Donate Help Contact The AHA Sign In Home
American Heart Association
Circulation
Search: search_blue_button Advanced Search
Circulation. 1999;100:756-760

This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow Request Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Carrozza, J. P.
Right arrow Articles by Baim, D. S.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Carrozza, J. P., Jr
Right arrow Articles by Baim, D. S.
Related Collections
Right arrow Catheter-based coronary interventions: stents
Right arrow Coronary imaging: angiography/ultrasound/Doppler/CC

(Circulation. 1999;100:756-760.)
© 1999 American Heart Association, Inc.


Basic Science Reports

In Vivo Assessment of Stent Expansion and Recoil in Normal Porcine Coronary Arteries

Differential Outcome by Stent Design

Joseph P. Carrozza, Jr, MD; Susanne E. Hosley, BS; David J. Cohen MD, MSc; Donald S. Baim, MD

From Beth Israel–Deaconess Medical Center, Boston, Mass (J.P.C., D.J.C., D.S.B.), and Boston Scientific Corp, Minneapolis, Minn (S.E.H.).

Correspondence to Donald S. Baim, MD, Cardiovascular Division, Beth Israel–Deaconess Medical Center, Boston, MA 02215. E-mail dbaim{at}bidmc.harvard.edu


*    Abstract
up arrowTop
*Abstract
down arrowIntroduction
down arrowMethods
down arrowResults
down arrowDiscussion
down arrowReferences
 
Background—Despite the routine use of high pressure, coronary stents generally fail to achieve a cross-sectional area (CSA) >60% to 80% of the nominal CSA of their dilating balloon. The extent to which incomplete balloon expansion and postdeflation stent recoil contribute to this failure has not been fully evaluated.

Methods and Results—Thirty-two stents (8 Gianturco-Roubin II, 8 Palmaz-Schatz, 8 MultiLink, and 8 NIR) were deployed in nondiseased coronary arteries of 8 Yorkshire pigs. All stents were then expanded according to 1 of 3 balloon strategies: appropriately sized compliant balloons, oversized compliant balloons, or oversized noncompliant balloons. Continuous ultrasound imaging was performed during stepwise balloon inflation and deflation, with an 0.018-in imaging core positioned within the guidewire lumen of the balloon. In these normal arteries, balloon underexpansion relative to the nominal size was not observed. After balloon deflation, however, all stents showed significant recoil from their maximum inflated CSA. Recoil was significantly greater for the coil Gianturco-Roubin II stent (30% CSA) than for the 3 slotted-tube stent designs (15% to 17% CSA).

Conclusions—In normal, compliant coronary arteries, stent recoil is the predominant mechanism by which stents fail to achieve the nominal CSA of their dilating balloon. The magnitude of this in vivo stent recoil is significantly greater than reported from bench testing and varies with stent design (coil versus slotted tube). Postdilatation strategies that result in controlled stent overexpansion are needed to overcome this recoil phenomenon and result in a final stent CSA that approximates the reference artery CSA.


Key Words: stents • arteries • balloon • ultrasonics


*    Introduction
up arrowTop
up arrowAbstract
*Introduction
down arrowMethods
down arrowResults
down arrowDiscussion
down arrowReferences
 
Multiple registries and randomized trials have demonstrated that favorable short-term and intermediate-term angiographic and clinical outcomes after stenting are contingent on optimal stent expansion.1 2 3 4 5 6 Yet seminal intravascular ultrasound (IVUS) studies in the early 1990s demonstrated that most stents remain underexpanded despite an excellent angiographic appearance. High-pressure stent postdilatation improves strut apposition, luminal cross-sectional area (CSA), and concentricity,7 so that routine adjunctive high-pressure postdilatation has become a standard part of stent implantation.

Although high-pressure dilatation increases the lumen CSA within the stent, final stent CSA still rarely exceeds 60% to 70% of the nominal balloon CSA.7 8 This may result from either failure of the balloon to expand fully to its nominal area or recoil of the stent after balloon deflation. The ability to distinguish between these mechanisms (or a combination thereof) has been limited by difficulties in accurately measuring in vivo balloon CSA as a function of pressure and quantifying the amount of stent area recoil that occurs immediately after balloon deflation.

To overcome these limitations, we used an 0.018-in high-frequency ultrasound imaging core positioned within the guidewire lumen of various balloons (both compliant and noncompliant) during stent dilatation. This allowed ultrasonic measurement of balloon CSA as a function of pressure during inflation and immediate quantification of any loss of stent CSA during balloon deflation for different stent designs.


*    Methods
up arrowTop
up arrowAbstract
up arrowIntroduction
*Methods
down arrowResults
down arrowDiscussion
down arrowReferences
 
Animal Preparation
Eight Yorkshire pigs (40 to 60 kg) of either sex were fasted on the day of the procedure. After sedation with ketamine 15 mg/kg IM and endotracheal intubation, general anesthesia was maintained with halothane 2%, with oxygen. An 8F sheath (Cordis Corp) was inserted via cutdown into the right femoral artery. Heparin 5000 U was given intravenously, and additional heparin was administered to maintain an activated clotting time >300 seconds. Arterial blood pressure and the surface ECG were monitored throughout the procedure. An 8F hockey-stick guiding catheter (Boston Scientific Corp) was engaged in the ostium of either the right or left coronary artery. The study was approved by the animal care committee at the Beth Israel–Deaconess Medical Center.

Stent Procedure
An 0.014-in angioplasty guidewire (Choice PT, Boston Scientific Corp) was advanced into either the distal right or circumflex coronary artery. After the administration of 200 µg nitroglycerin IC, standard IVUS was performed with a 2.9F catheter (UltraCross, CVIS Corp) to assess lumen size. Each stent design (see below) was then deployed at low pressure (<=6 atm) with a delivery balloon less than or equal to reference vessel size. Each stent was then assigned randomly to 1 of 3 postdilatation strategies: appropriate-sized compliant balloon (Cobra-18, Boston Scientific Corp) with nominal balloon-to-artery ratio {approx}1.0; oversized compliant balloon (Cobra-18, Boston Scientific Corp) with nominal balloon-to-artery ratio {approx}1.2; or oversized noncompliant balloon (NC Cobra-18, Boston Scientific Corp) with nominal balloon-to-artery ratio {approx}1.2.

The postdilating balloon was delivered into the stent over the angioplasty guidewire. The guidewire was then removed and its lumen flushed with heparinized saline. An 0.018-in 30-MHz rotating imaging core (Boston Scientific) was inserted into the guidewire lumen of the deflated balloon and positioned just distal to the balloon midmarker. The imaging wire was connected to the rotating motor drive unit, and all images were processed by use of the CVIS ClearView System. The balloon and the inflation device were filled with heparinized normal saline rather than iodinated contrast medium to avoid changes in density, and thus the speed of sound, in the intervening liquid medium during inflation. Continuous ultrasound imaging was performed as the balloon was gradually inflated in 2-atm steps from 2 to 12 atm. Pressure (and thus balloon volume) was allowed to equilibrate for at least 10 seconds at each pressure level. Ultrasound images were also acquired during slow, progressive balloon deflation. In a subset of stents (n=6), conventional IVUS imaging was performed with the CVIS catheter after balloon deflation and core wire imaging to assess correlation between the 2 imaging modalities.

Stents
Four different stent designs were evaluated: the coil Gianturco-Roubin II stent (Cook Cardiology), the slotted-tube Palmaz-Schatz coronary stent (Cordis Corp), the MultiLink stent (Guidant Corp), and the NIR stent (Medinol). To ensure unbiased distribution of each stent design in a particular location, each stent was implanted in each animal according to a matrix that randomly assigned the stent to vessel location and position (proximal, mid, or distal vessel).

Data Analysis and Statistics
All images were recorded on S-VHS videotape. CSA, minimum lumen diameter (MLD), and maximum lumen diameters were quantified with online, integrated measurement software. Elastic recoil was calculated in the following manner: % area recoil=1-(stent CSAdeflation/balloon CSA12 atm); % diameter recoil=1-(stent MLDdeflation/balloon MLD12 atm).

Data are expressed as mean±SEM. Comparisons were made by ANOVA and Student's t test with correction for multiple comparisons by the Bonferroni method. Correlation between continuous variables was assessed by linear regression. Data were analyzed by JMP software (SAS Institute).


*    Results
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
*Results
down arrowDiscussion
down arrowReferences
 
Balloon Expansion
A total of 32 stents (8 of each design) were implanted in 8 animals. Pressure-area relations for 3.5-mm compliant, 4.0-mm compliant, and 4.0-mm noncompliant balloons are shown in the TableDown. As predicted by the manufacturers' specifications, all 3 balloon groups demonstrated increases in CSA as a function of pressure. Pressure-area curves for compliant balloons that matched the reference vessel size and oversized noncompliant balloons closely approximated the relationships expected on the basis of these specifications (Figure 1Down). Compliant balloons that were oversized compared with the reference vessel (balloon-to-artery ratio {approx}1.2), however, had measured balloon areas that were smaller at high pressure than specified in the manufacturers' compliance charts.


View this table:
[in this window]
[in a new window]
 
Table 1. Pressure-Area Relations for Different Balloon Types Inflated Within Stented Pig Coronary Arteries



View larger version (16K):
[in this window]
[in a new window]
 
Figure 1. Measured pressure-area relations (solid lines) vs that expected from manufacturers' specifications (dotted lines) for 3 different balloons: 3.5-mm compliant balloons (mean balloon-to-artery ratio 1.05), 4.0-mm compliant balloons (mean balloon-to-artery ratio 1.15), and 4.0-mm noncompliant balloons (mean balloon-to-artery ratio 1.22). Note smaller-than-expected growth of oversized compliant balloon ({blacktriangleup}) vs manufacturers' specifications ({triangleup}) at high pressure.

Stent Recoil
In the subset of stents in which conventional IVUS imaging was performed after balloon deflation and imaging wire assessment, there was excellent correlation of stent CSA as measured by IVUS and the imaging core (Figure 2Down). The magnitude of acute stent recoil (from maximum expansion to postdeflation) ranged from 15% to 30% CSA and differed among the 4 stent types (Figure 3Down). Area recoil was significantly greater for the coil design (Gianturco-Roubin II stent) than for any of the slotted-tube configurations (Figure 4Down). The corresponding diameter recoil (17.8% versus 10.4%, P<0.01) was also significantly greater for the Gianturco-Roubin stent than for the reference Palmaz-Schatz coronary stent. As a result of these differences in stent recoil, the ratio of final stent to reference vessel area was significantly smaller for the coil than for the slotted-tube stents (0.94±0.06 versus 1.11±0.04; P=0.022) despite similar ratios of maximum measured balloon size to reference vessel CSA across all 4 stent designs (Figure 5Down). This greater recoil of the coil stent design is reflected graphically by a downward shift in the final stent–maximum balloon expansion curve when normalized by vessel size (Figure 6Down). To overcome this recoil effect in a coil design would require substantially greater overexpansion (to a balloon-to-artery area ratio of 1.4 by area [diameter ratio 1.2]) than in a slotted-tube design (area ratio 1.2 [diameter ratio 1.1]).



View larger version (13K):
[in this window]
[in a new window]
 
Figure 2. Correlation of stent CSA by IVUS (x axis) and imaging wire (y axis). Regression equation: y=1.028x+0.445.



View larger version (21K):
[in this window]
[in a new window]
 
Figure 3. Top, Acute area and bottom, diameter recoil for 4 different stents. GR II indicates Gianturco-Roubin II.



View larger version (70K):
[in this window]
[in a new window]
 
Figure 4. IVUS imaging with 30-MHz transducer positioned in guidewire lumen of balloon at 12 atm (top) and immediately after balloon deflations (bottom). Images were obtained during expansion of Palmaz-Schatz stent (left) and for Gianturco-Roubin II (GR II) (right). NC indicates noncompliant.



View larger version (16K):
[in this window]
[in a new window]
 
Figure 5. Stent expansion and recoil maximum balloon-to-artery CSA represented as top of stacked bar graph. Final stent-to-artery CSA ratio represented by top of each black bar. Difference between tops of white and black bars thus represents acute recoil as a percentage of balloon/artery CSA.



View larger version (14K):
[in this window]
[in a new window]
 
Figure 6. Regression lines relating balloon expansion and final stent-to-artery ratios for slotted-tube (•; solid line) and coil ({square}; dashed line) stents. y=0.73(slotted tube)+0.14; r2=0.69, P<0.001. y=0.83(coil)–0.17; r2=0.74, P=0.003. Coil stents require 41% area overexpansion (17% diameter overexpansion), whereas slotted-tube stents require only 17% area overexpansion (9% diameter overexpansion) to achieve a final CSA equivalent to reference CSA.


*    Discussion
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
up arrowResults
*Discussion
down arrowReferences
 
During the past several years, stenting has emerged as the predominant modality for percutaneous coronary revascularization, and it is now used in >50% of all interventions. The favorable short- and long-term outcomes after stenting relate to the ability of these metallic endovascular prostheses to seal dissection planes and resist elastic recoil, thereby achieving a large final lumen. Numerous studies have documented that more complete stent expansion is associated with reductions in subacute thrombosis and late restenosis.1 2 3 4 5 6 Although there has been a lack of consensus regarding what constitutes optimal stent expansion, angiographic and IVUS definitions usually stipulate that a ratio of stent to reference lumen CSA of 0.9 to 1.0 should be achieved.9 10 11

Nakamura et al7 demonstrated that despite excellent angiographic appearance, the majority of stents deployed at low pressure remain suboptimally expanded by IVUS criteria. Use of high-pressure postdilatation increased stent CSA and improved stent geometry, so that routine high-pressure postdilatation is now a universally accepted practice. Even so, final stent CSA rarely exceeds 60% to 80% of the nominal balloon CSA. Potential explanations for this persistent stent undersizing include balloon underexpansion (despite high inflation pressures), acute stent recoil, or a combination of the 2 mechanisms.

Haude et al12 reported that compared with conventional balloon dilatation, the slotted-tube Palmaz-Schatz coronary stent reduced diameter (3.5%) and area (5.1%) recoil. These findings have been interpreted to mean that stenting largely eliminates acute elastic recoil after coronary intervention. In that study, however, high-pressure dilatation was not performed uniformly, and recoil was calculated by use of geometric assumptions derived from angiographic variables. Lumen CSA was thus calculated from a single, measured diameter assuming a uniform, elliptical geometry. In addition, expansion was assumed to be uniform along the entire length of the balloon. Because the maximum balloon diameter was measured and equivalence was assumed in the diseased noncompliant segment of the artery, balloon size may have been significantly overestimated. More recently, Bermejo and colleagues13 used a combination of quantitative angiography and IVUS to assess the relative contribution of balloon underexpansion and stent recoil to the residual lumen stenosis after high-pressure dilatation. They measured minimum and maximum balloon diameters by quantitative coronary angiography and demonstrated a failure to reach nominal diameter as well as nonuniformity of balloon expansion within the diseased segments. By then performing IVUS in the same stented segments after high-pressure dilatation, they calculated area recoil of 15% for the Palmaz-Schatz stent and 26% for the coil Wiktor stent relative to the inflated balloon area.

In the present study, online IVUS was performed with an 0.018-in imaging core positioned in the guidewire lumen of the balloon to allow accurate assessment of both the balloon expansion during progressively higher-pressure inflation and acute stent recoil during balloon deflation. Balloon diameters and CSAs were thus measured directly, rather than being calculated from geometric assumptions and disparate imaging methodologies. In these normal (nondiseased) porcine arteries, balloon CSA closely matched that expected from the manufacturers' compliance charts. The only exception to this finding was for oversized (balloon-to-artery ratio >1.15) compliant balloons, for which some balloon underexpansion relative to specifications was seen at the highest pressures (12 atm).

The surprising finding, however, was the magnitude of stent recoil during balloon deflations. Acute stent recoil ranged from 15% to 17% area for the slotted-tube stents and 30% area for the coil stents. This corresponds to diameter recoil of 6.9% to 10.4% for slotted-tube stents, versus 17.8% for the coil stent. This 2-fold greater elastic recoil observed for the Gianturco-Roubin coil was similar to what Bermejo et al13 and Werner et al14 observed for the Wiktor coil stent compared with the Palmaz-Schatz slotted-tube design. These recoil data may explain the greater residual stenosis observed clinically after coil stent compared with slotted-tube placement.15 Despite similar maximum balloon-to-artery ratios with all 4 stent designs, the greater magnitude of recoil observed for the Gianturco-Roubin stent led to a significantly lower final ratio of stent to reference artery CSA than in the slotted-tube designs (0.9 versus 1.1, P=0.02). To achieve a final stent-to-artery ratio of 1.0, the coil stents would require overexpansion to a balloon-to-artery area ratio of 1.4 (diameter ratio 1.2), whereas slotted-tube stents would require overexpansion only to an area ratio of 1.2 (diameter ratio 1.1).

With the exception of oversized compliant balloons, balloon underexpansion was not observed in this study. This is in contradistinction to the findings of Bermejo et al, which documented up to a 33% balloon underexpansion in diseased human coronary arteries. The discrepancy between these studies may be due to different methodologies (quantitative coronary angiography versus IVUS balloon sizing) or because balloon underexpansion is more likely in diseased vessels, which are less compliant and often contain bulky fibrocalcific plaque (which physically limits balloon expansion), than in normal arteries as evaluated in the present study.16 17 It must be noted, however, that the magnitude of elastic stent recoil (20% to 30% of stent area) seen in this study is comparable to that seen by Bermejo et al and would alone be sufficient to explain the degree of final stent undersizing seen clinically. To compensate for this elastic stent recoil and achieve the optimal stent result (final stent-to-artery ratio of 1.0), stents must then be overexpanded by 10% to 20% diameter relative to the reference vessel diameter. This may be done by use of either an oversized noncompliant balloon or a correctly sized compliant balloon relative to the reference lumen that is inflated to high pressure. As long as the desired degree of balloon oversizing is achieved, however, inflation to high pressure is not per se a requirement for optimal stent results. This principle may explain the recent trend toward lower ({approx}12 atm) inflation pressures with balloons slightly oversized relative to the desired reference lumen diameter (including balloons sized to approach the diameter of the external elastic lamina measured by IVUS). In heavily diseased arteries, however, this calculation may increase the risk of edge dissection or frank vessel rupture, as was observed in the early experience of Colombo et al.9

Limitations
This study has several important limitations. First, all stents were expanded in nondiseased, compliant arteries. The forces resisting balloon expansion are clearly less in this model than in atherosclerotic vessels. The contribution of balloon undersizing to final stent underexpansion may thus have been underestimated significantly compared with that which occurs in diseased human arteries. Second, the degree of stent recoil may be even higher in diseased arteries, which may exert greater recoil force. Even so, the 15% area recoil observed for slotted-tube stents and 30% area recoil observed for coil stents in this study represent a "best-case scenario" for these devices and mandate intentional stent overexpansion to achieve the desired final stent luminal CSA.


*    Acknowledgments
 
This study was supported in part by an American Heart Association Clinician-Scientist Award to Dr Cohen.


*    Footnotes
 
S.E. Hosley is an employee of Boston Scientific, which distributes one of the study stents.

Received December 7, 1998; revision received April 14, 1999; accepted April 15, 1999.


*    References
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
up arrowResults
up arrowDiscussion
*References
 

  1. Fischman DL, Leon MB, Baim DS, Schatz RA, Penn I, Detre K, Savage MP, Veltri L, Ricci D, Nobuyoshi M, Cleman M, Heuser R, Almond D, Teirstein P, Fish D, Colombo A, Brinker J, Moses J, Hirshfeld J, Bailey S, Ellis S, Rake R, Goldberg S. A randomized comparison of coronary stent placement and balloon angioplasty in the treatment of coronary artery disease. N Engl J Med. 1994;331:496–501.[Abstract/Free Full Text]
  2. Serruys PW, de Jaegere P, Kiemeneij F, Macaya C, Rutsch W, Heyndrickx G, Emanuelsson H, Marco J, Legrand V, Materne P, Belardi J, Sigwart U, Colombo A, Goy J-J, van den Heuvel P, Delcan J, Morel MA, for the Benestent Study Group. A comparison of balloon-expandable stent implantation with balloon angioplasty in patients with coronary disease. N Engl J Med. 1994;331:489–495.[Abstract/Free Full Text]
  3. Carrozza JP Jr, Schatz RA, George CJ, Buchbinder M, Cleman MW, Curry RC, Goldberg S, Hirshfeld JW, Ivanhoe RJ, King SB III, Leon MB, Popma JJ, Ricci D, Safian RD, Baim DS, for the NACI Investigators. Acute and long-term outcome following Palmaz-Schatz stenting: analysis from the New Approaches to Coronary Intervention (NACI) registry. Am J Cardiol. 1997;80:78K–88K.[Medline] [Order article via Infotrieve]
  4. Kuntz RE, Safian RD, Carrozza JP, Fischman DL, Mansour M, Baim DS. The importance of acute lumen diameter in determining restenosis after coronary atherectomy or stenting. Circulation. 1992;86:1827–1835.[Abstract/Free Full Text]
  5. Dussaillant GR, Mintz GS, Pichard AD, Kent KM, Satler LF, Popma JJ, Wong SC, Leon MB. Small stent size and intimal hyperplasia contribute to restenosis: a volumetric intravascular ultrasound analysis. J Am Coll Cardiol. 1995;26:720–724.[Abstract]
  6. Moussa I, DiMario C, Reimers B, Akiyama T, Tobis J, Colombo A. Subacute stent thrombosis in the era of intravascular ultrasound-guided coronary stenting without anticoagulation: frequency, predictors, and clinical outcome. J Am Coll Cardiol. 1997;29:6–12.[Abstract]
  7. Nakamura S, Colombo A, Gaglione S, Almagor Y, Goldberg SL, Maiello L, Finci L, Tobis JM. Intracoronary ultrasound observations during stent implantation. Circulation. 1994;89:2026–2034.[Free Full Text]
  8. Carrozza JP Jr, Hermiller JB Jr, Linnemeier TJ, Popma JJ, Yock PG, Roubin GS, Dean LS, Kuntz RE, Robertson L, Ho KKL, Cutlip DE, Baim DS. Quantitative coronary angiographic and intravascular ultrasound assessment of a new non-articulated stent: report from the ACS MULTI-LINK Stent pilot study. J Am Coll Cardiol. 1998;31:50–56.[Abstract/Free Full Text]
  9. Colombo A, Hall P, Nakamura S, Almagor Y, Maiello L, Martini G, Gaglione A, Goldberg SL, Tobis JM. Intracoronary stenting without anticoagulation accomplished with intravascular ultrasound guidance. Circulation. 1995;91:1676–1688.[Abstract/Free Full Text]
  10. Fitzgerald PJ, Hayase M, Mintz GS, Kuntz RE, Moses JW, Diver DJ, Deutsch E, Simonton CA, Baim DS, Leon MB, Yock PG. CRUISE: can routine intravascular ultrasound influence stent expansion? Analysis of outcomes. J Am Coll Cardiol. 1998;31:3962. Abstract.
  11. de Jaegere P, Mudra H, Figulla H, Almagor Y, Doucet S, Penn I, Colombo A, Hamm C, Bartorelli A, Rothman M, Nobuyoshi M, Yamaguchi T, Voudris V, DiMario C, Makovski S, Hausmann D, Rowe S, Rabinovich S, Sunamura M, van Es GA. Intravascular ultrasound-guided optimized stent deployment: immediate and 6 months clinical and angiographic results from the Multicenter Ultrasound Stenting in Coronaries Study (MUSIC Study). Eur Heart J. 1998;19:1214–1223.[Abstract/Free Full Text]
  12. Haude M, Erbel R, Issa H, Meyer J. Quantitative analysis of elastic recoil after balloon angioplasty and after intracoronary implantation of balloon-expandable Palmaz-Schatz stents. J Am Coll Cardiol. 1993;21:26–34.[Abstract]
  13. Bermejo J, Botas J, García E, Elízaga J, Osende J, Soriano J, Abeytua M, Delcán JL. Mechanisms of residual lumen stenosis after high-pressure stent implantation: a quantitative coronary angiography and intravascular ultrasound study. Circulation. 1998;98:112–118.[Abstract/Free Full Text]
  14. Werner GS, Schünemann S, Ferrari M, Figulla HR, Kreuzer H. Comparison of slotted-tube and coil stents after high-pressure stent deployment by intravascular ultrasound. J Am Coll Cardiol. 1997;21:275. Abstract.
  15. Leon MB, Popma JJ, O'Shaughnessy C, Dean LS, Lansky AJ, Fry ETA, Curran MJ. Quantitative angiographic outcomes after Gianturco-Roubin II stent implantation in complex lesion subsets. Circulation. 1997;96(suppl I):I-653. Abstract.
  16. Nakatani S, Yamagishi M, Tamai J, Goto Y, Umeno T, Kawaguchi A, Yutani C, Miyatake K. Assessment of coronary artery distensibility by intravascular ultrasound: application of simultaneous measurement of luminal area and pressure. Circulation. 1995;91:2904–2910.[Abstract/Free Full Text]
  17. Albrecht D, Kaspers S, Fussl R, Hopp HW, Sechtem U. Coronary plaque morphology affects stent deployment: assessment by intracoronary ultrasound. Cathet Cardiovasc Diagn. 1996;38:229–235.[Medline] [Order article via Infotrieve]



This article has been cited by other articles:


Home page
HeartHome page
E. De Benedetti and P. Urban
Coronary stenting: why size matters
Heart, December 1, 2007; 93(12): 1500 - 1501.
[Abstract] [Full Text] [PDF]


Home page
HeartHome page
R. J Russo, P. D Silva, and M. Yeager
Coronary artery overexpansion increases neointimal hyperplasia after stent placement in a porcine model
Heart, December 1, 2007; 93(12): 1609 - 1615.
[Abstract] [Full Text] [PDF]


Home page
HeartHome page
J Gunn, N Arnold, K H Chan, L Shepherd, D C Cumberland, and D C Crossman
Coronary artery stretch versus deep injury in the development of in-stent neointima
Heart, October 1, 2002; 88(4): 401 - 405.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow Request Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Carrozza, J. P.
Right arrow Articles by Baim, D. S.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Carrozza, J. P., Jr
Right arrow Articles by Baim, D. S.
Related Collections
Right arrow Catheter-based coronary interventions: stents
Right arrow Coronary imaging: angiography/ultrasound/Doppler/CC