction with 2
defibrillation leads and a bipolar pacemaker lead. The leads were
introduced through the subclavian vein, and the defibrillation leads
were positioned in the right atrium (InControl/Guidant model 7205) and
distal coronary sinus (InControl/Guidant model 7109). The
pacemaker lead was placed in the right ventricle (Figure 1
). The earlier atrial defibrillator
generator (model 3000) had a maximum output of 300 V (
3 J). The
next-generation device (model 3020) also had a maximum output of 300 V,
but with a larger capacitor, the maximum delivered energy was
6 J.
Each model delivers a biphasic waveform shock. The defibrillator could
be programmed to provide
8 shocks, with individual voltages
programmable from 20 to 300 V in 10-V increments. The device uses a
dual algorithm to detect atrial fibrillation, and defibrillation shocks
are synchronized to the R wave.6 The device also manages
postshock bradycardia with VVI pacing.
|
The atrial defibrillator can be programmed to 1 of 5 operating modalities: automatic, patient-activated, monitor, pacing-only, or off mode. In the automatic mode, the device is programmed, from every minute to 120 minutes, to monitor the cardiac rhythm for the presence of atrial fibrillation. Therapy will be delivered if atrial fibrillation is present unless the patient inhibits device therapy with a conventional pacemaker magnet. The patient-activated mode allows the patient to activate the device with a pacemaker magnet. If the device detects atrial fibrillation, therapy will be delivered unless inhibited by a magnet. The monitor mode functions like the automatic mode with respect to atrial fibrillation detection; however, the device does not deliver defibrillation therapy. For the modes in which therapy is delivered, a warning shock (10 to 20 V) can be programmed to discharge before delivery of the therapeutic shock. When programmed to the pacing-only mode, the device paces in a VVI mode.
The device stores 3 electrograms for each episode of atrial
fibrillation. The atrial fibrillation detection electrogram displays 3
seconds of atrial and ventricular signals. The second
electrogram displays the ventricular electrogram
recorded for 3 seconds before the delivery of shock therapy. In
this electrogram, a mark is placed on the R wave to which shock therapy
was synchronized. The third electrogram displays atrial and
ventricular signals for 4 seconds after the shock (Figure 2
). These electrograms were reviewed to
assess the accuracy of device-based atrial fibrillation
detection,6 to confirm synchronization, to evaluate the
outcome of shock therapy, and to detect ventricular
proarrhythmia. Ventricular proarrhythmia
was defined as
3 consecutive ventricular beats at a cycle
length <600 ms.
|
Postimplantation Follow-Up
For at least the first 3 months after implant, the atrial
defibrillator was programmed to the monitor mode, and patients were
instructed to return to the hospital or clinic for outpatient treatment
of each symptomatic episode of atrial fibrillation.
Defibrillation was then performed under physician observation during
ECG recording. These therapies are reported as outpatient
defibrillator therapy. After
3 months of follow-up, the following
criteria were required to transition from outpatient defibrillator
therapy to ambulatory device therapy: an atrial defibrillation energy
requirement
260 V (
2.6 J for the 3000 model and 5.4 J for the 3020
model); no false-positive detection of atrial fibrillation; correct
R-wave synchronization; patient education regarding device
activation/deactivation; and
1 episode of atrial fibrillation treated
with outpatient defibrillator therapy under physician observation
fulfilling each of the following requirements: successful termination
with
300 V, therapy delivered during similar conditions as for
ambulatory use, patient acceptance of therapy, and sinus rhythm
maintained for 1 week after therapy,
After transition to ambulatory therapy, the device was programmed to the patient-activated or the automatic mode. Mode selection was left to the discretion of the physician in conference with the patient. Anxiolytics and/or narcotics were rarely prescribed.7
The patient was instructed to complete an event form and to return for clinical follow-up after each treated episode of atrial fibrillation. The event form recorded the patient-perceived onset of symptoms and assessed the patients perception of atrial fibrillation and device therapy. A visual-analog scale (1 represented "not at all," 10 represented "extremely") was used to score the answers to the following questions: "How severe were the symptoms from this episode of atrial fibrillation?" "How satisfied were you with the therapy?" "How uncomfortable was the therapy?" The fourth question required a yes/no response to, "Did device therapy relieve symptoms?"
Using the stored electrograms and the patients report of symptoms, the investigator determined whether an individual episode of atrial fibrillation was terminated. Ambulatory efficacy was defined as the percentage of episodes terminated by the device when treated away from the hospital/clinic. Therapy efficacy was defined as the overall percentage of episodes converted by the atrial defibrillator to stable sinus rhythm either with treatment delivered while the patient was ambulatory or when the patient was treated in the hospital/clinic.
Statistical Analysis
Continuous variables are expressed as mean±SD and were
compared by use of an ANOVA. Population safety and efficacy
calculations were made by fitting the distributions of these
variables in an individual patient (assuming a exponential
distribution for risk of proarrhythmia and a binomial
probability distribution for efficacy) and then determining the mean
across the population of patients, assuming a ß distribution.
Similarly, the risk of proarrhythmia was calculated for the
patient population. For all analyses, P<0.05
conferred statistical significance.
| Results |
|---|
|
|
|---|
|
|
Of the 105 patients enrolled, the device was explanted in 15 patients. Three devices (3%) were explanted because of lead complications, 1 (1%) because of infection, and 11 (10%) because of inadequate control of atrial fibrillation. In the 105 patients, 18 of the 270 leads (7%) implanted required repositioning (9 right atrial leads and 9 right ventricle leads). The coronary sinus lead did not dislodge in any patient. The mean follow-up was 11.7±6.8 months.
Outpatient Defibrillator Therapy Before Transition to
Ambulatory Therapy
Seventy-five patients had 412 episodes of atrial fibrillation for
which they sought therapy (Figure 3
). Of these 412 episodes, 388
were treated with 897 shocks delivered from the atrial defibrillator.
The atrial defibrillator restored sinus rhythm in 368 of the 388
episodes (95%); however, atrial fibrillation recurred within minutes
after successful cardioversion for 95 episodes (26%).5
For 54 of the 95 episodes in which early recurrence of atrial
fibrillation occurred during the treatment of an episode, sinus rhythm
was eventually restored with additional shocks from the atrial
defibrillator with or without the administration of antiarrhythmic
drugs. Taking into account the remaining 41 of the 95 episodes where
early recurrences were observed and could not be successfully
treated, the therapy efficacy was 85%.
|
Ambulatory Defibrillator Therapy
At the time of the most recent follow-up, 48 patients fulfilled
all criteria and agreed to receive ambulatory therapy (Figure 3
). The mean follow-up after transition to ambulatory therapy
was 8.4±3.7 months. For 6 patients, the device was programmed to the
automatic mode, and for 42 patients, the device was programmed to the
patient-activated mode. Table 4
outlines the programmed number of shocks and their intensity for 43 of
48 patients in whom these data were available. The mean number of
shocks programmed was 2.1±1.0, with a median of 2. No patients had >4
shocks programmed. The first shock voltage was programmed to the
maximum output of 300 V for 28 of the 43 patients (65%). Second,
third, and fourth shocks were always programmed to maximum output in
all patients. A warning shock was programmed for 18 patients (42%). In
most patients, this low-voltage shock was delivered within 5 minutes
before delivery of the therapeutic shock.
|
Of the 48 patients, 37 patients had 301 episodes of atrial fibrillation for which they desired therapy. Complete data to assess therapy outcome were available for 296 of 301 episodes (98%). The average number of episodes per patient was 8±10 (range, 1 to 43), with a median of 4 episodes per patient. The adjusted monthly episode rate for this patient population was 1.4±1.0 episodes per patient-month (range, 0.1 to 4.6), with a median of 1.2.
A total of 365 shocks were delivered for ambulatory therapy, and all shocks were appropriate. Atrial fibrillation was confirmed by stored electrograms. No shocks were delivered during sinus rhythm. The mean number of shocks required to treat each episode of atrial fibrillation was 1.6±1.2 (range, 1 to 8), with a median of 1 shock per episode. In 276 of the 296 episodes of atrial fibrillation in which the number of shocks delivered was known, 191 episodes (69%) were successfully treated with 1 shock, and only 22 episodes (8%) required >3 shocks.
The outcome of the 296 episodes is summarized in Figure 3
. For
265 of 296 episodes of atrial fibrillation (90%), the patient
attempted to treat the episode. The atrial defibrillator successfully
converted 199 of 265 episodes (75%), and therapy failed to
successfully convert 52 episodes (25%). The remaining 14 episodes did
not result in defibrillator therapy delivery because normal device
function precluded delivery of therapy as a result of a high
ventricular rate,6 7 the patient did not
correctly activate the device with the magnet, or the patient
deactivated the device after initial activation. Of the 66
episodes of atrial fibrillation that were not successfully treated or
for which no shock was delivered, 34 (51%) converted spontaneously
after failed device therapy (n=22 of 52) or after therapy was not
delivered (n=12 of 14).
For the remaining 31 of 296 episodes of atrial fibrillation, patients chose to return to the clinic or hospital for therapy for 27 episodes, and atrial fibrillation converted spontaneously for 4 episodes. Atrial fibrillation also converted spontaneously in 4 of the 27 episodes for which the patient chose to return to the clinic/hospital for device therapy.
The 55 sustained episodes of atrial fibrillation were treated under
physician observation (bottom of Figure 3
). Of these 55
episodes, 49 (89%) were successfully treated with the atrial
defibrillator, 5 episodes failed to be terminated and later reverted
spontaneously to sinus rhythm, and 1 episode terminated before device
activation.
Considering episodes of atrial fibrillation that were converted with the defibrillator in the ambulatory setting, the efficacy was 75% (199 of 265), and the therapy efficacy (including device therapy delivered in a clinic/hospital setting) was 90% (248 of 275). Therapy efficacy was 100% in 70% of patients. Among the 37 patients experiencing only 1 episode of atrial fibrillation, the atrial defibrillator failed to cardiovert the episode in only 3 patients (8%).
Safety of Shock Therapy
During this study, there were 5523 shocks delivered from the
atrial defibrillator; 365 shocks were delivered during ambulatory
therapy. An additional 897 shocks were delivered to treat episodes of
atrial fibrillation before the patients transition to ambulatory
therapy. The remaining 4261 shocks were delivered for atrial
fibrillation induction and defibrillation testing. All shocks were
delivered with appropriate synchronization, and there were no episodes
of ventricular proarrhythmia.6 8
Considering the number of shocks per patient and the number of patients
enrolled, the 95th percentile upper CI of risk of proarrhythmia
was 0.056% per shock. Assuming the median monthly episode rate and
shocks per episode as reported in this study, the maximum annual risk
of proarrhythmia was 0.81%.
Perception of Ambulatory Therapy
The impact of event scores was available from 143 treated episodes
(25 patients). The mean number of episodes per patient was 5.7±6.1
(range, 1 to 32), with a median of 4.
The average symptom severity score was not significantly different between episodes that were successfully treated and those that were unsuccessfully treated (6.3±2.0 versus 5.2±1.8, P>0.05). The overall severity index was 6.1±1.9. Similarly, there was no significant difference between successfully and unsuccessfully treated episodes with respect to discomfort (5.2±2.5 [median, 5.0] versus 4.2±2.2 [median, 4.0], P>0.05). Patients did have significantly higher satisfaction scores associated with successful compared with unsuccessful therapies (8.7±1.3 [median, 9] versus 3.4±3.3 [median, 2], P<0.05). Patients reported relief of symptoms after treatment of 83% of episodes.
| Discussion |
|---|
|
|
|---|
70% of the atrial fibrillation episodes were
successfully treated with a single shock. There was no
ventricular proarrhythmia. These findings are
considerably better than after 1 year of pharmacological therapy, in
which sinus rhythm is maintained in only 50% to 70% of
patients9 10 and the risk of proarrhythmia is
2% to 7%.11 12 In addition, successful atrial
defibrillation was associated with moderate discomfort and high patient
satisfaction scores. These findings suggest that ambulatory use of an
atrial defibrillator provides safe, successful, and tolerable
conversion of atrial fibrillation.
Absence of Proarrhythmia
With correct R-wave synchronization and timing to ensure that
shock therapy is not delivered during the ventricular
relative refractory period, no ventricular
proarrhythmia was observed with delivery of >5000 shocks.
Although the mean left ventricular ejection fraction was
normal, structural heart disease was present in most patients. The
results of this study therefore confirm the findings of other
studies5 6 8 that the risk of inducing a
ventricular tachyarrhythmia as a result of
atrial defibrillation is negligible and extends these findings to
ambulatory therapy.
Patient Tolerance of Shock Therapy
In previous studies, patients undergoing internal defibrillation
tolerated fewer high-intensity shocks (3 to 6 J) better than multiple
low-intensity shocks (1 to 3 J).5 13 14 In this study,
most atrial fibrillation episodes could be treated with a single shock,
and most first shocks were programmed to maximum output. This high rate
of single-shock success likely enhanced patient tolerance of
shock-induced discomfort.7 There are other factors that
may influence patient acceptance of shock therapy. Clinical,
social, and psychological factors, such as severity of symptoms
associated with atrial fibrillation, previous uncomfortable or painful
experiences, surrounding stimuli, and convenience of ambulatory and
timely therapy, may also affect patient tolerance.15 In
the present study, a questionnaire evaluating the impact of
ambulatory therapy showed high patient satisfaction with only moderate
discomfort.
Early Recurrence of Atrial Fibrillation
In an earlier study5 in which the atrial
defibrillator was tested at the time of implantation, an important
limitation was the 27% rate of early recurrence of atrial
fibrillation. The present study demonstrates that early
recurrence of atrial fibrillation is also an important clinical
problem when device therapy is delivered in the outpatient or
ambulatory setting. In the ambulatory setting, even though stored
electrograms occasionally documented early recurrences of
atrial fibrillation (Figure 2
), its exact incidence could not be
determined once the patient was treated away from the physician. From
the rate of early recurrences before transition, it is likely
that early recurrence of atrial fibrillation was mostly
responsible for the 25% failure rate of outpatient device
therapies.
Previous Studies
Studies of transvenous atrial defibrillation have demonstrated
that most patients with paroxysmal, persistent, and permanent atrial
fibrillation can be reliably cardioverted.2 3 4 13 16 17 18
Recently, the efficacy of the atrial defibrillator was demonstrated in
patients meeting strict implantation criteria.5 In this
initial study, atrial defibrillation was performed under physician
observation, and the efficacy was similar to the results in this study.
Therefore, the present study demonstrates that ambulatory therapy
with the atrial defibrillator is efficacious and is not adversely
affected by the absence of physician intervention.
Impact of Ambulatory Therapy
It is important to speculate on the impact of out-of-hospital
defibrillation on the cost of care and quality of life. Patients with
frequent episodes of atrial fibrillation, similar to the patients
included in this study who had frequent contact with the healthcare
system, often use the healthcare system at a significant
cost.16 17 In the present study, the efficacy of
ambulatory therapy was high, which would likely result in fewer
emergency room visits and hospitalizations. Even for episodes in which
ambulatory therapy was unsuccessful, the patient did not require
hospitalization to restore sinus rhythm. Furthermore, as demonstrated
in this study, cardioversion of atrial fibrillation relieved symptoms
and thus may also improve the patients quality of
life.18 Additional studies are needed to prospectively
evaluate the cost-effectiveness of ambulatory atrial defibrillation
with an implanted device.
Study Limitations
Antiarrhythmic therapy was not controlled in this study;
therefore, no evaluation of its effect on atrial fibrillation could be
made. However, despite antiarrhythmic therapy, patients continued to
experience symptomatic atrial fibrillation. A second
limitation is that these results are applicable only to patients with
frequently recurring, drug-refractory, symptomatic atrial
fibrillation. These results cannot be extended to patients with
depressed left ventricular ejection fraction or other
conditions that may be associated with a higher risk of
ventricular arrhythmias. A third limitation is that
fewer than the total patients implanted were transitioned to receive
ambulatory therapy. The primary reason for not transitioning to
ambulatory therapy was that strict criteria for transition were defined
in this preliminary trial to guard against life-threatening
ventricular proarrhythmia. Repeated confirmations
of successful detection and conversion of atrial fibrillation, without
early recurrence, were necessary before ambulatory therapy was
begun. However, the total number of delivered therapies was quite high
and confirms the safety of ambulatory use of an atrial defibrillator. A
final limitation is that patients were instructed to complete an
impact-of-event questionnaire immediately after delivery of ambulatory
therapy. A drawback of this type of survey is that the data collection
is dependent on nonmedical personnel to complete and return the
questionnaire. Compliance therefore was not 100%.
Clinical Implications
An implantable atrial defibrillator needs to satisfy several
requirements to achieve therapeutic success in the ambulatory treatment
of atrial fibrillation. The defibrillator must accurately detect atrial
fibrillation, properly synchronize shocks, and successfully cardiovert
atrial fibrillation in a tolerable manner, preferably one controlled by
the patient. This study confirms that the Metrix implanted atrial
defibrillator is effective at detecting and terminating atrial
fibrillation without induction of ventricular
proarrhythmia. Although patients perceived shock therapy as
moderately uncomfortable, patient satisfaction with the therapy was
high. Whether an atrial defibrillator is cost-effective and improves
quality of life remains to be determined.
| Footnotes |
|---|
Received November 5, 1999; revision received April 13, 2000; accepted April 19, 2000.
| References |
|---|
|
|
|---|
2. Murgatroyd FD, Slade AKB, Sopher SM, et al. Efficacy and tolerability of transvenous low energy cardioversion of paroxysmal atrial fibrillation in humans. J Am Coll Cardiol. 1995;25:13471353.[Abstract]
3. Schmitt C, Alt E, Plewan A, et al. Low energy intracardiac cardioversion after failed conventional external cardioversion of atrial fibrillation. J Am Coll Cardiol. 1996;28:994999.[Abstract]
4. Lévy S, Ricard P, Lau CP, et al. Multicenter low energy transvenous atrial defibrillation (XAD) trial results in different subsets of atrial fibrillation. J Am Coll Cardiol. 1997;29:750755.[Abstract]
5.
Wellens HJJ, Lau CP, Lüderitz B, et al, for the
Metrix Investigators. The atrial defibrillator: an implantable device
for the treatment of atrial fibrillation. Circulation. 1998;98:16511656.
6.
Tse HF, Lau CP, Sra JS, et al, for the Metrix
Investigators. Atrial fibrillation detection and R-wave synchronization
by Metrix implantable atrial defibrillator: implications for long-term
efficacy and safety. Circulation. 1999;99:14461451.
7.
Timmermans C, Nabar A, Rodriguez LM, et al. Use of
sedation during cardioversion with the implantable atrial
defibrillator. Circulation. 1999;100:14991501.
8.
Ayers GM, Alferness CA, Wagner DO, et al. Effect of
ventricular cycle length and shock strength on
ventricular vulnerability in a sheep model of transvenous
atrial defibrillation. Circulation. 1994;89:413422.
9. Nattel S, Hadjis T, Talajic M. The treatment of atrial fibrillation: an evaluation of drug therapy, electrical modalities and therapeutic considerations. Drugs. 1994;48:345371.[Medline] [Order article via Infotrieve]
10. Ganz LI, Antman EM. Antiarrhythmic drug therapy in the management of atrial fibrillation. J Cardiovasc Electrophysiol. 1997;8:11751189.[Medline] [Order article via Infotrieve]
11. Prystowsky EN. Proarrhythmia during drug treatment of supraventricular tachycardia: paradoxical risk of sinus rhythm for sudden death. Am J Cardiol. 1996;78:3541.[Medline] [Order article via Infotrieve]
12.
Middlekauff HR, Stevenson WG, Gornbein JA.
Antiarrhythmic prophylaxis vs warfarin anticoagulation to prevent
thromboembolic events among patients with atrial fibrillation: a
decision analysis. Arch Intern Med. 1995;155:913920.
13. Lok NS, Lau CP, Tse HF, et al. Clinical shock tolerability and effect of different right atrial electrode locations on efficacy of low energy human transvenous atrial defibrillation using an implantable lead system. J Am Coll Cardiol. 1997;30:13241330.[Abstract]
14. Ayers GM. How can atrial defibrillation be made more tolerable. In: Camm AJ, Murgatroyd F, eds. Non-Pharmacological Management of Atrial Fibrillation. Armonk, NY: Futura Publishing Co, Inc; 1997:475487.
15.
Timmermans C, Rodriguez LM, Ayers GM, et al. Effect of
butorphanol tartrate on shock-related discomfort during internal atrial
defibrillation. Circulation. 1999;99:18371842.
16. Maglio C, Sra J, Paquette M, et al. Measuring quality of life and symptom severity in patients with atrial fibrillation. Pacing Clin Electrophysiol. 1998;21:839. Abstract.
17. Bialy D, Lehrman MH, Schumaker DN, et al. Hospitalization for arrhythmias in the US: importance of atrial fibrillation. J Am Coll Cardiol. 1992;19:41A. Abstract.
18. Jung W, Lüderitz B. Quality of life in patients with atrial fibrillation. J Cardiovasc Electrophysiol. 1998;9:S177S186.[Medline] [Order article via Infotrieve]
This article has been cited by other articles:
![]() |
E. A. Stephenson and C. I. Berul Electrophysiological Interventions for Inherited Arrhythmia Syndromes Circulation, August 28, 2007; 116(9): 1062 - 1080. [Full Text] [PDF] |
||||
![]() |
Z. Goldberger and R. Lampert Implantable Cardioverter-Defibrillators: Expanding Indications and Technologies JAMA, February 15, 2006; 295(7): 809 - 818. [Abstract] [Full Text] [PDF] |
||||
![]() |
J.C. Geller, S. Reek, C. Timmermans, T. Kayser, H.-F. Tse, C. Wolpert, W. Jung, A.J. Camm, C.-P. Lau, H. J.J. Wellens, et al. Treatment of atrial fibrillation with an implantable atrial defibrillator -- long term results Eur. Heart J., December 1, 2003; 24(23): 2083 - 2089. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Manoharan, N. Evans, B. Kidwai, D. Allen, J. Anderson, and J. Adgey Novel Passive Implantable Atrial Defibrillator Using Transcutaneous Radiofrequency Energy Transmission Successfully Cardioverts Atrial Fibrillation Circulation, September 16, 2003; 108(11): 1382 - 1388. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Khairy and S. Nattel New insights into the mechanisms and management of atrial fibrillation Can. Med. Assoc. J., October 29, 2002; 167(9): 1012 - 1020. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Schwartzman, S. K. Musley, C. Swerdlow, R. H. Hoyt, and E. N. Warman early recurrence of atrial fibrillation after ambulatory shock conversion J. Am. Coll. Cardiol., July 3, 2002; 40(1): 93 - 99. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. M. Cooper, M. S. Katcher, and M. V. Orlov Implantable Devices for the Treatment of Atrial Fibrillation N. Engl. J. Med., June 27, 2002; 346(26): 2062 - 2068. [Full Text] [PDF] |
||||
![]() |
X. Zheng, M. E. Benser, G. P. Walcott, and R. E. Ideker Right Atrial Septal Electrode for Reducing the Atrial Defibrillation Threshold Circulation, August 28, 2001; 104(9): 1066 - 1070. [Abstract] [Full Text] [PDF] |
||||
![]() |
Ambulatory Use of Atrial Defibrillators -- Is It for Real? Journal Watch Cardiology, December 1, 2000; 2000(1201): 6 - 6. [Full Text] |
||||
![]() |
Early Experience with Atrial Defibrillator Favorable -- for Some Journal Watch (General), October 3, 2000; 2000(1003): 2 - 2. [Full Text] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2000 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |